K Number
K203667

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-02-02

(48 days)

Product Code
Regulation Number
892.1720
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EzRay M (Model: VMX-P300) is a portable general-purpose X-ray system that users can operate with one hand. The device uses a fixed tube current and voltage (kVp) and, therefore, is limited to taking diagnostic X-rays of extremities. It is intended to be used by a qualified and trained clinician on adult patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp), which may be required for full optimization of image quality and radiation exposure for different exam types.

Device Description

EzRay M (Model: VMX-P300), a medical portable X-ray system, operates on 21.6 Vdc supplied by a rechargeable Li-ion battery pack. The system is designed for medical examination and composed of an X-ray generating part with an X-ray tube including a device controller, a power controller, a user interface, a beam limiting part, and optional items. It is intended to be used by a qualified and trained clinician on adult patients. The device is intended to assist the diagnosis of bones and tissues through X-ray exposure using an imaging receptor. The image receptor (an integral part of a complete diagnostic system) is not part of this submission.

AI/ML Overview

This is a 510(k) premarket notification for a mobile x-ray system, EzRay M (Model: VMX-P300). The document does not describe the acceptance criteria and a study that proves a device meets acceptance criteria for performance relevant to AI/ML. Instead, it focuses on demonstrating substantial equivalence to a predicate device (NOMAD MD Handheld X-ray System, K140723) based on device characteristics, indications for use, and compliance with various international and federal safety standards for X-ray equipment.

Therefore, I cannot provide the requested information about acceptance criteria and a study proving device performance in the context of AI/ML, as this document does not contain that type of information. It is a regulatory submission for a traditional medical device (X-ray system), not an AI/ML powered device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 2, 2021

Vatech Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 7707 Fannin Street, Suite 200-V111 HOUSTON TX 77054

Re: K203667

Trade/Device Name: EzRay M (Model: VMX-P300) Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: December 10, 2020 Received: December 16, 2020

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203667

Device Name EzRay M (Model: VMX-P300)

Indications for Use (Describe)

EzRay M (Model: VMX-P300) is a portable general-purpose X-ray system that users can operate with one hand. The device uses a fixed tube current and voltage (kVp) and, therefore, is limited to taking diagnostic X-rays of extremities. It is intended to be used by a qualified and trained clinician on adult patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp), which may be required for full optimization of image quality and radiation exposure for different exam types.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 – 510(k) Summary K203667

1. Traditional 510(k) Summary

This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

2. Date 510K Summary prepared: January15, 2021

3. Administrative Information

Official Correspondent:Dave Kim / Mtech GroupAddress: 7707 Fannin St., Ste. 200-V111, Houston, TX 77054Tel: +713-467-2607Contact person: Mr. Dave Kim (davekim@mtech-inc.net)
510(k) Submitter:VATECH Co., Ltd.Address: 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, 18449, KoreaTel: +82-31-379-9492Fax: +82-31-379-9400Contact person: Daniel Kim / Manager (daniel.kim@vatech.com)

4. Device Information

Type of 510(k) Submission:Traditional
Trade or Proprietary Name:EzRay M (Model: VMX-P300)
Common or Usual Name:Medical Portable X-ray System
Regulation Classification:Mobile X-ray system (21 CFR 892.1720)
Product Code:IZL
Class of Device:Class II
Panel:Radiology

5. Predicate Device Information

Manufacturer:Aribex, Inc.
Trade or Proprietary Name:NOMAD MD Handheld X-ray System
Common or Usual Name:Nomad MD
Regulation Classification:Mobile X-ray system (21 CFR 892.1720)
Product Code:IZL
Class of Device:Class II
Panel:Radiology
510(k) Number:K140723

{4}------------------------------------------------

6. Device Description

EzRay M (Model: VMX-P300), a medical portable X-ray system, operates on 21.6 Vdc supplied by a rechargeable Li-ion battery pack. The system is designed for medical examination and composed of an X-ray generating part with an X-ray tube including a device controller, a power controller, a user interface, a beam limiting part, and optional items. It is intended to be used by a qualified and trained clinician on adult patients. The device is intended to assist the diagnosis of bones and tissues through X-ray exposure using an imaging receptor. The image receptor (an integral part of a complete diagnostic system) is not part of this submission.

7. Indication for use

EzRay M (Model: VMX-P300) is a portable general-purpose X-ray system that users can operate with one hand. The device uses a fixed tube current and voltage (KVp) and, therefore, is limited to taking diagnostic X-rays of extremities. It is intended to be used by a qualified and trained clinician on adult patients. It is not intended to replace a radiographic system with variable tube current and voltage (kVp), which may be required for full optimization of image quality and radiation exposure for different exam types.

{5}------------------------------------------------

8. Substantial Equivalence Chart

Subject DevicePredicate Device
Device NameEzRay M (Model: VMX-P300)NOMAD MD Handheld X-ray System
Applicant NameVATECH Co., Ltd.Aribex, Inc.
510(k) NumberK203667K140723
Device Classification NameMobile x-ray systemMobile x-ray system
Classification Product CodeIZLIZL
Regulation Number21 CFR 892.172021 CFR 892.1720
Regulation ClassIIII
Indications for UseEzRay M (Model: VMX-P300) is aportable general-purpose X-ray systemthat users can operate with one hand.The device uses a fixed tube currentand voltage (kVp) and, therefore, islimited to taking diagnostic X-rays ofextremities. It is intended to be usedby a qualified and trained clinician onadult patients. It is not intended toreplace a radiographic system withvariable tube current and voltage(kVp), which may be required for fulloptimization of image quality andradiation exposure for different examtypes.The NOMAD MD is a handheld andportable general purpose X-ray system.The device uses a fixed tube current andvoltage (kVp)and, therefore, is limited to takingdiagnostic X-rays of extremities. It isintended to be used by a qualified andtrained clinician on both adult andpediatric patients. It is not intended toreplace a radiographic system withvariable tube current and voltage (kVp)which may be required for fulloptimization of image quality andradiation exposure for different examtypes.
Principle of OperationGeneral Purpose Diagnostic X RayGeneral Purpose Diagnostic X Ray
Size: Body11.0" L x 5.4" W x 11.1" H(excluding Skin Distance Bar)9.5" H x 5.25" W x 10" L(excluding Source Skin guard)
Weight4.14 lbs11.0 lbs.
Source to skin distance30 cm with Skin Distance Bar30 cm
Focal Spot0.5 mm
TechnologicalCollimatorFour manually and steplesslyadjustable shutters with x-ray fieldindicatorFour manually and steplessly adjustableshutters with LED Light Field CenterIndicator
User InterfaceJog dial for operating mode selection.Additionally, several user-selectablepreset times with exposure timeselection icons on a display module.Up-down buttons for exposure timeselection, with timer display.
Energy sourceRechargeable 21.6 V DCLi-ion polymer battery packRechargeable 14.4 V DCNiCd battery pack
Exposure time0.05 - 1.0 secondsin 0.01 increments0.02 - 0.99 secondsin 0.01 increments
mA3.0 mA fixed
kVp65 kVp fixed75 kVp fixed

{6}------------------------------------------------

9. The differences between the subject device and the predicate device

The subject device described in this traditional 510(k) is similar to the predicate device (K140723) in its indications for use and technological application. Both the subject and predicate devices are taking diagnostic Xrays of extremities using a fixed tube current and voltage (kVp). The exposure parameters of the EzRay M device (KVp, mA, s) are similar to the predicate Nomad MD. For the energy source, battery pack, the subject device receives power from the built-in rechargeable Li-ion polymer battery pack whereas the predicate device (K140723) uses Rechargeable NiCd battery pack. Rechargeable Li-ion polymer battery pack has been tested and is in conformity with the standard IEC 62133. Other differences include device design such as exposure time, size and user interface.

10. Non clinical Testing

Testing was performed successfully according to the following standards:

  • IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012 ●
  • IEC 60601-1-2:2014 ●
  • IEC 60601-1-3:2008 (Second Edition) + A1:2013
  • IEC 60601-2-28:2017
  • IEC 60601-2-54:2009, AMD1:2015, AMD2:2018
  • IEC 62133:2012

Furthermore, the following Specific Guidance Document was utilized in the device development to ensure the safety of this device for both the operators and patients:

"The Content of Premarket Submissions for Software Contained in Medical Devices" "Radiation Safety Consideration for X-ray Equipment Designed for Hand-Held Use"

The device also conforms to the following:

21 CFR 1020 Subchapter J: Performance Standards for lonizing Radiation Emitting Products

  • 21 CFR 1020.30: Diagnostic x-ray system and their major components
    21 CFR 1020.31: Radiographic Equipment

11. Conclusions

In reference to the comparison provide in Substantial Equivalence Chart, the subject device and the predicate device have little difference with its technological features. As demonstrated in the performance bench testing, X-ray performance and X-ray Safety of the subject device was tested in accordance with Federal standard 21CFR Part 1020.30 and 31 as well as international standards such as IEC 60601-2-54. Quality assurance procedures are adhered to, and the specifications and functional requirements were met as the test results indicated.

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. VATECH Co., Ltd. concludes that EzRay M (Model: VMX-P300) is substantially equivalent to predicate device as described herein.

§ 892.1720 Mobile x-ray system.

(a)
Identification. A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.