K Number
K203188

Validate with FDA (Live)

Device Name
ADD
Manufacturer
Date Cleared
2021-03-12

(136 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.

Device Description

The ADD are wired or wireless digital flat panel detectors that have been designed for faster, more streamlined approach to digital radiography systems. The ADD detector utilize a combination of propriety TFT glass and scintillators(CsI), and those and electronics are housed in one package. The detectors support an auto-trigger signal sensing technology that allows the detectors to be used without generator integration. The flat panel sensors of the ADD are fabricated using thin film technology based on amorphous silicon technology. Electronically, the sensors are much like conventional photodiode arrays. Each pixel in the array consists of a light-sensing photodiode and a switching Thin Film Transistor (TFT) in the same electronic circuit. Amorphous silicon photodiodes are sensitive to visible light, with a response curve roughly comparable to human vision. The sensitivity of amorphous silicon photodiodes peaks in green wavelengths, well matched to scintillators such as CsI. The response has the excellent linearity of a charge-integrating-biased photodiode. SDK-MCW is the software of Detector that performs image acquisition, image correction, and preprocessing.

AI/ML Overview

Acceptance Criteria and Study for ADD Digital Flat Panel X-Ray Detector

This document describes the acceptance criteria and the study performed to demonstrate the substantial equivalence of the ADD Digital Flat Panel X-Ray Detector (K203188) to predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly present a table of "acceptance criteria" for clinical performance in the typical sense of a target metric to be achieved (e.g., sensitivity, specificity). Instead, the substantial equivalence is primarily based on technological characteristics and non-clinical performance tests, with clinical images provided as supportive evidence.

The relevant performance metrics and comparisons to predicate devices (Predicate Device#1: K181631, Predicate Device#2: K102349, Predicate Device#3: K140771) are summarized below from the provided text:

MetricAcceptance Criteria (Implied by Predicate Performance)Reported Device Performance (ADD)
ScintillatorCsICsI
Effective Pixel Area~345 x 425 mm to 426.3 x 432.0 mm345.24 x 425.6 mm
Total Pixel Number~1,994 x 2,430 pixels to 2,981 x 3,021 pixels2,560 x 3,072 pixels
Pixel Pitch~140 um to 175 um140 um
High Contrast Limiting Resolution (LP/mm)>= 3.5 (from Predicate Device#1)Max. 3.5
DQE (0.1lp/mm, min.)Typ. 40% to 70%50%
MTF (0.1lp/mm, min.)Typ. 95% to 97%97%
CommunicationWired/WirelessWired/Wireless
Exposure ModeNormal Mode (Manual) / AED Mode (Auto Exposure Detection)Normal Mode (Manual), AED Mode (Auto Exposure Detection)
WirelessIEEE 802.11a/b/g/n (from Predicate Device#1 & #2)IEEE 802.11a/b/g/n

The submission states, "subject device shows similar or better DQE," and generally, the "inherent technical characteristics and performance are comparable to the predicate devices."

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a distinct "test set" with a quantifiable sample size in the context of a prospective clinical trial. The "clinical images" were provided as supportive evidence, but the primary basis for substantial equivalence was non-clinical testing and technological comparison.

  • Sample Size for Clinical Images: Not specified. The document only mentions "Clinical images were provided."
  • Data Provenance: Not specified. However, the manufacturer is H&abyz Co., Ltd. from the Republic of Korea. It is common for such supportive clinical images to be collected retrospectively or prospectively within a clinical setting, but this is not detailed. The document explicitly states: "Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate (note TFT technology with CsI scintillator that is identical to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the subject digital detector works as intended." This suggests the clinical images were not the primary means of demonstrating equivalence.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not describe an expert-based ground truth establishment process for the clinical images. Given that the images were "not necessary to establish substantial equivalence" and were "further evidence... to show that the subject digital detector works as intended," it's highly improbable that a formal expert ground truth establishment for a test set was conducted as would be typical for an AI-enabled diagnostic device.

4. Adjudication Method for the Test Set

Since a formal expert-based ground truth establishment for a clinical test set is not described, there is no mention of an adjudication method (e.g., 2+1, 3+1).

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done or reported. The device is an X-ray detector, not an AI software for diagnostic interpretation. The submission is focused on demonstrating the detector's image quality and technical equivalence.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. The ADD is an X-ray detector, a hardware device for image acquisition, not a standalone algorithm. The software (SDK-MCW) performs image acquisition, correction, and preprocessing, but its performance is judged within the context of the detector's output, not as a standalone diagnostic algorithm.

7. Type of Ground Truth Used

For the non-clinical performance tests (DQE, MTF, resolution), the ground truth is established by physical measurements and standardized testing protocols as defined by IEC 62220-1.

For the supporting "clinical images," no specific ground truth (expert consensus, pathology, outcomes data) is mentioned or implied. The purpose of these images was to "show that the subject digital detector works as intended" in a real-world context, rather than to validate diagnostic accuracy against a definitive ground truth.

8. Sample Size for the Training Set

The document does not describe a training set in the context of machine learning. The ADD is a digital X-ray detector, and its software (SDK-MCW) performs image acquisition, correction, and preprocessing. While such software is developed and refined, the submission focuses on its validation rather than a "training set" for an AI model.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no mention of a training set for an AI model in this submission. The software validation followed a software development process.

{0}------------------------------------------------

March 12, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The Department of Health & Human Services logo is a circular emblem with an abstract human figure. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

H&abyz Co., Ltd. % Im Do Gyun Senior Consultant GMS Consulting Co., Ltd. 34. Sangamsan-ro. Mapo-gu Seoul. 03909 KOREA

Re: K203188

Trade/Device Name: ADD Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: January 14, 2021 Received: January 29, 2021

Dear Im Do Gyun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203188

Device Name ADD

Indications for Use (Describe)

ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic systems. Not to be used for mammography.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

{4}------------------------------------------------

510(k) Summary for K203188

[As Required by 21 CFR 807.92]

1. Date Prepared [21 CFR 807.92(a)(a)]

March 02, 2021

2. Submitter's Information [21 CFR 807.92(a)(1)]

  • Name of Manufacturer: H&abyz Co. Ltd.
    ● Address: 1F, 2-Dong, 41-16 Cheoinseong-Ro, Namsa-Myeon, Cheoin-Gu, Yongin-Si, Gyeonggi-Do, Republic of Korea [17118] ● Contact Name: Namkyu Hur / Quality Manager

  • Telephone No.: +82 070-4658-9300

  • Email Address: hnk@abyzr.com

  • Registration No.: 3016674851

3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

510(k) NumberK203188
Trade/Device/Model NameADD
Reference No.HAD1417MCW
Common NameDigital Flat Panel X-ray Detector
Device Classification NameStationary X-ray System
Regulation Number21 CFR 892.1680
Classification Product CodeMQB
Device ClassII
510(k) Review PanelRadiology

{5}------------------------------------------------

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate device within this submission are shown as follow;

Predicate Device#1

  • 510(k) Number: K181631
  • Applicant: Samsung Electronics Co., Ltd.
  • Trade/Device Name: S4335-AW (of GR40CW)
  • Common Name: Digital Diagnostic X-ray System
  • Classification Name: System. X-ray, Stationary
  • Regulation Number: 21 CFR 892.1680
  • Classification Product MQB Code
  • Device Class: II
  • 510(k) Review Panel: Radiology

Predicate Device#2

  • 510(k) Number: K102349
  • Applicant: Konica Minolta Medical & Graphic, Inc.
  • Trade/Device Name: AeroDR SYSTEM
  • Common Name: Digital Diagnostic X-ray System
  • Classification Name: System. X-ray, Stationary
  • Requlation Number: 21 CFR 892.1680
  • Classification Product KPR Code
  • Device Class: II
  • 510(k) Review Panel: Radiology

{6}------------------------------------------------

Predicate Device#3

Applicant:

  • 510(k) Number: K140771
  • PHILIPS ELEVA WORKSPOT Trade/Device Name:

Philips Medical Systems DMC GmbH

  • Common Name: Digital Diagnostic X-ray System
  • Classification Name: System. X-ray, Stationary
  • Regulation Number: 21 CFR 892.1680
  • Classification Product MQB, LLZ Code
  • Device Class: II
  • 510(k) Review Panel: Radiology

These predicate devices have not been subject to a design-related recall

{7}------------------------------------------------

5. Description of the Device [21 CFR 807.92(a)(4)]

The ADD are wired or wireless digital flat panel detectors that have been designed for faster, more streamlined approach to digital radiography systems. The ADD detector utilize a combination of propriety TFT glass and scintillators(CsI), and those and electronics are housed in one package. The detectors support an auto-trigger signal sensing technology that allows the detectors to be used without generator integration.

The flat panel sensors of the ADD are fabricated using thin film technology based on amorphous silicon technology. Electronically, the sensors are much like conventional photodiode arrays. Each pixel in the array consists of a light-sensing photodiode and a switching Thin Film Transistor (TFT) in the same electronic circuit. Amorphous silicon photodiodes are sensitive to visible light, with a response curve roughly comparable to human vision. The sensitivity of amorphous silicon photodiodes peaks in green wavelengths, well matched to scintillators such as CsI. The response has the excellent linearity of a charge-integrating-biased photodiode.

SDK-MCW is the software of Detector that performs image acquisition, image correction, and preprocessing. According to the FDA quidance document entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," it has a moderate level of concern. Also, it does not base on previously-cleared software and was originally coded.

{8}------------------------------------------------

6. Indications for use [21 CFR 807.92(a)(5)]

ADD (Digital Flat Panel X-Ray Detector) is indicated for digital imaging solution designed for providing general radiographic diagnosis of human anatomy targeting both adult and children. It is intended to replace film based radiographic diagnostic. Not to be used for manmography.

7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]

There are no significant differences in the technological characteristics of the predicate device which adversely affect safety or effectiveness. Provided below is a table summaring the technological characteristics of the ADD and the predicate device:

Proposed DevicePredicate Device#1Predicate Device#2Predicate Device#3
K NumberK203188K181631K102349K140771
ManufacturerH&abyz Co., Ltd.Samsung Electronics Co., Ltd.KONICA MINOLTA MEDICAL &GRAPHIC, INCPHILIPS MEDICAL SYSTEMSDMC GMBH
Trade NameADDS4335-AW (of GR40CW)AERODR SYSTEMPHILIPS ELEVA WORKSPOT
Common NameDigital Flat Panel X-ray DetectorDigital Flat Panel X-ray SystemDigital Flat Panel X-ray DetectorDigital Flat Panel X-ray Detector
Product CodeMQBMQBKPRMQB, LLZ
RegulationNumber21 CFR 892.168021 CFR 892.168021 CFR 892.168021 CFR 892.1680
510(k) ReviewPanelRadiologyRadiologyRadiologyRadiology
Indications forUseADD (Digital Flat Panel X-RayDetector) is indicated for digitalimaging solution designed forproviding general radiographicdiagnosis of human anatomytargeting both adult andchildren. It is intended toreplace film based radiographicdiagnostic systems. Not to beused for mammography.The GR40CW Digital X-rayImaging System is intended foruse in general projectionradiographic applicationswherever conventional screen-film systems or CR systems maybe used. This device is notintended for mammographicapplications.The AeroDR SYSTEM isindicated for use in generatingradiographic images of humananatomy. It is intended toreplace radiographic film/screensystems in general-purposediagnostic procedures. TheAeroDR SYSTEM is not indicatedfor use in mammography,fluoroscopy, tomography andangiography applications.As a part of a radiographicsystem, the Philips ElevaWorkspot is intended to acquire,process, store, display, andexport digital radiographicimages. The Philips ElevaWorkspot is suitable for allroutine radiography exams,including specialist areas likeintensive care, trauma, orpediatric work, excludingfluoroscopy, angiography andpediatric applications.
Proposed DevicePredicate Device#1Predicate Device#2Predicate Device#3
mammography.
ScintillatorCsICsICsICsI
Effective PixelArea345.24 x 425.6 mm345 x 425 mm348.95 x 425.25 mm426.3 x 432.0 mm
Total PixelNumber2,560 x 3,072 pixels2,460 x 3,040 pixels1,994 x 2,430 pixels2,981 x 3,021 pixels
Pixel Pitch140um140 um175um143um
High ContrastLimitingResolution(LP/mm)Max. 3.53.5Not publicly available dataNot publicly available data
CommunicationWired/WirelessWired/WirelessWired/WirelessWired/Wireless
DQE50% (0.1lp/mm, min.)Typ.70% @0.1lp/mmTyp. 40% @0.1lp/mmTyp. 50% @0.1lp/mm
MTF97% (0.1lp/mm, min.)Typ.95% @ 0.1lp/mmTyp.97% @ 0.1lp/mmTyp.95% @ 0.1lp/mm
AnatomicalGeneralGeneralGeneralGeneral
Proposed DevicePredicate Device#1Predicate Device#2Predicate Device#3
Sites
Exposure ModeNormal Mode (Manual),AED Mode (Auto ExposureDetection)Manual, Auto (AED Mode)Normal Mode (Manual),AED Mode (Auto ExposureDetection)Normal Mode (Manual),AED Mode (Auto ExposureDetection)
WirelessIEEE 802.11a/b/g/nIEEE 802.11a/b/g/nIEEE 802.11aNot publicly available data

[Table 1. Comparison of Proposed Device to Predicate Devices]

{9}------------------------------------------------

{10}------------------------------------------------

The proposed device is substantially equivalent to the previously cleared detector provided as part of the complete imaging system.

It is substantially equivalent to these devices in design, funciples, and intended use. The proposed device, ADD has been tested about electrical safety, EMC, and the software has been validated. Also, the clinical data has been provided to support the substantial equivalence to the predicate device.

There are no significant differences between the ADD and the predicate devices that would adversely affect the nonclinical & clinical test results. (MTF & DQE specification of the predicate devices are provided by their user manual and literature) These differences do not raise different questions of safety and effectiveness than the predicate.

{11}------------------------------------------------

8. Non-Clinical Test summary

The ADD complies with voluntary standards for electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:

  1. Electrical Safety, Electromagnetic Compatibility and Performance:

The ADD complies with the electrical safety and electromagnetic compatibility requirements established by the standards.

StandardsNo.StandardsOrganizationStandard TitleVersionPublicationYear
ES60601-1AAMI ANSIMedical Electrical Equipment - Part 1: GeneralRequirements for Basic Safety and EssentialPerformance (IEC 60601-1:2005, MOD)ES60601-1:2005(R)2012 andA1:20122014
60601-1-2IECMedical Electrical Equipment - Part 1-2: GeneralRequirements for Safety – Collateral Standard:Electromagnetic Compatibility - Requirementsand Tests60601-1-2Edition 4.02014-022016
-FDARadio Frequency Wireless Technology in MedicalDevicesAugust 142013

2) Software Validation

The ADD contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated. Software information is provided in accordance with FDA quidance:

  • The content of premarket submissions for software contained in medical devices, on May 11, 2005

3) Biocompatibility

  • ISO 10993-1 and series, Biological evaluation of medical devices

4) Performance Test

Imaging performance test has been conducted according to:

  • . IEC 62220-1, Medical Electrical Equipment – Characteristics of Digital X-ray Imaging Devices – Part 1-1: Determination of the Detective Quantum Efficiency – Detectors Used in Radiographic Imaging.
    We select the predicate device #2 and #3 in order to demonstrate adequate DQE performance of the ADD detector. According to the above comparison table, subject device shows similar or better DQE.

{12}------------------------------------------------

  1. Cybersecurity
  • Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, on October2, 2014
  1. Label
  • CFR Part 801
  • Pediatric Information for X-ray Imaging Device Premarket Notifications, on November 28, ● 2017

9. Clinical Test Summary

Clinical data has been provided according to FDA guidance document "Guidance for the Submission of 510(k)s for Solid Sate X-ray Imaging Devices". Clinical images were provided; these images were not necessary to establish substantial equivalence based on the differences from the predicate (note TFT technology with CsI scintillator that is identical to the predicate image plate) but they provide further evidence in addition to the laboratory performance data to show that the subject digital detector works as intended

10. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]

There are no significant differences between the ADD and the predicate device, K181631 that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.

11. Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the ADD is substantially equivalent in safety and effectiveness to the predicate device as described herein.

N/A