K Number
K202769

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-01-14

(115 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VELYS™ Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the femoral and tibial implant components intraoperatively, and preparing the bones during total knee arthroplasty.

The VELYS™ Robotic-Assisted Solution is indicated for use with the ATTUNE Total Knee System and its cleared indications for use.

Device Description

The VELYS™ Robotic-Assisted Solution is an image-free robotic-assisted surgery system for Total Knee Arthroplasty (TKA). It is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures and landmarks, planning the position of the femoral and tibial implant components intraoperatively, and preparing the bones during total knee arthroplasty.

The image-free system uses a dedicated optical tracking device to acquire anatomical landmarks intra-operatively. These landmarks are then used to plan the femoral and tibial implant locations based on the surgeon's preferred surgical technique and placement preferences. Following the planning step, the VELYS™ Robotic-Assisted Solution helps the surgeon to execute the bone preparation according to the plan.

The system includes a Robotic-Assisted Device that constrains the position and orientation of the saw handpiece and blade inside each plane corresponding to each resection on the patient's femur and tibia. The surgeon actuates and manipulates the saw handpiece, within the planned resection plane, to execute the bone resection. This is analogous to using manual instruments in TKA and the way the surgeon uses the saw handpiece in the predicate device. If the patient's leg moves during the resection, the Robotic-Assisted Device compensates for such movement in real-time.

The Robotic-Assisted Device is assembled with a Robotic-Assisted Device arm, mounted on the Operating Room (OR) bed rail, for a minimal footprint.

The VELYS™ Robotic-Assisted Solution incorporates several software sub-systems, including applications responsible for general operation of the system and a Clinical Application dedicated to the surgery workflow.

The users interact with the Clinical Application via a touchscreen and footswitch to navigate through the surgery steps. Case Reports including key surgical procedure information are stored on the system and can be retrieved by the surgeon for future use.

AI/ML Overview

The VELYS™ Robotic-Assisted Solution is an image-free robotic-assisted surgery system intended to aid in total knee arthroplasty (TKA). The acceptance criteria and the study proving the device meets these criteria are described below.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance MetricReported Device Performance
System Localization AccuracyConcluded to be the same as the predicate device (TOTAL KNEE SURGETICS Navigation System with iBlock) in an Accuracy and Repeatability study. (Specific numerical values for accuracy are not provided in the document, but the finding of "the same as the predicate" implies it met the predicate's established performance.)
Dynamic Compensation PerformancePerformance test conducted. (Specific numerical results not provided, but implicitly met requirements to support substantial equivalence.)
Saw Blade Position AccuracyAccuracy test bench study conducted to confirm. (Specific numerical results not provided, but implicitly met requirements to support substantial equivalence.)
Boundary Limit Essential PerformancePerformance test conducted. (Specific numerical results not provided, but implicitly met requirements to support substantial equivalence.)
Resection Accuracy (Cadaveric)Confirmed compared to conventional instruments in a Cadaveric Accuracy study. (Specific numerical results not provided, but implicitly met requirements to support substantial equivalence.)
Compliance to StandardsDemonstrated compliance with AAMI ANSI ISO 11137-1:2006 (R) 2015, ISO 11137-2:2013, ISO 11607-1:2019, ISO 11607-2:2019, ISO 10993-1:2018, ISO 10993-1:2009, IEC 60601-1:2005 + A1:2012, AAMI ES60601-1:2005 and A1:2012, IEC 60601-1-2:2014, IEC 80601-2-77:2019, ASTM F2554-18, ASTM F3107-14.
BiocompatibilityTested against ISO 10993-1 (2018 and 2009).
Electrical Safety & EMCTested against IEC 60601-1 (2005+A1:2012) and IEC 60601-1-2 (2014).
SterilizationTested against ISO 11137-1 (2006 (R) 2015) and ISO 11137-2 (2013).
Packaging & TransportTested against ISO 11607-1 (2019) and ISO 11607-2 (2019).
SoftwareVerification tests conducted.
UsabilityUsers (surgeons and Operating Room staff) successfully used the device on simulated knees and cadavers.

2. Sample size used for the test set and the data provenance

The document does not specify exact sample sizes for each test set regarding numerical values. However, it mentions:

  • Simulated Use Testing: Included testing on simulated knees (sawbones) and cadavers. The number of each is not specified.
  • Accuracy and Repeatability Study: Conducted to compare system localization accuracy to the predicate. The number of test instances or samples is not specified.
  • Accuracy Test Bench Study: Conducted to confirm saw blade position accuracy. The number of test instances or samples is not specified.
  • Cadaveric Accuracy Study: Conducted to confirm resection accuracy compared to conventional instruments. The number of cadavers used is not specified.

The data provenance is from non-clinical testing conducted by the manufacturer, DePuy Ireland UC. No specifics are given about the location where cadaveric studies were performed or if they represent diverse populations. The studies are by nature prospective as they are testing the newly developed device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document states that for usability testing, "Users included surgeons and Operating Room staff, who were able to successfully use the VELYS™ Robotic-Assisted Solution to implant ATTUNE Total Knee System per specification." However, it does not specify the number of experts, their qualifications (e.g., years of experience), or if these experts established a ground truth for the technical accuracy studies. For the technical accuracy tests, ground truth appears to be based on predefined targets or established metrology rather than expert judgment.

4. Adjudication method for the test set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for establishing ground truth or evaluating test results, as most of the performance tests are objective measurements against defined standards or established benchmarks rather than subjective expert interpretations.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The document explicitly states: "No clinical tests were conducted to demonstrate substantial equivalence." The comparison was primarily against the predicate device's technological characteristics and by demonstrating the device's technical performance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

The primary performance studies described are focused on the system's technical accuracy and performance, which does involve the robotic-assisted element but in the context of aiding a surgeon for bone preparation. For example, the dynamic-compensation performance, saw blade position accuracy, and boundary limit tests assess the automated and robotic components directly. While a human surgeon provides input and actuates the saw handpiece, these technical accuracy tests evaluate the device's capability to accurately position and maintain the resection plane. Therefore, these can be considered evaluations of the device's standalone technical performance within its intended functional scope.

7. The type of ground truth used

The ground truth for the performance studies appears to be based on:

  • Metrology/Engineering standards: For accuracy measurements (e.g., system localization, saw blade position, dynamic compensation, boundary limits), the ground truth would be defined by precise measurement tools and known physical targets.
  • Established specifications: For compliance with international standards (e.g., ISO, IEC, ASTM), the ground truth is the requirement set forth by these standards.
  • Conventional instruments/techniques: For the cadaveric accuracy study, the ground truth for resection accuracy was "compared to conventional instruments," implying that the accuracy achieved by established manual techniques served as a benchmark (either for equivalence or non-inferiority).

8. The sample size for the training set

The document describes non-clinical and simulated use testing. It does not mention a training set as this is a robotic-assisted surgical device, not an AI/ML diagnostic algorithm that requires training data in the traditional sense. The device's "training" would be its design, calibration, and verification against engineering specifications.

9. How the ground truth for the training set was established

Not applicable, as there is no mention of a training set for an AI/ML algorithm. The device's operational parameters and accuracy are based on engineering design and validation processes.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

January 14, 2021

DePuy Ireland UC Stéphanie Elvin Regulatory Affairs Manager Loughbeg, Ringaskiddy Ireland

Re: K202769

Trade/Device Name: VELYSTM Robotic-Assisted Solution Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: December 18, 2020 Received: December 22, 2020

Dear Stephanie Elvin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K202769

Device Name VELYSTM Robotic-Assisted Solution

Indications for Use (Describe)

The VELYS™ Robotic-Assisted Solution is intended for stereotaxic surgery to aid the surgeon in identifying the relative position and orientation of anatomical structures, planning the femoral and tibial implant components intraoperatively, and preparing the bones during total knee arthroplasty.

The VELYS™ Robotic-Assisted Solution is indicated for use with the ATTUNE Total Knee System and its cleared indications for use.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

(As required by 21 CFR 807.92)

Submitter Information
NameDePuy Ireland UC
AddressLoughbeg, RingaskiddyCo. Cork Munster, IRELAND
Phone number574-372-7211
Fax number574- 371-4987
Establishment RegistrationNumber3015516266
Name of contact personStéphanie Elvin selvin@jnj.its.com, Karen Mahoney kmahoney@its.jnj.com
Date prepared18 December 2020
Name of device
Trade or proprietary nameVELYSTM Robotic-Assisted Solution
Common or usual nameSurgical Navigation System
Classification nameOrthopedic Stereotaxic Instrument
ClassII
Classification panel87 Orthopedics
Regulation882.4560
Product Code(s)OLO
Legally marketed device(s) towhich equivalence is claimedTOTAL KNEE SURGETICS Navigation System with iBlock (K090953)
Reason for 510(k) submissionNew Device
Device descriptionThe VELYSTM Robotic-Assisted Solution is an image-free robotic-assistedsurgery system for Total Knee Arthroplasty (TKA). It is intended forstereotaxic surgery to aid the surgeon in identifying the relative position andorientation of anatomical structures and landmarks, planning the position ofthe femoral and tibial implant components intraoperatively, and preparing thebones during total knee arthroplasty.The image-free system uses a dedicated optical tracking device to acquireanatomical landmarks intra-operatively. These landmarks are then used to planthe femoral and tibial implant locations based on the surgeon's preferredsurgical technique and placement preferences. Following the planning step,the VELYSTM Robotic-Assisted Solution helps the surgeon to execute the bonepreparation according to the plan.
Indications for useThe system includes a Robotic-Assisted Device that constrains the positionand orientation of the saw handpiece and blade inside each planecorresponding to each resection on the patient's femur and tibia. The surgeonactuates and manipulates the saw handpiece, within the planned resectionplane, to execute the bone resection. This is analogous to using manualinstruments in TKA and the way the surgeon uses the saw handpiece in thepredicate device. If the patient's leg moves during the resection, the Robotic-Assisted Device compensates for such movement in real-time.The Robotic-Assisted Device is assembled with a Robotic-Assisted Devicearm, mounted on the Operating Room (OR) bed rail, for a minimal footprint.The VELYSTM Robotic-Assisted Solution incorporates several software sub-systems, including applications responsible for general operation of the systemand a Clinical Application dedicated to the surgery workflow.The users interact with the Clinical Application via a touchscreen andfootswitch to navigate through the surgery steps. Case Reports including keysurgical procedure information are stored on the system and can be retrievedby the surgeon for future use.The VELYSTM Robotic-Assisted Solution is intended for stereotaxic surgery toaid the surgeon in identifying the relative position and orientation of anatomicalstructures, planning the position of the femoral and tibial implant componentsintraoperatively, and preparing the bones during total knee arthroplasty.The VELYSTM Robotic-Assisted Solution is indicated for use with theATTUNE Total Knee System and its cleared indications for use.

{4}------------------------------------------------

{5}------------------------------------------------

Summary of the Technological Characteristics of the Device Compared to the Predicate Device
CharacteristicsSubject Device:VELYSTM Robotic-Assisted SolutionDePuy Ireland UCPredicate Device:Total Knee Surgetics Navigation Systemwith iBlock, OMNILife Science Inc.K090953 Cleared Jan 21, 2010
Indications forUseThe VELYSTM Robotic-Assisted Solutionis intended for stereotaxic surgery to aid thesurgeon in identifying the relative positionand orientation of anatomical structures,planning the position of the femoral andtibial implant components intraoperatively,and preparing the bones during total kneearthroplasty.The VELYSTM Robotic-Assisted Solutionis indicated for use with the ATTUNETMTotal Knee System and its clearedindications for use.The Total Knee Surgetics NavigationSystem with iBlock is intended for useduring stereotaxic surgery to aid thesurgeon in locating anatomical structuresand aligning endoprostheses with theanatomical structures.The Total Knee Surgetics NavigationSystem with iBlock is specificallyindicated for Total Knee Arthroplasty
Principles ofoperationUses established computer-assisted orthopaedic surgery technologies to prepare the bone Optical localization technology is used to collect intraoperative data and track relative position of the bones and instruments during surgery A motorized instrument controlled by the system automatically positions the resection plane (saw blade) within the planned plane for the bone resection and maintains alignment through dynamic compensation. Image-free technologyUses established computer-assisted orthopaedic surgery technologies to prepare the bone Optical localization technology is used to collect intraoperative data and track relative position of the bones and instruments during surgery A motorized instrument controlled by the system automatically positions the resection plane (cutting guide) to the planned plane for the bone resection. Image-free technology
StationsBase station: camera, power supply, Robot Control Unit, display monitor, foot pedal Satellite station: optional touchscreen, storage for Robotic-Assisted Device and Robotic-Assisted Device arm, transfer mechanism to assist with transfer to bedrailOne station: camera, power supply unit, control box for robot, computer unit, display monitor, foot pedal
SoftwareDedicated TKA application softwareDedicated TKA application software

{6}------------------------------------------------

Discussion of Similarities and Differences Between the Device and the Predicate Device

The VELYS™ Robotic-Assisted Solution (Subject device) and the TOTAL KNEE SURGETICS Navigation System with iBlock (Predicate device) have the same intended use and are both indicated for TKA surgery.

Both devices are image-free and use Stereotaxic / Computer Assisted Surgery (CAS) technology to localize bone structures and track their position relative to the instruments during surgery.

The systems use very similar setups with one or two stations holding a camera, computer units, power supply, and display monitors; dedicated instruments sets, containing in arrays with reflective markers; and instruments such as drill pins and clamps for attaching arrays to the bones. The VELYS™ Robotic-Assisted Solution includes a dedicated cabled Saw Handpiece.

Operation of the systems is very similar; the surgeon registers bone anatomy using a pointer, the system software provides a visual bone model and allows the surgeon to plan resections. Based on the surgical plan, the system positions either a cutting guide (in the predicate) or the Saw Handpiece (for the subject device) in the correct position to allow the surgeon to execute the planned resection.

The surgical flows are very similar, and the surgeon can choose the order of cuts to perform via the Graphic User Interface (GUI). Surgical techniques provided with the proposed and predicate devices are specific to the implants they are indicated for use with.

The devices differ slightly in the way they provide a resection plane. In the predicate device, a cutting guide is positioned by a motorized instrument and the surgeon places the sawblade through the slot in the guide to perform the resection. In the subject device, a motorized instrument holds the Saw Handpiece in the correct plane for the surgeon to execute the primary resections; there is no physical cutting guide, the cutting guide is effectively virtualized.

Software in the subject device establishes the desired resection plane, tracks the positioning of Arrays in real-time and adjusts the position of the Saw Blade to compensate for any movement of the patient leg. In the predicate device, the cutting guide is mounted on the patient's femur to ensure the cutting guide position is maintained relative to the bone.

Performance Data

Summary of Non-Clinical Tests Conducted for Determination of Substantial Equivalence

Design verification tests were performed on the VELYS™ Robotic-Assisted Solution in accordance with the product risk analysis and product requirements and to demonstrate substantial equivalence to the predicate device. Testing performed for assessing device performance include Performance Testing, Electrical Safety, Electromagnetic Compatibility, Biocompatibility, Sterilization, Packaging, Transport, Software and Usability. Simulated use testing included testing on simulated knees (sawbones) and cadavers. Users included surgeons and Operating Room staff, who were able to successfully use the VELYS™ Robotic-Assisted Solution to implant ATTUNE Total Knee System per specification.

{7}------------------------------------------------

Testing was performed that demonstrated compliance to the following standards:

AAMI ANSI ISO 11137-1:2006 (R) 2015 Sterilization of healthcare products - Radiation Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices

ISO 11137-2:2013 Sterilization of health care products - Radiation Part 2: Establishing the sterilization dose

ISO 11607-1:2019 Packaging for Terminally Sterilized Medical Devices – Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems

ISO 11607-2:2019 Packaging for Terminally Sterilized Medical Devices - Part 2: Validation Requirements for Forming. Sealing and Assembly Processes.

ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

IEC 60601-1:2005 + A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

AAMI ES60601-1:2005 and A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

IEC 60601-1-2:2014 Medical electrical equipment - Part 2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

IEC 80601-2-77:2019 Particular requirements for the BASIC SAFETY and essential performance of ROBOTICALLY ASSISTED SURGICAL EQUIPMENT

ASTM F2554-18 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems

ASTM F3107-14 Standard Test Method for Measuring Accuracy after Mechanical Disturbances on Reference Frames of Computer Assisted Surgery Systems

Subject and predicate devices were tested to the same international standards for key requirements such as Electrical Safety and EMC, as well as Biocompatibility, supporting substantial equivalence.

In addition to standard-driven testing, the following test was performed to compare the system navigation accuracy and support substantial equivalence to the predicate:

  • Accuracy and repeatability study, which concluded the system localization accuracy of the subject is . the same as the predicate.

{8}------------------------------------------------

The following tests were performed to support Safety and Efficacy of the proposed device as they pertain to features specific to the proposed device:

  • Dynamic-compensation performance test. ●
  • Accuracy test bench study to confirm the accuracy of the saw blade position. .
  • . Boundary Limit Essential Performance test.
  • . Cadaveric Accuracy study to confirm resection accuracy compared to conventional instruments.
  • . Design Validation study.

Non-Clinical testing supports Substantial Equivalence of the subject device VEL YS™ Robotic-Assisted Solution to the predicate device TOTAL KNEE SURGETICS Navigation System with iBlock

Summary of Clinical Tests Conducted for Determination of Substantial Equivalence and/or of Clinical Information

No clinical tests were conducted to demonstrate substantial equivalence.

Conclusions Drawn from Non-Clinical and Clinical Data

The subject VELYS™ Robotic-Assisted Solution is substantially equivalent to the predicate TOTAL KNEE SURGETICS Navigation System with iBlock.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).