K Number
K203174

Validate with FDA (Live)

Date Cleared
2021-08-02

(280 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Medical Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.

Device Description

Disposable Medical Mask is a single-use, three layer, flat-folded mask with ear loops and nose piece. The inner and outer layers are constructed of spun-bond polypropylene and the middle layer is constructed of melt blown polypropylene filter. The mask is held in place over the mouth and nose by two elastic loops welded to the facemask. The elastic loops are not made with natural rubber latex. The nose piece is made of malleable polyethylene with aluminum wire and allows the user to fit the facemask around their nose. Disposable Medical Mask is sold non-sterile and is intended to be a single use, disposable device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the "Disposable Medical Mask," based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

Test MethodologyPurposeAcceptance CriteriaReported Device Performance
Fluid Resistance (ASTM F1862)Evaluate fluid resistance performance.Level 3: 160 mm HgAll samples met the predetermined acceptance criteria.
Bacterial Filtration Efficiency (ASTM F2100-19)Evaluate bacterial filtration efficiency performance.$\geq$ 98%All samples met the predetermined acceptance criteria.
Differential Pressure (Delta P) (ASTM F2100-19)Evaluate differential pressure performance.< 6.0 mm H2O/cm²All samples met the predetermined acceptance criteria.
Particulate Filtration Efficiency (ASTM F2100-19)Evaluate particulate filtration efficiency performance.$\geq$ 98%All samples met the predetermined acceptance criteria.
Flammability (21 CFR 1610)Evaluate flammability performance.Class I, Does not IgniteAll samples met the predetermined acceptance criteria.
Cytotoxicity (ISO 10993-5)Evaluate cytotoxicity potential.Non-cytotoxicAll samples met the predetermined acceptance criteria.
Irritation (ISO 10993-10)Evaluate whether the device was an irritant to the animal model.Non-irritatingAll samples met the predetermined acceptance criteria.
Sensitization (ISO 10993-10)Evaluate whether the device was a sensitizer to the animal model.Non-sensitizingAll samples met the predetermined acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The non-clinical testing was performed using 3 nonconsecutive lots with a total of 96 samples evaluated.
  • Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. However, given that the manufacturer is Dong Guan Kinyet Metal Products Co., Ltd. in China, it's highly probable the testing was conducted prospectively in a laboratory, potentially in China or a facility used by the manufacturer.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable (N/A) for this type of device and study. The "ground truth" for the performance criteria of a medical mask (like filtration efficiency, fluid resistance, biocompatibility) is established through standardized laboratory test methods (e.g., ASTM, ISO standards), not through expert consensus on cases. The results are objective measurements.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

This information is N/A for this type of device and study, as the tests are objective measurements based on standardized protocols, not subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This information is N/A. This document describes the performance of a physical medical device (a mask), not an AI-powered diagnostic or assistive technology that would involve human readers or MRMC studies.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is N/A. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is established by objective, quantitative measurements obtained through adherence to recognized international and national standards (ASTM, ISO, 21 CFR). For example:

  • Fluid resistance is measured by the ability to resist penetration by synthetic blood at a certain pressure.
  • Bacterial filtration efficiency is measured by the percentage of bacteria filtered.
  • Biocompatibility (cytotoxicity, irritation, sensitization) is assessed through standardized in-vitro and in-vivo tests that yield measurable results (e.g., cell viability, observed reactions).

8. The Sample Size for the Training Set

This information is N/A. This is a physical medical device, not a machine learning model. Therefore, there is no "training set" in the context of AI.

9. How the Ground Truth for the Training Set Was Established

This information is N/A as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

August 2, 2021

Dong Guan Kinyet Metal Products Co., Ltd % Salon Chen System Engineer IMD Medical & Drug technology service institutions Tianbao office room 225,Sha Tai Road No.209 ShenZhen city, Guangdong province 518117 China

Re: K203174

Trade/Device Name: Disposable Medical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: July 23, 2021 Received: July 23, 2021

Dear Salon Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203174

Device Name Disposable Medical Mask

Indications for Use (Describe)

Disposable Medical Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary - K203174

Submitter's Information 1.

  • Company Name: Dong Guan Kinyet Metal Products Co., Ltd.

  • Establishment Registration Number: 3012119360

  • Address: NO.38, Weijian Road, Chashan Town, Dongguan City, Guangdong Province, China

  • Phone:+86-769-86487999

  • Fax: +86-769-86487999

  • Contact Person(Title): Emily Zhang (General Manager)

  • E-mail: emily.zhang@kinyet.com

  • Device Information 2.
  • Trade Name: Disposable Medical Mask

  • Common Name: Mask,Surgical

  • Classification Name: Surgical Apparel

  • Model: KLY-301

    1. Classification
  • Classification Product Code: FXX

  • Regulation Number: 21 CFR 878.4040

  • Classification: Class II

  • Review Panel: General Hospital

4. Predicate Device Information:

  • 510(k) Number: K201137 >
  • Predicate Device Name: ASEPT® Surgical Face Mask Page 1 of 7

{4}------------------------------------------------

  • Manufacturer: PFM Medical, Inc.

  • This predicate has not been subject to a design-related recall

  • No reference devices were used in this submission.

5. Application Correspondent

  • Company Name: IMD Medical & Drug technology service institutions

  • Phone: +86-18613190779

  • Fax: +86-755-62809168

  • Contact Person(Title): Salon Chen (System engineer)

  • E-mail: 33999439@qq.com

6. Device Description

Disposable Medical Mask is a single-use, three layer, flat-folded mask with ear loops and nose piece. The inner and outer layers are constructed of spun-bond polypropylene and the middle layer is constructed of melt blown polypropylene filter. The mask is held in place over the mouth and nose by two elastic loops welded to the facemask. The elastic loops are not made with natural rubber latex. The nose piece is made of malleable polyethylene with aluminum wire and allows the user to fit the facemask around their nose. Disposable Medical Mask is sold non-sterile and is intended to be a single use, disposable device.

7. Indications for Use Statement

Disposable Medical Mask is indicated as a protective nose and mouth covering for healthcare workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situation where there is a risk of exposure to microorganisms and body fluids.

8. Technological Characteristic Comparison

Elements of ComparisonProposed DevicePredicate DeviceComparison
Company NameDong Guan KinyetMetal Products Co., Ltd.PFM Medical, Inc./

{5}------------------------------------------------

Device NameDisposable Medical MaskASEPT® Surgical Face Mask/
Classification Product CodeFXXFXXSame
Regulation21 CFR 878.404021 CFR 878.4040Same
Classification NameSurgical ApparelSurgical ApparelSame
ClassIIIISame
Prescription or OTCOTCOTCSame
Intended Use & Indications for UseDisposable Medical Maskis indicated as a protectivenose and mouth coveringfor healthcare workers andpatients involved inmedical and surgicalprocedures. The masksare indicated in anyprocedure or situationwhere there is a risk ofexposure tomicroorganisms and bodyfluids.The ASEPT® SurgicalFace Mask is indicated asa protective nose andmouth covering forhealthcare workers andpatients involved inmedical and surgicalprocedures. The masksare indicated in anyprocedure or situationwhere there is a risk ofexposure tomicroorganisms and bodyfluids.Similar
MaterialsInner and Outer LayersSpun-bond polypropyleneSpun-bond polypropylene
Middle LayerMelt blown polypropylene filterMelt blown polypropylene filter
Ear LoopsPolyesterPolyester

{6}------------------------------------------------

Nose PieceMalleable polyethylenewith aluminum wireMalleable polyethylenewith aluminum wire
Dimension17.5cm length x 9.5cmheight17.5cm length x 9.5cmheightSame
Mask StylePleatedPleatedSame
Design FeaturesMalleable nosepiece, flatpleated,elastic ear loopsMalleable nosepiece, flatpleated,elastic ear loopsSame
SterilityNon-sterileNon-sterileSame
UseSingle Use,DisposableSingle Use,DisposableSame
ColorBlue and WhiteBlue and WhiteSame
ASTM F2100 levelLevel 3Level 2Different
BiocompatibilityCytotoxicity,ISO10993-5:2009Non-cytotoxicNon-cytotoxic
Irritation,ISO10993-10:2002Non-irritatingNon-irritatingSame
Sensitization,ISO10993010:2002Non-sensitizingNon-sensitizing

9. Summary of Non-Clinical Test

Non-clinical data:

Per FDA document Guidance for Industry and FDA Staff: Surgical Masks – Premarket Notification [510(k)] Submissions, the below testing has been completed on the subject device: The testing provided below was performed using 3 nonconsecutive lots with a total of 96 samples evaluated.

{7}------------------------------------------------

Test MethodologyPurposeAcceptance CriteriaResults
Fluid Resistance PerformanceASTM F1862:Standard Test Method forResistance of MedicalFace Masks to Penetrationby Synthetic BloodThe purpose was toevaluate the fluidresistanceperformance.Level 3:160 mm HgAll samples met thepredetermined acceptancecriteria.
Bacterial Filtration EfficiencyASTM F2100-19:Standard Specification forPerformance of MaterialsUsed in Medical Face MasksThe purpose was toevaluate thebacterial filtrationefficiencyperformance.$\geq$ 98%All samples met thepredetermined acceptancecriteria.
Differential Pressure(Delta P)ASTM F2100-19:Standard Specification forPerformance of MaterialsUsed in Medical FaceMasksThe purpose was toevaluate thedifferential pressureperformance.< 6.0 mm H2 O/cm²All samples met thepredetermined acceptancecriteria.
Particulate Filtration EfficiencyASTM F2100-19:Standard Specification forPerformance of MaterialsUsed in Medical Face MasksThe purpose was toevaluate theparticulate filtrationefficeincyperformance.$\geq$ 98%All samples met thepredetermined acceptancecriteria.
Flammability21 CFR 1610The purpose wasto evaluate theflammabilityperformance.Class I, Does not IgniteAll samples met thepredetermined acceptancecriteria.

{8}------------------------------------------------

CytotoxicityISO 10993-5:Biological evaluationof medical devices— Part 5: Testsfor in vitro cytotoxicityThe purpose was toevaluate the cytoxicitypotential.Non-cytotoxicAll samples met thepredetermined acceptancecriteria.
IrritationISO 10993-10:Biologicalevaluation of medical devices -Part 10: Tests for irritation andskin sensitizationThe purpose was toevaluate whether thedevice was an irritantto the animal model.Non-irritatingAll samples met thepredeterminedacceptance criteria.

{9}------------------------------------------------

SensitizationISO 10993-10:Biologicalevaluation of medical devices –Part 10: Tests for irritation andskin sensitizationThe purpose was toevaluate whether thedevice was asensitizer to theanimal model.All samples met thepredetermined acceptancecriteria.
Non-sensitizing

10. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) Disposable Medical Mask, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K201137.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.