K Number
K203141

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-09-17

(332 days)

Product Code
Regulation Number
876.1075
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Uramix Curaway Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as liver, kidneys, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone, or breast.

Device Description

The URAMIX CuraWay Automatic Core Biopsy Instrument is a single use only core biopsy device and is individually packaged and sterilized. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified to maintain the sterilization condition of the device. It is available in several needle gauge sizes and lengths. The side and rear actuator buttons are color coded according to the various gauge sizes. The 18 gauge is the one used for current intended marketing, for prostate and kidney biopsies.

AI/ML Overview

The URAMIX CuraWay Automatic Core Biopsy Instrument is a medical device intended for obtaining biopsies from soft tissues. The provided text outlines a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study of the device's diagnostic performance against specific acceptance criteria. Therefore, several of the requested sections (e.g., effect size of human readers, standalone performance, training set details) are not applicable or cannot be extracted from the given document as it primarily covers non-clinical engineering and design comparisons.

Here's an analysis of the provided information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state "acceptance criteria" in the traditional sense of diagnostic performance metrics (like sensitivity, specificity, or AUC). Instead, it provides a comparison of engineering and functional characteristics to a predicate device, aiming to show substantial equivalence. The "performance" reported is primarily in relation to these comparative metrics.

Feature/TestAcceptance Criteria (Implied by Predicate/Standards)Reported Device Performance (Subject Device)Outcome
Intended UseObtaining biopsies from soft tissues (liver, kidney, prostate, spleen, lymph nodes, various soft tissue tumors). Not for bone.Intended for use in obtaining biopsies from soft tissues (liver, kidney, prostate, spleen, lymph nodes, various soft tissue tumors). Not for bone or breast.Similar clinical condition, same intended purpose and site. (Note: Subject device adds "not for breast")
MaterialStainless steel + ABSShaft: Medical grade stainless steel (SUS304/06Cr19Ni10). Handle/Actuator: Acrylonitrile Butadiene Styrene (ABS).Same material used.
Types/Sizes (Gauges)14G, 16G, 18G, 20G12G, 14G, 16G, 18G, 20G12G also available, no significant influence on clinical use.
Length (mm)100mm, 160mm, 200mm, 250mm80mm, 100mm, 130mm, 1600mm, 200mm, 250mmDifferent lengths available, no significant influence on clinical use.
Slot size18mm18mmSame.
BiocompatibilityISO 10993-1, ISO 10993-5, ISO 10993-10ISO 10993-1, ISO 10993-5, ISO 10993-10Same standards met.
Sterilization MethodEO sterilizationEO sterilizationSame.
Puncture Force ValuePredicate: 5.3NSubject: 2.7NSubject device has lower puncture force.
Resistance during operationPredicate: noneSubject: noneSame.
Abnormal noise during operationPredicate: noneSubject: noneSame.
Connection firmness (needle tube + plastic parts)Predicate: 181 to 230NSubject: 190 to 212NWithin comparable range.
Ex-vivo tissue studies (sample length & weight)Similar length, weight between predicate and subject.Equivalent tissue samples in terms of length and weight (porcine muscle, liver).Equivalent.
CytotoxicityPer ISO 10993-5In vitro cytotoxicity performed.Demonstrated compliance.
Irritation and Skin SensitizationPer ISO 10993-10Tests performed.Demonstrated compliance.
Systemic ToxicityPer ISO 10883-11Test performed.Demonstrated compliance.
Sterility TestPer ISO 11138-2, ISO 11737-2Sterility test performed.Demonstrated compliance.
EO ResidualsPer EN ISO 10993-71.13 mg/deviceDemonstrated compliance.
Packaging & Shelf-lifePer ISTA2A and ASTM F1980-16 for 3-year shelf life.3-year shelf life.Demonstrated compliance.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "ex-vivo tissue studies (porcine muscle, liver)" which suggests an experimental test set. However, it does not specify the sample size for these ex-vivo studies beyond stating "equivalent tissue samples." The data provenance is implied to be laboratory testing rather than human clinical data. No country of origin for this specific testing data is given, but the manufacturer is based in China. This is not a clinical study on patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. The comparisons and tests described are for physical and material characteristics of the device, not diagnostic performance requiring expert interpretation of results.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This device is not an imaging or diagnostic AI device requiring adjudication of human expert interpretations. The testing focuses on mechanical and material properties.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The URAMIX CuraWay Automatic Core Biopsy Instrument is a physical medical instrument (a biopsy needle), not an AI device or an imaging/diagnostic AI software. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant to this device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an algorithm or AI-driven device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the ex-vivo tissue studies, the ground truth was implied by the objective measurements of sample length and weight taken from the porcine tissues. For other non-clinical tests (biocompatibility, sterility, etc.), the ground truth is compliance with recognized international standards (e.g., ISO, ASTM).

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. No training set is relevant for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below the words "U.S. FOOD & DRUG".

September 17, 2021

Uramix, Inc. % Michael Mooreville, MD Medical Director Uramix, Inc 272 N. Lansdowne Ave. Lansdowne, Pennsylvania 19050

Re: K203141

Trade/Device Name: URAMIX CuraWay Automatic Core Biopsy Instrument Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: July 21, 2021 Received: July 21, 2021

Dear Dr. Mooreville :

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approv al of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203141

Device Name

Uramix Curaway Automatic Core Biopsy Instrument

Indications for Use (Describe)

The Uramix Curaway Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as liver, kidneys, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone, or breast.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SUBMITTER

Uramix, Inc 272 N. Lansdowne Ave. Lansdowne, PA 19050

Contact Person:Michael Mooreville, MD
Medical Director
Telephone:(610) 304-9070
Email:uramix@comcast.net

DEVICE

Name of the Device: URAMIX CuraWay Automatic Core Biopsy Instrument Regulation Number: 21 CFR 876.1075 Regulation Name: Instrument, Biopsy/Regulatory Class II Regulatory Specialty: Gastroenterology/Urology KNW Product code:

Manufacturer: Zhejiang CuraWay Medical Technology Co., Ltd Company Address: Room 106, Building 1, No.600, 21st Street, Qiantang New Area, Hangzhou, Zhejiang, China

PREDICATE DEVICE

K133948, BARD® MAX-CORE® Disposable Core Biopsy Primary:

DEVICE DESCRIPTION

The URAMIX CuraWay Automatic Core Biopsy Instrument is a single use only core biopsy device and is individually packaged and sterilized. The packaging is compatible with the product's EO sterilization method. The sterilization validation confirms the packaging is qualified to maintain the sterilization condition of the device. It is available in several needle gauge sizes and lengths. The side and rear actuator buttons are color coded according to the various gauge sizes. The 18 gauge is the one used for current intended marketing, for prostate and kidney biopsies.

{4}------------------------------------------------

K203141

510(k) Summary

ABN 1208,ABN 1208AABN 1408,ABN 1408AABN 1608,ABN 1608AABN 1808,ABN 1808AABN 2008,ABN 2008A
ABN 1210,ABN 1210AABN 1410,ABN 1410AABN 1610,ABN 1610AABN 1810,ABN 1810AABN 2010,ABN 2010A
ABN 1213,ABN 1213AABN 1413,ABN 1413AABN 1613,ABN 1613AABN 1813,ABN 1813AABN 2013,ABN 2013A
ABN 1216,ABN 1216AABN 1416,ABN 1416ABN 1616,ABN 1616AABN 1816,ABN 1816AABN 2016,ABN 2016A
ABN 1220,ABN 1220AABN 1420,ABN 1420AABN 1620,ABN 1620AABN 1820,ABN 1820AABN 2020,ABN 2020A
ABN 1225,ABN 1225AABN 1425,ABN 1425AABN 1625,ABN 1625AABN 1825,ABN 1825AABN 2025,ABN 2025A
12G(2.7mm)Blue14G(2.1mm)Green16G(1.7mm)Violet18G(1.2mm)Pink20G(0.9mm)Yellow

According to gauge and length, the following variants are available:

The operating handle is color coded and the Type A is a variant in the position of the needle markings, which provide reference for depth of placement.

INDUCATIONS FOR USE

The URAMIX CuraWay Automatic Core Biopsy Instrument is intended for use in obtaining core biopsy samples from soft tissue such as kidney, liver, prostrate, spleen, lymph nodes, and various soft tissue tumors. It is not intended for use in bone, or breast.

{5}------------------------------------------------

COMPARISON TO PREDICATE

Predicate DeviceSubject Device
ItemBARD® MAX-CORE® DisposableCore Biopsy Instrument, K133948URAMIX CuraWay Automatic CoreBiopsy Instrument, K203141Outcome
IntendedUseObtaining biopsies from soft tissuessuch as liver, kidney, thyroid,prostate, spleen, lymph nodes andvarious soft tissue tumors.It is not intended for use in bone.The URAMIX Curaway AutomaticCore Biopsy Instrument isintended for use in obtainingbiopsies from soft tissues suchas liver, kidney, prostate, spleen,lymph nodes and various softtissue tumors. It is not intendedfor use in bone, or breast.It was used for the sameclinical condition, and sameintended purpose, and at thesame site.
MaterialStainless steel + ABSThe shaft of cutting cannula and innerstylet is made of medical grade stainlesssteel (SUS304/06Cr19Ni10). The operatinghandle and actuator button are made ofacrylonitrile butadiene styrene (ABS).Same material used
Types/Sizes14G, 16G, 18G, 20G12G, 14G, 16G, 18G, 20G12G is also available fromZhejiang CuraWay, this hasno significant influence onclinical use.
DesignAttributesEach needle has numerically orderedcentimeter markings on the outercannula to provide reference for depthplacementNeedles feature an adjustable needlestop which allows the user to restrictforward movement, localizing theneedle tip to the biopsy siteColor coded stylet hubs indicategauge size of needles available in avariety of gauge sizes and centimeterlengthsAvailable with an echogenic tip on theouter cannula to promote accurateplacement under ultrasound guidanceAutomatic Core Biopsy Instrument is apuncture needle consisting of an innerstylet and a cutting cannula with etched tip(s. Figure 1). The cutting cannula hasnumerically ordered centimeter markingson it to provide reference for depthplacement. In the distal area of the cuttingcannula, an ultrasound enhancement isavailable to promote accurate placementunder ultrasound or X-ray guidance. Anadjustable depth stopper allows the user torestrict forward movement, localizing theneedle tip to the biopsy site. The Type A isa variant in position of the depth stopper.The operating handle is color-coded, whichindicates gauge size of the needle. Theneedle is protected in a needle sheath.Automatic Core Biopsy Instrument isavailable in several needle gauge sizesand lengths.Very similar design
Length(mm)100mm, 160mm, 200mm, 250mm80mm, 100mm, 130mm, 1600mm, 200mm,250mmThis has no significantinfluence on clinical use.
Slot size18mm18mmSame
Structurea. Photo(examples)Very similar structure anddesign concept
PackagingTyvek 1073B + PETyvek 1073B + PET/PECritical component Tyvek isthe same.Both packaging systems arecommon packaging.
BiocompatibilityISO 10093-1, ISO 10993-5 and ISO10993-10ISO 10093-1, ISO 10993-5 and ISO 10993-10Same.
Sterilization methodEO sterilizationEO sterilizationSame.
ConclusionAfter comparison of the device in question with the predicate device, K133948 regardingclinical, technical and biological characteristics, it can be concluded that both devices havesimilar design, perform the same function under the same clinical conditions. They use thesame materials and have the same durability. Tissue penetration and sample size are the sameand the device is in contact with the same human tissues and body fluids. They can beconsidered as equivalent devices.

{6}------------------------------------------------

SUMMARY OF NON CLINICAL TESTING

The following testing was conducted to demonstrate the safe and effective use of the URAMIX Curaway Automatic Core Biopsy Instrument:

  • Biocompatibility Testing per ISO 10993-5, including in vitro cytotoxicity ●
  • ISO 10993-10 tests for irritation and skin sensitization ●
  • ISO 10883-11 test for systemic toxicity ●
  • Sterilization Testing per ISO 11138-2, and ISO 11737-2 sterility test ●
  • Residuals per EN ISO 10993-7 EO residuals = 1.13 mg/device ●
  • Packaging and Shelf-life per ISTA2A and ASTM F1980-16 3-year shelf life ●

{7}------------------------------------------------

  • Puncture Force Value: Predicate 5.3N; Subject 2.7N ●
  • Resistance during operation: Predicate none; Subject none ●
  • Abnormal noise during operation: Predicate none: Subject none .
  • Connection firmness (needle tube + plastic parts): P 181 to 230N; S 190 to 212N ●
  • Samples of various tissues: similar length, weight between Predicate and Subject .
  • The ex-vivo tissue studies (porcine muscle, liver) demonstrate equivalent tissue . samples in terms of length and weight.

CONCLUSION

Based upon the performance data provided in this submission and comparing indicated use, design, materials, principle of operation and overall technological characterisitics, the URAMIX Curaway Automatic Core Biopsy Instrument has been determined to be substantially equivalent to the predicate device, BARD Max-Core, K133948

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.