K Number
K203350

Validate with FDA (Live)

Device Name
ENDOtube
Date Cleared
2021-09-29

(320 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ENDOtube devices are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/ or air to an endoscope during gastrointestinal endoscopic procedure. These devices are intended to be used by healtheare professional only. In procedure room, all ENDOtube devices are 24 hours use disposable products except the one-way valves which are single use products and the insufflation connectors which are multiple patient use.

Device Description

The ENDOtube kits are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/or air to an endoscope during a GI endoscopic procedure. The ENDOtube products are compatible with various endoscopes (FujiFilm® 500/600 series and 700 series, OLYMPUS® 140, 160, 180 and 190 and PENTAX®), and irrigation pumps (OLYMPUS®, Erbe® and EMED®). The ENDOtube products include several families: ENDOtube for Wash Bottle, ENDOtube for Insufflation, ENDOtube for Wash Pump, and ENDOtube for Wash Bottle CO2 Insufflation.

AI/ML Overview

Based on the provided text, the device in question, ENDOtube, is an endoscopic tubing kit, not an AI/ML-driven device. Therefore, the information requested about acceptance criteria and studies that prove the device meets those criteria (especially points related to AI/ML, such as ground truth, expert adjudication, MRMC studies, and training/test sets) is not applicable to this document.

The document describes the submission to the FDA for 510(k) premarket notification for a medical device (ENDOtube) which is a combination of tubes, fittings, and valves. The acceptance criteria and the studies performed are related to the performance, safety, and compatibility of this physical medical device, not software or AI/ML algorithms.

Here's an analysis of the "acceptance criteria" and "proof" based on the provided text, adapted to the nature of the device:

1. Acceptance Criteria and Reported Device Performance (Non-AI/ML)

The "acceptance criteria" for this device are primarily demonstrated through non-clinical testing to ensure its functionality, safety, and compatibility with other medical equipment. The document reports "Pass" for all tested parameters, indicating the device meets the established criteria.

Parameter/Acceptance Criteria (Implicit)Reported Device Performance
Endoscope Column Compatibility:
OLYMPUS® GI endoscopes 140, 160, 180, 190 compatibilityPass
PENTAX® GI endoscopes compatibilityPass
FUJIFILM® GI 500/600 series and 700 series endoscopes compatibilityPass
Irrigation Peristaltic Pump Compatibility:
Olympus®, Erbe®, EMED® pump compatibilityPass
Insufflator Compatibility:
OLYMPUS® UCR, FUJIFILM® GW-1 and GW-100 compatibilityPass
Universal Water Bottle Cap:
Presence and effectiveness with common sterile water bottle brandsPass
Fluid Supply:
Sterile water; Air; CO2 supply (leak test and usability)Pass
One-way valve (for irrigation pump compatible device):
Presence and effectiveness (cross-contamination test)Pass
Tube Use Period:
Multi-patient use, without reprocessing, up to 24 hours (durability)Pass
One-way Valve Use Period:
Single use, up to 2 hours (operational)Pass
Device Shelf-Life:
2 years performancePass
Insufflation Connector Use Period:
Multi-patient use (operational)Pass
Insufflation Connector Processing:
Reprocessable (performance)Pass
Tube and Valve Sterilization:
EO sterilization validationPass
Main Materials:
Biocompatibility (AAMI/ISO 10993)Pass
Inert metallic materials, elastomers, thermoplastic polymer (biocompatibility)Pass
No DEHP (chemical characterization)Pass

2. Sample Size and Data Provenance for Test Set:

  • Sample Size: The document does not specify exact sample sizes for each "Connection and usability testing," "Leak test and usability testing," "Cross-contamination test," "Durability test," "Operational test," "Performance test," "Sterilization validation," or "Biocompatibility testing." These are typically standardized tests performed on a representative number of units or materials according to relevant standards (e.g., ISO, AAMI).
  • Data Provenance: The tests are non-clinical (laboratory/bench) tests. The manufacturer, Nissha Medical Technologies SAS, is based in France. The tests would likely have been conducted in controlled laboratory environments. The data is prospective in the sense that it's generated for the purpose of the 510(k) submission.

3. Number of Experts and Qualifications for Ground Truth:

  • Not applicable. This is a non-AI/ML device. "Ground truth" in the context of device performance refers to verified operational parameters and safety compliance, which are established through adherence to engineering specifications and regulatory standards, not expert medical opinion on image interpretation.

4. Adjudication Method for Test Set:

  • Not applicable. As this is not an AI/ML diagnostic device, there is no "adjudication" of expert interpretations. Device performance is objectively measured against engineering and safety specifications.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Not applicable. This type of study is relevant for AI/ML diagnostic aids that assist human readers in interpreting medical images. The ENDOtube is a physical medical accessory.

6. Standalone (Algorithm Only) Performance Study:

  • Not applicable. The ENDOtube is a physical device, not an algorithm.

7. Type of Ground Truth Used:

  • Not applicable in the AI/ML sense. The "ground truth" for this device's performance is established by engineering specifications, relevant industry standards (e.g., ISO 10993 for biocompatibility), and regulatory requirements. Functional tests confirm the device operates as intended (e.g., no leaks, proper connections, durability).

8. Sample Size for Training Set:

  • Not applicable. This device does not involve a training set for an AI/ML algorithm.

9. How Ground Truth for Training Set Was Established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.

September 29, 2021

Nissha Medical Technologies SAS Roch Mignot Ouality & Regulatory Affairs Manager ZA Des Boutries 12 Rue Des Cayennes Conflans STE Honorine, Yvelines 78700 FRANCE

Re: K203350

Trade/Device Name: ENDOtube Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OCX Dated: August 12, 2021 Received: August 19, 2021

Dear Roch Mignot:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203350

Device Name ENDOtube

Indications for Use (Describe)

The ENDOtube devices are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/ or air to an endoscope during gastrointestinal endoscopic procedure. These devices are intended to be used by healtheare professional only. In procedure room, all ENDOtube devices are 24 hours use disposable products except the one-way valves which are single use products and the insufflation connectors which are multiple patient use.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for "NISSHA MEDICAL TECHNOLOGIES". The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" written in smaller, light gray letters. The logo is simple and modern.

SUBMITTER l.

NISSHA MEDICAL TECHNOLOGIES SAS

Z.A. DES BOUTRIES 12, RUE DES CAYENNES B.P. 310 78703 CONFLANS SAINTE HONORINE CEDEX France

Phone: +33 1 39 72 66 66 Fax: +33 1 39 72 61 61

Under registration number: 3004762958

A division of NISSHA MEDICAL TECHNOLOGIES

400 EXCHANGE STREET BUFFALO, NY 14204 United States of America Tel: 716.853.7500

Contact Person:

Roch MIGNOT, Quality and Requlatory Affairs Manager - France Tel: +33 1 39 72 77 - Cellphone: +33 7 85 19 39 07 Email: rmignot@nisshamedical.com

Date Prepared: October 30, 2020 Date Revised: December 30, 2020 Date Latest Reviewed: August 03, 2021

II. DEVICE

Device / Property NameENDOtube
Common / Usual NameEndoscopic tubing kit
Classification / Regulation Number 21CFR876.1500
Classification NameEndoscope and accessories
Product CodeOCX
Product Code NameEndoscopic Irrigation/Suction System
Regulatory ClassII
Device PanelGastroenterology/Urology

The ENDOtube products include several families as detailed bellow in VOL_002_Submission Context in the file 005_Product List and bellow.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" in smaller, gray letters. The logo is simple and modern, and the colors are clean and professional.

FamilyReferenceDescription
ENDOtube for Wash BottleS-FLO000001Sterile disposable tubing - Olympus Gl endoscopes
S-FLP000002Sterile disposable tubing - Pentax Gl endoscopes
S-FLF000003Sterile disposable tubing - Fujifilm Gl endoscopes 5-600 series
S-FLF000004Sterile disposable tubing - Fujifilm Gl endoscopes 700 series
ENDOtube for Wash BottleS-FLO001CO2Sterile disposable tubing + CO2 option - Olympus Gl endoscopes
ENDOtube for Wash BottleS-FLP002CO2Sterile disposable tubing + CO2 option - Pentax Gl endoscopes
ENDOtube for Wash BottleS-FLF003CO2Sterile disposable tubing + CO2 option - Fujifilm Gl endoscopes 5-600 series
ENDOtube for Wash BottleS-FLF004CO2Sterile disposable tubing + CO2 option - Fujifilm Gl endoscopes 700 series
ENDOtube for InsufflationS-TUBCO2LUERCO2 insufflation sterile tubing with hydrophobic filter - Luer connection
S-TUBCO2FUJICO2 insufflation sterile tubing with hydrophobic filter - Adapter connection
04628Connection adapter for insufflator FUJIFILM GW-1
ENDOtube for Insufflation04700Connection adapter for insufflator OLYMPUS UCR
ENDOtube for Insufflation04712Connection adapter for insufflator FUJIFILM GW-100
ENDOtube for Wash PumpS-PLO000100Sterile disposable irrigation tubing - Emed pump/Olympus Gl endoscopes
S-PLO000300Sterile disposable irrigation tubing - Olympus pump/Olympus Gl endoscopes
S-PLO000400Sterile disposable irrigation tubing - Endogator pump/Olympus Gl endoscopes
S-PLO000600Sterile disposable irrigation tubing - Erbe pump/Olympus Gl endoscopes
S-PLPPLF200Sterile disposable irrigation tubing - Emed pump/Pentax-Fujifilm Gl endoscopes
S-PLPPLF500Sterile disposable irrigation tubing - Endogator pump/Pentax-Fujifilm Glendoscopes
S-PLPPLF700Sterile disposable irrigation tubing - Erbe pump/Pentax-Fujifilm Gl endoscopes
ENDOtube for Wash PumpS-100100Sterile single use one-way water jet connector for Olympus
ENDOtube for Wash PumpS-100300Sterile single use one-way water jet connector for Fujifilm
ENDOtube for Wash BottleCO2 InsufflationS-FLOCO2LUERSterile ENDOtube for Wash Bottle with CO2 tubing - for Olympus/Luer InsufflatorConnection
S-FLOCO2FUJISterile ENDOtube for Wash Bottle with CO2 tubing - for Olympus/AdapterInsufflator Connection
S-FLPCO2LUERSterile ENDOtube for Wash Bottle with CO2 tubing - for Pentax/Luer InsufflatorConnection
S-FLPCO2FUJISterile ENDOtube for Wash Bottle with CO2 tubing - for Pentax/AdapterInsufflator Connection
S-FLFCO2LUERSterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 5-600/LuerInsufflator Connection
S-FLFCO2FUJISterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 5-600/AdapterInsufflator Connection
S-FLF7CO2LUERSterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 700/LuerInsufflator Connection
S-FLF7CO2FUJISterile ENDOtube for Wash Bottle with CO2 tubing - for Fujifilm 700/AdapterInsufflator Connection

PREDICATE DEVICE lll.

In order to properly cover the entire product range of the application, two predicates have been selected. The primary predicate is ENDOGATOR™ (K092429).

Device NameENDOGATOR™ENDO SMART CAP™
510(k) NumberK092429K093665
ClassIIII
Indications For UseThe ENDOGATOR™ system (tubingand accessories to accommodatevarious endoscopes and irrigationpumps) is intended to provideirrigation via sterile water supplyduring GI endoscopic procedureswhen used in conjunction with anirrigation pump (or cautery unit).The ENDO SMARTCAP™ isintended to be used with anair or CO2 source and/or apump along with a sterilewater source to supply air orCO2 and sterile water to anendoscope during endoscopicprocedures.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA", the words "MEDICAL TECHNOLOGIES" are written in smaller, gray letters. The logo is simple and modern.

IV. Device Description

IV.1. General

The ENDOtube kits are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/or air to an endoscope during a GI endoscopic procedure.

Visual description of ENDOtube products in the endoscopic column can be found in the Figure 1.

Image /page/5/Figure/7 description: The image shows a medical device with different tubes and components. There are three tubes labeled "Endo tube for insufflation", "Endo tube for wash bottle", and "Endo tube for wash pump". The tubes are connected to the medical device, which includes a monitor, a control panel, and a pump.

Figure 1: Visual description of the ENDOtube

The ENDOtube products are compatible with various endoscopes (FujiFilm® 500/600 series and 700 series, OLYMPUS® 140, 160, 180 and 190 and PENTAX®), and irrigation pumps (OLYMPUS®, Erbe® and EMED®).

IV.2. ENDOtube for Wash Bottle

The ENDOtube for Wash Bottle kits are a combination of tubes, fittings and valves intended to provide sterile water, air or CO2 to an endoscope during a GI endoscopic procedure.

The product is composed of at least two tubes bounded to a hermetic universal cap. The cap is screwed on a sterile water bottle. One tube is connected to the air output of the endoscopic column, and drive air into the bottle to put the sterile water bottle under is therefore forced into a second tube that is connected to the air/water channel of the endoscope. The device can also provide directly air or CO2 to the endoscope through another channel.

IV.3. ENDOtube for Wash Pump

The ENDOtube for Wash Pump kits are a combination of tubes, fittings and valves intended to provide sterile water to an endoscope through a peristaltic pump during a GI endoscopic procedure.

The product is composed of one or more tubes bounded one after another. On one extremity, the tube is inserted into sterile water bottle. On the other extremity the tube is connected to the water-jet channel of the endoscope. The central part is inserted into the pumping part of a peristaltic pump.

{6}------------------------------------------------

The ENDOtube for Wash Pump family contains also two independent single use one-way valves, used as an interface to connect the ENDOtube for Wash Pump tube to the endoscope waterjet channel.

ENDOtube for Insufflation IV.4.

The ENDOtube for Insufflation kits are a combination of tubes, fittings and filters intended to provide CO2 or air to an endoscope during a GI endoscopic procedure.

Form on side, the tube is connected to an independent CO2 or air supply. On the other side, it's connected to an ENDOtube for Wash Bottle. The product drive air or CO2 to put the sterile water bottle under pressure, to power it through the ENDOtube for Wash Bottle to the endoscope.

The ENDOtube for Insufflation family contains also three independent insufflation connectors used to adapt the ENDOtube for Insufflation tube to common independent CO2 or air supply devices.

IV.5. ENDOtube for Wash Bottle CO2 Insufflation

The ENDOtube for Wash Bottle CO2 Insufflation kits are a combination of tubes, fittings, valves and filters intended to provide sterile water and/or CO2 and/or air to an endoscope during a GI endoscopic procedure.

It's the combination of an ENDOtube for Wash Pump and an ENDOtube for Insufflation, bounded through a valve.

Indications for Use V.

The ENDOtube devices are a combination of tubes, fittings and valves intended to provide sterile water and/or CO2 and/or air to an endoscope during gastrointestinal endoscopic procedure. These devices are intended to be used by healthcare professional only. In procedure room, all ENDOtube devices are 24 hours use disposable products except the one-way valves which are single use products and the insufflation connectors which are multiple patient use.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is in large, blue, sans-serif font. Below that, in a smaller, gray, sans-serif font, are the words "MEDICAL TECHNOLOGIES."

VI. Technological characteristics comparison

Trade NameENDOtube (subjectdevice)ENDOGATOR™(primary predicate)ENDO SMARTCAP™
510(k) numberK203350K092429K093665
Intended UseEndoscopic AccessoriesEndoscopic AccessoriesEndoscopic Accessories
Endoscope columncompatibilityOLYMPUS® GIendoscopes 140, 160,180 and 190OLYMPUS® GIendoscopes 140, 160,180 and 190OLYMPUS® GIendoscopes 140, 160and 180
PENTAX® GI endoscopesPENTAX® GI endoscopesPENTAX® GI endoscopes
FUJIFILM® GI 500/600series and 700 seriesendoscopesN/AFUJIFILM® GI 500/600series and 700 seriesendoscopes
OLYMPUS®, Erbe®,EMED®OLYMPUS®, Erbe®,EMED®OLYMPUS®, Erbe®,EMED®
Irrigation peristaticpump compatibilityOLYMPUS®, Erbe®,EMED®OLYMPUS®, Erbe®,EMED®OLYMPUS®, Erbe®,EMED®
Insufflator compatibilityOLYMPUS® UCR,FUJIFILM® GW-1 andGW-100Not compatible withOLYMPUS® or FUJIFILM®GW-1 and GW-100N/A
Universal water bottlecapCommon sterile waterbottle brandsCommon sterile waterbottle brandsCommon sterile waterbottle brands
Fluid supplySterile water; Air; CO2Sterile waterSterile water; Air; CO2
One-way valve (forirrigation pumpcompatible device)PresentPresentPresent
Tube use periodMulti-patient use,without reprocessingUp to 24 hoursMulti-patient use,without reprocessingUp to 24 hoursMulti-patient use,without reprocessingUp to 24 hours
One-way valve useperiodSingle useUp to 2 hoursSingle useNot communicated butexpected to be up to 2hoursSingle useNot communicated butexpected to be up to 2hours
Insufflation connectoruse periodMulti-patient useN/AN/A
Insufflation connectorprocessReprocessableN/AN/A
Device shelf-life2 years2 years2 years
Tube and valvesterilizationEO sterilizationEO sterilizationEO sterilization
Insufflation connectorsterilizationNon-sterileN/AN/A
Main materialsBiocompatibleBiocompatibleBiocompatible
Inert metallicmaterials,Inert metallicmaterials,Inert metallic materials,
elastomers andelastomers andelastomers and
thermoplasticthermoplasticthermoplastic polymer
polymerpolymer
Not made with DEHPContain DEHPContain DEHP
Device tissue contactIntestinal mucosaindirect contact for thepatientIntestinal mucosaindirect contact for thepatientIntestinal mucosaindirect contact for thepatient
Skin direct contact forthe end-userSkin direct contact forthe end-userSkin direct contact forthe end-user
PVC/SiliconePVC/SiliconePVC/Silicone
Tube materialsPVC/SiliconePVC/SiliconePVC/Silicone
Connector and valvematerialsSAN/ABSSAN/ABSSAN/ABS
Trade NameENDOtube (subjectdevice)ENDOGATOR™(primary predicate)ENDO SMARTCAP™
510(k) numberK203350K092429K093665
Bottle cap materialPCPE/SANPE/SAN
Bottle cap seal materialTPUSiliconeSilicone
TestingFluid and PressureTesting, flow rated dataNot provided in 510(k)summaryNot provided in 510(k)summary

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" written in smaller, gray letters. The logo is simple and modern.

Substantial Equivalence Discussion VII.

Based on the above comparison of intended use, endoscope compatibilities, testing methods and construction materials employed, the subject device ENDOtube (multiple models) are substantially equivalent to the two named predicates. The main asset of a device that is intended to provide sterile water, air or CO2 to an endoscope are the safety of use and the compatibility of the product with the environment of a medical theater.

The comparison between ENDOtube devices and the predicates are mainly based on these attributes.

  • Compatibility with the main endoscope, pump and sterile water bottle available on the market
  • Ability to provide sterile water, air or CO2
  • Sterilized, cross-contamination proof devices
  • Main materials used

VIII. Non-clinical Testing Summary

DescriptionParameterTest descriptionConclusion
EndoscopecolumncompatibilityOLYMPUS® GI endoscopes140, 160, 180 and 190Connection and usability testingPass
PENTAX® GI endoscopesConnection and usability testingPass
FUJIFILM® GI 500/600series and 700 seriesendoscopesConnection and usability testingPass
Irrigation peristalticpump compatibilityOlympus®, Erbe®, EMED®Connection and usability testingPass
InsufflatorcompatibilityOLYMPUS® UCR,FUJIFILM® GW-1 and GW-100Connection and usability testingPass
Universal waterbottle capPresence and effectivenessConnection and usability testingPass
Fluid supplySterile water; Air; CO2Leak test and usability testingPass
One-way valve (forirrigation pumpcompatible device)Presence and effectivenessCross-contamination testPass
Tube use periodMulti-patient use, withoutreprocessingUp to 24 hoursDurability testPass
One-way valve useSingle useOperational testPass
  • CONFIDENTIAL AND PROPRIETARY -ENDOtube 510(k) Traditional Submission

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Nissha Medical Technologies. The word "NISSHA" is written in large, bold, blue letters. Below the word "NISSHA" is the phrase "MEDICAL TECHNOLOGIES" written in smaller, light gray letters. The logo is simple and modern.

DescriptionParameterTest descriptionConclusion
periodUp to 2 hours
Device shelf-life2 yearsPerformance testPass
Insufflation connectoruse periodMulti-patient useOperational testPass
Insufflation connectorprocessingReprocessablePerformance testPass
Tube and valvesterilizationEO sterilizationSterilization validationPass
Main materialsBiocompatibleBiocompatibility testing (AAMI/ISO 10993)Pass
Inert metallic materials,elastomers andthermoplastic polymerBiocompatibility testing (AAMI/ISO 10993)Pass
No DEHPChemical characterizationPass

Clinical Testing IX.

Not required for a finding of substantial equivalence.

X. Conclusion

Based on the above comparisons of intended use, endoscope, testing methods and construction materials employed, the subject device ENDOtube (multiple models) are substantially equivalent to the two named predicates. The main asset of a device that is intended to provide sterile water, air or CO2 to an endoscope are the safety of use and the compatibility of the product with the environment of a medical theater. The comparison between ENDOtube devices and the predicates are mainly based on these following attributes:

  • Compatibility with the main endoscope, pump and sterile water bottle available on the market;
  • Ability to provide sterile water, air or CO2;
  • Sterilized, cross-contamination proof devices;
  • Main materials used.

Therefore, the ENDOtube is substantially equivalent to the named predicate devices.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.