K Number
K203299

Validate with FDA (Live)

Device Name
CellFX System
Date Cleared
2021-02-02

(85 days)

Product Code
Regulation Number
878.4400
Age Range
12 - 120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.

Device Description

The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.

The CellFX Console is capable of delivering short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.

AI/ML Overview

The provided text describes the CellFX® System, an electrosurgical cutting and coagulation device, and its acceptance criteria for FDA clearance. The information focuses on safety and performance, drawing comparisons to a predicate device, Fractora by Invasix Ltd.

Here's a breakdown of the requested information based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

For Biocompatibility Testing:

Acceptance CriteriaReported Device Performance
All tests met respective acceptance criteria as per FDA Guidance (June 16, 2016): Use of International Standard ISO 10993-1.All biocompatibility tests met their respective acceptance criteria.

For Electrical Safety and Electromagnetic Compatibility (EMC):

Acceptance CriteriaReported Device Performance
Compliance with IEC 60601-1, IEC 60601-1-2, 60601-2-2, IEC 60601-1-6, IEC 62366, and IEC 62304.All electrical safety and EMC tests passed.

For Software Verification and Validation Testing:

Acceptance CriteriaReported Device Performance
Software testing supports the safety and effectiveness of the device. Cybersecurity controls mitigate malware risk as per FDA Draft Guidance.Completed software testing supports the safety and effectiveness of the device. Cybersecurity controls have been implemented.

For Animal Study (Safety):

Acceptance CriteriaReported Device Performance
No procedure-related complications; CellFX treatments considered safe based on intra-procedural treatments, gross evaluation, histopathological analysis, clinical pathology, and in-life physical examinations.All animals survived to their termination date with no procedure-related complications. CellFX treatments are considered to be safe.

For Animal Study (Performance):

Acceptance CriteriaReported Device Performance
All test articles meet the acceptance criteria for device performance. Treatment zones meet acceptance criteria. Successful degeneration and necrosis of active treatment sites while the epidermis remained intact.All test articles met the acceptance criteria for device performance. The treatment zones met the acceptance criteria. The CellFX treatment sites successfully achieved degeneration and necrosis of active treatment sites while the epidermis remained intact. The device is expected to perform as intended.

2. Sample Size Used for the Test Set and Data Provenance

The primary testing described for effectiveness and safety is an Animal Study.

  • Sample Size: 15 Yucatan pigs.
  • Data Provenance: The study was conducted as a GLP (Good Laboratory Practice) animal safety study. While the geographical origin isn't specified, GLP studies are typically conducted in controlled laboratory environments. It is a prospective study since animals underwent treatments specifically for the study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number or qualifications of experts for establishing ground truth within the animal study. However, the evaluation included:

  • Gross evaluation
  • Histopathological analysis
  • Clinical pathology
  • In-life physical examinations

These types of evaluations would typically be performed by qualified veterinary pathologists, clinicians, and support staff, but specific numbers and qualifications are not provided in this summary.

4. Adjudication Method for the Test Set

The document does not specify an adjudication method like 2+1 or 3+1. The evaluations for the animal study (gross, histopathological, clinical pathology, physical exams) suggest a comprehensive assessment, likely involving consensus among the relevant scientific and medical professionals involved in the study, but a formal adjudication process is not detailed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study involving human readers or interpretation of results is not mentioned or described in this document. The study described is an animal study validating device safety and performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This device is not an AI algorithm. It's a physical electrosurgical system. Therefore, the concept of "standalone (algorithm only)" is not applicable here. The performance described is that of the physical device in an animal model.

7. The Type of Ground Truth Used

For the animal study, the ground truth was established through a combination of:

  • Histopathology: Microscopic examination of tissue samples.
  • Clinical Pathology: Analysis of blood or other body fluids.
  • Gross Evaluation: Macroscopic observation of tissues and organs.
  • In-life Physical Examinations: Direct clinical assessment of the animals.
  • Survival Data: Absence of procedure-related complications.

This constitutes a comprehensive set of biological and pathological observations.

8. The Sample Size for the Training Set

The document does not describe a training set. This is a premarket notification for a physical medical device, not an AI/ML algorithm that typically undergoes a training phase with a dataset.

9. How the Ground Truth for the Training Set Was Established

As there is no training set described for this device, information on how its ground truth was established is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

February 2, 2021

Pulse Biosciences, Inc. William Knape VP Regulatory, Clinical, & Quality Affairs 3957 Point Eden Way Hayward. California 94545

Re: K203299

Trade/Device Name: CellFX® System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 6, 2020 Received: November 9, 2020

Dear William Knape:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrb/ofdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203299

Device Name CellFX® System

Indications for Use (Describe)

The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

510(k) Summary

I. Submitter

Applicant:Pulse Biosciences, Inc.3957 Point Eden WayHayward, CA 94545Phone: (510) 906-4649
Contact Person:William A. KnapeVP Regulatory, Clinical, & Quality AffairsPulse Biosciences, Inc.Phone: (919) 757-2033
Date Prepared:February 2, 2021

DEVICE INFORMATION II.

Trade Name:CellFX® System
Regulation Number21 CFR § 878.4400
Regulation Name:Electrosurgical Cutting and Coagulation Device andAccessories
Regulation Class:Class II
Product Code:GEI
Classification Panel:General and Plastic Surgery

III. PREDICATE DEVICE

K102461 - Fractora by Invasix Ltd. No reference devices were used in this submission.

IV. DEVICE DESCRIPTION

The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.

The CellFX Console is capable of delivering short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The

{4}------------------------------------------------

treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.

V. INDICATIONS FOR USE STATEMENT

The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

To support a determination of substantial equivalence, Pulse Biosciences performed verification and validation testing demonstrating the subject device performs as intended; thus, even though the CellFX System has different technological characteristics than the predicate device, the differences do not raise new or different questions of safety and effectiveness.

Subject DevicePredicate Device (K102461)
Trade NameCellFX® SystemFractora
Prescription OnlyYesYes
Regulation Number878.4400878.4400
Product CodeGEIGEI
Indications for UseIndications for Use
Indications for Use StatementThe CellFX System is intended for dermatological procedures requiring ablation and resurfacing of the skin.Fractora is intended for dermatological procedures requiring ablation and resurfacing of the skin.
Technological Characteristics
Mechanism of ActionNano-Pulse Stimulation: ultrafast nanosecond electrical pulses to the targeted tissue via treatment tips with an array of microneedlesNon-homogeneous (fractional) pulsed RF energy delivered to skin via an array of multi-electrode pins resulting in heating of skin directly below the electrodes
Pulse Frequency1-10 Pulses per secondUp to 2 Pulses per second
Pulse Amplitude/Voltage300V to 15kV~300 Volts
Pulse Width100 ns – 700 nsNot Applicable
Power Input100-240 VAC, 50/60 Hz, 2A100-240 VAC, 50-60 Hz, 1.4 A
Power Output30 Watts75 Watts
Energy RangeMinimum: 0.5 Joules @ 0.06 WattsMaximum: 4.3 Joules @ 0.6 WattsMinimum: 0.2 Joules @ 75 WattsMaximum: 3.7 Joules @ 75 Watts
Integrated Skin Temperature SensorNot IncludedIncluded
Use of Local AnesthesiaYesYes
Treatment TipYesYes
HandpieceYesYes

{5}------------------------------------------------

CELLFX® SYSTEM 510(k) Premarket Notification

System Physical Characteristics
Height132 cm100 cm
Length46 cm36 cm
Width53 cm36 cm
Weight54 kg15 kg
Power Input100-240 VAC, 50/60 Hz, 2A100-240 VAC, 50-60 Hz, 1.4 A
Tip Characteristics
MicroneedleConfiguration1.5x1.5mm: 2 rows of 3 pins per row2.5x2.5mm: 2 rows of 3 pins per row5.0x5.0mm: 2 rows of 4 pins per row24 pin tip: 6 rows of 4 pins per row60 pin tip: 10 rows of 6 pins per row

PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility Testing

The testing was conducted in accordance with FDA Guidance, issued June 16, 2016: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." All biocompatibility tests met their respective acceptance criteria.

The battery of testing included the following:

  • Material-Mediated Pyrogenicity ●
  • Systemic Toxicity ●
  • Sensitization ●
  • Irritation ●
  • Cytotoxicity
  • . Hemolysis

All biocompatibility tests met their respective acceptance criteria.

Electrical safety and electromagnetic compatibility (EMC)

The CellFX System complies with IEC 60601-1, IEC 60601-1-2, 60601-2-2. IEC 60601-1-6, IEC 62366, and IEC 62304. All test conditions were performed as outlined within the IEC standards. All electrical safety and EMC tests passed

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since a failure of the device could directly result in minor injury to the patient or operator. Completed software testing supports the safety and effectiveness of the device.

Cybersecurity controls have been implemented to mitigate the risk of malware being introduced into the CellFX System as recommended by FDA's Draft Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."

{6}------------------------------------------------

Animal Study

A GLP animal safety study was conducted to validate the safety and performance of the CellFX System. The study was conducted on 15 Yucatan pigs that underwent skin treatments with either the CellFX System (test) or the predicate device (control). Animals were divided into 5 cohorts of 3 animals each for the following timepoints: 0, 2, 6, 14, and 30-days.

Safety

  • All animals survived to their termination date with no procedure related . complications. Based on the results of intra-procedural treatments, gross evaluation, histopathological analysis, clinical pathology, and in-life physical examinations the CellFX treatments are considered to be safe.

Performance

  • All test articles met the acceptance criteria for device performance. The treatment . zones met the acceptance criteria. The CellFX treatment sites successfully achieved degeneration and necrosis of active treatment sites while the epidermis remained intact. The device is expected to perform as intended.

Summary

Based on the GLP animal safety study, the CellFX System was found to have a safety and performance profile that is equivalent to the predicate device.

CONCLUSION

The CellFX® System has the same intended use as the predicate. The differences in technological characteristics between the CellFX System and predicate device do not raise any different questions of safety and effectiveness. The non-clinical performance testing and in vivo animal data provided in this submission demonstrate and support that the CellFX System is as safe and as effective as the predicate device. Therefore, the CellFX System is substantially equivalent to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.