K Number
K202491

Validate with FDA (Live)

Date Cleared
2021-01-06

(128 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Device Description

Disposable Surgical Face Mask is composed of three layers and is flat-pleated. The mask materials consist of an outer layer (spun-bond polypropylene), a middle layer, between the outer layer and inner layers (melt-blown polypropylene), and an inner layer (spun-bond polypropylene). Each mask contains tie strings (spun-bond polypropylene) or ear loops (Spandex elastic cord) to secure the mask over the users' mouth and face and includes a malleable nose piece (iron wire with plastic covering) to provide a firm fit over the nose.

AI/ML Overview

This document is a 510(k) summary for a Disposable Surgical Face Mask. To provide a comprehensive answer regarding acceptance criteria and study proving device meets them, I will focus on the non-clinical performance tests which are detailed for this device. This document does not describe a clinical study or an AI/human-in-the-loop system.

Here's a breakdown based on the provided text, addressing your points where information is available:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are established by various ASTM standards (F2100, F1862, F2299, F2101) and others (MIL-M-36954C, 16 CFR Part 1610, ISO 10993 series). The device performance is compared to these standards across different performance levels (Level 1, Level 2, Level 3).

Acceptance Criterion (Standard & Requirement)Subject Device Performance (Disposable Surgical Face Mask K202491)Predicate Device Performance (Surgical Face Mask K160269)Result
Fluid Resistance (ASTM F1862)
Pass at 80 mmHg (Level 1, 2)Pass at 80 mmHgPass at 80 mmHgSame
Pass at 120 mmHg (Level 2)Pass at 120 mmHgPass at 120 mmHgSame
Pass at 160 mmHg (Level 3)Pass at 160 mmHgPass at 160 mmHgSame
Particulate Filtration Efficiency (PFE) (ASTM F2299)
Level 1Pass at >99.8%Pass at 99.6%Same / Better
Level 2Pass at 99.6%Pass at 99.7%Same
Level 3Pass at 99.6%Pass at 99.7%Same
Bacterial Filtration Efficiency (BFE) (ASTM F2101)
Level 1Pass at ≥99.8%Pass at >98%Same / Better
Level 2Pass at >98%Pass at >98%Same
Level 3Pass at >99%Pass at >99%Same
Differential Pressure (Delta P) (MIL-M-36954C)
Level 1Pass at <4.2mmH2O/cm²SameSimilar
Level 2Pass at 2.0 mmH2O /cm²SimilarSimilar
Level 3Pass at 1.6 mmH2O/cm²Pass at 2.5 mmH2O/cm²Same / Better
Flammability (16 CFR PART 1610)
Class 1Class 1 Non-FlammableClass 1Same
Biocompatibility (ISO 10993-1, 5, 10)
CytotoxicityNon-cytotoxicNon-cytotoxicSame
IrritationNon-irritatingNon-irritatingSame
SensitizationNon-sensitizingNon-sensitizingSame

Note: The table above extracts the relevant quantitative and qualitative performance data. "Same" implies performance is within acceptable limits relative to the predicate or standard. "Better" implies the subject device shows a numerically superior result where higher numbers are better (e.g., higher PFE/BFE) or lower numbers are better (e.g., lower Delta P, indicating better breathability).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not specify the exact sample sizes (e.g., number of masks, test replicates) used for each non-clinical performance test. It states "Non-clinical tests were conducted..." without providing the number of units tested.
  • Data Provenance: The tests are non-clinical (laboratory) rather than patient data. The manufacturer is Jiangsu New Value Medical Products Co., Ltd. in China, implying the tests were likely conducted in China or by labs contracted by the manufacturer. The data is retrospective in the sense that it was collected prior to this submission for regulatory review.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

  • This evaluation is based on non-clinical, objective laboratory testing against established performance standards (ASTM, MIL-M, ISO, CFR). Therefore, there isn't a concept of a "ground truth established by experts" in the traditional sense of medical image interpretation or clinical outcomes. The "ground truth" is defined by the passing criteria of the referenced standards.
  • The qualifications of the laboratory personnel performing these tests are not explicitly stated in this document, but they would be expected to be trained and qualified in laboratory testing and quality control for medical devices.

4. Adjudication Method for the Test Set

  • Not applicable. As described above, this is objective laboratory testing against defined standards, not a subjective assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No. This document explicitly states in "Clinical Test Conclusion": "No clinical study is included in this submission."
  • Therefore, an MRMC comparative effectiveness study was not performed, and there is no information on how human readers would improve with AI assistance, as AI is not part of this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • No. This is a physical medical device (surgical face mask), not an AI algorithm. Therefore, the concept of standalone algorithm performance is not applicable.

7. The Type of Ground Truth Used

  • For this 510(k) submission, the "ground truth" for demonstrating substantial equivalence is based on objective, standardized laboratory test measurements that meet pre-defined performance criteria outlined in recognized consensus standards (e.g., ASTM F2100, ASTM F1862, ASTM F2299, ASTM F2101, MIL-M-36954C, 16 CFR Part 1610, ISO 10993 series). The comparison is also made against a legally marketed predicate device (K160269) for which similar data would have been accepted.

8. The Sample Size for the Training Set

  • Not applicable. This is a physical device, not an AI/machine learning model, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

  • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

January 6, 2021

Jiangsu New Value Medical Products Co., Ltd. % Julie Chen Technical Manager Shanghai Medical Business Consulting Co., Ltd. Room 304, No 170 Huajiang Road, Jiading District Shanghai. 201803 China

Re: K202491

Trade/Device Name: Disposable Surgical Face Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: November 16, 2020 Received: December 7, 2020

Dear Julie Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

Device Name Disposable Surgical Face Mask

Indications for Use (Describe)

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable device, provided non-sterile.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary — K202491

I. SUBMITTER:

Jiangsu NewValue Medical Products Co., Ltd. Building G35, No.1 Avenue, China Medical City, Taizhou, Jiangsu, 225300, China Contact Person: Zhou Zhengguo Title: Quality Manager Tel: +86(0523)86867878 Email: zhouzhg@new-value.cn

Submission Correspondent: Julie Chen Email: julie.chen@medicalbc.net Tel:+86 139 1804 5781

Summary prepared: 01/04/2021

II. DEVICE

Name of Device: Disposable Surgical Face Mask Regulation Number: 21 CFR PART 878.4040 Common Name: Surgical Mask Classification Name: Surgical Mask Regulatory Class: II Product Code: FXX

PREDICATE DEVICE III.

Primary predicate device: Surgical Face Masks (K160269)

{4}------------------------------------------------

DEVICE DESCRIPTION IV.

Disposable Surgical Face Mask is composed of three layers and is flat-pleated. The mask materials consist of an outer layer (spun-bond polypropylene), a middle layer, between the outer layer and inner layers (melt-blown polypropylene), and an inner layer (spun-bond polypropylene). Each mask contains tie strings (spun-bond polypropylene) or ear loops (Spandex elastic cord) to secure the mask over the users' mouth and face and includes a malleable nose piece (iron wire with plastic covering) to provide a firm fit over the nose.

REF No.ProductSizeModel DescriptionMask Style
Blue MaskEar LoopLevel 1Level 2Level 3
EL-M01145×95mmXXX
EL-M02145×95mmXXX
EL-M03145×95mmXXX
EL-L01175×95mmXXX
EL-L02175×95mmXXX
EL-L03175×95mmXXX
  • V. Available Model

VI. INDICATIONS FOR USE

The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single-use, disposable-device, provided non-sterile.

VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Disposable Surgical Face Masks are compared with the predicate device (Surgical Face Masks (K160269)). The results are shown below in the Technological Characteristics Comparison Table:

{5}------------------------------------------------

DEVICESubject DevicePrimary Predicate DeviceSubstantialEquivalence
Disposable Surgical FaceMask (K202491)Surgical Face Mask(K160269)
Level 1Level 2Level 3Level 1Level 2Level 3
Intended UseThe Disposable SurgicalFace Masks are intended tobe worn to protect both thepatient and healthcarepersonnel from transfer ofmicroorganisms, bodyfluids and particulatematerial. These face masksare intended for use ininfection control practicesto reduce the potentialexposure to blood and bodyfluids. The DisposableSurgical Face Masks aresingle use, disposabledevice, providednon-sterile.The surgical face masksare intended to be wornto protect both thepatient and healthcarepersonnel from transferof microorganisms,body fluids, andparticulate material.These face masks areintended for use ininfection controlpractices to reduce thepotential exposure toblood and body fluids.This is a single-use,disposable device,provided non-sterile.Same
Classification ProductCodeFXXFXXSame
Ear Loop Model andTie-on ModelEar LoopsEar Loops, Tie-onSimilar
Materials
Outer Facing LayerSpun-bond Polypropylenenon-woven fabricPolypropyleneSimilar
Middle LayerMelt-blown polypropylene1. Polypropylene spunbond2. PolypropylenemeltblownSimilar
Inner Facing LayerSpun-bond Polypropylenenon-woven fabricPolyethylene/PolyesterSimilar
Nose PieceMalleable iron wire withplastic coveringPolyethylene coated steelwireSimilar
Ear LoopsSpandex Elastic cordEar loops: Polyester,polyurethaneSimilar

{6}------------------------------------------------

Polyester spunbond (ear loops mask only)
Design Features
ColorBlueBlue, whiteSimilar
StyleFlat - PleatedFlat - foldedSame
Multiple Layers3 Layers4 LayersDifferent
Single UseSingle useSingle useSame
Sterility
SterileNon-sterileNon-sterileSame
Dimensions
Length x Width145×95mm (±5mm)175×95mm (±5mm)Length: 90 ± 3 mmWidth:175 ± 5mmLength: 90 ± 3 mmWidth:180 ± 5mmSimilar
Technological Characteristics Product Barrier Specifications Per ASTM F2100 – Meets Level 1 Leve2, Level 3
Level 1Level 2Level 3
Fluid Resistance ASTM F1862Pass at 80 mmHg;Pass at 120 mmHgPass at 160mmHg;Pass at 80 mmHgPass at 120 mmHgPass at 160 mmHgSame
Particulate Filtration Efficiency (PFE) ASTM F2299Pass at >99.8%Pass at 99.6%Pass at 99.6%Pass at 99.7%Same
Bacterial Filtration Efficiency (BFE) ASTM F2101Pass at ≥99.8%Pass at >98%Pass at >98%Pass at >99%Same
Differential Pressure (Delta P) MIL-M-36954CPass at <4.2mmH2O/cm²Pass at 2.0 mmH2O /cm²Pass at 1.6 mmH2O/cm²Pass at 2.5 mmH2O/cm²Same
Flammability 16 CFR PART 1610Class 1 Non-FlammableClass 1Class 1Class 1Same
Biocompatibility
CytotoxicityNon-cytotoxicNon-cytotoxicSame
IrritationNon-sensitizingNon-sensitizingSame
SensitizationNon-irritatingNon-irritatingSame

{7}------------------------------------------------

VIII. PERFORMANCE DATA

Non-Clinical Performance Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the Disposable Surgical Face Masks complies with the following standards:

  • ASTM F2100-19 Standard Specification for Performance of Materials Used . in Medical Face Masks
  • ASTM F1862 Standard Test Method for Resistance of Medical Face Masks to ● Penetration by
  • Synthetic Blood (Horizontal Projection of Fixed Volume at a Known ● Velocity)
  • ASTM F2299 Standard Test Method for Determining the Initial Efficiency of . Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres
  • ASTM F2101 Standard Method for Evaluating the Bacterial Filtration ● Efficiency (BFE) of Medical Face Mask Materials, Using a Biological Aerosol of Staphylococcus aureus
  • MIL-M- 36954C Military Specification, Mask, Surgical, Disposable ●
  • 16 CFR Part 1610 Standard for the Flammability of Clothing ●
  • ISO10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
  • . ISO10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity of medical devices
  • ISO10993-10 Biological Evaluation of Medical Devices Part 10: Tests for . Irritation and Skin Sensitization

Biocompatibility

Under the conditions of the study, the proposed device extract was determined to be non-cytotoxic, non-sensitizing and non-irritating.

Clinical Test Conclusion

No clinical study is included in this submission.

DISCUSSION AND CONCLUSION IX.

Although the design features of proposed device is slightly different with the predicate device, the data drawn from the non-clinical tests demonstrate that the proposed device, the Disposable Surgical Face Mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device Surgical Face

{8}------------------------------------------------

Masks (K160269).

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.