K Number
K203671

Validate with FDA (Live)

Date Cleared
2021-01-13

(28 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The M.U.S.T. Sacral Illac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacroiliac joint disruptions and degenerative sacroiliitis.

Device Description

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is designed for sacroiliac joint fusion in degenerative sacroiliitis, as well as for the fixation of small and long bone fractures in trauma cases.

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System consists of various sized screws, manufactured from Ti6A14V ELI and coated with rough Hydroxyapatite (HA). The HA coating allows for biological fixation and potentially leads to arthrodesis.

The sacroiliac (SI) joint screws are hollow-body threaded fusion devices with a multiplefenestrated pattern shaft to promote arthrodesis, a self-tapping design to facilitate screw insertion, and a tapered tip to aid in guidance through pilot hole. Radial windowed slots along the shaft facilitate bone ingrowth after implantation and "bone filling" after insertion.

The SI joint screws are provided in standard and headless designs. The standard screw can be coupled to modular washers of different diameters to optimize and stabilize the contact between the screw head and the cortical bone and to improve the compression activity of the screw. The standard screws are provided sterile in three diameters (8, 9, and 10 mm) and multiple lengths (25 – 80 mm).

The headless screw has an anatomical shape which allows for full insertion into the bone. The headless screws are provided sterile in three diameters (7.5, 9, and 11 mm) and multiple lengths (30 - 75 mm). The M.U.S.T. Sacral Iliac Screws Extension introduces new sizes of these screws keeping the same diameters (7.5, 9, and 11 mm) with new lengths from 80 mm (5 mm increment).

AI/ML Overview

The provided FDA 510(k) document does not describe acceptance criteria for a device, nor does it present the results of a study designed to prove the device meets specific acceptance criteria in the context of device performance as one might expect for AI/ML or diagnostic devices.

Instead, this document is a 510(k) submission for a medical device (M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System), seeking to demonstrate substantial equivalence to a previously cleared predicate device. Substantial equivalence is a regulatory pathway for medical devices in the US, where a new device is compared to a legally marketed predicate device, and it's shown that the new device is as safe and effective as the predicate.

The "performance data" section in this document refers to design verification and validation testing related to the physical characteristics and safety of the bone screws, not diagnostic accuracy or clinical effectiveness in the way an AI/ML device would be evaluated.

Therefore, many of the questions in your prompt are not applicable to the information contained in this 510(k) submission.

Here's an attempt to answer the applicable parts based on the provided text, while clearly indicating where information is not present:


1. Table of acceptance criteria and the reported device performance

This document does not define specific "acceptance criteria" in terms of performance metrics like sensitivity or specificity. Instead, the "acceptance" for this device is based on demonstrating substantial equivalence to a predicate device through engineering and material testing.

The document states that the introduction of new sizes for the M.U.S.T. SI-Joint Headless screws does not create a new worst case compared to the predicate device. Therefore, a comparative analysis was performed to determine if the line extension created a new worst-case product size, and it was determined that the subject devices are substantially equivalent to the previously cleared predicate device.

The "performance data" section lists the following tests that were leveraged from the predicate device:

Test/StandardReported Performance/Outcome (Implied Acceptance)
ASTM F543-13 (Metallic Medical Bone Screws)(Implied compliance for mechanical properties)
ASTM F2193-14 (Components in Spinal Skeletal System Fixation)(Implied compliance for mechanical properties)
ISO 13779-3: 2008 (Hydroxyapatite - Chemical Analysis & Characterization)(Implied compliance for HA coating quality)
ASTM F1185-03 (Reapproved 2014) (Composition of Hydroxyapatite)(Implied compliance for HA coating composition)
Sterilization Validation(Implied successful validation of sterilization method)
Shelf-life Testing(Implied successful validation of shelf life)
Cadaver/Sawbones Workshop(Implied successful demonstration of usability/functionality in simulated environment)
Comparative analysis to predicate device for new sizes (M.U.S.T. SI-Joint Headless screws)Determined that new sizes do not create a new worst case and are substantially equivalent to predicate.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size: Not explicitly stated for each test. These are typically engineering tests performed on a limited number of physical samples (e.g., screws) according to the respective ASTM/ISO standards.
  • Data provenance: Not specified, but likely from internal Medacta International SA testing (Switzerland). These are not clinical studies with patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable. This device is a bone fixation fastener, not a diagnostic or AI/ML device requiring expert ground truth for performance evaluation in the clinical sense. The "ground truth" here is compliance with engineering standards and demonstration of substantial equivalence.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not a study involving human reader consensus for "ground truth."


5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML or diagnostic device.


6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

Not applicable. This is not an AI/ML device.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" in this context is adherence to established engineering standards (ASTM, ISO) for material properties, mechanical performance, sterilization, and biocompatibility, as well as the device's physical and mechanical comparability to the predicate device. It does not involve clinical outcomes data or expert consensus in a diagnostic sense.


8. The sample size for the training set

Not applicable. This is not an AI/ML device, so there is no "training set."


9. How the ground truth for the training set was established

Not applicable. As above, no training set.

{0}------------------------------------------------

May 4, 2021

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Chris Lussier Senior Director, Quality and Regulatory Medacta International SA 3973 Delp Street Memphis, Tennessee 38118

Re: K203671

Trade/Device Name: M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR, HWC, HTN

Dear Chris Lussier:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 13, 2021. Specifically, FDA is updating this SE Letter as an administrative correction for an incorrect Indications for Use Statement.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Colin O'Neill, M.B.E., OHT6: Office of Orthopedic Devices, (301) 796-6428, colin.oneill@fda.hhs.gov.

Sincerely.

Colin O'neill -S

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

January 13, 2021

Image /page/1/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration" with the word "Administration" on the second line.

Chris Lussier Senior Director, Quality and Regulatory Medacta International SA 3973 Delp Street Memphis, Tennessee 38118

Re: K203671

Trade/Device Name: M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: OUR, HWC, HTN Dated: December 15, 2020 Received: December 16, 2020

Dear Mr. Lussier:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{2}------------------------------------------------

801 and Part 809; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'neill -S

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203671

Device Name

M.U.S.T. Sacro-Iliac Screws and Pelvic Trauma System

Indications for Use (Describe)

The M.U.S.T. Sacral Illac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacroiliac joint disruptions and degenerative sacroiliitis.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

2.0 510(k) Summary

I. Submitter

Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66

Contact Person: Stefano Baj, Regulatory Affairs Manager, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director Quality / Regulatory, Medacta USA Date Prepared: December 15, 2020

II. Device

Device Proprietary Name:M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System
Common or Usual Name:Pelvic Joint Fixation
Classification Name:Smooth or threaded metallic bone fixation fastener
Primary Product Code:OUR
Secondary Product Codes:HWC, HTN
Regulation Number:21 CFR 888.3040, 21 CFR 888.3030
Device ClassificationII

III. Predicate Device

Substantial equivalence is claimed to the following device:

Primary Predicate Device:

  • . M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System, K193083, Medacta International SA

Reference Predicate Device

  • M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System, K171595, Medacta ● International SA

{5}------------------------------------------------

IV. Device Description

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is designed for sacroiliac joint fusion in degenerative sacroiliitis, as well as for the fixation of small and long bone fractures in trauma cases.

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System consists of various sized screws, manufactured from Ti6A14V ELI and coated with rough Hydroxyapatite (HA). The HA coating allows for biological fixation and potentially leads to arthrodesis.

The sacroiliac (SI) joint screws are hollow-body threaded fusion devices with a multiplefenestrated pattern shaft to promote arthrodesis, a self-tapping design to facilitate screw insertion, and a tapered tip to aid in guidance through pilot hole. Radial windowed slots along the shaft facilitate bone ingrowth after implantation and "bone filling" after insertion.

The SI joint screws are provided in standard and headless designs. The standard screw can be coupled to modular washers of different diameters to optimize and stabilize the contact between the screw head and the cortical bone and to improve the compression activity of the screw. The standard screws are provided sterile in three diameters (8, 9, and 10 mm) and multiple lengths (25 – 80 mm).

The headless screw has an anatomical shape which allows for full insertion into the bone. The headless screws are provided sterile in three diameters (7.5, 9, and 11 mm) and multiple lengths (30 - 75 mm). The M.U.S.T. Sacral Iliac Screws Extension introduces new sizes of these screws keeping the same diameters (7.5, 9, and 11 mm) with new lengths from 80 mm (5 mm increment).

Indications for Use V.

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System is intended for use in skeletally mature patients for fracture fixation of small and long bones of the pelvis, and for sacroiliac joint fusion for patients suffering from sacroiliac joint disruptions and degenerative sacroillitis.

VI. Comparison of Technological Characteristics

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System SI joint screws and the predicate devices share the following characteristics:

  • material of construction; ●
  • diameters; ●
  • coating and coating composition; ●
  • biocompatibility; ●

{6}------------------------------------------------

  • sterility;
  • shelf life; and ●
  • packaging. ●

The M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System SI joint screws are technologically different from the predicate device with respect to length of the screw and number of the radial windowed slots on the screw shaft.

These differences do not raise different questions of safety or effectiveness and design verification testing supports there are no new risks associated with the subject devices.

Performance Data VII.

The introduction of new sizes of M.U.S.T. SI-Joint Headless screws does not create a new worst case; therefore, the following performance testing from the predicate device was leveraged to support this submission:

  • ASTM F543-13 Standard Specification and Test Methods for Metallic Medical Bone ● Screws;
  • ASTM F2193-14 Standard Specifications and Test Methods for Components Used in ● the Surgical Fixation of the Spinal Skeletal System;
  • ISO 13779-3: 2008 Implants for Surgery Hydroxyapatite Part 3: Chemical Analysis ● and Characterization of Crystallinity and Phase Purity;
  • ASTM F1185-03 (Reapproved 2014) Standard Specification for Composition of ● Hydroxyapatite for Surgical Implants;
  • sterilization validation; and ●
  • shelf-life testing ●
  • Cadaver/Sawbones Workshop ●

A comparative analysis of the subject devices to the identified predicate and reference devices was performed to determine if this line-extension created a new worst-case product size. It was determined that the subject devices are substantially equivalent to the previously cleared predicate device and that the line-extension can be considered within the scope of the design verification and validation testing previously completed for the initial M.U.S.T. Sacral Iliac Screw and Pelvic Trauma System product range.

VIII. Conclusion

The information provided within this submission supports that the M.U.S.T. SI-Joint Headless screws are are substantially equivalent to the predicate device.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.