K Number
K202861

Validate with FDA (Live)

Date Cleared
2021-08-27

(333 days)

Product Code
Regulation Number
N/A
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

Device Description

Needle Stimulator is an electro-acupuncture device for acupuncture therapy powered by 6 pieces of 1.5V batteries or AC 100-240V. It is composed of a console and 6 channels of electrode cables with alligator type connectors. The console has the operating controls including function knobs or buttons. Needle Stimulator does not equip with acupuncture needles. The practitioners should select legally marketed needles.

AI/ML Overview

Based on the provided text, the device in question is a "Needle Stimulator," an electro-acupuncture stimulator. The document is a 510(k) Summary, which aims to demonstrate that a new device is substantially equivalent to a legally marketed predicate device. This type of submission relies on comparisons to demonstrate safety and effectiveness rather than conducting extensive clinical studies typically associated with novel devices.

Therefore, the acceptance criteria and the "study" proving the device meets them are primarily based on equivalence to a predicate device through engineering and performance testing, rather than a clinical efficacy study with human subjects.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the parameters of the predicate device (K130768) and compliance with relevant safety standards. The reported device performance is presented as a comparison to these predicate specifications.

ParameterAcceptance Criteria (Predicate Device K130768)Reported Device Performance (Model CMNS6-1)Reported Device Performance (Model CMNS6-2)Remarks (from text)
Intended UseElectro-acupuncture stimulator for qualified practitionersSameSameSame
Type of UsePrescription useSameSameSame
Power Source(s)DC 1.5Vx6 Type R14DC 1.5Vx6 Type R14 or AC 100-240VDC 1.5Vx6 Type R14 or AC 100-240VNote 1: AC power passed IEC 60601-1 and EMC tests
Method of Line Current IsolationType BFType BFType BFSame
Patient Leakage Current (Normal)2µA2µA2µASame
Patient Leakage Current (Single Fault)≤50µA≤50µA≤50µASame
Avg DC current through electrodes (no pulses)N/AN/AN/ASame
Number of Output Modes3 (continuous/interrupted/dense-disperse wave)SameSameSame
Number of Output channels6 (3 channels at most work together on single patient)SameSameSame
Synchronous or AlternatingSynchronousSynchronousSynchronousSame
Method of Channel IsolationTransformerTransformerTransformerSame
Regulated Current or Voltage?Voltage ControlVoltage ControlVoltage ControlSame
Software/Firmware/Microprocessor Control?YesYesYesSame
Automatic Overload Trip?NoNoNoSame
Automatic No-Load Trip?NoNoNoSame
Automatic Shut Off?YesYesYesSame
User Override Control?YesYesYesSame
On/Off Status Indicator?YesYesYesSame
Low Battery Indicator?YesYesYesSame
Voltage/Current Level Indicator?YesYesYesSame
Timer Range (minutes)0-60min0-60min1-99minNote 2 (for CMNS6-2): Difference insignificant
Compliance with Voluntary Standards?IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2SameSameSame
Compliance with 21 CFR 898?YesYesYesSame
Weight (grams)740g740gapprox. 657gNote 2 (for CMNS6-2): Difference insignificant
Dimensions [W x H x D]23015555mm23015555mm23818475mmNote 2 (for CMNS6-2): Difference insignificant
Housing Materials & ConstructionABS; Injection moldedABS; Injection moldedABS; Injection moldedSame
WaveformBiphasicBiphasicBiphasicSame
ShapeAsymmetric biphasic square waveAsymmetric biphasic square waveAsymmetric biphasic square waveSame
Max Output Voltage (@500Ω)27V±10%27V±10%27V±10%Same
Max Output Voltage (@2kΩ)60.4V±10%60.4V±10%60.4V±10%Same
Max Output Voltage (@10kΩ)75V±10%75V±10%75V±10%Same
Max Output Current (@500Ω)54mA±10%54mA±10%54mA±10%Same
Max Output Current (@2kΩ)30.2mA±10%30.2mA±10%30.2mA±10%Same
Max Output Current (@10kΩ)7.5mA±10%7.5mA±10%7.5mA±10%Same
Pulse width (Positive)200µs±10%200µs±10%175µs±10%Note 3 (for CMNS6-2): Passed IEC 60601-2-10 tests
Pulse width (Negative)1030µs (5.15 x (+Phase))1030µs (5.15 x (+Phase))1051µs (6 x (+Phase))Note 3 (for CMNS6-2): Passed IEC 60601-2-10 tests
Pulse Period (msec)10~1000ms10~1000ms10~1000msSame
Max. pulse frequency (Rate)1~100Hz±10%1~100Hz±10%1~100Hz±10%Same
Net Charge (µC per pulse)0µC@500Ω, + and – pulses cancel0µC@500Ω, + and – pulses cancel0µC@500Ω, + and - pulses cancelSame
Maximum Phase Charge (µC)9.4µC @500Ω9.4µC @500Ω8.225µC @500ΩNote 4 (for CMNS6-2): Complies with standards
Maximum Average Current (mA)1.08mA @500Ω1.08mA @500Ω0.945mA @500ΩNote 4 (for CMNS6-2): Better safety
Maximum Current Density (mA/cm², r.m.s.)13.3mA/cm²@500Ω9.4mA/cm²@500Ω8.225mA/cm² @500ΩNote 2 (for K130768 vs CMNS6-1), Note 4 (for K130768 vs CMNS6-2). Calculated by different needle surface areas, within safety limit.
Maximum Average Power Density (W/cm²)0.2W/cm²@500Ω0.141W/cm²@500Ω0.1234W/cm²@500ΩNote 2 (for K130768 vs CMNS6-1), Note 4 (for K130768 vs CMNS6-2). Calculated by different needle surface areas, within safety limit.
BiocompatibilityISO10993-5, ISO 10993-10SameSameSame
AccessoriesLead wires, Alligator type connectorsSameSameSame

2. Sample size used for the test set and the data provenance

The document does not describe a "test set" in the context of clinical data for performance evaluation. Instead, the "testing" refers to engineering and compliance testing of the device prototypes themselves. Therefore, the "sample size" is the manufactured device unit or units tested for compliance.

  • Sample Size: Not applicable in the context of a clinical test set. The testing was performed on the device models themselves (CMNS6-1 and CMNS6-2).
  • Data Provenance: Not specified as clinical data. The tests were likely conducted by the manufacturer in China, given their location. The studies are retrospective in the sense that they are proving equivalency to a device already on the market, but the testing on the new device would have been done prospectively for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the ground truth for this type of submission is established through:

  • Engineering specifications and design requirements.
  • Compliance with recognized national and international safety standards (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10).
  • Demonstration of substantial equivalence to a legally marketed predicate device.

No human expert interpretation of clinical data (like radiology reads) is involved in establishing "ground truth" for this device's safety and performance claim.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable as there is no clinical test set requiring human adjudication. The "adjudication" is done through engineering measurements and conformity assessments to established standards and predicate specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. The device is an electro-acupuncture stimulator, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was conducted or required for this type of device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. The device is a physical medical instrument, not a software algorithm. Its performance is inherent in its electrical and mechanical properties, not an algorithm's output.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this submission is:

  • Engineering specifications and performance characteristics of the predicate device (K130768).
  • International and national safety standards (ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10), which define acceptable limits for various electrical and physical parameters, ensuring patient and user safety.
  • The premise that if the new device operates within the same safety and performance parameters as a previously cleared predicate and complies with relevant standards, it is "safe and effective."

8. The sample size for the training set

This is not applicable. This is not an AI/ML device that requires a "training set."

9. How the ground truth for the training set was established

This is not applicable. As it is not an AI/ML device, there is no training set or associated ground truth establishment process.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services - USA. To the right of the seal is the FDA logo in blue, with the letters "FDA" in a bold, sans-serif font. Underneath the letters "FDA" is the text "U.S. FOOD & DRUG ADMINISTRATION" in a smaller, sans-serif font, also in blue. The logo is simple and professional, and it is easily recognizable.

August 27, 2021

Wuxi Jiajian Medical Instrument Co., Ltd Caihong Sun Regulatory Assurance No. 35 Baiqiao Rd., Ehu Town, Xishan District, Wuxi, Jiangsu 214116 China

Re: K202861

Trade/Device Name: Needle Stimulator Regulatory Class: Unclassified Product Code: BWK Dated: August 9, 2021 Received: August 13, 2021

Dear Caihong Sun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jitendra V. Virani -S

For

Amber Ballard, PhD Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202861

Device Name Needle Stimulator

Indications for Use (Describe)

Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary [As required by 21 CFR 807.92]

1. Submission Information: 510(k) Number: K202861 Date: April 1st, 2021 Type of 510(k) Submission: Traditional Modified device Basis for 510(k) Submission: Submitter/Manufacturer: Wuxi Jiajian Medical Instrument Co., Ltd No. 35 Baiqiao Rd., Ehu Town, Xishan District, Wuxi, Jiangsu, China 214116 Contact: Doris Dong [Consultant, from Shanghai CV Technology Co., Ltd.] Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China E-mail: doris.d(@ceve.org.cn Tel: 86 21-31261348 / Fax: 86 21-57712250

2. Device Description:

Proprietary Name:Needle Stimulator
Model:CMNS6-1, CMNS6-2
Common Name:Electro-Acupuncture
Product Code:BWK
Device Class:Unclassified
Review Panel: DeviceNeurology
Description:Needle Stimulator is an electro-acupuncture device for acupuncture therapy powered by 6 pieces of 1.5V batteries or AC 100-240V. It is composed of a console and 6 channels of electrode cables with alligator type connectors. The console has the operating controls including function knobs or buttons. Needle Stimulator does not equip with acupuncture needles. The practitioners should select legally marketed needles.
Indications for use:Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

3. Substantial Equivalence to Predicate device:

Detailed comparison data is included in "Section 10.1 - Substantial Equivalence Discussion" of this 510(k) submission.

Table 1-

ParametersModified DevicePredicate DeviceRemark
1510(k) Number:K202861K130768--
2Marketing clearancedate:NoAugust 16, 2013--
3Device NameNeedle StimulatorJiajian® CMN Stimulator--
4ModelCMNS6-1/--
5ManufacturerWuxi Jiajian Medical Instrument Co.,LtdWuxi Jiajian Medical Instrument Co.,LtdSame
6Intended useNeedle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.Jiajian® CMN Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.Same
7Type of usePrescription usePrescription useSame
Power Source(s)DC 1.5Vx6 Type R14 or AC 100-240VDC 1.5Vx6 Type R14SameNote 1
Method of Line Current IsolationType BFType BFSame
8Patient Leakage-Current----Same
Normal Condition (μΑ)2μΑ2μΑ
Single Fault-Condition (μΑ)≤50μΑ≤50μΑ
9Average DC current through electrodes when device is on but no pulses are being applied (μΑ)N/AN/ASame
10Number of Output Modes3 (continuous wave/ interrupted wave/ Dense-disperse wave)3 (continuous wave/ interrupted wave/ Dense-disperse wave)Same
Number of Output channels:6 (3 channels at most work together on single patient)6 (3 channels at most work together on single patient)Same
11- Synchronous or Alternating?SynchronousSynchronousSame
Method of Channel IsolationTransformerTransformerSame
12Regulated Current or Regulated Voltage?Voltage ControlVoltage ControlSame
13Software/Firmware/Microprocessor Control?YesYesSame
14Automatic Overload Trip?NoNoSame
15Automatic No-Load Trip?NoNoSame
16Automatic Shut Off?YesYesSame
17User OverrideYesYesSame
Control?
18On/OffStatus?YesYesSame
LowBattery?YesYesSame
IndicatorDisplay:Voltage/CurrentLevel?YesYesSame
19Timer(minutes)Range0-60min0-60min
20Compliance withVoluntary Standards?IEC 60601-1, IEC 60601-2-10,IEC 60601-1-2IEC 60601-1, IEC 60601-2-10,IEC 60601-1-2Same
21Compliance with 21CFR 898?YesYesSame
22Weight (grams)740g740gSame
23Dimensions [W x H x D]23015555mm23015555mmSame
24Housing Materials &ConstructionABS; Injection moldedABS; Injection moldedSame
25WaveformBiphasicBiphasicSame
26ShapeAsymmetric biphasic squarewaveAsymmetric biphasic squarewaveSame
27Maximum OutputVoltage (volts)27V±10% @500Ω27V±10% @500ΩSame
60.4V±10% @2kΩ60.4V±10% @2kΩSame
75V±10% @10kΩ75V±10% @10kΩSame
28Maximum OutputCurrent (specify units)54mA±10% @500Ω54mA±10% @500ΩSame
30.2mA±10% @2kΩ30.2mA±10% @2kΩSame
7.5mA±10% @10kΩ7.5mA±10% @10kΩSame
29Pulsewidth(µsec)Positive200µs±10%200µs±10%Same
Negative1030µs (5.15 x (+Phase) )1030µs (5.15 x (+Phase) )Same
30PulsePeriod (msec)10~1000ms10~1000msSame
31Max. pulse frequency(Hz) [or Rate (pps)]1~100Hz±10%1~100Hz±10%Same
32Net Charge (µC perpulse)0µC@500Ω, + and – pulsescancel0µC@500Ω, + and – pulsescancelSame
33MaximumCharge, (µC)Phase9.4µC @500Ω9.4µC @500ΩSame
34Maximum AverageCurrent, (mA)1.08mA @500Ω1.08mA @500ΩSame
35Maximum CurrentDensity, (mA/cm², r.m.s.)9.4mA/cm² @500Ω13.3mA/cm² @500ΩSameNote 2
36Maximum Average0.141W/cm²@500Ω0.2W/cm²@500Ω
Power(W/cm²)Density,
37BiocompatibilityISO10993-5, ISO 10993-10ISO10993-5, ISO 10993-10Same
38AccessoriesLead wires, Alligator type connectorsLead wires, Alligator type connectorsSame

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

Note 1:

Compared with the predicate device, the proposed device can be connected to AC power. Both of the AC power and DC power of the proposed device have passed IEC 60601-1 and EMC tests, and the description of AC power supply has been added to user manual. Therefore, this difference doesn't raise any safety and effectiveness issues.

Note 2:

The maximum current density and maximum power density are different between the proposed device and the predicate device because they are calculated by different needle surface areas. After calculation, both of them are within the safety limit. Therefore, this difference doesn't raise any safety and effectiveness issues.

Detailed comparison data is included in "Section 10.2 - Substantial Equivalence Discussion" of this 510(k) submission.

Table 2-

ParametersModified DevicePredicate DeviceRemark
1510(k) Number:K202861K130768--
2Marketing clearance date:NoAugust 16, 2013--
3Device NameNeedle StimulatorJiajian® CMN Stimulator--
4ModelCMNS6-2/--
5ManufacturerWuxi Jiajian Medical Instrument Co.,LtdWuxi Jiajian Medical Instrument Co.,LtdSame
6Intended useNeedle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.Jiajian® CMN Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.Same
7Type of usePrescription usePrescription useSame
Power Source(s)DC 1.5Vx6 Type R14 or AC 100-240VDC 1.5Vx6 Type R14SameNote 1
- Method of Line Current IsolationType BFType BFSame
8- Patient Leakage Current----Same
-- Normal Condition (μΑ)2μΑ2μΑ
-- Single Fault Condition (μΑ)≤50μΑ≤50μΑ
9Average DC currentN/AN/ASame
throughelectrodes
when device is on but
no pulses are being
applied (µA)
10Number of Output Modes3 (continuous wave/interrupted wave/Dense-disperse wave)3 (continuous wave/interrupted wave/Dense-disperse wave)Same
11Number of Output channels:6 (3 channels at most worktogether on single patient)6 (3 channels at most worktogether on single patient)Same
- Synchronous or Alternating?SynchronousSynchronousSame
- Method of Channel IsolationTransformerTransformerSame
12Regulated Current or Regulated Voltage?Voltage ControlVoltage ControlSame
Software/Firmware/Microprocessor Control?YesYesSame
13
14Automatic Overload Trip?NoNoSame
15Automatic No-Load Trip?NoNoSame
16Automatic Shut Off?YesYesSame
17User Override Control?YesYesSame
18Indicator Display:
On/Off Status?YesYesSame
Low Battery?YesYesSame
Voltage/Current Level?YesYesSame
19Timer (minutes) Range1-99min0-60minSameNote 2
20Compliance with Voluntary Standards?IEC 60601-1, IEC 60601-2-10,IEC 60601-1-2IEC 60601-1, IEC 60601-2-10,IEC 60601-1-2Same
21Compliance with 21 CFR 898?YesYesSame
22Weight (grams)approx. 657g740gSame
23Dimensions [W x H x D]23818475mm23015555mmSameNote 2
24Housing Materials & ConstructionABS; Injection moldedABS; Injection moldedSame
25WaveformBiphasicBiphasicSame
26ShapeAsymmetric biphasic squareAsymmetric biphasic squareSame
wavewave
27MaximumOutputVoltage (volts)27V±10% @500Ω27V±10% @500ΩSame
60.4V±10% @2kΩ60.4V±10% @2kΩSame
75V±10% @10kΩ75V±10% @10kΩSame
28MaximumOutput54mA±10% @500Ω54mA±10% @500ΩSame
Current(specify30.2mA±10% @2kΩ30.2mA±10% @2kΩSame
units)7.5mA±10% @10kΩ7.5mA±10% @10kΩSame
29PulsePositive175µs±10%200µs±10%Same
width(µsec)Negative1051µs (6 x (+Phase) )1030µs (5.15 x (+Phase) )Note 3
30PulsePeriod (msec)10~1000ms10~1000msSame
31Max. pulse frequency(Hz) [or Rate (pps)]1~100Hz±10%1~100Hz±10%Same
32Net Charge (µC perpulse)0µC@500Ω, + and - pulsescancel0µC@500Ω, + and - pulsescancelSame
33MaximumPhaseCharge, (µC)8.225μC @500Ω9.4μC @500ΩSameNote 4
34MaximumAverageCurrent, (mA)0.945mA @500Ω1.08mA @500Ω
35MaximumDensity,r.m.s.)Current(mA/cm²,8.225mA/cm² @500Ω13.3mA/cm² @500Ω
36MaximumPower(W/cm²)AverageDensity,0.1234W/cm²@500Ω0.2W/cm²@500Ω
37BiocompatibilityISO10993-5, ISO 10993-10ISO10993-5, ISO 10993-10Same
38AccessoriesLeadwires, AlligatortypeconnectorsLeadwires, AlligatortypeconnectorsSame

{7}------------------------------------------------

{8}------------------------------------------------

Note 1:

Compared with the predicate device, the proposed device can be connected to AC power. Both of the AC power and DC power of the proposed device have passed IEC 60601-1 and EMC tests, and the description of AC power supply has been added to user manual. Therefore, this difference doesn't raise any safety and effectiveness issues.

Note 2:

The weight, dimensions and appearance of the proposed device are a little different from the predicate device K130768, but these differences are insignificant in the terms of safety or effectiveness. Besides, there is a difference in treatment time range between the proposed device and predicate device. The treatment time can be adjusted by user as they want. So, the difference of treatment time range will not raise any safety or effectiveness issue.

Note 3:

There is a difference in the pulse width between the proposed device and predicate device. Based on the calculation of maximum current density, maximum power density, these parameters don't exceed the safety limit. All deviation and the worst case have been considered in risk analysis report, and all parameters have passed IEC 60601-2-10 test codes. So, the difference doesn't raise any new safety and effectiveness issues.

{9}------------------------------------------------

Note 4:

The Maximum Phase Charge of the proposed device are similar to the predicate device, and both of them comply with IEC 60601-1 and IEC 60601-2-10 requirements. So, the differences will not raise any safety or effectiveness issue.The maximum average current of the proposed device is smaller than that of the predicate device which means the better safety. The maximum current density and maximum power density are different between the proposed device and the predicate device because they are calculated by different needle surface areas. After calculation, both of them are within the safety limit. Therefore, this difference doesn't raise any safety and effectiveness issues.

4. Test summary:

Needle Stimulator is safe and effective as the predicate devices cited above. The new devices have passed testings according to the following standards:

  1. ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD);

  2. IEC 60601-1-2: Edition 4.0 2014-02 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements And Tests;

  3. IEC 60601-2-10 Edition 2.1 2016-04 Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators;

The conclusion drawn from the testings are that the new devices are substantially equivalent to the predicate device. Furthermore, the new device complies with the recognized standards and performs its intended tasks as well as the legally marketed predicate device.

N/A