K Number
K203396

Validate with FDA (Live)

Device Name
DEKA SMARTPERIO
Date Cleared
2021-08-23

(278 days)

Product Code
Regulation Number
878.4810
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DEKA SmartPerio system is intended to perform the following types of intraoral procedures: soft tissue dental, general, oral maxillofacial, and cosmetic surgery. The DEKA SmartPerio is intended for ablating, incising, excising, vaporization and coagulation of soft tissues using a contact fiber-optic delivery system. The device is intended to be used in dentistry. The following are the oropharyngeal indications for use: Abscess Incision and Drainage Apthous Ulcers Treatment Biopsies Excision and Incision Crown lengthening Hemostatic assistance Fibroma Removal Frenectomy Frenotomy Gingival Incision and Excision Gingivectomy Gingivoplasty Operculectomy Oral Papillectomy Tissue retraction for Impression Vestibuloplasty. Selective ablation of enamel (first degree) caries Exposure of unerupted / partially erupted teeth Implant recovery Lesion (tumor) removal Leukoplakia Pulpotomy Pulpotomy as adjunct to root canal therapy Removal of filling material such as gutta percha or resin as adjunct treatment during root canal retreatment Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment level and tooth mobility tooth mobility. Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium.)

Device Description

The DEKA SmartPerio is a Nd:YAG laser device for soft-tissues intraoral treatments. The DEKA SmartPerio system delivers laser through an optical fiber that is guided to the target tissue with the aid of an handpiece and a tip. The DEKA Smartperio device consists of: An AC/DC power supply unit CPU controller LASER source Cooling system User interface with LCD touch screen Beam delivery system Laser activation is controlled by footswitch.

AI/ML Overview

The provided document is a 510(k) summary for the DEKA SmartPerio laser device. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving the device meets specific acceptance criteria with quantifiable performance metrics.

Therefore, much of the requested information cannot be extracted directly from this document. However, I can provide what is available, noting where information is absent.

1. A table of acceptance criteria and the reported device performance

This document does not specify quantitative acceptance criteria typically seen in performance studies for new medical devices, nor does it report specific device performance metrics in detail beyond a comparison of technical specifications with a predicate device.

Instead, the document's primary "acceptance criteria" appear to be meeting electrical safety and electromagnetic compliance standards, as well as demonstrating substantial equivalence to a legally marketed predicate device.

Acceptance Criteria (Inferred from regulatory context)Reported Device Performance (Summary from comparison)
Substantial Equivalence:
- Indications for useDEKA SmartPerio's indications are a subset of the predicate device's indications.
- Technological Characteristics (Laser Wavelength, Energy per Pulse, Pulses per Second, Pulse duration, Average Power, Aiming Beam wavelength, Aiming beam power, Delivery system, Power Requirements, Weight, Dimensions)Most are identical or similar, with differences noted as not affecting safety and effectiveness.
Safety Standards Compliance:
- AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, cl:2009/(R)2012 and a2:2010/(R)2012 (Medical Electrical Equipment)Compliant
- IEC 60601-1-2 Ed. 4 (Electromagnetic disturbance)Compliant
- IEC60601-2-22 Ed 3.1 (Laser equipment safety)Compliant
- IEC 60825-1 Ed. 3.0 (Safety of laser products)Compliant
Biocompatibility:Subject device relies on biocompatibility of already FDA-cleared accessories.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

The document explicitly states "Clinical Performance Data: None" and "Non-Clinical Performance Data" only refers to compliance with electrical and laser safety standards. This implies there was no "test set" of patient data in the sense of a clinical trial or performance study. The data provenance and sample size for such a test set are therefore not applicable or not provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

Not applicable, as no clinical test set or ground truth adjudication using experts is mentioned.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable, as no clinical test set or ground truth adjudication is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The DEKA SmartPerio is a laser surgical instrument, not an AI-powered diagnostic or assistive device that would involve human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. The DEKA SmartPerio is a physical medical device (laser surgical instrument), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable, as no clinical performance study requiring a "ground truth" (e.g., for diagnostic accuracy) was performed or reported.

8. The sample size for the training set

Not applicable, as the DEKA SmartPerio is a physical laser device and not an AI or machine learning algorithm that requires a "training set."

9. How the ground truth for the training set was established

Not applicable, as the DEKA SmartPerio is a physical laser device and not an AI or machine learning algorithm.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 23, 2021

El.En. Electronic Engineering Spa Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, FI 50041 Italy

Re: K203396

Trade/Device Name: Deka Smartperio Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: NVK, GEX Dated: August 2, 2021 Received: August 6, 2021

Dear Paolo Peruzzi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K203396

Device Name DEKA SMARTPERIO

Indications for Use (Describe)The DEKA SmartPerio system is intended to perform the following types of intraoral procedures: soft tissue dental, general, oral maxillofacial, and cosmetic surgery. The DEKA SmartPerio is intended for ablating, incising, excising, vaporization and coagulation of soft tissues using a contact fiber-optic delivery system. The device is intended to be used in dentistry. The following are the oropharyngeal indications for use: Abscess Incision and Drainage Apthous Ulcers Treatment Biopsies Excision and Incision Crown lengthening Hemostatic assistance Fibroma Removal Frenectomy Frenotomy Gingival Incision and Excision Gingivectomy Gingivoplasty Operculectomy Oral Papillectomy Tissue retraction for Impression Vestibuloplasty. Selective ablation of enamel (first degree) caries Exposure of unerupted / partially erupted teeth Implant recovery Lesion (tumor) removal Leukoplakia Pulpotomy Pulpotomy as adjunct to root canal therapy Removal of filling material such as gutta percha or resin as adjunct treatment during root canal retreatment Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment level and tooth mobility tooth mobility. Laser assisted new attachment procedure (cementum-mediated periodontal ligament new-attachment to the root surface in the absence of long junctional epithelium.)
Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(K) Summary

DEKA SMARTPERIO

Submitter:

El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy

Contact:

Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it

Date Summary Prepared:

August 2, 2021

Device Trade Name:

DEKA SMARTPERIO

Common Name:

Nd:YAG Pulsed Laser

Classification Name:

Laser surgical instrument for use in general and plastic surgery and in dermatology

Product Code :

GEX

Classification Number: 21 CFR 878.4810

Equivalent Device:

Millennium Dental Technologies, Inc : PerioLase Dental Laser (K030290) (Primary predicate)

Device Description:

The DEKA SmartPerio is a Nd:YAG laser device for soft-tissues intraoral treatments. The DEKA SmartPerio system delivers laser through an optical fiber that is guided to the target tissue with the aid of an handpiece and a tip.

The DEKA Smartperio device consist s of:

  • An AC/DC power supply unit
  • CPU controller

{5}------------------------------------------------

K203396

  • LASER source
  • . Cooling system
  • . User interface with LCD touch screen
  • . Beam delivery system

Laser activation is controlled by footswitch.

Electrical specifications are:

115V ~ single phase, 50/60 Hz, Absorbed electric power1840 VA (max)

Indications for Use:

The DEKA SmartPerio system is intended to perform the following types of intraoral procedures: soft tissue dental, general, oral maxillofacial, and cosmetic surgery. The DEKA SmartPerio is intended for ablating, incising, excising, vaporization and coagulation of soft tissues using a contact fiber-optic delivery system.

The device is intended to be used in dentistry.

The following are the oropharyngeal indications for use:

Abscess Incision and Drainage

Apthous Ulcers Treatment

Biopsies Excision and Incision

Crown lengthening

Hemostatic assistance

Fibroma Removal

Frenectomy

Frenotomy

Gingival Incision and Excision

Gingivectomy

Gingivoplasty

Operculectomy

Oral Papillectomy

Tissue retraction for Impression

Vestibuloplasty.

Selective ablation of enamel (first degree) caries

Exposure of unerupted / partially erupted teeth

Implant recovery

Lesion (tumor) removal

{6}------------------------------------------------

Leukoplakia

Pulpotomy

Pulpotomy as adjunct to root canal therapy

Removal of filling material such as gutta percha or resin as adjunct treatment during root canal retreatment Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment level and tooth mobility tooth mobility.

Laser assisted new attachment procedure (cementum-mediated periodontal ligament newattachment to the root surface in the absence of long junctional epithelium.)

Comparison with the predicate device:

DEKA SMARTPERIO is substantially equivalent to the Millennium Dental Technologies, Inc PerioLase Dental Laser System (K030290):

Proposed 510(k) DevicePrimary Predicate DeviceComparison
DeviceTradeNameDEKA SMARTPERIOK203396Millennium Dental Technologies,Inc : PerioLase Dental LaserK030290
Indications for useIntraoral procedures: soft tissuedental, general, oral maxillofacial,and cosmetic surgery.Intended for ablating, incising,excising, vaporization andcoagulation of soft tissues using acontact fiber-optic delivery system.The device is intended to be used indentistry.The following are the oropharyngealindications for use:Abscess Incision and DrainageApthous Ulcers TreatmentBiopsies Excision and IncisionCrown lengtheningIntraoral soft tissue dental, general,oral maxillo-facial and cosmeticsurgery.Intended for ablating, incising,excising, vaporization andcoagulation of soft tissues using acontact fiber optic delivery system.The device will be used in thefollowing areas: general andcosmetic dentistry, otolaryngology,arthroscopy, gastroenterology,general surgery, dermatology &plastic surgery, neurosurgery,gynecology, urology, ophthalmologyand pulmonary surgery.The following are the oropharyngealindications for use:Abscess Incision and DrainageApthous Ulcers TreatmentBiopsies Excision and IncisionCrown lengtheningIndications for use ofproposed device are a subset of indications for useof the predicate device.Proposed device is notintended forotolaryngology,arthroscopy,gastroenterology, generalsurgery, dermatology &plastic surgery,neurosurgery,gynecology,urology,ophthalmology andpulmonary surgery.
Hemostatic assistanceHemostatic assistance
Fibroma RemovalFibroma Removal
FrenectomyFrenectomy
FrenotomyFrenotomy
Gingival Incision and ExcisionGingival Incision and Excision
GingivectomyGingivectomy
GingivoplastyGingivoplasty
OperculectomyOperculectomy
Oral PapillectomyOral Papillectomy
Tissue retraction for ImpressionTissue retraction for Impression
Vestibuloplasty.Vestibuloplasty.
Selective ablation of enamel (firstdegree) cariesSelective ablation of enamel (firstdegree) caries
Exposure of unerupted / partiallyerupted teethExposure of unerupted / partiallyerupted teeth
Implant recoveryImplant recovery
Lesion (tumor) removalLesion (tumor) removal
LeukoplakiaLeukoplakia
PulpotomyPulpotomy
Pulpotomy as adjunct to root canaltherapyPulpotomy as adjunct to root canaltherapy
Removal of filling material such asgutta percha or resin as adjuncttreatment during root canalretreatment Sulcular debridement(removal of diseased or inflamed softtissue in the periodontal pocket) toimprove clinical indices includinggingival index, gingival bleedingindex, probe depth, attachment leveland tooth mobility tooth mobility.Removal of filling material such asgutta percha or resin as adjuncttreatment during root canalretreatment Sulcular debridement(removal of diseased or inflamed softtissue in the periodontal pocket) toimprove clinical indices includinggingival index, gingival bleedingindex, probe depth, attachment leveland tooth mobility tooth mobility.
Laser assisted new attachmentprocedure (cementum-mediatedperiodontal ligament new-attachment to the root surface in theabsence of long junctionalepithelium.)Laser assisted new attachmentprocedure (cementum-mediatedperiodontal ligament new-attachment to the root surface in theabsence of long junctionalepithelium.)

{7}------------------------------------------------

{8}------------------------------------------------

Product codeGEXGEXIdentical
Laser Wavelength1064 nm1064 nmIdentical
Energy per Pulse20-300 mJ20-300 mJIdentical
Pulses per Second10-100 Hz10-100 HzIdentical
Pulse duration100-650 µs100-650 µsIdentical
Average Power0.2-6W0.2-6WIdentical
AimingBeamwavelength635 nm635 nmIdentical
Power of Aimingbeam3 mW5mWSimilar, difference inaiming beam power doesnot affect safety andeffectiveness.
Delivery systemoptical fiber (cleared with K124003,K200234) guided to the target tissuewith the aid of handpiece and tip(cleared with K952006)optical fiber guided to the targettissue with the aid of handpiece andtipSimilar delivery systemaccessories. Subjectdevice uses already FDAcleared accessories, whilefor predicate device theyare included in the deviceFDA clearance. Thisdifference does not affectsafety and effectiveness.
PowerRequirements115Vac, 50/60 Hz, 16Amps110-120 VAC, 50/60 Hz, 8 Amps,Single phase; 200 - 240 VAC, 50 Hz, 4AmpsBoth devices had powerrequirements consistentwith US standards.Difference in power inputcurrent does not affectsafety and effectiveness.
Weight38 Kg45 lbs. [20.4 Kgs]/38 lbs. [17.2 Kgs]Difference in weight doesnot affect safety andeffectiveness.
Dimensions68x23x65cm11" (width) x 19" (depth) x 25"(height)Difference in dimensionsdoes not affect safetyand effectiveness.
BiocompatibilityYes- the only parts that can get incontact with patient have alreadybeen cleared by FDA (K124003,K200234, K952006)YES (predicate device cleared byFDA)Subject device relies onbiocompatibility ofaccessories alreadycleared by FDA; no dataavailable from predicatedevice

{9}------------------------------------------------

Clinical Performance Data:

None

Non-Clinical Performance Data:

The DEKA SMARTPERIO has been tested and found in compliance with following standards:

. AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012, cl:2009/(R)2012 and a2:2010/(R)2012 – Medical Electrical Equipment – Part 1: general requirements for basic safety and essential performance.

. IEC 60601-1-2 Ed. 4 - Medical electrical equipment - part 1-2: general requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbance – Requirements and tests.

. IEC60601-2-22 Ed 3.1 Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.

. IEC 60825-1 Ed. 3.0 – Safety of laser products – Part 1 : Equipment classification, and requirements.

Conclusion:

Based on the comparison of indications for use and the technological characteristics, and on the outcome of non-clinical performance data provided , we can conclude that the DEKA SMARTPERIO is as safe, as effective, and performs as well as the legally marketed predicate device.

Additional Information:

None.

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.