(60 days)
The Ortho Solutions Plating System is indicated for use in stabilization of fractures or osteotomies, revision procedures, joint fusion, and reconstruction of small bones of the feet and ankles, including the toes, talus, and calcaneus; the distal fibula and tibia; as well as the fingers, hands, and wrists.
In addition, the non-locking screws and washers are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
The Ortho Solutions Plating System is a set of implantable metallic (Titanium Alloy) bone plates, locking and non-locking screws of various lengths. The system consists of a series of tubular plates of varying lengths as well as anatomically contoured plates. Locking and non-locking screws are provided in Ø2.7mm, Ø3.5mm, & Ø4.0mm in various lengths and are compatible with all plates. All implants are made from Titanium Alloy (Ti-6AL-4V ELI) per ASTM F136 and are provided non-sterile.
This document is a 510(k) Premarket Notification from the FDA regarding the "Ortho Solutions Plating System." As such, it does not contain the information requested about acceptance criteria for an AI/device performance study. The document is about a bone fixation device, not an AI or diagnostic tool.
The requested information:
- A table of acceptance criteria and the reported device performance
- Sample sized used for the test set and the data provenance
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Adjudication method for the test set
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- The type of ground truth used
- The sample size for the training set
- How the ground truth for the training set was established
None of this information is available in the provided text because the device is a physical bone plating system, not a software or AI-driven diagnostic device.
The "Performance Analysis" section on page 3 simply states:
"Performance testing per ASTM F382, engineering analysis and rationale was completed for the plates to demonstrate substantial equivalence for static and fatigue bending strength. Performance testing was completed for the screws per ASTM F543 to demonstrate substantial equivalence with respect to torsional strength, insertion torque, and pullout strength."
This refers to mechanical engineering tests for physical product performance, not clinical or algorithmic performance, and it does not describe study methodologies relevant to AI/software.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2020
Ortho Solutions UK Ltd Andy Leither US Technical and Development Director West Station Business Park Spital Road Essex, CM9 6FF United Kingdom
Re: K202764
Trade/Device Name: Ortho Solutions Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC, HTN Dated: September 16, 2020 Received: September 21, 2020
Dear Andy Leither:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K202764
Device Name Ortho Solutions Plating System
Indications for Use (Describe)
The Ortho Solutions Plating System is indicated for use in stabilization of fractures or osteotomies, revision procedures, joint fusion, and reconstruction of small bones of the feet and ankles, including the toes, talus, and calcaneus; the distal fibula and tibia; as well as the fingers, hands, and wrists.
In addition, the non-locking screws and washers are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for OrthoSolutions Group. The words "OrthoSolutions" are in red, with the word "Group" in gray underneath. To the right of the text is a stylized "S" symbol, which is also in red and gray.
K202764 510(k) SUMMARY
Submitter Information
| Submitters Name:Address: | Ortho Solutions UK LtdWest Station Business ParkSpital RoadEssex, CM9 6FFUnited Kingdom |
|---|---|
| Telephone Number:Fax Number:Prepared By:Contact Person:Date Prepared:Device Information | +44 1621 843 599+44 1621 858 953Andy LeitherAndy Leither09/16/2020 |
| Trade Name: | Ortho Solutions Plating System |
| Common Name: | Plate, Fixation, BoneScrew, Fixation, BoneWasher, Bolt Nut |
| Device Classification: | Class II per 21 CFR 888.3030 (Primary)Single/multiple component metallic bone fixation appliances andaccessories.Panel: Orthopedic |
| Class II per 21 CFR 888.3040Smooth or threaded metallic bone fixation fastener.Panel: Orthopedic | |
| Class II per 21 CFR 888.3030Single/multiple component metallic bone fixation appliances andaccessories.Panel: Orthopedic | |
| Product Code: | HRS (Primary)HWCHTN |
| Material Composition: | Titanium Alloy (Ti-6AL-4V ELI) per ASTM F136 |
| Device Description: | The Ortho Solutions Plating System is a set of implantable metallic(Titanium Alloy) bone plates, locking and non-locking screws of variouslengths. The system consists of a series of tubular plates of varyinglengths as well as anatomically contoured plates. Locking and non-locking screws are provided in Ø2.7mm, Ø3.5mm, & Ø4.0mm in variouslengths and are compatible with all plates. All implants are made fromTitanium Alloy (Ti-6AL-4V ELI) per ASTM F136 and are provided non-sterile. |
| Primary Predicate Device: | Ortho Solutions Trauma Plates for Osteosynthesis, K120360 |
| Additional Predicate Devices: | Ortho Solutions UltOS Plating System, K141784 |
| Wright Medical ORTHOLOC 3Di Ankle Fracture Plating System, K163044 | |
| Indications for Use: | The Ortho Solutions Plating System is indicated for use in stabilization andfixation of fractures or osteotomies, revision procedures, joint fusion, andreconstruction of small bones of the feet and ankles, including the toes,talus, and calcaneus; the distal fibula and tibia; as well as the fingers,hands, and wrists. |
| In addition, the non-locking screws and washers are indicated for use inbone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair,and fracture fixation, appropriate for the size of the device. | |
| Technological Characteristics: | The Ortho Solutions Plating System is substantially equivalent inmaterial, geometry, design, and indications to the predicate devices. |
| Performance Analysis: | Performance testing per ASTM F382, engineering analysis and rationalewas completed for the plates to demonstrate substantial equivalencefor static and fatigue bending strength. Performance testing wascompleted for the screws per ASTM F543 to demonstrate substantialequivalence with respect to torsional strength, insertion torque, andpullout strength. |
| Conclusion: | The Ortho Solutions Plating System has been compared to thepreviously cleared predicate devices with respect to indications for use,materials, mechanical performance, and technological characteristics.The design characteristics of the subject device do not raise new typesof questions regarding safety or efficacy. These comparisonsdemonstrate that the Ortho Solutions Plating System is substantiallyequivalence to the predicate devices. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OrthoSolutions Group. The words "OrthoSolutions" are in red, with the word "Group" in gray underneath. To the right of the text is a stylized silver and red symbol that appears to be an abstract letter S.
N/A