K Number
K203055

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-02-09

(124 days)

Product Code
Regulation Number
882.5320
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PEEK Customized Cranial Implant Kit is indicated for the augmentation of bony and/or soft tissue deformities in the cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone); including but not limited to, the correction and prevention of persistent temporal hollowing (PTH) in patients 3.5 years of age and older.

Device Description

The PEEK Customized Cranial Implant Kit product provides a customized cranial or craniofacial patient specific implant solution based on patient CT/CBCT data and surgeon input. This submission is to expand the Stryker PEEK CCI product portfolio to include the single stage option.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Stryker PEEK Customized Cranial Implant Kit (K203055). This submission expands an existing product portfolio to include a single-stage option. The device is a custom implant for cranial and craniofacial bone and soft tissue deformities.

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:

1. A table of acceptance criteria and the reported device performance:

Acceptance CriteriaReported Device Performance
BiocompatibilityNo change in material, duration of contact, or reprocessing methods compared to the predicate device.
SterilityNo change in manufacturing processes, materials, or packaging processes compared to the predicate device.
Performance Bench Testing - Implant FitAll V&V activities that were performed met their respective acceptance criteria. All end-user validation tasks were completed, passed successfully, and supports the substantial equivalence of the subject device to the predicate device.
Performance Bench Testing - Single Stage Surgical Procedure with/without surgical aid options (Marking guides/guided; Virtual template/Navigation system; w/out surgical aid)All V&V activities that were performed met their respective acceptance criteria. All end-user validation tasks were completed, passed successfully, and supports the substantial equivalence of the subject device to the predicate device.

Study Proving Acceptance Criteria:

The study that proves the device meets the acceptance criteria is an End-User Validation Lab as part of Performance Bench Testing. The document explicitly states: "Additionally, an end-user validation lab was conducted to evaluate the subject device and to support the basis for substantial equivalence."

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document does not specify the sample size (number of implants, tests, or end-users) used for the end-user validation lab.
  • Data Provenance: The document does not explicitly state the country of origin of the data or whether it was retrospective or prospective. It describes a "lab" setting, suggesting a controlled, prospective evaluation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document. While an "end-user validation lab" implies involvement of relevant personnel (likely surgeons or qualified medical professionals), the number and specific qualifications are not detailed.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

The document does not specify any adjudication method for the end-user validation lab.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No, an MRMC comparative effectiveness study was not conducted. This device is a passive implant, not an AI-assisted diagnostic or therapeutic tool.
  • Effect Size: Not applicable.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. The device itself is an implant, not a standalone algorithm. The "single stage surgical procedure with/without surgical aid options" described in the performance bench testing evaluates the implant's fit and the use of aids, which inherently involves human interaction.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for the end-user validation lab appears to be based on successful completion of tasks and meeting pre-defined acceptance criteria for implant fit and surgical procedure performance, likely evaluated by the participating end-users and study observers. The document doesn't mention pathology or long-term outcomes data for this specific validation.

8. The sample size for the training set:

Not applicable. This device is a physical implant, not an AI or machine learning model that requires a training set in the conventional sense.

9. How the ground truth for the training set was established:

Not applicable, as there is no training set for this type of device. The design and manufacturing processes are validated against established standards and predicate device characteristics.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

February 9, 2021

Stryker Zainab Amini Senior Regulatory Affairs Specialist 750 Trade Centre Way - Suite 200 Portage, Michigan 49002

Re: K203055

Trade/Device Name: Stryker PEEK Customized Cranial Implant Kit Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: December 10, 2020 Received: December 11, 2020

Dear Zainab Amini:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K203055

Device Name Stryker PEEK Customized Cranial Implant Kit

Indications for Use (Describe)

The PEEK Customized Cranial Implant Kit is indicated for the augmentation of bony and/or soft tissue deformities in the cranial and craniofacial skeleton (orbital rim, zygoma, and adjacent bone); including but not limited to, the correction and prevention of persistent temporal hollowing (PTH) in patients 3.5 years of age and older.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Stryker logo with a line underneath it. Below the line is the text "K203055, Page 1 of 3". The logo is in black and white.

510(k) Summary

This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.

SUBMITTER I.

510(k) Owner:Stryker Leibinger GmbH & Co. KGBoetzinger Strasse 41D-79111 Freiburg, Germany
Submitter/Author/ContactPerson:Zainab AminiSenior Regulatory Affairs SpecialistStryker Craniomaxillofacial750 Trade Centre WayPortage, MI 49002Phone: 269-290-8072Fax: 877-648-7114
Date prepared:February 8, 2021
II.DEVICE
Trade Name:PEEK Customized Cranial Implant Kit
Common or Usualname:Customized Cranial Implant
Classificationname:Preformed alterable cranioplasty plate 21 CFR §882.5320
Regulatory Class:Class II
Product Code:GWO

III. PREDICATE DEVICE

Predicate: K190229, Stryker PEEK Customized Cranial Implant Kit

This predicate has not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The PEEK Customized Cranial Implant Kit product provides a customized cranial or craniofacial patient specific implant solution based on patient CT/CBCT data and surgeon input. This submission is to expand the Stryker PEEK CCI product portfolio to include the single stage option.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the word "stryker" in a bold, sans-serif font. Below the word "stryker" is a horizontal line. Underneath the line is the text "K203055, Page 2 of 3". The text appears to be part of a document or report.

V. INDICATIONS FOR USE

The PEEK Customized Cranial Implant Kit is indicated for the augmentation and/or restoration of bony and/or soft tissue deformities in the cranial and craniofacial skeleton (orbital rim, zygoma. and adjacent bone); including but not limited to, the correction and prevention of persistent temporal hollowing (PTH) in patients 3.5 years of age and older.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The PEEK Customized Cranial Implant Kit is compared to its predicate device for substantial equivalence based on the following criteria:

  • A. Principle of Operation
  • B. Technological Characteristics

A. Principle of Operation

The basic principle of operation for the PEEK Customized Cranial Implant Kit is to fill bony voids, defects, and contour irregularities in non-load bearing regions of the cranial and craniofacial skeleton.

B. Technological and Operational Characteristics

Both the Subject and Predicate Devices have similar technological characteristics, and the principle of operation, mode of fixation, and design specification and manufacturing processes are all identical. Additionally, the subject device can be used in a single stage surgical procedure with or without marking guides or a navigation system to transfer the resection outline.

VII. PERFORMANCE DATA

Biocompatibility Testing

Biocompatibility and sterility testing of the device is a basis for substantial equivalence. There is no change in the material, duration of contact, or reprocessing methods for the PEEK Customized Cranial Implant Kit. As the design and manufacturing processes, materials, and packaging processes are identical for both the predicate and the subject devices.

Performance Bench Testing

Both the subject and predicate devices are designed similarly and manufactured identically. The majority of the Performance Bench testing of the predicate device is valid for the subject device. Additionally, an end-user validation lab was conducted to evaluate the subject device and to support the basis for substantial equivalence. Table below provides a summary of the end-user validation.

TestTest Methods SummaryResults
End-UserValidation• Single Stage surgical procedurewith/&without the following surgicalaid options:- Marking guides/guided; Virtualtemplate/Navigation system;w/out surgical aid• Implant FitAll V&V activities that were performedmet their respective acceptance criteria.All end-user validation tasks werecompleted, passed successfully, andsupports the substantial equivalence ofthe subject device to the predicatedevice.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Stryker logo in bold, black font. Below the logo, there is a line of text that reads "K203055, Page 3 of 3". The text is in a smaller font size than the logo and is also in black.

Animal Testing

Animal testing was not required as a basis for substantial equivalence.

Clinical Testing

Clinical testing was not required as a basis for substantial equivalence.

VIII. CONCLUSIONS

The evidence presented demonstrates that the subject device is as safe and effective as the legally marketed predicate and does not change nor raise any new questions of safety and effectiveness. In conclusion, in accordance to the requirement of 21 CFR 807.87, the PEEK Customized Cranial Implant Kit is substantially equivalent with respect to the safety and effectiveness to the legally marketed predicate device, K190229.

§ 882.5320 Preformed alterable cranioplasty plate.

(a)
Identification. A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.(b)
Classification. Class II (performance standards).