K Number
K203012

Validate with FDA (Live)

Device Name
Surgical Mask
Manufacturer
Date Cleared
2021-04-11

(192 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. This is a single use device, provided non-sterile.

Device Description

The device is a surgical mask which can be described as non-sterile, single-use, 3 layer, flat-pleated style with ear loops and a nose piece. The outer layer of the surgical mask consists of blue nonwoven spunbond polypropylene and the inner layer consists of carded non-woven polyethylene/polyester. The middle layer consists of melt-blown polypropylene filter. The ear loops consists of 140D nylon and spandex elastic string not made with natural rubber latex. The nose piece is made with white galvanized iron with polyethylene coating.

AI/ML Overview

The provided document is a 510(k) premarket notification for a Surgical Mask. It details the device's characteristics and its comparison to a predicate device, focusing on non-clinical performance data to establish substantial equivalence.

Here's an analysis based on your requested information:

1. Table of Acceptance Criteria and Reported Device Performance

Test ItemTest MethodAcceptance CriteriaReported Device Performance
Fluid Resistance PerformanceASTM F1862Fluid resistant, pass at 80 mmHgLot K-1820-01: Pass, 29 of 32 Passed at 80 mmHg Lot J-0720-02: Pass, 30 of 32 Passed at 80 mmHg Lot J-0920-03: Pass, 29 of 32 Passed at 80 mmHg
Particulate Filtration EfficiencyASTM F2299/F2299M-03≥95%Lot J-0620-01: Pass, ≥97.6% Lot J-0720-02: Pass, ≥98.8% Lot J-0920-03: Pass, ≥96.8%
Bacterial Filtration EfficiencyASTM F2101-01(2019)≥95%Lot J-0620-01: Pass, ≥99.63% Lot J-0720-02: Pass, ≥99.74% Lot J-0920-03: Pass, ≥99.57%
Differential Pressure (Delta-P)ASTM F2100-19 (EN 14683)<5.0 mmH2O/cm²Lot J-0620-01: Pass, Average 3.1 mmH2O/cm² (All < 5.0 mmH2O/cm²) Lot J-0720-02: Pass, Average 3.5 mmH2O/cm² (All < 5.0 mmH2O/cm²) Lot J-0920-03: Pass, Average 4.1 mmH2O/cm² (All < 5.0 mmH2O/cm²)
Flammability16 CFR Part 1610Class 1Lot J-0620-01: Pass, 32 samples did not ignite Lot J-0720-02: Pass, 32 samples did not ignite Lot J-0920-03: Pass, 32 samples did not ignite
Biocompatibility (ISO 10993)(Not directly specified in table, but listed)Non-cytotoxic, Non-sensitizing, Non-irritatingThe document states the device is "Non-cytotoxic, Non-sensitizing, Non-irritating," implying it meets these criteria. Specific test results are not detailed in the provided table but are asserted.

2. Sample Size Used for the Test Set and the Data Provenance

  • Fluid Resistance Performance: 32 surgical masks per lot, tested across 3 lots (K-1820-01, J-0720-02, J-0920-03).
  • Particulate Filtration Efficiency: 32 surgical masks per lot, tested across 3 lots (J-0620-01, J-0720-02, J-0920-03).
  • Bacterial Filtration Efficiency: 32 surgical masks per lot, tested across 3 lots (J-0620-01, J-0720-02, J-0920-03).
  • Differential Pressure (Delta-P): 32 surgical masks per lot, tested across 3 lots (J-0620-01, J-0720-02, J-0920-03).
  • Flammability: 32 samples per lot, tested across 3 lots (J-0620-01, J-0720-02, J-0920-03).

The data provenance is non-clinical performance data generated from laboratory testing of the device for various physical and biological properties. The document does not specify the country of origin where the testing was conducted, nor does it classify the data as "retrospective" or "prospective" in the context of clinical studies, as these are non-clinical lab tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable to this submission. The "ground truth" for the performance criteria of a surgical mask (e.g., filtration efficiency, fluid resistance) is established by industry standards (like ASTM F2100-19, ASTM F1862, ASTM F2299/F2299M-03, ASTM F2101-01, 16 CFR Part 1610, ISO 10993) and regulatory guidance, not by expert consensus on individual cases. The tests are objective measurements performed in laboratories.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies involving human interpretation (e.g., radiology reads) to resolve discrepancies among experts. For objective laboratory performance tests of a surgical mask, there is no human interpretation or adjudication process in the same sense. The results are quantitative measurements against predefined pass/fail criteria.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic or decision support systems where human readers interpret medical images or data. The submitted device is a surgical mask and does not involve AI or human readers for its intended use.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This concept applies to AI-driven devices. The surgical mask is a physical product, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the performance criteria of the surgical mask is based on established industry standards and regulatory requirements (e.g., ASTM standards, CFR regulations). These standards define the acceptable performance levels for specific physical and biological properties of surgical masks. The test results are objective measurements compared against these predefined numerical criteria.

8. The Sample Size for the Training Set

This information is not applicable. The device is a physical product (surgical mask), not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable, as there is no training set for a physical device like a surgical mask.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 11, 2021

Customfab, INC. % Laura Nygard RAQA Consultant, Lean RAQA Lean RAQA, LLCX 12602 N Summer Wind Drive Marana, Arizona 85658

Re: K203012

Trade/Device Name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: March 8, 2021 Received: March 9, 2021

Dear Laura Nygard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega -S

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K203012

Device Name Surgical Mask

Indications for Use (Describe)

The Surgical Mask is intended to be worn to protent and healtheare personnel from transfer of microorganisms, body fluids and particulate material in infection control practices to reduce of the wearer to blood and body fluids. This is a single use device, provided non-sterile.

Type of Use (Select one or both, as applicable)

Over-The-Counter Use (21 CFR 801 Subpart C) _ Prescription Use (Part 21 CFR 801 Subpart D)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displaysa currently valid OMB number."

{3}------------------------------------------------

1. Applicant Information

Company Name Establishment Registration Number Phone Number Company Street Address Fax Number City State Country Zip Code

CUSTOM FAB, INC. 2031000 (714) 891-9119 7345 Orangewood Ave 714-891-1699 GARDEN GROVE CA US 92841

2. Contact Person

Full NameErentia Gillmer
Job TitleBrand Director
Phone(714) 891-9119 X 231
Emailerentia@customfabusa.com

3. Correspondent Information

Full NameLaura Nygard
Job TitleRAQA Consultant, Lean RAQA
Phone(734) 807-1282
Emaillauran@leanraqa.com

4. Date of Preparation

Date of Preparation

04/05/2021

5. Device Information

Table 1 - Device Information

Trade NameSurgical Mask
Common or Usual NameSurgical Apparel
Classification Name21 CFR 878.4040
Regulatory Class2
Product CodeFXX

{4}------------------------------------------------

6. Predicate Devices

Table 2 - Predicate Device(s)

Predicate Type510(k) NumberName of DeviceName of Manufacturer
Primary DeviceK201729Medical MaskZhende Medical Co., Ltd

7. Device Description

The device is a surgical mask which can be described as non-sterile, single-use, 3 layer, flat-pleated style with ear loops and a nose piece. The outer layer of the surgical mask consists of blue nonwoven spunbond polypropylene and the inner layer consists of carded non-woven polyethylene/polyester. The middle layer consists of melt-blown polypropylene filter. The ear loops consists of 140D nylon and spandex elastic string not made with natural rubber latex. The nose piece is made with white galvanized iron with polyethylene coating.

8. Intended Use/ Indications for Use

The Surgical Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material in infection control practices to reduce the potential exposure of the wearer to blood and body fluids. This is a single use device, provided non-sterile.

9. Comparison of Technological Characteristics

DeviceProposed DevicePredicate DeviceRemark
ManufacturerCUSTOMFAB, INC.Zhende Medical Co., LtdN/A
510(k) NumberK203012K201729N/A
Product Common NameSurgical maskMedical Face MaskSimilar
Product CodeFXXFXXSame
ClassificationClass II 21 CFR 878.4040Class II 21 CFR 878.4040Same
Surgical apparel.Surgical apparel.
The Surgical Mask isThe Medical Masks areSame
intended to be worn tointended to be worn to
protect both the patientprotect both the patient
and healthcare personneland healthcare personnel
from transfer offrom transfer of
microorganisms, bodymicroorganisms, body
fluids and particulatefluids and particulate
material in infectionmaterial. These face masks
control practices toare intended for use in
reduce the potentialinfection control practices to
exposure of the wearer toreduce the potential exposure
blood and body fluids.of the wearer to blood and
This is a single use
device, provided non-body fluids. This is a single
sterile.use, disposable device(s),
provided non-sterile.
Model1 model of the level 1surgical mask.L1171801N/A
Materials
DeviceSurgical MaskSurgical MaskSame
Outer Facing LayerBlue non-wovenspunbondpolypropyleneSpun-bond polypropyleneSimilar
Middle LayerMelt-blownpolypropyleneMelt blownpolypropylene filterSame
Inner Facing LayerNon-wovenpolyester/polypropyleneSpun-bond polypropyleneSimilar
Nose PieceWhite galvanized steelwithpolyethylene coatingMalleable polypropylenewith iron wireSimilar
Ear LoopsNylon and spandexelasticStringPolyester and spandexSimilar
Color(s)Blue and WhiteBlueSimilar
Specifications and Dimensions
Dimension (width)9.5 +/- 0.25 cm9.5 +/- 1.0 cmSimilar
Dimension (length)17.5 +/- 0.25 cm18.0 +/- 1.0 cmSimilar
Mask Style
ShapeFlat-pleatedFlat-pleatedSame
Design Features
Ear Loops or TiesEar loopsEar loopsSame
NosePiece(ifapplicable)YesYesSame
ASTM F2100 LevelLevel 1Level 1Same
UseSingle UseSingle UseSame
OTC UseYesYesSame
SterilityNon-SterileNon-SterileSame
Performance Data
StandardProposed DevicePredicate DeviceRemarks
Fluid ResistancePerformance ASTM F1862Fluid resistant, pass at 80mmHgFluid resistant, pass at 80mmHgSame
Particulate FiltrationEfficiencyASTM F2299/F2299M-03≥95%≥95%Same
Bacterial FiltrationEfficiency ASTM F2101-01(2019)≥95%≥95%Same
Differential Pressure (Delta-P) ASTM F2100-19 (EN14683)16 CFR Part 1610<5.0 mmH20/cm²<5.0 mmH20/cm²Same
Class 1Class 1Same
Biological Performance
StandardProposed DevicePredicate DeviceRemarks
Biocompatibility (ISO 10993)Non-cytotoxic, Non-sensitizing, Non-irritatingNon-cytotoxic, Non-sensitizing, Non-irritatingSame

Table 3 - Comparison of Technological Characteristics

{5}------------------------------------------------

{6}------------------------------------------------

10. Performance Data

Non-Clinical Performance Data 10.1

The proposed Surgical Mask was tested to ASTM F2100-19 Level 1 requirements and pursuant to the FDA Guidance "Surgical Mask - Premarket Notification [510(k)] Submission".

Table 4- Performance Testing Summary

Test ItemTest MethodPass CriteriaResults
Lot Number:K-1820-01Sample Size:32 surgical masksFluid ResistancePerformance ASTMF1862Fluid resistant,pass at 80mmHg,Pass,29 of 32 Passed at 80 mmHg
Lot Number:J-0720-02Sample Size:32 surgical masksPass,30 of 32 Passed at 80 mmHg
Lot Number:J-0920-03Sample Size:32 surgical masksPass,29 of 32 Passed at 80mmHg
Lot Number:J-0620-01Sample Size:32 surgical masksParticulate FiltrationEfficiencyASTM F2299/F2299M-03≥95%Pass,≥97.6%
Lot Number:J-0720-02Sample Size:32 surgical masksPass,≥98.8%
Lot Number:J-0920-03Sample Size:32 surgical masksPass,≥96.8%
Lot Number:J-0620-01Sample Size:32 surgical masksBacterial FiltrationEfficiency ASTM F2101-01(2019)≥95%Pass,≥99.63%
Lot Number:J-0720-02Sample Size:32 surgical masksPass,≥99.74%

{7}------------------------------------------------

Lot Number:J-0920-03Sample Size:32 surgical masksPass,≥99.57%
Lot Number:J-0620-01Sample Size:32 surgical masksDifferential Pressure(Delta-P) ASTM F2100-19 (EN 14683)<5.0 mmH20/cm²Pass,Average 3.1 mmH20/cm²All < 5.0 mmH20/cm²
Lot Number:J-0720-02Sample Size:32 surgical masksPass,Average 3.5 mmH20/cm²All < 5.0 mmH20/cm²
Lot Number:J-0920-03Sample Size:32 surgical masksPass,Average 4.1mmH20/cm²All< 5.0 mmH20/cm²
Lot Number:J-0620-01Sample Size:32 surgical masks16 CFR Part 1610Class 1Pass,32 samples did not ignite
Lot Number:J-0720-02Sample Size:32 surgical masksPass,32 samples did not ignite
Lot Number:J-0920-03Sample Size:32 surgical masksPass,32 samples did not ignite

10.2 Clinical Performance Data

No Clinical study is included in this submission

11. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.