K Number
K203111

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-09-16

(336 days)

Product Code
Regulation Number
870.4350
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to 6 hours during CPB surgery.

The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures. The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures. The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

The Affinity Pixie hollow fiber oxygenator with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.

The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration. The Affinity Pixie cardiotomy/venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.

The Affinity Pixie Hollow Fiber Oxygenator with Balance Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/min. The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Pixie Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/ min. The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Pixie Cardiotomy/Venous Reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.

Device Description

The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration. The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Cortiva BioActive Surface or Balance Biosurface bonded to the blood contacting surface of the device. The integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery. Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.

The Affinity Pixie Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration. The Affinity Pixie hollow fiber oxygenator is a single-use, microporous, hollow-fiber, gas exchange device with plasma-resistant fiber and integrated heat exchanger. The primary blood-contacting surfaces of the oxygenator are coated with either Balance Biosurface or Cortiva BioActive Surface. Some models are packaged with an Affinity Pixie Cardiotomy/Venous Reservoir (CVR) which are designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Pixie CVR associated with the Balance or Cortiva coated Oxygenators are also Balance or Cortiva coated respectively. The Affinity Pixie CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Pixie CVR may be used during vacuum assisted venous drainage (VAVD) procedures. All are provided sterile, nonpyrogenic, and for single use.

AI/ML Overview

I am sorry, but the provided text does not contain information about acceptance criteria or a study proving device performance in the context of an AI/ML device. The document is a 510(k) premarket notification for a medical device (cardiopulmonary bypass oxygenators) and focuses on the device description, indications for use, and comparison to predicate devices, primarily regarding a material modification. It discusses the device's function (oxygenation, CO2 removal, temperature control, filtering emboli, blood collection) but does not include any performance metrics for AI or ML algorithms, sample sizes for test sets, expert ground truth adjudication, or comparative effectiveness studies involving AI. Therefore, I cannot generate the requested information based on the given input.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 16, 2021

Medtronic, Inc. Kaitlin Cady Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428

Re: K203111

Trade/Device Name: Affinity Fusion Oxygenator with Integrated Arterial Filter and Cortiva™ BioActive Surface (Model CB811), Affinity Fusion Oxygenator with Integrated Arterial Filter and Cortiva™ BioActive Surface and Cardiotomy/Venous Reservoir with Balance™ Biosurface (Model CB841), Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Model BB811). Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BB841), Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211), Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241), Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211), Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241) Regulation Number: 21 CFR 870.4350

Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: August 9, 2021 Received: August 10, 2021

Dear Kaitlin Cady:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and

{1}------------------------------------------------

adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Nicole Gillette Assistant Director (Acting) DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cortiva™ BioActive Surface (Model CB811)

Indications for Use (Describe)

The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.

The Affinity Fusion oxygenator with inter and Cortiva bioactive surface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to 6 hours during CPB surgery.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{3}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Fusion Oxygenator with Integrated Arterial Filter and CortivalM BioActive Surface and Cardiotomy/Venous Reservor with Balance™ Biosurface (Model CB841)

Indications for Use (Describe) Oxygenator with Integrated Arterial Filter

The Affinity Fusion oxygenator with inter and Cortiva bioactive surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.

The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to 6 hours during CPB surgery.

Cardiotomy/Venous Reservoir

The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.

The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{4}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Model BB811)

Indications for Use (Describe)

The Affinity Fusion Oxygenator with Integrated Arter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Inter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Type of Use ( Select one or both, as applicable )
----------------------------------------------------------

|X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

{5}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BB841)

Indications for Use (Describe) Oxygenator with Integrated Arterial Filter

The Affinity Fusion Qxvgenator with Inter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Inter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Cardiotomy/Venous Reservoir

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{6}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211)

Indications for Use (Describe)

The Affinity Pixie hollow fiber oxygenator with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.

The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.

Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{7}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241)

Indications for Use (Describe)

The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatic patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.

The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.

The Affinity Pixie cardiotomy/venous reservoir is intended to be used in an extracorporeal perfusion orcuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{8}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211)

Indications for Use (Describe)

The Affinity Pixie Hollow Fiber Oxygenator with Balance Biosurface are indicated for use for neonate, infant, and small pediativ patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/min.

The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 206 Subpart D)☐ Over-The-Counter Use (21 CFR 201 Subpart G)
------------------------------------------------------------------------------------------------------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{9}------------------------------------------------

510(k) Number (if known): K203111

Device Name

Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241)

Indications for Use (Describe)

The Affinity Pixie Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/ min.

The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Pixie Cardiotomy/Venous Reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

{10}------------------------------------------------

510(k) Summary

Date Prepared: September 16, 2021

Submitter:Medtronic, Inc.Medtronic Perfusion Systems7611 Northland Dr NorthBrooklyn Park, MN 55428Establishment Registration Number: 2184009
Contact Person:Kaitlin CadyRegulatory Affairs SpecialistMedtronic Cardiac SurgeryPhone : 763.526.0278Email : kaitlin.cady@medtronic.com
Alternate Contact :Wendy PinorPrincipal Regulatory Affairs SpecialistMedtronic Cardiac SurgeryPhone : 763.526.3309Email : wendy.pinor@medtronic.com

Device Name and Classification

Trade Name:

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cortiva™ BioActive Surface (Model CB811)

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cortiva™ BioActive Surface and Cardiotomy/Venous Reservoir with Balance™ Biosurface (Model CB841)

Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Model BB811)

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BB841)

Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211)

Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241)

Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211)

Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241)

{11}------------------------------------------------

Common Name:Oxygenator
Regulation Number:21 CFR 870.4350
Product Code:DTZ
Product Classification:II

Name of Predicate Device

K183490 Affinity Fusion Oxygenator K183511 Affinity Pixie Oxygenation System

Device Description

Affinity Fusion Oxygenator

The Affinity Fusion Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator contains both an integrated arterial filter and integrated heat exchanger. The Affinity Fusion Oxygenator is a microporous, hollow-fiber, gas-exchange devices available with Cortiva BioActive Surface or Balance Biosurface bonded to the blood contacting surface of the device.

The integrated arterial filter is designed to filter from the circuit microemboli larger than the specified micron size from the circuit for periods up to six hours during cardiopulmonary bypass surgery.

Some models of the Affinity Fusion Oxygenator are packaged with an Affinity Fusion Cardiotomy/Venous Reservoir (CVR) with Balance Biosurface which is designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Affinity Fusion CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Fusion CVR may be used during vacuum assisted venous drainage (VAVD) procedures and collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement during open heart surgery.

Affinity Pixie Oxygenator

The Affinity Pixie Oxygenator is intended to be used in an extracorporeal perfusion blood circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.

The Affinity Pixie hollow fiber oxygenator is a single-use, microporous, hollow-fiber, gas exchange device with plasma-resistant fiber and integrated heat exchanger. The primary blood-contacting surfaces of the oxygenator are coated with either Balance Biosurface or Cortiva BioActive Surface.

Some models are packaged with an Affinity Pixie Cardiotomy/Venous Reservoir (CVR) which are designed to be an integral part of a cardiopulmonary bypass circuit for use during cardiac surgery. The Pixie CVR associated with the Balance or Cortiva coated Oxygenators are also Balance or Cortiva coated respectively. The Affinity Pixie CVR is designed to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to six (6) hours in duration. Additionally, the Affinity Pixie CVR may be used during vacuum assisted venous drainage (VAVD) procedures.

{12}------------------------------------------------

All are provided sterile, nonpyrogenic, and for single use.

Indications for Use

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cortiva™ BioActive Surface (Model CB811)

The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.

The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to 6 hours during CPB surgery.

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cortiva™ BioActive Surface and Cardiotomy/Venous Reservoir with Balance™ Biosurface (Model CB841)

Oxygenator with Integrated Arterial Filter

The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.

The Affinity Fusion oxygenator with integrated arterial filter and Cortiva bioactive surface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to 6 hours during CPB surgery.

Cardiotomy/Venous Reservoir

The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuumassisted venous drainage (VAVD) procedures.

The Affinity Fusion cardiotomy/venous reservoir with Balance biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface (Model BB811)

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

{13}------------------------------------------------

Affinity Fusion Oxygenator with Integrated Arterial Filter and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BB841)

Oxygenator with Integrated Arterial Filter

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface is designed to filter from the circuit microemboli larger than the specified micron size for periods up to six hours during cardiopulmonary bypass surgery.

Cardiotomy/Venous Reservoir

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.

The Affinity Fusion Cardiotomy/Venous Reservoir with Balance Biosurface is also intended for use after open heart surgery to collect autologous blood from the chest and to aseptically return the blood to the patient for blood volume replacement.

Affinity Pixie Oxygenator with Cortiva BioActive Surface (Model CBP211)

The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.

The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.

Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Cortiva BioActive Surface (Model CBP241)

The Affinity Pixie hollow fiber oxygenator and cardiotomy/venous reservoir with Cortiva bioactive surface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass (CPB) procedures requiring a blood flow rate up to 2.0 L/min.

The Affinity Pixie hollow fiber oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine CPB procedures up to 6 hours in duration.

The Affinity Pixie cardiotomy/venous reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum-assisted venous drainage (VAVD) procedures.

{14}------------------------------------------------

Affinity Pixie Oxygenator with Balance Biosurface (Model BBP211)

The Affinity Pixie Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/min.

The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Affinity Pixie Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface (Model BBP241)

The Affinity Pixie Hollow Fiber Oxygenator and Cardiotomy/Venous Reservoir with Balance Biosurface are indicated for use for neonate, infant, and small pediatric patients undergoing cardiopulmonary bypass procedures requiring a blood flow rate up to 2.0 L/min.

The Affinity Pixie Hollow Fiber Oxygenator is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

The Affinity Pixie Cardiotomy/Venous Reservoir is intended to be used in an extracorporeal perfusion circuit to collect venous and cardiotomy suctioned blood during routine cardiopulmonary procedures up to 6 hours in duration. The CVR is also intended for use during vacuum assisted venous drainage (VAVD) procedures.

Comparison to Predicate Devices

A comparison of the modified product to the currently marketed predicate products (K183490, K183511) indicates the following similarities:

  • . Intended Use
  • Technological Characteristics
  • Operating Principle
  • Design Features
  • Base Materials
  • Shelf Life

The purpose of this submission is a modification to use an alternative resin used to mold the temperature monitor adapter component of the oxygenator used on both the Medtronic Affinity Fusion and Pixie Oxygenator product lines. Biocompatibility and design verification testing were submitted to demonstrate that the proposed modifications are similar to the predicate device.

Conclusion

Medtronic has demonstrated that the modifications made to the Affinity Fusion Oxygenator System and the Affinity Pixie Oxygenation System product family described in this submission result in a substantially equivalent device because the fundamental scientific principle, operating principle, design features and intended use are unchanged from the predicate devices.

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”