K Number
K202678

Validate with FDA (Live)

Date Cleared
2021-03-01

(167 days)

Product Code
Regulation Number
880.5725
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OMNI® Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma.

Device Description

The Sight Sciences OMNI Surgical System (with modified indication) is a handheld, manually operated device used by ophthalmologists to access, microcatheterize, and viscodilate Schlemm's canal ("canaloplasty") and to re-access Schlemm's canal and cut trabecular meshwork tissue ("trabeculotomy"). The OMNI Surgical System is provided sterile and disposed after single-patient use. The device is fabricated from biocompatible materials standard to the medical device industry. Each OMNI Surgical System device dispenses fluid on the principle of exchanging volumes much like a syringe and is designed to function with commercially available cohesive viscoelastic fluids (also known as ophthalmic viscosurgical device, or "OVD").

The OMNI Surgical System device includes a stainless-steel cannula, polymeric microcatheter, removable priming lock, internal reservoir and plunger tube, a Luer fitting for direct connection with an OVD cartridge to prime the internal reservoir, and two advancement wheels. A single advancement wheel is located on each side of the handle. This allows the OMNI Surgical System device to be used in either eye (OD or OS) and in either hand of the surgeon (left or right), by turning the device 180 degrees along its vertical axis. These wheels are used to advance and retract the microcatheter.

The stainless-steel cannula has a curved shape with a beveled tip for entry through the trabecular meshwork into Schlemm's canal. To perform the combined and sequential canaloplasty/trabeculotomy procedures, the canaloplasty is performed first, followed by trabeculotomy as explained in further detail below.

Performing Canaloplasty First: the microcatheter is advanced into Schlemm's canal up to 180 degrees (one hemisphere) by rotating the advancement wheel forward until the wheel stops (about 20mm). When the device is being used to deliver viscoelastic fluid, retraction of the microcatheter causes the plunger tube to advance into the viscoelastic fluid reservoir thereby automatically dispensing viscoelastic fluid along the length of Schlemm's canal and collector channels. The microcatheter can be advanced/retracted up to 20 mm per cycle by manually rotating the advancement wheel. The microcatheter can be fully advanced/retracted multiple times, however, viscoelastic fluid can only be dispensed during the first two advancement/retraction cycles in order to dispense viscoelastic fluid along each hemisphere of Schlemm's canal. Thus, the OMNI Surgical System device is designed to be used twice within Schlemm's canal to deliver a controlled volume of viscoelastic fluid along the first 180 degrees of the canal, followed by a second delivery of viscoelastic fluid along the other 180 degrees. The OMNI Surgical System delivers a total viscoelastic fluid volume of 11 microliters throughout Schlemm's canal (approximately 5.5 microliters for each of the first two advancement/retraction cycles).

Performing Trabeculotomy Second: the beveled tip of the curved stainless-steel cannula is repositioned into the same Schlemm's canal location after finishing canaloplasty. The polymeric microcatheter is re-advanced into Schlemm's canal up to 180 degrees (one hemisphere) by rotating the advancement wheel forward until the wheel stops (about 20 mm). With the microcatheter resting in the canal, the cannula is removed from the corneal incision and out of the eye causing the microcatheter to cut through the trabecular meshwork. This process can be repeated in the second Schlemm's hemisphere.

AI/ML Overview

The provided text describes the OMNI® Surgical System and its substantial equivalence to a predicate device, focusing on bench testing and clinical evidence rather than explicitly defining acceptance criteria and subsequent studies in the typical AI/ML context. This document is a 510(k) summary for a medical device (infusion pump), not an AI/ML device. Therefore, the questions related to AI/ML device performance (like experts for ground truth, adjudication methods, MRMC studies, standalone performance, training sets) are not directly applicable or answerable from the provided text.

However, I can extract information related to the device's performance based on the clinical study.

1. Table of Acceptance Criteria and Reported Device Performance

As this is a 510(k) for a non-AI/ML medical device, there aren't explicit acceptance criteria in the way one would define them for an AI/ML algorithm (e.g., target specificity, sensitivity, etc.). Instead, the device's performance is demonstrated through its ability to reduce intraocular pressure (IOP) and its safety profile, which are compared against a historical control and expected outcomes. The key performance indicators evaluated were IOP reduction and medication reduction.

Performance MetricAcceptance Criteria (Implicit from Study Design/Objectives)Reported Device Performance (ROMEO Study)
Effectiveness: IOP ReductionClinically significant IOP reduction in adult patients with primary open-angle glaucoma, sustained up to 12 months. (Compared against Lewis 2007 literature control).+Cataract Group (Baseline IOP ≥ 16 mmHg, n=45): - Baseline Mean IOP: 19.5 ± 3.8 mmHg - 12 Month Mean IOP: 15.2 ± 3.0 mmHg Standalone Group (Baseline IOP ≥ 16 mmHg, n=38): - Baseline Mean IOP: 20.0 ± 3.6 mmHg - 12 Month Mean IOP: 15.3 ± 2.7 mmHg Responders (≥ 20% IOP reduction at 12 months, no medication increase, no secondary surgery): - Standalone (Pre-op IOP > 18 mmHg): 58.3% (14/24) - Standalone (Lewis criteria, Baseline IOP ≥ 16 mmHg): 45.7% (16/35) - +Cataract (BL > 18 mmHg): 62.5% (15/24) - +Cataract (Lewis criteria, Baseline IOP ≥ 16 mmHg): 43.5% (20/46)
Effectiveness: Medication ReductionReduction in the number of ocular hypotensive medications or at least no increase.All Patients (n=129): - Baseline Mean Medications: 1.8 ± 1.3 - 12 Month Mean Medications: 1.1 ± 1.2
Safety and Adverse EventsAdverse events should be infrequent, mild, non-serious, transient, and consistent with those expected in the target population. No serious device-related adverse events.Adverse events were generally infrequent, mild, non-serious, transient, and resolved with or without treatment. No serious adverse events or serious device-related adverse events reported. Most common (12-month visit): Posterior capsular opacity (14.7%), Mild anterior chamber inflammation (10.9%), Cystoid macular edema (5.4%), Corneal edema (4.7%), IOP increase ≥ 10 mmHg above baseline >30 days postoperative (4.7%), Hyphema > 1 mm (3.9%).
Secondary Surgical InterventionsAcceptable rate of re-intervention.7.0% (9/129 eyes) required secondary surgical intervention to reduce IOP. - +Cataract: 4.9% (4/81) - Standalone: 10.4% (5/48)

2. Sample Size for the Test Set and Data Provenance

  • Sample Size:
    • Clinical Study (ROMEO): 129 patients with a single qualifying eye.
      • +Cataract group: 81 procedures
      • Standalone group: 48 procedures
    • Bench Testing (Cadaver Eyes): 4 cadaver eyes, using 8 OMNI Surgical System devices.
  • Data Provenance:
    • Clinical Study (ROMEO): Retrospective, observational, multi-center, single-arm, consecutive case series study conducted at 11 sites throughout the U.S.
    • Bench Testing: Human cadaver eyes.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This is not applicable as the device is a surgical instrument, not an AI/ML diagnostic or prognostic device requiring expert-established ground truth on a test set. The clinical performance data involves measuring physiological parameters (IOP) and observing real-world outcomes over time. The "ground truth" here is the clinical efficacy and safety observed in patients. The document mentions the study was performed by an ophthalmologist and a physician assistant during cadaver eye testing.

4. Adjudication Method for the Test Set

Not applicable for this type of medical device study. Data collected in the ROMEO study would have followed standard clinical trial data collection and monitoring practices, but not specific adjudication for ground truth in the AI/ML sense.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This is a manually operated surgical instrument.

7. The Type of Ground Truth Used

  • Clinical Study (ROMEO): The "ground truth" for effectiveness was direct physiological measurement (intraocular pressure, number of medications) and observed clinical outcomes (adverse events, need for secondary surgery) in live human patients. The study compared its findings to a reference literature control (Lewis RA, von Wolff K, Tetz M, Kearney JR, Shingleton B, Samuelson TW. Canaloplasty: circumferential viscodilation and tensioning of Schlemm's canal using a flexible microcatheter for the treatment of open-angle glaucoma in adults: interim clinical study analysis. J Cataract Refract Surg. 2007 Jul;33(7):1217-26.)
  • Bench Testing (Cadaver Eyes): The ground truth was the validated ability of an ophthalmologist to perform the intended surgical steps (access, microcatheterize, viscodilate, cut trabecular meshwork) using the device in cadaveric tissue as per the Instructions For Use.

8. The Sample Size for the Training Set

Not applicable. This is a non-AI/ML device.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This is a non-AI/ML device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

March 1, 2021

Sight Sciences, Inc. Edward Sinclair Vice President, Regulatory Affairs 4040 Campbell Ave, Suite 100 Menlo Park, California 94025

Re: K202678

Trade/Device Name: OMNI® Surgical System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: MRH, HMZ Dated: January 18, 2021 Received: January 21, 2021

Dear Edward Sinclair:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LT Charles Chiang Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202678

Device Name OMNI® Surgical System

Indications for Use (Describe)

The OMNI® Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Submitter Information

510(k) Owner:Sight Sciences, Inc.4040 Campbell Ave., Suite 100Menlo Park, CA 94025Tel: (877) 266-1144
Contact Person:Edward J. SinclairVice President, Regulatory Affairs4040 Campbell Ave., Suite 100Menlo Park, CA 94025Tel: 650-218-9149
Date Prepared:March 1, 2021

Device Name and Classification

Trade Name:OMNI® Surgical System
Common Name:Ophthalmic Infusion Pump
Classification Name:Infusion Pump
Regulation Number:21 CFR 880.5725
Device Classification:Class II
Primary Product Code:MRH
Secondary Product Code:HMZ

Primary Predicate Device

Device Name:OMNI® Surgical System
510(k) Holder:Sight Sciences, Inc.
510(k) Number:K173332
Clearance Date:December 21, 2017

Reference Device

Device Name:iScience Interventional Canaloplasty Microcatheter (iTrack™ Surgical System)
510(k) Holder:Ellex iScience, Inc.
510(k) Number:K080067
Clearance Date:July 18, 2008

Intended Use

{4}------------------------------------------------

The OMNI® Surgical System is an ophthalmic surgical tool for the delivery of controlled amounts of viscoelastic fluid into the anterior segment and the cutting of trabecular meshwork when a trabeculotomy is indicated.

Indications for Use

The OMNI® Surgical System is indicated for canaloplasty (microcatheterization and transluminal viscodilation of Schlemm's canal) followed by trabeculotomy (cutting of trabecular meshwork) to reduce intraocular pressure in adult patients with primary open-angle glaucoma.

Device Description

The Sight Sciences OMNI Surgical System (with modified indication) is a handheld, manually operated device used by ophthalmologists to access, microcatheterize, and viscodilate Schlemm's canal ("canaloplasty") and to re-access Schlemm's canal and cut trabecular meshwork tissue ("trabeculotomy"). The OMNI Surgical System is provided sterile and disposed after single-patient use. The device is fabricated from biocompatible materials standard to the medical device industry. Each OMNI Surgical System device dispenses fluid on the principle of exchanging volumes much like a syringe and is designed to function with commercially available cohesive viscoelastic fluids (also known as ophthalmic viscosurgical device, or "OVD").

The OMNI Surgical System device includes a stainless-steel cannula, polymeric microcatheter, removable priming lock, internal reservoir and plunger tube, a Luer fitting for direct connection with an OVD cartridge to prime the internal reservoir, and two advancement wheels. A single advancement wheel is located on each side of the handle. This allows the OMNI Surgical System device to be used in either eye (OD or OS) and in either hand of the surgeon (left or right), by turning the device 180 degrees along its vertical axis. These wheels are used to advance and retract the microcatheter.

The stainless-steel cannula has a curved shape with a beveled tip for entry through the trabecular meshwork into Schlemm's canal. To perform the combined and sequential canaloplasty/trabeculotomy procedures, the canaloplasty is performed first, followed by trabeculotomy as explained in further detail below.

Performing Canaloplasty First: the microcatheter is advanced into Schlemm's canal up to 180 degrees (one hemisphere) by rotating the advancement wheel forward until the wheel stops (about 20mm). When the device is being used to deliver viscoelastic fluid, retraction of the microcatheter causes the plunger tube to advance into the viscoelastic fluid reservoir thereby automatically dispensing viscoelastic fluid along the length of Schlemm's canal and collector channels. The microcatheter can be advanced/retracted up to 20 mm per cycle by manually rotating the advancement wheel. The microcatheter can be fully advanced/retracted multiple times, however, viscoelastic fluid can only be dispensed during the first two advancement/retraction cycles in order to dispense viscoelastic fluid along each hemisphere of

{5}------------------------------------------------

Schlemm's canal. Thus, the OMNI Surgical System device is designed to be used twice within Schlemm's canal to deliver a controlled volume of viscoelastic fluid along the first 180 degrees of the canal, followed by a second delivery of viscoelastic fluid along the other 180 degrees. The OMNI Surgical System delivers a total viscoelastic fluid volume of 11 microliters throughout Schlemm's canal (approximately 5.5 microliters for each of the first two advancement/retraction cycles).

Performing Trabeculotomy Second: the beveled tip of the curved stainless-steel cannula is repositioned into the same Schlemm's canal location after finishing canaloplasty. The polymeric microcatheter is re-advanced into Schlemm's canal up to 180 degrees (one hemisphere) by rotating the advancement wheel forward until the wheel stops (about 20 mm). With the microcatheter resting in the canal, the cannula is removed from the corneal incision and out of the eye causing the microcatheter to cut through the trabecular meshwork. This process can be repeated in the second Schlemm's hemisphere.

Comparison of Technological Characteristics with the Predicate

The primary predicate in this 510(k) submission is the OMNI Surgical System described in K173332 and K201953, based on substantial equivalence to the Ellex iScience Interventional Canaloplasty Microcatheter ("iTrack Surgical System") cleared in K080067. The same iTrack Surgical System device (K080067) is also used as a reference device in this 510(k) submission.

The Sight Sciences OMNI Surgical System subject device complies with the same product design requirements and applicable standards as the predicate OMNI Surgical System and shares the identical principle of operation, intended use, and key technological characteristics.

Additionally, the technical features of the subject OMNI Surgical System are similar, but not identical to the commercially available Ellex iTrack Surgical System reference device. The different technological characteristics do not raise different questions of safety and effectiveness. The subject OMNI Surgical System with modified indications and the Ellex iTrack device are both manually operated devices that utilize a microcatheter to access Schlemm's canal and deliver viscoelastic fluid. Fluid is dispensed from each system on the principle of exchanging volumes much like a syringe. The iTrack device serves as a reference device to support the scientific methods to assess the safety and effectiveness of the device to lower IOP in glaucoma patients. The iTrack device is used in the same anatomical location as the OMNI device and is intended for use to perform canaloplasty to lower IOP in glaucoma patients, which is similar to the intended use proposed for the OMNI device.

A comparison of the attributes of the subject OMNI Surgical System with the predicate OMNI Surgical System and the Ellex iTrack Surgical System (iScience Interventional Canaloplasty Microcatheter) reference device are listed in Table 1 below.

Table 1. Technological Characteristics Comparison

{6}------------------------------------------------

CharacteristicOMNI Surgical SystemOMNI Surgical System(K173332)Ellex iTrack Surgical System(iScience InterventionalCanaloplasty MicrocatheterK080067)REFERENCE DEVICE
Intended UseSUBJECT DEVICEOphthalmic surgical tool fordelivery of controlled amountsof viscoelastic fluid into theanterior segment and used tocut trabecular meshworkwhen a trabeculotomy isindicatedPRIMARY PREDICATE DEVICEOphthalmic surgical tool fordelivery of controlled amountsof viscoelastic fluid into theanterior segment and used tocut trabecular meshworkwhen a trabeculotomy isindicatedDelivery of controlled amountsof viscoelastic fluid duringophthalmic surgery
Indications forUseThe OMNI® Surgical System isindicated for canaloplasty(micro-catheterization andtransluminal viscodilation ofSchlemm's canal) followed bytrabeculotomy (cutting oftrabecular meshwork) toreduce intraocular pressure inadult patients with open-angleglaucomaThe OMNI® Surgical System isa manually operated devicefor delivery of small amountsof viscoelastic fluid, forexample Healon® or HealonGV® from Abbott MedicalOptics, Amvisc® from Bausch& Lomb, or PROVISC® fromAlcon, during ophthalmicsurgery. It is also indicated tocut trabecular meshworktissue during trabeculotomyproceduresThe iScience InterventionalCanaloplasty Microcatheter isindicated for fluid infusion andaspiration during surgery. TheiScience InterventionalCanaloplasty Microcatheter isindicated for catheterizationand viscodilation of Schlemm'scanal to reduce intraocularpressure in adult patients withopen angle glaucoma
Regulation880.5725 (Infusion Pump)880.5725 (Infusion Pump)886.4350 (Manual ophthalmicsurgical instrument)876.1500 (Endoscope andaccessories)
Device ClassClass IIClass IIClass II
Product CodePrimary: MRH (OphthalmicInfusion Pump)Secondary: HMZ(Trabeculotome)Primary: MRH (OphthalmicInfusion Pump)Secondary: HMZ(Trabeculotome)Primary: MPA (Endoscope)Secondary: HMX (ManualOphthalmic SurgicalInstrument)
PrescriptionStatusPrescription use onlyPrescription use onlyPrescription use only
TargetAnatomySchlemm's Canal andTrabecular MeshworkSchlemm's Canal andTrabecular MeshworkSchlemm's Canal andTrabecular Meshwork
OperatingPrincipleManualManual• Manual (microcatheter)• Powered (endoilluminator)
Design/Mechanism ofAction• Stainless-steel cannula hassharp tip that can be used topierce the trabecularmeshwork and provide accessinto Schlemm's canal. Minordimensional changes tocannula tip height and radius• Flexible microcatheter withrounded, atraumatic tip fordispensing of viscoelastic• Microcatheter is blue color tofacilitate its visibility inSchlemm's canal as it isadvanced/retracted throughthe cannula• Microcatheter allows accessto 360° of Schlemm's canal intwo 180° segments• Proximal handle changed toovoid shape with elastomericmaterial for added grip• Proximal handle allowsambidextrous use in eitherpatient eye• Internal viscoelastic reservoirand plunger tube withdimensional changes to allowdispensing of viscoelastic• Ovoid handle shape allowsadvancement wheels (fingerwheels) to be reduced to twofor advancing and retractingmicrocatheter up to 20mmusing a rack and pinionmechanism• Tactile and audible clicksindicate precise advancement• Viscoelastic dispensed duringretraction of first two cyclesafter priming with viscoelasticfluid• Flexible microcatheterintroduced into Schlemm'scanal and pulled through tocut trabecular meshwork• Priming Lock moved toaccommodate new Luerfitting and prevents accidentaldispensing during viscoelasticpriming• Changes were made to the• Stainless-steel cannula hassharp tip that can be used topierce the trabecularmeshwork and provide accessinto Schlemm's canal• Flexible microcatheter withrounded, atraumatic tip fordispensing of viscoelastic• Microcatheter is blue color tofacilitate its visibility inSchlemm's canal as it isadvanced/retracted throughthe cannula• Microcatheter allows accessto 360° of Schlemm's canal intwo 180° segments• Proximal handle hasrectangular shape• Proximal handle allowsambidextrous use in eitherpatient eye• Internal viscoelastic reservoirand plunger tube to allowdispensing of viscoelastic• Four advancement wheels(finger wheels) for advancingand retracting microcatheterup to 20mm using a rack andpinion mechanism• Tactile and audible clicksindicate precise advancement• Viscoelastic dispensed duringretraction of first two cyclesafter priming with viscoelasticfluid• Flexible microcatheterintroduced into Schlemm'scanal and pulled through tocut trabecular meshwork• Priming Lock Pin preventsaccidental dispensing duringviscoelastic priming• Flexible microcatheter fordispensation of viscoelasticfluid• Microcatheter has a round,bolus, atraumatic tip• Microcatheter allows accessto 360° of Schlemm's canal inone pass• Microcatheter has internalcatheter support wire• Tactile and audible clicksindicate precise advancement• Viscoelastic manuallydispensed during retractionmicrocatheter• No cannula – a 27 ga needle(not included) is used topierce the trabecularmeshwork• Centrally located hub• Microcatheter attaches atproximal end to EllexViscoInjector• Viscoelastic cartridge placedinto ViscoInjector• Microcatheter manuallyadvanced/retracted bysurgeon using microsurgicalforceps (not included)• Knob on ViscoInjector turnedto dispense viscoelasticduring microcatheterretraction• Microcatheter is clear andtransparent but incorporatesa red light at its tip tofacilitate visibility inSchlemm's canal
CharacteristicOMNI Surgical SystemSUBJECT DEVICEOMNI Surgical System(K173332)PRIMARY PREDICATE DEVICEEllex iTrack Surgical System(iScience InterventionalCanaloplasty MicrocatheterK080067)REFERENCE DEVICE
reservoir, Luer fitting and anew bonding adhesive wasused
DispensingControlAfter priming, viscoelasticdispensing control occursthrough manual rotation ofeither advancement wheel atthe distal end of the device.Synchronization of the twowheels was reversed for easeof useAfter priming, viscoelasticdispensing control occursthrough manual rotation ofeither of the advancementwheels at the distal end of thedeviceRotational action via a knob onthe ViscoInjector to dispenseviscoelastic fluid
DispensingMechanismInternal reservoir with plungertube (syringe-like volumeexchange). Three internalcomponent dimensionsmodified to reduce air bubbleformation during primingInternal reservoir with plungertube (syringe-like volumeexchange).OVD cartridge attaches to thedevice inside the EllexViscoInjector (syringe-likevolume exchange)
ViscoelasticFluid (OVD)and PrimingMethodCohesive viscoelastic fluid(OVD or ophthalmicviscosurgical device) is suppliedseparately. Viscoelastic loadedinto device (primed) prior touse by attaching OVD cartridgedirectly to a Luer fitting thatreplaces the cap on proximalend of OMNI device handle forease of primingCohesive viscoelastic fluid(OVD or ophthalmicviscosurgical device) is suppliedseparately. Viscoelastic loadedinto device (primed) prior touse by attaching OVD cartridgeto a supplied Nozzle that isinserted into a on proximal endof OMNI device handleCohesive viscoelastic fluid(OVD or ophthalmicviscosurgical device) is suppliedseparately. OVD cartridgeattaches to the device insidethe Ellex ViscoInjector
OVD VolumeDispensed11 μL9 μLNot stated
MaterialsMedical grade materials,including ABS, polycarbonate,stainless steel, silicone,parylene coating,cyanoacrylate, acrylatedurethane, polyimideMedical grade materials,including ABS, polycarbonate,stainless steel, silicone,parylene coating,cyanoacrylate, acrylatedurethane, polyimideMicrocatheter - polyimidetubing and an outer sheath ofpolyethylene terephthalate(PET) shrink tubing, with alubricious coating
User InterfaceHandheldHandheldHandheld
MicrocatheterShaft OuterDiameter200 microns200 microns200 microns
MicrocatheterTip OuterDiameter Range0.0090 to 0.0110 inches0.0095 to 0.0110 inches0.0098 inches
Sterile andSingle UseProvided sterile. Single patientuseProvided sterile. Single patientuseProvided sterile. Singlepatient use
CharacteristicOMNI Surgical SystemOMNI Surgical System(K173332)Ellex iTrack Surgical System(iScience InterventionalCanaloplasty MicrocatheterK080067)
SUBJECT DEVICEPRIMARY PREDICATE DEVICEREFERENCE DEVICE
SterilizationMethodGamma radiationGamma radiationGamma radiation
SterilityAssurance Level10-610-6
PackagingThermoformed plastic traywith heat-sealed Tyvek lidTyvek pouch with a polymertray card
Shelf Life13 Months6 Months

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Risk Analysis

The risk management process at Sight Sciences complies with EN ISO 14971:2012 "Medical devices -- Application of risk management to medical devices." As required by this standard, risk analyses are conducted according to defined procedures, using experienced, qualified personnel from multiple functions throughout the organization with prior experience in risk assessment.

Risk analysis activities were first documented in a Risk Management Plan. This plan specified the risk management activities to be conducted based on labeling changes to the cleared OMNI Surgical System. The following activities were completed:

  • The Clinical Hazards for the OMNI Surgical System were assessed by qualified personnel with respect to the proposed changes in indications for use. The result of this analysis indicated that no new hazards were identified.
  • . In accordance with the Plan, all Failure Modes Effect Analyses (FMEA) were assessed by the qualified team of personnel. This included a review of the design FMEA, process FMEA and Use FMEA. In order to evaluate the hazards related to the proposed changes to indications for use, it was determined that design control activities, including bench performance testing and clinical performance testing, were required to verify that potential harm and effects resulting from the previously identified hazards did not result in any new failure modes, causes or effects and did not change the severity and occurrence rankings.
  • Design control verification activities resulting from assessment of the FMEA's included nonclinical bench performance testing according to a written protocol with appropriate sample sizes and pre-determined acceptance criteria. Clinical performance testing was performed according to a written protocol.

Based on the risk management documentation and supported by successful non-clinical bench performance testing and clinical performance testing, all identified clinical hazards remained mitigated to an acceptable level of residual risk. The review of the design, process and use FMEA's determined that no changes to the risk management file were required in order to

{10}------------------------------------------------

implement the changes to the indications for use. Taking into account the modified indications of the OMNI Surgical System, the potential benefits to patients outweigh the low residual risk.

Performance Data

A. Bench Testing

Design verification was already established and did not need to be repeated. Based on the risk analysis, bench simulated use testing and clinical performance testing were determined to be needed. Simulated use testing in human cadaver eyes was performed using the OMNI Surgical System. The results of this nonclinical testing demonstrate that the OMNI Surgical System meets the defined specifications and functioned as intended.

Simulated Use - Cadaver Eye Testing

The changes to the indications for use were supported by simulated use testing in human cadaver eyes. Cadaver eyes were used to validate the OMNI Surgical System's ability to perform its indicated use, including viscodilation of Schlemm's canal followed by trabeculotomy. The study was performed by an ophthalmologist along with a physician assistant. A total of eight OMNI Surgical System devices were tested in four cadaver eyes. Following the Instructions For Use (with the modified indications) the surgeon was able to consistently access, microcatheterize, and viscodilate Schlemm's canal ("canaloplasty") and re-access, microcatheterize, and cut trabecular meshwork ("trabeculotomy") in human cadaveric tissue.

Conclusions of Bench Performance Testing

The human eye performance testing results demonstrate the ability of the surgeon's assistant to prepare the OMNI Surgical System device, and the ophthalmic surgeon to (a) access Schlemm's canal, (b) navigate Schlemm's canal with the microcatheter, (c) deliver viscoelastic fluid into Schlemm's canal upon retraction of the microcatheter from the canal, and (d) regain access to Schlemm's canal and perform a trabeculotomy.

B. Clinical Evidence Supporting Substantial Equivalence

A retrospective, observational, multi-center, single-arm, consecutive case series study ("ROMEO") was conducted at 11 sites throughout the U.S. where surgeons used the OMNI Surgical System for the microcatheterization and viscodilation of Schlemm's canal ("canaloplasty") followed by trabeculotomy as a standalone procedure or in combination with cataract extraction. Results from this study were compared with a literature control.1 In this study, a prior iteration of the OMNI device was used. Minor system technological differences between the device used in this study (the predicate device K173332) and the

4 Lewis RA, von Wolff K, Tetz M, Kearney JR, Shingleton B, Samuelson TW. Canaloplasty: circumferential viscodilation and tensioning of Schlemm's canal using a flexible microcatheter for the treatment of open-angle glaucoma in adults: interim clinical study analysis. J Cataract Refract Surg. 2007 Jul;33(7):1217-26.

{11}------------------------------------------------

subject device are described in Table 1. Each site had collected data on intraocular pressure (IOP), the use of ocular hypotensive medications, and safety outcomes in adult patients with open angle glaucoma. Endpoint stratification was performed by type of procedure; combined with cataract extraction, or stand-alone. A total of 129 patients with a single qualifying eye were treated with the OMNI Surgical System and followed post-operatively at months 1, 6 and 12. There were 81 (63%) procedures combined with cataract surgery, and 48 (37%) standalone procedures.

Effectiveness Results

Descriptive statistics were assessed over time for the subgroup of patients who met the entry criteria of the literature control (i.e. a baseline IOP ≥ 16 mmHg).

Table 2. OMNI +Cataract and Standalone Results (Subgroup using the Lewis 2007eligibility, baseline IOP ≥ 16 mmHg)
+CataractStandalone
VisitMean IOP ± SDRangenMean IOP ± SDRangen
Baseline19.5 ± 3.816.0-33.04520.0 ± 3.616.0-31.038
1 month15.7 ± 4.06.7-25.04415.3 ± 4.48.0-27.036
6 month15.1 ± 2.910.0-22.04015.6 ± 3.09.0-20.037
12 month15.2 ± 3.010.0-22.04215.3 ± 2.77.0-19.536
Table 3. Number of Medications Required Over Time (ROMEO – All patients)
VisitMean ± SDRangen
Baseline1.8 ± 1.30-4129
1 month1.3 ± 1.30-5125
6 month1.1 ± 1.20-5115
12 month1.1 ± 1.20-5120

A post-hoc responder analysis was also performed to assess the proportion of patients who experienced a ≥ 20% reduction in IOP at Month 12, no increase in medication, no secondary surgery. Refer to Table 4 for results in patients who had OMNI performed as a Standalone procedure and Table 5 for results in patients who had OMNI performed in conjunction with cataract surgery.

Table 4. Proportion of Standalone Subjects with ≥ 20% Reduction in IOP at Month 12, noincrease in medication, no secondary surgery
GroupProportion

{12}------------------------------------------------

Pre-op IOP > 18 mmHg14/2458.3%
Pre-op IOP ≤ 18 mmHg4/2416.7%*
All Standalone18/4837.5%*
All meeting Lewis criteria (Pre-Op IOP ≥16 mmHg)16/3545.7%*
  • 20% responder analysis is not appropriate for patients with baseline IOP ≤ 18 mmHg because baseline IOP was controlled.

Table 5. Proportion of +Cataract Subjects with ≥ 20% Reduction in IOP at Month 12, no increase in medication, no secondary surgery

GroupnProportion
BL > 18 mmHg15/2462.5%
BL ≤ 18 mmHg10/5717.5%*
All Combined with cataract25/8130.9%*
All meeting Lewis criteria (BL ≥16 mmHg)20/4643.5%*
  • 20% responder analysis is not appropriate for patients with baseline IOP ≤ 18 mmHg because baseline IOP was controlled.

Safety Results

All 129 patients treated with the OMNI Surgical system were included in the safety analyses. Adverse events were generally infrequent, mild, non-serious, transient in nature, resolved with or without treatment and were consistent with those expected in the target population. The most common adverse events reported prior to and including the 12month visit were: posterior capsular opacity, mild anterior chamber inflammation, secondary surgical intervention for IOP control, cystoid macular edema, IOP spike, and hyphema, as listed in Table 6. There were no serious adverse events or serious device-related adverse events reported in the study.

Adverse EventStandalone(n=48)n (%)+Cataract (n=81)n (%)ALL AdverseEvents (n=129)n (%)
Posterior capsule opacity5 (10.4)14 (17.3)19 (14.7)
Mild anterior chamber inflammation6 (12.5)8 (9.9)14 (10.9)
Cystoid macular edema3 (6.3)4 (4.9)7 (5.4)
Corneal edema2 (4.2)14 (4.9)26 (4.7)
IOP increase ≥ 10 mmHg above baseline>30 days postoperative3 (6.3)3 (3.7)6 (4.7)
Hyphema > 1 mm2 (4.2)3 (3.7)5 (3.9)

Table 6. Adverse Events, by Procedure Sub-Group

{13}------------------------------------------------

Adverse EventStandalone(n=48)n (%)+Cataract (n=81)n (%)ALL AdverseEvents (n=129)n (%)
Worsening of visual field mean deviation≥ 2 dB3 (6.3)1 (1.2)4 (3.1)
BCVA loss of ≥ 2 lines Snellen at or after 3months post-op2 (4.2)1 (1.2)3 (2.3)
Cataract surgery complication1 (2.1)31 (1.2)42 (1.6)
Choroidal effusion01 (1.2)1 (0.8)
Macular degeneration (dry)1 (2.1)01 (0.8)
Epiretinal membrane peel1 (2.1)01 (0.8)
Ocular allergic reaction01 (1.2)1 (0.8)
Posterior vitreous detachment01 (1.2)1 (0.8)
Vitreous hemorrhage1 (2.1)501 (0.8)
Cyclodialysis01 (1.2)1 (0.8)
Lid edema1 (2.1)01 (0.8)
Late hypotony01 (1.2)61 (0.8)
Loss of light perception000
Chronic anterior iritis as defined in theFDA MIGS guidance000
TOTAL314374

1 One subject developed corneal edema at 4-6 months post-procedure which was noted to have resolved one year later.

  • 2 Pre-existing Fuch's dystrophy in one subject which worsened required a DSAEK.
  • 3 AE was an IOL dislocation from prior cataract surgery.
  • 4 AE was lens fragment.
  • 5 Vitreous hemorrhage verbatim description was "peripheral retinal hemorrhage" and resolved without treatment in 32 days by the Investigator.
  • 6 One subject underwent post-surgical anterior chamber reformation with Healon viscoelastic fluid for a shallow chamber with failed laser cycloplexy for a related cyclodialysis cleft; multiple paracentesis was performed to remove viscoelastic causing an IOP spike and subsequent surgical cycloplexy to repair the cleft. Small choroidal effusions resolved with closure of the cleft.

Secondary Surgical Interventions

There were 9/129 eyes (7.0%) required a secondary surgical intervention to reduce IOP in the medical judgment of the Investigator. There were 4/81 secondary surgical interventions in the +cataract group (4.9%) and 5/48 in the standalone group (10.4%). The reinterventions were SLT (n=3/9, 33%), glaucoma drainage device (tube or valve) (n=3/9, 33%), trabeculectomy including Express device (n=2/9, 22%), and paracentesis (n=1/9, 11%).

In addition to secondary surgical intervention for IOP, two subjects underwent additional interventions in the follow-up period, one in the standalone OMNI group (n=1/48, 2.1%)

{14}------------------------------------------------

and one in combined Omni with cataract extraction group (n=1/81, 1.2%). The subject in the standalone OMNI group had pre-existing Fuch's dystrophy and experienced a worsening requiring a Descemet Stripping Automated Endothelial Keratoplasty (DSAEK) procedure which resolved the corneal edema. The second subject who underwent a combined cataract extraction and OMNI procedure was noted to have a shallow, but not a flat chamber. Small choroidal effusions were noted as well as a cyclodialysis cleft which was not notable on exam due to the shallow chamber. Chamber reformation using Healon was performed , partially for therapeutic reasons but mostly to facilitate view of the angle. This allowed gonioscopic visualization of the angle which revealed the location of the cleft and laser cycloplexy was attempted to close the cleft. An IOP spike ≥ 10 mmHg above baseline was secondary to the Healon injection and paracentesis were performed to remove viscoelastic material from the eye. When laser cycloplexy failed to close the cleft, the surgeon performed a surgical cycloplexy. Small choroidal effusions resolved with closure of the cleft.

Clinical Study Conclusions

The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is substantially equivalent to the legally marketed OMNI Surgical System.

The clinical performance data collected in the ROMEO study for the OMNI Surgical System suggest clinically significant IOP reduction in both the cataract and standalone arms for up to 12 months for the intended population of adult patients with primary open-angle glaucoma.

Conclusions

The Sight Sciences OMNI Surgical System subject device complies with the same product design requirements and applicable standards as the predicate OMNI Surgical System and shares the identical principle of operation, intended use, and key technological characteristics. The indications for use differ from that of the predicate OMNI Surgical System, however, the differences do not alter the intended use and are supported by bench simulated use and clinical performance data that demonstrate the ability of the OMNI Surgical System to access Schlemm's canal, navigate the canal with the microcatheter, perform canaloplasty by delivering commercially available viscoelastic fluid upon retraction of the microcatheter from the canal, and regain access to Schlemm's canal with the microcatheter to perform a trabeculotomy.

The conclusions drawn from the nonclinical tests demonstrate that the device is substantially equivalent to the legally marketed OMNI Surgical System.

The clinical performance data collected in the ROMEO study for the OMNI Surgical System suggest clinically significant IOP reduction in both the cataract and standalone arms for up to 12 months for the intended population of adult patients with primary open-angle glaucoma.

The OMNI Surgical System is substantially equivalent to the legally-marketed OMNI Surgical System predicate device cleared under 510(k) K173332.

{15}------------------------------------------------

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).