K Number
K202872

Validate with FDA (Live)

Date Cleared
2021-02-04

(129 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Prase PEEK Anterior Cervical Interbody Spacer is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Prase PEEK Anterior Cervical Interbody Spacer is used to facilitate intervertebral body fusion in the cervical spine from the C3 to C7 disc levels using autograft bone. Prase PEEK Anterior Cervical Interbody Spacer is to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage.

Device Description

The Prase PEEK Anterior Cervical Interbody Spacer is intervertebral body fixation devices intended for use as an aid in spinal fusion. The Prase PEEK Anterior Cervical Interbody Spacer is made from PEEK as per ASTM F2026 and Tantalum marker as per ASTM F560. X-ray markers system on the cages permits the identification of cage position and allows post-operative assessment. The devices are available in a variety of different sizes and heights to match more closely the patient's anatomy. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates. The devices are supplied non-sterile. The device is supplied with their specific instruments. The specific instruments are supplied dedicated tray and non-sterile. The device must be used in combination with the dedicated instruments supplied.

AI/ML Overview

The provided document, K202872, is a 510(k) premarket notification for a medical device called "Prase PEEK Anterior Cervical Interbody Spacer." This document describes the device, its indications for use, and its substantial equivalence to previously marketed predicate devices.

However, this document does NOT contain information about acceptance criteria or a study proving that an AI/software device meets acceptance criteria related to its performance.

The "Performance Testing" section (Section 7) of the summary states that mechanical tests were performed according to ASTM standards for intervertebral body fusion devices (e.g., static compression, fatigue, subsidence). These are physical performance tests for the implant itself, not a study evaluating the performance of a digital health or AI-driven device.

Therefore, I cannot provide the requested information regarding:

  • A table of acceptance criteria and reported device performance for an algorithm.
  • Sample sizes, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, or ground truth details for an AI-related study.

The document discusses the regulatory clearance of a physical medical implant (a PEEK interbody spacer), not an AI/software as a medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 4, 2021

GBS Commonwealth Co., Ltd. Mr. Jimmy Kim #C-309, 168 Gasan Digital 1-ro Geumcheon-Gu Seoul South Korea

Re: K202872

Trade/Device Name: Prase PEEK Anterior Cervical Interbody Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP Dated: December 4, 2020 Received: December 11, 2020

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202872

Device Name Prase PEEK Anterior Cervical Interbody Spacer

Indications for Use (Describe)

Prase PEEK Anterior Cervical Interbody Spacer is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Prase PEEK Anterior Cervical Interbody Spacer is used to facilitate intervertebral body fusion in the cervical spine from the C3 to C7 disc levels using autograft bone. Prase PEEK Anterior Cervical Interbody Spacer is to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CER 801 Subpart D)
Over-The-Counter Use (21 CER 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K202872 Page 1 of 3" at the top, indicating a document identifier and page number. Below the text is a logo with the letters "GBS" in large, bold, blue font. Underneath the letters, the word "COMMONWEALTH" is written in a smaller font, also in blue. The logo appears to represent an organization or company named GBS Commonwealth.

SUMMARY

The following summary is being submitted as required by 21 CFR 807.92(a):

Device Identification 1.

Submitter:GBS Commonwealth Co., Ltd.
#C-309, 168 Gasan Digital 1-ro, Geumcheon-Gu Seoul,
South Korea
Phone. 82-2-6925-4469
e-mail: Jimmy.kim@gbscommonwealth.com
Contact Person:Jimmy Kim
Date preparedSeptember, 25, 2020
Trade NamePrase PEEK Anterior Cervical Interbody Spacer
Regulatory ClassClass II
RegulationName/Common NameIntervertebral fusion device with bone graft,Cervical
Classification NameIntervertebral body fusion device(21 CFR 888.3080)
PanelOrthopedic
Product CodeODP

Purpose of 510(k) 2.

The GBS Commonwealth Co. Ltd., submits this submission for Initial product Introduction of Prase PEEK Anterior Cervical Interbody Spacer

3. Predicate or legally marketed devices which are substantially equivalent

  1. Primary Predicate Device : K151677 LnK Cervical Interbody Fusion Cage System

  2. Additional Predicate Device : K192026 Peridot Interverbral body fusion system K200592 The GS Medical Anyplus PEEK Cage system

Description of the Device 4.

The Prase PEEK Anterior Cervical Interbody Spacer is intervertebral body

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text 'K202872 Page 2 of 3' at the top. Below the text is a logo with the letters 'GBS' in large, bold, blue font. Underneath the letters, the word 'COMMONWEALTH' is written in a smaller, bold, blue font. The logo appears to be for a company or organization called GBS Commonwealth.

fixation devices intended for use as an aid in spinal fusion. The Prase PEEK Anterior Cervical Interbody Spacer is made from PEEK as per ASTM F2026 and Tantalum marker as per ASTM F560.

X-ray markers system on the cages permits the identification of cage position and allows post-operative assessment.

The devices are available in a variety of different sizes and heights to match more closely the patient's anatomy. The ends of the implants have machined teeth which are designed to engage with the vertebral body end plates. The devices are supplied non-sterile.

The device is supplied with their specific instruments. The specific instruments are supplied dedicated tray and non-sterile. The device must be used in combination with the dedicated instruments supplied.

5. Indication for Use

Prase PEEK Anterior Cervical Interbody Spacer is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. Prase PEEK Anterior Cervical Interbody Spacer is used to facilitate intervertebral body fusion in the cervical spine from the C3 to C7 disc levels using autograft bone. Prase PEEK Anterior Cervical Interbody Spacer is to be used with supplemental fixation. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage.

6. Comparison of the technological characteristics of the subject and predicate devices

Prase PEEK Anterior Cervical Interbody Spacer is considered substantially equivalent to the primary predicate LnK Cervical Interbody Fusion Cage System K151677and additional predicate device. They are similar in design, material, scientific technologies and indications for use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "K202872 Page 3 of 3" at the top. Below the text is a logo for GBS Commonwealth. The GBS is in large, bold, blue letters, and the word "COMMONWEALTH" is in smaller, bold, blue letters below the GBS.

7. Performance Testing

Mechanical testing includes static compression, static torsion, fatigue compression, fatigue torsion test were performed according to ASTM F2077-18, subsidence testing performed according to ASTM F2267-04.

Conclusion 8.

Prase PEEK Anterior Cervical Interbody Spacer is substantially equivalent to legally marketed predicates.

N/A