K Number
K202794

Validate with FDA (Live)

Date Cleared
2021-12-13

(446 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Surgical Face Masks are is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Surgical Face Masks are single use, three-layer, flat -folded masks with ear loops and nose piece. The Surgical Face Masks are manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of PVC wrapped with metal wire. The surgical face masks will be provided in blue. The surgical face masks are sold non-sterile and are intended to be single use, disposable devices.

AI/ML Overview

The provided text describes the acceptance criteria and the study results for the "Surgical Face Mask: F-Y1-A".

1. Table of acceptance criteria and the reported device performance:

ItemAcceptance CriteriaReported Device Performance
Fluid Resistance Performance (ASTM F1862)29 out of 32 pass at 120 mmHgPass
Particulate Filtration Efficiency (ASTM F2299)≥ 98%Lot1: 99.64%, Lot2: 99.70%, Lot3: 99.63% (Pass)
Bacterial Filtration Efficiency (ASTM F2101)≥ 98%Lot1: 99.6%, Lot2: 99.8%, Lot3: 99.4% (Pass)
Differential Pressure (Delta P) (MIL-M-36954C)< 5.0mmH2O/cm²Lot1: 3.4mmH2O/cm², Lot2: 3.2mmH2O/cm², Lot3: 3.4mmH2O/cm² (Pass)
Flammability (16 CFR 1610)Class 1Class 1 (Pass)
Irritation (ISO 10993-10)Under the conditions of the study, the device is non-irritatingUnder the conditions of the study, the device is non-irritating (Pass)
Sensitization (ISO 10993-10)Under the conditions of the study, the device is non-sensitizingUnder the conditions of the study, the device is non-sensitizing (Pass)
Cytotoxicity (ISO 10993-5)Under the conditions of the study, the device is non-cytotoxicUnder the conditions of the study, the device is non-cytotoxic (Pass)

2. Sample size used for the test set and the data provenance:

  • Sample Size: For Fluid Resistance, Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, and Flammability, 3 non-consecutive lots of the device were tested. For Irritation, Sensitization, and Cytotoxicity, the testing refers to "the conditions of the study" without specifying a particular numerical sample size, though these are typically conducted on biological samples.
  • Data Provenance: The document does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. It is a submission to the FDA from a Chinese manufacturer (Jiande Chaomei Daily Chemicals Co., Ltd. in China), so it is likely that the testing was conducted in China or by a laboratory recognized by the manufacturer. The studies are non-clinical (laboratory-based) and would be considered prospective for the purpose of demonstrating compliance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable as the studies are non-clinical (laboratory performance tests for a surgical face mask) and do not involve expert interpretation or ground truth establishment in the way a diagnostic imaging study would. The acceptance criteria are based on established standards (e.g., ASTM, MIL-M, ISO, CFR).

4. Adjudication method for the test set:

This information is not applicable as the studies are non-clinical performance tests for a surgical face mask, not studies requiring expert adjudication methods. The results are quantitative measurements or assessments against predefined criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable. The device is a surgical face mask, not an AI-assisted diagnostic tool, and therefore, an MRMC comparative effectiveness study involving human readers and AI assistance would not be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not applicable. The device is a physical product (surgical face mask), not an algorithm or software requiring standalone performance testing.

7. The type of ground truth used:

The "ground truth" for this device is based on established industry standards and regulatory requirements (e.g., ASTM F1862, ASTM F2299, ASTM F2101, MIL-M-36954C, 16 CFR 1610, ISO 10993-10, ISO 10993-5). The device's performance is measured against the specified thresholds and criteria outlined in these standards.

8. The sample size for the training set:

This information is not applicable. The device is a physical product (surgical face mask) undergoing performance testing, not a machine learning model that requires a training set.

9. How the ground truth for the training set was established:

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. The word "ADMINISTRATION" is written below "U.S. FOOD & DRUG".

December 13, 2021

Jiande Chaomei Daily Chemicals Co., Ltd Sam Lin Official Correspondent Shanghai Spica Management Consulting Co., Ltd. 609Room, NO.133 Shengang Avenue, Pudong New District Shanghai, 200120 China

Re: K202794

Trade/Device Name: Surgical Face Mask: F-Y1-A Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 26, 2021 Received: November 26, 2021

Dear Sam Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202794

Device Name Surgical face mask: F-Y1-A

Indications for Use (Describe)

The Surgical Face Masks are is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Type of submissionTraditional
Date preparedMay 1, 2021
Submission sponsor
Manufacturer NameJiande Chaomei Daily Chemicals Co., Ltd.
AddressShangshan Village, Yancunqiao Town, Jiande City, ZhejiangProvince, China
Tel86-18858158190
Emailzhuliping@cmmask.com
Contact PersonZhu Lipin
Device identification
Classification NameMask, Surgical
Trade NameSurgical face mask: F-Y1-A
Device ClassificationClass II
Regulation Number21 CFR 878.4040
PanelGeneral Hospital
Product CodeFXX
Previous SubmissionsNone
Application correspondent
Company NameShanghai Spica Management Consulting Co., Ltd.
Address609 Room, No.133 Shengang Avenue, Pudong New District,Shanghai, China
Tel86-15626132181
Emailsam@spicagloble.com
Contact PersonSam Lin

{4}------------------------------------------------

Predicate device information

SponsorWUHAN DYMEX HEALTHCARE CO., LTD.
Trade/Device NameSURGICAL FACE MASK
510(K) numberK182515
Regulation Number21 CFR 878.4040

Indications for use

The Surgical Face Masks are is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device description

The Surgical Face Masks are single use, three-layer, flat -folded masks with ear loops and nose piece. The Surgical Face Masks are manufactured with three layers, the inner and outer layers are made of spunbond polypropylene, and the middle layer is made of melt blown polypropylene filter. The ear loops are held in place over the users' mouth and nose by two elastic ear loops welded to the facemask. The elastic ear loops are not made with natural rubber latex. The nose piece in the layers of facemask is to allow the user to fit the facemask around their nose, which is made of PVC wrapped with metal wire. The surgical face masks will be provided in blue. The surgical face masks are sold non-sterile and are intended to be single use, disposable devices.

Technological Characteristic Comparison

Provided below is a comparison of the subject device with the predicate device.

{5}------------------------------------------------

Table 6A: General Comparison
Proposed DevicePredicate DeviceDifferences
Discussion
Device nameSurgical face mask: F-Y1-ASURGICAL FACE MASKN/A
510(k) numberK202794K182515N/A
ManufacturerJiande Chaomei Daily Chemicals Co., Ltd.Wuhan Dymex Healthcare Co., LtdN/A
Product regulation21 CFR 878.404021 CFR 878.4040Same
Classification nameMask, SurgicalMask, SurgicalSame
Regulation classClass IIClass IISame
Product codeFXXFXXSame
Indications for useThe Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.The Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterileSame
Rx or OTCOTCOTCSame
ModelEar Loops, Flat Pleated, 3 layersEar Loops, Flat Pleated, 3 layersSame
MaterialOuter facing layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Middle layerMelt blown polypropylene filterMelt blown polypropylene filterSame
Inner facing layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Nose piecePVC wrapped with metal wireMalleable polyethylene wireSimilar
Ear loopsLow elastic spandex yarnSpandexSimilar
ColorBlueYellowSimilar
Dimension(Width)17.5cm±0.5cm17.5cm±0.2cmSame
Dimension(Length)9.0cm±0.2cm9.5cm±0.2cmSame
SterilityNon-sterileNon-sterileSame
UseSingle Use, DisposableSingle Use, DisposableSame
ASTM F2100 LevelLevel 2Level 2Same

{6}------------------------------------------------

{7}------------------------------------------------

Summary of Non-Clinical Testing: provided below is a summary of the subject devices to demonstrate that the device meets the specification or acceptance criteria of the standards and test methods shown below.

ItemProposed deviceAcceptance CriteriaResult
Fluid Resistance PerformanceASTM F18623 non-consecutive lots tested29 out of 32 pass at 120 mmHgPass
Particulate Filtration EfficiencyASTM F22993 non-consecutive lots testedLot1: 99.64%Lot2: 99.70%Lot3: 99.63%≥ 98%Pass
Bacterial Filtration EfficiencyASTM F21013 non-consecutive lots testedLot1: 99.6%Lot2: 99.8%Lot3: 99.4%≥ 98%Pass
Differential Pressure (Delta P)MIL-M-36954C3 non-consecutive lots testedLot1: 3.4mmH2O/cm²Lot2: 3.2mmH2O/cm²Lot3: 3.4mmH2O/cm²< 5.0mmH2O/cm²Pass
Flammability 16 CFR 1610Class 1, 3 non-consecutive lots testedClass 1
IrritationISO 10993-10Under the conditions of the study, the device is non-irritatingUnder the conditions of the study, the device is non-irritatingPass
SensitizationISO 10993-10Under the conditions of the study,Under the conditions of the study,Pass
the device is non-sensitizingthe device is non-sensitizing
CytotoxicityISO 10993-5Under the conditions of the study,Under the conditions of the study,Pass
the device is non-cytotoxic.the device is non-cytotoxic.

Table 6B: Comparison of Non-clinical testing

{8}------------------------------------------------

{9}------------------------------------------------

Performance Testing - Clinical

NOT Applicable.

Performance Testing - Animal

NOT Applicable.

Conclusion

The conclusions drawn from the non-clinical test demonstrate that "Surgical face mask: F-Y 1-A (K202794)" is as safe, as effective, and performs as well as the legally marketed predicate devices, "SURGICAL FACE MASK (K182515)".

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.