K Number
K203748

Validate with FDA (Live)

Device Name
SeptiCyte RAPID
Manufacturer
Date Cleared
2021-11-29

(341 days)

Product Code
Regulation Number
866.3215
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SeptiCyte ® RAPID test is a gene expression assay using reverse transcription polymerase chain reaction to quantify the relative expression levels of host response genes isolated from whole blood collected in the PAXgene ® Blood RNA Tube. The SeptiCyte ® RAPID test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infectionnegative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The SeptiCyte ® RAPID test generates a score (SeptiScore®) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infectionpositive systemic inflammation. SeptiCyte ® RAPID is intended for in-vitro diagnostic use on the Biocartis IdyllaTM System.

Device Description

SeptiCyte RAPID is an in vitro diagnostic test for simultaneous amplification and detection of two RNA transcripts (PLA2G7 and PLAC8) using total RNA extracted from human blood. The test has been designed, manufactured, and validated for use on the Biocartis Idylla real-time PCR system. The SeptiCyte RAPID test is performed with an Idylla Cartridge, a single-use, disposable, multi-chambered fluidic cartridge that runs on the Biocartis Idylla System. In an automated fashion, all reaction steps take place within the cartridge, including sample extraction/purification, RT-qPCR for the detection and relative quantification of the two human mRNA targets PLAC8. PLA2G7. Test results (measured Cg values and calculated SeptiScore) are available in about 65 minutes.

The specimen used for the SeptiCyte RAPID is a sample of whole blood collected in a PAXgene blood RNA tube (FDA 510k number K042613). The cartridge contains all of the necessary reagents to perform RNA isolation from the sample.

SeptiCyte RAPID uses quantitative, real-time determination of the amount of each transcript in the sample based on the detection of fluorescence by the Biocartis Idylla qPCR instrument function. The cartridge includes the reagents for reverse transcription and PCR. Transcripts PLAC8 and PLA2G7 are amplified and quantified. These values are combined to produce the SeptiScore, which is interpreted and categorized into four discrete bands, which are associated with a sequentially higher likelihood of sepsis.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Immunexpress, Inc. SeptiCyte RAPID device, based on the provided FDA 510(k) summary:

1. Acceptance Criteria and Reported Device Performance

The clinical studies for SeptiCyte RAPID had the following acceptance criteria for the primary endpoint:

  1. A monotonic increase in the probability of sepsis across the four SeptiScore interpretation bands.
  2. Non-overlapping 80% confidence intervals for the probability of sepsis between non-adjacent SeptiScore interpretation bands (i.e., between bands 1 and 3; and between bands 2 and 4).

Here's the performance reported using the Forced RPD Method (as it covers the full dataset and shows the trend across all bands):

SeptiScore Interpretation BandAcceptance Criteria 1: Monotonic Increase (Observed Probability of Sepsis)Acceptance Criteria 2: Non-overlapping 80% Confidence Intervals (Observed Probability)
Band 1 (0 - 4.9)0.12(0.08 - 0.18)
Band 2 (5.0 - 6.1)0.24(0.18 - 0.3)
Band 3 (6.2 - 7.3)0.48(0.4 - 0.56)
Band 4 (7.4 - 15.0)0.80(0.74 - 0.84)

Proof of Acceptance:
The reported device performance demonstrates:

  1. Monotonic Increase: The observed probabilities of sepsis (0.12, 0.24, 0.48, 0.80) clearly increase monotonically across the four bands, meeting the first criterion.
  2. Non-overlapping 80% Confidence Intervals:
    • Band 1 (0.08 - 0.18) vs. Band 3 (0.4 - 0.56): The intervals do not overlap.
    • Band 2 (0.18 - 0.3) vs. Band 4 (0.74 - 0.84): The intervals do not overlap.
      This meets the second criterion.

The analysis using Consensus and Unanimous RPD methods also showed similar results, fulfilling the acceptance criteria.

2. Sample Size and Data Provenance for the Test Set

  • Sample Size: N = 386
    • N = 356 from retrospective studies
    • N = 30 from a prospective study
  • Data Provenance:
    • Retrospective Study: Netherlands-based Molecular diagnosis And Risk Stratification of sepsis (MARS) clinical trial (NCT01905033)
    • Retrospective Study: USA-based Validation of septic gene Expression Using SeptiCyte (VENUS) clinical trial (NCT02127502)
    • Prospective Study: USA-based NEar PatienT MolecUlar TestiNg in Sepsis (NEPTUNE) clinical trial

3. Number of Experts and Qualifications for Ground Truth Establishment

  • Number of Experts: Three-member expert panel.
  • Qualifications of Experts: Not explicitly stated beyond "external three-member expert panel not involved in the care of the patients." It does not specify their medical specialty (e.g., critical care physicians, infectious disease specialists) or years of experience.

4. Adjudication Method for the Test Set

Three different methods of Retrospective Physician Diagnosis (RPD) were used:

  • Consensus: Two or three RPD panelists agree that the patient is either sepsis or SIRS. Indeterminates are excluded.
  • Unanimous: All three RPD panelists, as well as the consensus discharge evaluation of the site investigators, agree that the patient is either sepsis or SIRS.
  • Forced: Applied to patients initially called "indeterminate" by RPD panelists. The panelists are then forced to make a consensus or unanimous call of sepsis or SIRS.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. The device is an in vitro diagnostic (IVD) assay that provides a score to aid in diagnosis, not an imaging AI or a system designed to assist human readers in interpretation. Therefore, the question of human readers improving with AI vs. without AI assistance is not applicable in this context.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire clinical evaluation is based on the SeptiCyte RAPID test (algorithm only) generating a SeptiScore and its associated interpretation bands. The clinical studies aim to validate the device's ability to differentiate infection-positive (sepsis) from infection-negative systemic inflammation based solely on its output.

7. Type of Ground Truth Used

The ground truth used was expert consensus or retrospective physician diagnosis (RPD). This involved an external three-member expert panel reviewing clinical cases to determine whether a subject's clinical status was SIRS, sepsis, or indeterminate.

8. Sample Size for the Training Set

The document does not explicitly state a separate sample size for a training set. The clinical studies described (MARS, VENUS, NEPTUNE) were primarily for validation and 510(k) clearance of the already developed SeptiCyte RAPID test. It's common for gene expression assays to be developed using earlier research cohorts, but this document focuses on the validation of the specific device. The clinical studies might have been used for final validation and refinement of the SeptiScore cut-offs, but a distinct "training set" for the algorithm itself is not mentioned here.

9. How the Ground Truth for the Training Set Was Established

As a distinct training set is not explicitly detailed, the method for establishing its ground truth is also not described. However, given that the clinical validation ground truth was established by expert consensus/RPD, it is highly probable that any development or training earlier in the product lifecycle would have used similar methodologies.

{0}------------------------------------------------

November 29, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.

Immunexpress, Inc Lakesha Hunt-Dickens Director, Quality and Regulatory Affairs 425 Pontius Avenue North, Suite 470 Seattle, Washington 98109

Re: K203748

Trade/Device Name: SeptiCyte RAPID Regulation Number: 21 CFR 866.3215 Regulation Name: Rt-Qpcr Assay For Mrna Transcript Immune Biomarkers Regulatory Class: Class II Product Code: PRE Dated: December 21, 2020 Received: December 23, 2020

Dear Lakesha Hunt-Dickens:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Maria Ines Garcia, Ph.D. Assistant Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203748

Device Name SeptiCyte® RAPID

Indications for Use (Describe)

The SeptiCyte ® RAPID test is a gene expression assay using reverse transcription polymerase chain reaction to quantify the relative expression levels of host response genes isolated from whole blood collected in the PAXgene ® Blood RNA Tube. The SeptiCyte ® RAPID test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infectionnegative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The SeptiCyte ® RAPID test generates a score (SeptiScore®) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infectionpositive systemic inflammation. SeptiCyte ® RAPID is intended for in-vitro diagnostic use on the Biocartis IdyllaTM System.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

7. 510(k) Summary

A. 510(k) Number:

K203748

  • B. Purpose for Submission:
    To obtain a substantial equivalence determination for SeptiCyte® RAPID

  • C. Measurand:
    Two mRNA transcript immune biomarkers: PLA2G7, PLAC8

D. Type of Test:

Reverse transcription + quantitative PCR (RT-qPCR)

E. Applicant:

Immunexpress, Inc.

  • F. Proprietary and Established Names:
    SeptiCyte® RAPID

G. Regulatory Information:

    1. Regulation section:
      21 CFR 866.3215; Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis
    1. Classification:
      Class II (Special Controls)
    1. Product code: PRE
    1. Panel:

83: Microbiology

H. Intended Use:

    1. Intended use(s):
      The SeptiCyte® RAPID test is a gene expression assay using reverse transcription polymerase chain reaction to quantify the relative expression levels of host response genes isolated from whole blood collected in the PAXgene® Blood RNA Tube. The SeptiCyte® RAPID test is used in conjunction with clinical assessments and other laboratory findings as an aid to differentiate infection-positive (sepsis) from infectionnegative systemic inflammation in patients suspected of sepsis on their first day of ICU admission. The SeptiCyte® RAPID test generates a score (SeptiScore®) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infectionpositive systemic inflammation. SeptiCyte® RAPID is intended for in-vitro diagnostic use on the Biocartis Idylla™ System.

{4}------------------------------------------------

    1. Indication(s) for use:
      Same as Intended Use.
    1. Special conditions for use statement(s):
      Whole blood is collected in a PAXgene® blood RNA tube (K042613)
    1. Special instrument requirements:
      SeptiCyte RAPID is designed to run on the Biocartis Idylla system.

Device Description: -

SeptiCyte RAPID is an in vitro diagnostic test for simultaneous amplification and detection of two RNA transcripts (PLA2G7 and PLAC8) using total RNA extracted from human blood. The test has been designed, manufactured, and validated for use on the Biocartis Idylla real-time PCR system. The SeptiCyte RAPID test is performed with an Idylla Cartridge, a single-use, disposable, multi-chambered fluidic cartridge that runs on the Biocartis Idylla System. In an automated fashion, all reaction steps take place within the cartridge, including sample extraction/purification, RT-qPCR for the detection and relative quantification of the two human mRNA targets PLAC8. PLA2G7. Test results (measured Cg values and calculated SeptiScore) are available in about 65 minutes.

The specimen used for the SeptiCyte RAPID is a sample of whole blood collected in a PAXgene blood RNA tube (FDA 510k number K042613). The cartridge contains all of the necessary reagents to perform RNA isolation from the sample.

SeptiCyte RAPID uses quantitative, real-time determination of the amount of each transcript in the sample based on the detection of fluorescence by the Biocartis Idylla qPCR instrument function. The cartridge includes the reagents for reverse transcription and PCR. Transcripts PLAC8 and PLA2G7 are amplified and quantified. These values are combined to produce the SeptiScore, which is interpreted and categorized into four discrete bands, which are associated with a sequentially higher likelihood of sepsis.

J. Substantial Equivalence Information:

    1. Predicate device name(s): SeptiCyte LAB
    1. Predicate 510(k) number(s): K163260

{5}------------------------------------------------

3. Comparison with predicate:

Similarities:

ItemProposed Device:SeptiCyte RAPIDPredicate device:SeptiCyte LAB
Assay MethodSameReverse transcription + quantitative PCR (RT-qPCR)
Assay PrincipleSameQuantitative gene expression assay, based on real-time generation of fluorescence from hydrolysis of dye-quencher hydrolysis probes during cycles of PCR amplification of nucleic acid templates
Detection MethodSameMulti-channel fluorescence detection
Intended Use PopulationSamePatients suspected of sepsis on their first day of Intensive Care Unit (ICU) admission
Specimen TypeSameWhole blood collected in PAXgene Blood RNA tube
Analyte(s)Two mRNA transcript immune biomarkers: PLA2G7 and PLAC8Four mRNA transcript immune biomarkers: LAMP1, CEACAM4, PLA2G7, PLAC8

Differences:

ItemProposed Device:SeptiCyte RAPIDPredicate device:SeptiCyte LAB
Specimen ProcessingAutomated extraction of materialusing the Idylla System1. Manual extraction of material inPAXgene Blood RNA Tube, usingthe IVD version of the QIAGENPAXgene Blood RNA Kit2. Semi-automated extractionPAXgene Blood RNA Tube, usingthe QIAGEN QIAcube System
PCR ChemistryOne-step RT-qPCR; ThreeSingleplex reactions; Dried formatTwo-Step RT-qPCR; one Singleplex,one Triplex reaction; Wet format
Fluorescent 5'-Probesand 3'-QuenchersPLA2G7: 5'-FAM / 3'-BHQ1PLAC8: 5'-ATTO532 / 3'-BHQ1MS2: 5'-ATTO647N / 3'-BHQ2PLA2G7: 5'-FAM / 3'-BHQ1PLAC8: 5'-Q670 / 3'-BHQ3CEACAM4: 5'-JOE / 3'-BHQ1LAMP1: 5'-FAM / 3'-BHQ1
ControlsMS2 bacteriophage particles, servingas sample processing control (SPC),i.e., a within-cartridge positive controlfor both the sample extraction stepand the coupled RT-qPCR stepHigh Positive Control, Low PositiveControl and Negative Control for eachin vitro transcript (LAMP1, CEACAM4,PLA2G7, PLAC8 (IVTs) formulated inneutral buffer, with IVT concentrationsdesigned to produce high, medium orlow SeptiScore values
Instrument PlatformBiocartis Idylla SystemApplied Biosystems 7500 Fast DxReal-Time PCR Instrument

{6}------------------------------------------------

Standard NumberStandard Name
EP05-A3Evaluation of Precision of Quantitative Measurement Procedures
EP07-A2Interference Testing in Clinical Chemistry
EP17-A2Evaluation of Detection Capability for Clinical Laboratory
Measurement Procedures
EP25-AEvaluation of Stability of In Vitro Diagnostic Reagents
Guidance
FDA Guidance for the Content of Premarket Submissions for Software Contained in
Medical Devices (issued 11 May 2005)

K. Standard/Guidance Document Referenced (if applicable):

L. Test Principle:

The SeptiCyte RAPID test employs a coupled reverse transcription - quantitative polymerase chain reaction (RT-qPCR) system to measure the expression levels (Cq values) of the host response genes PLA2G7 and PLAC8 isolated from whole blood. Amplicons generated by the RT-qPCR process are detected and quantitated by fluorescent dyes, upon exonucleolytic release of the dyes from oligonucleotide probes that specifically bind to the amplicons. The test is configured to run on the Biocartis Idylla System. The assay-specific software uses the measured PLA2G7 and PLAC8 Cq values to generate a SeptiScore, which falls into four discrete bands associated with a sequentially higher likelihood of sepsis.

M. Performance Characteristics:

1. Analytical Performance:

a. Precision/Reproducibility

Repeatability and intermediate precision were measured according to CLSI EP5-A3, using replicate PAXgene blood samples prepared from low, medium and high SeptiScore pools of PAXgene blood. Standard deviations were calculated for different subsets of data across operators, days, and instruments. Two sub-studies were performed:

Sub-Study #1 - The primary purpose of this sub-study was to evaluate repeatability (within-run precision). The study was performed using the Low sample pool, and was performed over three days, by two operators running three replicates per day, with all runs on a single instrument. Sub-study #1 was performed with a single lot of cartridges.

{7}------------------------------------------------

Sub-Study #2 - The purpose of this sub-study was to evaluate intermediate precision. Sub-study #2 was performed with one cartridge lot, over a total span of 12 days, using three instruments and two operators. Each operator's runs were conducted over 4 non-sequential days within the 12-day study window. Each combination of day, operator and instrument was tested with two replicates. At each level of sub-study #2, aggregate means and standard deviations were calculated. These were then further summarized by taking the average of the aggregates as presented in the summary table.

Table 7-1 presents a summary of results for the repeatability and intermediate precision studies for SeptiScore. Low CVs were also observed for the individual PLA2G7 and PLAC8 Cq's.

Sub-study-1Sub-study-2
RepeatabilityAcrossDaysAcrossOperatorsAcrossInstrumentsTotalVariation
PoolMEANCV(%)MEANCV(%)MEANCV(%)MEANCV(%)MEANCV(%)
Low4.564.04.484.54.484.84.484.94.485.0
MedNANA6.334.06.333.96.334.36.334.3
HighNANA10.331.810.331.910.331.910.332.0

Table 7-1: Summary of Repeatability and Intermediate Precision Results

-NA's: Only the Low pool was used for Repeatability (Sub-study-1), Medium and High pools were not run for Repeatability

b. Lot-to-Lot reproducibility

Lot-to-lot reproducibility was performed with the three sample types (Low, Medium, and High pools, as described in the previous study) using three cartridge lots, over 3 non-consecutive days, using two instruments and one operator. Each combination of lot, day, and instrument was tested with two replicates.

The samples used in the study were prepared from banked samples in PAXgene Blood RNA Tubes. The banked samples were pooled to prepare a High Pool, Medium Pool, and Low Pool, with high, medium, and low SeptiScore results, respectively. Table 7-2 summarizes the results for the lot-to-lot reproducibility study for SeptiScore. Low CVs were also observed for the individual PLA2G7 and PLAC8 Cq's.

{8}------------------------------------------------

SeptiScore
PoolMeanSDCV (%)
Low4.360.143.3
Medium6.250.193.0
High10.200.191.8

Table 7-2: Lot-to-lot reproducibility of SeptiCyte RAPID

ﻥ Linearity/assay reportable range

A specimen panel to test the linearity of quantitation of the two mRNA targets (PLA2G7, PLAC8) was constructed by spiking varying amounts of white blood cells (WBC) into a matrix consisting of platelet-reduced plasma. A broad range of WBC input amounts (25 cells/uL to 50,000 cells/μL) was tested. An inverse log-linear response (Pearson's r2 > 0.99) was observed for the individual SeptiCyte RAPID assay targets (PLA2G7 and PLAC8) when calculated across the entire input WBC concentration range. At the level of the individual analytes (PLA2G7, PLAC8), there appears to be a small Hook Effect (non-linearity) at the highest input concentrations (25,000 cells/uL and above). However, the SeptiScore, which is calculated based on relative expression of the individual analytes, compensates for the Hook Effect at the high input concentrations.

d. Traceability, Stability, Expected Values (controls, calibrators, or methods)

Controls:

A sample processing control is present in each cartridge. This consists of a predefined quantity of dried bacteriophage MS2 particles, which become mixed with the sample after sample addition but before further processing. The sample processing control is taken through the entire sample processing path, ultimately generating a PCR curve and Cq value, thereby serving as a positive control for both the extraction process and for the correct general functioning of the RT-qPCR enzymology.

Real-Time Stability:

Three manufactured lots of SeptiCyte RAPID cartridges were found to be stable for a minimum of 12 months when stored between 2°C and 30°C in their packaging as delivered (i.e., in their sealed pouches) .

In-Use Stability:

The results of the in-use stability study confirm that the SeptiCyte RAPID cartridge exhibits stability under two tested conditions:

  • When unwrapped for up to 3 days before loading the sample and running on the . ldylla instrument, and

{9}------------------------------------------------

  • When held for up to 9 hours after receiving a sample, before running on the Idylla . instrument.
    These tests were conducted at the upper end of the range of expected laboratory conditions (30 ± 2 °C and 75 ± 5 % relative humidity).

Shipping Stability:

The SeptiCyte RAPID test is packaged in secondary packaging that is consistent with other tests manufactured for the Idylla platform. The shipping stability study consisted of thermal stress and shock/vibration components. Cartridges subjected to thermal stress or shock vibration had similar SeptiScore values when compared to their respective reference condition (cartridges tested during QC release). All Cq and SeptiScore values were valid and within ±10% of their respective reference values obtained at the time of QC release.

  • e. Detection limit
    To determine the LoD and LoQ, platelet-reduced plasma was spiked with serial dilutions of white blood cells (WBCs) and 20 replicates of each level of WBC tested with the SeptiCyte RAPID Test. The lowest level at which 19/20 replicates generated a SeptiScore is reported as the LoD. The LoQ is the lowest level of WBC for which 19/20 replicates generate a SeptiScore with a standard deviation of 1.0 Score units or less. The LoD and LoQ were both determined to be 25 WBC/uL blood. Additionally, a platelet-reduced plasma sample with no spiked WBC (matrix only) was included and tested with 10 replicates to demonstrate that a SeptiScore would not be generated for any of the matrix replicates.

  • f. Analytical specificity

Interfering Substances

Based on the CLSI Guidance EP07-A2, "Interference Testing in Clinical Chemistry" SeptiCyte RAPID was evaluated in the presence of interfering substances. Each substance was added to a blood sample in a PAXgene Blood RNA tube. No interference was found for any of the substances in Table 7-3 at the concentration listed.

{10}------------------------------------------------

Potential InterferentConcentration
Rheumatoid Factor45 IU/mL
Heparin3000 U/L
Imipenem1.2 mg/mL
Bilirubin20 mg/dL
Triglycerides500 mg/dL
Vancomycin70 µmol/L
Cefotaxime670 µmol/L
Dopamine6.0 µmol/L
C-reactive protein (CRP)4 mg/dL
Noradrenaline700 pg/mL
Dobutamine12 µg/mL
Hemoglobin20 g/dL
Albumin5 g/dL
Furosemide180 µmol/L
soluble CD14 (sCD14)5 µg/mL
Interleukin 6 (IL-6)15 pg/mL
Lipopolysaccharide binding protein (LBP)45 µg/mL

Table 7-3: Interferents and Testing Levels

Lack of Reactivity Towards Blanks or Genomic DNA

This analytical specificity study was designed to confirm that the SeptiCyte RAPID test does not generate a valid signal against blank samples or in the presence of genomic DNA (gDNA).

Blank samples were prepared by adding a volume of nuclease-free phosphate buffered saline solution (PBS) to PAXgene stabilizer at the same volume ratio of specimen to RNA stabilizer expected during the standard sample collection process for PAXgene Blood RNA tubes. Twenty aliquots (0.9 mL) of this blank preparation were then tested with individual SeptiCyte RAPID cartridges. In addition to the blank samples, 10 cartridges were run empty, with no sample added to the lysis chamber

{11}------------------------------------------------

prior to running on the Biocartis Idylla system. For absence of reactivity towards gDNA. testing was conducted on 10 replicates of each of two commercially procured total human gDNA samples. Each lot was tested at 500 ng and 1000 ng gDNA input per reaction.

The results of the study conclude that (1) the SeptiCyte RAPID test does not exhibit reactivity toward blank samples and (2) the SeptiCyte RAPID test does not exhibit reactivity toward human genomic DNA, at input quantities of either 500 ng or 1000 ng per reaction.

g. Carryover

To evaluate the potential for carryover events, a series of 21 cartridges were tested with alternating blank samples, low SeptiScore positive samples, and high SeptiScore positive samples. The results of the study showed that all blank samples failed to generate a valid SeptiScore. All of the low SeptiScore samples generated a result in Band 1, and all of the high SeptiScore samples generated a result in Band 4. The results from the carryover study confirm that no carryover contamination events occur during the processing of individual SeptiCyte RAPID test cartridges.

h. PAXgene Blood RNA Tube Stability

Two matched PAXgene Blood RNA samples were collected from each of 15 healthy donors. The first sample was processed in real time, by removing two 0.9 mL aliquots for SeptiCyte RAPID analysis at the following times after phlebotomy: immediately (i.e., less than 5 min) and after 120 minutes. During this period of time, both tubes were held at room temperature (18 - 25°C) to approximate the sample handling conditions in a near-patient setting (e.g., STAT laboratory). The second PAXgene Blood RNA sample was frozen and stored at < -70°C for one month before being thawed, processed, and analyzed by SeptiCyte RAPID, to determine if storage at < -70°C, combined with one freeze/thaw cycle, has any effect on the resultant Cq and SeptiScore. The study demonstrated that matched healthy donor PAXgene blood samples did not differ significantly, when a 2-hour sample incubation step was either included or omitted. Additionally, SeptiScore values for matched healthy donor PAXgene blood samples were comparable, when samples were processed fresh (no incubation after blood draw), versus after being frozen and thawed once.

  • Assay cut-off: See clinical cutoff i.

2. Comparison studies:

  • a. Method comparison with predicate device: Not applicable
  • b. Matrix comparison:
    • Not applicable

{12}------------------------------------------------

3. Clinical Studies:

a. Prospective and Retrospective Studies

The validation and 510(k) clearance of SeptiCyte RAPID was based on samples and clinical data collected from the following observational trials, which recruited critically ill subjects suspected of sepsis:

  • Netherlands-based Molecular diagnosis And Risk Stratification of sepsis (MARS) ● clinical trial (NCT01905033) - retrospective study
  • . USA-based Validation of septic gene Expression Using SeptiCvte (VENUS) clinical trial (NCT02127502) - retrospective study
  • USA-based NEar PatienT MolecUlar TestiNg in Sepsis (NEPTUNE) clinical trial-. prospective study

The overall clinical accuracy of the SeptiCyte RAPID test was established by combining data from the retrospective (N= 356, frozen specimens) and prospective (N = 30, fresh specimens) clinical studies for a combined dataset of size N = 386.The prospective study was terminated early due to enrollment restrictions during the COVID-19 pandemic.

In both retrospective and prospective studies, the clinical status of each subject (SIRS, sepsis or indeterminate) was determined by clinical case review or retrospective physician diagnosis (RPD) conducted by an external three-member expert panel not involved in the care of the patients. Three different methods of RPD were used in analysis, as briefly described below:

Consensus: Two or three RPD panelists agree that patient is either sepsis or SIRS. Indeterminates are excluded.

Unanimous: All three RPD panelists, as well as the consensus discharge evaluation of the site investigators, agree that patient is either sepsis or SIRS. (A unanimous call of indeterminate is also theoretically possible).

Forced: Applies to patients that are initially called "indeterminate" by the RPD panelists. The RPD panelists are then forced to make a consensus or unanimous call of sepsis or SIRS.

The combined retrospective and prospective studies had the following acceptance criteria for the primary endpoint:

    1. A monotonic increase in probability of sepsis across the four SeptiScore interpretation bands;

{13}------------------------------------------------

    1. Non-overlapping 80% confidence intervals for the probability of sepsis between non-adjacent SeptiScore interpretation bands (i.e., between bands 1 and 3; and between bands 2 and 4).
      The clinical data and primary endpoint analysis are summarized in Figure 7-1 and Table 7-4:

Figure 7-1: Analysis of clinical performance data. Results are presented for (A) Forced. (B) Consensus, and (C) Unanimous RPD methods. The percentage (prevalence) of sepsis is indicated for each Band with error bars representing 80% confidence intervals. The number (N) of sepsis and SIRS in each Band is indicated to the right of the point estimate (Nseps: MsRS) as determined by each RPD method.

Image /page/13/Figure/3 description: The image shows three plots comparing the probability of sepsis based on SeptiScore bands under different decision-making conditions: Forced, Consensus, and Unanimous. Each plot displays four data points representing the probability of sepsis for SeptiScore bands 1 through 4. The Forced plot (n=386) shows probabilities of 0.125, 0.237, 0.48, and 0.797 for bands 1 to 4, respectively. The Consensus plot (n=348) shows probabilities of 0.1, 0.221, 0.453, and 0.806, while the Unanimous plot (n=254) shows probabilities of 0.111, 0.203, 0.458, and 0.821.

Table 7-4: Analysis of clinical performance data by interpretation band for each RPD Method.

Prevalence of Sepsis
RPD MethodN(80% Confidence Interval)
Band 1(0-4.9)Band 2(5.0-6.1)Band 3(6.2-7.3)Band 4(7.4-15)
Forced3860.12(0.08 - 0.18)0.24(0.18 - 0.3)0.48(0.4 - 0.56)0.80(0.74 - 0.84)
Consensus3480.1(0.06 - 0.15)0.22(0.16 - 0.29)0.45(0.37 - 0.54)0.81(0.75 - 0.85)
Unanimous2540.11(0.06 - 0.18)0.20(0.14 - 0.29)0.46(0.36 - 0.56)0.82(0.76 - 0.87)

{14}------------------------------------------------

4. Clinical interpretation

SeptiCyte RAPID generates a score (SeptiScore) that falls within one of four discrete Interpretation Bands based on the increasing likelihood of infectionpositive systemic inflammation.

SeptiScore cut-off values (i.e., the interpretation band thresholds) were established prior to the clinical trial. The following SeptiScore Interpretation tables are the results from validation clinical trials using the Forced Diagnosis (Table 7-5), Consensus Diagnosis (Table 7-6), or Unanimous RPD Diagnosis (Table 7-7).

SeptiScoreInterpretation BandPrevalence*Sepsis LikelihoodRatio
SIRSSepsis
Band 4 (7.4 - 15.0)High risk of sepsis20.379.75.23
Band 3 (6.2 - 7.3)52.048.01.24
Band 2 (5.0 - 6.1)76.323.70.41
Band 1 (0 - 4.9)Low risk of sepsis87.512.50.19

Table 7-5: Results observed using the Forced Diagnosis

Table 7-6: Results observed using the Consensus Diagnosis
SeptiScoreInterpretation BandPrevalence*Sepsis LikelihoodRatio
SIRSSepsis
Band 4 (7.4 - 15.0)High risk of sepsis19.480.65.86
Band 3 (6.2 – 7.3)54.745.31.17
Band 2 (5.0 – 6.1)77.922.10.40
Band 1 (0 – 4.9)Low risk of sepsis90.010.00.16

{15}------------------------------------------------

Prevalence*Sepsis LikelihoodRatio
SeptiScoreInterpretation BandSIRSSepsis
Band 4 (7.4 – 15.0)High risk of sepsis17.982.16.03
Band 3 (6.2 – 7.3)54.245.81.10
Band 2 (5.0 – 6.1)79.720.30.33
Band 1 (0 – 4.9)Low risk of sepsis88.911.10.16

Table 7-7: Results observed using the Unanimous Diagnosis

*Distribution observed from clinical trial

5. Expected values/Reference range:

Test samples were collected from a total of 30 (29 for Hispanic donors) individual adult donors (≥20 years of age) for each race/ethnicity (White, African American, Asian, Hispanic). The donor cohort consisted of an equal number of males (15) and females (15). SeptiCyte RAPID is not intended for use in patients not suspected of sepsis. Healthy individuals may generate scores above Band 1.

Race / EthnicitySeptiScore Interpretation Band (SeptiScoreRange)
1 (0-4.9)2 (5.0-6.1)3 (6.2-7.3)4 (7.4-15)
Asian-AmericanN42150
%13.370.016.70
African-AmericanN31683
%10.053.326.710.0
HispanicN131420
%44.848.36.90
WhiteN181110
%60.036.73.30
TotalN3862163

Table 7-8: Healthy Reference Range Band Values by Ethnic Subgroup

{16}------------------------------------------------

%31.952.113.42.5
--------------------------

N. Instrument Name:

Biocartis Idylla™ System

O. System Descriptions:

    1. Modes of Operation: See Device Description (Section I) above
      Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes

    1. Software: FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
      Yes X
    1. Specimen Identification:
      The Idylla Console and its software provide users with a Graphical User Interface for workflow management and barcode scanner to scan Sample and Cartridge barcodes. Refer to K163628 for more information.
    1. Specimen Sampling and Handling: See Section I- Device description
    1. Calibration: No calibrators are run as part of the SeptiCyte RAPID protocol.
    1. Quality Control: See Section M.1.d. Traceability, Stability, Expected values (controls, calibrators, or methods):

P. Proposed Labeling:

The proposed labeling can be found in Section 14 and satisfies the requirements of 21 CFR Part 809.10.

Q. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3215 Device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis.

(a)
Identification. A device to detect and measure non-microbial analyte(s) in human clinical specimens to aid in assessment of patients with suspected sepsis is identified as an in vitro device intended for the detection and qualitative and/or quantitative measurement of one or more non-microbial analytes in human clinical specimens to aid in the assessment of patients with suspected sepsis when used in conjunction with clinical signs and symptoms and other clinical and laboratory findings.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the device's detailed Indications for Use statement describing what the device detects and measures, the results provided to the user, whether the measure is qualitative and/or quantitative, the clinical indications for which the test is to be used, and the specific population(s) for which the device use is intended.
(2) Premarket notification submissions must include detailed documentation of the device description, including (as applicable), all device components, software, ancillary reagents required but not provided, explanation of the device principle and methodology, and for molecular devices include detailed documentation of the primer/probe sequence, design, and rationale for sequence selection.
(3) Premarket notification submissions must include detailed documentation of applicable analytical studies, such as, analytical sensitivity (Limit of Detection, Limit of Blank, and Limit of Quantitation), precision, reproducibility, analytical measuring range, interference, cross-reactivity, and specimen stability.
(4) Premarket notification submissions must include detailed documentation of a prospective clinical study or, if appropriate, results from an equivalent sample set. This detailed documentation must include the following information:
(i) Results must demonstrate adequate device performance relative to a well-accepted comparator.
(ii) Clinical sample results must demonstrate consistency of device output throughout the device measuring range likely to be encountered in the Intended Use population.
(iii) Clinical study documentation must include the original study protocol (including predefined statistical analysis plan), study report documenting support for the Indications for Use(s), and results of all statistical analyses.
(5) Premarket notification submissions must include evaluation of the level of the non-microbial analyte in asymptomatic patients with demographic characteristics (
e.g., age, racial, ethnic, and gender distribution) similar to the Intended Use population.(6) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling, and a detailed explanation of the interpretation of the limitations of the samples (
e.g., collected on day of diagnosis) must be included in the device's 21 CFR 809.10(b)(10) compliant labeling.