K Number
K203613

Validate with FDA (Live)

Date Cleared
2021-08-27

(260 days)

Product Code
Regulation Number
876.5980
Age Range
2 - 12
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.

Device Description

The proposed device with Enfit® connector (1ml-60ml) are standard piston style syringes consisting of syringe barrel with integral EnFit syringe tip, syringe plunger and ring type piston. The proposed device are supplied non-sterile, and reusable. The sizes rang from 1ml to 60ml nominal capacity. The integral syringe tip is a female ENFit connector which is compatible only with enteral access devices or accessories having ENFit compliant or compatible male connectors to form a dedicated system that prevents wrong-route administration of fluids.

AI/ML Overview

This is a 510(k) summary for a medical device (ENFit Reusable Enteral Syringe). 510(k) submissions typically demonstrate substantial equivalence to a predicate device through non-clinical testing, rather than extensive clinical efficacy studies with acceptance criteria in the manner of AI/software devices. Therefore, the information requested in the prompt, especially related to AI-specific metrics, reader studies, and ground truth establishment, is not applicable or present in this document.

However, I can extract the non-clinical tests performed and mention that these tests serve as the "acceptance criteria" for demonstrating substantial equivalence for this type of device.

1. A table of acceptance criteria and the reported device performance

The document lists various non-clinical tests performed to verify that the proposed device met design specifications and was substantially equivalent to the predicate device. These tests, and the implication that the device met their requirements, serve as the "reported device performance" and "acceptance criteria" for a 510(k) submission. No specific numerical thresholds are provided in this summary, but the successful completion of these tests indicates compliance.

Acceptance Criteria (Non-Clinical Tests Conducted)Reported Device Performance
A. Finished Device (Performance Testing per ISO 7886-2:2020)
Risk management report in accordance with ISO 14971:2007Met (implied by conclusion of SE)
Usability testing for ENFit Reusable Enteral Syringe (human factor and usability validation)Met (implied by conclusion of SE)
AppearanceMet (implied by conclusion of SE)
Pump ForceMet (implied by conclusion of SE)
Short-term Flow Rate ErrorMet (implied by conclusion of SE)
Syringe ComplianceMet (implied by conclusion of SE)
Syringe Design critical dimensionMet (implied by conclusion of SE)
B. Reusability
Cleaning Instructions Validation and Use Cycle Parameters StudyMet (implied by conclusion of SE)
C. Biocompatibility
ISO 10993-5:2009 (Tests for Vitro Cytotoxicity)No Cytotoxicity (explicitly stated in comparison table, and implied by conclusion of SE)
ISO 10993-10:2010 (Tests for irritation and skin sensitization)No Irritation and Sensitization (explicitly stated in comparison table, and implied by conclusion of SE)
ISO 10993-11:2017 (Tests for systemic toxicity)No Acute Toxicity (explicitly stated in comparison table, and implied by conclusion of SE)
D. Shelf Life
Package verification test reportMet (implied by conclusion of SE)
Bench test after real-time aging (ISO 80369-3:2016)
Fluid LeakageMet (implied by conclusion of SE)
Sub-atmospheric Pressure Air LeakageMet (implied by conclusion of SE)
Stress CrackingMet (implied by conclusion of SE)
Resistance to Separation from Axial LoadMet (implied by conclusion of SE)
Resistance to Separation from UnscrewingMet (implied by conclusion of SE)
Resistance to OverridingMet (implied by conclusion of SE)
Disconnection by UnscrewingMet (implied by conclusion of SE)
Bench test after real-time aging (ISO 7886-1:2017)
AppearanceMet (implied by conclusion of SE)
Overall Length of Scale to Nominal Capacity LineMet (implied by conclusion of SE)
Push-button DistanceMet (implied by conclusion of SE)
Force Required to Operate PlungerMet (implied by conclusion of SE)
Freedom from Air and Liquid Leakage past PistonMet (implied by conclusion of SE)
Limits for Acidity or AlkalinityMet (implied by conclusion of SE)
Limits for Extractable MetalsMet (implied by conclusion of SE)
Tolerance on Graduated CapacityMet (implied by conclusion of SE)
Maximum Dead SpaceMet (implied by conclusion of SE)
E. Finished Device Performance Test (not related to aging)
Critical Dimension verificationMet (implied by conclusion of SE)
Ink AdhesionMet (implied by conclusion of SE)
Bench test (ISO 80369-3:2016)(These are repeated from the "after real-time aging" section, but without the "after real-time aging" qualifier, suggesting they are also tested in the non-aged state.)
Fluid LeakageMet (implied by conclusion of SE)
Sub-atmospheric Pressure Air LeakageMet (implied by conclusion of SE)
Stress CrackingMet (implied by conclusion of SE)
Resistance to Separation from Axial LoadMet (implied by conclusion of SE)
Resistance to Separation from UnscrewingMet (implied by conclusion of SE)
Resistance to OverridingMet (implied by conclusion of SE)
Disconnection by UnscrewingMet (implied by conclusion of SE)
Bench test (ISO 7886-1:2017)(These are repeated from the "after real-time aging" section, but without the "after real-time aging" qualifier, suggesting they are also tested in the non-aged state.)
AppearanceMet (implied by conclusion of SE)
Overall Length of Scale to Nominal Capacity LineMet (implied by conclusion of SE)
Push-button DistanceMet (implied by conclusion of SE)
Force Required to Operate PlungerMet (implied by conclusion of SE)
Freedom from Air and Liquid Leakage past PistonMet (implied by conclusion of SE)
Limits for Acidity or AlkalinityMet (implied by conclusion of SE)
Limits for Extractable MetalsMet (implied by conclusion of SE)
Tolerance on Graduated CapacityMet (implied by conclusion of SE)
Maximum Dead SpaceMet (implied by conclusion of SE)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes for each non-clinical test conducted. It states that "Non-clinical tests were conducted to verify that the proposed device met all design specifications." The data provenance (country of origin, retrospective/prospective) is not mentioned, as these are material performance tests rather than clinical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is a physical medical device (syringe), not an AI/software device that requires expert ground truth for classification or diagnosis. Therefore, this information is not applicable and not present in the document.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable, as this is not an AI/software device or a study involving human interpretation with adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a physical medical device. No MRMC study was conducted or reported.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For these physical device tests, the "ground truth" is defined by the established industry standards (e.g., ISO 14971, ISO 7886-2, ISO 10993 series, ISO 80369-3) and pre-defined specifications for device performance (e.g., freedom from leakage, specific dimensions, resistance to force). The device's performance against these verifiable physical and chemical metrics serves as the ground truth.

8. The sample size for the training set

Not applicable. This is a physical medical device, not an AI/software device requiring a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

August 27, 2021

Shantou Wealy Medical Instrument Co., Ltd. % Eva Li Consultant Shanghai Sungo Management Consulting Company Limited Room 1309, Dongfang Building, 1500#Century Ave Shanghai, Shanghai 200122 CHINA

Re: K203613

Trade/Device Name: ENFit Reusable Enteral Syringe Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: PNR Dated: July 30, 2021 Received: July 30, 2021

Dear Eva Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general control's provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K203613

Device Name ENFit® Reusable Enteral Syringe

Indications for Use (Describe)

The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

510(K) Summary

A. Applicant

Shantou Wealy Medical Instrument Co., Ltd. Address: North Jinhuan Road (near of Qishan Mid-school) 515064 Shantou, China Contact Person: Yang Songling Tel: +8615816758600 Fax: 0086-754-82121654

Date Prepared: July 27, 2021

Submission Correspondent Primary contact: Ms. Eva Li Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: eatereva@hotmail.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

B. Device

Trade Name: ENFit® Reusable Enteral Syringe Common Name: Enteral Syringe with Enteral Specific connector Model: 1mlLD, 2.5mlLD, 5ml, 10ml, 20ml, 60ml Product Series Code: WR

Regulatory Information Classification Name: Gastrointestinal tube and accessories Regulator Class: Class II Product code: PNR Regulation Number: 876.5980 Review Panel: Gastroenterology/Urology

C. Predicate device:

K183540

Oral/Enteral Syringes with ENFit® connector (12 mL to 60 mL) and Low Dose Tip Oral/Enteral Syringes with ENFit® connector (1 mL to 6 mL) NeoMed, Inc.

D. Intended use of the device:

{4}------------------------------------------------

North Jinhuan Road(near Qishan mid-school), 515064 Shantou, China The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.

E. Device Description:

The proposed device with Enfit® connector (1ml-60ml) are standard piston style syringes consisting of syringe barrel with integral EnFit syringe tip, syringe plunger and ring type piston. The proposed device are supplied non-sterile, and reusable. The sizes rang from 1ml to 60ml nominal capacity. The integral syringe tip is a female ENFit connector which is compatible only with enteral access devices or accessories having ENFit compliant or compatible male connectors to form a dedicated system that prevents wrong-route administration of fluids.

F. Comparison with predicate device

DeviceProposed DevicePredicate DeviceComparison
ManufacturerShantou Wealy Medical Instrument Co., Ltd.NeoMed, Inc.--
510(K) numberK203613K183540--
Product CodePNRPNRSame
Regulation NumberCFR 876.5980CFR 876.5980Same
Intend useThe ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.The device is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.Same
ConfigurationBarrel with ENFit connectorPurple PlungerPistonBarrel with ENFit connectorPurple PlungerPistonSame
SizeLow dose tipENFit™ syringe1ml, 2.5ml,Low dose tip ENFit™ 1ml-6mlSimilar*

Table 1 General Comparison

{5}------------------------------------------------

Standard ENFit™syringe5ml-60mlStandard ENFit™syringe12ml-60ml
SterileNoNoSame
ReusableYesYesSame
BiocompatibilityNo CytotoxicityNo CytotoxicitySame
No Irritation and SensitizationNo Irritation and Sensitization
No Acute ToxicityNo Acute Toxicity

*Similar discussion

The size of the proposed device of the low dose tip syringe or standard syringe is similar, all the size of the proposed are pass the acceptance criteria of the related performance standard requirements, so it will not affect the safety and effectiveness of the proposed device.

G. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following performance tests are conducted:

  • A Finished Device
    • Risk management report in accordance with ISO 14971:2007 Medical devices -● Application of risk management to medical devices
    • . Summary of the Usability testing for ENfit Reusable Enteral Syringe(human factor and usability validation)
    • . Performance testing with ISO 7886-2:2020 "Sterile hypodermic syringes for single use Part 2: Syringes for use with power-driven syringe pumps
      • ——Appearance
      • -- Pump Force
      • -- Short-term Flow Rate Error
      • -- Syringe Compliance
      • ——Syringe Design critical dimension
  • A Reusability
    • . Cleaning Instructions Validation and Use Cycle Parameters Study
  • Biocompatibility

    • ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity;
    • ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization;
    • . ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity
  • A Shelf life
    • Package verification test report 0
    • . Bench test after real-time aging(ISO 80369-3:2016, Small-bore connectors for liquids and gases in healthcare application-Part3: Connectors for enteral applications ) ——Fluid Leakage

{6}------------------------------------------------

K203613 Page 4 of 5

  • North Jinhuan Road(near Qishan mid-school), 515064 Shantou, China
  • -- Sub-atmospheric Pressure Air Leakage
  • -- Stress Cracking
  • -- Resistance to Separation from Axial Load
  • -- Resistance to Separation from Unscrewing
  • -- Resistance to Overriding
  • -- Disconnection by Unscrewing
  • . Bench test after real-time aging(ISO 7886-1:2017 Sterile hypodermic syringe for single use-Part 1: Syringe for manual use )
    • -- Appearance
    • ——Overall Length of Scale to Nominal Capacity Line
    • -- Push-button Distance
    • -- Force Required to Operate Plunger
    • ——Freedom from Air and Liquid Leakage past Piston
    • -- Limits for Acidity or Alkalinity
    • ——Limits for Extractable Metals
    • -- Tolerance on Graducated Capacity
    • ——Maximum Dead Space
  • Finished Device performance Test
    • Critical Dimension verification
    • Ink Adhesion
  • . Bench test (ISO 80369-3:2016, Small-bore connectors for liquids and gases in healthcare application-Part3: Connectors for enteral applications )
    • --Fluid Leakage
    • -- Sub-atmospheric Pressure Air Leakage
    • -- Stress Cracking
    • ——Resistance to Separation from Axial Load
    • -- Resistance to Separation from Unscrewing
    • -- Resistance to Overriding
    • -- Disconnection by Unscrewing
  • . Bench test(ISO 7886-1:2017 Sterile hypodermic syringe for single use-Part 1: Syringe for manual use )
    • ——Appearance
    • ——Overall Length of Scale to Nominal Capacity Line
    • ——Push-button Distance
    • -- Force Required to Operate Plunger
    • ——Freedom from Air and Liquid Leakage past Piston
    • -- Limits for Acidity or Alkalinity
    • ——Limits for Extractable Metals
    • ——Tolerance on Graduated Capacity
    • ——Maximum Dead Space

H. Clinical Test Conclusion

No clinical study is included in this submission.

{7}------------------------------------------------

North Jinhuan Road(near Qishan mid-school), 515064 Shantou, China

I. Conclusion

Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Oral/Enteral Syringes with ENFit® connector (12 mL to 60 mL) and Low Dose Tip Oral/Enteral Syringes with ENFit® connector (1 mL to 6 mL) cleared under K183540.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.