K Number
K202627

Validate with FDA (Live)

Date Cleared
2021-02-03

(145 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Device Description

The Medical Face Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Medical Face Masks are manufactured with three layers. The outer layer is made of spun-bonded polypropylene (PP) non-woven fabric. The middle layer with filtration function is made of melt blown polypropylene (PP) non-woven fabric. The inner layer contact with face is made of spun-bonded polypropylene (PP) non-woven fabric. The Medical Face Masks are single use, disposable device, provided non-sterile.

AI/ML Overview

The provided text describes the acceptance criteria and the study conducted for "Medical Face Masks" (K202627).

Here's the extracted information:

  1. Table of Acceptance Criteria and Reported Device Performance:

    Methodology/StandardPurposeAcceptance CriteriaReported Device Performance
    ASTM F1862M-17Fluid Resistance Performance29 out of 32 pass at 120mmHg29 out of 32 pass at 120mmHg
    ASTM F2299Particulate Filtration Efficiency≥98%Average 99.82% at 0.1μm
    ASTM F2101-19Bacterial Filtration Efficiency≥98%Average 99.78%
    EN 14683:2019 Annex CDifferential Pressure<6mmH2O/cm²Average 3.3 mmH2O/cm²
    16 CFR 1610FlammabilityClass I Non FlammableMeet Class I
    ISO 10993-5In Vitro CytotoxicityDevice is non-cytotoxicDevice is non-cytotoxic
    ISO 10993-10Skin IrritationDevice is non-irritatingDevice is non-irritating
    ISO 10993-10Skin SensitizationDevice is non-sensitizingDevice is non-sensitizing
  2. Sample size used for the test set and the data provenance:

    • Fluid Resistance Performance (ASTM F1862M-17): The test involved 32 samples (29 out of 32 passed).
    • For other performance tests (Particulate Filtration Efficiency, Bacterial Filtration Efficiency, Differential Pressure, Flammability, Biocompatibility), specific sample sizes are not explicitly stated in the provided text.
    • Data Provenance: Not specified, but the device manufacturer is based in China (Guilin HBM Sanitary Protections, Inc). The testing appears to be non-clinical laboratory testing.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. These are non-clinical performance and material characterization tests, not reliant on expert clinical interpretation or ground truth establishment in the traditional sense of medical image analysis or diagnostic studies.
  4. Adjudication method for the test set:

    • Not applicable. The tests are standardized laboratory procedures with objective measurements.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is for a Medical Face Mask, not an AI diagnostic device. No MRMC study was conducted.
  6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is for a Medical Face Mask. The tests performed are standalone performance tests for the physical product.
  7. The type of ground truth used:

    • The "ground truth" for these tests refers to established scientific and regulatory standards (e.g., ASTM, EN, 16 CFR, ISO standards) against which the device's physical properties and performance characteristics are measured. For example, for "Bacterial Filtration Efficiency," the ground truth is the specified bacterial filtration efficiency percentage (≥98%).
  8. The sample size for the training set:

    • Not applicable. This is a medical device (face mask) and not an AI/ML algorithm that requires a training set.
  9. How the ground truth for the training set was established:

    • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 3, 2021

Guilin HBM Sanitary Protections, Inc % Shelley Li Director Shanghai Landlink Medical Information Technology Co., Ltd. Room 703, 705, Baohua International Plaza, West Guangzhong Road 555, Jingan Shanghai, 200071 China

Re: K202627

Trade/Device Name: Medical Face Masks Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 9, 2020 Received: November 13, 2020

Dear Shelley Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202627

Device Name Medical Face Masks

Indications for Use (Describe)

The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

Type of Use (Select one or both, as applicable)

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary: K202627

I. Submitter

Guilin HBM Sanitary Protections, Inc No.1-2, Shuijing East Road, Economic and Technological Development Area, Guilin 541805, China Contact person: Liu Xin Position: General Manager Assistant Tel.: +86-13761713489 E-mail: 724095871@qq.com

Preparation date: Jan.30, 2021

Submission Correspondent

Ms. Shelley Li Tel.: +86-021-80317637 Shanghai Landlink Medical Information Technology Co., Ltd. E-mail: shelley.li@landlink-healthcare.com

II. Proposed Device

Trade Name of Device:Medical Face Masks
Common name:Surgical Mask
Regulation Number:21 CFR 878.4040
Regulatory Class:Class II
Product code:FXX
Review PanelGeneral Hospital

III. Predicate Devices

510(k) Number:K153496
Trade name:Disposable Surgical Face Mask
Common name:Surgical Mask
Classification:Class II
Product Code:FXX
ManufacturerXiantao Rayxin Medical Product Co., Ltd

IV. Device description

{4}------------------------------------------------

The Medical Face Masks are Flat Pleated type mask, utilizing Ear Loops way for wearing, and they all have Nose Piece design for fitting the face mask around the nose. The Medical Face Masks are manufactured with three layers. The outer layer is made of spun-bonded polypropylene (PP) non-woven fabric. The middle layer with filtration function is made of melt blown polypropylene (PP) non-woven fabric. The inner layer contact with face is made of spun-bonded polypropylene (PP) non-woven fabric.

The Medical Face Masks are single use, disposable device, provided non-sterile.

V. Indication for use

The Medical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.

VI. Comparison of technological characteristics with the predicate devices

ItemProposed device(K202627)Predicate device(K153496)Comparison
Product nameMedial Face MasksDisposable SurgicalFace MaskSimilar
Product CodeFXXFXXSame
Regulation No.21 CFR 878.404021 CFR 878.4040Same
ClassClass IIClass IISame
Mask styleFlat-pleated, ear loop,3 layersFlat-pleated, ear loop,tie-on, 3 LayersSimilar
Indication for useThe Medical FaceMasks are intendedto be worn to protectboth the patient andhealthcare personnelfrom transfer ofmicroorganisms,body fluids andparticulate material.The DisposableSurgical Face Masksare intended to beworn to protect boththe patient andhealthcare personnelfrom transfer ofmicroorganisms,body fluids andSame
These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device(s), provided non-sterile.
MaterialsInner layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Middle layerMelt blown polypropylene filterMelt blown polypropylene filterSame
Outer layerSpun-bond polypropyleneSpun-bond polypropyleneSame
Ear loop85% Spandex + 15% PolyesterPolyesterDifferent1
Nose pieceIron wire covered polypropyleneMalleable aluminum wireDifferent1
ColorBlueBlueSame
Length17.5cm17.5±1cmSame
Width9.5cm9.5±1cmSame
OTC useYesYesSame
SterileNon-sterileNon-sterileSame
Single for useYesYesSame
BiocompatibilityConfirm to the requirements of ISO 10993 series standardsConfirm to the requirements of ISO 10993 series standardsSame
Fluid Resistance PerformanceASTM E1862Pass at 120mmHgpass at 120mmHgSimilar
ParticulateFiltration EfficiencyASTM F2299Average 99.82% at0.1μmAverage 99.74% at0.1μm
Bacterial FiltrationEfficiencyASTM F2101Average 99.78%Average 99.4%
DifferentialPressureAverage 3.3mmH2O/cm²Average 2.7mmH2O/cm²
Flammability16 CFR 1610Class INon FlammableClass 1Non Flammable
In VitroCytotoxicityISO 10993-5Under the conditionsof this study thedevice isnon-cytotoxicUnder the conditionsof this study thedevice isnon-cytotoxicSame
Skin IrritationISO 10993-10Under the conditionsof this study thedevice is non-irritatingUnder the conditionsof this study thedevice is non-irritatingSame
Skin SensitizationISO 10993-10Under the conditionsof this study thedevice isnon-sensitizingUnder the conditionsof this study thedevice isnon-sensitizingSame

Table 1 Comparison of characteristics

{5}------------------------------------------------

{6}------------------------------------------------

1 The difference in the materials does not raise additional questions for safety and effectiveness of the device. The biocompatibility evaluation test of the subject devices have been performed on the final finished device which includes all construction materials and color additives. The test results shows pass the requirements.

VII. Summary of Non-Clinical Testing

Non-clinical performance tests were conducted to verify that the proposed device met all design specifications. The below table shows the test results of test article, which demonstrated that the proposed device complies with the related standards:

Methodology/StandardPurposeAcceptanceCriteriaResults
ASTMF1862M-17Fluid ResistancePerformance29 out of 32 pass at120mmHg29 out of 32pass at

{7}------------------------------------------------

120mmHg
ASTM F2299Particulate Filtration Efficiency≥98%Average99.82% at0.1μm
ASTMF2101-19Bacterial Filtration Efficiency≥98%Average99.78%
EN 14683:2019Annex CDifferential Pressure<6mmH2O/cm²Average 3.3mmH2O/cm²
16 CFR 1610FlammabilityClass INon FlammableMeet Class I

VIII. Clinical Testing

No clinical study is included in this submission.

IX. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe as effective, and performs as well as or better than the legally marketed predicate deviceK153496.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.