K Number
K203198

Validate with FDA (Live)

Date Cleared
2021-07-06

(250 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KARL STORZ Metal Sterilization Trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in pre-vacuum steam sterilization cycle (132ºC for four minutes with a 20 minute dry time).

When used in conjunction with an FDA cleared sterility of the enclosed medical device is maintained until used.

The system has been validated with devices with working channels >= 0.2mm and a maximum length of 240mm.

Device Description

The KARL STORZ-ENDOSCOPE Metal Sterilization Trays are intended only for use to encase various KARL STORZ reusable medical devices for sterilization in steam cycles. The sterilization trays are not intended to maintain sterility by themselves. Prior to sterilization. the trays must be double-wrapped with an appropriate FDA-cleared sterilization wrap to provide a microbial barrier which allows sterilant to permeate throughout the interior of the loaded tray.

The tray configurations are all baskets, available in various sizes, which are designed to encase KARL STORZ medical devices, such as light cables, instruments, rigid telescopes, semi-rigid telescopes, etc. All systems consist of a stainless steel metal mesh base and lid. Lids are attached to the trays with assembled hardware.

The sterilization trays are constructed from mesh material to allow for permeation of sterilant during sterilization. The mesh basket design has a higher percentage of open cells than metal mesh allowing for complete permeation. The trays have latches designed to fasten the lid onto the base. Other tray components include silicone instrument holders to secure instruments and provide protection of the medical devices in the sterilization tray.

AI/ML Overview

The document describes the acceptance criteria and the results of non-clinical testing for the KARL STORZ Metal Sterilization Trays. It does not describe a study involving a medical device that uses artificial intelligence or machine learning. Therefore, many of the requested elements of the response (e.g., sample size for the test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set size, ground truth for the training set) are not applicable or cannot be extracted from the provided text.

However, I can extract the acceptance criteria and reported device performance from the "Summary of Non-Clinical Testing" table.

Here's the relevant information that can be extracted:

1. A table of acceptance criteria and the reported device performance

Type of TestingPurposeAcceptance CriteriaReported Device Performance/Result
Pre-vacuum sterilization efficacy AAMI ST77 Containment Devices for Reusable Medical Device Sterilization; ISO 17665-1 Sterilization of Health Care Products Moist Heat Part 1 Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical DevicesDemonstrate sterilization capabilities.10-6 SAL (Sterility Assurance Level)PASSED
Pre-vacuum dry time AAMI ST77 Containment Devices for Reusable Medical Device Sterilization; ISO 17665-1 Sterilization of Health Care Products Moist Heat Part 1 Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical DevicesTo establish minimum dry time.Pre and Post Sterilization weight difference <3% after dryingPASSED (20 minutes dry time)
Manual Cleaning - Protein, Hemoglobin AAMI TIR 30 A Compendium of Processes, Materials, Test Methods, and Acceptance Criteria for Cleaning Reusable Medical DevicesTo demonstrate manual cleaning.< 6.4 $\mu g/cm^2$ protein and < 2.2 $\mu g/cm^2$ hemoglobin on device after cleaningPASS
Mechanical Cleaning - Protein, Hemoglobin AAMI TIR 30 A Compendium of Processes, Materials, Test Methods, and Acceptance Criteria for Cleaning Reusable Medical DevicesTo demonstrate mechanical cleaning.< 6.4 $\mu g/cm^2$ protein and < 2.2 $\mu g/cm^2$ hemoglobin on device after cleaningPASS
Material biocompatibility ISO 10993-5 Biological Evaluation of Medical Devices Part 5: Tests for in vitro cytotoxicityTo demonstrate no Cytotoxic properties.Cytotoxicity: No evidence of lysisPASS

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The document refers to non-clinical testing performed with the subject device and standards used, but details on sample sizes for specific tests or data provenance are absent.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as the document describes non-clinical performance and safety testing of a physical medical device (sterilization trays), not an AI/ML diagnostic or prognostic device that would require expert-established ground truths.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable for the same reasons as point 3.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study is not applicable as the device is a sterilization tray, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Standalone performance refers to AI algorithm performance. This is not applicable to the KARL STORZ Metal Sterilization Trays.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the non-clinical tests described, the "ground truth" is established by the specified standards and test methods (e.g., AAMI ST77, ISO 17665-1, AAMI TIR 30, ISO 10993-5) which define what constitutes a "passed" result based on quantifiable metrics (e.g., 10-6 SAL, weight difference <3%, specific protein/hemoglobin concentrations, no evidence of lysis). These are objective measurements rather than expert consensus on a clinical diagnosis.

8. The sample size for the training set

This information is not applicable as the device is not an AI/ML product that would require a training set.

9. How the ground truth for the training set was established

This information is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

July 6, 2021

Karl Storz Endoscopy America Inc % David Furr Consultant Toscano Consulting 8708 Capehart Cove Austin, Texas 78733

Re: K203198

Trade/Device Name: KARL STORZ Metal Sterilization Trays Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: May 26, 2021 Received: June 1, 2021

Dear David Furr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203198

Device Name KARL STORZ Metal Sterilization Trays

Indications for Use (Describe)

The KARL STORZ Metal Sterilization Trays are intended only for use to encase and protect specific KARL STORZ reusable medical devices for sterilization in pre-vacuum steam sterilization cycle (132ºC for four minutes with a 20 minute dry time).

When used in conjunction with an FDA cleared sterility of the enclosed medical device is maintained until used.

The system has been validated with devices with working channels ≥ 0.2mm and a maximum length of 240mm. Intended Load & Maximum Weight Trav Catalog Number Dimensions

11580A7"x5.31"x1.46"Sterilization of probes and dilators 1lb. (Karl Storz Instruments Only)
11580B10.8"x6.9"x1.46"Telescope sterilization tray 2.4lbs.
11580C13.5"x7"x1.38"Telescope sterilization tray 5lbs.
11580D10.8"x6.9"x1.46"Telescope sterilization tray 2.75lbs.
39501A111.4"x2.36"x2"Cleaning/sterilization basket: Rigid scope 1.25lbs.
39501A213.9"x4.92"x2.13"Cleaning/sterilization basket: 2 rigid scopes and LT cable 3.75lbs.
39501B219"x4.9"x2.1"Cleaning/Sterilization basket; 2 rigid scopes/1 LT cable 5lbs.
39501BEC18.9"x4.9"x2.1"Cleaning/sterilization basket: Scope & LT cable 3lbs
39501C26.3"x3.1"x2"Cleaning/sterilization basket for telescopes up to 670mm 2.5lbs
39501CEC23.8"x4.9"x2.1"Cleaning/sterilization basket: Scope & LT cable 3lbs
39501F22.4"x3.15x2.05"Cleaning/sterilization basket: Rigid endoscope 3.5lbs.
39501LC210.24"x4.72"x6.7"Cleaning/sterilization basket: Laryngoscope blades and module 3.2lbs
39501X25.3"x5.9"x3.1"Cleaning/sterilization basket for telescopes up to 43cm 2.4lbs
39501XK18.11"x5.9"x3.15"Cleaning/sterilization basket: telescopes 2.4lbs
39501XP18.11"x5.9"x3.15"Cleaning/Sterilization basket for telescopes 2.4lbs
39550A18.9"x9.84"x2.36"Cleaning/Sterilization basket: shaver and cable 3.25lbs.

l ype of Use elect one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary Pursuant to 21 CFR 807.92 K203198

Date: May 26, 2021

1.Submitted By:KARL STORZ Endoscopy-America, Inc.2151 E. Grand Ave.El Segundo, California 90245424-218-8376
2.Contact:David C. FurrToscano Consulting Group8708 Capehart CoveAustin, Texas 78733512-906-9654
3.Product:KARL STORZ Sterilization Trays- Metal K203198Product code: KCT - Class II (21 CFR 880.6850)
4.Common/ClassificationName:Predicate devices:Sterilization wrap/containerSymmetry Medical Polyvac Surgical Instrument DeliverySystem K012105

Description:

The KARL STORZ-ENDOSCOPE Metal Sterilization Trays are intended only for use to encase various KARL STORZ reusable medical devices for sterilization in steam cycles. The sterilization trays are not intended to maintain sterility by themselves. Prior to sterilization. the trays must be double-wrapped with an appropriate FDA-cleared sterilization wrap to provide a microbial barrier which allows sterilant to permeate throughout the interior of the loaded tray.

The tray configurations are all baskets, available in various sizes, which are designed to encase KARL STORZ medical devices, such as light cables, instruments, rigid telescopes, semi-rigid telescopes, etc. All systems consist of a stainless steel metal mesh base and lid. Lids are attached to the trays with assembled hardware.

The sterilization trays are constructed from mesh material to allow for permeation of sterilant during sterilization. The mesh basket design has a higher percentage of open cells than metal mesh allowing for complete permeation. The trays have latches designed to fasten the lid onto the base. Other tray components include silicone instrument holders to secure instruments and provide protection of the medical devices in the sterilization tray.

{5}------------------------------------------------

510(k) Premarket Notification KARL STORZ Sterilization Travs- Metal

Intended Use:

The KARL STORZ Metal Sterilization Trays are intended to protect medical device instrumentation and facilitate the sterilization process by sterilant penetration and air removal. When used in conjunction with an FDA cleared sterilization wrap, sterility of the enclosed medical device is maintained until used.

KARL STORZ has validated the KARL STORZ Metal Sterilization Trays for use in steam sterilizers. Most of the metal sterilization trays are used with rigid telescopes, instruments and cables that do not have lumens. The system was validated with a miniature scope with an irrigation channel lumen diameter of 0.2mm and a length of 240mm.

Comparison of Technological Characteristics:

The KARL STORZ Metal Sterilization Trays are comparable to Symmetry Medical Polyvac Surgical Instrument Delivery System (K012105). The Symmetry Medical system also includes plastic and hybrid trays however only the Symmetry Medical metal tray are considered as a predicate.

Steam enters the KARL STORZ metal sterilization trays through mesh openings in the tray base and lid. After sterilization, sterility is maintained by the FDA cleared sterilization wrap. All of these characteristics are the same as the predicate device.

The devices and the predicate have essentially the same indications for use and perform in a similar manner.

Element ofComparison510(k) Device:KARL STORZ MetalSterilization TraysPredicate Device:Symmetry MedicalPolyvac SurgicalInstrument DeliverySystem (K012105)Explanation of Differences
Regulation andProduct ClassificationCode21 CFR 880.6850KCT21 CFR 880.6850KCTNone
Indications for UseThe KARL STORZ MetalSterilization Trays areintended only for use toencase and protectspecific KARL STORZreusable medical devicesfor sterilization in pre-vacuum steamsterilization cycles (132°Cfor four minutes with a 20minute dry time).When used in conjunctionwith an FDA clearedsterilization wrap, sterilityof the enclosed medicaldevice is maintained untilused.Polyvac's DeliverySystems consist ofperforated trays withlids, which areintended to enclose andprotect medical deviceinstrumentation, and tofacilitate thesterilization processingby allowing steampenetration and airremoval, when used inconjunction with anapproved sterilizationwrap. Sterility of theenclosed medicaldevices maintainedSimilar

{6}------------------------------------------------

The system has beenvalidated with deviceswith working channels $\geq$0.2mm and a maximumlength of 240mm.(See IFU for table of partnumbers & contents)until used. Polyvac'sDelivery Systems areto be sterilized in oneof the following cycles:pre vacuum steam,132°C 4 minutesminimum, gravitysteam 132°C 30minutes minimum andgravity steam 121°C 55minutes minimum.
Principal Material ofConstructionStainless steel metalmeshThermoformed RadelPolyphenylsulfone,stainless steel oraluminumDifferent

{7}------------------------------------------------

KARL STORZ ModelNumbers16 models 11580A 11580B 11580C 11580D 39501A1 39501A2 39501B2 39501BEC 39501C 39501CEC 39501F 39501LC2 39501X 39501XK 39501XP 39550AN/ADifferent
DimensionalConfiguration RangeSizes available in inches: 7"x5.31"x1.46" 10.8"x6.9"x1.46" 13.5"x7"x1.38" 10.8"x6.9"x1.46" 11.4"x2.36"x2" 13.9"x4.92"x2.13" 19"x4.9"x2.1" 18.9"x4.9"x2.1" 26.3"x3.1"x2" 23.8"x4.9"x2.1" 22.4"x3.15x2.05" 10.24"x4.72"x6.7" 25.3"x5.9"x3.1" 18.11"x5.9"x3.15" 18.11"x5.9"x3.15" 18.9"x9.84"x2.36"Approximate sizesavailable in inches: 7x2x1 7x3x1 8x4x1 11x7x1 21x10x4 17x10x4 17x8x2 15x10x1.5 20.5x9.7x5 20x10x3 26x9x6 17x3.5x1.5 And othersSimilar
Device ManufacturerContract manufacturedfor KARL STORZ byHelmut ZepfMedizintechnik GmbHand HUPFERMetallwerke GmbHManufactured byTecomet (formerlySymmetry Medical)Different
Sterilization CyclesPrevacuum Steam 4minute cycle 132°CDrying time 20 minutesPrevacuum Steam 4minute cycle 132°C 20-40 minutes dryingtime; gravity steam132°C for 30 minutesor 121°C for 55minutes 20-50 minutesdrying timeSimiliar
LoadTrays are to be loadedwith KARL STORZinstrumentsVarious loads up to 25lbs.KARL STORZ trays are used withKARL STORZ instruments with a totalload of <5 lbs.

{8}------------------------------------------------

5 10(k) Premarket Notification KARL STORZ Sterilization Trays- Metal

Summary of Non-Clinical Testing:

Shown below is the non-clinical testing performed with the subject device and the standards and test method used to demonstrate the subject device met the acceptance criteria of each standard.

Type of TestingPurposeAcceptanceCriteriaResult
Pre-vacuum sterilizationefficacy AAMI ST77Containment Devices forReusable MedicalDevice SterilizationISO 17665-1 Sterilization ofHealth Care Products MoistHeat Part 1 Requirementsfor the Development,Validation, and Routine Controlof a Sterilization Process forMedical DevicesDemonstratesterilizationcapabilities.10-6 SALPASSED
Pre-vacuum dry timeAAMI ST77Containment Devices forReusable MedicalDevice SterilizationISO 17665-1 Sterilization ofHealth Care Products MoistHeat Part 1 Requirementsfor the Development,Validation, and Routine Controlof a Sterilization Process forMedical DevicesTo establish minimumdry time.Pre and Post Sterilizationweightdifference <3% after dryingPASSED20 minutes dry time

{9}------------------------------------------------

5 10(k) Premarket Notification KARL STORZ Sterilization Trays- Metal

Manual Cleaning - Protein, HemoglobinAAMI TIR 30 A Compendium of Processes,Materials, Test Methods, and AcceptanceCriteria for Cleaning Reusable MedicalDevicesTo demonstratemanual cleaning< 6.4 $\mu g/cm^2$ proteinand< 2.2 $\mu g/cm^2$hemoglobin ondevice after cleaningPASS
Mechanical Cleaning - Protein, HemoglobinAAMI TIR 30 A Compendium of Processes,Materials, Test Methods, and AcceptanceCriteria for Cleaning Reusable MedicalDevicesTo demonstratemechanical cleaning< 6.4 $\mu g/cm^2$ proteinand< 2.2 $\mu g/cm^2$hemoglobin ondevice after cleaningPASS
Material biocompatibilityISO 10993-5 Biological Evaluation of MedicalDevices Part 5: Tests for in vitro cytotoxicityTo demonstrate noCytotoxic propertiesCytotoxicity Noevidence of lysisPASS

Conclusion:

The conclusions drawn from the non-clinical tests demonstrate that the proposed device is as safe, as effective, and performed as well as or better than the legally marketed predicate devices.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).