(667 days)
The device is intended to measure, save, and view continuous electrocardiogram (ECG) information for long-term recording (up to 14 days) from ambulatory patients by attaching to the skin surface. ECG records are stored in the the device for review after the recording period (up to 14 days) is completed, and are not intended for for real-time monitoring. The device does not include automated ECG analysis functions, and the recorded ECGs are not intended for automated analysis. The device allows patient and clinicians to view the ECG signal recorded in real-time solely for the purpose of visual assessment of the recording quality; the ECG signal displayed in real time is not intended for any clinical or diagnostic use. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, fatigue, or anxiety.
AT-Patch consists of 1) AT-Patch device, 2) Software for ECG viewing (ATR-C130), 3) App (ATN-C130) that communicates with the device and allows the patient to view the ECG waveform for visual assessment of the recording quality, and included parts (Dedicated Cable, BLE Dongle and USB memory). AT-Patch an ambulatory ECG recorder, designed to continuously acquire and store patient's ECG for up to 14 days. The AT-Patch can connect to the device applicaitn (ATN-C130) torecord symptoms with the App during operation and to check the ECG recording. The device is not intended for real-time monitoring.
The provided text does not contain detailed acceptance criteria for the device (AT-Patch) based on specific performance metrics, nor does it describe a study specifically designed to prove that the device meets such criteria. Instead, it focuses on demonstrating substantial equivalence to a predicate device (Zio SkyRunner (SR) ECG Monitoring service, K143513) through various tests and comparisons.
The document discusses performance testing in a general sense, confirming compliance with several standards, but it does not provide a table of acceptance criteria with reported device performance results.
Here's an analysis of what can be extracted from the provided text and what is missing based on your request:
1. A table of acceptance criteria and the reported device performance:
- Not provided in the text. The document lists various standards (e.g., IEC 60601-1, ANSI/AAMI EC12) that the device was tested against and states that "The test results met the electrical safety and EMC requirements" and "The Test results met the performance requirements." However, it does not specify quantitative acceptance criteria or the measured performance values for specific device functions (e.g., accuracy of ECG measurement, signal-to-noise ratio).
2. Sample size used for the test set and the data provenance:
- Not provided in the text. The document mentions various performance and safety tests but does not specify the sample size for any of these tests or the origin of any data used (e.g., number of patients, retrospective/prospective study design, country of origin).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided in the text. The AT-Patch device "does not include automated ECG analysis functions, and the recorded ECGs are not intended for automated analysis." Its primary function is to measure, save, and view ECG data for manual review. Therefore, there's no mention of establishing ground truth for automated detection algorithms, and consequently, no information about experts for this purpose.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not provided in the text. As there is no automated analysis requiring expert adjudication to establish ground truth for algorithm performance, this information is not present.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, this was not done. The device "does not include automated ECG analysis functions" and "is not intended for automated analysis." Therefore, a comparative effectiveness study involving AI assistance for human readers is not relevant to this device's stated capabilities and was not performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No, this was not done. The device explicitly states that "The device does not include automated ECG analysis functions, and the recorded ECGs are not intended for automated analysis." Thus, there is no standalone algorithm performance to evaluate.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable/Not provided in the text. Since the device does not perform automated analysis, the concept of "ground truth" for algorithm performance evaluation, as it pertains to AI/CADe devices, is not relevant here. The tests performed are related to the physical and electrical performance of the ECG recorder itself (e.g., battery life, adhesion, electrical safety, EMC, biocompatibility).
8. The sample size for the training set:
- Not applicable/Not provided in the text. As there is no automated analysis or AI component, there is no training set mentioned or required.
9. How the ground truth for the training set was established:
- Not applicable/Not provided in the text. For the same reasons as above, this information is not relevant.
Summary of available information regarding criteria and studies:
The document focuses on:
- Safety and Performance Standards Compliance: The device underwent tests for electrical safety (IEC 60601-1, IEC 60601-2-47, IEC 60601-1-2, IEC 60601-1-11), biocompatibility (ISO 10993 series), software validation (IEC 62304), usability (IEC 62366-1), cybersecurity, battery life, patch adhesion persistence, and ECG electrode performance (ANSI/AAMI EC12:2000 (R2015)).
- Verification of Intended Use Period: "As a result of testing according to the internal criteria, it has been verified that this device can be use for up 14 days and satisfies the warranty period of 1 year." This indicates an internal test for battery life/recording duration.
- QoS (Quality of Service) testing: Performed according to ANSI TIR 69:2017.
- Defibrillation Safety: Tested using an alternative method of IEC 60601-1 Clause 8.5.5.2, demonstrating safety but not being defibrillation-proof.
The key takeaway is that the AT-Patch is an ECG recorder for manual review and does not incorporate automated analysis, which simplifies the regulatory requirements in terms of AI/CADe typical acceptance criteria and study designs for algorithmic performance. The studies and tests performed are primarily related to the physical, electrical, and data recording integrity and safety of the device, rather than the accuracy of an automated diagnostic or detection algorithm.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 12, 2022
ATsens Co., Ltd % Do Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul. Seoul 06210 Korea, South
Re: K203638
Trade/Device Name: AT-Patch Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: DSH Dated: September 13, 2022 Received: October 7, 2022
Dear Do Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K203638
Device Name AT-Patch
Indications for Use (Describe)
The device is intended to measure, save, and view continuous electrocardiogram (ECG) information for long-term recording (up to 14 days) from ambulatory patients by attaching to the skin surface. ECG records are stored in the the device for review after the recording period (up to 14 days) is completed, and are not intended for for real-time monitoring. The device does not include automated ECG analysis functions, and the recorded ECGs are not intended for automated analysis. The device allows patient and clinicians to view the ECG signal recorded in real-time solely for the purpose of visual assessment of the recording quality; the ECG signal displayed in real time is not intended for any clinical or diagnostic use. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, fatigue, or anxiety.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
Preparation Date: October 7, 2022
1. Applicant / Submitter
Applicant/Submitter: ATsens Co.,Ltd/KeonHoon Lee Address: Point Town 803, 11, Gumi-ro, Bundang-gu, Seongnam-si, Gyeonggi-do, 13637 Korea Tel: +82-70-5220-0220 Fax: +82-70-8270-0738 Email: khlee@atsens.com
2. Submission Contact Person
Contact Person: Do Hyun Kim, BT Solutions, Inc. Address: 904, Eonju-ro 86-gil 5, Gangnam-gu, Seoul, 06210, Republic of Korea Tel: +82-2-538-9140 Fax: +82-2-539-9140 Email: ceo@btsolutions.co.kr
3. Device Information
| Trade Name | AT-Patch |
|---|---|
| Common Name | AT-Patch |
| Classification Name | Medical magnetic Tape Recorder |
| Classification Regulation | 21CFR870.2800 |
| Device Class | 2 |
| Product Code | DSH |
4. Predicate Device
| Manufacturer | iRhythm technologies, Inc. |
|---|---|
| Device Name | Zio SkyRunner(SR) ECG Monitoring service |
| 510(k) number | K143513 |
5. Description
{4}------------------------------------------------
AT-Patch consists of 1) AT-Patch device, 2) Software for ECG viewing (ATR-C130), 3) App (ATN-C130) that communicates with the device and allows the patient to view the ECG waveform for visual assessment of the recording quality, and included parts (Dedicated Cable, BLE Dongle and USB memory).
AT-Patch an ambulatory ECG recorder, designed to continuously acquire and store patient's ECG for up to 14 days. The AT-Patch can connect to the device applicaitn (ATN-C130) torecord symptoms with the App during operation and to check the ECG recording. The device is not intended for real-time monitoring.
| ATP-C130 | ||
|---|---|---|
| Item | Description | |
| ECG | Type | BF type |
| Sampling Rate | 250 sample/sec | |
| Input Offset Dynamic Range | $\pm$ 300mV | |
| Channel | 1 channel | |
| ADC Resolution | 10 bits | |
| Input Impedance | $\geq$ 10MΩ | |
| Frequency Response | 0.05Hz to 40Hz | |
| Electrode | AC impedance | Less than 3KΩ (10Hz) |
| RF | RF communication | 2.4GHz BLE 4.2 |
| Effective Radiated Power | <1mW | |
| RF Frequency Band of TX | 2.4GHz | |
| Bandwidth of the Receiver | 2400 ~ 2480MHz |
6. Technological Characteristic
| CPU | ARM Cortex-M4 | |
|---|---|---|
| S/W | Supported App | Android 8.1 orhigher iOS 12.0 orhigher |
| Supported PC S/W Version | Window 10 (64bit) | |
| Power Requirement | Power Supply | DC 3V, Coin Battery (CR2032) |
| Intended use period | Up to 14 days |
{5}------------------------------------------------
| Total Size (L x W x H: mm) | 95 x 52.6 x 8.3 | |
|---|---|---|
| PhysicalCharacteristics | Main Body Size (L x W x H:mm) | 31 x 39 x 8.3 |
| Weight (g) | Main Body: Below 13g | |
| Lifetime | 12 months | |
| QoS | Latency | ATN-C130: less than130msec ATR-C130: lessthan 150msec |
| Throughput | ATN-C130: more than12,000bps ATR-C130: morethan 4,500bps | |
| Distance | Maximum 7 (m) |
7. Intended Use / Indication for use
The device is intended to measure, save, and view continuous electrocardiogram (ECG) information for long-term recording (up to 14 days) from ambulatory patients by attaching to the skin surface. ECG records are stored in the the device for review after the recording period (up to 14 days) is completed, and are not intended for for real-time monitoring. The device does not include automated ECG analysis functions, and the recorded ECGs are not intended for automated analysis. The device allows patient and clinicians to view the ECG signal recorded in realtime solely for the purpose of visual assessment of the recording quality; the ECG signal displayed in real time is not intended for any clinical or diagnostic use. It is indicated for use on patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, fatigue, or anxiety.
8. Device Components
-
- AT-Patch Device (ATP-C130): The device that acquires and stores the patient's ECG signal for up to 14 days.
- App (ATN-C130): App to check that the device is operating normally and allows the patient and caregiver to manually record symptoms. The app has the following features.
2-1). Check if the device is operating normally (View ECG data - not intended for clinical or diagnostic uses)
2-2). Symptom note recording (allows to manually note that a patient feels an abnormality while wearing it)
2-3). Log-in function (user ID, password)
2-4). Operation connection for start the
device record 2-5). Software lock function
-
Live on view 60 minutes limit function per days
-
30 minutes access limit in case of 5 consecutive login failures
{6}------------------------------------------------
-
Login execution function when not in use for a long time (30 minutes)
-
- AT-Report Software for ECG viewing (ATR-C130): Software that downloads the patient's ECG data saved in the device's memory and the symptom note record recorded by the patient from the App (ATN-C130) through a specific cable provided. The software also allows to visualize the ECG signal for manual interpretation. The software does not include automated ECG analysis. This software also provides the main functions of ATN-130 App via a BlueTooth dongle, allowing the use of the device by patients not having a mobile phone compatible with ATN-130 App.
Accessories:
- Cable
- BlueTooth Low Energy dongle
9. Substantial Equivalence Discussion
AT-Patch is substantially equivalent to the predicate device, Zio SR ECG Monitoring service (K143513). The following comparison table is presented to demonstrate substantial equivalence.
- A. Comparison with predicate device
| DescriptiveInformation | Subject Device | Predicate Device |
|---|---|---|
| Manufacturer | ATsens Co.,Ltd | iRhythm technologies, Inc. |
| Device Name | AT-Patch | Zio SkyRunner (SR)Electrocardiogram(ECG) MonitoringService |
| 510(k) number | K203638 | K143513 |
| ClassificationProduct Code /RegulatoryNumber | DSH (21CFR §870.2800) | DSH(21CFR §870.2800)DQK(21CFR § 870.1425)DXH(21CFR § 870.2920) |
| Regulatory Class | 2 | 2 |
| Patientenvironment | Ambulatory | Ambulatory |
| Patient population | Non-pediatric,Non-critical care patients | Non-pediatric,Non-critical care patients |
{7}------------------------------------------------
| Indications forUse | The device is intended tomeasure, save, and viewcontinuous electrocardiogram(ECG) information for long-term recording (up to 14 days)from ambulatory patients byattaching to the skin surface.ECG records are stored in thethe device for review after therecording period (up to 14days) is completed, and are notintended for for real-timemonitoring. The device doesnot include automated ECGanalysis functions, and therecorded ECGs are notintended for automatedanalysis. The device allowspatient and clinicians to viewthe ECG signal recorded inreal-time solely for the purposeof visual assessment of therecording quality; the ECGsignal displayed in real time isnot intended for any clinical ordiagnostic use. It is indicatedfor use on patients 18 years orolder who may beasymptomatic or who maysuffer from transient symptomssuch as palpitations, shortnessof breath, dizziness, light-headedness, fatigue, or anxiety. | The ZIO SR ECG MonitoringService is intended to capture,analyze, and reportsymptomatic and/orcontinuous electrocardiogram(ECG) information for long-term monitoring (up to 14days).It is indicated for use on adultpatients 18 years or older whomay be asymptomatic or whomay suffer from transientsymptoms such as palpitations,shortness of breath, dizziness,light-headedness, pre-syncope, syncope, fatigue, oranxiety. The reported ECGmetrics include single-leadanalysis on a beat by beatbasis, heart rate measurementand rhythm analysis. Thereport does not containdiagnostic interpretation; thereported analysis is providedfor review by the intended userto render a diagnosis based onclinical judgment andexperience. |
|---|---|---|
| Prescription orOTC | Prescription | Prescription |
| Technological characteristics | ||
| Key systemcomponents | 1) AT-Patch2) App3) AT-Report | 1) Zio SR Patch Recorderwith Bluetooth technology2) Zio SR Wireless Gateway3) ZIO ECG UtilizationService System |
| Wear period | Up to 14 days | Up to 14 days |
| ECG channel | 1 channel | 1 channel |
| Recording Format | Continuous | Continuous |
| Input Impedance | ≥ 10MΩ | ≥ 10MΩ |
| RFCommunication | 2.4 GHz Bluetooth | 2.4 GHz Bluetooth |
| Frequency of band | 2400 ~ 2480 MHz | 2400 ~ 2480 MHz |
| Power supply | DC 3V, Coin battery (CR2032) | DC 3V, Coin battery (CR2032)(2ea) |
| ECG analysis | View ECG | Rhythmtypes BeatHeartratesRuns |
| Operationaltemperature | 41 to 104° F | 41 to 104° F |
| Operationalhumidity | 10% to 95% (non-condensing) | 10% to 95% (non-condensing) |
| App function | 1. Check whether the deviceis working normally2. Document symptom events | 1. Document symptom events |
| LOC of software | Major | Moderate |
{8}------------------------------------------------
1. The same between Subject device and Predicate Device.
1) Product Code
The proposed product code of the subject device is DSH. This is the same product code of the predicate device in K143513.
2) Regulatory number and classification
The proposed regulatory number of the subject device is 878.2800 and the classification is 2. Compared to the predicate device, it does not contain 870.1425 and 870.2920. It is the same regulatory number and classification as the predicate device in K143513.
3) Indications for Use
The subject device is indicated for use to measure, save, and view electrocardiogram (ECG) information for long-term recording (up to 14 days) by attaching to the skin surface. The predicate device is indicated for use to capture, analyze, and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring. It is similar as the predicate device in K143513.
{9}------------------------------------------------
4) Prescription Use
The subject device is a prescription use device. It is the same as the predicate devices in K143513.
5) Patients environment
The subject device is a wearable device. So, the patient can use this device while moving in daily life. It is the same as the predicate devices in K143513.
6) Technical characteristic
The key functions of the subject device are similar in most functions except in some components and App functions compared to the predicate device. The technological characteristics are similar compared to the predicate devices in K143513.
| Performance test standard | Subject device | Predicate device |
|---|---|---|
| IEC 60601-1 | 0 | 0 |
| IEC 60601-1-2 | 0 | 0 |
| IEC 60601-1-6 | 0 | 0 |
| IEC 60601-1-11 | 0 | 0 |
| IEC 62304 | 0 | 0 |
| IEC 62366 | 0 | 0 |
| IEC 60601-2-47 | 0 | 0 |
| ANSI/AAMI EC12 | 0 | 0 |
| Cybersecurity test | 0 | 0 |
| Battery using time performancetest | 0 | 0 |
| ISO 10993-1 | 0 | 0 |
| ISO 10993-5 | 0 | 0 |
| ISO 10993-10 | 0 | 0 |
| ISO 10993-12 | 0 | 0 |
| ISO 10993-23 | 0 | - |
1.2 Difference between Subject device and Predicate Device
1) Key system components
The subject device has different system components (AT-Patch, AT-Report and App). The subject device does not have wireless gateway. The wireless gateway of the predicate device role is a transmitter that sends a real-time recording such as ECG waveforms recorded in the app when the device is operated in iRhythm's monitoring
{10}------------------------------------------------
center for reporting. The subject device shows the electrocardiogram record in real time as an App solely for visual signal quality assessment, but does not have the function of sending it to a diagnostic center such as a hospital, therefore a wireless gateway is not needed for the subject device. ECG signals recorded by the subject device are not intended for automated analysis, and the subject device does not include automated analysis functions.
| Subject device | Predicate device | |
|---|---|---|
| SameComponent | 1. AT-Patch2. AT REPORT (ATR-C130)3. App (ATN-C130) | 1. Zio SR Patch recorder2. Zio SR wireless Gateway3. Zio ECG Utilization Service system |
| DifferentComponent | - | Zio SR gateway |
The presence or absence of a gateway does not affect the measurement, recording and analysis of data after use for 14 days, so subject device and predicate devices are not different due to this difference.
2) App function
: The subject device has a function to check whether the device is working normally through ECG waveform on the App compared to the predicated device. The subject device is a recorder, whereas the predicate device is a monitoring device. The predicate device can view the patient ECG waveform in the monitoring center for report of iRhythm technologies, Inc. The difference between the subject device and the predicate device is whether the corresponding function is implemented in the App or the monitoring center for reporting. The function of App allows the patient to check whether the device they are wearing is operating normally.
| Subject device | Predicate device | |
|---|---|---|
| Same function | 1. Document symptom events | 1. Document symptom events |
| Differentfunction | 1. To check whether the device isworking normally | - |
※ The patient's waveform through the App is simply a means to check whether the device is working normally. The record is not saved. It is impossible to analyze any symptoms such as arrhythmia through this. Therefore, whether the waveform is viewed by the patient or at the center. it is a difference in convenience and not in function that has any influence on the purpose of use of the product. Therefore, the difference in App function does not affect the equivalence with the predicate device.
3) Indication for use
: The subject device is indicated for use to measure, save, and view electrocardiogram (ECG) information for long-term recording (up to 14 days) by attaching to the skin
{11}------------------------------------------------
surface. The predicate device is indicated for use to capture, analyze, and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information for long-term monitoring.
| Subject device | Predicate device | |
|---|---|---|
| Same part | continuously measure, save andview patient's ECG | continuously measure, save andview patient's ECG |
| Differentpart | View ECG information forlong- term recording (up to 14days) by attaching to the skinsurface. | analyze, and report symptomaticand asymptomatic cardiac eventsand continuouselectrocardiograminformation for long-termmonitoring |
4) ECG Analysis software
The subject device does not have ECG analysis functions. By receiving the patient's data stored in the device, you can manage information including the patient's ECG data and view the patient's ECG. On the other hand, the predicate device can analyze, and report symptomatic and asymptomatic cardiac events and continuous electrocardiogram information in their ECG Analysis Software.
| Subject device | Predicate device | |
|---|---|---|
| Samefunction | 1. Patientinformationmanagement2. View patients ECG data | 1. Patient information management2. View patients ECG data |
| Differentfunction | - | Analyze, and report patients ECGdata |
The subject device has been tested about electrical safety, EMC, biocompatibility, and performance. The results show that these differences do not raise any problems in the safety and effectiveness
10. Electrical Testing and EMC Testing
The subject device is an electrical medical device. Electrical hazard, mechanical hazard and high temperature hazard are included within the device.
The electrical and EMC tests were performed in accordance with the FDA recognized standards.
{12}------------------------------------------------
- IEC 60601-1:2005/2012, Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-47:2012, Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic system.
- -IEC 60601-1-2:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- -IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
The test results met the electrical safety and EMC requirements.
11. Performance Testing - Nonclinical
1) Biocompatibility
The biocompatibility tests were performed in accordance with the FDA recognized standards:
- ISO 10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process.
- -ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cvtotoxicity
- -ISO 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
- ISO 10993-12:2012 Biological evaluation of medical devices Part 12: Sample preparation and reference materials
- -ISO 10993-23:2021(E) Biological evaluation of medical devices - Part 23: Tests for irritation
2) Sterilization and Shelf life test
This device is not sterile.
The shelf life of device was verified in accordance with the internal criteria.
As a result of testing according to the internal criteria, it has been verified that this device can be use for up 14 days and satisfies the warranty period of 1 year.
- ATP-C130 intended using time test
{13}------------------------------------------------
As a result of testing according to the internal criteria, it has been verified that this device can be used for up 14 days.
- Patch adhesion persistence test
Patch adhesion persistence test was performed in accordance with FDA Standard
- ANSI/AAMI EC12:2000 (R2015) Disposable ECG Electrodes
- ECG Electrodes performance test
ECG Electrode was performed according to FDA Standard
- ANSI/AAMI EC12:2000 (R2015) Disposable ECG Electrodes
- Cybersecurity test
Cybersecurity tests were performed in accordance the FDA recognized guidance.
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices issued in 2018.
- Extractable Substance Test
The extractable substance tests were performed about PH, Potassium permanganatereducible substances, Residue on evaporation, UV spectrum and Heavy metals.
- ECG Electrodes test
ECG Electrodes tests were performed in accordance 'ANSI/AAMI EC12:2000(R2015)'.
- Software
Software validation was evaluated according to IEC62304:2015.
10) Energy reduction
Energy reduction tests were performed using alternative method of IEC 60601-1 Clause 8.5.5.2. The device was demonstrated to be safe in case of defibrillation, however the device is not defibrillation-proof.
- QoS
Qos test are performed according to ANSI TIR 69:2017
- Usability
Usability formative & summative test were performed according to IEC 62366-1:2015 and FDA guidance.
The Test results met the performance requirements.
{14}------------------------------------------------
12. Conclusion
In comparison between the subject device and the predicate device, there are the similar indications for use, principle of operation, and technological characteristics. Although there are some differences (e.g., App function, system components, medical process and ECG Analysis function), they do not raise new or different questions of safety and effectiveness as compared to the predicate device. Results of the tests performed support the claims of substantial equivalence.
In this regard, we conclude that the subject device is substantially equivalent to the predicate device.
§ 870.2800 Medical magnetic tape recorder.
(a)
Identification. A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.(b)
Classification. Class II (performance standards).