(145 days)
AI Segmentation uses CT images to segment patient anatomy for use in radiation therapy treatment planning. AI Segmentation utilizes a pre-defined set of organ structures in the following regions: head and neck, thorax, pelvis, abdomen. Segmentation results are subject to review and editing by qualified, expert radiation therapy treatment planners. Results of AI Segmentation are utilized in the Eclipse Treatment Planning System where it is the responsibility of a qualified physician to further review, edit as needed, and approve each structure.
AI segmentation is a web-based application, running in the cloud, that provides a combined deep learning and classical-based approach for automated segmentation of organs at risk, along with tools for structure visualization. This software medical device product is used by trained medical professionals and consists of a web application user interface where the results from the automated segmentation can be reviewed and selected for export into the compatible treatment planning system. AI Segmentation is not intended to provide clinical decisions, medical advice, or evaluations of radiation plans or treatment procedures.
The provided text describes that the AI Segmentation device does not include clinical data in its premarket submission. The document explicitly states: "No animal studies or clinical tests have been included in this pre-market submission." Therefore, it is not possible to provide acceptance criteria or a study proving the device meets the acceptance criteria using the requested information (e.g., sample size for the test set, number of experts, adjudication method, MRMC study, standalone performance, type of ground truth for test and training sets, and training set size).
Instead, the submission for AI Segmentation focused on non-clinical data, specifically software verification and validation testing, and conformance to relevant standards.
Here is what can be extracted from the document regarding the non-clinical evaluation:
1. Table of Acceptance Criteria and Reported Device Performance:
Since no clinical study data is available, a table of acceptance criteria and reported device performance in terms of clinical accuracy (e.g., Dice score, sensitivity, specificity) cannot be provided. The performance data presented is focused on software quality and adherence to regulatory standards.
| Acceptance Criterion (Non-Clinical) | Reported Device Performance |
|---|---|
| Conformance to applicable software requirements and specifications | "Test results demonstrate conformance to applicable requirements and specifications." (Page 5) |
| Software level of concern assessment | Assessed as "major" level of concern. (Page 5) |
| Conformance to IEC 62304 Edition 1.1 2015-06 (Medical device software - Software life cycle processes) | Conforms in whole or in part. (Page 5) |
| Conformance to IEC 62366-1 Edition 1.0 2015-02 (Application of usability engineering to medical devices) | Conforms in whole or in part. (Page 6) |
| Conformance to IEC 62083 Edition 2.0 2009-09 (Requirements for the safety of radiotherapy treatment planning systems) | Conforms in whole or in part. (Page 6) |
| Conformance to IEC 82304-1 Edition 1.0 2016-10 (Health software Part 1: General requirements for product safety) | Conforms in whole or in part. (Page 6) |
| Absence of Safety or Customer Intolerable Discrepancy Reports (DRs) | "There were no remaining discrepancy reports (DRs) which could be classified as Safety or Customer Intolerable." (Page 6) |
2. Sample size used for the test set and the data provenance: Not applicable, as no clinical test set data from patients was submitted. The evaluation was based on software testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as no clinical test set requiring expert ground truth was submitted.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable, as no clinical test set requiring adjudication was submitted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was done, as explicitly stated that no clinical tests were included.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The document states, "Segmentation results are subject to review and editing by qualified, expert radiation therapy treatment planners. Results of AI Segmentation are utilized in the Eclipse Treatment Planning System where it is the responsibility of a qualified physician to further review, edit as needed, and approve each structure." This indicates the device is intended for human-in-the-loop use. However, no clinical performance data (standalone or otherwise) was provided. The software verification and validation would have tested the algorithm's output without human intervention, but these were functional tests, not clinical performance studies.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable for clinical ground truth, as no clinical studies were submitted. For software verification, ground truth would be against predetermined functional requirements and expected outputs established during software development.
8. The sample size for the training set: Not provided in the document.
9. How the ground truth for the training set was established: Not provided in the document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration".
Varian Medical Systems Inc. % Mr. Peter Coronado Director, Varian Oncology Systems Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
April 19, 2021
Re: K203469
Trade/Device Name: AI Segmentation Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ Dated: March 18, 2021 Received: March 19, 2021
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K203469
Device Name AI Segmentation
Indications for Use (Describe)
AI Segmentation uses CT images to segment patient anatomy for use in radiation therapy treatment planning. AI Segmentation utilizes a pre-defined set of organ structures in the following regions: head and neck, thorax, pelvis, abdomen. Segmentation results are subject to review and editing by qualified, expert radiation therapy treatment planners. Results of AI Segmentation are utilized in the Eclipse Treatment Planning System where it is the responsibility of a qualified physician to further review, edit as needed, and approve each structure.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "K203469" in black font above three blue circles. The circles are arranged horizontally and are evenly spaced. The text is positioned above the circles, with the "K" aligned with the left edge of the leftmost circle and the "9" aligned with the right edge of the rightmost circle.
varian
Varian Medical Systems 3100 Hansen Way Palo Alto, CA 94304
510(K) SUMMARY
The following information is provided as required by 21 CFR 807.92.
| SUBMITTER | |
|---|---|
| Name and Address: | Varian Medical Systems3100 Hansen Way, m/s E110Palo Alto, CA 94304 |
| Contact Person: | Peter J. CoronadoSr. Director, Regulatory AffairsPhone: 650-424-6320 Fax: 650-646-9200submissions.support@varian.com |
| Date Prepared: | 18 March 2021 |
| DEVICE | |
| Subject Device Name:Common/Usual Name:Product Code and Classification: | Al Segmentationmedical image segmentation softwareMedical charged-particle radiation therapy systemMUJ 21 CFR 892.5050 Class II |
| PREDICATE DEVICE | |
| Predicate Device Name:Reference Device: | Smart Segmentation Knowledge Based Contouring v2.2 (K141248)Ethos Treatment Planning v1.0 (K192377)Note: Referenced for the use of existing, established methods toevaluate performance of Al-based algorithms and technology |
DEVICE DESCRIPTION
Al segmentation is a web-based application, running in the cloud, that provides a combined deep learning and classical-based approach for automated segmentation of organs at risk, along with tools for structure visualization. This software medical device product is used by trained medical professionals and consists of a web application user interface where the results from the automated segmentation can be reviewed and selected for export into the compatible treatment planning system. Al Segmentation is not intended to provide clinical decisions, medical advice, or evaluations of radiation plans or treatment procedures.
INDICATIONS FOR USE
Al Segmentation uses CT images to segment patient anatomy for use in radiation therapy treatment planning. Al Segmentation utilizes a pre-defined set of organ structures in the following regions: head and neck, thorax, pelvis, abdomen. Segmentation results are subject to review and editing by qualified, expert radiation therapy
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "varian" in a bold, sans-serif font. The letters are all lowercase and evenly spaced. The color of the text is black, and the background is white. The word is centered in the image.
treatment planners. Results of Al Segmentation are utilized in the Eclipse Treatment Planning System where it is the responsibility of a qualified physician to further review, edit as needed, and approve each structure.
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The new device, referred to as the "subject device" throughout this summary, is release version v1.0 (version 1.0) of Al Segmentation. The subject device has some technological and functional similarity and differences with the predicate device, Smart Segmentation Knowledge Based Contouring version v2.2 (K141248),
At a high level, both the predicate device and the subject device are based on the same characteristics:
- Both devices are software-only medical devices. .
- 9 Both devices are intended for use by medical professionals within the context of supporting radiotherapy treatment planning.
- . Both devices contain automated segmentation algorithms used to process radiological images in order to generate contouring of structures for a variety of anatomical sites.
- . Both devices include review interfaces and tools for users to independently assess the output.
- . Both devices are compatible with the Eclipse Treatment Planning System, which is Varian's radiotherapy treatment planning software.
The significant differences in the subject device compared with the predicate device are:
-
- Use of Al-based algorithms for automated segmentation and contouring
- Note: These algorithms are static and non-adaptive; they do not alter their behavior over time a. based on user input.
-
- Cloud deployment and hosting of software components and algorithms
-
- Absence of manual editing or contouring tools
PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination.
Software Verification and Validation Testing
Software verification and validation testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern.
Test results demonstrate conformance to applicable requirements and specifications. No animal studies or clinical tests have been included in this pre-market submission.
Standards Conformance
The subject device conforms in whole or in part with the following standards that address software development, safety, and usability:
- . IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes
{5}------------------------------------------------
orign
- . IEC 62366-1 Edition 1.0 2015-02 Application of usability engineering to medical devices
- . IEC 62083 Edition 2.0 2009-09 Requirements for the safety of radiotherapy treatment planning systems
- IEC 82304-1 Edition 1.0 2016-10 Health software Part 1: General requirements for product safety .
Argument for substantial equivalence to the predicate device
A subset of software features and characteristics of the subject device are different from the predicate device. However, Varian considers these differences to be enhancements of the predicate, while the principle of operation of the subject device is the same as that of the existing predicate device. Verification and validation testing demonstrate that the subject device performs its intended use as designed through the product's functional, usability, and safety requirements. Varian therefore believes that the subject device is substantially equivalent to the predicate device.
CONCLUSION
The predicate device was cleared based only on non-clinical testing, and no animal or clinical studies were performed for the subject device. The non-clinical data supports the safety of the device, and verification and validation testing demonstrate that the subject device should perform as intended in the specified use conditions. There were no remaining discrepancy reports (DRs) which could be classified as Safety or Customer Intolerable.
Therefore, Varian considers AI Segmentation to be substantially equivalent to the predicate device, Smart Segmentation Knowledge Based Contouring (K141248).
N/A