K Number
K202574

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-02-18

(167 days)

Product Code
Regulation Number
866.3830
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LIAISON® Lyme IgG assay uses chemiluminescent immunoassay (CLIA) technology for the qualitative detection of IgG antibodies to Borrelia burgdorferi in human serum and plasma specimens (K2-EDTA, Li-heparin). This assay is intended for use on samples from patients with signs and symptoms consistent with or patients suspected of having Lyme disease to assess the presence of IgG antibodies and exposure to Borrelia burgdorferi. In addition, the LIAISON® Lyme Ig G assay may be used as a confirmatory test in the modified two-tier test (MTTT) in combination with the DiaSorin LIAISON® Lyme Total Antibody Plus assay.

If used as a first stage test, positive or equivocal results with the LIAISON® Lyme IgG assay should be confirmed through additional testing with a Standard two-tier test (STT) methodology using an IgG Borrelia burgdorferi Western blot test following current guidelines.

Positive supplemental results are supportive evidence of the presence of antibodies and exposure to Borrelia burgdorferi and may be used along with patient history, symptoms and other laboratory data to support a clinical diagnosis of Lyme disease.

Negative results by the LIAISON® Lyme IgG assay should not be used to exclude Lyme disease.

The test must be performed on the LIAISON® XL Analyzer.

The DiaSorin LIAISON® Lyme IgG Control Set is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Lyme IgG assay. The performance characteristics of LIAISON® Lyme IgG controls have not been established for any other assays or instrument platforms different from the LIAISON® XL.

Device Description

The LIAISON® Lyme IgG assay is an indirect chemilyminescence immunoassay (CLIA) for the qualitative detection of IgG antibodies to Borrelia burgdorferi in human serum and plasma samples. All assay steps (with the exception of magnetic particle resuspension) and incubations are performed by the Analyzer. The principal components of the test are magnetic particles (solid phase) coated with recombinant Borrelia antigens and a conjugate reagent containing an anti-human IgG mouse monoclonal antibody linked to an isoluminol derivative (isoluminol-antibody coniugate). During the first incubation, anti-Borrelia burgdorferi antibodies present in calibrators, samples or controls bind to the solid phase. Unbound material is then removed with a wash cycle. During the second incubation, the antibody conjugate reacts with anti-Borrelia burgdorferi IgG antibodies that have bound to the solid phase. Excess antibody conjugate is then removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is induced. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of the presence of Borrelia burgdorferi IgG antibodies present in calibrators, samples or controls.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the LIAISON® Lyme IgG assay, based on the provided FDA 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly defines acceptance criteria through its performance study results, specifically for percentage agreement (sensitivity and specificity) when compared to a predicate device and STTT/MTTT protocols.

Acceptance Criteria (Implicitly defined by performance targets in studies)

Performance MetricAcceptance Criteria (from study context)Reported Device Performance (LIAISON® Lyme IgG)Study Context
First Tier Agreement with Predicate Device (ZEUS ELISA B. burgdorferi IgG Test System)
Positive % AgreementImplied high agreement (no specific % mentioned but demonstrated acceptability)55.1% (166/301)Comparison to Predicate ELISA IgG
Negative % AgreementImplied high agreement (no specific % mentioned but demonstrated acceptability)96.5% (2210/2320)Comparison to Predicate ELISA IgG
Standard Two-Tier Testing Methodology (STTT - WB IgG)
2nd Tier PPA (Positive Percent Agreement)Implied high PPA89.0% (97/109)Comparison to STTT WB IgG
2nd Tier NPA (Negative Percent Agreement)Implied high NPA99.5% (2500/2512)Comparison to STTT WB IgG
Modified Two-Tier Testing Methodology (MTTT - Prospective Population - STTT WB IgG as GT)
PPA (Positive Percent Agreement)Implied high PPA96.9% (93/96)Comparison to STTT WB IgG
NPA (Negative Percent Agreement)Implied high NPA30.2% (39/129)Comparison to STTT WB IgG
Modified Two-Tier Testing Methodology (MTTT - Retrospective CDC Panel - STTT WB IgG as GT)
Sensitivity (Stage I)Implied high Sensitivity80%Comparison to STTT WB IgG
Sensitivity (Stage II)Implied high Sensitivity90%Comparison to STTT WB IgG
Sensitivity (Stage III)Implied high Sensitivity100%Comparison to STTT WB IgG
Specificity (Healthy Controls)Implied high Specificity100%Comparison to STTT WB IgG
Specificity (Disease Controls)Implied high Specificity97.8%Comparison to STTT WB IgG

Note on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria for each metric (e.g., "PPA must be >X%"). Instead, it presents the device's performance results and concludes substantial equivalence, implying that these results met the FDA's expectations for equivalence to legally marketed predicate devices.

2. Sample Size Used for the Test Set and Data Provenance:

  • Method Comparison (First Tier and STTT):
    • Sample Size: 2621 human serum specimens.
    • Data Provenance: Collected in 14 states across five (5) distinct U.S. geographical regions. This was a prospective study ("All 2621 prospective (all comer) specimens").
  • Modified Two Tier Testing Methodology (Prospective Population):
    • Sample Size: 225 specimens (from the initial 2621 that were positive or equivocal by the first-tier assay).
    • Data Provenance: U.S. geographical regions, prospective.
  • Characterized Lyme Panel (CDC Panel):
    • Sample Size: 280 samples.
    • Data Provenance: Acquired from the CDC (Centers for Disease Control), making it a retrospective panel.
  • Modified Two Tier Testing Methodology (Retrospective Population from CDC Panel):
    • Sample Size: 279 samples (1 sample from the 280 CDC panel lacked sufficient volume). 82 of these were further tested with STTT WB IgG or MTTT.
    • Data Provenance: CDC, retrospective.
  • Cross-reactivity Study:
    • Sample Size: 222 specimens (from 22 different disease states).
    • Data Provenance: Not specified, but likely obtained from various sources or commercial vendors for specific disease states. Retrospective by nature of obtaining specific disease samples.
  • Matrix Equivalence Study:
    • Sample Size: 32 matched patient sets.
    • Data Provenance: Not specified.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

The document does not explicitly state the "number of experts" or their specific "qualifications" for establishing ground truth. However, the ground truth for some studies is based on established laboratory methods:

  • For the Method Comparison and STTT/MTTT studies: The ground truth for confirming Borrelia burgdorferi infection or exposure is primarily established by Western blot testing (IgG Borrelia burgdorferi Western blot test) following current guidelines. This is a recognized laboratory "gold standard" or highly accepted confirmatory method for Lyme disease. While experts interpret WBs, the document doesn't detail the number or qualifications of individuals performing or interpreting these specific Western blots.
  • For the Characterized Lyme Panel (CDC Panel): The samples were "characterized" by the CDC. This implies that the CDC itself, a leading public health agency with extensive expertise in infectious diseases, established the reference classification for these samples. The ground truth here is the CDC's reference classification, which would involve robust testing and expert consensus within the CDC, though specific expert numbers or qualifications are not provided in this submission summary.

4. Adjudication Method for the Test Set:

Not explicitly described in the document. The studies primarily compare the LIAISON® Lyme IgG assay results against either a predicate ELISA assay or Western Blot results (STTT/MTTT). There isn't mention of a separate adjudication process beyond the defined "ground truth" method (e.g., WB results).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study, as typically understood for AI-assisted image analysis where human readers improve with AI vs without AI assistance, was not explicitly mentioned or performed. This device is an in-vitro diagnostic (IVD) assay designed to detect antibodies, not an AI-powered diagnostic imaging system requiring human interpretation with AI assistance. Its performance is measured directly against other laboratory tests or characterized panels.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Yes, the performance studies described are essentially standalone evaluations of the LIAISON® Lyme IgG assay. The assay is an automated chemiluminescent immunoassay (CLIA) performed on the LIAISON® XL Analyzer. The results (Index values) lead to a qualitative positive, negative, or equivocal determination by the instrument's algorithm based on a predefined cutoff. There is no human-in-the-loop performance described in the context of interpreting the assay's primary output. Human involvement is in running the test and interpreting the final qualitative result provided by the system within a clinical context.

7. The Type of Ground Truth Used:

  • Expert Consensus / Established Gold Standard:
    • For the prospective and retrospective studies that compare the LIAISON® assay to STTT (Standard Two-Tier Test) or MTTT (Modified Two-Tier Test), the ground truth is primarily based on IgG Borrelia burgdorferi Western Blot testing following current guidelines. Western blot is generally considered the confirmatory "gold standard" for Lyme disease serology.
    • For the "Characterized Lyme Panel" (from CDC), the ground truth is the CDC Reference Classification, which would be based on comprehensive testing and expert knowledge.
  • Predicate Device Comparison: In the initial method comparison, the ground truth for calculating agreement percentages is the ZEUS ELISA B. burgdorferi IgG Test System (predicate device).

8. The Sample Size for the Training Set:

The document describes performance studies (method comparison, STTT, MTTT, cross-reactivity, precision, reproducibility, etc.) of the LIAISON® Lyme IgG assay. It does not explicitly mention a "training set" or "validation set" in the context of an algorithm or AI model development. The reported studies represent the testing and validation of the already developed assay.

  • However, for the assay development process itself (not detailed in this summary), a training phase would typically occur during the R&D of the assay to establish optimal reagent concentrations, cut-offs, and assay parameters. This information is generally proprietary to the manufacturer and not typically included in a 510(k) summary unless the device is a novel AI/ML product.
  • The calibration described ("Two-point verification (in triplicate) of stored 10 point master curve") indicates a specific calibration approach for the instrument, which would have been established during initial assay development using a set of characterized samples. The number of samples for developing this master curve is not provided.

9. How the Ground Truth for the Training Set Was Established:

As noted above, a distinct "training set" for an AI/ML algorithm is not described here, as this is an immunoassay. For the development of the assay itself (e.g., establishing optimal cut-offs), the ground truth would have been established using well-characterized samples from known Lyme disease patients and healthy controls, likely confirmed by established methods like Western blot and/or clinical diagnosis, similar to the methods used to establish the ground truth for the test sets presented. The specifics of this internal R&D process are not detailed in the FDA summary.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it, also in blue.

February 18, 2021

DiaSorin Inc. Carol Depouw Principal Regulatory Affairs Specialist 1951 Northwestern Avenue Stillwater, Minnesota 55082

Re: K202574

Trade/Device Name: LIAISON Lyme IgG, LIAISON Lyme IgG Control Set, LIAISON Lyme Total Antibody Plus Regulation Number: 21 CFR 866.3830 Regulation Name: Treponema Pallidum Treponemal Test Reagents Regulatory Class: Class II Product Code: LSR Dated: September 2, 2020 Received: September 4, 2020

Dear Carol Depouw:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Maria Ines Garcia, Ph.D. Branch Chief Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202574

Device Name LIAISON® Lyme IgG assay: LIAISON® Lyme IgG Control Set

Indications for Use (Describe)

The LIAISON® Lyme IgG assay uses chemiluminescent immunoassay (CLIA) technology for the qualitative detection of IgG antibodies to Borrelia burgdorferi in human serum and plasma specimens (K2-EDTA, Li-heparin). This assay is intended for use on samples from patients with signs and symptoms consistent with or patients suspected of having Lyme disease to assess the presence of IgG antibodies and exposure to Borrelia burgdorferi. In addition, the LIAISON® Lyme Ig G assay may be used as a confirmatory test in the modified two-tier test (MTTT) in combination with the DiaSorin LIAISON® Lyme Total Antibody Plus assay.

If used as a first stage test, positive or equivocal results with the LIAISON® Lyme IgG assay should be confirmed through additional testing with a Standard two-tier test (STT) methodology using an IgG Borrelia burgdorferi Western blot test following current guidelines.

Positive supplemental results are supportive evidence of the presence of antibodies and exposure to Borrelia burgdorferi and may be used along with patient history, symptoms and other laboratory data to support a clinical diagnosis of Lyme disease.

Negative results by the LIAISON® Lyme IgG assay should not be used to exclude Lyme disease.

The test must be performed on the LIAISON® XL Analyzer.

The DiaSorin LIAISON® Lyme IgG Control Set is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Lyme IgG assay. The performance characteristics of LIAISON® Lyme IgG controls have not been established for any other assays or instrument platforms different from the LIAISON® XL.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.0 510(k) SUMMARY

SUBMITTED BY:Carol A. DePouwDiaSorin Inc.1951 Northwestern AvenueStillwater, MN 55082-0285Phone (651) 439-9710Fax (651) 351-5669Email carol.depouw@diasorin.com
DATE PREPARED:September 01, 2020
NAME OF DEVICE:Trade Name:LIAISON® Lyme IgGLIAISON® Lyme IgG Control Set
Common Names/Descriptions:Borrelia burgdorferi IgG assay andBorrelia burgdorferi IgG controls
Classification Names:Treponema pallidum; treponemal test reagentsClass II, 21 CFR: 866.3830; Microbiology
Product Code:LSR and QCH
Predicate Device:ZEUS ELISA B. burgdorferi IgG Test System(K895292)/modified (K191398)

INTENDED USE:

The LIAISON® Lyme IgG assay uses chemiluminescent immunoassay (CLIA) technology for the qualitative detection of IqG antibodies to Borrelia burqdorferi in human serum and plasma specimens (K2-EDTA, Li-heparin). This assay is intended for use on samples from patients with signs and symptoms consistent with or patients suspected of having Lyme disease to assess the presence of IgG antibodies and exposure to Borrelia burgdorferi. In addition, the LIAISON® Lyme IgG assay may be used as a confirmatory test in the modified two-tier test (MTT) in combination with the DiaSorin LIAISON® Lyme Total Antibody Plus assay.

If used as a first stage test, positive or equivocal results with the LIAISON® Lyme IgG assay should be confirmed through additional testing with a Standard two-tier test (STTT) methodology using an IgG Borrelia burgdorferi Western blot assay following current quidelines.

Positive supplemental results are supportive evidence of the presence of antibodies and exposure to Borrelia burgdorferi and may be used along with patient history, symptoms and other laboratory data to support a clinical diagnosis of Lyme disease.

Negative results by the LIAISON® Lyme IgG assay should not be used to exclude Lyme disease.

The test must be performed on the LIAISON® XL Analyzer.

The LIAISON® Lyme IgG Control Set is intended for use as assayed quality control samples to monitor the performance of the LIAISON® Lyme IgG assay. The performance characteristics of LIAISON® Lyme IgG controls have not been established for any other assays or instrument platforms different from the LIAISON® XL.

{4}------------------------------------------------

KIT DESCRIPTION:

The LIAISON® Lyme IgG assay is an indirect chemilyminescence immunoassay (CLIA) for the qualitative detection of IgG antibodies to Borrelia burgdorferi in human serum and plasma samples. All assay steps (with the exception of magnetic particle resuspension) and incubations are performed by the Analyzer. The principal components of the test are magnetic particles (solid phase) coated with recombinant Borrelia antigens and a conjugate reagent containing an anti-human IgG mouse monoclonal antibody linked to an isoluminol derivative (isoluminol-antibody coniugate). During the first incubation, anti-Borrelia burgdorferi antibodies present in calibrators, samples or controls bind to the solid phase. Unbound material is then removed with a wash cycle. During the second incubation, the antibody conjugate reacts with anti-Borrelia burgdorferi IgG antibodies that have bound to the solid phase. Excess antibody conjugate is then removed with a wash cycle. Subsequently, the starter reagents are added and a flash chemiluminescence reaction is induced. The light signal, and hence the amount of isoluminol-antibody conjugate, is measured by a photomultiplier as relative light units (RLU) and is indicative of the presence of Borrelia burgdorferi IgG antibodies present in calibrators, samples or controls.

Table 1: Table of Similarities
CharacteristicCandidate DeviceLIAISON® Lyme IgG
Intended UseQualitative detection of IgG antibodies to Borreliaburgdorferi in human serum and plasmaspecimens. This assay is intended for use onsamples from patients with signs and symptomsconsistent with or patients suspected of havingLyme disease to assess the presence of IgGantibodies and exposure to Borrelia burgdorferi.In addition, the LIAISON® Lyme IgG assay may beused as a confirmatory test in the modified two-tiertest (MTTT) in combination with the DiaSorinLIAISON® Lyme Total Antibody Plus assay.If used as a first stage test, positive or equivocalresults with the LIAISON® Lyme IgG assay shouldbe confirmed through additional testing with aStandard two-tier test (STTT) methodology usingan IgM Borrelia burgdorferi Western blot testfollowing current guidelines.Positive supplemental results are supportiveevidence of the presence of antibodies andexposure to Borrelia burgdorferi and may be usedalong with patient history, symptoms and otherlaboratory data to support a clinical diagnosis ofLyme disease.Negative results by the LIAISON® Lyme IgG assayshould not be used to exclude Lyme disease.
Predicate DeviceZEUS ELISABorrelia burgdorferi IgG Test System –K895292/K191398
Qualitative detection of IgG class antibody toBorrelia burgdorferi in human serum.The assay is intended for testing serumsamples from symptomatic patients or thosesuspected of Lyme Disease.Positive and equivocal test results with theZEUS ELISA Borrelia burgdorferi IgG TestSystem for the presence of Borrelia burgdorferiantibodies must be confirmed through additionaltesting by one of the following approaches:(1) Standard two-tier test methodology (STTT)using IgG Western blot testing following currentguidelines; or -(2) Modified two-tier test methodology using theZUES ELISA Borrelia VIsE1/pepC10 IgG/IgMTest System.Positive test results by either the STTT or MTTTmethodology are supportive evidence for thepresence of antibodies and exposure to Borreliaburgdorferi, the cause of Lyme disease.A diagnosis of Lyme disease should be madebased on the presence of Borrelia burgdorferiantibodies history, symptoms and otherlaboratory data.
ResultsQualitative
Same

COMPARISON TO PREDICATE DEVICE

{5}------------------------------------------------

CharacteristicCandidate DeviceLIAISON® Lyme IgGPredicate DeviceZEUS ELISABorrelia burgdorferi IgM Test System –K895292/K191398
MeasurandIgG antibodies to Borrelia burgdorferiSame
IntendedPopulationPatients with signs and symptomsconsistent with Borrelia infection(Lyme disease)Same
AssayPrincipleUses Borrelia antigens coated on a solid phaseto capture specific patient IgG antibodies.Uses B. burgdorferi antigen on coatedsolid phase (wells) to bind withIgG antibodies in patient sample
ConjugateantibodyspecificitiesAnti-human IgGSame
Assay OutputIndexSame
Table 2: Table of Differences
FeatureCandidate DeviceLIAISON® Lyme IgGPredicate DeviceZEUS ELISABorrelia burgdorferi IgM Test System –K895292/K191398
Test FormatCLIA (indirect chemiluminescentassay)ELISA
Sample TypeHuman serum, serum separatortubes, K2-EDTA, lithium heparin plasmaHuman serum
Reporter MoleculeIsoluminol derivative conjugated toanti-human IgGTMB (as a substrate for Horseradishperoxidase conjugated to anti-human IgG).
AntigenRecombinant antigens:VIsE ( B. burgdorferi strain B31)Peptide for C6 region of VIsEWhole cell antigen fromB. burgdorferi (B31 strain)
Assay ProcedureAutomated(on the LIAISON® XL Analyzer)Manual
CalibrationTwo-point verification (in triplicate) ofstored 10 point master curveSingle Cut-off Calibrator assayed in triplicate
Output SignalFlash chemiluminescent response isintegrated over a 3 second readingperiod to generate a relative light unitMicrotiter well O.D. (450 nm) is measuredafter the enzyme reaction is halted by1M H2SO4/0.7M HCl.
MeasurementSystemPhotomultiplier(flash chemiluminescence reader)Spectrophotometer(EIA Microtiter plate reader)

PERFORMANCE DATA: METHOD COMPARISON:

Two thousand six hundred twenty one (2621) human serum specimens were collected in 14 states which represented five (5) distinct U.S. geographical regions. Of the 2621 samples: 44.1% were male, 55.7% were female, 0.2% gender unknown, the ages ranged from 2 years to 103 years of age.

{6}------------------------------------------------

Testing with the LIAISON® Lyme IgG assay on the LIAISON® XL was performed in three (3) laboratories (2 external and internally at DiaSorin).

Predicate Assay (ELISA IgG)
LIAISON Lyme IgGPositiveEquivocalNegativeTotal
Positive156486246
Equivocal602430
Negative1191622102345
Total2812023202621
Table 3: First Tier Percent Agreement with Predicate Device
---------------------------------------------------------------------

Agreement Results

Positive % Agreement*55.1% (166/301)95% CI: 49.5% - 60.7%
Negative % Agreement96.5% (2210/2320)95% CI: 94.3% - 96.1%
*Includes Positive and Equivocal combined

Standard Two-Tier Testing Methodology:

Western blot testing was performed on the samples that were positive or equivocal by the test device and the predicate following the current quidelines for Standard Two-Tier testing methodology. The following results were obtained:

Table 4: Standard Two-Tier Western Blot
Test SystemTier 1 + or EqvWB IgG +WB IgG -
Predicate assay301109192
LIAISON® Lyme IgG276109167
Predicate assay + LIAISON® Lyme IgG1669769

Agreement Results:

2nd Tier PPA89.0% (97/109)95%CI: 81.7% - 93.6%
2nd Tier NPA99.5% (2500/2512)95%CI: 99.2% - 99.7%

Modified Two Tier Testing Methodology – Prospective Population

All 2621 prospective (all comer) specimens were tested with the first-tier assay, LIAISON® Lyme Total Antibody Plus. There were 202 positive and 23 equivocal results. In the STTT protocol, the specimens that are positive or equivocal (n=225) are then tested with a B. burgdorferi (IgG) Western Blot.

Using the MTTT algorithm, the positive/equivocal specimens (n=225) were tested on the LIAISON® Lyme IgG assay. The second-tier LIAISON® Lyme IgG equivocal and positive results were considered positive. The equivocal and positive results were added together, and the results compared with the STTT positive results. The results obtained are shown in Table 5.

{7}------------------------------------------------

WB-STTT (IgG)
+-Total
XL -MTTT (IgG)+9390183
-33942
Total96129225

Table 5: MTTT | IAISON | vme lgG compared to STTT WB lgG

Agreement Results

PPA:96.9% (93/96)95% CI: 91.2% - 98.9%
NPA:30.2% (39/129)95% CI: 23.0% - 38.6%

Characterized Lyme Panel:

Two hundred eighty samples of various reactivity were acquired from the CDC and evaluated internally at the manufacture's site. The results of the testing are presented here as a means of conveying further information on the performance of the LIAISON® Lyme IgG assay with a characterized serum panel. This does not imply an endorsement of the assay by the CDC.

CDC Reference Classification
Sample CategoryNLIAISON Lyme IgGPredicate IgG
PosNegEqvPosNegEqvPPA95% Wilson CI
Acute39281012315174.4% (29/39)58.9% - 85.4%
Convalescent312920226393.5% (29/31)79.3% - 98.2%
Late2020002000100% (20/20)83.9% - 100%
Look-alike Diseases909810977490% (81/90)82.1% - 94.6%
Healthy controls1003970198197.0% (97/100)91.5% - 99.0%

Table 6: Testing of CDC Lyme Reference Sera

Modified Two Tier Testing Methodology-Retrospective population

The 280 retrospective samples from the CDC were first tested with the LIAISON® Lyme Total Antibody Plus. One (1) sample, belonging to endemic negative control group, did not have sufficient volume for testing; therefore 279 retrospective samples were evaluated. The LIAISON® Lyme Total Antibody Plus assay vielded 82 positive and equivocal samples. The 82 positive and equivocal samples were then tested by the IgG Western Blot (STTT) or the LIAISON® Lyme IgG assay (MTTT).

{8}------------------------------------------------

Stage I (n=60)Stage II (n=10)Stage III (n=20)Healthy Controls (n=99)Disease Controls (n=90)
STTT-WB IgGMTTT-XL IgGSTTT-WB IgGMTTT-XL IgGSTTT-WB IgGMTTT-XL IgGSTTT-WB IgGMTTT-XL IgGSTTT-WB IgGMTTT-XL IgG
Positive18485920200002
Negative4212510099999088
Sensitivity or PPA30%80%50%90%100%100%N/AN/AN/AN/A
Specificity or NPAN/AN/AN/AN/AN/AN/A100%100%100%97.8%

The results of MTTT-IqG compared to WB-STTT (IgG) are in the following table.

PRECISION STUDY

A 12 day precision/repeatability was conducted at DiaSorin Inc. on two (2) lots of the LIAISON® Lyme IgG assay. Six (6) serum samples and two (2) lots of LIAISON® Lyme IgG Controls were tested for 12 days, 2 runs/day, and 2 reps per run by multiple technicians for a total of 48 replicates per lot spanning 2 calibration cycles. One (1) lot is presented below. The CLSI Document EP5-A3 was consulted in the preparation of the testing protocol.

Sample IDNMeanWithin RunBetween RunBetween DayTOTAL(Within-lot)
SD%CVSD%CVSD%CVSD%CV
Negative Control Lot V1480.460.024.2%0.024.7%0.000.0%0.035.9%
Positive Control Lot V1482.220.104.4%0.114.8%0.031.5%0.156.7%
Negative Control LotV2480.090.004.1%0.003.5%0.004.7%0.017.2%
Positive Control Lot V2482.390.072.9%0.218.6%0.000.0%0.177.2%
LG-QC1480.770.033.9%0.045.4%0.000.0%0.056.1%
LG-QC3481.370.042.7%0.053.8%0.042.7%0.075.4%
LG-QC4484.890.112.3%0.112.3%0.112.2%0.193.9%
LG-QC5480.820.056.4%0.045.2%0.045.4%0.089.8%
LG-QC6481.470.053.3%0.064.2%0.000.0%0.085.2%
LG-QC7480.480.023.3%0.011.7%0.012.3%0.024.4%

REPRODUCIBILITY STUDY

A five (5) day precision/reproducibility study was performed internally at DiaSorin Inc. and at two (2) external U.S. laboratories with one (1) lot of the LIAISON® Lyme IgG assay.

The study was performed for 5 days, 2 runs/day, and 3 replicates/run. Each day, two operators, at each testing site performed the testing for a total of 30 replicates at each site. The combined results from the 3 sites are provided. CLSI document EP15-A3 was consulted in the preparation of the testing protocol.

SampleIDnmeanWithin RunBetween DayBetween RunBetween SiteTOTAL
SD%CVSD%CVSD%CVSD%CVSD%CV
Negative Control900.1280.0064.3%0.0043.3%0.000.0%0.0064.8%0.0097.3%
Positive Control902.250.0642.9%0.0482.1%0.0371.7%0.1135.0%0.1446.4%
LG-QC1900.8200.0182.2%0.0324.0%0.0162.0%0.0435.3%0.0607.3%
LG-QC3901.400.0392.8%0.0352.5%0.0412.9%0.0392.8%0.0775.5%
LG-QC4905.050.1132.2%0.1422.8%0.0681.3%0.2034.0%0.2805.5%
LG-QC5900.7350.0212.8%0.0304.1%0.0172.2%0.0344.6%0.0537.2%
LG-QC6901.630.0483.0%0.0291.8%0.0352.1%0.0654.0%0.0925.7%
LG-QC7900.5420.0132.5%0.0203.7%0.0061.1%0.0336.2%0.0417.7%

{9}------------------------------------------------

CROSS-REACTIVITY STUDY

The cross-reactivity study was designed to evaluate 222 specimens from twenty-two (22) disease states either known to contain potentially cross reactive antibodies to B. burgdorferi or from patients with diagnoses that can exhibit signs and symptoms similar to Lyme disease and cause false positive results.

Organism/Disease StateSamplesTested(n)LIAISON® Lyme IgGPos or Eqv
Tick Borne Diseases*
Babesiosis IgG105
Autoimmune Disorders
Anti-Nuclear Antibodies (ANA)100
Multiple Sclerosis100
Viral Diseases
Cytomegalovirus (CMV) IgG100
Epstein-Barr Virus (EBV) IgG100
Epstein-Barr Virus (EBV) EBNA IgG101
Epstein-Barr Virus (VCA) IgG100
Herpes Simplex Virus (HSV) IgG100
Human Immunodeficiency Virus (HIV)100
Influenza Virus100
Parvovirus IgG100
Varicella Zoster Virus (VZV) IgG100
Bacterial Diseases
H. pylori101
Syphilis100
Leptospirosis21
Rheumatic Diseases
Fibromyalgia100
Rheumatoid Arthritis100
Rheumatoid Factor100
Systemic Lupus Erythematosus (SLE)100
Sjogrens Syndrome101
Additional Markers
Chronic Fatigue Syndrome101
Human Anti-mouse Antibodies (HAMA)100
E. coli100
Total22210

{10}------------------------------------------------

INTERFERING SUBSTANCES

"Controlled studies of potentially interfering substances from endogenous interferents spiked into serum specimens containing B. burgdorferi IgG antibodies at levels near the cut-off showed that assay performance was not affected at the concentration for each substance listed below. The testing was based on CLSI-EP7-A3.

SubstancesTestedConcentrations
Hemoglobin1000 mg/dL
Triglycerides1500 mg/dL
Bilirubin40 mg/dL
Total protein15 g/dL
Cholesterol400 mg/dL
Biotin3600 ng/mL

MATRIX EQUIVALENCE STUDY:

Thirty-two (32) matched patient sets of serum, SST serum, K2-EDTA plasma and lithium heparin plasma samples were tested to determine if these sample types provide equivalent results. Sample regression analysis was done by Passing & Bablok regression. All sample types met acceptance criteria for use in the LIAISON® Lyme IgG assay. A summary of the results are shown in the following table.

Comparison to SerumBias95% Cl
Serum SSTConstant-0.01-0.04 to 0.00
Proportional1.041.01 to 1.10
EDTA PlasmaConstant0.00-0.01 to 0.03
Proportional0.960.93 to 1.01
Lithium Heparin PlasmaConstant0.01-0.04 to 0.04
Proportional0.970.91 to 1.03

CONCLUSION:

The material submitted in this premarket notification supports a substantial equivalence decision. The labelling satisfies the requirements of 21CFR 809.10.

§ 866.3830

Treponema pallidum treponemal test reagents.(a)
Identification. Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), theTreponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies toTreponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genusTreponema and provides epidemiological information on syphilis.(b)
Classification. Class II (performance standards).