K Number
K202755

Validate with FDA (Live)

Date Cleared
2022-11-05

(775 days)

Product Code
Regulation Number
866.3181
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DiaSorin Molecular Simplexa™ Congenital CMV Direct is a real-time PCR assay intended for use on the LIAISON® MDX instrument for the in vitro qualitative detection of cytomegalovirus (CMV) from saliva swabs and urine from infants less than 21 days of age. Positive results from saliva are presumptive and should be confirmed with urine. The results of the Simplexa™ Congenital CMV Direct assay should be used in conjunction with the results of other clinical findings as an aid in the diagnosis of congenital CMV infection.

This test has not been cleared for screening of blood products for the presence of CMV or for use with samples other than urine and saliva swabs.

DiaSorin Molecular's Simplexa™ Congenital CMV Positive Control Pack is intended to be used as a control with the Simplexa Congenital CMV Direct kit for use on the LIAISON MDX instrument. This control is not intended for use with other assays or systems.

Device Description

The Simplexa™ Congenital CMV Direct assay is a real-time PCR system that enables the direct amplification and detection of CMV DNA from either saliva swab or urine specimens without nucleic acid extraction. The system consists of the Simplexa™ Congenital CMV Direct Reaction Mix, the LIAISON® MDX (with LIAISON® MDX Studio Software), the Direct Amplification Disc (DAD) and associated accessories.

In the Simplexa™ Congenital CMV Direct assay, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify CMV DNA. A well-conserved region of the CMV UL83 gene is targeted to identify CMV DNA. An internal control is used to detect PCR failure and/or inhibition.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state quantitative acceptance criteria in a dedicated table format. However, the reported performance metrics imply the criteria for acceptance. For the purpose of this response, I'll infer the implicit acceptance criteria based on the demonstrated performance, generally implying "high agreement" for positive and negative cases.

Metric (Implicit Acceptance Criteria)Saliva Swab Performance (Retrospective)Urine Performance (Retrospective)Saliva Swab Performance (Prospective)Urine Performance (Prospective)
Positive Percent Agreement (PPA)100.0% (95% CI: 93% - 100%)100.0% (95% CI: 93% - 100%)94.1% (95% CI: 73% - 99%)95.3% (95% CI: 85% - 99%)
Negative Percent Agreement (NPA)100.0% (95% CI: 97% - 100%)98.4% (95% CI: 94% - 100%)99.9% (95% CI: 100% - 100%)100.0% (95% CI: 100% - 100%)
Reproducibility (%CV)Between 0.5% and 1.6%Between 0.7% and 1.5%N/A (not directly from this study)N/A (not directly from this study)
Analytical Sensitivity (LoD)500 Copies/mL (AD-169, Towne, Merlin) in UTM400 Copies/mL (AD-169), 800 Copies/mL (Towne), 6400 IU/mL (Merlin)N/AN/A
Cross-Reactivity100% agreement (no cross-reactivity)100% agreement (no cross-reactivity)N/AN/A
Interference100% agreement (no interference)100% agreement (no interference)N/AN/A
Microbial Inhibition100% agreement (no inhibition)100% agreement (no inhibition)N/AN/A

2. Sample Size and Data Provenance

  • Retrospective Study:

    • Saliva Swab: 173 total specimens (3 removed due to indeterminate CRM result, 170 analyzed)
    • Urine: 173 total specimens
    • Provenance: "collected during the clinical study", "stored at a central site", then distributed to three (3) laboratories. The document doesn't explicitly state the country of origin for these retrospective samples, though the testing sites were in the USA. These were pre-selected positive and negative samples based on routine laboratory results.
  • Prospective Study:

    • Saliva Swab: 1,859 initially collected, 6 deemed ineligible, resulting in 1,853 analyzed specimens.
    • Urine: 1,656 initially collected, 32 deemed ineligible, resulting in 1,624 analyzed specimens.
    • Provenance: Prospectively collected (frozen and/or fresh) from ten (10) collection sites across the USA and two (2) collection sites outside the USA. Testing was performed at six (6) testing sites located in the USA.

3. Number of Experts and Qualifications

The document states that a "Composite Reference Method (CRM)" was used, which involved "two (2) validated PCR followed by bi-directional sequencing assays." One (1) central laboratory performed these comparator assays.

  • The document does not specify the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with 10 years of experience). It relies on the validation of the PCR and bi-directional sequencing assays as the basis for ground truth, implying that these are established and reliable laboratory methods.

4. Adjudication Method

The ground truth for the clinical agreement studies (both retrospective and prospective) was established via a "Composite Reference Method (CRM)".
This CRM "utilized two (2) validated PCR followed by bi-directional sequencing assays. A sample had a final sequencing result of 'Detected' if one or both sequencing results were 'Detected'. Conversely a sample had a final sequencing result of 'Not Detected' if both results were 'Not Detected'." This implies a form of 2+0 or 1+1 adjudication model where if either reference method detects CMV, the sample is considered positive, and both must be negative for the sample to be considered negative.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This study focuses on the diagnostic performance of the device itself against a laboratory-based reference method, not on how human readers/clinicians improve with AI assistance.

6. Standalone (Algorithm Only) Performance

Yes, a standalone performance evaluation was done. The Simplexa™ Congenital CMV Direct assay is a real-time PCR assay and its performance was evaluated directly against the Composite Reference Method (CRM) without human-in-the-loop assistance. The reported PPA and NPA values represent the algorithm's standalone performance.

7. Type of Ground Truth Used

The ground truth used was expert consensus on laboratory results, specifically based on a "Composite Reference Method (CRM) utilized two (2) validated PCR followed by bi-directional sequencing assays." This is a highly robust and objective form of ground truth for nucleic acid detection devices, often considered a gold standard in molecular diagnostics.

8. Sample Size for the Training Set

The document does not provide information on the sample size used for the training set. This is typical for submissions of this nature, where the focus is on the validation of the final device/algorithm using a separate, independent test set, rather than details of the developmental (training) phase.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for the training set was established, as details about the training phase are not included in this summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 5, 2022

DiaSorin Molecular LLC Sharon Young Principal Regulatory Affairs Specialist 11331 Valley View Street Cypress, California 90630

Re: K202755

Trade/Device Name: Simplexa Congenital CMV Direct and Simplexa Congenital CMV Positive Control Pack Regulation Number: 21 CFR 866.3181 Regulation Name: Cytomegalovirus Nucleic Acid Detection Device For Congenital Cytomegalovirus Infection Regulatory Class: Class II Product Code: ODZ Dated: September 17, 2020 Received: September 21, 2020

Dear Sharon Young:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Maria I. Garcia -S

Maria Garcia, Ph.D. Assistant Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202755

Device Name

Simplexa™ Congenital CMV Direct and Simplexa™ Congenital CMV Positive Control Pack

Indications for Use (Describe)

Simplexa™ Congenital CMV Direct Catalog Number MOL2250

The DiaSorin Molecular Simplexa™ Congenital CMV Direct is a real-time PCR assay intended for use on the LIAISON® MDX instrument for the in vitro qualitative detection of cytomegalovirus (CMV) from saliva swabs and urine from infants less than 21 days of age. Positive results from saliva are presumptive and should be confirmed with urine. The results of the Simplexa™ Congenital CMV Direct assay should be used in conjunction with the results of other clinical findings as an aid in the diagnosis of congenital CMV infection.

This test has not been cleared for screening of blood products for the presence of CMV or for use with samples other than urine and saliva swabs

Simplexa™ Congenital CMV Positive Control Pack Catalog Number MOL2260 DiaSorin Molecular's Simplexa™ Congental CMV Positive Control Pack is intended to be used as a control with the Simplexa Congenital CMV Direct kit for use on the LIAISON MDX instrument. This control is not intended for use with other assays or systems.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, with the word "Molecular" underneath in a lighter green color.

510(k) Summarv

Simplexa™ Congenital CMV Direct MOL2250 Simplexa ™ Congenital CMV Positive Control Pack. MOL2260 Nov 2. 2022

Page 1 of 17

ApplicantDiaSorin Molecular LLC.11331 Valley View StreetCypress, California 90630USA
EstablishmentRegistration No.2023365
Contact PersonTara Viviani, RACSr. Directory Molecular Requlatory AffairsTel. 562.240.6680Tara. Viviani@DiaSorin.com
Summary DateNovember 2, 2022
Proprietary NameSimplexa™ Congenital CMV Direct andSimplexa™ Congenital CMV Positive Control Pack
US ProductCodes/Names andRegulation NumbersQDZ - Cytomegalovirus nucleic acid detection device for congenitalcytomegalovirus infection 21 CFR § 866.3281OOI - Instrumentation for clinical multiplex test systems 21 CFR § 862.2570
ClassificationClass II
Predicate DevicesAlethia® CMV Assay Test System DEN180040 (for saliva swab). Currently,there is not a predicate device for urine as a compatible specimen type.

Intended Use

Simplexa™ Congenital CMV Direct REF MOL2255

The DiaSorin Molecular Simplexa™ Congenital CMV Direct is a real-time PCR assay intended for use on the LIAISON® MDX instrument for the in vitro qualitative detection of cytomegalovirus (CMV) from saliva swabs and urine from infants less than 21 days of age. Positive results from saliva are presumptive and should be confirmed with urine. The Simplexa™ Congenital CMV Direct assay should be used in conjunction with the results of other clinical findings as an aid in the diagnosis of congenital CMV infection.

This test has not been cleared for screening of blood or blood products for the presence of CMV or for use with samples other than urine and saliva swabs.

Simplexa™ Congenital CMVPositive Control Pack REF MOL2265

DiaSorin Molecular's Simplexa™ Congenital CMV Positive Control Pack is intended to be used as a control with the Simplexa™ Congenital CMV Direct kit for use on the LIAISON® MDX instrument. This control is not intended for use with other assays or systems.

Device Description

The Simplexa™ Congenital CMV Direct assay is a real-time PCR system that enables the direct amplification and detection of CMV DNA from either saliva swab or urine specimens without nucleic acid extraction. The system consists of the Simplexa™ Congenital CMV Direct Reaction Mix, the LIAISON® MDX (with LIAISON® MDX Studio Software), the Direct Amplification Disc (DAD) and associated accessories.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, stacked above the word "Molecular" in a lighter green, sans-serif font.

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022 Page 2 of 17

In the Simplexa™ Congenital CMV Direct assay, bi-functional fluorescent probe-primers are used together with corresponding reverse primers to amplify CMV DNA. A well-conserved region of the CMV UL83 gene is targeted to identify CMV DNA. An internal control is used to detect PCR failure and/or inhibition.

Simplexa™ Congenital CMV Direct REF MOL2255

Component NameREFEC Symbol on LabelAbbreviated NameCap ColorNumber of VialsReactions per Vial/KitVolume per Vial
Simplexa™ Congenital CMV Direct Reaction MixMOL2256REAGCRMWhite241/2450 µL

Simplexa™ Congenital CMV Direct Components and Descriptions

Kit ComponentContents
DNA polymerase, buffer, dNTPs, template DNA (Internal Control), dye-labeled fluorescent probesand primers specific for detection of CMV and for the DNA Internal Control.
Simplexa™Congenital CMVDirect ReactionMix (RM)TargetProbeFluorophore(Dye)Excitation (nm)Emission (nm)Targeted Gene
CMVFAM495520UL83 gene
InternalControl RNA(IC)Q670644670N/A
Simplexa™Congenital CMVDirect BarcodeCardAssay specific parameters and lot information

Simplexa™ Congenital CMV Positive Control Pack REF MOL2265 Component and Description

Component NameREFDescriptionCapColorNumberof VialReactionsperVial/KitsVolumeper Vial
Simplexa™Congenital CMVDirect PositiveControlMOL2256Inactivated CMVRed101/1050µL

Materials Supplied Separately

Direct Amplification Disc Kit (REF MOL1455) Direct Amplification Discs for use on the LIAISON® MDX

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a green and blue DNA helix on the left, with the words "DiaSorin" in blue on the top right and the word "Molecular" in green on the bottom right. The logo is simple and modern, and it effectively communicates the company's focus on molecular diagnostics.

510(K) Summary Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022 Page 3 of 17

Comparison to Predicate Device

Comparison toPredicate DevicePredicate Device:Alethia CMV Assay Test System(DEN180040)Candidate Device:Simplexa™ Congenital CMV Direct andSimplexa™ Congenital CMV PositiveControl Pack
Product CodeQDZSame
RegulationNumber andDescription21 CFR § 866.3381 – Cytomegalovirusnucleic acid detection device for congenitalcytomegalovirus infection.Same
OrganismDetectedcytomegalovirusSame
MeasurandgeneA well-conserved region of the CMV UL83
Intended Use KitThe Alethia CMV Assay Test Systemincludes separately provided test kits for theAlethia CMV DNA Amplification Assay andthe Alethia CMV External Control Reagents.The Alethia CMV DNA Amplification Assay,performed on the Alethia instrument, is aqualitative, in vitro diagnostic test system forthe direct detection of Cytomegalovirus(CMV) DNA in saliva samples from neonatesyounger than 21 days of age. The test is usedas an aid in the diagnosis of congenital CMVinfection. The results of this test should beused in conjunction with the results of otherclinical findings.Flocked swabs should be used to collectsaliva from neonates. The swab can becollected dry, without viral transport media(VTM), or placed in no more than 1 mL VTM.The Alethia CMV External Control Reagentsare used as part of a routine quality controlprogram to aid the user in detection ofunexpected conditions that may lead to testerrors. The external controls are intended foruse with the Alethia CMV DNA AmplificationAssay; the controls are not intended for usewith other assays or systems.The DiaSorin Molecular Simplexa™Congenital CMV Direct is a real-time PCRintended for use on the LIAISON MDXinstrument for the in vitro qualitative detectionof cytomegalovirus (CMV) from saliva swabsand urine from infants less than 21 days ofage. The Simplexa™ Congenital CMV Directis an aid in the diagnosis of congenital CMVinfection.
Intended UseControl PackThe Alethia CMV Assay Test Systemincludes separately provided test kits for theAlethia CMV DNA Amplification Assay andthe Alethia CMV External Control Reagents.The Alethia CMV External Control Reagentsare used as part of a routine quality controlprogram to aid the user in detection ofunexpected conditions that may lead to testerrors. The external controls are intended foruse with the Alethia CMV DNA AmplificationAssay; the controls are not intended for usewith other assays or systems.The Simplexa™ Congenital CMV PositiveControl Pack is intended to be used as acontrol with the Simplexa™ Congenital CMVDirect kit. This control is not intended for usewith other assays or systems.
Comparison toPredicate DevicePredicate Device:Alethia CMV Assay Test System(DEN180040)Candidate Device:Simplexa™ Congenital CMV Direct andSimplexa™ Congenital CMV PositiveControl Pack
AutomatedSystem (Sampleto Answer)YesSame
InstrumentationAlethia™ Instrument; Meridian Bioscience,Inc.LIAISON® MDX
SampleTypes/MediaTypeDry flocked saliva swab or saliva swab in 1 mLVTM from infants < 21 days old.Saliva swab in BD Universal Viral Transport(UVT), Copan UTM® (1mL or 3mL), RemelM4RT®, M4®, and M6® transport and urinefrom infants less than 21 days old.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, stacked above the word "Molecular" in a lighter green, sans-serif font.

Simplexa™ Congenital CMV Direct Simplexa™ Congenital CMV Positive Control Pack. M Nov 2. 2022

Page 4 of 17

CLINICAL AGREEMENT

The performance of the Simplexa™ Congenital CMV Direct assay was established in a clinical study that included two (2) cohorts based on sample status. Specifically, prospective (pre-selected positive and negative samples based on routine laboratory results) samples from infants less than twentyone (21) days of age, were tested in the clinical agreement study.

Retrospective Study

A total of 346 retrospective specimens were collected during the clinical study. Specimens were enrolled, aliquoted and shipped to a central site where they were distributed for Simplexa™ Congenital CMV Direct testing at three (3) laboratories. One (1) central laboratory performed both comparator PCR/bi-directional sequencing assays for the two (2) part Composite Reference Method (CRM). The Composite Reference Method (CRM) utilized two (2) validated PCR followed by bi-directional sequencing assays. A sample had a final sequencing result of 'Detected' if one or both sequencing results were 'Detected'. Conversely a sample had a final sequencing result of 'Not Detected' if both results were 'Not Detected'.

Of the 346 specimens tested, 170 results were generated for saliva swab specimens and 173 results were generated for urine specimens. The results are presented in Tables 3a and 3b.

The saliva swab study included 173 specimens in the analysis. Three (3) specimens were removed from analysis. The three (3) samples yielded an indeterminate final result with the Composite Reference Method (CRM) CMV-2 PCR/Bi-directional sequencing assay preventing the algorithm from producing a result for use as a comparator.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in shades of green and blue on the left. To the right of the DNA graphic are the words "DiaSorin" in dark blue, with the word "Molecular" underneath in green.

510(k) Summarv

Simplexa™ Congenital CMV Direct MOL2250 Simplexa ™ Congenital CMV Positive Control Pack. MOL2260 Nov 2. 2022

Page 5 of 17

Table 3a. Simplexa™ Congenital CMV Direct Clinical Agreement - Saliva Swab (Retrospective)

Clinical AgreementPPANPA
Simplexa™ CongenitalCMV Direct ResultsComposite Reference Method (CRM)
DetectedNot DetectedTotal
Detected53053100.0%(53/53)95% CI:93% - 100%100.0%(117/117)95% CI: 97% - 100%
Not Detected0117117
Total53117170

PPA = Positive Percent Agreement, NPA = Negative Percent. 95% Cl = 95% Confidence Interval

The 95% confidence intervals (CI) were calculated following the Wilson Score method

Table 3b. Simplexa™ Congenital CMV Direct Clinical Agreement - Urine (Retrospective)

Clinical AgreementPPANPA
Simplexa™ CongenitalCMV Direct ResultsDetectedNot DetectedTotal
Detected492a51100.0%(49/49)95% CI:93% - 100%98.4%(122/124)95% CI: 94%- 100%
Not Detected0122122
Total49124173

a Two (2) urine samples were positive by routine methodology.

PPA = Positive Percent Agreement, NPA = Negative Percent Agreement. 95% Cl = 95% Confidence Interval

The 95% confidence intervals (CI) were calculated following the Wilson Score method.

Prospective Study

A total of one thousand eight hundred fifty-nine (1,859) saliva swab specimens and/or one thousand six hundred fifty-six (1,656) urine specimens were prospectively collected as frozen and/or fresh specimens. Of these collected specimens, six (6) saliva swab and thirty-two (32) urine specimens were deemed ineligible and removed from analysis. Specimens were collected from ten (10) collection sites across the USA and two (2) collection sites outside the USA. Testing was performed at six (6) testing sites located in the USA. One (1) central laboratory performed both comparator PCR/bi-directional sequencing assays for the two (2) part Composite Reference Method (CRM).

The Composite Reference Method (CRM) utilized two (2) validated PCR followed by bi-directional sequencing assays. A sample had a final sequencing result of 'Detected' if one or both sequencing results were 'Detected'. Conversely a sample had a final sequencing result of 'Not Detected' if both results were 'Not Detected'.

Prospective clinical agreement was based on a total of one thousand eight hundred fifty-three (1,853) saliva swab specimens and one thousand six hundred twenty-four (1,624) urine specimens. The results are presented in Tables 4a and 4b.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for DiaSorin Molecular. The word "DiaSorin" is in large, bold, blue font on the top line. The word "Molecular" is in a smaller, green font on the second line. To the left of the words is a green and blue image that looks like a strand of DNA.

510(k) Summarv

Simplexa™ Congenital CMV Direct MOL2250 Simplexa ™ Congenital CMV Positive Control Pack. MOL2260

Nov 2. 2022

Table 4a. Simplexa™ Congenital CMV Direct Clinical Agreement - Saliva Swab (Prospective)

Clinical AgreementPPANPA
Simplexa™ CongenitalCMV Direct ResultsDetectedNot DetectedTotal
Detected1611794.1%(16/17)95% CI:73% - 99%99.9%(1835/1836)95% CI:100% -100%
Not Detected1a18351836
Total1718361853

ª One (1) saliva swab specimen was negative by routine methodology.

PPA = Positive Percent Agreement, NPA = Negative Percent. 95% Cl = 95% Confidence Interval

The 95% confidence intervals (CI) were calculated following the Wilson Score method.

Clinical AgreementPPANPA
SimplexaTM CongenitalCMV Direct ResultsComposite Reference Method (CRM)
DetectedNot DetectedTotal
Detected4104195.3%(41/43)95% CI:85% - 99%100%(1581/1581)95% CI:100% -100%
Not Detected2a15811583
Total4315811624

Table 4b. Simplexa™ Congenital CMV Direct Clinical Agreement - Urine (Prospective)

a Two (2) urine specimens were negative by routine methodology.

PPA = Positive Percent Agreement, NPA = Negative Percent Agreement. 95% Cl = 95% Confidence Interval

The 95% confidence intervals (CI) were calculated following the Wilson Score method.

REPRODUCIBILITY

Reproducibility for the Simplexa™ Congenital CMV Direct assay was evaluated. Three (3) investigative sites assessed the device's inter-site, inter-day and inter/intra-assay reproducibility. Each of the sites tested the Simplexa™ Congenital CMV Direct Positive Control, No Template Control (NTC), a negative urine sample, a negative saliva swab in UTM sample and eight (8) contrived samples spiked into a negative matrix of either saliva swabs in UTM or urine. The eight (8) contrived samples consisted of a low positive (LP) contrived at approximately 3X the limit of detection (LoD) and a medium positive (MP) contrived at approximately 10X LOD for each of the following: CMV strain AD169 in saliva swabs in UTM, CMV strain AD169 in urine, CMV strain Towne in saliva swabs in UTM, and CMV strain Towne in urine. Each contrived sample was prepared by spiking a specific concentration of the strain into CMV negative urine or a CMV neqative saliva swab in UTM. The samples were tested in quadruplicate on nine (9) different days. Each site had three (3) operators who each assayed the entire sample panel and Positive Control twice per day, for a total of two (2) sets of data per day on one (1) LIAISON® MDX instrument, per site. The combined results for all sites are presented in Table 5. The results show the reproducibility of the Simplexa™ Congenital CMV Direct % CV ranged between 0.4-1.6%.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for DiaSorin Molecular. The word "DiaSorin" is written in a dark blue font on the top line, and the word "Molecular" is written in a green font on the second line. To the left of the text is a green and blue DNA helix.

510(k) Summary

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022

Page 7 of 17

Table 5. Simplexa™ Congenital CMV Direct Reproducibility

Site 1Site 2Site 3All Sites
SampleCMV Strainand MatrixAgreementwith expectedresultsAvg. Ct ±SD (%CV)Agreementwith expectedresultsAvg. Ct ±SD (%CV)Agreementwith expectedresultsAvg. Ct ±SD (%CV)AgreementwithexpectedresultsAvg. Ct ±SD (%CV)
SalivaTowne_LP100.0%(36/36)34.3 ±0.32(0.9%)100.0%(36/36)34.5 ±0.26(0.7%)100.0%(36/36)34.3 ±0.32(0.9%)100.0%(108/108)34.4 ±0.32(0.9%)
SalivaTowne MP100.0%(36/36)32.2 ±0.25(0.8%)100.0%(36/36)32.5 ±0.21(0.6%)100.0%(36/36)32.2 ±0.22(0.7%)100.0%(108/108)32.3 ±0.25(0.8%)
SalivaAD-169_LP100.0%(36/36)33.4 ±0.34(1.0%)100.0%(36/36)33.8 ±0.23(0.7%)100.0%(36/36)33.5 ±0.25(0.8%)100.0%(108/108)33.6 ±0.32(1.0%)
SalivaAD-169 MP100.0%(36/36)32.2 ±0.22(0.7%)100.0%(36/36)32.3 ±0.27(0.8%)100.0%(36/36)32.1 ±0.15(0.5%)100.0%(108/108)32.2 ±0.23(0.7%)
UrineTowne_LP100.0%(36/36)33.9 ±0.51(1.5%)100.0%(36/36)34.0 ±0.48(1.4%)100.0%(36/36)33.9 ±0.45(1.3%)100.0%(108/108)34.0 ±0.48(1.4%)
UrineTowne MP100.0%(36/36)32.1 ±0.26(0.8%)100.0%(36/36)32.3 ±0.26(0.8%)100.0%(36/36)32.1 ±0.21(0.7%)100.0%(108/108)32.2 ±0.26(0.8%)
UrineAD-169 LP100.0%(36/36)35.9 ±0.57(1.6%)100.0%(36/36)36.1 ±0.59(1.6%)97.2%(35/36)35.9 ±0.46(1.3%)99.1%(107/108)36.0 ±0.54(1.5%)
UrineAD-169 MP100.0%(36/36)34.0 ±0.42(1.2%)100.0%(36/36)34.3 ±0.37(1.1%)100.0%(36/36)34.0 ±0.33(1.0%)100.0%(108/108)34.1 ±0.40(1.2%)
SalivaNegative*100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(108/108)0.0 ± 0.00(N/A%)
UrineNegative*100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(36/36)0.0 ±0.00(N/A%)100.0%(108/108)0.0 ±0.00(N/A%)
NTC (UTM)100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(36/36)0.0 ± 0.00(N/A%)100.0%(108/108)0.0 ±0.00(N/A%)
PC as is100.0%(36/36)29.7 ±0.14(0.5%)100.0%(36/36)29.9 ±0.13(0.4%)100.0%(36/36)29.6 ±0.22(0.7%)100.0%(108/108)29.7 ±0.20(0.7%)

LP = Low Positive, MP = Moderate Positive, UTM = Universal Transport Media, NTC = No Template Control, SD = Standard Deviation, %CV = Percent Coefficient of Variation, Ct = Cycle Threshold

*Expected result for these samples is negative.

ANALYTICAL SENSITIVITY/LIMIT OF DETECTION

The limit of detection (LoD) was determined for the Simplexa™ Congenital CMV Direct assay using quantified stocks of three (3) CMV strains (AD169, Towne and Merlin) serially diluted in negative human saliva swab and urine matrices. The LoD was determined to be the lowest concentration that could be detected positive > 95% of the time. The LoD for each matrix is presented in Tables 6a and 6b.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a green and blue DNA helix graphic on the left, followed by the text "DiaSorin" in blue and "Molecular" in green. The text is aligned to the right of the DNA helix graphic.

Simplexa™ Congenital CMV Direct I Simplexa™ Congenital CMV Positive Control Pack. M Nov 2. 2022

Page 8 of 17

CMV strainLoD - Saliva in 1mL UTMLoD - Saliva in 3mL UTMLoD - Saliva in 3mL M4RT
Copies/mLin SalivaCopies/mLin UTMCopies/mLin salivaCopies/mLin UTMCopies/mLin salivaCopies/mLin M4RT
AD-1696,75050019,25050019,250500
Towne6,75050019,25050019,250500
CMV strainIU/mL insalivaIU/mL inUTMIU/mL insalivaIU/mL inUTMIU/mL insalivaIU/mL inM4RT
Merlin6,75050019,25050019,250500
Table 6b. Simplexa™ Congenital CMV Direct Limit of Detection - Urine
CMV strainCopies /mL
AD-169400 Copies/mL
Towne800 Copies/mL
CMV strainIU/mL
Merlin6,400 IU/mL

ANALYTICAL REACTIVITY/CROSS REACTIVITY

Analytical Reactivity

The analytical reactivity of the Simplexa™ Congenital CMV Direct assay was evaluated using different strains/genotypes of CMV that were not used in the determination of the limit of detection (LoD) for the assay. Quantified CMV was spiked at 1x LoD into negative adult saliva swab and negative urine from neonates less than 21 days of age. For the saliva swabs the preparations were spiked onto a flocked swab and transferred to each tube of UTM. The results are presented in Table 7. No genotype 4 (gB4) strains were available for testing. In addition to the strains that were physically tested, in silico BLAST analysis demonstrated that the assay should detect at least 327 CMV sequences available in the NCBI database, including the four (4) CMV genotypes gB1, gB2, gB3 and gB4.

Table 7. Simplexa™ Congenital CMV Direct Analytical Reactivity

CMV (gB3) Strain andMatrixAgreement withExpected Results(#Detected/#Total)
CMV Toledo Strain(Saliva Swab)100% (10/10)
CMV Toledo Strain(Urine)100% (10/10)

Cross-Reactivity (Analytical Specificity)

The Simplexa™ Congenital CMV Direct assay's analytical specificity was evaluated by testing the ability of the assay to exclusively identify CMV virus with no cross-reactivity to organisms that are closely related, cause similar clinical symptoms or may be present in saliva swabs and urine. Forty-one (41) potential crossreactants were spiked into negative saliva swab and 13 potential cross reactants were spiked into negative

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in green and blue on the left, followed by the company name "DiaSorin" in blue and "Molecular" in green on the right. The text is clean and professional, suggesting a focus on scientific and technological innovation.

510(k) Summary

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022 Page 9 of 17

urine. The samples were assayed in triplicate. No cross-reactivity was observed. The results are presented in Tables 8a and 8b.

No.OrganismTested Concentration% Agreement**(# Expected Results/ #Tested)
1Acinetobacter baumannii1x106 CFU/mL100.0% (3/3)
2Actinomyces odontolyticus1x106 CFU/mL100.0% (3/3)
3Bordetella pertussis1x106 CFU/mL100.0% (3/3)
4Coronavirus 229E1x105 TCID50/mL100.0% (3/3)
5CoxsackievirusA91x105 TCID50/mL100.0% (3/3)
6Epstein-Barr Virus1x105 IU/mL100.0% (3/3)
7Enterovirus 71*1x104 TCID50/mL100.0% (3/3)
8FLU A/ Michigan/45/20151x105 Cps/mL100.0% (3/3)
9FLU B/ Phuket/3073/20131x105 Cps/mL100.0% (3/3)
10Fusobacterium nucleatum1x106 CFU/mL100.0% (3/3)
11Adenovirus (C1)1x105 Cps/mL100.0% (3/3)
12Haemophilus influenza1x106 CFU/mL100.0% (3/3)
13Haemophilus parainfluenzae1x106 CFU/mL100.0% (3/3)
14Herpes Simplex Virus 11x105 IU/mL100.0% (3/3)
15Human herpesvirus 6A1x105 Cps/mL100.0% (3/3)
16Human herpesvirus 6B1x105 Cps/mL100.0% (3/3)
17Human herpesvirus 71x105 Cps/mL100.0% (3/3)
18Human herpesvirus 81x105 Cps/mL100.0% (3/3)
19Human Genomic DNA1x106 Cps/mL100.0% (3/3)
20Human metapneumovirus1x105 TCID50/mL100.0% (3/3)
21Klebsiella oxytoca1x106 CFU/mL100.0% (3/3)
22Klebsiella pneumoniae1x106 CFU/mL100.0% (3/3)
23Moraxella catarrhalis1x106 CFU/mL100.0% (3/3)
24Mycoplasma pneumoniae1x106 CCU/mL100.0% (3/3)
25Parainfluenza virus 11x105 U/mL100.0% (3/3)
26Parainfluenza virus 21x105 U/mL100.0% (3/3)
27Parainfluenza virus 31x105 U/mL100.0% (3/3)
28Porphyromonas gingivalis1x106 CFU/mL100.0% (3/3)
29Pseudomonas aeruginosa1x106 CFU/mL100.0% (3/3)
30Respiratory syncytial virus A1x105 IU/mL100.0% (3/3)
31Respiratory syncytial virus B1x105 TCID50/mL100.0% (3/3)
32Rhinovirus1x105 U/mL100.0% (3/3)
33Staphylococcus aureus1x106 CFU/mL100.0% (3/3)
34Staphylococcus epidermidis1x106 CFU/mL100.0% (3/3)
35Streptococcus anginosus1x106 CFU/mL100.0% (3/3)
36Streptococcus oralis1x106 CFU/mL100.0% (3/3)
No.OrganismTested Concentration% Agreement**(# Expected Results/ #Tested)
37Streptococcus mitis1x106 CFU/mL100.0% (3/3)
38Streptococcus pneumoniae1x106 CFU/mL100.0% (3/3)
39Streptococcus salivarius1x106 CFU/mL100.0% (3/3)
40Streptococcus sanguinis1x106 CFU/mL100.0% (3/3)
41Varicella Zoster Virus1x105 Cps/mL100.0% (3/3)

Table 8a. Simplexa™ Congenital CMV Direct Cross-Reactivity (Analytical Specificity) - Saliva Swab

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with the word "Molecular" underneath in a lighter green color.

510(k) Summarv

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2. 2022

Page 10 of 17

** Expected result for all organisms is negative.

*Enterovirus 71 was tested at a concentration lower than 1x10 TCIDs/mL due to the low concentration of the virus stock that was available. In silico (BLAST) analysis was also performed. The BLAST analysis showed that no crossreactivity is expected with this microorganism.

No.OrganismTested Concentration% Agreement(# Expected Results/ #Tested)
1Candida albicans1x106 CFU/mL100.0% (3/3)
2Enterobacter aerogenes1x106 CFU/mL100.0% (3/3)
3Enterobacter cloacae1x106 CFU/mL100.0% (3/3)
4Enterococcus faecium1x106 CFU/mL100.0% (3/3)
5Enterococcus faecalis1x106 CFU/mL100.0% (3/3)
6Escherichia coli1x106 CFU/mL100.0% (3/3)
7Herpes Simplex Virus 21x105 cps/mL100.0% (3/3)
8Lactobacillus acidophilus1x106 CFU/mL100.0% (3/3)
9Morganella morganii1x106 CFU/mL100.0% (3/3)
10Proteus mirabilis1x106 CFU/mL100.0% (3/3)
11Proteus vulgaris1x106 CFU/mL100.0% (3/3)
12Streptococcus agalactiae (GBS)1x106 CFU/mL100.0% (3/3)
13Enterovirus 71*1x104 TCID50/mL100.0% (3/3)

Table 8b. Simplexa™ Congenital CMV Direct Cross-Reactivity (Analytical Specificity) - Urine

*Expected result for all organisms is negative.

*Enterovirus 71 was tested at a concentration lower than 1x10 TCIDs/mL due to the virus stock that was available. In silico (BLAST) analysis was also performed. The BLAST analysis showed that no cross-reactivity is expected with this microorganism.

INTERFERENCE

The performance of the Simplexa™ Congenital CMV Direct assay was evaluated with potentially interfering substances that may be present in saliva swabs and urine samples at the concentrations indicated in the table below. A total of 17 potentially interfering substances were tested for saliva swabs and seven (7) potentially interfering substances were tested for urine in a low positive CMV sample at approximately three times the limit of detection (3X LoD) in saliva swab and urine matrices and assayed in triplicate. No interference was observed. The results are presented in Tables 9a and 9b.

Table 9a. Simplexa™ Congenital CMV Direct Interference for saliva swab

No.Potential InterferentCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
1Acetylsalicylic acidAD1690.65 mg/mL100.0% (3/3)

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue above the word "Molecular" in green.

510(k) Summary

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022

Page 11 of 17

No.Potential InterferentCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
1Towne0.65 mg/mL100.0% (3/3)
2Breast milkAD16910% (v/v)100.0% (3/3)
Towne10% (v/v)100.0% (3/3)
3CaffeineAD1690.06 mg/mL100.0% (3/3)
Towne0.06 mg/mL100.0% (3/3)
4CaseinAD16910 mg/mL100.0% (3/3)
Towne10 mg/mL100.0% (3/3)
5Enfamil Poly-vi-sol withIronAD1691.5 mg/mL100.0% (3/3)
Towne1.5 mg/mL100.0% (3/3)
6Enfamil™ formula neuropro®AD16910% (v/v)100.0% (3/3)
Towne10% (v/v)100.0% (3/3)
7Enfamil™ Tri-Vi-Sol®AD1698% (v/v)100.0% (3/3)
Towne8% (v/v)100.0% (3/3)
8Gaviscon® (SodiumAlginate)AD1691.2 mg/mL100.0% (3/3)
Towne1.2 mg/mL100.0% (3/3)
9Infants' Pain & Fever(Acetaminophen)AD1690.4 mg/mL100.0% (3/3)
Towne0.2 mg/mL100.0% (3/3)
10Infants' Mylicon® GasRelief (Simethicone)AD1696.7 mg/mL100.0% (3/3)
Towne6.7 mg/mL100.0% (3/3)
11Little Remedies SalineDropsAD16910% (v/v)100.0% (3/3)
Towne10% (v/v)100.0% (3/3)
12Motrin Infant Drops(Infants' Ibuprofen)AD1690.5 mg/mL100.0% (3/3)
Towne0.5 mg/mL100.0% (3/3)
13MucinAD16925 mg/mL100.0% (3/3)
Towne25 mg/mL100.0% (3/3)
14NystatinAD1691,727 U/mL100.0% (3/3)
Towne1,727 U/mL100.0% (3/3)
15PrednisoneAD1690.0003 mg/mL100.0% (3/3)
Towne0.0003 mg/mL100.0% (3/3)
16White Blood CellAD16910% (v/v)100.0% (3/3)
Towne10% (v/v)100.0% (3/3)
17Whole bloodAD16910% (v/v)100.0% (3/3)
Towne10% (v/v)100.0% (3/3)

**Expected result for all potential interferents is positive.

Table 9b. Simplexa™ Congenital CMV Direct Interference for urine
-----------------------------------------------------------------------------
No.OrganismCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
1Baby powderAD16910% (w/v)100% (3/3)

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, stacked above the word "Molecular" in a lighter green, sans-serif font.

510(k) Summarv

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2. 2022

Page 12 of 17

No.OrganismCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
Towne10% (w/v)100% (3/3)
2Johnson's Baby Oil(Mineral Oil)AD16910% (v/v)100% (3/3)
Towne10% (v/v)100% (3/3)
MeconiumAD1690.5% (w/v)100% (3/3)
3Towne1% (w/v)100% (3/3)
Moist Towelettes withBenzalkonium ChlorideAD16910% (w/v)100% (3/3)
4Towne10% (w/v)100% (3/3)
5NystatinAD1690.3mg/mL100% (3/3)
Towne0.3mg/mL100% (3/3)
6StoolAD1692% (w/v)100% (3/3)
Towne2% (w/v)100% (3/3)
7Whole bloodAD1695% (v/v)100% (3/3)
Towne10% (v/v)100% (3/3)

**Expected result for all potential interferents is positive.

INHIBITION BY OTHER MICROORGANISMS

The Simplexa™ Congenital CMV Direct assay was evaluated by testing the ability to identify CMV virus when other potentially inhibitory organisms are present. A panel of forty-one (41) potentially inhibitory organisms was individually spiked into a pool with a low concentration of CMV at approximately three times the limit of detection (3X LoD) in saliva. Thirteen (13) potentially inhibitory organisms were individually spiked into a pool with a low concentration of CMV at approximately three the limit of detection (3X LoD) in urine. Potentially inhibiting organisms were tested at the concentrations specified in Tables 10a and 10b. No inhibition by other organisms was observed.

Table 10a Simplexa™ Congenital CMV Direct Microbial Interference – Saliva Swab

No.OrganismCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
1Acinetobacter baumanniiAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
2Actinomyces odontolyticusAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
3Bordetella pertussisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
4Coronavirus 229EAD1691x105 TCID50/mL100.0% (3/3)
Towne1x105 TCID50/mL100.0% (3/3)
5CoxsackievirusA9AD1691x105 TCID50/mL100.0% (3/3)
Towne1x105 TCID50/mL100.0% (3/3)
6Epstein-Barr VirusAD1691x105 IU/mL100.0% (3/3)
Towne1x105 IU/mL100.0% (3/3)
7Enterovirus 71*AD1691x104 TCID50/mL100.0% (3/3)
No.OrganismCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
8FLU A/ Michigan/45/2015Towne1x104 TCID50/mL100.0% (3/3)
AD1691x105 Cps/mL100.0% (3/3)
9FLU B/ Phuket/3073/2013Towne1x105 Cps/mL100.0% (3/3)
AD1691x105 Cps/mL100.0% (3/3)
10Fusobacterium nucleatumTowne1x106 CFU/mL100.0% (3/3)
AD1691x106 CFU/mL100.0% (3/3)
11Adenovirus (C1)Towne1x105 Cps/mL100.0% (3/3)
AD1691x105 Cps/mL100.0% (3/3)
12Haemophilus influenzaTowne1x106 CFU/mL100.0% (3/3)
AD1691x106 CFU/mL100.0% (3/3)
13Haemophilus parainfluen zaeTowne1x106 CFU/mL100.0% (3/3)
AD1691x106 CFU/mL100.0% (3/3)
14Herpes Simplex Virus 1Towne1x105 IU/mL100.0% (3/3)
AD1691x105 IU/mL100.0% (3/3)
15Human herpesvirus 6ATowne1x105 Cps/mL100.0% (3/3)
AD1691x105 Cps/mL100.0% (3/3)
16Human herpesvirus 6BTowne1x105 Cps/mL100.0% (3/3)
AD1691x105 Cps/mL100.0% (3/3)
17Human herpesvirus 7Towne1x105 Cps/mL100.0% (3/3)
AD1691x105 Cps/mL100.0% (3/3)
18Human herpesvirus 8Towne1x105 Cps/mL100.0% (3/3)
AD1691x105 Cps/mL100.0% (3/3)
19Human Genomic DNATowne1x106 Cps/mL100.0% (3/3)
AD1691x106 Cps/mL100.0% (3/3)
20Human metapneumovirusTowne1x105 TCID50/mL100.0% (3/3)
AD1691x105 TCID50/mL100.0% (3/3)
21Klebsiella oxytocaTowne1x106 CFU/mL100.0% (3/3)
AD1691x106 CFU/mL100.0% (3/3)
22Klebsiella pneumoniaeTowne1x106 CFU/mL100.0% (3/3)
AD1691x106 CFU/mL100.0% (3/3)
23Moraxella catarrhalisTowne1x106 CFU/mL100.0% (3/3)
AD1691x106 CFU/mL100.0% (3/3)
24Mycoplasma pneumoniaeTowne1x106 CCU/mL100.0% (3/3)
AD1691x106 CCU/mL100.0% (3/3)
25Parainfluenza virus 1Towne1x105 U/mL100.0% (3/3)
AD1691x105 U/mL100.0% (3/3)
26Parainfluenza virus 2Towne1x105 U/mL100.0% (3/3)
AD1691x105 U/mL100.0% (3/3)
27Parainfluenza virus 3Towne1x105 U/mL100.0% (3/3)
AD1691x105 U/mL100.0% (3/3)
No.OrganismCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
28Porphyromonas gingivalisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
29Pseudomonas aeruginosaAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
30Respiratory syncytial virus AAD1691x105 IU/mL100.0% (3/3)
Towne1x105 IU/mL100.0% (3/3)
31Respiratory syncytial virus BAD1691x105 TCID50/mL100.0% (3/3)
Towne1x105 TCID50/mL100.0% (3/3)
32RhinovirusAD1691x105 U/mL100.0% (3/3)
Towne1x105 U/mL100.0% (3/3)
33Staphylococcus aureusAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
34Staphylococcus epidermidisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
35Streptococcus anginosusAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
36Streptococcus oralisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
37Streptococcus mitisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
38Streptococcus pneumoniaeAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
39Streptococcus salivariusAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
40Streptococcus sanguinisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
41Varicella Zoster VirusAD1691x105 Cps/mL100.0% (3/3)
Towne1x105 Cps/mL100.0% (3/3)

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in dark blue, with the word "Molecular" underneath in green.

510(k) Simplexa™ Congenital CMV Direct MOL2250
Simplexa™ Congenital CMV Direct MOL2250
Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022

Page 13 of 17

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the alphanumeric string "K202755" in a bold, sans-serif font. The characters are arranged horizontally and are uniformly sized. The string appears to be a code or identifier, possibly a product or serial number.

Image /page/16/Picture/1 description: The image contains the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, with the word "Molecular" underneath in a lighter green color.

510(k) Summary

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022

Page 14 of 17

*Expected result for all organisms is positive.

  • Enterovirus 71 was tested at a concentration lower than the 1x105 TCID50/mL due to the virus stock that was available. In silico (BLAST)

analysis was also performed. The results of the BLAST analysis showed that no microbial inhibition is expected with this microorganism.

-----------------------------------AMI OL-----Tested% Agreeme14 F -------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Table 10b Simplexa™ Congenital CMV Direct Microbial Interference - Urine

No.OrganismCMV StrainTestedConcentration% Agreement(# Expected Results/ #Tested)
1Candida albicansAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo features a stylized DNA double helix in shades of green and blue on the left. To the right of the helix are the words "DiaSorin" in a dark blue, sans-serif font, stacked above the word "Molecular" in a lighter green, sans-serif font.

510(k) Summarv

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. M 2260 Nov 2. 2022

Page 15 of 17

No.OrganismCMV StrainTestedConcentration% Agreement**(# Expected Results/ #Tested)
2Enterobacter aerogenesAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
3Enterobacter cloacaeAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
4Enterococcus faeciumAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
5Enterococcus faecalisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
6Escherichia coli ATCCAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
7Herpes Simplex Virus 2AD1691x105 Cps/mL100.0% (3/3)
Towne1x105 Cps/mL100.0% (3/3)
8Lactobacillus acidophilusAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
9Morganella morganiiAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
10Proteus mirabilisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
11Proteus vulgarisAD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
12Streptococcusagalactiae (GBS)AD1691x106 CFU/mL100.0% (3/3)
Towne1x106 CFU/mL100.0% (3/3)
13Enterovirus 71*AD1691x104 TCID50/mL100.0% (3/3)
Towne1x104 TCID50/mL100.0% (3/3)

** Expected result for all organisms is positive.

  • Enterovirus 71 was tested at a concentration lower than the 1x10s TCIDs; mL due to the virus stock that was available. In silico (BLAST) analysis was also performed. The BLAST analysis showed that no microbial inhibition is expected with this microorganism.

CARRY-OVER CONTAMINATION

Amplification carry-over for the Simplexa™ assays has been assessed. The study was designed by alternately placing high positive and negative samples on each disc. No evidence of carry-over contamination was observed.

The performance of the Simplexa™ Congenital CMV Direct assay was established in a clinical study that included two (2) cohorts based on sample status. Specifically, prospective samples from infants less than twenty-one (21) days of age were tested in the clinical agreement study.

EXPECTED VALUES

The prevalence of CMV as determined by the Simplexa™ Congenital CMV Direct assay in a multi-site clinical study with prospectively collected specimens was 1.19% for saliva swabs and 2.59% for urine. Table

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image contains the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix on the left, rendered in shades of green and blue. To the right of the helix are the words "DiaSorin" in a bold, dark blue font, stacked above the word "Molecular" in a lighter green font.

510(k) Sui mmarv

Simplexa™ Congenital CMV Direct MOL2250 Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022 Page 16 of 17

8 shows the prevalence of saliva swabs by collection site and Table 9 shows the prevalence of urine by collection site.

Table 8. Prospective Results: Simplexa™ Congenital CMV Direct Expected Values for Saliva Swabs by Collection Site

Site IDTotal SpecimensSimplexa™ CongenitalCMV Direct DetectedCMV Prevalence
714200.00%
81616.25%
963230.47%
10900.00%
121400.00%
131119.09%
14900.00%
151000.00%
18800.00%
191002121.20%
All1853170.92%

Table 9. Prospective Results: Simplexa™ Congenital CMV Direct Expected Values for Urine by Collection Site

Site IDAllSimplexa™ CongenitalCMV Direct DetectedCMV Prevalence
517121.17%
61336362.69%
7500.00%
84724.26%
10800.00%
121300.00%
131119.09%
14900.00%
151000.00%
181100.00%
19300.00%
All1624412.52%

Conclusion

The analytical and method comparison studies have demonstrated that the Simplexa™ Congenital CMV Direct is Substantially Equivalent to the predicate device (DEN180040).The device labeling is compliant with 21 CFR § 809.10.

{19}------------------------------------------------

Image /page/19/Picture/0 description: The image shows the logo for DiaSorin Molecular. The logo consists of a stylized DNA double helix on the left, with the words "DiaSorin" in blue on the right. Below "DiaSorin" is the word "Molecular" in green.

510(k) Simplexa™ Congenital CMV Direct MOL2250
Simplexa™ Congenital CMV Direct MOL2250
Simplexa™ Congenital CMV Positive Control Pack. MOL2260 Nov 2, 2022 Page 17 of 17

§ 866.3181 Cytomegalovirus nucleic acid detection device for congenital cytomegalovirus infection.

(a)
Identification. A cytomegalovirus nucleic acid detection device for congenital cytomegalovirus infection is an in vitro diagnostic device intended for the qualitative detection of cytomegalovirus DNA in clinical samples from newborn babies to aid in the diagnosis of congenital cytomegalovirus infection. Negative results do not preclude infection and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. Positive results should be interpreted with consideration of other clinical information and laboratory findings and should not be used as the sole basis for treatment or other patient management decisions.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The labeling required under § 809.10(b) of this chapter must include:
(i) An intended use with a detailed description of what the device detects, the type of results provided to the user, the clinical indications appropriate for test use, and the specific population(s) to be tested.
(ii) A detailed device description, including all device components, instrument requirements, ancillary reagents required but not provided, and an explanation of the methodology, including all pre-analytical methods for specimen processing.
(iii) Performance characteristics from analytical and clinical studies required under paragraphs (b)(2)(ii) and (iii) of this section.
(iv) A detailed explanation of the interpretation of results and criteria for validity of results.
(v) A limiting statement that device results are not intended to be used as the sole basis for diagnosis, treatment, or other patient management decisions.
(vi) As applicable, a limiting statement and specific sample collection recommendations to indicate that breast milk can result in false positive results for saliva samples if samples are collected less than 1 hour after breastfeeding. Sample collection a minimum of 1 hour from breastfeeding must be recommended.
(vii) Detailed instructions for use that minimize the risk of generating a false result.
(2) Design verification and validation must include:
(i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for sequence selection, and computational path from collected raw data to reported result (
e.g., how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies including characterization of the cutoff, analytical sensitivity (limit of detection), inclusivity, reproducibility, interference, cross reactivity, instrument and method carryover/cross contamination, and sample stability and handling.
(iii) Detailed documentation from a clinical study documenting sensitivity and specificity of the device; if the number of positive samples in the clinical study is insufficient to properly estimate device sensitivity, additional pre-selected positive samples must be evaluated to supplement the study. Clinical study subjects must be consistent with the intended use population (
i.e., infants younger than 21 days of age), and device results must be compared to FDA-accepted comparator methods. Documentation from the clinical study must include the clinical study protocol, the clinical study report, testing results, and results of all statistical analyses.(iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software.