K Number
K202837

Validate with FDA (Live)

Device Name
GoSpiro
Date Cleared
2020-12-04

(70 days)

Product Code
Regulation Number
868.1840
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.

Device Description

The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.

The GoSpiro spirometer transmits real-time lung function data to computers, tablets or smartphones over a Bluetooth connection for tele-healthcare applications. The GoSpiro performs full flow-volume loops including inspiratory and expiratory data. The internal program performs all of the calculations for measurements to meet American Thoracic Society and European Respiratory Society requirements. It has built-in quality control measurements and transmits indices of measurement quality including time to peak flow, back-extrapolated volume, total expiratory time, and end-expiratory flow detection.

It is used with the GoSpiro App display and communications software on a smartphone or tablet.

The GoSpiro is powered by an internal rechargeable Lithium battery and is charged via its USB charging station connected to a USB power source. The device complies with ES 60601-1, IEC 60601-1-2, and IEC 60601-1-11.

The fundamental technology to measure flow is a vertical turbine volume sensor. The turbine transducer measures expired air directly at B.T.P.S. (body temperature and pressure with saturated water vapor) thus avoiding the requirement for temperature correction on exhalation. An electronic temperature sensor on the device PCB measures atmospheric temperature, thus enabling correction of inspired volumes and flows. This transducer is insensitive to the effects of condensation and temperature and avoids the need for individual calibration prior to performing a test.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the GoSpiro® device, which is a diagnostic spirometer. The primary change introduced in this submission (K202837) is the ability to clean, disinfect, and reuse the turbine for multi-patient, multi-use, as opposed to the predicate device (K163249) where the turbine had to be replaced if no filter was used.

Here's an analysis of the provided information concerning acceptance criteria and studies:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance results for all device functionalities. However, it implicitly uses the predicate device's specifications as the standard for equivalence. The key performance parameters related to spirometry and their stated values are:

Acceptance Criteria (Implied from Predicate)Reported Device Performance (GoSpiro®)
Flow Range: ±14 l/sFlow Range: ±14 l/s
Flow Accuracy: ±5% or 200 mL/sFlow Accuracy: ±5% or 200 mL/s
Flow Resistance: 137 Pascals (Pa) per Liter per second, measured at 14 Liters per second (Lps)Flow Resistance: 137 Pascals (Pa) per Liter per second, measured at 14 Liters per second (Lps)
Volume Range: 0-8 litersVolume Range: 0-8 liters
Volume Accuracy: ±3% of reading, or 0.05 liters, whichever is greaterVolume Accuracy: ±3% of reading, or 0.05 liters, whichever is greater
Compliance with ATS Spirometry guidelinesComplies with ATS Spirometry guidelines
Compliance with AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11Complies with AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-11
Water Ingress Protection: IP22Water Ingress Protection: IP22
Biocompatibility: Externally communicating (Indirect), Tissue and Surface Contact, Mucosa, limited exposureSimilar patient contact and identical materials (Biocompatibility evaluated per ISO 10993-1)
Cleaning of mouthport (predicate)Cleaning of mouthport (Same mouthport); Cleaning and disinfection of Turbine (Proposed)

The primary acceptance criterion that needed separate validation for the proposed device was the effectiveness of turbine cleaning and disinfection. The document states: "The validation testing has been performed demonstrating that the turbine may be cleaned, disinfected and reused." This implies that the acceptance criteria for cleaning and disinfection (e.g., reduction of microbial load, retention of device functionality, no adverse material degradation) were met, though the specific quantitative criteria are not detailed in this excerpt.

2. Sample Size Used for the Test Set and the Data Provenance

The document mentions "High level disinfection validation", "Post-disinfecting performance", and "Post-disinfecting biocompatibility" as types of performance testing.

  • Sample Size: The document does not specify the sample size used for these validation tests (e.g., number of turbines subjected to disinfection, number of cycles tested, or number of biological or performance tests).
  • Data Provenance: The document does not specify the country of origin of the data or whether the studies were retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not provided in the excerpt. The studies mentioned (disinfection, performance, biocompatibility) are laboratory/engineering tests, not clinical studies that typically involve expert interpretation for ground truth.

4. Adjudication Method for the Test Set

This information is not applicable and therefore not provided. The studies described are validation tests (e.g., disinfection efficacy, material properties, device function) rather than clinical evaluations requiring expert adjudication of results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC study is not applicable as the GoSpiro® is a diagnostic spirometer for measuring lung function, not an AI-assisted diagnostic tool that relies on human readers interpreting images or data with and without AI. It directly measures physiological parameters.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This concept is not applicable in the context of the GoSpiro® device. It is a physical medical device that measures spirometry parameters. Its "performance" is determined by the accuracy and precision of its physical measurements, not by an algorithm's standalone diagnostic capability in the absence of human interaction (beyond the user performing the spirometry maneuver). The device and its internal program perform calculations to meet ATS/ERS requirements, and its accuracy targets are detailed above.

7. The Type of Ground Truth Used

For the core spirometry measurements (Flow, Volume, Accuracy, Resistance), the ground truth is implicitly based on established physical and physiological standards and measurement principles (e.g., traceable reference spirometers or calibrated flow/volume sources). For the specific validation related to the new submission:

  • Cleaning and Disinfection Validation: The ground truth would likely be established through standard microbiological testing methods (e.g., reduction of specified microorganisms to acceptable levels) and material compatibility assessments according to recognized standards (e.g., AAMI TIR12, ISO 17664, or specific FDA guidance).
  • Post-disinfecting performance: The ground truth would be that the device, after disinfection, continues to meet its specified flow/volume accuracy and resistance criteria.
  • Post-disinfecting biocompatibility: The ground truth would be that the materials, after disinfection, remain biocompatible as per ISO 10993-1.

8. The Sample Size for the Training Set

This information is not applicable as the GoSpiro® is a physical measurement device and its underlying technology (bidirectional turbine) and internal program for calculations are not described as based on machine learning or AI models requiring a "training set" in the conventional sense. The "internal program" would be developed based on physics and engineering principles, not statistical learning from data.

9. How the Ground Truth for the Training Set was Established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

December 4, 2020

Monitored Therapeutics, Inc. % Paul Dryden President - consultant Monitored Therapeutics, Inc. c/o ProMedic, LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704

Re: K202837

Trade/Device Name: GoSpiro® Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: November 3, 2020 Received: November 4, 2020

Dear Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K202837

Device Name

GoSpiro®

Indications for Use (Describe)

The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

PSC Publishing Services (301) 443-6740

вн

{3}------------------------------------------------

Sponsor:Michael Taylor - CEOMonitored Therapeutics, Inc.5995 Shier Rings Road, Ste ADublin, OH 43016Tel: +614-761-3555
Submission Correspondent:Paul DrydenProMedic, LLC
Proprietary or Trade Name:GoSpiro®
Common/Usual Name:Diagnostic Spirometer
Regulation Number:Regulation Code:Product Code:Regulatory Class:21CFR 868.1840Diagnostic SpirometerBZGII
Predicate Device:Monitored Therapeutics, Inc. – K163249 – GoSpiro®

Modification:

Changing the turbine from single patient, multi-patient, multi-patient, multi-use.

Device Description:

The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.

The GoSpiro spirometer transmits real-time lung function data to computers, tablets or smartphones over a Bluetooth connection for tele-healthcare applications. The GoSpiro performs full flow-volume loops including inspiratory and expiratory data. The internal program performs all of the calculations for measurements to meet American Thoracic Society and European Respiratory Society requirements. It has built-in quality control measurements and transmits indices of measurement quality including time to peak flow, back-extrapolated volume, total expiratory time, and end-expiratory flow detection.

It is used with the GoSpiro App display and communications software on a smartphone or tablet.

The GoSpiro is powered by an internal rechargeable Lithium battery and is charged via its USB charging station connected to a USB power source. The device complies with ES 60601-1, IEC 60601-1-2, and IEC 60601-1-11.

The fundamental technology to measure flow is a vertical turbine volume sensor. The turbine transducer measures expired air directly at B.T.P.S. (body temperature and pressure with saturated water vapor) thus avoiding the requirement for temperature correction on exhalation. An electronic temperature sensor on the device PCB measures atmospheric temperature, thus enabling correction of inspired volumes and flows. This transducer is insensitive to the effects of condensation and temperature and avoids the need for individual calibration prior to performing a test.

{4}------------------------------------------------

Indications for Use:

The GoSpiro® is intended to be used by adults and children over 5 years old in physician's offices, clinics and home settings to conduct basic lung function and spirometry testing.

Device Comparison:

Table 1 compares the subject device to the predicate Monitored Therapeutics – GoSpiro®, K163249.

Table 1 - Comparison Predicate to Proposed Device
-------------------------------------------------------------
Technical feature/specificationPredicateK163249 - GoSpiro®ProposedGoSpiro®
Indications for UseThe GoSpiro® is intended to be used byadults and children over 5 years old inphysician's offices, clinics and homesettings to conduct basic lung functionand spirometry testing. It is single-patient use device.The GoSpiro® is intended to be usedby adults and children over 5 years oldin physician's offices, clinics andhome settings to conduct basic lungfunction and spirometry testing.
Environment of usePhysician's offices, clinics and homesettingsPhysician's offices, clinics and homesettings
Patient PopulationAdults and pediatric patients over 5years oldAdults and pediatric patients over 5years old
Use with PFT filter after each useYesYes
Turbine may be cleaned / disinfectedNo, had to be replaced if no filter wasusedCleaned and Disinfected
Technology for measure flow andvolumeBidirectional TurbineBidirectional Turbine
Energy TypeRechargeable 3.7V, 500 mAh LithiumbatteryRechargeable 3.7V, 500 mAh Lithiumbattery
Physical configurationTouchscreen on smartphone or tablet andhard buttonsTouchscreen on smartphone or tabletand hard buttons
Operating ConditionsTemperature - 17 to 35°CHumidity - 30%RH to 75%RHTemperature - 17 to 35°CHumidity - 30%RH to 75%RH
Weight10.5 oz.10.5 oz.
Size3.5 x 4.5"3.5 x 4.5"
Flow Range±14 l/s±14 l/s
Flow Accuracy±5% or 200 mL/s±5% or 200 mL/s
Flow Resistance137 Pascals (Pa) per Liter per second,measured at 14 Liters per second (Lps)137 Pascals (Pa) per Liter per second,measured at 14 Liters per second (Lps)
Volume Range0-8 liters0-8 liters
Technical feature/specificationPredicateProposed
K163249 - GoSpiro®GoSpiro®
Volume accuracy±3% of reading, or 0.05 liters, whicheveris greater.±3% of reading, or 0.05 liters,whichever is greater.
Used with PFT filter and / ormouthpieceYesYes
CommunicationBluetooth to Wi-Fi or CellularBluetooth to Wi-Fi or Cellular
CoachingDisplay of flow-time curves, 6 secondcount-down timer, tone at 6 seconds.Display of flow-time curves, 6 secondcount-down timer, tone at 6 seconds.
Feedback of test qualityYes- Trend dataYes- Trend data
Test storedYesYes
Typical battery life per chargeApproximately 140 measurementsApproximately 140 measurements
Auditory / visual alarmsYesYes
Allows inspiration before forcedexhalationYesYes
Tests Performed*FVC*FVC
FEV0.75, *FEV1, FEV3, FEV6FEV0.75, *FEV1, FEV3, FEV6
FEV/FVC (FER) for 0.75 /1 /3 / 6FEV/FVC (FER) for 0.75 /1 /3 / 6
*PEF*PEF
MMEFMMEF
FEF 25/50/75FEF 25/50/75
FEF25-75 (MEF)FEF25-75 (MEF)
FIV1FIV1
FIVCFIVC
PIFPIF
FIF25-75 (MIF25-75)FIF25-75 (MIF25-75)
FIF 25/50/75FIF 25/50/75
MET25-75MET25-75
FEV0.75/FEV6FEV0.75/FEV6
FEV1/FEV6FEV1/FEV6
FEF50/FVCFEF50/FVC
MMEF/FVC (FEF25-75/FVC)MMEF/FVC (FEF25-75/FVC)
FIV1/FIVC (FIR)FIV1/FIVC (FIR)
R50 (FEF50/FIF50)R50 (FEF50/FIF50)
FETFET
MVV (ind)MVV (ind)
VextVext
Time to PEFTime to PEF
Possible CoughPossible Cough
LAST500VLAST500V
Flow / Volume Curve (on server)Flow / Volume Curve (on server)
Volume / Time Curve (on server)Volume / Time Curve (on server)
Flow / Time CurveVolume / Time Curve (on server)
Page 4 of 4
Technical feature/specificationPredicateProposed
K163249 - GoSpiro®GoSpiro®
ATS Spirometry guidelinesYesYes
Standards ComplianceAAMI ANSI ES 60601-1IEC 60601-1-2IEC 60601-1-11AAMI ANSI ES 60601-1IEC 60601-1-2IEC 60601-1-11
Water Ingress ProtectionIP22IP22
Biocompatibility and Patient ContactExternally communicating (Indirect),Tissue and Surface Contact, Mucosa,limited exposureSimilar patient contact and identicalmaterials
Cleaning and Disinfection validationCleaning of mouthportSame mouthportCleaning and disinfection of Turbine

{5}------------------------------------------------

510(k) Summary

December 4, 2020

{6}------------------------------------------------

510(k) Summary December 4, 2020

  • Indicates parameters displayed for patients only.

Differences vs. Predicate:

The only difference between the predicate and the subject device is the labeling to allow for cleaning, disinfection and reuse of the turbine. The validation testing has been performed demonstrating that the turbine may be cleaned, disinfected and reused.

All other features and specifications are identical to the cleared K163249 device.

Substantial Equivalence Discussion:

As the devices are identical the only discussion relates to the change in the ability to clean, disinfect and reuse the turbine for multi-patient, multi-use. Disinfection is only required when a filter is not used with each patient.

All other comparative features are identical.

  • Indications for Use / Patient Population / Environment of Use ●
  • Prescriptive
  • Design and Technology ●
  • Performance and Specifications ●
  • Compliance with Standards
  • Performance Testing ●
    • High level disinfection validation
    • Post-disinfecting performance o
    • Post-disinfecting biocompatibility о
  • . Biocompatibility
    • The materials have been evaluated per ISO 10993-1 o

Substantial Equivalence Conclusion:

The GoSpiro® is substantially equivalent to the cleared GoSpiro® predicate, K163249.

§ 868.1840 Diagnostic spirometer.

(a)
Identification. A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.(b)
Classification. Class II (performance standards).