K Number
K203388

Validate with FDA (Live)

Date Cleared
2021-06-04

(198 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable medical masks are intended for use by healthcare workers during procedures to protect both patients and healthcare workers against transfer of microorganisms, bodily fluids, and particulate materials. This device is single-use and provided non-sterile.

Device Description

The Disposable medical mask is flat pleated type mask, utilizing ear loops way for wearing, and they all have nose clip design for fitting the facemask around the nose. The Disposable medical mask is only one model: WSKZ-001 and the color is blue. The Disposable medical mask is manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, only the outer layers' color is blue (colorant: Pigment Blue, CAS number: 147-14-8), and the middle layer is made of melt blown fabric. Ear loops, which is held to cover the users' mouth and nose by two Ammonium nylon elastic bands ultrasonic welded to the Disposable Medical Mask. The elastic ear loops are not made with natural rubber latex. The nose clip contained in the Disposable medical mask is in the middle layer of Disposable medical mask to allow the user to fit the Disposable mask around their noses, which is made of Polypropylene and metallic iron. The Disposable medical mask is sold non-sterile and is intended to be single use, and the intended environment is professional healthcare facility environment. The shelf life of each Disposable medical mask is 2 years. The dimensions of each Disposable medical mask are length 175mm and width 95mm. The dimension of nose clip is length 100mm, and the ear loop is length 170mm.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for a medical mask (K203388) seeking FDA clearance, primarily through comparison to a predicate device (K153409).

Here's the breakdown of the information requested:

1. A table of acceptance criteria and the reported device performance

Test ItemAcceptance CriteriaReported Device Performance
Bacterial Filtration Efficiency≥ 98%≥99.2% (Passed 32/32)
Differential Pressure (Delta-P)<5.0 mm H2O/cm²<4.0 mm H2O/cm² (Passed 32/32)
Particulate Filtration Efficiency≥ 95%≥96.1% (Passed 32/32)
Resistance to penetration by synthetic bloodFluid resistant claimed at 80 mmHg80 mmHg (Passed 32/32)
Flame SpreadClass 1≥3 Seconds burn Time-Class 1 (Passed 32/32)
In vitro Cytotoxicity TestNon-cytotoxicNon-cytotoxic (Pass)
Skin Sensitization TestNon-sensitizingNon-sensitizing (Pass)
Skin Irritation TestNon-irritatingNon-irritating (Pass)

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for performance tests: 32 samples were used for each of the bacterial filtration efficiency, differential pressure, particulate filtration efficiency, resistance to synthetic blood, and flame spread tests.
  • Sample Size for biocompatibility tests: The document refers to "the subject device extract" for cytotoxicity, irritation, and sensitization tests, implying a sufficient sample was prepared according to the respective ISO standards. No specific number of devices or extracts is provided.
  • Data Provenance: The document does not explicitly state the country of origin of the testing data or whether it was retrospective or prospective. However, the submitter, Everwin Toys (Dongguan)., Ltd., is located in China, and the application correspondent is in Guangzhou, China. The testing standards (ASTM, EN, ISO) are international.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This device is a medical mask and the evaluation involves objective performance metrics (e.g., filtration efficiency, pressure, flammability) and biocompatibility, not diagnostic interpretation. Therefore, there are no "experts" in the sense of clinicians or radiologists establishing a "ground truth" for a test set. The ground truth is established by the results of standardized laboratory tests by qualified personnel following the specified ASTM, EN, and ISO methodologies.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. The evaluation is based on objective laboratory measurements against predefined thresholds, not subjective interpretation requiring adjudication among experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical mask, not an AI-powered diagnostic device. No human reader or AI assistance studies were performed or are relevant to this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical mask, not an algorithm or software device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

The ground truth for the performance and biocompatibility tests is established by the results obtained from adherence to recognized international standards (ASTM F2101, EN 14683, ASTM F2299, ASTM F1862, 16 CFR Part 1610, ISO 10993-5, ISO 10993-10). These are objective, quantitative measurements and qualitative assessments (e.g., non-cytotoxic) conducted in a laboratory setting.

8. The sample size for the training set

Not applicable. This is a physical medical device (mask), not an AI algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As
described above, this is a medical mask and does not involve AI algorithms or training sets.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 4, 2021

Everwin Toys (Dongguan)., Ltd. % Cassie Lee Manager Share Info (Guangzhou) Medical Consultant Ltd. No. 1919-1920. Building D3, Miniie Plaza, Shuixi Road, Huangpu District Guangzhou, Guangdong 510700 China

Re: K203388

Trade/Device Name: Disposable medical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: April 25, 2021 Received: May 5, 2021

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence Murray III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203388

Device Name Disposable medical mask

Indications for Use (Describe)

The Disposable medical masks are intended for use by healthcare workers to procedures to protect both patients and healthcare workers against transfer of microorganisms, bodily fluids, and particulate materials. This device is single-use and provided non-sterile.

Model: WSKZ-001 Color : Blue

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary for K203388

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

1. Date of the summary prepared: June 04, 2021

2. Submitter's Information

510(k) Owner's Name: EVERWIN TOYS (DONGGUAN)., LTD. Establishment Registration Number: 3012360847 Address: Xiekeng Village,Qingxi Town ,Dongguan City ,Guangdong Province , China Post Code : 523000 Contact name: Huang Zibin Tel: +86 0769-87314418-6118 Phone: +86 13790595177 E-mail: zhaogang@wahshing.com

Application Correspondent :

Contact Person : Ms. Cassie Lee Share Info (Guangzhou) Medical Consultant Ltd. Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China Tel : +86 20 8266 2446 Email : regulatory@glomed-info.com

3. Subject Device Information

Type of 510(k): Traditional Classification Name: Mask, Surgical Common name: Surgical Mask Trade Name: Disposable medical mask Model: WSKZ-001 Review Panel: General Hospital Product Code: FXX Regulation Number: 21 CFR 878.4040 Regulatory Class: II

Predicate Device Information 4.

Sponsor: Protect U Guard, LLC Device Name: Protect U Guard Earloop and Tie-On Mask (Blue, White or Green) Model: P10031, P10032, P10033, P10041, P10042, P10043 Classification Name: Mask, Surgical 510(K) Number: K153409 Review Panel: General Hospital Product Code: FXX Requlation Number: 21 CFR 878.4040 Regulation Class: II

5. Indications for Use

The Disposable medical masks are intended for use by healthcare workers during procedures to protect both patients and healthcare workers against transfer of microorganisms, bodily fluids, and particulate materials. This device is single-use and provided non-sterile.

Model: WSKZ-001 Color: Blue

{4}------------------------------------------------

The Disposable medical mask is flat pleated type mask, utilizing ear loops way for wearing, and they all have nose clip design for fitting the facemask around the nose. The Disposable medical mask is only one model: WSKZ-001 and the color is blue.

The Disposable medical mask is manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, only the outer layers' color is blue (colorant: Pigment Blue, CAS number: 147-14-8), and the middle layer is made of melt blown fabric.

Ear loops, which is held to cover the users' mouth and nose by two Ammonium nylon elastic bands ultrasonic welded to the Disposable Medical Mask. The elastic ear loops are not made with natural rubber latex.

The nose clip contained in the Disposable medical mask is in the middle layer of Disposable medical mask to allow the user to fit the Disposable mask around their noses, which is made of Polypropylene and metallic iron.

The Disposable medical mask is sold non-sterile and is intended to be single use, and the intended environment is professional healthcare facility environment. The shelf life of each Disposable medical mask is 2 years.

The dimensions of each Disposable medical mask are length 175mm and width 95mm. The dimension of nose clip is length 100mm, and the ear loop is length 170mm.

7. Comparison of Technological Characteristics

The following table identifies technological characteristics shared between the subject device and Predicate:

Elements ofComparisonSubject Device (K203388)Predicate Device (K153409)Comparison
CompanyEVERWIN TOYS (DONGGUAN)., LTD.Protect U Guard, LLC--
510 (k)K203388K153409--
Trade NameDisposable medical maskProtect U Guard Earloop and Tie-On Mask (Blue, White or Green)--
ModelWSKZ-001Earloop Models: Blue (P10031),White (P10032), or Green(P10033)Tie-On Models: Blue (P10041),White (P10042), or Green(P10043)--
ClassificationNameMask, SurgicalMask, SurgicalSame
ClassificationClass II Device, FXX(21 CFR878.4040)Class II Device, FXX (21CFR878.4040)Same
Intended useThe Disposable medical masks areintended for use by healthcareworkers during procedures to protectboth patients and healthcare workersagainst transfer of microorganisms,bodily fluids, and particulatematerials. This device is single-useand provided non-sterile.Earloop Mask and Tie-On Maskis intended for use by healthcareworkers during procedures toprotect both patients andhealthcare workers againsttransfer of microorganisms,bodily fluids, and airborneparticles. This device is single-use and provided non-sterile.Earloop Models: Blue (P10031),White (P10032), or Green(P10033)Tie-On Models: Blue (P10041),White (P10042), or Green(P10043)Over-The-Counter UseSimilar
Material
Elements ofComparisonSubject Device (K203388)Predicate Device (K153409)Comparison
Outer facinglayerSpunbond PolypropyleneSpunbond PolypropyleneSame
Middle layerMelt blown polypropyleneMelt blown polypropyleneSame
Inner facinglayerSpunbond PolypropyleneSpunbond PolypropyleneSame
Nose clipPolypropylene and metallic ironAluminum stripSimilarNote 1
Ear loopsAmmonium nylon elasticUrethane elastic fiberSimilarNote 1
DesignfeaturesColor: BlueEar loopsColor: Blue, White,Green Ear loopsSame
Dimension17.5cm x 9.5 cm17.7cm x 9.5 cmSimilarNote 1
OTC useYesYesSame
SterilityNon-SterileNon-SterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
PerformanceTestingLevel 1Level 1Same
FluidResistancePerformancePass at 80 mmHgPass at 80 mmHgSame
ParticulateFiltrationEfficiency≥96.1%99.18% at 0.1 micronSimilarNote 2
BacterialFiltrationEfficiency≥99.2%99.17%SimilarNote 2
DifferentialPressure<4.0 mm H2O/cm23.79 mm H2O/cm²SimilarNote 2
FlammabilityClass 1Class 1Same
Biocompatibility
CytotoxicityUnder the conditions of the study, thesubject device extract was determinedto be non-cytotoxic.Under the conditions of thestudy, the subject deviceextract wasdetermined to be non-cytotoxic.Same
IrritationUnder the conditions of the study, thesubject device non-polar and polarextracts were determined to be non-irritating.Under the conditions of thestudy, the subject device non-polar and polar extracts weredetermined to be non-irritating.Same
SensitizationUnder the conditions of the study, thesubject device non-polar and polarextracts were determined to be non-sensitizing.Under the conditions of thestudy, the subject device non-polar and polar extracts weredetermined to be non-sensitizing.Same

{5}------------------------------------------------

{6}------------------------------------------------

Comparison in Detail(s):

Note 1:

Although the "Ear loops", "Nose clip" and "Dimension" of subject device is a little different from predicate device, they all meet the requirements of essential performance standard ASTM F2100 and ISO 10993 series. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.

Note 2:

Although the "Particulate Filtration Efficiency", "Bacterial Filtration Efficiency" and "Differential Pressure" of subject device is a little different from predicate device, they all meet the level 1 requirements of essential performance standard ASTM F2100. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.

8. Non-Clinical Performance Testing Summary:

Test item(PerformanceLevel 1)Test methodPasscriteriaTest results
Bacterial filtrationefficiencyASTM F2101-14 StandardTest Method for Evaluating theBacterial Filtration Efficiency(BFE) of Medical Face MaskMaterials, Using a BiologicalAerosol of Staphylococcusaureus according to ASTMF2100:2019≥ 98%32/32 Passed at ≥99.2%/ Pass
Differentialpressure (Delta-P)EN 14683: 2019, Annex CMedical face masks -Requirements and testmethods according to ASTMF2100:2019<5.0 mmH2O/cm²32/32 Passed at <4.0mm H2O/cm²/ Pass
ParticulateFiltrationEfficiencyASTM F2299-03 StandardTest Method for Determiningthe Initial Efficiency ofMaterials Used in MedicalFace Masks to Penetration byParticulates Using LatexSpheres according to ASTMF2100:2019≥ 95%32/32 Passed at ≥96.1%/ Pass

{7}------------------------------------------------

Resistance topenetration bysyntheticblood, minimumpressure inmmHgfor pass resultASTM F1862/F1862M-17Standard Test Method forResistance of Medical FaceMasks to Penetration bySynthetic Blood (HorizontalProjection of Fixed Volume ata Known Velocity) accordingto ASTM F2100:2019Fluidresistantclaimedat 80mmHg32/32 Passed at 80mmHg/ Pass
Flame spread16 CFR Part 1610 Standardfor the Flammability ofClothing according to ASTMF2100:2019Class 132/32 Passed ≥3Seconds burnTime-Class 1 / Pass

Biocompatibility Testing Summary:

According to ISO 10993-1: 2018, the nature of body contact for the subject device is the Surface Device category, Skin Contact, and duration of the contact is A-Limited (<24 h). The following tests for the subject device were conducted to demonstrate that the subject device is biocompatible and safe for its intended use:

Title of the testPurpose of the testThe source of references (Test method)Acceptance criteriaTest results
In vitro Cytotoxicity TestUnder the research conditions, determine whether the target device extract is cytotoxic.ISO 10993-5:2009Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicityUnder the conditions of the study, the subject device extract was determined to be non-cytotoxic.Pass
Skin Sensitization TestUnder the research conditions, determine whether the non-polar and polar extracts of the target device are sensitive.ISO 10993-10:2010Biological evaluation of medical devices— Part 10: Tests for irritation and skin sensitizationUnder the conditions of the study, the subject device non-polar and polar extracts were determined to be non-sensitizing.Pass
Skin Irritation TestUnder the research conditions, determine whether the non-polar and polar extracts of the target device are irritating.ISO 10993-10:2010Biological evaluation of medical devices— Part 10: Tests for irritation and skin sensitizationUnder the conditions of the study, the subject device non-polar and polar extracts were determined to be non-irritating.Pass

9. Summary of Clinical Performance Test

No clinical study is included in this submission.

10. Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the subject device, Disposable medical mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device, K153409, Protect U Guard Earloop and Tie-On Mask (Blue, White, or Green).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.