K Number
K203249

Validate with FDA (Live)

Device Name
Nucleus.io
Manufacturer
Date Cleared
2020-12-04

(30 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Nucleus.io is a software-based image management solution and PACS to be used by radiologists, other medical personnel and patients. Nucleus.io is comprised of software modules that provide image receipt, diagnostic viewing, storage, distribution, enhancement, sharing, manipulation, and networking of medical 2D/3D images at distributed locations.

Device Description

Nucleus.io is a PACS which, when integrated with standard off-the-shelf hardware/software, acts similarly to its predicate device and other industry standard PACS systems. The Nucleus System has the following primary features and functions: Zero-footprint HTML5 medical image acceptance (upload) and transfer of medical images between facilities. Easy, real-time access to images for all participants in the healthcare process, including radiologists, imaging technicians, workflow coordinators, physicians, nurses, and other patient care facilitators. High-speed, zero-footprint diagnostic review of medical images using industry-standard tools for manipulation, annotation, measurement, and comparison. Simultaneous information review with multiple parties, including radiologists and coordinators. Sharing of medical images between healthcare providers, facilities, and patients. Encrypted transmission of medical images through secured wired and wireless networks. Encrypted storage of medical images. Organization and matching of multiple medical images based on patient name, medical record number, facility, etc. Quality assurance review of studies by coordinators and preparation of information for study interpretation by radiologists. Notification of care coordinators that studies need to be transferred to client systems. HIPAA-compliant data management and LDAP security integration with external systems Management of users, roles, permissions, and organization accounts Configuration of image acceptance (upload) and transfer settings. The Nucleus.io System consists of the following primary components: Nucleus.io Image Exchange (iX)™ - Exam acceptance (upload), transfer, sharing, and management. Nucleus.io Viewer (dX)™ - Zero-footprint streaming viewer suitable for diagnostic image review of all modalities. Nucleus.io Infinite Store (iS)™ (formerly Image Store) - Image storage and archiving. Nucleus.io PACS™ - Radiology workflow and worklist features. Nucleus PaaS - Vendor-neutral Class I platform that can be incorporated as a component of other devices.

AI/ML Overview

The document describes the Nucleus.io, a Picture Archiving & Communications System (PACS), and its equivalency to a predicate device, Visage PACS/CS (K082269), particularly concerning the diagnosis of mammography.

Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

1. Table of acceptance criteria and the reported device performance:

The document does not provide a formal table of quantitative acceptance criteria and reported device performance metrics in the way one might expect for an AI/ML diagnostic algorithm (e.g., sensitivity, specificity, AUC). Instead, the acceptance criteria are implicitly defined by the goal of demonstrating substantial equivalence to the predicate device, particularly for mammography diagnosis. The reported performance is framed as the availability of features that are comparable to those in the predicate device.

Implicit Acceptance Criteria (based on comparison to predicate device):

  • Availability of core PACS functionalities (image receipt, diagnostic viewing, storage, distribution, enhancement, sharing, manipulation, networking).
  • Compatibility with various modalities (CT, MR, US, CR, DX, NM, PET, XA).
  • Ability to handle uncompressed and non-lossy compressed images for primary image diagnosis in mammography.
  • Support for the same mammography-specific viewing features as the predicate (Quadrant Navigation, Next/Previous Prior Exam, Change between 2D, Projection, and 3D (Tomo) Display).
  • Meeting general safety and effectiveness requirements for a Class II medical device.

Reported Device Performance (as demonstrated by feature comparison):

Feature/FunctionSubject Device (Nucleus.io)Comparable Device (Visage PACS/CS)Assessment
Product NameNucleus.ioVisage PACS/CSEquivalent in function, different name.
510(k) NumberN/A - ProposedK082269
ManufacturerNucleus Health, LLCVisage Imaging, Inc.
Web Sitewww.nucleushealth.iowww.visageimaging.com
Overview of devicesPACS/clinical viewing/image sharing/storage/long term archivingPACS/clinical viewing/image sharing/storage/long term archivingEquivalent.
User TypesHospital Administrative Staff, Technicians, Radiologists, PatientsHospital Administrative Staff, Technicians, Radiologists, PatientsEquivalent, with explicit mention of patients for non-diagnostic purposes in Nucleus.io.
Supports DICOM image transferYesYesEquivalent.
Software basedYes (moderate concern)Yes (moderate concern)Equivalent risk classification by both.
LAN/WAN supportYesYesEquivalent.
HTML/Web based image transferYesYesEquivalent.
Client HardwareZero-footprint (browser) minimumThin Client minimumDifferent technology (web-based vs. thin client), but serves equivalent purpose.
Image streamingYesNoDifferent technology, claimed to be an improvement (faster, safer, efficient).
Secure image sharing to outside locationsYesYesEquivalent.
Dual monitor supportYesYesEquivalent.
Diagnostic/Clinical viewingYesYesEquivalent.
Off the shelf hardwareYesYesEquivalent.
JPEG/industry standard lossy/lossless compressionYesYesEquivalent.
Zoom/panning featuresYesYesEquivalent.
Image flip/rotate capabilityYesYesEquivalent.
Text annotationYesYesEquivalent.
Statistical reportingYesYesEquivalent.
ROIYesYesEquivalent.
Window level by regionYesYesEquivalent.
HIS/RIS connectivityYesYesEquivalent.
Off-site viewing (including teleradiology)YesYesEquivalent.
Fax supportYesYesEquivalent.
Scalable platformYesYesEquivalent.
HIPAA compliant (encryption)YesYesEquivalent.
HL-7 integrationYesYesEquivalent.
LDAP security integrationYesYesEquivalent.
Archiving to cloud based or network serversYesYesEquivalent.
General admin features (assign, relate studies, etc.)YesYesEquivalent.
Secure log inYesYesEquivalent.
Support for MammographyYesYesKey equivalence for this submission.
Quadrant Navigation (Mammography Specific)YesYesEquivalent.
Next/Previous Prior Exam (Mammography Specific)YesYesEquivalent.
Change between 2D, Projection, and 3D (Tomo) Display (Mammography Specific)YesYesEquivalent.

2. Sample size used for the test set and the data provenance:

The document does not describe a clinical performance study with a specific "test set" in the context of an AI/ML algorithm's diagnostic performance evaluation (e.g., a set of mammograms for classification). Instead, the "study" is a feature-by-feature comparison and technical validation against the predicate device.

  • Sample size: Not applicable in the context of a dataset for AI evaluation. The "sample" here refers to the device itself and its implemented features.
  • Data provenance: Not applicable. The document focuses on the features of the device rather than testing it against a clinical dataset.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This device is a PACS/viewer system, not an AI diagnostic algorithm that requires expert-established ground truth for a clinical dataset. The "ground truth" for this submission is established by the functional specifications and features of the predicate device (Visage PACS/CS K082269) and the general requirements for PACS systems.

4. Adjudication method for the test set:

Not applicable. No clinical dataset requiring adjudication was used for this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No MRMC study was done or described. This is not an AI-assisted diagnostic device that augments human readers but rather a platform for viewing and managing medical images.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. Nucleus.io is a PACS/viewer, not a standalone diagnostic algorithm.

7. The type of ground truth used:

The "ground truth" for this 510(k) submission, particularly regarding the claim for mammography diagnosis, is the functional equivalency to a legally marketed predicate device. This is demonstrated by showing that Nucleus.io possesses the same relevant features and capabilities as Visage PACS/CS, especially those related to mammography viewing and processing, and that it adheres to industry standards (e.g., DICOM, handling uncompressed/lossless images for primary diagnosis). The risk management and verification/validation processes serve to confirm that the implemented features meet specifications and do not introduce new safety or effectiveness concerns.

8. The sample size for the training set:

Not applicable. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable. This is not an AI/ML device that requires a training set.

In summary, this 510(k) submission for Nucleus.io is primarily based on demonstrating substantial equivalence to a predicate PACS system (Visage PACS/CS) by comparing functional features, particularly those supporting mammography diagnosis. It is not an submission for an AI/ML-driven diagnostic device, and therefore the traditional metrics and study designs (e.g., test sets, ground truth establishment, MRMC studies) typically associated with AI performance evaluation are not present in this documentation. The acceptance criteria are qualitative, focusing on feature parity and adherence to general medical device safety and performance standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "U.S. FOOD & DRUG" in a larger font size above the word "ADMINISTRATION" in a smaller font size.

December 4, 2020

Nucleus Health, LLC % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114

Re: K203249

Trade/Device Name: Nucleus.io Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: November 3, 2020 Received: November 4, 2020

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203249

Device Name Nucleus.io

Indications for Use (Describe)

Nucleus.io is a software-based image management solution and PACS to be used by radiologists, other medical personnel and patients. Nucleus.io is comprised of software modules that provide image receipt. diagnostic viewing, storage, distribution, enhancement, sharing, manipulation, and networking of medical 2D/3D images at distributed locations. All modules of Nucleus.io are web-based and can operate on off-the-shelf hardware, as needed. Nucleus.io consists of the following primary components: Nucleys.io Viewer (dX) with image streaming technology for use by medical professionals for diagnostic and clinical image review-Nucleus.io Image Exchange (iX) for image acceptance, transfer, and sharing with hospitals/clinics as well as between facilities , Nucleus.io Infinite Store (iS) for secure cloud-based image storage and management through HPAA compliant encryption-Nucleus.io PACS for radiology workflow and worklist features when used in conjunction with Nucleus.io dX and, optionally , Nucleus.io PaaS, which provides a vendorneutral Class I platform that can be incorporated as a component of other devices.

Nucleus io interfaces with health information systems (HIS) using industry-standard image exchange protocols-such as DICOM, HL7, and HTML-through web-based networked gateways and local and wide area networks. Nucleus.io is compatible with modalities including Computed Tomography (CT), Magnetic Resonance Imaging (MR), Ultrasound (US), Computed Radiography (DX), Digital Radiography (DX), Nuclear Medicine (NM), Positron Emission Tomography (PET), and X-Ray Angiography (XA). When appropriate, Nucleus.io provides and installs software and, optionally, off-the-shelf server hardware at client facilities to facilitate secure, web-based connections for image transmission to and from hospital central servers via the Internet. Additionally, industry standard HTTPS, VPNs, and other encryption methodologies are utilized to allow for optimal, secure, rapid streaming of image compression and encryption adhere to standard industry protocols. Nucleus.io can be used as a full-featured PACS or as an independent viewer in clinical settings.

Although the web and thin client technologies allow the software to be run on a variety of hardware platforms, for diagnostic purposes the user must make sure that the display hardware used for reading the images complies with state of- the-art diagnostic requirements and currently valid laws.

Only DICOM for presentation images can be used on an FDA approved monitor for primary image diagnosis.

Only uncompressed or non-lossy compressed images must be used for primary image diagnosis in mammography. Not for diagnostic use on mobile devices.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

< Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid 0MB number."

{3}------------------------------------------------

1 Introduction and Administrative Information

This 510(k) Summary provides a high-level summary of the contents of the Nucleus.io Premarket Notification (510(k)), including a comparison of Nucleus.io to existing legally marketed medical devices.

This Premarket Notification (510(k)) Summary contains no confidential or trade secret information. For additional information, please contact the Establishment's contact listed below.

This summary was prepared on July 09, 2020.

Submitter 1.1

SubmitterNucleusHealth, LLC
Submitter Address13280 Evening Creek Dr. SSan Diego CA, 92128
Submitter Phone(858) 251-3400
Submitter Websitewww.nucleushealth.io
Establishment Number3008008144
Establishment ContactZachary Wright
Contact TitleManager, Regulatory Affairs and Quality
Contact Phone(858) 251-3400
Contact Fax(858) 546-3900
Contact Emailzwright@nucleushealth.io

1.2 Device Identification

Proprietary Name(s):Nucleus.io
Classification Name:Picture Archiving & Communications System (PACS)
Common / Usual Name:Medical Image Processing Software

1.3 Device Classification

Nucleus.io is product code LLZ, which is a Class II medical device per 21 CFR 892.2050.

Similarly, classification of Nucleus.io per IEC 62304 Edition 1.0 2015-06 was assigned by assessing the highest risk classification first:

  • Death or serious injury is not possible. Thus, the device is not Class C.

{4}------------------------------------------------

  • . Non-serious injury is possible. Thus, the device is Class B.
  • . As some injury or damage to health is possible, the device is not Class A.

1.4 Related Previous Submissions

Previously cleared as "Nucleus Imagement System" in the United States as a Class II medical device under 510(K) K171130.

Predicate Device Identification 1.5

Proprietary Name(s):Visage PACS/CS
510(k) File:K082269
Classification Name:Picture Archiving & Communications System (PACS)

No reference devices were used in this submission.

2 Device Description and Intended Use

2.1 Device Description

Nucleus.io is a PACS which, when integrated with standard off-the-shelf hardware/software, acts similarly to its predicate device and other industry standard PACS systems.

The Nucleus System has the following primary features and functions:

  • Zero-footprint HTML5 medical image acceptance (upload) and transfer of medical . images between facilities.
  • . Easy, real-time access to images for all participants in the healthcare process, including radiologists, imaging
  • technicians, workflow coordinators, physicians, nurses, and other patient care facilitators.
  • High-speed, zero-footprint diagnostic review of medical images using industry-standard ● tools for manipulation, annotation, measurement, and comparison.
  • . Simultaneous information review with multiple parties, including radiologists and coordinators.
  • . Sharing of medical images between healthcare providers, facilities, and patients.
  • . Encrypted transmission of medical images through secured wired and wireless networks.
  • . Encrypted storage of medical images.
  • Orqanization and matching of multiple medical images based on patient name, medical . record number, facility, etc.
  • . Quality assurance review of studies by coordinators and preparation of information for study interpretation by radiologists.
  • Notification of care coordinators that studies need to be transferred to client systems. ●
  • . HIPAA-compliant data management and LDAP security integration with external systems
  • Management of users, roles, permissions, and organization accounts
  • Configuration of image acceptance (upload) and transfer settings ●

{5}------------------------------------------------

The Nucleus.io System consists of the following primary components:

Nucleus.io Image Exchange (iX)™

  • Exam acceptance (upload), transfer, sharing, and management .

Nucleus.io Viewer (dX)™

  • Zero-footprint streaming viewer suitable for diagnostic image review of all modalities .

Nucleus.io Infinite Store (iS)™ (formerly Image Store)

  • Image storage and archiving .

Nucleus.io PACS™

  • Radiology workflow and worklist features .

Nucleus PaaS

  • Vendor-neutral Class I platform that can be incorporated as a component of other ● devices
    Note: When Nucleus is used as a PaaS, the resulting device carries its own intended use, and must be filed as a separate medical device which uses Nucleus as a component, as appropriate. PaaS is mentioned here only to assist in understanding the features present in Nucleus IMS.

The graphic below provides a high-level architectural view of how the product interacts with external users and systems:

Image /page/5/Figure/15 description: The image shows a diagram of the Nucleus.io application suite. The suite includes applications for users, such as patients, physicians, and operators, as well as applications for health systems, such as PACS, modality, and RIS/EMR. The Nucleus.io application suite is connected to Microsoft Azure via HTTPS and API. The diagram also shows third-party apps that are connected to the Nucleus.io application suite.

Figure 1. Marketing Description, Nucleus.io architecture

{6}------------------------------------------------

The following images provide visual examples of features used in the most common user workflows. Many additional features are available, and are fully described in the User Manuals, discussed in the Proposed Labeling (Section D of the submission).

The interface depicted in Figure 2, below, is displayed when the user logs in. This interface provides access to the most common features used by hospital staff and Radiologists.

nucleus.ioChange HealthcareSUSystem User
ACTIONSPATIENTEXAM DATEMODDESCRIPTIONIMAGESRECEIVEDSOURCEPUSH TO PACS
ClearAdvanced >MRN: NDX-DEMO-15 DOB: 02-02-09:38 AM
111-25-2014OB: 02-02- 09:45 AMCTCT CHEST W CONTRAST56308-18-202012:35 PMUploaderNever Pushed
2Exam InfoHIPAA LogPush HistoryShares OutDownloadDelete
3MRNPATIENT NAMEDATE OF BIRTHGENDERACC. #EXAM DESCRIPTION
NDX-DEMO-15Hanging, Harriet02-02-1985Female000092218CT CHEST W CONTRAST
EXAM DATEEXAM TIMEMOD
11-25-201409:45 AMCT
More Info >Edit
4Healy, JoeMRN: 7487619 DOB: 03-23-195210-31-201712:07 AMXAXA Brain708-18-202012:35 PMUploaderNever Pushed
5

Figure 2. Nucleus.io Primary User Interface

ltemDescription
By selecting the ellipsis icon () on an exam line item, an authorized user canshare the exam to another Organization, to the patient, to another professional, ordirectly to an email address.
2Detailed exam information can be viewed by selecting the arrow icon (>) on anexam line item.
3From this expanded view, an authorized user can edit, download or delete theexam. Exam history, including all access to Personal Health Information (PHI) isalso available.
By selecting the eye icon ( ® ), an authorized user can view the exam. See Figure4 for the interface displayed when viewing exams.
5By selecting the button in the Push to PACS column, an authorized user can sendthe exam to a configured external PACS system. The icon displays the currentpush status for the exam (e.g., Never Pushed, In Progress, Complete).

Figure 3 depicts the clinical exam viewer available from the main user interface. This viewer meets the needs of most system users. The "Dx" viewer, also available within Nucleus.io, has additional image manipulation and diagnostic features, but shares a similar interface.

{7}------------------------------------------------

Image /page/7/Figure/2 description: The image shows a medical imaging software interface, likely for viewing CT scans. The interface displays a series of thumbnails on the left, labeled with information such as "Hanging, Harriet NDX-DEMO-15" and "CT CHEST W CONTRAST". A larger image of a full body scan is displayed in the center, with a scale on the right indicating a location of 342.0 and a thickness of 2.0.

Figure 3. Nucleus.io Clinical Viewer User Interface

ItemDescription
1Icons in the tool bar at the top of the screen allow the user to work with the imageson screen. This includes features such as changing contrast, rotating, zoom,measuring, and playback. The user may also configure the layout of the viewer todetermine settings such as how many series/images can be displayed at once.
2The windows at the left show the exam series available, as specified when theexam was performed and compiled into a DICOM-compatible format.
3The main window shows the selected image series (per Item 2) and may beconfigured to show more than one image at a time.
4The scroll bar serves multiple functions, including the ability to scroll and a "loading"bar to show download (or "streaming" progress of the displays series. In thisexample, the blue portion of the scroll bar indicates images which have loaded, andthe darker portion of the scroll bar indicates images which are still loading.

Intended Use/Indications for Use 2.2

Nucleus.io is a software-based image management solution and PACS to be used by radiologists, other medical personnel and patients. Nucleus.io is comprised of software modules that provide image receipt, diagnostic viewing, storage, distribution, enhancement, sharing, manipulation, and networking of medical 2D/3D images at distributed locations.

All modules of Nucleus.io are web-based and can operate on off-the-shelf hardware, as needed. Nucleus.io consists of the following primary components: Nucleus.io Viewer (dX) with image streaming technology for use by medical professionals for diagnostic and clinical image review—Nucleus.io Image Exchange (iX) for image acceptance, transfer, and sharing with hospitals/clinics as well as between facilities, Nucleus.io Infinite Store (iS) for secure cloud

{8}------------------------------------------------

based image storage and management through HIPAA compliant encryption—Nucleus.io PACS for radiology workflow and worklist features when used in conjunction with Nucleus.io dX and, optionally, Nucleus.io iS-and Nucleus.io PaaS, which provides a vendor-neutral Class I platform that can be incorporated as a component of other devices.

Nucleus.io interfaces with health information systems (HIS) using industry-standard image transfer and data exchange protocols-such as DICOM, HL7, and HTML-through web-based networked gateways and local and wide area networks. Nucleus.io is compatible with modalities including Computed Tomography (CT), Magnetic Resonance Imaging (MR), Ultrasound (US), Computed Radiography (CR), Digital Radiography (DX), Nuclear Medicine (NM), Positron Emission Tomography (PET), and X-Ray Angiography (XA). When appropriate Nucleus.io provides and installs software and, optionally, off-the-shelf server hardware at client facilities to facilitate secure, web-based connections for image transmission to and from hospital central servers via the Internet. Additionally, industry standard HTTPS, VPNs, and other encryption methodologies are utilized to allow for optimal, secure, rapid streaming of images. Lossless image compression and encryption adhere to standard industry protocols. Nucleus.jo can be used as a full-featured PACS or as an independent viewer in clinical settings.

Although the web and thin client technologies allow the software to be run on a variety of hardware platforms, for diagnostic purposes the user must make sure that the display hardware used for reading the images complies with state of- the-art diagnostic requirements and currently valid laws.

Only DICOM for presentation images can be used on an FDA approved monitor for mammography for primary image diagnosis.

Only uncompressed or non-lossy compressed images must be used for primary image diagnosis in mammography.

Not for diagnostic use on mobile devices.

3 Predicate Device Comparison

3.1 Context for Predicate Device Selection

Nucleus.io was previously cleared under 510(k) K171130 as Nucleus Image Management System, contraindicated for diagnosis of mammography. This submission seeks to remove that contraindication now that features are available to support this indication. Within the K171130 submission, StatPACS (K092134) was identified as a predicate device. StatPACS is a client-server-based PACS which is contraindicated for diagnosis of mammography.

Within the StatPACS (K092134) submission, Visage PACS/CS (K082269) was identified as a predicate device. Visage PACS/CS is also a client-server-based PACS but includes an indication for diagnosis of mammography. A key difference between StatPACS and Visage PACS/CS is that Visage PACS/CS is indicated for diagnosing mammography. Note: This difference was not noted in K092134 as the StatPACS indications for use did not include diagnosis of mammography.

In conclusion, the predicate device selected for this submission is Visage PACS/CS (K082269) as the technological differences have previously been shown to not introduce additional safety

{9}------------------------------------------------

or effectiveness concerns; and because the features provided for diagnosis of mammography are similar to those provided in Nucleus.io.

Device Comparison 3.2

The features in Nucleus.io related to diagnosis of mammography were developed based on the principle features and operations within its predicate device. Visage PACS/CS (K082269). There are two primary differences between the products:

    1. Nucleus.io is modular, allowing the user to gain access to either minimal features needed in smaller/niche environments, or scale to a full PACS,
    1. Nucleus.io uses cloud architecture to replace traditional server hosting, and web browser technology to replace thick client applications, in order to increase the availability of the system for medical professionals in remote locations and to access to health records for patients.

Both Visage PACS/CS and Nucleus.io run on off-the-shelf PC hardware utilizing industrystandard operating systems and interfaces to all hospital modalities.

Both products support non-compressed, lossless compressed, and lossy compressed image/data transmission. The default mode of operation is non-compressed, lossless compression. Changes to the standards which may affect the compression coding will be evaluated to worth and merit and if adopted, validated and verified to show compliance to the standard. This method fully preserves image data integrity but allows reduced transmission times and reduced storage requirements.

As PACS (Picture Archiving and Communications Systems) are multi-site/location systems, both Visage PACS/CS and Nucleus.io support various industry standard connectivity including, but not limited to LAN/WAN/VPN/Wireless technology, which enables data/images to be transmitted with speed and accuracy. Nucleus.io has adopted image transfer using streaming technology, which enables the medical professional to obtain the images in a fast, safe and efficient manner as compared to the traditional file download features used in Visage PACS/CS.

3.2.1 Indications for Use Comparison

The intended uses/indications for use of the devices, shown below, include comparable purposes. Differences include:

  • . Patients are specifically mentioned as users in the Nucleus.io documentation, as patients may access their records for non-diagnostic purposes similar to the common practice of a physician providing the patient a CD-ROM.
  • . Non-exhaustive lists of modalities compatible with the devices are provided; and different modalities were selected to include in the list by each company. However, all modalities listed for each can be found in the DICOM Conformance Statements for each device; these differences are limited to the wording used and specific items to be included in the non-exhaustive list. Both DICOM Conformance statements are provided in Section C of this submission. In particular, the following Classes confirm their equivalence with respect to the Intended Uses:
    • 1.2.840.10008.5.1.4.1.1.3.1 O
    • 1.2.840.10008.5.1.4.1.1.6.1 O

{10}------------------------------------------------

  • 1.2.840.10008.5.1.4.1.1.1 O
  • O 1.2.840.10008.5.1.4.1.1.20
  • O 1.2.840.10008.5.1.4.1.1.128
  • O 1.2.840.10008.5.1.4.1.1.12.1
  • Two warnings are included at the bottom of the Visage PACS/CS intended use which . are not explicitly mentioned in the Nucleus.io intended use. Nucleus.io has implemented these warnings differently in documentation, including these within the User Manuals.
Subject DeviceComparable Device
Nucleus.io (NA – Proposed Device)Visage PACS/CS (K082269)
Nucleus.io is a software-based image management solution and PACS to be used by radiologists, other medical personnel, and patients. Nucleus.io is comprised of software modules that provide image receipt, diagnostic viewing, storage, distribution, enhancement, sharing, manipulation, and networking of medical 2D/3D images at distributed locations. All modules of Nucleus.io are web-based and can operate on off-the shelf hardware, as needed. The Nucleus.io IMS consists of the following primary components: Nucleus.io Viewer (dX) with image streaming technology for use by medical professionals for diagnostic and clinical image review—Nucleus.io Image Exchange (iX) for image acceptance, transfer, and sharing with hospitals/clinics as well as between facilities—Nucleus.io Infinite Store (iS) for secure cloud based image storage and management through HIPAA compliant encryption—Nucleus.io PACS for radiology workflow and worklist features when used in conjunction with Nucleus.io dX and, optionally, Nucleus.io iS—and Nucleus.io PaaS, which provides a vendor-neutral Class I platform that can be incorporated as a component of other devices.Visage PACS/CS is a system for distributing, viewing, processing, and archiving medical images within and outside health care environments.The Visage PACS/CS server receives images data in DICOM format via the hospital network. This provides universal connections to archives, modalities, and workstations. The supported modalities are listed in the DICOM Conformance Statement.Besides general image interpretation and processing tools, Visage PACS/CS provides specific tool sets for several clinical applications, including:- CT/MR angiography, e.g., for vascular analysis and stent planning- Cardiac analysis, including calcium scoring and functional assessment of cardiac CT data- Neuroradiology, including CT and MR brain perfusion analysis
Nucleus.io interfaces with health information systems (HIS) using industry standard image transfer and data exchange protocols—such as DICOM, HL7, and HTML—through web-based networked gateways and local and wide area networks. Nucleus.io is compatible with modalities including, but not limited to: Computed Tomography (CT), Magnetic Resonance Imaging (MR), Ultrasound (US), Computed Radiography (CR), Digital Radiography (DX), Nuclear Medicine (NM), Positron Emission Tomography (PT), and X-Ray Angiography (XA). When appropriate,Visage PACS/CS is to be used only by trained and instructed health care professionals. It can support physicians and/or their medical staff in providing their own diagnosis for medical cases. The final decision regarding diagnoses, however, resides with the doctors and/or their medical staff in their own area of responsibility.Although the web and thin client technologies allow the software to be run on a variety of hardware platforms, for diagnostic purposes the user must make sure that the display hardware used for reading the images complies with state-of-the-art diagnostic requirements and currently valid laws.
Subject DeviceComparable Device
Nucleus.io provides and installs software and,optionally, off-the-shelf server hardware at clientfacilities to facilitate secure, web-basedconnections for image transmission to and fromhospital central servers via the Internet.Additionally, industry standard HTTPS, VPNs, andother methodologies are utilized to allow foroptimal, secure, rapid streaming of images.Lossless image compression and encryptionadhere to standard industry protocols. Nucleus.iocan be used as a full-featured PACS or as anindependent viewer in clinical settings.Only DICOM for presentation images can beused on an FDA approved monitor formammography for primary image diagnosis.Only uncompressed or non-lossy compressedimages must be used for primary imagediagnosis in mammography.

{11}------------------------------------------------

3.2.2 Tabular Comparison of Features

The table below provides a side-by-side comparison of key features.

Feature/FunctionSubject DeviceComparable Device
Product NameNucleus.ioVisage PACS/CS
510(k) NumberN/A - ProposedDeviceK082269Primary PredicateDevice
ManufacturerNucleus Health, LLCVisage Imaging, Inc.
Web Sitewww.nucleushealth.iowww.visageimaging.com
Overview of devicesPACS/clinicalviewing/imagesharing/storage/longterm archivingPACS/clinicalviewing/imagesharing/storage/longterm archiving
User TypesHospitalAdministrative Staff,Technicians,Radiologists,PatientsHospital AdministrativeStaff, Technicians,Radiologists, Patients
Supports DICOM image transferYesYes
Software basedYes (moderateconcern)Yes (moderate concern)
LAN/WAN supportYesYes
Feature/FunctionSubject DeviceComparable Device
HTML/Web based image transferYesYes
Client HardwareZero-footprint(browser) minimumThin Client minimum
Image streamingYesNo
Secure image sharing to outside locationsYesYes
Dual monitor supportYesYes
Diagnostic/Clinical viewingYesYes
Off the shelf hardwareYesYes
JPEG/industry standard lossy/losslesscompressionYesYes
Zoom/panning featuresYesYes
Image flip/rotate capabilityYesYes
Text annotationYesYes
Statistical reportingYesYes
ROIYesYes
Window level by regionYesYes
HIS/RIS connectivityYesYes
Off-site viewing (including teleradiology)YesYes
Fax supportYesYes
Scalable platformYesYes
HIPAA compliant (encryption)YesYes
HL-7 integrationYesYes
LDAP security integrationYesYes
Archiving to cloud based or network serversYesYes
Feature/FunctionSubject DeviceComparable Device
General admin features (assign, relate studies,etc.)YesYes
Secure log inYesYes
Support for MammographyYesYes

{12}------------------------------------------------

··· nucleushealth"

{13}------------------------------------------------

Features specifically implemented for viewing mammography studies are shown below.

Feature/FunctionSubject DeviceComparable Device
Product NameNucleus.ioVisage PACS/CS
Quadrant NavigationYesYes
Next/Previous Prior ExamYesYes
Change between 2D, Projection, and 3D (Tomo)DisplayYesYes

4 Performance Data

4.1 Risk Management

The Nucleus.io Hazard and Risk Management Plan, Analysis, and Report was performed to determine and evaluate all potential health and safety hazards associated with use of the platform. All foreseeable system hazards, effects, and causes have been evaluated to determine necessary and appropriate risk mitigation and validation, risk mitigation traceability, and design review, have been performed to ensure effective implementation of the stated risk mitigations.

4.2 Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was classified as a Moderate Level of Concern software medical device, due to the low potential for injury possible in the event of software/hardware failure, both directly and indirectly. Per the Hazard and Risk Management Plan, Analysis, and Report, no pre-mitigation items are life-threatening, and all have been mitigated to "minor" (results in temporary injury or impairment not requiring professional medical intervention) or lower.

Verification and Validation Protocols have been executed to ensure adequate testing of all defined product design requirements and specifications. Verification and Validation Test Reports have been created to evaluate the acceptability of test results and product module / product release preparedness. All applicable design and development and verification and

{14}------------------------------------------------

validation activities and records have been completed to ensure safety and effectiveness of the final Nucleus.io platform

All technological characteristics and performance requirements identified in the Predicate Device Comparison have been tested to ensure success of the design implementation. There are no new or different issues of safety or effectiveness introduced by the stated design change. Risk Management and Verification and Validation activities confirm that Nucleus.io meets its specifications.

5 Conclusions

The comparison above in tabular format enables one to see how Nucleus.io compares in intent and usage to its predicate device, Visage PACS/CS. Specifically, all features identified for diagnosing mammography by Visage PACS/CS are also available within Nucleus.io. Therefore, Nucleus.io raises no new issues of safety or effectiveness from its predicate device.

N/A