K Number
K203321

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-12-09

(27 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cystoject Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.

Device Description

The Cystoject Syringe Lever is a reusable device that is intended for use in the delivery of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of the physician. The Cystoject Syringe Lever body contains a housing for holding a standard syringe and a lever that pushes the syringe's plunger when the user rotates its knob. A clutch mechanism allows the user to quickly move the lever to the syringe plunger's position. The Cystoject Syringe Lever does not require an energy source, does not utilize electrical components, and does not utilize software or hardware. The Cystoject Syringe Lever does not come into direct or indirect contact with the patient nor with fluids intended for instillation. The Cystoject Syringe Lever is provided non-sterile as a reusable, reprocessable device to be cleaned and sterilized or high-level disinfected before each use. The Cystoject Syringe Lever indicated for use with a 30 mL syringe.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Cystoject Syringe Lever." This is a device for administering sterile materials, not an AI/ML powered device, therefore the typical acceptance criteria and study information for AI/ML devices, such as performance metrics, sample sizes for test and training sets, expert qualifications, and MRMC studies, are not applicable.

The submission focuses on demonstrating substantial equivalence to a predicate device (Uroject12 Syringe Lever) based on intended use, technological characteristics, and non-clinical performance tests.

Here's a breakdown of the available information:

1. A table of acceptance criteria and the reported device performance:

The document states that "The results of the validation testing demonstrate that the Cystoject Syringe Lever met all predetermined acceptance criteria." However, the specific acceptance criteria and detailed performance results are not publicly available in this FDA letter/summary. The performed tests were:

Acceptance Criteria CategoryReported Device Performance
Residual Fluid Volume post dispensingMet predetermined criteria (no specific values provided)
Load/Unload TestingMet predetermined criteria (no specific values provided)

2. Sample size used for the test set and the data provenance:

Not applicable. The described tests are non-clinical (validation testing of the physical device), not an evaluation of diagnostic accuracy using a test set of medical data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This is not an AI/ML device that requires expert-established ground truth for a test set.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

Not applicable. The "ground truth" for a mechanical device's performance would be engineering specifications and functional requirements.

8. The sample size for the training set:

Not applicable. This is not an AI/ML device that utilizes a training set.

9. How the ground truth for the training set was established:

Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 9, 2020

UroGen Pharma Ltd. % James G. Ottinger, R.Ph. Executive Vice President, Regulatory and Quality UroGen Pharma, Inc. 400 Alexander Park Drive Princeton, NJ 08540

Re: K203321

Trade/Device Name: Cystoject Syringe Lever Regulation Number: 21 CFR§ 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: QBL Dated: November 11, 2020 Received: November 12, 2020

Dear James G. Ottinger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino, M.S. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203321

Device Name Cystoject Syringe Lever

Indications for Use (Describe)

The Cystoject Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for UroGen Pharma. The logo consists of a stylized blue circle with a branching design inside, resembling a molecule or cell structure. To the right of the circle are the words "UroGen" in a bold, blue font, with the trademark symbol "®" next to it. Below "UroGen" is the word "Pharma" in a smaller, lighter blue font.

9 Ha'Ta'asiya St. POB 2397 Ra'anana 4365405, Israel Tel.: +972 - 9 - 770 - 7600 Fax:+972 - 77 - 417 – 1410 e-mail: info@urogen.com

510(K) SUMMARY [SPECIAL 510(K)] Cystoject Syringe Lever

510(k) Number: K203321

SUBMITTER I.

Applicant's Name and address:

UroGen Pharma Ltd. 9 HaTaasiya Street Ra'anana, Israel, 4365405 Tel: +972-9-770-7600 Fax: +972-77-417-1410

Primary Contact Person:

James G. Ottinger, R.Ph. Executive Vice President , Regulatory and Quality UroGen Pharma, Inc. Telephone: +1-646-768-9780 Mobile: +1-610-745-1230 Fax: +1-646-918-1262 e-mail: jim.ottinger@urogen.com

Date Prepared: December 1, 2020

II. DEVICE

Trade Name:Cystoject Syringe Lever
Common or Usual Name:Syringe Lever
Classification Name:
Regulation:Piston Syringe (21 CFR 880.5860)
Class:II
Product Code:QBL (Piston Syringe Lever)
Panel:General Hospital

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for UroGen Pharma. The logo consists of a blue abstract graphic on the left, resembling a cell or branching structure within a partial circle. To the right of the graphic is the text "UroGen" in a bold, blue sans-serif font. Below "UroGen" is the word "Pharma" in a smaller, lighter blue font.

K203321 Page 2 of 3

III. PREDICATE DEVICES

Predicate device Uroject12 Syringe Lever, by UroGen Pharma, Ltd., Product code QBL, cleared under: K190987.

This predicate has not been subject to a design-related recall.

510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Act of 1990 and 21 CFR 807.92.

IV. DEVICE DESCRIPTION

The Cystoject Syringe Lever is a reusable device that is intended for use in the delivery of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of the physician. The Cystoject Syringe Lever body contains a housing for holding a standard syringe and a lever that pushes the syringe's plunger when the user rotates its knob. A clutch mechanism allows the user to quickly move the lever to the syringe plunger's position. The Cystoject Syringe Lever does not require an energy source, does not utilize electrical components, and does not utilize software or hardware. The Cystoject Syringe Lever does not come into direct or indirect contact with the patient nor with fluids intended for instillation.

The Cystoject Syringe Lever is provided non-sterile as a reusable, reprocessable device to be cleaned and sterilized or high-level disinfected before each use.

The Cystoject Syringe Lever indicated for use with a 30 mL syringe.

V. INDICATIONS FOR USE

The Cystoject Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.

VI. SUBSTANTIAL EQUIVALENCE TO PREDICATE DEVICE

The Cystoject Syringe Lever is substantially equivalent to the predicate device based on the following:

Intended Use

The intended use of the subject and predicate devices is identical.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for UroGen Pharma. The logo consists of a blue circular design on the left, resembling a stylized molecule or cell. To the right of the design is the text "UroGen" in a larger, bolder blue font, with the word "Pharma" underneath in a smaller font. The overall design is clean and professional, conveying a sense of innovation and scientific expertise.

Ha'Ta'asiva St. POB : a'anana 4365405 Israe 972 - 77 - 417 e-mail: info@urogen.com

Technology

Both the Cystoject Syringe Lever and the Uroject12 Syringe Lever have the same principle and mode of operation. The Cystoject Syringe Lever has the same general design as the Uroject12 Syringe Lever, except for modifications to enable a different compatible syringe size (i.e., 30 mL with the Uroject12 Syringe Lever) and change in the exterior device color. These differences in technological characteristics do not raise different questions of safety and effectiveness.

VII. PERFORMANCE DATA

The summary of design control activities identified the following non-clinical performance tests to support the substantial equivalence determination.

Validation Testing

The only testing included in this submission that pertains to the validation activities for the additional device are the Residual Fluid Volume post dispensing and Load/Unload Testing.

The results of the validation testing demonstrated that the Cystoject Syringe Lever is considered safe and effective for its intended use.

Summary

The results of the validation testing demonstrate that the Cystoject Syringe Lever met all predetermined acceptance criteria.

VIII. CONCLUSION

The performance testing demonstrates that the Cystoject Syringe Lever is substantially equivalent to the predicate device.

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).