K Number
K202395

Validate with FDA (Live)

Device Name
SP-1000, SP-2000
Date Cleared
2021-09-01

(376 days)

Product Code
Regulation Number
890.5650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SP-1000/2000 is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health.

Device Description

Operating principal of this device is that the air output from the air motor is delivered to a sleeve composed of four air chambers which are sequentially inflated from the first air chamber to the fourth chamber. SP-1000 can adjust the pressure values of all chambers at once. The SP-2000 can adjust the pressure value for each channel. This device is composed of a power supply, a pump, a control unit, and a wearable unit. The power applied from the power supply is applied to the pump. And user can adjust the time through time control knob, during which time power is supplied to the air pump. When the set time is over, the pump will stop working. Air pump creates air pressure and transmits it to the Synchro motor. The air delivered to the synchro motor distributes the air to each channel of the sleeve. And there is a pressure control valve between the synchro motor, the user can adjust the pressure of each sleeve through the pressure control knob. The sleeve prevents blood from accumulating by repeatedly applying inflation/deflation to the wearing to the type of the sleeve, it can apply to the leg, lower body, and arm.

AI/ML Overview

The provided text is a 510(k) Summary for the SP-1000 and SP-2000 devices. It states that clinical testing was NOT required to demonstrate substantial equivalence to the predicate device. Therefore, a study proving the device meets acceptance criteria, as per the typical structure of an FDA clearance for AI/ML or diagnostic devices that require clinical validation, was not performed.

The document primarily focuses on demonstrating substantial equivalence through:

  • Comparison of Indications for Use: Both the proposed devices and the predicate device are indicated for "temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health."
  • Comparison of Technological Characteristics: Similar operating principles (air inflation, 4 chambers), power sources, time settings, pressure ranges, and accessories (leg, arm, body sleeves).
  • Non-Clinical Testing: Compliance with electrical safety and electromagnetic compatibility standards (IEC 60601 series) and usability.

Since no clinical study was conducted for this 510(k) clearance, I cannot provide details on:

  1. Acceptance Criteria Table: No clinical acceptance criteria were defined or met through a clinical study.
  2. Sample Size/Data Provenance: No test set or patient data was used for clinical evaluation.
  3. Number of Experts/Qualifications: Not applicable as no ground truth was established from clinical data.
  4. Adjudication Method: Not applicable.
  5. MRMC Study/Effect Size: Not applicable.
  6. Standalone Performance: Not applicable.
  7. Type of Ground Truth: Not applicable.
  8. Training Set Sample Size: Not applicable as the device is not an AI/ML product and did not involve a training set as understood in AI/ML validation.
  9. Ground Truth for Training Set: Not applicable.

In summary, the FDA cleared the SP-1000/SP-2000 based on substantial equivalence to a legally marketed predicate device (Rapid Reboot Compression Therapy System) and compliance with relevant non-clinical (electrical safety, EMC, usability) standards, without the need for a clinical study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 1, 2021

DaeSung Maref Co., Ltd % Do Gyun Lim Senior Consultant Global Medical Standard Consulting Co., Ltd. 34, Sangamsan-ro, Mapo-gu Seoul, 03909 Korea. South

Re: K202395

Trade/Device Name: Sp-1000, Sp-2000 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: September 9, 2020 Received: September 10, 2020

Dear Do Gyun Lim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Amber Ballard Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202395

Device Name SP-1000, SP-2000

Indications for Use (Describe)

SP-1000/2000 is indicated for the temporary relief of minor muscle aches and pains and for temporaryincrease in circulation to the treated areas in people who are in good health.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K202395 510(K) Summary

1. Data Prepared [21 CFR 807.92(a)(a)|

May 08, 2021

2. Submitter's Information [21 CFR 807.92(a)(1)]

  • Name of Manufacturer :
    DAESUNG MAREF CO., LTD.

  • Address :
    298-24, Gongdan-ro Gunpo-si, Gyeonggido Republic of Korea

  • Contact Name :
    Hyoung Ju, Cho

  • Telephone No. :
    82-31-459-7211

  • Fax No. :
    82-31-459-7215

  • Email Adress :
    rndra@dsmaref.com

  • · Registration No. :

    • 3004116008

3. Trade Name, Regulation Name, Classification [21 CFR 807.92(a)(2)|

Trade / Device NameSP-1000, SP-2000
Common NamePowered Inflatable Tube Massager
Regulation Number21 CFR 890.5650
Regulation NamePowered Inflatable Tube Massager
Regulation ClassII
Product CodeIRP

4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)|

Predicate Device

  • 510(k) Number :
    K182668

  • Applicant :
    Rapid Reboot Recovery Product, LLC

  • Trade / Device Name :
    Rapid Reboot Compression Therapy System

  • · Regulation Number :
    21 CFR 890.5650

  • · Regulation Name :
    Powered Inflatable Tube Massager

  • · Regulation Class:

    • II
  • Product Code:

IRP

Predicate device has not been subject to a design-related recall.

5. Description of the Device [21 CFR 807.92(a)(4)]

Operating principal of this device is that the air output from the air motor is delivered to a sleeve composed of four air chambers which are sequentially inflated from the first air chamber to the fourth chamber.

SP-1000 can adjust the pressure values of all chambers at once. The SP-2000 can adjust the pressure value for each channel. This device is composed of a power supply, a pump, a control unit, and a wearable unit. The power applied from the power supply is applied to the pump. And user can adjust the time through time control knob, during which time power is supplied to the air pump. When the set time is over, the pump will stop working. Air pump creates air pressure and transmits it to the Synchro motor. The air delivered to the synchro motor distributes the air to each channel of the sleeve. And there is a pressure control valve between the synchro motor, the user can adjust the pressure of each sleeve through the pressure control knob. The sleeve prevents blood from accumulating by repeatedly applying inflation/deflation to the wearing to the type of the sleeve, it can apply to the leg, lower body, and arm.

{4}------------------------------------------------

K202395 510(K) Summary

6. Indications For Use [21 CFR 807(a)(5)]

SP-1000/2000 is indicated for the temporary relief of minor muscle aches and pains and for temporary increase in circulation to the treated areas in people who are in good health.

7. Determination of Substantial Equivalence

Summary of technological characteristics of the device compared to the predicate device. [21CFR

807.92(a)(6)]

The SP-1000/2000 is substantially equivalent to legally marketed predicate device (Rapid Reboot Compression Therapy System) with respect to indications for use and technology characteristics. The table below presents comparisons for devices:

[Table 1. Comparison of Proposed Device to Predicate Devices]

Proposed DevicePredicate Device
Model NameSP-1000 / SP-2000Rapid Reboot Compression Therapy System
510(k) NumberK202395K182668
ManufacturerDAESUNG MAREF CO., LTD.Raid Reboot Recovery Product, LLC
Product CodeIRPIRP
Device ClassIIII
Regulation Number21 CFR 890.565021 CFR 890.5650
Regulation NamePowered Inflatable Tube MassagerPowered Inflatable Tube Massager
Indications For UseSP-1000/2000 is indicated for the temporary relief of minormuscle aches and pains and for temporary increase incirculation to the treated areas in people who are in goodhealth.The Rapid Reboot Compression Therapy System is intended for thetemporary relief of minor muscle aches and pains and for the temporaryincrease in circulation to the treated areas in people who are in goodhealth. The Rapid Reboot Compression Therapy System simulateskneading and stroking of tissues by using an inflatable garment.
Intended Use environmentProfessional healthcare environment & Home environmentClinics, hospital, athlete training, and home environments
Type of UseOTCOTC
AccessoriesLeg sleevesArm sleeveCenter body sleeveLeg attachmentArm attachmentHip attachment
Specifications
Power Source100-127V~, 50/60Hz200-240V~, 50/60Hz100V, 60 Hz
Time0-30 min0-30 min
Pressure0-200mmHg ± 15% mmHg0-200mmHg
Number of chamber44

The table also provides rationale for a little difference in support of substantial equivalence to the Predicate devices.

[Table 2. Little difference with Predicate Device]

Justification to Support Substantial Equivalence

There are no significant differences between SP-1000 and the predicate devices that would adversely affect its use. The subject device declares the safety and effectiveness of the device through tests for appropriate standards and the accessories, though named differently, are used on the same anatomical locations as the predicate device. Therefore, It is substantially equivalent in indications for use and technology characteristics.

Non-Clinical Test Summary

The SP-1000 / SP-2000 comply with voluntary standards for electromagnetic compatibility, use in the home healthcare environment and usability.

The following data were provided in support of the substantial equivalence determination :

  1. Electrical Safety, Electromagnetic Compatibility and Performance

The SP-1000 / SP-2000 comply with the electrical safety and electromagnetic compatibility requirements established by the standards.

  • IEC 60601 : 2005/A1:2012, Medical Electrical Equipment:Part 1: General Requirements for Basic Safety and Essential Performance
  • · IEC 60601-1-2:2014, Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements and Tests
  • · IEC 60601-1-11:2015, Medical Electrical Equipment Part 1-11: General Requirements for Basic Safety and Essential Per formance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment

{5}------------------------------------------------

K202395 510(K) Summary

Clinical Test Summary

Clinical testing was not required to demonstrate the substantial equivalence of the SP-1000 / SP-2000 to its predicate device

8. Conclusion [21 CFR 807.92(b)(3)]

The SP-1000 / SP-2000 have similar intended use and technical characteristics to the predicate devices. Based on that information, we conclude that the differences between the proposed device and predicate devices do not introduce a new intended use and do not raise of safety and effectiveness. Therefore, the subject device is substantially equivalent to the predicate device.

§ 890.5650 Powered inflatable tube massager.

(a)
Identification. A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.(b)
Classification. Class II (performance standards).