K Number
K203647

Validate with FDA (Live)

Date Cleared
2021-12-22

(373 days)

Product Code
Regulation Number
862.3610
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For In Vitro Diagnostic Use.

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay is an enzyme immunoassay with a cutoff of 50 ng/mL in neat oral fluid collected by Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of methamphetamine in human oral fluid with clinical analyzers. This assay is calibrated against d-methamphetamine.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

Device Description

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay is an in-vitro test to detect the presence of methamphetamine in human oral fluid samples collected by Quantisal or Quantisal II Oral Fluid Collection Device.

AI/ML Overview

The provided document describes the performance characteristics of the Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay for the detection of methamphetamine in human oral fluid. This is a medical device, and the data presented supports its substantial equivalence to a legally marketed predicate device (LZI Oral Fluid Methamphetamine Enzyme Immunoassay [K131652]).

The document details various studies, primarily focusing on analytical performance rather than diagnostic accuracy involving human subjects. Therefore, many of the typical acceptance criteria and study aspects for AI-powered diagnostic devices (e.g., expert consensus for ground truth, MRMC studies, human-in-the-loop performance) are not applicable here. This device performs a chemical analysis.

Here's an analysis based on the provided text, addressing the relevant points:


Acceptance Criteria and Reported Device Performance

The acceptance criteria for this type of in-vitro diagnostic device are generally defined by demonstrating analytical performance metrics such as precision, specificity, linearity, and stability, with results falling within acceptable ranges. Substantial equivalence is often shown by comparing these metrics to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

Since this is an in-vitro diagnostic device, the "acceptance criteria" are implied by the expected performance common for such assays, aiming for high agreement with confirmed methods and consistent results. The document states a design goal of ">95% agreement" for method comparison.

Performance CharacteristicAcceptance Criteria (Implied/General for IVDs)Reported Device Performance (SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay)
Precision (Qualitative)Consistent classification (Negative/Positive) at specific concentrations, especially around the cutoff.Quantisal:- 0-37.5 ng/mL: 100% Negative (60/60)- 50 ng/mL (Cutoff): 26 Neg/34 Pos- 62.5-100 ng/mL: 100% Positive (60/60)Quantisal II Pad A:- 0-37.5 ng/mL: 100% Negative (60/60)- 50 ng/mL (Cutoff): 34 Neg/26 Pos- 62.5-100 ng/mL: 100% Positive (60/60)Quantisal II Pad B:- 0-37.5 ng/mL: 100% Negative (60/60)- 50 ng/mL (Cutoff): 31 Neg/29 Pos- 62.5-100 ng/mL: 100% Positive (60/60)
Precision (Semi-Quantitative)Mean concentration values close to expected spiked concentrations, and consistent classification.Quantisal: Mean concentrations generally close to spiked values (e.g., 50 ng/mL mean 50.2 ng/mL). Classification performance similar to qualitative.Quantisal II Pad A: Mean concentrations generally close to spiked values (e.g., 50 ng/mL mean 48.6 ng/mL). Classification performance similar to qualitative.Quantisal II Pad B: Mean concentrations generally close to spiked values (e.g., 50 ng/mL mean 49.0 ng/mL). Classification performance similar to qualitative.
Specificity/Cross-ReactivityMinimal to no cross-reactivity with structurally unrelated compounds; expected cross-reactivity with structurally similar compounds.- Structurally Similar: Varies. High cross-reactivity with MDMA (90.9%), (±)-3,4-Methylenedioxyethylamphetamine (45.5%), PMMA (180.2%), and some with d,l-Methamphetamine (45.5%), l-Ephedrine (1.2%), Fenfluramine (1.0%), MDA (0.8%), Methylone (0.2%), PMA (1.5%), d-Pseudoephedrine (0.3%), l-Pseudoephedrine (0.1%), d-Amphetamine (0.6%), l-Methamphetamine (0.7%).- Structurally Unrelated: No interference observed at tested high concentrations (Table 9 shows a wide range of compounds tested up to 40,000 ng/mL).
Interference (Endogenous/Exogenous)No interference from common endogenous or exogenous substances.No interference observed for a wide range of endogenous (e.g., Albumin, Bilirubin, Hemoglobin, Salivary-alpha-amylase) and exogenous (e.g., Acetylsalicylic Acid, Caffeine, Alcohol, Mouthwash, Toothpaste) compounds at tested levels.
Interference (pH)Performance maintained across physiological pH range.No interference observed across pH 3.0-11.0.
Linearity/RecoveryRecovery percentage within an acceptable range (e.g., 90-110%) across the linear range.Linear range confirmed for 20-200 ng/mL. Recovery percentages: Quantisal (93.9-107.9%), Quantisal II "A" (96.8-104.0%), Quantisal II "B" (96.2-109.2%).
Methamphetamine StabilityStability of analyte in collected samples over time under specified storage conditions.Oral fluid samples stable for up to 12 months at 2°C - 8°C. Data for 10-day stability at ambient temperature (8°C - 25°C) referenced in previous submissions (K183048, K200801).
Calibration DurationConsistent performance over defined calibration interval.Achieved acceptance criteria up to 10 days. Recommended frequency: 7 days.
Method Comparison (Qualitative & Semi-Quantitative)Agreement with confirmatory method (LC-MS/MS) greater than 95%.Achieved 100% agreement (40/40) for both positive and negative results when compared to LC-MS/MS across all three collection devices (Quantisal, Quantisal II "A", Quantisal II "B"). The total samples tested were 80, partitioned into 40 positive and 40 negative categories by LC-MS/MS.

2. Sample Size and Data Provenance:

  • Precision Study: 60 determinations for each concentration level, replicating across 15 days, 2 runs/day, 2 collection devices/run (N=60 per conc.). An additional 20-day study on 3 reagent lots for repeatability.
    • Provenance: Not explicitly stated (e.g., country of origin). The study states "Drug free negative oral fluid was spiked," implying a controlled laboratory setting (prospective creation of samples).
  • Specificity and Cross-Reactivity: Compounds spiked into drug-free pooled oral fluid.
    • Provenance: Controlled laboratory (prospective creation of samples).
  • Interference (Structurally Unrelated): Compounds spiked into drug-free oral fluid containing methamphetamine at ±25% of cutoff.
    • Provenance: Controlled laboratory (prospective creation of samples).
  • Interference (Endogenous/Exogenous): Spiking into drug-free oral fluid; "Additional orally used products were tested by collecting oral fluid... from volunteers after use of the substances."
    • Provenance: Mix of controlled laboratory (prospective creation) and possibly prospective volunteer studies.
  • Interference (pH): Spiked samples at various pH levels.
    • Provenance: Controlled laboratory (prospective creation of samples).
  • Linearity/Recovery: Serially diluted spiked samples.
    • Provenance: Controlled laboratory (prospective creation of samples).
  • Methamphetamine Stability: Spiked samples stored over time.
    • Provenance: Controlled laboratory (prospective creation of samples).
  • Calibration Duration: Spiked samples tested over time.
    • Provenance: Controlled laboratory (prospective creation of samples).
  • Method Comparison: 80 deidentified, unaltered clinical oral fluid samples collected by Quantisal II Oral Fluid Collection Devices.
    • Provenance: "Obtained from clinical research facilities," suggesting real-world clinical samples, likely retrospective. Country of origin not specified.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is an in-vitro diagnostic (IVD) device for chemical analysis, not an imaging AI or clinical decision support system that relies on human expert interpretation of complex clinical data. Therefore, the concept of "experts" establishing ground truth in the traditional sense (e.g., radiologists, pathologists) is not directly applicable.

For the analytical studies performed, the "ground truth" is established very precisely through:

  • Spiking concentrations: Known amounts of methamphetamine or other compounds are added to drug-free oral fluid. The exact concentration is the ground truth.
  • LC-MS/MS (Liquid Chromatography-Tandem Mass Spectrometry): This is a highly accurate and widely accepted gold-standard method for confirming drug presence and concentration in biological samples. The results from LC-MS/MS served as the "ground truth" for the method comparison study. The laboratory performing this analysis would follow strict protocols and be staffed by trained analytical chemists or toxicologists, but no "expert consensus" process among multiple human interpreters is involved as it would be for an image-based diagnosis.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable. Ground truth for an IVD drug test is established by precise chemical methods (spiking, LC-MS/MS), not through subjective human interpretation requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is an automated enzyme immunoassay (chemical test), not an AI-powered diagnostic assist tool for human readers/interpreters. There are no human "readers" in the loop.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

The device (assay performed on a clinical analyzer) functions as a standalone test. Its performance is reported solely based on its analytical output against the chemically defined ground truth (spiked concentrations, LC-MS/MS).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth was established through:

  • Known Spiked Concentrations: For precision, specificity, interference, linearity, stability, and calibration studies.
  • Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS): For the method comparison study, LC-MS/MS results served as the definitive ground truth for "deidentified, unaltered clinical oral fluid samples".

8. The sample size for the training set:

Not applicable. This device is an enzyme immunoassay, a biochemical assay, not a machine learning/AI model that requires a "training set" in the computational sense. Its "training" is inherent in the chemical reactions and calibration curves established by the manufacturer, validated through the performance studies described.

9. How the ground truth for the training set was established:

Not applicable, as there is no "training set" in the context of an AI/ML algorithm. The assay's analytical characteristics are determined through standard laboratory validation practices using materials with known characteristics (e.g., calibrated standards, spiked samples).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

December 22, 2021

Immunalysis Corporation Elina Arroyo Manager Regulatory Affairs 829 Towne Center Drive Pomona, California 91767

Re: K203647

Trade/Device Name: SEFRIATM Methamphetamine Oral Fluid Enzyme Immunoassay Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine Test System Regulatory Class: Class II Product Code: LAF Dated: October 7, 2021 Received: October 8, 2021

Dear Elina Arroyo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Marianela Perez-Torres, Ph.D. Deputy Director Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiologial Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203647

Device Name

SEFRIA™ Methamphetamine Oral Fluid Enzyme Immunoassay

Indications for Use (Describe) For In Vitro Diagnostic Use.

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay is an enzyme immunoassay with a cutoff of 50 ng/mL in neat oral fluid collected by Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of methamphetamine in human oral fluid with clinical analyzers. This assay is calibrated against d-methamphetamine.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red arrow. The arrow is pointing to the right. The word is written in all capital letters and is centered on the arrow.

510(K) SUMMARY

510(k) Number: K203647

A. GENERAL INFORMATION

Applicant Name:Immunalysis Corporation829 Towne Center DrivePomona, CA 91767Establishment # 2020952
Company Contact:Elina Arroyo, Manager Regulatory AffairsImmunalysis Corporation

829 Towne Center Drive, Pomona, CA 91767 USA (224) 361-7080 elina.arroyo@aabbott.com

October 7, 2021 Date Prepared:

B. DEVICE IDENTIFICATION

Trade or Proprietary Names: SEFRIA™ Methamphetamine Oral Fluid Enzyme Immunoassay

Common Name: Methamphetamine Oral Fluid Enzyme Immunoassay

C. REGULATORY INFORMATION

Device Classification Name: Gas Chromatography, Methamphetamine
Product Codes:LAF
Regulatory Class:Class II
Classification Regulation:862.3610
Panel:Toxicology (91)
Predicate Device:LZI Oral Fluid Methamphetamine Enzyme Immunoassay, LZI Oral Fluid

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, arrow-shaped background. The word is written in all capital letters and is centered on the arrow. The arrow is pointing to the right.

Methamphetamine Calibrators, LZI Oral Fluid Methamphetamine

Controls [K131652]

D. DEVICE DESCRIPTION

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay is an in-vitro test to detect the presence of methamphetamine in human oral fluid samples collected by Quantisal or Quantisal II Oral Fluid Collection Device.

Methamphetamine is a stimulant drug usually used as a white, bitter-tasting powder or a pill. It is highly addictive and is rarely prescribed. When smoked, the vapor moves from the lungs to the bloodstream causing euphoria, increased energy and alertness, higher confidence, and motivation. Side effects include paranoia, tremors, loss of appetite, mood swings, and neuropsychological effects including deficits in episodic memory, executive functions, information processing speed, motor skills, language, and visuoconstructional abilities. [1] The disposition of methamphetamine in oral fluid has been well documented showing the main drug found after intake is the parent compound. Following single dosing of 10 mg and 20 mg, concentrations over 50 ng/mL of methamphetamine were detected in oral fluid.121 Oral fluid has been shown to be a useful matrix for the analysis of methamphetamine in several areas including workplace drug testing, pain management compliance monitoring, and driving under the influence of drugs. [3],[4],[5]

E. INTENDED USE

For In Vitro Diagnostic Use.

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 50 ng/mL in neat oral fluid collected by Quantisal II Oral Fluid Collection Device. The assay is intended for the qualitative and semi-quantitative analysis of methamphetamine in human oral fluid with clinical analyzers. This assay is calibrated against dmethamphetamine.

The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) or permitting laboratories to establish quality control procedures.

The Immunalysis SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to

4 Cobb Scott J, Woods SP, Matt GE, Meyer RA, Heaton RK, Atkinson JH, Grant I. Neurocognitive effects of methamphetamine: A critical review and meta-analysis. Neuropsych Review. 2007;17(3):275-297.

2 Huestis MA、Cone EJ. Methamphetamine disposition in oral fluid, plasma and urine. Ann NY Acad Sci. 2007:1098:104-121.

3 Moore C. Oral fluid and hair in workplace drug testing programs: new technology for immunoassays. Drug Test Anal. 2011;3(3):166-168.

4 Moore C. Drug testing and adherence monitoring in pain management: Oral fluid testing. J Opioid Manage. 2015;11(1):69-75.

5 Veitenheimer AM, Wagner JR. Evaluation of oral fluid as a specimen for DUID. J Anal Toxicol. 2017;41(6):517-522.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is written in all capital letters. The background is a red arrow shape.

obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any test result, particularly when preliminary positive results are used.

F. COMPARISON WITH PREDICATE

AttributeCandidate DeviceSEFRIA Methamphetamine OralFluid Enzyme ImmunoassayPredicate DeviceLZI Oral FluidMethamphetamine EnzymeImmunoassay [K131652]
Similarities
Test PrincipleIdenticalHomogeneous enzymeimmunoassay
CalibratedAgainstIdenticald-methamphetamine
Assay MaterialsIdenticalantibody reagent, drug conjugatereagent
Cutoff LevelIdentical50 ng/mL
UserEnvironmentIdenticalFor use in laboratories
Sample MatrixIdenticalHuman oral fluid
Reagent StorageIdentical2-8°C until expiration date
InstrumentationIdenticalAutomated clinical chemistryanalyzer
MassSpectrometryConfirmationIdenticalRequired for preliminary positiveanalytical results
Differences
Intended UseQualitative and semi-quantitativeanalysis of methamphetamine inhuman oral fluid collected byQuantisal or Quantisal II Oral FluidCollection DeviceQualitative and semi-quantitativedetermination of d-methamphetamine in neat humanoral fluid collected into the LZIOral Fluid Collector
SampleCollectionDeviceOral fluid is collected with theQuantisal or Quantisal II Oral FluidCollection Device. Sample is storedin a plastic tube containingpreservative buffer with snap cap.Oral fluid is collected with the LZIOral Fluid Collector. Sample isstored in a plastic tube.

G. PERFORMANCE CHARACTERISTICS

The following laboratory performance studies were performed to determine substantial equivalence of the SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay to the predicate device. Assay performance was established using the Beckman Coulter AU480 chemistry analyzer.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image is a red arrow pointing to the right. Inside the arrow, the word "IMMUNALYSIS" is written in white, bold letters. The arrow and the text are the only elements in the image.

1. Precision

Precision study was performed over 15 days, 2 runs per day with 2 collection devices per run (N=60), one replicate per collection device on 1 lot of reagent and 1 lot of Quantisal and 1 lot of Quantisal II oral fluid collection devices. Drug free negative oral fluid was spiked to concentrations of assay cutoff and ±25%, ±50%, ±75%, ±100% of the cutoff and was collected using the collection devices. The spiked concentrations were confirmed by mass spectrometry (LC-MS/MS) before collection. The study established the repeatability of the testing system. including assay and oral fluid collection device. Test results in qualitative and semi-quantitative modes are presented in Tables 1 to 6.

An additional 20-day study was performed on 3 lots of assay reagent to demonstrate the repeatability across multiple reagent lots.

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%6060 Negative
12.5-75%6060 Negative
25-50%6060 Negative
37.5-25%6060 Negative
50Cutoff6026 Neg/34 Pos
62.5+25%6060 Positive
75+50%6060 Positive
87.5+75%6060 Positive
100+100%6060 Positive

Table 1. Precision - Qualitative - Quantisal

Table 2. Precision - Semi-Quantitative - Quantisal
--------------------------------------------------------
Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%601.160 Negative
12.5-75%6014.260 Negative
25-50%6025.360 Negative
37.5-25%6039.360 Negative
50Cutoff6050.237 Neg/23 Pos
62.5+25%6068.960 Positive
75+50%6078.960 Positive
87.5+75%6093.860 Positive
100+100%60112.260 Positive

Table 3. Precision - Qualitative - Quantisal II Pad A

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%6060 Negative

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is in the shape of a stretched hexagon. The text is centered in the hexagon.

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
12.5-75%6060 Negative
25-50%6060 Negative
37.5-25%6060 Negative
50Cutoff6034 Neg/26 Pos
62.5+25%6060 Positive
75+50%6060 Positive
87.5+75%6060 Positive
100+100%6060 Positive

Table 4. Precision - Semi-Quantitative – Quantisal II Pad A

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%602.460 Negative
12.5-75%6014.760 Negative
25-50%6025.160 Negative
37.5-25%6038.460 Negative
50Cutoff6048.644 Neg/16 Pos
62.5+25%6063.860 Positive
75+50%6074.960 Positive
87.5+75%6086.860 Positive
100+100%60109.360 Positive

Table 5. Precision - Qualitative - Quantisal II Pad B

Concentration(ng/mL)% of Cutoff# of DeterminationsResult
0-100%6060 Negative
12.5-75%6060 Negative
25-50%6060 Negative
37.5-25%6060 Negative
50Cutoff6031 Neg/29 Pos
62.5+25%6060 Positive
75+50%6060 Positive
87.5+75%6060 Positive
100+100%6060 Positive

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, hexagon-shaped background. The word is written in all capital letters and is centered on the background. The red background is a bright, saturated color. The font is sans-serif and bold.

Concentration(ng/mL)% ofCutoff# ofDeterminationsMean Conc.(ng/mL)Result
0-100%603.860 Negative
12.5-75%6014.760 Negative
25-50%6024.960 Negative
37.5-25%6040.460 Negative
50Cutoff6049.042 Neg/18 Pos
62.5+25%6065.860 Positive
75+50%6078.360 Positive
87.5+75%6089.060 Positive
100+100%60111.960 Positive

Table 6. Precision - Semi-Quantitative - Quantisal II Pad B

2. Specificity and Cross-Reactivity

Structurally and functionally similar compounds were spiked into drug free pooled oral fluid at levels that will yield a result that is equivalent to the cutoff, if cross reacting. The study verified the cross reactivity of the methamphetamine assay to related drugs and drug metabolites, in both the qualitative and semi-quantitative modes. Cross-reactivity test results in qualitative mode are presented in Table 7. Cross-reactivity test results in semi-quantitative mode are presented in Table 8.

Table 7. Cross-Reactivitv – Qualitative
---------------------------------------------
CompoundCompoundConc. (ng/mL)MethamphetamineEquivalent Conc.(ng/mL)ResultCross-Reactivity(%)
l-Methamphetamine7,50050POS0.7
d,l-Methamphetamine11050POS45.5
d-Amphetamine8,00050POS0.6
l-Amphetamine40,000<50NEG<0.1
Diphenhydramine40,000<50NEG<0.1
Doxylamine40,000<50NEG<0.1
d-Ephedrine40,000<50NEG<0.1
l-Ephedrine4,25050POS1.2
Fenfluramine5,00050POS1.0
Methylenedioxymethamphetamine(MDMA)5550POS90.9
(±)-3,4-Methylenedioxyethylamphetamine(MDEA)11050POS45.5
Methylenedioxyamphetamine(MDA)6,00050POS0.8
Methylone27,50050POS0.2
4-Methoxyamphetamine (PMA)3,25050POS1.5
Para-Methoxy-N-Methylamphetamine (PMMA)27.7550POS180.2

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The background is shaped like an arrow pointing to the right. The word is written in all capital letters and is centered on the arrow.

CompoundCompoundConc. (ng/mL)MethamphetamineEquivalent Conc.(ng/mL)ResultCross-Reactivity(%)
Phenethylamine40,000<50NEG<0.1
Phenylephrine40,000<50POS<0.1
Phentermine40,000<50POS<0.1
Phenylpropanolamine (PPA)40,000<50POS<0.1
d-Pseudoephedrine15,00050POS0.3
l-Pseudoephedrine40,00050NEG0.1*
Tyramine40,000<50NEG<0.1

*As the Semi-Quantitative result is positive, the % cross reactivity is calculated based on Semi-Quantitative result.

CompoundCompoundConc.(ng/mL)MethamphetamineEquivalent Conc.(ng/mL)MeanValue(ng/mL)ResultCross-Reactivity(%)
l-Methamphetamine7,5005054.4POS0.7
d,l-Methamphetamine1105054.1POS45.5
d-Amphetamine8,0005051.2POS0.6
l-Amphetamine40,000<5028.9NEG<0.1
Diphenhydramine40,000<502.8NEG<0.1
Doxylamine40,000<502.3NEG<0.1
d-Ephedrine40,000<5022.9NEG<0.1
l-Ephedrine4,2505050.0POS1.2
Fenfluramine5,0005054.3POS1.0
Methylenedioxymethamphetamine (MDMA)555051.3POS90.9
(±)-3,4-Methylenedioxyethylamphetamine (MDEA)1105050.0POS45.5
Methylenedioxyamphetamine (MDA)6,0005053.7POS0.8
Methylone27,5005052.2POS0.2
Methoxyamphetamine(PMA)3,2505053.6POS1.5
Para-Methoxy-N-Methylamphetamine(PMMA)27.755051.5POS180.2
Phenethylamine40,000<5014.1NEG<0.1
Phenylephrine40,000<5014.8POS<0.1
Phentermine40,000<5035.2POS<0.1
Phenylpropanolamine(PPA)40,000<5015.9POS<0.1
d-Pseudoephedrine15,0005052.5POS0.3
l-Pseudoephedrine40,0005050.0POS0.1
Tyramine40,000<5015.4NEG<0.1

Table 8. Cross-Reactivity - Semi-Quantitative

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is written in all capital letters. The background is a solid red color. The image is simple and easy to read.

3. Interference - Structurally Unrelated Compounds

Structurally unrelated compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free oral fluid containing methamphetamine at ±25% of the cutoff. At the levels tested, there was no interference with structurally unrelated compounds. The concentration levels of structurally unrelated compounds are presented in Table 9.

CompoundConc. Tested (ng/mL)
4-Bromo-2,5,Dimethoxyphenethylamine40,000
Acetaminophen40,000
6-Acetylcodeine40,000
6-Acetylmorphine40,000
Alprazolam40,000
7-Aminoclonazepam40,000
7-Aminoflunitrazepam40,000
7-Aminonitrazepam40,000
Amitriptyline40,000
Amobarbital40,000
Benzylpiperazine5,000
Bromazepam40,000
Buprenorphine40,000
Bupropion40,000
Butabarbital40,000
Butalbital40,000
Cannabidiol40,000
Cannabinol40,000
Carbamazepine40,000
Carisoprodol40,000
Chlordiazepoxide40,000
Chlorpromazine40,000
cis-Tramadol40,000
Clobazam40,000
Clomipramine40,000
Clonazepam40,000
Cocaine40,000
Clozapine40,000
Codeine40,000
Cotinine40,000
Cyclobenzaprine40,000
Demoxepam40,000
CompoundConc. Tested (ng/mL)
Desakylflurazepam40,000
Desipramine15,000
Desomorphine40,000
Dextromethorphan40,000
Dihydrocodeine40,000
Diazepam40,000
Digoxin40,000
Dehydronorketamine40,000
Delta-9-THC40,000
Doxepin35,000
Ecgonine40,000
Ecgonine Methyl Ester40,000
EDDP40,000
EMDP40,000
Ethyl-β-D-Glucuronide40,000
Ethylmorphine40,000
Fentanyl40,000
Flunitrazepam40,000
Fluoxetine15,000
Flurazepam40,000
Haloperidol40,000
Heroin40,000
Hydrocodone40,000
Hydromorphone40,000
11-hydroxy-delta-9-THC40,000
Imipramine40,000
Ketamine40,000
Lamotrigine40,000
Levorphanol40,000
Lidocaine40,000
Lorazepam40,000
Lorazepam Glucuronide40,000
Lormetazepam40,000
LSD40,000
Maprotiline40,000
Meperidine40,000
Meprobamate40,000
Methadone40,000
Methaqualone40,000
CompoundConc. Tested (ng/mL)
Methoxetamine40,000
Methylphenidate40,000
Midazolam40,000
Morphine40,000
Morphine-3-Glucuronide40,000
Morphine-6-Glucuronide40,000
N-desmethyl tapentadol40,000
N-desmethyl tramadol40,000
N-desmethyl venlafaxine40,000
Nalorphine40,000
Naloxone40,000
Naltrexone40,000
Naproxen40,000
Nitrazepam40,000
11-nor-9 carboxy THC40,000
Norbuprenorphine40,000
Norcodeine40,000
Nordiazepam40,000
Norketamine40,000
Normorphine40,000
Noroxycodone40,000
Noroxymorphone40,000
Norpropoxyphene40,000
Norpseudoephedrine40,000
Nortriptyline20,000
O-desmethyl tramadol40,000
O-desmethyl venlafaxine40,000
Oxycodone40,000
Oxymorphone40,000
Oxycodone-3-β-Glucuronide40,000
Olanzapine40,000
Oxazepam40,000
PCP40,000
Pentazocine40,000
Pentobarbital40,000
Phenobarbital40,000
Phenytoin40,000
Prazepam40,000
Propranolol40,000
CompoundConc. Tested (ng/mL)
Propoxyphene40,000
Protriptyline20,000
Ritalinic Acid40,000
Salicylic Acid40,000
Secobarbital40,000
Sertraline40,000
Sufentanil40,000
Tapentadol40,000
Temazepam40,000
Theophylline40,000
Thioridazine40,000
Trazadone40,000
Triazolam40,000
3-Trifluoromethylphenyl-piperazine20,000
Trimipramine20,000
Venlafaxine40,000
Verapamil30,000
Zolpidem40,000

Table 9. Non-Interfering Structurally Unrelated Compounds

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red hexagon. The letters are bold and sans-serif. In the bottom right corner of the hexagon, there is a small "TM" symbol.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red hexagon. The letters are all capitalized and in a sans-serif font. The letters are evenly spaced and easy to read. In the bottom right corner of the image, there is a small "TM" symbol.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, horizontally-oriented hexagon. The word is written in all capital letters. The hexagon is a bright red color.

4. Interference - Endogenous Compounds and Exogenous Compounds

Endogenous compounds and exogenous compounds were evaluated in qualitative and semi-quantitative modes by spiking the potential interferent into drug free oral fluid containing methamphetamine at ±25% of the cutoff. Additional orally used products were tested by collecting oral fluid using Quantisal and Quantisal II Oral Fluid Collection Devices from volunteers after use of the substances. At the levels tested, there was no interference observed with endogenous compounds, exogenous compounds and orally used compounds. Endogenous compounds and exogenous compounds are presented in Tables 10 and 11. Orally used compounds tested are presented in Table 12.

CompoundConcentration Tested
Ascorbic Acid2 mg/mL
Bilirubin0.15 mg/mL
Cholesterol0.45 mg/mL
γ-Globulin0.8 mg/mL
Hemoglobin2 mg/mL
Human Serum Albumin15 mg/mL
IgA1 mg/mL
IgG1 mg/mL
IgM0.5 mg/mL
Salivary-α-amylase1000 U/mL

Table 10. Non-interfering Endogenous Compounds

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, horizontally-oriented hexagon. The word is written in all capital letters. The hexagon is a bright red color, and the white letters stand out against the red background.

CompoundConcentration Tested
Acetylsalicylic Acid0.01 mg/mL
Baking Soda0.6% v/v
Denture Adhesive0.6% w/v
Ibuprofen0.01 mg/mL
Alcohol (Ethanol)6% v/v
Caffeine0.01 mg/mL
Coffee6% v/v
Cranberry Juice6% v/v
Milk1% v/v
Mouthwash6% v/v
Naproxen0.01 mg/mL
Orange Juice2% v/v
Soft Drink (Pepsi)6% v/v
Sodium Chloride18 mg/mL
Sugar10 mg/mL
Tea6% v/v
Toothpaste6% w/v

Table 11. Non-interfering Exogenous Compounds

Table 12. Non-interfering Orally Used Exogenous Products

CompoundConcentration Tested
Teeth Whitener2 strips
Cigarette1 cigarette
Hard Candy1 piece
Chewing Gum1 piece
Hydrogen Peroxide (3% OTC)Neat (2 min. mouth rinse)
Sugar2 Teaspoons
Cough Syrup2 Teaspoons
Milk100 mL
Orange Juice100 mL
Ibuprofen200 mg
Acetaminophen1000 mg

5. Interference - pH

To evaluate potential interference from the effect of oral fluid pH, device performance in the qualitative and semi-quantitative modes was tested using a range of oral fluid pH values (3.0, 4.0, 5.0, 8.0, 9.0, 10.0 and 11.0). All test samples were prepared in drug free oral fluid containing methamphetamine at ±25% of the cutoff. At the pH levels tested, there was no interference observed for each test mode.

6. Linearity/Recovery

Assay linearity was evaluated in the semi-quantitative mode by spiking a drug free oral fluid pool with a high concentration of methamphetamine. Additional pools were made by serially diluting the high concentration specimen with drug free oral fluid to achieve concentrations ranging from 20 ng/mL to

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red background. The word is centered in the image and is written in all capital letters. The background is a solid red color, and the letters are a bright white color. The image is simple and easy to read.

220 ng/mL. The 0 ng/mL specimen was made from drug free oral fluid. Each pool was collected by Quantisal and Quantisal II oral fluid collection devices and tested in triplicate to calculate the mean concentration values that were used to calculate drug recovery. Linearity test results in semi-quantitative mode are presented in Tables 13 to 15. The study confirmed the linear range to be 20-200 ng/mL with a drug recovery percentage of 93.9% to 109.2% across the collection devices.

Expected ConcentrationMean ConcentrationRecovery (%)
(ng/mL)(ng/mL)
01.0N/A
2020.7103.3
4043.1107.8
5051.9103.8
6064.8107.9
8081.2101.5
10096.896.8
120127.7106.4
140141.4101.0
160150.393.9
180184.6102.6
200194.897.4
220225.3102.4

Table 13. Linearity/Recovery - Quantisal

Table 14, Linearity/Recoverv – Quantisal II "A"
---------------------------------------------------
Expected Concentration(ng/mL)Mean Concentration(ng/mL)Recovery (%)
0-1.4N/A
2020.5102.5
4041.6103.9
5048.496.8
6059.398.9
8078.297.8
10097.397.3
120123.2102.6
140145.6104.0
160159.099.4
180180.0100.0
200201.0100.5
220219.9100.0

Table 15. Linearity/Recovery - Quantisal II "B"

Expected Concentration(ng/mL)Mean Concentration(ng/mL)Recovery (%)
00.5N/A
2021.8109.2
4042.9107.3
5049.198.3

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, horizontally-oriented hexagon. The word is written in all capital letters. The hexagon is a bright red color.

Expected Concentration(ng/mL)Mean Concentration(ng/mL)Recovery (%)
6060.9101.5
8078.297.8
10094.894.8
120119.399.4
140140.7100.5
160163.6102.2
180176.398.0
200192.496.2
220214.597.5

7. Methamphetamine Stability in Oral Fluid

Drug free negative oral fluid spiked with methamphetamine at +50% of the 50 ng/mL cutoff were collected and stored in Ouantisal II Oral Fluid Collection Devices at 2°C - 8°C. tested by LC-MS/MS at each time point and compared to the baseline concentration result. The test results indicate that oral fluid samples containing methamphetamine are stable for up to 12 months stored in Quantisal or Quantisal II Oral Fluid Collection Device at 2°C - 8°C.

Data to support 10-day storage in Quantisal II Oral Fluid Collection Device at ambient temperature 8°C - 25°C were reported in K183048 and K200801.

8. Calibration Duration

Drug free negative oral fluid spiked with methamphetamine at ±25% of the cutoff were tested in qualitative mode at time points up to 14 days and in semi-quantitative at time points up to 10 days. At the initial time point, a two-point calibration curve was established in qualitative mode and multi-points calibration curve was established in semi-quantitative mode. These calibrations were used through the duration of the study. The test results met acceptance criteria up to 10 days. The recommended frequency of calibration is 7 days.

9. Method Comparison

80 deidentified, unaltered clinical oral fluid samples collected by Quantisal II Oral Fluid Collection Devices were obtained from clinical research facilities, analyzed for methamphetamine at assay cutoff with the SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay in both qualitative and semi-quantitative modes and compared to Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) results. The instruments used were the Beckman Coulter AU480 chemistry analyzer and an Agilent 6430 Liquid Chromatography-Tandem Mass Spectrometry. The data demonstrate that the design goal of greater than 95% agreement was achieved. Method comparison test results in qualitative and semi-quantitative modes are presented from Tables 16 to 18.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the word "IMMUNALYSIS" in white letters on a red, hexagon-shaped background. The word is centered in the hexagon. The hexagon is oriented horizontally.

ImmunoassayResultLC-MS/MS Methamphetamine Concentration
< 25 ng/mL(less than-50%cutoff)25 – 49 ng/mL(between -50%cutoff andcutoff)50 – 75 ng/mL(betweencutoff and+50% cutoff)> 75 ng/mL(greaterthan +50%cutoff)Agreement (%)
Qual.Positive00436100% (40/40)
Negative36400100% (40/40)
Semi-Quant.Positive00436100% (40/40)
Negative36400100% (40/40)

Table 16. Method Comparison – Quantisal

Table 17. Method Comparison – Quantisal II "A"

ImmunoassayResultLC-MS/MS Methamphetamine Concentration
< 25 ng/mL(less than-50%cutoff)25 – 49 ng/mL(between -50%cutoff andcutoff)50 – 75 ng/mL(betweencutoff and+50% cutoff)> 75 ng/mL(greaterthan +50%cutoff)Agreement (%)
Qual.Positive00535100% (40/40)
Qual.Negative36400100% (40/40)
Semi-Quant.Positive00535100% (40/40)
Semi-Quant.Negative36400100% (40/40)

Table 18. Method Comparison - Quantisal II "B"

ImmunoassayResultLC-MS/MS Methamphetamine ConcentrationAgreement (%)
< 25 ng/mL(less than-50%cutoff)25 – 49 ng/mL(between -50%cutoff andcutoff)50 – 75 ng/mL(betweencutoff and+50% cutoff)> 75 ng/mL(greaterthan +50%cutoff)
Qual.Positive00535100% (40/40)
Negative36400100% (40/40)
Semi-Quant.Positive00535100% (40/40)
Negative36400100% (40/40)

H. CONCLUSION

The information provided in this pre-market notification demonstrates that the SEFRIA Methamphetamine Oral Fluid Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its intended use.

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).