K Number
K203776

Validate with FDA (Live)

Date Cleared
2021-03-25

(87 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Surgical Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Surgical Masks are manufactured with three/four layers. The surgical mask has two models which are SNN200640 and MN12. They are basically the same, the only difference is the SNN200640 has the three layers and MN112 has the fours layers. The Surgical Masks are single use, disposable device, provided non-sterile.

AI/ML Overview

The document you provided is a 510(k) Premarket Notification for surgical masks, which focuses on demonstrating substantial equivalence to a predicate device. It includes non-clinical testing results against specific standards, but it does not describe a study in the context of evaluating an AI/ML-driven device's performance against clinical acceptance criteria. The questions you've asked are typically relevant for AI/ML medical devices.

Since the provided text describes a medical device (surgical masks) that is evaluated based on physical and filtration properties, and not an AI/ML device, many of the requested categories (like number of experts, ground truth, AI assistance effect size, training set) are not applicable.

However, I can extract the acceptance criteria and performance data for the surgical masks from the "Non-Clinical Testing" section.

Here's the information derived from the document, acknowledging that it's for a traditional medical device, not an AI/ML system:

1. A table of acceptance criteria and the reported device performance

StandardPurposeAcceptance Criteria (Level 1)Acceptance Criteria (Level 3)Reported Performance
ASTM F1862M-17Fluid Resistance Performance29 out of 32 pass at 80mmHg29 out of 32 pass at 160mmHgPass
ASTM F2299Particulate Filtration Efficiency≥95%≥98%Pass
ASTM F2101-19Bacterial Filtration Efficiency≥95%≥98%Pass
EN 14683:2019 Annex CDifferential Pressure<5.0mmH2O/cm²<6.0mmH2O/cm²Pass
16 CFR 1610FlammabilityClass I non flammableN/APass

2. Sample size used for the test set and the data provenance

  • Sample Size:
    • For Fluid Resistance (ASTM F1862M-17), the acceptance criterion mentions "29 out of 32 pass", implying a sample size of 32 items tested for this specific criterion.
    • For other tests (Particulate Filtration, Bacterial Filtration, Differential Pressure, Flammability), the document states "Pass" without specifying the exact number of units tested.
  • Data Provenance: The document does not specify the country of origin of the data or whether the tests were retrospective or prospective. These are standard non-clinical tests performed in a lab setting rather than clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable as the device is a surgical mask, and its performance is evaluated against standardized test methods for physical properties (like filtration efficiency and fluid resistance), not against expert interpretations of clinical images or data. Ground truth in this context is established by the methods outlined in standards such as ASTM F1862M-17, ASTM F2299, ASTM F2101-19, EN 14683:2019, and 16 CFR 1610.

4. Adjudication method for the test set

This question is not applicable. The performance is measured directly by scientific instruments and defined protocols within the specified ASTM and EN standards, not through human adjudication or consensus.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable as the device is a surgical mask and not an AI-assisted diagnostic or interpretive system. No human reader studies with or without AI assistance were conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable as the device is a surgical mask and does not involve any algorithm or AI component performing in a standalone capacity.

7. The type of ground truth used

The "ground truth" for the performance evaluation of these surgical masks is established by physical and material testing standards (e.g., filtration rates, fluid resistance under specified pressures, flammability class) as defined by organizations like ASTM and EN.

8. The sample size for the training set

This question is not applicable. The document describes non-clinical performance testing for a physical product, not the development of an AI/ML model. Therefore, there is no "training set."

9. How the ground truth for the training set was established

This question is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "U.S. FOOD & DRUG" on top of the word "ADMINISTRATION".

March 25, 2021

SQ Medical Supplies Inc. % Shelly Li Official Correspondent Landlink Healthcare Technology (Shanghai) Co., Ltd. Room 703, 705, Building 1, West Guangzhong Road 555 Shanghai. 200072 China

Re: K203776

Trade/Device Name: Surgical Masks, Model SNN200640 & MN112 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FXX Dated: December 23, 2020 Received: December 28, 2020

Dear Shelly Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203776

Device Name Surgical Masks

Indications for Use (Describe)

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K203776

l. Submitter

SQ Medical Supplies Inc. 8889 West Olympic Blvd. Suite 1000 Beverly Hills, CA 90211. Establishment Registration Number: 3016757871

Contact person: Jonathan Lim Position: Chief Executive Officer Tel.: +1-888-912-8168 E-mail: jonlim@sqmedicalsupplies.com

Preparation date: Feb. 26, 2021

II. Proposed Device

Trade Name of Device:Surgical Masks
Common name:Surgical Mask
Regulation Number:21 CFR 878.4040
Regulatory Class:Class II
Product code:FXX
Review PanelGeneral Hospital

III. Predicate Devices

510(k) Number:K160269
Trade name:Surgical Face masks (Ear loops and Tie-on)
Common name:Surgical Mask
Classification:Class II
Product Code:FXX
ManufacturerSan-M Package Co., Ltd.

IV. Device Description

The Surgical Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Surgical Masks are manufactured with three/four layers. The surgical mask has two models which are SNN200640 and MN12. They are basically the same, the only difference is the SNN200640 has the three layers and MN112 has the fours layers.

{4}------------------------------------------------

The Surgical Masks are single use, disposable device, provided non-sterile.

V. Indication for Use

The Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

ItemProposed Device(K203776)Predicate Device(K160269)Decision
Trade NameSurgical MasksSurgical Face masks (Ear - loops and Tie-on)
Product CodeFXXFXXSame
Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassClass IIClass IISame
Mask StyleFlat-pleated, ear loops, 3/4 layersFlat-pleated, ear loops or tie-on, 4 layersSimilar
Indication for UseThe Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.The Surgical facemasks are intended to be worn to protect the patient and healthcare personnel from transfer of microorganisms, blood fluid, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluid. This is a single-use, disposable devices provided non-sterile.Same
MaterialInner layer: White double spun-bond polypropylenePolypropylene
Middle layer: Layer #1: PolypropyleneLayer #2: Melt blownLayer #1. Polypropylene spun-bondSame

VI. Comparison of Technological Characteristics with the Predicate Devices

{5}------------------------------------------------

polypropylene filter(Middle Layer #1 is not applicableSNN200640)Layer #2: Polypropylene meltblown
OuterlayerBlue double spun-bondpolypropylenePolypropyleneSame
Ear loopsSpandex+NylonEar loops: Polyester,polyurethane;Side tapes: Polyesterspun-bond (ear loops maskonly)Ties tapes: Polypropylenespun-bond or polyesterspun-bondDifferent
NosepieceSteel wire coated byPolyethyleneSteel wire coated byPolyethyleneSame
ColorBlueBlue ,whiteSimilar
Dimension17.5cmx9.5cm17.5cmx9.0cm18.0cmx9.0cmSame
OTC UseYesYesSame
SterilityNon-sterileNon-sterileSame
For single UseYesYesSame
ASTM F2100LevelLevel 1: Pass at 80mmHgLevel 3: Pass at 160mmHgLevel 1: Pass at 80mmHgLevel 2: Pass at 120mmHgLevel 3: Pass at 160mmHgSimilar
BiocompatibilityConfirm to the requirements ofISO 10993 series standardsConfirm to the requirements ofISO 10993 series standardsSame

VII. Non-Clinical Testing

Non clinical performance tests were conducted to verify that the proposed device met all design specifications. The below table shows the test results of test article, which demonstrated that the proposed device complies with the standards of ASTM F2100-19:

StandardPurposeAcceptance CriteriaResults
Level 1Level 3
ASTMF1862M-17FluidResistancePerformance29 out of 32 passat 80mmHg29 out of 32 passat 160mmHgPass
ASTM F2299Particulate≥95%≥98%Pass

{6}------------------------------------------------

FiltrationEfficiency
ASTMF2101-19BacterialFiltrationEfficiency≥95%≥98%Pass
EN14683:2019Annex CDifferentialPressure<5.0mmH2O/cm²<6.0mmH2O/cm²Pass
16 CFR 1610FlammabilityClass I non flammablePass

VIII. Clinical Testing

No clinical study is included in this submission.

IX. Conclusion

The conclusion drawn from the non-clinical performance testing data demonstrates that the subject device is as safe, as effective, and performs as well as or better than the predicate device, Surgical Face masks (K160269).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.