K Number
K203682

Validate with FDA (Live)

Date Cleared
2021-05-17

(151 days)

Product Code
Regulation Number
878.4400
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ESG-400 is an electrosurgical generator intended for tissue cutting and coagulation in open, laparoscopic and endoscopic surgery in conjunction with electrosurgical accessories and ancillary equipment.

The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, sleeve gastrectomy, hysterectomy, oophorectomy.

The POWERSEAL Sealer and Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the POWERSEAL devices for these procedures.

Device Description

The subject device ESG-400 is a reusable, non-sterile electrosurgical generator that features different mono- and bipolar cutting and coagulation modes. The maximum output power is 320 W. The front panel features a touch screen GUI that displays current settings, connection status, and allows access to menus for editing settings, procedures, and preferences. Compatible accessories include previously cleared footswitches.

The POWERSEAL 5mm laparoscopic curved jaw sealer divider is an electrosurgical bipolar device with an integral extending cutting blade. It features a pistol grip handle and will be provided in shaft lengths of 23, 37, and 44 cm. The POWERSEAL devices are sterile, single-use, hand-held bipolar electrosurgical instruments designed for use with Olympus electrosurgical generators to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics. The jaws are designed to seal vessels, and grasp and dissect tissue during open and minimally invasive general surgical procedures using high frequency (HF) energy. A hand actuated mechanism opens and closes the jaws. A second control initiates bipolar energy delivery for sealing. A separate control activates a blade for tissue division.

AI/ML Overview

The provided text details the 510(k) submission for the Olympus Electrosurgical Generator ESG-400 and Accessories, and the POWERSEAL Curved Jaw Sealer and Divider, Double Action. The study described focuses on demonstrating substantial equivalence to predicate devices, rather than establishing specific acceptance criteria and proving direct performance against them in a traditional sense. The performance data provided is primarily in the context of comparative testing to established predicate devices.

Here's an attempt to extract the information requested, with indications where the information is not explicitly available in the provided document:

1. Table of acceptance criteria and the reported device performance

The document doesn't explicitly state specific numerical acceptance criteria for performance metrics (e.g., burst pressure in mmHG, or a specific range for coagulation time). Instead, it states that the device "met all acceptance criteria" for biocompatibility and that "performance requirements defined in the User Requirements Specification and Design Specification were met for both subject devices, and that they exhibit comparable performance characteristics to the predicate device and reference devices."

The performance is described qualitatively as being "comparable" or having "equivalent technology and performance" to predicate/reference devices. For example, for vessel sealing, the key performance indicator mentioned is "vessel burst pressure testing."

Therefore, a table of acceptance criteria and reported numerical performance values cannot be fully constructed from the provided text in the typical quantitative manner.

Table 1: Acceptance Criteria and Reported Device Performance (as inferred and stated qualitatively)

Performance AspectAcceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (Qualitative Statement)
BiocompatibilityMet all criteria of ISO 10993Met all acceptance criteria for Cytotoxicity, Material Mediate Pyrogen, ISO Acute Systemic Injection Test, ISO Intracutaneous Irritation Test, ISO Guinea Pig Maximization Sensitization
Electrical Safety & EMCCompliance with IEC standardsDesign of subject devices comply with recognized standards (AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-2)
Thermal SafetyCompliance with recognized standardsDesign of subject devices comply with recognized standards (referencing Table 11, which includes IEC 60601-1, IEC 60601-2-2)
Software ValidationCompliance with FDA Guidance for "Major Level of Concern"Software validation activities performed; existing functionalities not influenced by new mode; electrical waveforms of "predicate modes" verified.
Vessel Sealing PerformanceComparable to predicate devices in animal and bench testsDemonstrated substantial equivalence to predicate device in chronic and acute animal studies. Performance requirements met, comparable to predicate and reference devices in ex-vivo vessel burst pressure testing.
Shelf Life and SterilizationCompliance with ISO 11607-1, ASTM F1980-16, ISO 11135Stability evaluation supports three-year shelf life; accelerated aging test conducted as required.
Risk ManagementAcceptable residual risk per ISO 14971Risk analysis carried out, residual risk evaluated as acceptable.

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: The document does not specify exact sample sizes for the "test set" in terms of number of cases or samples for the animal studies or bench testing (e.g., how many vessels were sealed for burst pressure testing). It only states that "ex-vivo Vessel Burst Pressure testing was conducted."
  • Data Provenance (Country of Origin, Retrospective/Prospective): Not explicitly stated. The studies were likely conducted in a controlled lab or animal facility. The document refers to FDA guidance documents, which are for U.S. regulatory submissions, but doesn't specify where the actual testing took place. It also doesn't specify if the animal studies or bench tests were retrospective or prospective, though performance testing is generally prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This type of information (number of experts, their qualifications, and their role in establishing ground truth) is typically related to clinical studies involving human interpretation or pathology. The reported studies are primarily bench (ex-vivo) and animal studies. Therefore, this information is not applicable in the context of the provided document. The "ground truth" for these studies would be objective measurements (e.g., burst pressure from instrumentation) rather than expert consensus on clinical findings.

4. Adjudication method for the test set

This is also typically relevant for clinical studies with human interpretation. For bench and animal studies (e.g., measuring vessel burst pressure), the "adjudication" is typically through objective measurements and statistical analysis, not a consensus process among experts as described by methods like 2+1 or 3+1. Therefore, this information is not applicable in the context of the provided document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study Done: No. This study is for an electrosurgical device, not an AI-assisted diagnostic tool involving human readers.
  • Effect Size of Human Readers: Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Standalone Performance: Not applicable to this type of device. The device itself is an electrosurgical tool, not an algorithm, and it is used by a human surgeon.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the performance testing, the ground truth was established through objective physical measurements in bench and animal studies:

  • Biocompatibility: In vitro and in vivo (animal tissue/systems) tests following ISO 10993 standards.
  • Electrical/Thermal Safety & EMC: Measurements against IEC standards.
  • Vessel Sealing Performance: Chronic and acute animal studies demonstrating seal performance, and ex-vivo vessel burst pressure testing. The "ground truth" here would be the measured burst pressure values and observed tissue effects.

8. The sample size for the training set

The document describes an electrosurgical device, not a machine learning or AI algorithm in the context of diagnostic imaging. Therefore, the concept of a "training set" in this context is not applicable. The "training" for such a device involves engineering design, prototyping, and testing against specifications and regulatory standards.

9. How the ground truth for the training set was established

Not applicable, as there is no "training set" in the context of an AI/ML algorithm being developed.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

May 17, 2021

Olympus Winter & Ibe GmbH % Christina Flores Regulatory Affairs Manager Olympus Surgical Technologies America 118 Turnpike Road Southborough, Massachusetts 01772

Re: K203682

Trade/Device Name: Electrosurgical Generator ESG-400 and Accessories, POWERSEAL Curved Jaw Sealer and Divider, Double Action (PS-0523CJDA, PS-0537CJDA, PS-0544CJDA). Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 16, 2020 Received: December 17, 2020

Dear Christina Flores:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/odth/cfdocs/cfpmn/pmn.cfm identifies coombination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and. if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203682

Device Name

Electrosurgical Generator ESG-400 and Accessories

Indications for Use (Describe)

The ESG-400 is an electrosurgical generator intended for tissue cutting and coagulation in open, laparoscopic and endoscopic surgery in conjunction with electrosurgical accessories and ancillary equipment.

Type of Use (Select one or both, as applicable)

☑Prescription Use (Part 21 CFR 801 Subpart D)☐Over The Counter Use (21 CFR 801 Subpart D)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

|X | Prescription Use (Part 21 CFR 801 Subpart D)

|__| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203682

Device Name

POWERSEAL Curved Jaw Sealer and Divider, Double Action (PS-0523CJDA, PS-0537CJDA, PS-0544CJDA)

Indications for Use (Describe)

The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, cholecystectomy, adhesiolysis, sleeve gastrectomy, hysterectomy, oophorectomy.

The POWERSEAL Sealer and Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the POWERSEAL devices for these procedures.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary of Safety and Effectiveness

Date Prepared: December 16, 2020

General Information

Manufacturer(ESG-400 Generator):Olympus Winter & Ibe GmbHKuehnstr. 6122045 Hamburg GermanyEstablishment Registration Number: 9610773
Manufacturer (POWERSEALSealer and Divider)Olympus Surgical Technologies America9600 Louisiana Ave. NorthBrooklyn Park, MN 55455USAEstablishment Registration Number: 3011050570
Official Correspondent:Christina Flores, RACManager, Regulatory AffairsOlympus Surgical Technologies America118 Turnpike RoadSouthborough, MA 01772Phone: 508.808-3341Email: christina.flores@olympus.comEstablishment Registration Number: 3003790304
Device Identification
Proprietary names:Electrosurgical Generator ESG-400POWERSEAL Curved Jaw Sealer and Divider,Double Action (PS-0523CJDA, PS-0537CJDA, PS-0544CJDA)
Device Classification name:Electrosurgical cutting and coagulation device andaccessories
Regulation Medical Specialty:General & Plastic Surgery
Regulations Number:Regulatory class:21 CFR 878.4400Class II
Product code:GEI
Generic/Common Name:Electrosurgical, Cutting & Coagulation & Accessories

{5}------------------------------------------------

Predicate Device

ESG-400 Generator

The initial version of the proposed medical device ESG-400 was cleared by the FDA in 2011 (K103032). This submission is based on a modification of the initial device version ESG-400 cleared by the FDA in 2014 (K141225). The 2014 clearance is the latest clearance for the device to date. The subject device ESG-400 therefore is a modification of the already legally marketed predicate electrosurgical generator ESG400 (K141225).

Predicate DeviceManufacturer510(k) No
ESG-400Olympus Winter & Ibe GmbHK141225

Table 1: Identification of predicate device

This submission has been triggered by the development of new compatible POWERSEAL instruments, and the subsequent required software update of the generator.

The ESG-400 software update extends the instrument portfolio which can be driven by the ESG-400. It adds a new coagulation mode to support the POWERSEAL technology to be used with sealing instruments for open and laparoscopic surgery.

The following reference device has been chosen to support the substantial equivalence claim in terms of safety and effectiveness for the bipolar vessel sealing mode "POWERSEAL" only.

Reference deviceManufacturer510(k) No
Valleylab™ FT10 ElectrosurgicalPlatformCovidienK151649

Table 2: Reference devices for one specific feature

POWERSEAL Sealer and Divider, Curved Jaw

The primary predicate device of the subject POWERSEAL Sealer/Divider is the Ligasure Maryland Jaw One-Step Sealer/Divider, Curved; and the reference predicate is the Ligasure Maryland Jaw Thoracic Sealer/Divider One-Step Sealing, Nano-Coated. The subject POWERSEAL shares the same fundamental technologies, including design, intended use, and principle of operation with both the predicate and reference device.

{6}------------------------------------------------

Predicate DeviceManufacturer510(k) No
Ligasure Maryland Jaw One-StepSealer/Divider, CurvedCovidienK141153
Reference deviceManufacturer510(k) No
Ligasure Maryland Jaw ThoracicSealer/Divider One-Step Sealing,CovidienK181085

Product Description

ESG-400 Generator

The subject device ESG-400 is a reusable, non-sterile electrosurgical generator that features different mono- and bipolar cutting and coagulation modes. The maximum output power is 320 W.

The front panel of the proposed ESG-400 features a touch screen GUI (graphical user interface) that displays the current settings of the chosen output mode, the connection status of accessories and peripherals connected to the electrosurgical generator.

Push buttons are placed next to GUI to switch between the output sockets, to enter the Menu in order to edit settings/ procedures (e.g. save or delete a procedure), to edit preferences (e.g. select language, touch tone control, output volume, or brightness) and to show service options (e.g. software version identifier, for service and maintenance purposes).

Compatible accessories, two footswitches a double-pedal footswitch, which is delivered with the generator and may also be ordered separately, and an optional singlepedal footswitch have been previously cleared (K141225) and are not impacted by the SW update described in this submission. Changes to the footswitches that have been made since original clearance are outlined below and discussed in Device Description and Specifications:

  • Labeling Changes -
  • Improvement of bend relief -
  • Packaging change -

POWERSEAL Sealer and Divider

The POWERSEAL 5mm laparoscopic curved jaw sealer divider is an electrosurgical bipolar device with an integral extending cutting blade. It features a pistol grip handle and will be provided in shaft lengths of 23, 37, and 44 cm.

{7}------------------------------------------------

The subject POWERSEAL devices will be provided as sterile, single-use, hand-held bipolar electrosurgical instruments designed for use with Olympus electrosurgical generators to ligate (seal) and divide (cut) vessels, tissue bundles, and lymphatics. Similar to the predicate LigaSure Maryland Jaw, the jaws of the POWERSEAL are designed to seal vessels, and grasp and dissect tissue during open and minimally invasive general surgical procedures using high frequency (HF) energy. A hand actuated mechanism allows the user to open and close the instrument jaws. When the instrument jaws are correctly placed over tissue or vessel to be sealed, the user operates a second control to initiate delivery of bipolar energy, which seals the tissue. When the sealing is complete, the user operates a separate control to activate a blade, which divides the tissue along the seal line.

The subject devices, ESG-400 and accessories and the POWERSEAL Sealer and Divider are class II medical devices under the regulation number 878.4400 and the product code GEI - "Electrosurgical cutting and coagulation device and accessories". Regulation Medical Specialty: General & Plastic Surgery.

They are compliant with FDA recognized consensus safety standards as listed in Table 11.

Indications for Use

ESG-400 Generator

The ESG-400 is an electrosurgical generator intended for tissue cutting and coagulation in open, laparoscopic and endoscopic surgery in conjunction with electrosurgical accessories and ancillary equipment.

POWERSEAL Sealer and Divider

The POWERSEAL Sealer and Divider is a bipolar electrosurgical device intended for use in laparoscopic/minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. POWERSEAL devices can be used on vessels (arteries and veins, pulmonary arteries, pulmonary veins) up to and including 7 mm, lymphatics, and tissue bundles. POWERSEAL devices are indicated for use in general surgery and such surgical specialties as urologic, colorectal, bariatric, vascular, thoracic, and gynecologic. Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, sleeve gastrectomy, hysterectomy, oophorectomy.

The POWERSEAL Sealer and Divider has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the POWERSEAL devices for these procedures.

Technological Characteristics

{8}------------------------------------------------

ESG-400 Generator

The ESG-400 has the same intended use and technological characteristics as the predicate device ESG-400 (K141225).

Various instruments can be connected to various output sockets: two monopolar socket, one bipolar socket and one universal socket. In addition, dedicated Olympus instruments or Olympus cables can be connected to the universal socket with instrument recognition.

With respect to the predicate device ESG-400 (K141225) the basic design philosophy of the User Interface (UI) and GUI flow chart concept was not changed. User interface modifications were the result of implementing the new output mode to support the POWERSEAL technology as well as improvement of GUI usability derived from market surveillance. POWERSEAL technology related GUI changes have been implemented. They follow the general GUI flow chart concept of the ESG-400 referring to All/ Set/ Mode Screens.

Output modes in comparison to the predicate device ESG-400

In this chapter the different unchanged monopolar and bipolar cutting and coagulation modes of the subject and predicate device are listed:

Subject Device:ESG-400Predicate Device:ESG-400 (K141225)
PureCutPureCut
BlendCutBlendCut
PulseCut SlowPulseCut Slow
PulseCut FastPulseCut Fast
FineCutFineCut
Table 3: Monopolar Cut Modes
Subject Device:ESG-400Predicate Device:ESG-400 (K141225)
SoftCoagSoftCoag
ForcedCoagForcedCoag
SprayCoagSprayCoag
PowerCoagPowerCoag
Table 4: Monopolar Coagulation Modes
Subject Device:ESG-400Predicate Device:ESG-400 (K141225)
BipolarCutBipolarCut
SalineCutSalineCut
Table 5: Bipolar Cut Modes
Subject Device:ESG-400Predicate Device:ESG-400 (K141225)
BiSoftCoagBiSoftCoag

{9}------------------------------------------------

AutoCoagAutoCoag
Subject Device:ESG-400Predicate Device:ESG-400 (K141225)
SalineCoagSalineCoag
HardCoagHardCoag
FineCoagFineCoag
RFCoag (RCAP)RFCoag (RCAP)

Table 6: Bipolar Coagulation Modes

Subject Device:ESG-400Predicate Device:ESG-400 (K141225)
PK PureCutPK PureCut
PK SoftCutPK SoftCut
PK LoopCutPK LoopCut
PK MorceCutPK MorceCut

Table 7: Bipolar Plasmakinetic Cut Modes

Subject Device:ESG-400Predicate Device:ESG-400 (K141225)
PK CoagPK Coag
PK SoftCoagPK SoftCoag
PK AutoCoagPK AutoCoag

Table 8: Bipolar Plasmakinetic Coagulation Modes

The range of output waveforms and the power levels are identical in comparison to the FDA cleared ESG-400 electrosurgical generator, K141225.

Output modes in comparison to the reference device

As stated above, the reference device, Covidien Valleylab FT10 Electrosurgical Platform (K151649), is solely used for the newly implemented bipolar vessel sealing mode "POWERSEAL", that is comparable in the subject device and in the reference device.

Subject Device:ESG-400Reference DeviceCOVIDIEN Valleylab™ FT10Electrosurgical Platform (K151649)
SEALLigaSure

Table 9: Bipolar Coagulation Mode - Reference Device - COVIDIEN Valleylab™ FT10 Electrosurgical Platform

{10}------------------------------------------------

The range of output waveforms and the power levels of the POWERSEAL mode are comparable to the LigaSure mode of the FDA cleared COVIDIEN Valleylab™ FT10 Electrosurgical Platform, K151649.

POWERSEAL Sealer and Divider

The basic fundamental technology, including design, intended use, and principles of operation are the same between the subject and predicate Ligasure devices. Both the subject and predicate devices connect to a radiofrequency (RF) energy source The Indications for Use statement is also similar but identifies additional surgical specialties that otherwise fall under the broader indications of the predicate devices. The following table outlines the similarities and differences between the subject and predicate devices.

{11}------------------------------------------------

K203682 Electrosurgical Generator ESG-400 and Accessories Olympus Winter & Ibe GmbH

Traditional 510(k) Notification

FeatureSubject DevicePOWERSEAL Sealer and DividerPredicate DeviceLigaSure Maryland Jaw Sealer/Divider(#K141153)Reference DeviceLigasure Maryland Jaw ThoracicSealer/Divider One-Step Sealing, Nano-Coated (#K181085)
Indications forUseThe POWERSEAL Sealer and Divider is a bipolarelectrosurgical device intended for use inlaparoscopic/minimally invasive or open surgicalprocedures where ligation and division of vessels,tissue bundles, and lymphatics is desired.POWERSEAL devices can be used on vessels(arteries and veins, pulmonary arteries, pulmonaryveins) up to and including 7 mm, lymphatics, andtissue bundles. POWERSEAL devices areindicated for use in general surgery and suchsurgical specialties as urologic, colorectal,bariatric, vascular, thoracic, and gynecologic.Procedures may include, but are not limited to,Nissen fundoplication, colectomy,cholecystectomy, adhesiolysis, sleeve gastrectomy,hysterectomy, oophorectomy. The POWERSEALSealer and Divider has not been shown to beeffective for tubal sterilization or tubal coagulationfor sterilization procedures. Do not use thePOWERSEAL devices for these procedures.The LigaSure Sealer/Divider is a bipolarelectrosurgical instrument intended for usein minimally invasive or open surgicalprocedures where ligation and division ofvessels, tissue bundles, and lymphatics isdesired. The LigaSure Sealer/Divider can beused on vessels (arteries and veins) up to andincluding 7 mm. It is indicated for use ingeneral surgery and such surgicalspecialities as urologic, vascular, thoracic,and gynecologic. Procedures may include,but are not limited to, Nissen fundoplication,colectomy, cholecystectomy, adhesiolysis,hysterectomy, oophorectomy, etc. TheLigaSure system has not been shown to beeffective for tubal sterilization or tubalcoagulation for sterilization procedures. Donot use the LigaSure system for theseprocedures.The LigaSure Sealer/Divider is a bipolarelectrosurgical instrument intended for use inminimally invasive or open surgicalprocedures where ligation and division ofvessels, tissue bundles, and lymphatics isdesired. The LigaSure Sealer/Divider can beused on vessels (arteries and veins, pulmonaryarteries, pulmonary veins) up to and including7 mm. It is indicated for use in generalsurgery and such surgical specialities asurologic, vascular, thoracic, and gynecologic.Procedures may include, but are not limitedto. Nissen fundoplication, colectomy,cholecystectomy, adhesiolysis, hysterectomy,oophorectomy, etc.The LigaSure system has not been shown tobe effective for tubal sterilization or tubalcoagulation for sterilization procedures. Donot use the LigaSure system for theseprocedures.
DesignPistol Grip with Shaft and curved jawPistol Grip with Shaft and curved jawPistol Grip with Shaft and curved jaw
Size Range5mm diameter in lengths of 23, 37, 44cm5mm diameter in lengths of 23, 37, 44cm5mm diameter in length of 30cm
Output modeand nameBipolar Sealing = POWERSEAL (Coag)Bipolar Sealing = LigaSure (Coag)Bipolar Sealing = LigaSure (Coag)
Sterility, singlepatient use,disposableYesYesYes

{12}------------------------------------------------

K203682 Electrosurgical Generator ESG-400 and Accessories Olympus Winter & Ibe GmbH

Table 10: Subject and Predicate comparison table

{13}------------------------------------------------

Substantial Equivalence

ESG-400 Generator

Substantial equivalence is demonstrated by acknowledged verification/validation methodologies. The subject device is a modified version of the predicate device ESG400 (K141225) and has equivalent technology, performance, dimensions and materials. The difference to the predicate device ESG-400 is:

  • The newly implemented bipolar coagulation mode to support the POWERSEAL technology to be used with sealing instruments for open and laparoscopic surgery.
    Regarding the additionally implemented bipolar coagulation mode "POWERSEAL". one reference devices has been chosen, because of its specific output mode. For this reference devices the substantial equivalence is demonstrated by acknowledged verification/ validation methodologies. The reference device has equivalent technology and performance in respect to the compared mode.

POWERSEAL Sealer and Divider

In establishing substantial equivalence of the subject POWERSEAL Sealer and Divider to the predicate Ligasure Maryland devices, an evaluation of the indications for use, intended use and technological characteristics was conducted. The subject and predicate devices have similar technology, performance, dimensions and materials. Performance testing confirmed that the subject device is as safe and effective as the predicate device for the proposed indications for use.

Performance Testing

The following performance data were provided in support of the substantial equivalence determination. All standards applied are FDA recognized conformance standards.

For the ESG-400 Generator, all data was prepared in accordance with the FDA guidance, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" Guidance for Industry and Food and Drug Administration Staff, issued on March 9, 2020. The guidance was followed for all relevant sections.

For the POWERSEAL Sealer and Divider, all data was prepared in accordance with the following FDA guidance documents: "Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery" Guidance for Industry and Food and Drug Administration Staff, issued on August 15, 2016; and "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgerv" Guidance for Industry and Food and Drug Administration Staff, issued on March 9, 2020. The guidance documents were followed for all relevant sections.

{14}------------------------------------------------

Biocompatibility testing

ESG-400 Generator

The ESG-400 and its accessories do not contain components that come directly or indirectly in patient contact. Biocompatibility testing according to ISO 10993 is not required.

POWERSEAL Sealer and Divider

In accordance with ISO 10993-1 and the 2016 FDA guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", the subject devices met all acceptance criteria for the following biocompatibility evaluations:

CytotoxicityISO 10993-5:2009 Biological Evaluation of Medical Devices, Part5: Tests for In Vitro Cytotoxicity
Material Mediate PyrogenISO 10993-11:2017. Biological Evaluation of Medical Devices, Part11: Tests for Systemic Toxicity.
ISO Acute SystemicInjection TestISO 10993-11: 2017 Biological Evaluation of Medical Devices Part11: Tests for Systemic Toxicity.
ISO Intracutaneous IrritationTestISO 10993-10: 2010 Standard, Biological Evaluation of MedicalDevices, Part 10: Tests forIrritation and Skin Sensitization.
ISO Guinea PigMaximization SensitizationISO 10993-10:2010. Biological Evaluation of Medical Devices,Part 10: Tests for Irritation and Skin Sensitization.

Sterilization and Shelf Life Discussion

ESG-400 and accessories

The ESG-400 generator and its accessories are not provide sterile. They are reusable devices.

POWERSEAL Sealer and Divider

Sterilization testing for the POWERSEAL Sealer and Divider was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile". Stability evaluation for the sterile packaging and for product performance supports the three year shelf life. Accelerated aging test was conducted as required per ISO 11607-1 and in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices.

{15}------------------------------------------------

Electrical safety and electromagnetic compatibility (EMC)

Basic safety and performance testing was performed in accordance with IEC standards. The design of the subject ESG-400 and footswitches and the subject POWERSEAL Sealer and Divider comply with recognized standards as listed in Table 11, respectively.

The FDA guidance "Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices", CDRH July 11, 2016 has been followed.

Thermal Safety

The design of the subject ESG-400 Generator and the subject POWERSEAL Sealer and Divider, comply with recognized standards as listed in Table 11.

Clinical and Animal Studies

Clinical studies were not necessary for the subject devices.

Animal Studies, including Acute and Chronic testing conducted demonstrate substantial equivalence of the subject POWERSEAL Sealer and Divider to the predicate device.

All data was prepared in accordance with the FDA guidance. "Premarket Notification (510/k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgerv" Guidance for Industry and Food and Drug Administration Staff, issued on August 15, 2016; and "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery'' Guidance for Industry and Food and Drug Administration Staff, issued on March 9, 2020. The guidance documents were followed for all relevant sections.

TestContents
Chronic Animal StudyChronic animal study was conducted onboth the subject and predicate devices todemonstrate seal performance
Acute Animal StudyAcute animal study was conducted onboth the subject and predicate devices todemonstrate seal performance andsafety

{16}------------------------------------------------

Software

The software validation activities were performed for the subject ESG-400 Generator in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", issued May 11, 2005. The device software is considered a "Major Level of Concern".

The output modes of the subject device to the predicate device (K141225) have not been changed. Software regression tests confirmed that existing functionalities are not influenced and remain unaffected/ unchanged by the addition of the new coagulation mode. The electrical waveforms of these "predicate modes" have been verified.

The POWERSEAL Sealer and Divider does not contain software.

Performance Testing Bench

Verification and comparison bench studies were conducted to evaluate the functional performance of the newly implemented "POWERSEAL" mode. Ex-vivo Vessel Burst Pressure testing was conducted on both the subject and predicate devices to demonstrate vessel sealing performance.

Testing demonstrated that the performance requirements defined in the User Requirements Specification and Design Specification were met for both subject devices, and that they exhibit comparable performance characteristics to the predicate device and reference devices.

Bench testing results support the claim of substantial equivalence of the subject ESG400 and the subject POWERSEAL Sealer and Divider, to the predicate and reference devices.

The following non-clinical and preclinical tests were conducted:

    1. non-clinical (electrical, mechanical, functional)
    1. preclinical (simulated use) evaluation and testing of tissue effects and thermal safety and vessel burst pressure testing and vessel thermal margin

Usability and user interface were also assessed according to the risk management plan. Use-related hazardous situations were assessed and risk mitigation measures in terms of usability design for safety were defined. The residual risk was evaluated as acceptable.

Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971.

{17}------------------------------------------------

Reprocessing

Required cleaning, disinfecting and drying procedures are described in the instructions for use for the ESG-400 Generator and accessories.

The POWERSEAL Sealer and Divider is provided sterile, for single-use. It is not intended to be reprocessed.

Standard No.Standard TitleFDA-Recognitionno + dateApplicability
AAMI/ANSI ES60601-1:2005/(R)2012and C1:2009/(R)2012and, A2:2010/(R)2012Medical electrical equipment - Part 1: Generalrequirements for basic safety and essentialperformance (IEC 60601-1:2005, MOD)19-407/09/2014-ESG-400 Generator-POWERSEALInstrument
IEC 60601-1-2 Ed. 4.0:2014Medical electrical equipment - Part 1-2:General requirements for basic safety andessential performance - Collateral Standard:Electromagnetic disturbances - Requirementsand tests19-809/17/2018-ESG-400 Generator-POWERSEALInstrument
IEC 60601-1-8 Ed. 2.1:2012Medical electrical equipment - Part 1-8:General requirements for basic safety andessential performance - Collateral Standard:General requirements, tests and guidance foralarm systems in medical electrical equipmentand medical electrical systems5-7608/06/2013-ESG-400-POWERSEALInstrument
IEC 60601-2-2 Ed. 6.0:2017Medical electrical equipment - Part 2-2:Particular requirements for the basic safety andessential performance of high frequencysurgical equipment and high frequencysurgical accessories6-38908/21/2017-ESG-400 Generator-POWERSEALInstrument
IEC62304Ed. 1.12015consolidated versionMedical device software - Software life cycleprocesses13-7901/14/2019-ESG-400
IEC 60601-1-6 Ed. 3.1:2013Medical electrical equipment Part 1-6: Generalrequirements for basic safety and essentialperformance - Collateral standard:Usability5-8906/27/2016-ESG-400
IEC 62366-1Edition 1.02015Medical devices - Part 1: Application ofusability engineering to medical devices[Including CORRIGENDUM 1 (2016)]5-11412/23/2016-ESG-400 Generator-POWERSEALInstrument
ISO 14971 secondedition 2007/2019Medical devices – Application of riskmanagement to medical devices5-4006/27/2016-ESG-400 Generator-POWERSEAL
Instrument
ISO 11135: 2014Sterilization of health care products - Ethyleneoxide - Requirements for development,validation and routine control of a sterilizationprocess for medical devices.14-52907/15/2019-POWERSEALInstrument
ISO 11607-1Second edition 2019-02Packaging for terminally sterilized medicaldevices - Part 1: Requirements for materials,sterile barrier systems and packaging systems14-53007/15/2019-POWERSEALInstrument
ASTM F1980-16Standard Guide for Accelerated Aging ofSterile Barrier Systems for Medical Devices14-49712/23/2016-POWERSEALInstrument
ISO 10993-5: 2009Biological Evaluation of Medical Devices,Part 5: Tests for In Vitro Cytotoxicity2-24512/23/2016-POWERSEALInstrument
Standard No.Standard TitleFDA-Recognitionno +dateApplicability
ISO 10993-10: 2010Biological Evaluation of Medical Devices,Part 10: Tests forIrritation and Skin Sensitization2-17407/26/2016-POWERSEALInstrument
ISO 10993-11: Thirdedition 2017-09Biological Evaluation of Medical Devices,Part 11: Tests for Systemic Toxicity2-25509/17/2018-POWERSEALInstrument
ISO 10993-7: Secondedition 2008-10-15Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals14-40801/30/2014-POWERSEALInstrument
ISO 10993-18:2020Biological evaluation of medical devices —Part 18: Chemical characterization of medicaldevice materials within a risk managementprocess2-27607/06/2020-POWERSEALInstrument

Applied standards

{18}------------------------------------------------

K203682 Electrosurgical Generator ESG-400 and Accessories Olympus Winter & Ibe GmbH

Traditional 510(k) Notification

Table 11: Applied standards

Conclusion

The performance data supports the safety of the device and demonstrate that the subject devices comply with the recognized standards as specified.

In summary, Olympus believes the ESG-400 Generator and accessories and the POWERSEAL Sealer and Divider are substantially equivalent to the predicate devices with respect to the general design approach, function, and the intended use. The subject ESG-400 Generator and the POWERSEAL Sealer and Divider raise no new concerns of safety or effectiveness when compared to the predicates and the reference devices.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.