K Number
K203596

Validate with FDA (Live)

Date Cleared
2021-04-20

(132 days)

Product Code
Regulation Number
N/A
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Panthera Occlusal Appliance is indicated for protections from the forces and damage of parafunctions like bruxism, prevention of noise associated with bruxism, and for alleviation of temporomandibular joint, muscle and tension headache pains associated with temporomandibular disorders.

Device Description

The Panthera Occlusal Appliance is a removable intraoral device used for protecting teeth and restorations against the forces of bruxism. It consists of patient-specific splint that fit over the upper or lower teeth. This device functions as a protective barrier for teeth and restorations by creating physical separation between upper and lower tooth surfaces preventing tooth damage caused by bruxism (like grinding and clenching) and alleviating temporomandibular joint, jaw, and muscle and tension headache pains. There are various occlusal designs for this application, all with varying degrees of function and suitability depending on the patient's needs. The Panthera Occlusal Appliance is customized for each patient based on the clinician's prescription, and is to be worn every day during hours of sleep.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Panthera Occlusal Appliance. This document focuses on demonstrating substantial equivalence to previously marketed devices rather than proving the performance of a novel AI/software-driven medical device against specific acceptance criteria using a study.

Therefore, the information required to answer the prompt regarding AI/software performance, such as specific acceptance criteria for algorithm performance, sample sizes for test/training sets, expert adjudication methods, MRMC studies, standalone performance, and ground truth establishment, is not present in the provided document.

The document states:

  • "Human clinical study was not deemed necessary to support substantial equivalence." (page 7)
  • "The Panthera Occlusal Appliance does not use design dissimilar from the primary predicate and reference devices, does not use new technologies different from the primary predicate and reference devices, and does not deviate from the indications for use of the primary predicate and reference devices." (page 7)

This indicates that the submission relies on demonstrating that the new device is fundamentally similar in design, technology, and indications for use to existing, legally marketed devices (predicates and references). The "bench testing" mentioned (page 7) is for physical properties and intended use, not for AI/algorithm performance.

In summary, the provided document does not contain the information requested in the prompt, as it is a 510(k) for a physical dental appliance based on substantial equivalence, not an AI/software medical device requiring performance studies against specific acceptance criteria.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's role in protecting public health by ensuring the safety, efficacy, and security of human and veterinary drugs, biological products, and medical devices.

April 20, 2021

Martine Fortin, Director Regulatory Affairs and Quality Assurance Panthera Dental Inc. 9105, rue John-Simons Quebec, Quebec G2B OS6 Canada

Re: K203596

Trade/Device Name: Panthera Occlusal Appliance Regulatory Class: Unclassified Product Code: MQC, OCO Dated: March 19, 2021 Received: March 22, 2021

Dear Martine Fortin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adiodha. M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203596

Device Name Panthera Occlusal Appliance

Indications for Use (Describe)

The Panthera Occlusal Appliance is indicated for protections from the forces and damage of parafunctions like bruxism, prevention of noise associated with bruxism, and for alleviation of temporomandibular joint, muscle and tension headache pains associated with temporomandibular disorders.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo for a dental practice. The logo features a black panther in a crouching position above the words "PANTHERA DENTAL" in purple font. The word "PANTHERA" is in a larger font than the word "DENTAL". The panther is the main focus of the logo, and it is positioned to look sleek and powerful.

510 (k) Summary

Owner Information 1.

Owner's namePanthera Dental Inc.
Owner's address9105 rue John-SimonsQuébec (QC) G2B 0S6Canada
Telephone(418) 527-0388
Fax(418) 431-9942
Official ContactMartine FortinDirector, Quality Assurance and Regulatory Affairsregulatory@panthera.com
Date PreparedApril 19, 2021

Device Name 2.

Trade NamePanthera Occlusal Appliance
Common NameMouthguard, Prescription
ClassificationUnclassified, Pre-Amendment
Review PanelDental
Product CodeMQC
Subsequent Product CodeOCO
510(k) numberK203596

3. Predicate devices

Primary PredicateThe POD (K182820)
Reference DevicesSomnoBrux - Various versions of SomnoBrux (K102909)
Panthera Anti-Snoring Device (K143244)

4. Description

The Panthera Occlusal Appliance is a removable intraoral device used for protecting teeth and restorations against the forces of bruxism. It consists of patient-specific splint that fit over the upper or lower teeth. This device functions as a protective barrier for teeth and restorations by creating physical separation between upper and lower tooth surfaces

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther, the word "PANTHERA" is written in a bold, sans-serif font, with a purple line underneath. Below "PANTHERA", the word "DENTAL" is written in a smaller, sans-serif font.

preventing tooth damage caused by bruxism (like grinding and clenching) and alleviating temporomandibular joint, jaw, and muscle and tension headache pains. There are various occlusal designs for this application, all with varying degrees of function and suitability depending on the patient's needs. The Panthera Occlusal Appliance is customized for each patient based on the clinician's prescription, and is to be worn every day during hours of sleep.

5. Indications for Use

The Panthera Occlusal Appliance is indicated for protection of teeth and restorations from the forces and damage of parafunctions like bruxism, prevention of noise associated with bruxism, and for alleviation of temporomandibular joint, muscle and tension headache pains associated with temporomandibular disorders.

Technological characteristics 6.

The technical characteristics of the Panthera Occlusal Appliance are substantially equivalent to the three (3) previously marketed devices. The table below compares the technological aspects of the new, the predicate, and the references devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a stalking pose above the words "PANTHERA DENTAL" in a stylized font. The word "PANTHERA" is in a larger, purple font, while "DENTAL" is in a smaller, gray font below it.

FeaturePanthera OcclusalAppliance(proposed device)The POD(Primary Predicate)K182820SomnoBrux Splint(Reference device)K102909The Panthera Anti-Snoring Device(Reference device)K143244
ClassificationUnclassifiedSameSameN/Ap.
Product codeMQC, OCOSameMQCN/Ap.
Indications forUseThe Panthera OcclusalAppliance is indicatedfor protection of teethand restorations fromthe forces and damageof parafunctions likebruxism, prevention ofnoise associated withbruxism, and foralleviation oftemporomandibularjoint, muscle andtension headache painsassociated withtemporomandibulardisorders.For the amelioration ofclenching and bruxingassociatedwith TMD and is tobe used to aid in the reliefof symptoms ofTMD/TMJ.Used for the protection ofteeth and restorationsfrom the forces ofbruxismN/Ap.
Materials ofconstructionMade from polymers(polyamide type 12),metal-freeMethylmethacrylate,medical gradestainless steelAcrylicSame
Prescription/OTCPrescription onlySameSameSame

Table 1: Comparison chart between the submitted device and the predicates.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its head lowered and its tail curled up. Below the panther, the word "PANTHERA" is written in a stylized font, with the word "DENTAL" written in a smaller font below it. The word "PANTHERA" is purple.

Traditional 510(k) – Panthera Occlusal Appliance

FeaturePanthera OcclusalAppliance(proposed device)The POD(Primary Predicate)K182820SomnoBrux Splint(Reference device)K102909The Panthera Anti-Snoring Device(Reference device)K143244
Fixed/removableRemovableSameSameSame
Suppliedsterile/nonsterileNon sterileSameSameSame
TargetpopulationAdult patientsSameSameSame
Single Use/reusableReusableSameSameSame
DesignComputer-aideddesign (CAD) frompatient dental model orstl fileCustom-designed frompatient dental model bydental techniciansCustom-designed frompatient dental model bydental techniciansSame
ManufacturingComputer-aidedmanufacturing (CAM)N/Ap.N/Ap.Same

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position above the words "PANTHERA DENTAL". The word "PANTHERA" is in a stylized font, and the word "DENTAL" is in a smaller, simpler font. A purple line is underneath the word "PANTHERA".

Non-clinical performance data 7.

Bench testing

The bench testing includes assessment of the physical properties of the Panthera Occlusal Appliance and its ability to achieve its intended use. The Panthera Occlusal Appliance meets the same specifications as set for the primary predicate.

Biocompatibility testing

The biocompatibility testing conducted for the reference device (K143244) is valid because the final product of the proposed device (including material composition and manufacturing) is the same as the final product of the reference device (K143244). Therefore, no additional biocompatibility testing was conducted.

In addition, a risk analysis was performed as per ISO 14971, and found no new concerns for the subject device.

8. Clinical performance data

Human clinical study was not deemed necessary to support substantial equivalence. The Panthera Occlusal Appliance does not use design dissimilar from the primary predicate and reference devices, does not use new technologies different from the primary predicate and reference devices, and does not deviate from the indications for use of the primary predicate and reference devices.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a logo for Panthera Dental. The logo features a black panther in a crouching position, with its tail curled up. Below the panther, the word "PANTHERA" is written in a bold, sans-serif font, with the word "DENTAL" written in a smaller font to the right.

Substantial Equivalence Conclusion 9.

The new device, the Panthera Occlusal Appliance, is considered substantially equivalent to the predicate and the reference devices. The POA has the same indications for use as the predicate device (K182820); it is indicated for the same user population as the predicate and the reference devices; it uses biocompatible material for the manufacturing, as the reference device (K143244) does, and it has equivalent technological characteristics as the ones for the predicate and the reference devices.

N/A