K Number
K203566

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-05-13

(157 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Smith & Nephew Tablet Application is indicated for use to provide wireless control of Smith & Nephew compatible surgical and endoscopic devices within the operating room including camera/camera control unit, patient information system, mechanical resection system, fluid management system and RF coblation system. These controls consist of adjusting parameter settings only.

Device Description

The Smith & Nephew Tablet Application is a software application that provides a Wi-Fi connection between compatible medical devices. Once connected the Tablet Application has the ability to provide limited remote control to the connected devices.

AI/ML Overview

The provided document is a 510(k) premarket notification for the Smith & Nephew Tablet Application, which is a software application for wireless control of surgical and endoscopic devices. However, it does not contain specific acceptance criteria, performance data, or details about a study proving the device meets those criteria. The document focuses on demonstrating substantial equivalence to a predicate device (K192876) based on technological characteristics and intended use.

Therefore, I cannot provide the requested information from the given text.

Here's why and what information is missing:

  1. Table of acceptance criteria and reported device performance: The document states that "Testing demonstrated that the Smith & Nephew Tablet Application has met the performance specifications" (Page 5, Section H). However, it does not list what those specific "performance specifications" (acceptance criteria) are, nor does it provide the reported device performance against those criteria (e.g., specific accuracy, latency, reliability metrics).
  2. Sample size and data provenance: There is no mention of a test set, its sample size, or its provenance (e.g., country of origin, retrospective/prospective).
  3. Number/qualifications of experts for ground truth: Since there's no mention of a clinical study or performance evaluation involving human interpretation of data where ground truth would be established by experts, this information is absent.
  4. Adjudication method: Similarly, no adjudication method is described because no expert-based ground truth establishment is detailed.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study: The document does not describe any study involving human readers or AI assistance for human readers. The device's function is to control surgical parameters, not to assist in image interpretation or diagnosis.
  6. Standalone performance: While the software itself performs wirelessly, the concept of "standalone performance" in the context of an AI device (i.e., algorithm only without human-in-the-loop performance) is not applicable here, as its function is control, not interpretation or diagnosis. No specific performance metrics like sensitivity, specificity, or AUC for a diagnostic task are provided.
  7. Type of ground truth: No ground truth is described because the device's function is control, not diagnosis or interpretation that would require a ground truth for evaluation. The "performance specifications" it met would likely relate to functional requirements, responsiveness, and connectivity, not diagnostic accuracy.
  8. Sample size for the training set: There is no mention of a training set as this is not an AI/ML diagnostic or predictive algorithm.
  9. How ground truth for the training set was established: Not applicable, as there's no training set mentioned.

In summary, the provided document is a 510(k) summary for a device that controls surgical equipment wirelessly, not an AI/ML diagnostic or interpretative device. Therefore, the types of studies and performance metrics typically associated with AI/ML devices (like those requiring ground truth, expert readers, MRMC studies, etc.) are not present in this submission. The "performance data" mentioned (Page 5, Section H) refers to "Software validations" and "Cybersecurity testing," which are distinct from the type of performance data requested in the prompt for AI/ML devices.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 13, 2021

Smith & Nephew Sean Reynolds Director, Regulatory Affairs Specialist 150 Minuteman Rd Andover, Massachusetts 01810

Re: K203566

Trade/Device Name: Tablet Application Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: ODA Dated: April 14, 2021 Received: April 15, 2021

Dear Sean Reynolds:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203566

Device Name Smith & Nephew Tablet Application 4.0

Indications for Use (Describe)

The Smith & Nephew Tablet Application is indicated for use to provide wireless control of Smith & Nephew compatible surgical and endoscopic devices within the operating room including camera control unit, patient information system, mechanical resection system, fluid management system and RF coblation system. These controls consist of adjusting parameter settings only.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

T: + 1 978 749 1000 T: + 1 800 343 8386 (USA toll free) www.smith-nephew.com

K203566

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which the substantial equivalence is based.

Smith & Nephew: Smith+Nephew Tablet Application

Date Prepared: May 12, 2021

A. Submitter's Name:

Smith+Nephew, Inc., Endoscopy Division 150 Minuteman Road, Andover MA. 01810

B. Company Contact:

Sean Reynolds Director, Regulatory Affairs T 978-749-1173

C. Device Name

Trade Name: Smith+Nephew Tablet Application Common Name: Application Classification Name: Endoscopes and Accessories, 876.1500 Regulatory class: II Product Code: ODA

D. Predicate Devices

The Smith & Nephew Tablet Application presented in this submission is equivalent in intended use and fundamental scientific technology to the following legally marketed device currently in commercial distribution: Smith+Nephew Tablet Application cleared in K192876.

E. Description of Device

The Smith & Nephew Tablet Application is a software application that provides a Wi-Fi connection between compatible medical devices. Once connected the Tablet Application has the ability to provide limited remote control to the connected devices.

{4}------------------------------------------------

F. Intended Use

Indication for Use:

The Smith & Nephew Tablet Application is indicated for use to provide wireless control of Smith & Nephew compatible surgical and endoscopic devices within the operating room including camera/camera control unit, patient information system, mechanical resection system, fluid management system and RF coblation system. These controls consist of adjusting parameter settings only.

G. Comparison of Technological Characteristics

The Smith & Nephew Tablet Application has the following similarities as the predicate device, Smith & Nephew Tablet Application, cleared in K192876.

PropertyProposed DevicePredicate Device K192876
Intended UseThe Smith & Nephew TabletApplication is intended for use forwireless control of compatible Smith& Nephew surgical and endoscopicdevices.The Smith & Nephew TabletApplication is intended for use forwireless control of compatible Smith& Nephew surgical and endoscopicdevices.
Indications for useThe Smith & Nephew TabletApplication is indicated for use toprovide wireless control of Smith &Nephew compatible surgical andendoscopic devices within theoperating room includingcamera/camera control unit, patientinformation system, mechanicalresection system, fluid managementsystem and RF coblation system.These controls consist of adjustingparameter settings only.The Smith & Nephew TabletApplication is indicated for use toprovide wireless control ofparameters settings for compatibleSmith & Nephew surgical andendoscopic devices within theoperating room includingcamera/camera control unit, patientinformation system, DYONICSPOWER II Control Unit and theWerewolf Controller.
Regulation Number876.1500876.1500
ClassificationClass IIClass II
Pro CodeODAODA
Device NameTablet Application 4.0INTELLIO Tablet Application
Software4.003.05.49
WiFiYesYes

Both Applications utilizes Wi-Fi capability for connection with compatible Wi-Fi devices. The difference between the proposed Smith & Nephew Tablet Application and the currently cleared predicate Tablet Application (K192876) is the addition of arthroscopic inflow/ outflow fluid management system which communicates with the Tablet Application. These controls consist of adjusting parameter settings only.

{5}------------------------------------------------

K203566

H. Performance Data

Testing demonstrated that the Smith & Nephew Tablet Application has met the performance specifications, compatibility to the Arthroscopic inflow/ outflow fluid management system and required Cybersecurity testing therefore, is substantially equivalent to the predicate device cleared in K192876.

The following Software validations were conducted:

  • Software verification ●
  • . Software validation

G. Conclusion:

The Smith & Nephew Tablet Application has met all specified criteria and does not raise new safety or effectiveness questions. The substantial equivalence of the modified device is based on the same fundamental technology. Based on the similarities to the predicate device cleared in K192876, the Smith & Nephew Tablet Application, the proposed Smith+Nephew Tablet is substantially equivalent to its predicate device cleared in K192876.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.