K Number
K202689

Validate with FDA (Live)

Device Name
Dentilube Spray
Manufacturer
Date Cleared
2021-09-03

(353 days)

Product Code
Regulation Number
N/A
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Dentilube dry mouth spray is indicated for the symptomatic relief from the effects of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors and other oral symptoms associated with dry mouth.

Device Description

Dentilube dry mouth spray is a viscous pink electrolyte-containing solution that is designed to mimic the compositions and actions of saliva to moisturize, and refresh the mouth. It provides a lubricating and moisturizing coating inside the mouth, thereby relieving symptoms of dry mouth. The product is supplied in 50mL or 100 mL PET bottles.

AI/ML Overview

The provided text describes the acceptance criteria and performance data for the Dentilube dry mouth spray (K202689). The study is a non-clinical performance test to demonstrate substantial equivalence to a predicate device, not a comparative effectiveness study with human readers or a standalone algorithm performance study.

Here's the breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are implicitly defined by demonstrating "similarity" or "sameness" to the predicate device, BioXtra® Moisturizing Gel (K072306), across various attributes and physical properties. The performance is reported against these comparative attributes.

Attribute / Acceptance Criteria (Similarity to Predicate)Reported Device Performance (Dentilube dry mouth spray)
ProductDentilube dry mouth spray
ManufacturerOxy2plus, LLC
510(K) #K202689
Product CodeLFD (Same as predicate)
RegulationPre-Amendment (Same as predicate)
ClassUnclassified (Same as predicate)
Review PanelDental (Same as predicate)
Indications for UseSymptomatic relief from effects of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors, and other oral symptoms associated with dry mouth. (Same as predicate)
Mode of ActionMoisturizing and lubricating oral dryness (Same as predicate)
Dosage FormOral spray (Same as predicate)
Type of ProductLiquid solution (Same as predicate)
Area of UseOral cavity (Same as predicate)
Presentation Rx/OTCNon-sterile OTC. (Predicate is Non-sterile Rx/OTC; subject device is OTC only)
Disease StateXerostomia (Same as predicate)
Method of UseReady to use gel spray (Same as predicate)
Application per DayAs needed (Same as predicate)
SolventPurified water (Same as predicate)
BuffersDipotassium hydrogen phosphate, Potassium dihydrogen phosphate. (Similar to predicate's Potassium phosphate, dibasic & Sodium Chloride; both are GRAS and provide similar pH)
Humectants/ThickenersSodium carboxymethyl-cellulose. (Similar to predicate's Hydroxyethyl cellulose, Polyacrylic acid, Sodium polyacrylate, Hydrogenated starch hydrolysates (HSH); both are GRAS and provide similar viscosity/moisture-retaining quality)
PreservativesMethyl P-hydroxybenzoate (Methylparaben). (Similar to predicate's Sodium methylparaben, Sodium propylparaben, Sodium benzoate, Potassium sorbate, Citric acid monohydrate, Lactoperoxidase; both are GRAS and used for maintaining/extending shelf life)
ColorantFD&C Red 40 Dye (2%) (Predicate has None; colorant does not affect use)
SweetenersSorbitol. (Similar to predicate's Sorbitol, Xylitol, Sodium Saccharin; both use sorbitol to balance solution and provide denser solution)
ElectrolytesCalcium chloride anhydrous, Magnesium Chloride anhydrous, Potassium Chloride, Sodium fluoride. (Similar to predicate's Calcium chloride dihydrate, Magnesium chloride hexahydrate, Potassium Chloride, Sodium monofluorophosphate; both contain electrolytes similar to natural saliva)
Environment of UseHome and Clinic (Same as predicate)
FlavorsLemon / Mint (Predicate is Mint; not considered a significant difference affecting intended use)
AppearanceClear Viscous Liquid (Same as predicate)
ColorReddish Pink (Predicate is Light Straw; different color does not affect use)
pH6.35 (Predicate is 6.82; similar, both near natural saliva pH 5.3-7.8)
SolubilityWater Soluble (Same as predicate)
Viscosity @25°C [RVT Spindle #5@10rpm]3500 cps (3.5 Pa-s) (Predicate is 2900 cps (2.9 Pa-s); similar, greater than natural saliva viscosity 0.0078 Pa-s)
Specific gravity (SPG)1.02 (Predicate is 1.05; similar, not significantly different from saliva SPG 1.000-1.010)
SterilityNon-sterile (Same as predicate)
Shelf Life2 years (Same as predicate)
BiocompatibilityConforms with ISO 10993-1 (cytotoxicity, sensitization, irritation) (Same as predicate)

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not detail a specific "test set" in the context of a clinical study with human subjects. The performance testing described is primarily non-clinical performance testing (e.g., pH, viscosity, specific gravity, shelf-life, biocompatibility). It aims to demonstrate that the physical and chemical properties of the device are substantially equivalent to the predicate device. Therefore, a sample size for a clinical test set is not applicable here.

The data provenance is not specified beyond being generated by Oxy2plus, LLC for the purpose of this 510(k) submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. This is not a study involving expert assessment of images or clinical outcomes. The "ground truth" for the non-clinical tests are standard physical and chemical measurements (e.g., pH meter readings, viscometer readings).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No expert adjudication process is described as this is a non-clinical, comparative assessment of product characteristics.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This document describes a medical device (dry mouth spray), not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study is not relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

No. This is a physical product, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical performance tests:

  • Physical Property Measurements: The ground truth for properties like pH, viscosity, specific gravity, and solubility are the direct measurements obtained through standard laboratory instruments and methods.
  • Shelf-Life/Stability: Determined by degradation studies over time.
  • Biocompatibility: Determined by adherence to ISO 10993 standards (e.g., cytotoxicity assays, sensitization studies, irritation tests).

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device.

9. How the ground truth for the training set was established

Not applicable. This is not a machine learning or AI device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

September 3, 2021

Oxy2plus, LLC Chun Lin Co-Founder 4120 Rosemead Blvd Rosemead, California 91770

Re: K202689

Trade/Device Name: Dentilube Spray Regulatory Class: Unclassified Product Code: LFD Dated: June 28, 2021 Received: July 7, 2021

Dear Chun Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Michael E. Adiodha. M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202689

Device Name Dentilube dry mouth sprav

Indications for Use (Describe)

Dentilube dry mouth spray is indicated for the symptomatic relief from the effects of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors and other oral symptoms associated with dry mouth.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY (K202689)

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

SUBMITTER INFORMATION 1.

Name:Oxy2plus, LLC
Address:4120 Rosemead
Blvd Rosemead,
CA 91770
Contact person:Chun Nan Lin, DDS, MS
Telephone:626-286-7800
FAX:626-286-7600
Email:oxy2plus@outlook.com
Date Summary Prepared:June 26, 2021

2. DEVICE NAME

Device Name:Dentilube dry mouth spray
Trade or Proprietary Name:Dentilube dry mouth spray
Common or Usual Name:Saliva, Artificial Classification
Name:Saliva, Artificial
Product Code:LFD
Classification:Unclassified
Panel:Dental

3. IDENTIFICATION OF EQUIVALENCE (Predicate Device):

BIO-X HEALTHCARE S.A. BioXtra® Moisturizing Gel (K072306)

4. DEVICE DESCRIPTION

Dentilube dry mouth spray is a viscous pink electrolyte-containing solution that is designed to mimic the compositions and actions of saliva to moisturize, and refresh the mouth. It provides a lubricating and moisturizing coating inside the mouth, thereby relieving symptoms of dry mouth.

The product is supplied in 50mL or 100 mL PET bottles.

5. STATEMENT OF INTENDED USE

{4}------------------------------------------------

Dentilube dry mouth spray is indicated for the symptomatic relief from the effects of chronic or temporary xerostomia (dry mouth), mouth discomfort, mouth odors and other oral symptoms associated with dry mouth.

6. SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS

AttributesSubject DevicePredicate DeviceComparison
ProductDentilube dry mouthsprayBioXtra® MoisturizingGel-
ManufacturerOxy2plus, LLCBio-X Healthcare S.A.-
510(K) #K202689K072306-
Product CodeLFDLFDSame
RegulationPre-AmendmentPre-AmendmentSame
ClassUnclassifiedUnclassifiedSame
Review PanelDentalDentalSame
Indications forUseDentilube dry mouth sprayis indicated for thesymptomatic relief fromthe effects of chronic ortemporary xerostomia (drymouth), mouth discomfort,mouth odors and other oralsymptoms associated withdry mouth.BioXtra® is indicatedfor the symptomaticrelief from the effects ofchronic or temporaryxerostomia (dry mouth),mouth discomfort,mouth odors and otheroral symptomsassociated with drymouth.Same; both areindicated forrelief ofsymptoms ofxerostomia (drymouth).

Characteristics of the device compared to the predicate device Substantial Equivalence Comparison Chart

Conditions of UseMode ofActionMoisturizing andlubricating oral drynessMoisturizing andlubricating oral drynessSame
Dosage FormOral sprayOral spraySameType ofProductLiquid solutionLiquid solutionSame
Area of UseOral cavityOral cavitySamePresentationRx/OTCNon-sterileOTCNon-sterileRx/OTCsameThe subjectdevice is OTConly
Disease StateXerostomiaXerostomiaSameComposition
Method of UseReady to use gel sprayReady to use gel spraySameSolventPurified waterPurified watersame
Applicationper DayAs neededAs neededSameBuffersDipotassium hydrogenphosphatePotassium dihydrogenphosphatePotassium phosphate,dibasicSodium ChlorideSimilar. Thesedifferentbuffering agentsare generallyrecognized assafe (GRAS) andprovide similarpH levels for theboth devices.
Packaging50 mL and 100 mL inPET bottle with sprayhead50 mL in PET bottlewith spray headSameHumectants/ThickenersSodium carboxymethyl-celluloseHydroxyethyl cellulosePolyacrylic acidSodium polyacrylateHydrogenated starchhydrolysates (HSH)Similar. Thesedifferent thickeneragents are generallyrecognized as safe(GRAS) and providesimilar viscosity andmoisture-retainingquality for bothdevices.
Area of UseOral cavityOral cavitySamePreservativesMethyl P-hydroxybenzoate(Methylparaben)Sodium methylparabenSodium propylparabenSodium benzoatePotassium sorbateCitric acid monohydrateLactoperoxidase(Hydrogen-peroxideoxidoreductase)Similar. Thesepreservative agentsare generallyrecognized as safe(GRAS) by theUSFDA and are usedfor maintaining andextending shelf life.
Environmentof UseHome and ClinicHome and ClinicSameColorantFD&C Red 40 Dye(2%)NoneColorant is notused in thepredicate device
SweetenersSorbitolSorbitolXylitolSodium SaccharinSimilar. Thepredicate has moresweeteners. Bothdevices use sorbitolto balance thesolution and toprovide a densersolution.

{5}------------------------------------------------

{6}------------------------------------------------

ElectrolytesCalcium chloride,anhydrousMagnesium Chloride,anhydrousPotassium ChlorideSodium fluorideCalcium chloride,dihydrateMagnesium chloride,hexahydratePotassium ChlorideSodium monofluoro-phosphateSimilar. Both devicescontain electrolytessimilar to thosefound in naturalsaliva.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

The subject device, Dentilube dry mouth spray, and the predicate device have the same indications and method of use. They also share many conditions of use. Further, both employ the same fundamental scientific technology (a formulation of water, humectants or moisturizers, thickening/binding agents, buffering agents, sweeteners, flavor, surfactants and preservatives).

Physical Properties (Non-Clinical Performance)
FlavorsLemon / MintMint
AppearanceClear Viscous LiquidClear Viscous LiquidSame
ColorReddish PinkLight StrawDifferent colordoes not affectuse
pH6.356.82Similar. Bothsubject andpredicatedevices have pHvalues near thatof natural saliva(5.3 - 7.8)
SolubilityWater SolubleWater SolubleSame
Viscosity@25°C [RVTSpindle#5@10rpm]3500 cps (3.5 Pa-s)2900 cps (2.9 Pa-s)Similar. Thevalues aregreater viscosityof natural saliva(0.0078 Pa-s).
Specific gravity(SPG)1.021.05Similar. Thevalues are notsignificantlydifferent whencompared to theSPG of thesaliva (1.000-1.010).

SUMMARY OF NON-CLINICAL PERFORMANCE TESTING 7.

{7}------------------------------------------------

SterilityNon-sterileNon-sterileSame
Shelf Life2 years2 yearsSame
BiocompatibilityConforms with ISO10993-1• cytotoxicity• sensitization• irritationConforms with ISO10993-1• cytotoxicity• sensitization• irritationSame

Dentilube dry mouth spray has been tested for shelf-life/stability, which provides for a 24-month shelf life, same as the primary predicate device. Biocompatibility assessments has been done in accordance with ISO 10993 for cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), and irritation (ISO 10993-10) and has been shown to be safe for the intended use. No other clinical tests were performed other than a Use Study for this submission.

DISCUSSION AND CONCLUSION 8.

Based on the comparison of intended use and technical characteristics, as well as non-clinical performance testing, we conclude that Dentilube dry mouth spray is substantially equivalent to the predicate device.

N/A