K Number
K203132

Validate with FDA (Live)

Date Cleared
2021-04-15

(178 days)

Product Code
Regulation Number
876.5015
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AXIOS™ Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 davs and should be removed upon confirmation of pseudocyst resolution.

Device Description

The AXIOS Stent is a flexible, MR conditional, fully-covered self-expanding braided nitinol stent, which comes preloaded into the delivery system. The AXIOS stent is designed with two flanges on each end to prevent migration and to enable tissue plane apposition and a "saddle" in between the flanges to span the tissue implant distance.

The AXIOS Electrocautery-Enhanced Delivery System consists of a catheter and an integrated handle with manual controls for positioning and deploving the AXIOS Stent. The AXIOS Electrocautery-Enhanced Delivery System is designed to be used in the gastrointestinal tract with commercially available echoendoscopes with a 3.7 mm diameter or larger working channel and is compatible with commercially available 0.035-inch insulated endoscopic guidewires.

The Electrocautery-Enhanced Delivery System connects with an off-the-shelf electrosurgical unit or generator that is compliant to IEC 60601-1-2 and IEC 60601-2-2.

The AXIOS Stent and Electrocautery-Enhanced Delivery System is provided sterile, disposable and intended for single use.

AI/ML Overview

This FDA 510(k) summary describes a medical device, the AXIOS Stent and Electrocautery-Enhanced Delivery System, and its performance. However, it does not involve an AI/ML algorithm, and therefore the acceptance criteria and study information related to AI/ML performance metrics (such as effect size for human readers with AI assistance, type of ground truth, training set size, etc.) are not applicable here.

The document focuses on demonstrating substantial equivalence to a predicate device (K181905) through engineering performance data (bench testing) rather than clinical study data involving AI/ML.

Here's a breakdown of the available information based on your request, with an emphasis on what is not present due to the nature of this device submission:

1. A table of acceptance criteria and the reported device performance

The document provides a table summarizing the types of bench tests performed and their results. The "acceptance criteria" are implied by the "Pass" result for each test, indicating that the device met the pre-defined product specifications for these engineering parameters. Specific numerical acceptance criteria values are not provided, only the pass/fail outcome.

Test ItemReported Result (Pass/Fail)
AXIOS Stent
Deployed Stent Saddle LengthPass
Deployed Stent Saddle Outer DiameterPass
Deployed Stent Flange WidthPass
Stent Pull-out ForcePass
Stent (Saddle) Radial Strength - in compression & expansionPass
Deployment ForcePass
Implant Anchor Function- Retention (tensile)Pass
Magnetic Resonance TestingPass
Fatigue TestingPass
AXIOS 9 Fr. Electrocautery-Enhanced Delivery System
Delivery System Working LengthPass
Catheter ExtensionPass
Nose Lock Hold ForcePass
Slider Lock Hold ForcePass
Nose Lock CyclingPass
Slider Lock CyclingPass
Distal Pusher Catheter to Distal Nose Joint StrengthPass
Pusher Catheter to Hypotube Joint StrengthPass
Outer Sheath to Hypotube Joint StrengthPass
Pusher Hypotube to Handle Joint StrengthPass
Outer Sheath Hypotube to Handle Joint StrengthPass
Distal Pusher to Proximal Pusher Catheter Joint StrengthPass
Outer Sheath to Handle Torque StrengthPass
Luer to Nose Joint StrengthPass
Tracking ForcePass

2. Sample sized used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Sample Size: Not specified for the bench tests. For medical devices, bench testing typically involves a sufficient number of samples to ensure statistical confidence in the results, but the exact N is not disclosed in this summary.
  • Data Provenance: Not applicable as this relates to engineering bench testing of physical device components, not clinical data sets.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

Not applicable. This device is not an AI/ML diagnostic tool requiring expert interpretation for ground truth establishment. The "ground truth" for the performance tests are engineering specifications verified through physical measurement and functional testing.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable. This is not a clinical study involving human interpretation or adjudication of cases.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML powered device, and no MRMC study was conducted or mentioned.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. There is no algorithm to test in a standalone configuration.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for these tests are the pre-defined product specifications and engineering requirements for the device's physical and functional characteristics. This is established through engineering design, material science, and manufacturing standards.

8. The sample size for the training set

Not applicable. This is not an AI/ML device, so there is no training set.

9. How the ground truth for the training set was established

Not applicable. As there is no AI/ML component, there is no training set or associated ground truth establishment process.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health and Human Services seal, and on the right is the FDA acronym in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The text is stacked, with "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below.

April 15, 2021

Boston Scientific Corporation Aparna Philip Regulatory Affairs Specialist II 100 Boston Scientific Way Marlborough, MA 01752

Re: K203132

Trade/Device Name: AXIOS Stent and Electrocautery Enhanced Delivery System Regulation Number: 21 CFR§ 876.5015 Regulation Name: Pancreatic Drainage Stent and Delivery System Regulatory Class: II Product Code: PCU, KNS Dated: March 15, 2021 Received: March 16, 2021

Dear Aparna Philip:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Je An -S Digitally signed
by Je An -S

Je Hi An, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203132

Device Name

AXIOSTM Stent and Electrocautery-Enhanced Delivery System

Indications for Use (Describe)

The AXIOS™ Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 davs and should be removed upon confirmation of pseudocyst resolution.

Type of Use (Select one or both, as applicable):

Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K203132 Page 1 of 5

SECTION 5: 510(K) SUMMARY

510(k) SUMMARY

1. Submitter:

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752

Primary Contact: Aparna Philip Regulatory Affairs Specialist II Telephone: 508-382-0259 E-mail: Aparna.Philip@bsci.com

Date Prepared: October 16, 2020

2. Device:

Trade Name:

Device Common Name:

Classification Name: Regulation Number:

Product Code: Classification:

3. Predicate Device

Trade Name:

510(k) Number: Device Common Name:

Classification Name:

Regulation Number:

Product Code:

Classification:

AXIOSTM Stent and Electrocautery-Enhanced Delivery System Pancreatic drainage stent and delivery system & endoscopic electrosurgery device Pancreatic Drainage Stent and Delivery System 21CFR 876.5015 21CFR 876.4300 PCU/KNS Class II

AXIOSTM Stent and Electrocautery-Enhanced Delivery System & AXIOS Stent and Delivery System K181905 Pancreatic drainage stent and delivery system & endoscopic electrosurgery device Pancreatic drainage stent and accessories and endoscopic electrosurgery accessories 21CFR 876.5015 21CFR 876.4300 PCU/ KNS Class II

{4}------------------------------------------------

4. Device Description

AXIOS Stent:

The AXIOS Stent is a flexible, MR conditional, fully-covered self-expanding braided nitinol stent, which comes preloaded into the delivery system. The AXIOS stent is designed with two flanges on each end to prevent migration and to enable tissue plane apposition and a "saddle" in between the flanges to span the tissue implant distance.

Electrocautery-Enhanced Delivery System:

The AXIOS Electrocautery-Enhanced Delivery System consists of a catheter and an integrated handle with manual controls for positioning and deploving the AXIOS Stent. The AXIOS Electrocautery-Enhanced Delivery System is designed to be used in the gastrointestinal tract with commercially available echoendoscopes with a 3.7 mm diameter or larger working channel and is compatible with commercially available 0.035-inch insulated endoscopic guidewires.

The Electrocautery-Enhanced Delivery System connects with an off-the-shelf electrosurgical unit or generator that is compliant to IEC 60601-1-2 and IEC 60601-2-2.

The AXIOS Stent and Electrocautery-Enhanced Delivery System is provided sterile, disposable and intended for single use. Table 5-1 below discusses the main features of the AXIOS Stent and Electrocautery-Enhanced Delivery System.

Component/DesignFeature Description
CatheterProvided sterile, for single-patient use●Working length: 138 cm Electrocautery-Enhanced Delivery SystemOuter Diameter 9 FrFluoroscopy: AXIOS Stent is containedbetween two (2) Platinum IridiumMarkersElectrocautery Tip for precise cuttingMonopolar 750VP or 1500Vp-p RatedAccessory VoltageIEC 60601-1 compliantO
HandleStaged delivery system for precise stentplacement=>Two (2)-step release of each flange,including a full "stop"=>Lock-out after the release of the firstflange, preventing unintendeddeployment of the second flange
Guidewire Compatibility0.035" insulated guidewires
Endoscope CompatibilityCompatible with 3.7 mm diameter orlarger working channel
Delivery system is luer-locked to the proximal end of the biopsy port of the endoscope
Electrosurgical Unit or GeneratorSelect an electrosurgical unit or generator that is compliant to IEC 60601-1-2 and IEC 60601-2-2. Recommended electrosurgical units or generators include: ERBE VIO 300D ERBE ICC 200 ERBE VIO 300S ERBE VIO 200D
AXIOS Stent DesignBi-flange or double anchor for Staged and Precise positioning Flange/anchor designed to: ⇒hold tissue layers in apposition ⇒prevent migration MR Conditional Provided sterile, for single-patient use
AXIOS Stent LumenStent lumen diameter and short flow path/conduit to ⇒Facilitate passive efficient pseudocyst drainage
AXIOS Stent MaterialNitinol (Nickel-Titanium) ⇒Shape memory and superelasticity for controlled placement and optimal deployment ⇒Corrosion resistant and biocompatible
AXIOS Stent CoveringFully covered with Silicone ⇒Well tolerated by surrounding tissue to minimize tissue ingrowth ⇒Provides leak protection and minimizes tissue ingrowth allowing for atraumatic stent removal
AXIOS Stent VisualizationThe Stent is delivered constrained within a delivery system and deployed under visualization ⇒EUS confirmation of first flange deployment ⇒Direct endoscopic or EUS viewing of second flange deployment ⇒Radiopacity of Nitinol allows fluoroscopy of deployed stent

Table 5-1: AXIOS Stent and Electrocautery-Enhanced Delivery System- Main Features

{5}------------------------------------------------

{6}------------------------------------------------

5. Indications for Use:

The AXIOS Stent and Electrocautery-Enhanced Delivery System is indicated for use to facilitate transgastric or transduodenal endoscopic drainage of symptomatic pancreatic pseudocysts ≥ 6 cm in size, that are adherent to the gastric or bowel wall and are free of solid debris. The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst resolution.

6. Technological Characteristics

The proposed AXIOS Stents (6 mm x 8 mm and 8 mm x 8 mm) and Electrocautery-Enhanced Delivery System retains similar performance compared to the predicate AXIOS Stent (10 mm x 10 mm) despite the dimensional differences of the stent and delivery system. The intended use, and over all mode of operation remains identical to the predicate AXIOS Stent and Electrocautery-Enhanced Delivery system cleared via K181905.

7. Performance Data

The proposed AXIOS Stent and Electrocautery-Enhanced Delivery System successfully passed all pre-defined product specifications for the tests performed. Below is a summary of the tests performed to show the proposed device satisfied all design verification and validation requirements.

Table 5-2 Summary of Bench Tests and Results
TestResults (Pass/ Fail)
AXIOS Stent
Deployed Stent Saddle LengthPass
Deployed Stent Saddle Outer DiameterPass
Deployed Stent Flange WidthPass
Stent Pull-out ForcePass
Stent (Saddle) Radial Strength -in compression & expansionPass
Deployment ForcePass
Implant Anchor Function- Retention (tensile)Pass
Magnetic Resonance TestingPass
Fatigue TestingPass
AXIOS 9 Fr. Electrocautery-Enhanced Delivery System
Delivery System Working LengthPass
Catheter ExtensionPass
Nose Lock Hold ForcePass
Slider Lock Hold ForcePass
Nose Lock CyclingPass
Slider Lock CyclingPass
Distal Pusher Catheter to Distal Nose Joint StrengthPass
Pusher Catheter to Hypotube Joint StrengthPass
Outer Sheath to Hypotube Joint StrengthPass
Pusher Hypotube to Handle Joint StrengthPass
Outer Sheath Hypotube to Handle Joint StrengthPass

{7}------------------------------------------------

Distal Pusher to Proximal Pusher Catheter Joint StrengthPass
Outer Sheath to Handle Torque StrengthPass
Luer to Nose Joint StrengthPass
Tracking ForcePass

Conclusion

The information Boston Scientific Corporation provided in this submission demonstrates that the proposed AXIOS 6 mm x 8 mm x 8 mm x 8 mm Stent and Electrocautery-Enhanced Delivery System is substantially equivalent to the currently cleared AXIOS 10 mm x 10 mm Stent and Electrocautery-Enhanced Delivery System K181905.

§ 876.5015 Pancreatic drainage stent and delivery system.

(a)
Identification. A pancreatic drainage stent is a prescription device that consists of a self-expanding, covered, metallic stent, intended for placement to facilitate transmural endoscopic drainage of pancreatic pseudocysts. This stent is intended to be removed upon confirmation of pseudocyst resolution. This device may also include a delivery system.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The device and elements of the delivery device that may contact the patient must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of patient-contacting components of the device.
(3) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the requested shelf life.
(4) Non-clinical testing data must demonstrate that the stent and delivery system perform as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing of the stent and delivery system must be conducted under simulated use conditions.
(ii) Removal force testing must be conducted. The removal force testing must demonstrate that the stent can be safely removed, and that the stent will remain in place when subjected to forces encountered during use.
(iii) Expansion force testing must be conducted. The expansion force must demonstrate that the forces exerted by the stent will not damage the tissue surrounding the stent.
(iv) Compression force testing must be conducted. The compression force must demonstrate that the stent will withstand the forces encountered during use.
(v) Dimensional verification testing must be conducted.
(vi) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(vii) Fatigue testing must be conducted. Material strength must demonstrate that the stent will withstand forces encountered during use.
(viii) Corrosion testing must be conducted. Corrosion resistance must demonstrate that the stent will withstand conditions encountered during use.
(5) Non-clinical testing must evaluate the compatibility of the stent in a magnetic resonance (MR) environment.
(6) Well-documented clinical experience must demonstrate safe and effective use, and capture any adverse events observed during clinical use.
(7) Labeling must include the following:
(i) Appropriate instructions, warnings, cautions, limitations, and information related to the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device.
(ii) A warning that the safety and patency of the stent has not been established beyond the duration of the documented clinical experience.
(iii) Specific instructions and the qualifications and clinical training needed for the safe use of the device, including deployment of the device, maintenance of the drainage lumen, and removal of the device.
(iv) Information on the patient population for which the device has been demonstrated to be effective.
(v) A detailed summary of the clinical experience pertinent to use of the device.
(vi) A detailed summary of the device technical parameters.
(vii) A detailed summary of the device- and procedure-related complications pertinent to use of the device.
(viii) An expiration date/shelf life.