K Number
K202139

Validate with FDA (Live)

Date Cleared
2020-11-20

(112 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

This product is indicated for infection control practices in the health care industry. When worn properly, the Medical Surgical Mask is intended to protect both patient and wearer from the transfer of microorganisms, body fluids and particulate material.

Device Description

The Medical surgical mask is a flat style face mask with ear loops and nose clip for fitting around the nose and mouth. The Medical surgical mask has three layers are made of polypropylene nonwoven and the middle layer is made of polypropylene melt-blown. The outer layer is blue, and the colorant material is identified as Pigment Blue K6911D /CAS number: 12239-87-1. The face mask is held in place over the users' nose and mouth by two polyester and spandex elastic bands as ear loops welded to the face mask. The nose clip is made of iron-cored polypropylene, which allows the users to fit the mask around their nose area.

The dimensions of each mask are 175±5 mm in width. The density of the inner and outer layer is 25 gsm, and the density of the middle layer is 35 gsm. The dimensions of nosepiece are 100±5 mm in length and 3±0.5 mm in width. The ear loop is 175±10 mm in length and 3.5±0.5 mm in width. The Medical surgical mask is sold nonsterile and is intended to be single use, disposable device.

AI/ML Overview

The provided text describes the non-clinical testing performed for the Medical Surgical Mask (Model: KF-B P05(L3)) to demonstrate its substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance:

Test PerformanceAcceptance Criteria (Predicate)Reported Device Performance (Subject Device)
Bacterial Filtration Efficiency$\ge 98%$99.65%
Differential Pressure (Delta-P)$< 6.0$ mm H2O/cm²3.72 mm H2O/cm²
Sub-micron particulate filtration efficiency at 0.1 µm of Polystyrene Latex Spheres$\ge 98%$99.9%
Resistance to penetration by synthetic bloodFluid resistant claimed at 160 mm HgFluid Resistant claimed at 160 mm Hg
Flame SpreadClass 1Class 1

Note: The table above primarily lists the performance criteria for Level 3 surgical masks, which both the subject and predicate devices claim to meet. Some direct predicate values are not explicitly stated, but the subject device's performance is compared against the general standard.

2. Sample size used for the test set and the data provenance:

  • The document does not specify the exact sample sizes used for each non-clinical test.
  • The data provenance is from non-clinical tests performed in a laboratory setting, likely conducted by or for the manufacturer, GUANGDONG KINGFA SCI. & TECH.CO., LTD., which is based in China. The nature of these tests (e.g., ASTM F2101-14, EN 14683: 2019, ASTM F2299-03, ASTM F1862/F1862M-17, 16 CFR Part 1610) indicates they are retrospective laboratory evaluations against established standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

  • This information is not applicable as the tests are non-clinical, objective measurements of physical properties based on established ASTM and EN standards. There are no human experts establishing a "ground truth" in the clinical sense for these performance tests. The "ground truth" is defined by the technical specifications of the standards themselves.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • This information is not applicable as the tests are objective, quantitative measurements following standardized protocols, not subjective human assessments requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This information is not applicable. The device is a medical surgical mask, not an AI-powered diagnostic or assistive tool for human readers. No MRMC study was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This information is not applicable. The device is a medical surgical mask, which is a physical product, not an algorithm. Standalone performance testing refers to the inherent physical measurement capabilities of the mask itself, as detailed in the performance table.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for the non-clinical tests is based on established international and national standards and test methods (e.g., ASTM F2101-14, EN 14683: 2019, ASTM F2299-03, ASTM F1862/F1862M-17, 16 CFR Part 1610). These standards define the procedures and acceptance criteria for measuring specific physical properties of medical masks.

8. The sample size for the training set:

  • This information is not applicable. The device is a physical product, not an AI model that requires a training set.

9. How the ground truth for the training set was established:

  • This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 20, 2020

Guangdong Kingfa SCI.&Tech.Co., Ltd. % Cassie Lee Manager Share Info (Guangzhou) Medical Consultant Ltd. No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District Guangzhou, Guangdong 510700 China

Re: K202139

Trade/Device Name: Medical surgical mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: October 15, 2020 Received: October 22, 2020

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For CAPT Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202139

Device Name Medical Surgical Mask (Model: KF-B P05(L3))

Indications for Use (Describe)

This product is indicated for infection control practices in the health care industry. When worn properly, the Medical Surgical Mask is intended to protect both patient and wearer from the transfer of microorganisms, body fluids and particulate material.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

K202139

prepared in accordance with the requirement of 21 CFR 807.92.

1. Submitter's Information

510(k) Owner's Name: GUANGDONG KINGFA SCI. & TECH.CO., LTD.

Establishment Registration Number: 3016785267

Address: NO.28 Delong Avenue, Shijiao Town, Qingcheng District, Qingyuan City, Guangdong Province, China

Post Code: 511500

Contact Person: Yu Xiaoge

Tel: +86 13570952157

Fax: +0763-3203108

Email: yuxiaoge@kingfa.com.cn

Application Correspondent:

Contact Person: Ms. Cassie Lee, Share Info (Guangzhou) Medical Consultant Ltd.

Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China

Tel: +86 20 8266 2446

Email: regulatory@glomed-info.com

Date of preparation: October 15, 2020

2. Subject Device Information

Type of 510(k): Traditional

Classification Name: Mask, Surgical

Trade Name: Medical surgical mask

Model Name: KF-B P05(L3)

Review Panel: General Hospital

Product Code: FXX

Regulation Number: 878.4040

Regulatory Class: 2

3. Predicate Device Information

Sponsor: H&H RESEARCH COMPANY Trade Name: The New Medical Mask

Classification Name: Mask, Surgical

510(K) Number: K093179

Review Panel: General Hospital Product Code: FXX

Regulation Number: 878.4040

Regulation Class: 2

{4}------------------------------------------------

4. Device Description

The Medical surgical mask is a flat style face mask with ear loops and nose clip for fitting around the nose and mouth. The Medical surgical mask has three layers are made of polypropylene nonwoven and the middle layer is made of polypropylene melt-blown. The outer layer is blue, and the colorant material is identified as Pigment Blue K6911D /CAS number: 12239-87-1. The face mask is held in place over the users' nose and mouth by two polyester and spandex elastic bands as ear loops welded to the face mask. The nose clip is made of iron-cored polypropylene, which allows the users to fit the mask around their nose area.

The dimensions of each mask are 175±5 mm in width. The density of the inner and outer layer is 25 gsm, and the density of the middle layer is 35 gsm. The dimensions of nosepiece are 100±5 mm in length and 3±0.5 mm in width. The ear loop is 175±10 mm in length and 3.5±0.5 mm in width. The Medical surgical mask is sold nonsterile and is intended to be single use, disposable device.

5. Intended Use / Indications for Use

The Medical surgical mask is indicated for infection control practices in the health care facilities. When worn properly, the Medical surgical mask is intended to protect both patient and wearer from the transfer of microorganisms, body fluids and particulate material.

Elements ofSubject DevicePredicate DeviceVerdict
Comparison
CompanyGUANGDONG KINGFA SCI.&TECH.CO.,LTD.H&H RESEARCH COMPANY--
510 (k)K202139K093179--
Trade NameMedical surgical maskThe New Medical Mask--
ClassificationNameMask, SurgicalMask, SurgicalIdentical
ClassificationClass II Device, FXX (21CFR878.4040)Class II Device, FXX (21CFR878.4040)Identical
Intended use/Indications forUseThis product is indicated for infectioncontrol practices in the health careindustry. When worn properly, TheMedical surgical mask is intended toprotect both patient and wearer from thetransfer of microorganisms, body fluidsand particulate material.This product is indicated for infectioncontrol practices in the health careindustry. When worn properly, TheNew Medical Mask is intended toprotect both patient and wearer fromthe transfer of microorganisms, bodyfluids and particulate material.Identical
Material
Outer facinglayerPolypropylenePolypropyleneIdentical
Middle layerPolypropylene melt-blownPolypropyleneSimilarNote 1
Inner facinglayerPolypropylenePolypropyleneIdentical
Nose clipIron core polypropylene stripAdhesive tapeDifferentNote 1
Ear LoopsPolyester and spandex elastic bandsNon-latex elastic ear bandsDifferentNote 1
DesignfeaturesColor: blueColor: blueIdentical
Mask StyleEar loop flat styleEar loop flat styleIdentical
SpecificationandDimension17.5cm×9.5cmLength: 7.1 inches (18 cm)Width: 3.9 inches (10 cm)SimilarNote 1

6. Comparison to predicate device and conclusion

{5}------------------------------------------------

OTC useYesYesIdentical
SterilityNon-SterileNon-SterileIdentical
UseSingle Use, DisposableSingle Use, DisposableIdentical
ProtectionlevelLevel 3Level 3Identical
FluidResistancePerformancePass at 160 mmHgPass at 160 mmHgIdentical
ParticulateFiltrationEfficiency99.65%99.9%SimilarNote 2
BacterialFiltrationEfficiency>99.9%>99.9%Identical
DifferentialPressureOn average of 3.72 mm H2O/cm²Pass at 2.7 mmH2O/cm²SimilarNote 2
FlammabilityClass 1Class 1Identical
LatexNot Made with Natural Rubber LatexNot Made with Natural Rubber LatexIdentical
Biocompatibility
CytotoxicityUnder the conditions of the study, thesubject device extract was determined tobe non-cytotoxic.Under the conditions of the study, thesubject device extract was determinedto be non-cytotoxic.Identical
IrritationUnder the conditions of the study, thesubject device non-polar and polarextracts were determined to be non-irritating.Under the conditions of the study, thesubject device non-polar and polarextracts were determined to be non-irritating.Identical
SensitizationUnder the conditions of the study, thesubject device non-polar and polarextracts were determined to benon-sensitizing.Under the conditions of the study, thesubject device non-polar and polarextracts were determined to be non-sensitizing.Identical

Note 1:

Although the "Middle layer", "Nose clip", "Ear Loops" and "Specification and Dimension" of subject device is slightly difference with predicate device, it meets the requirement standard ISO 10993. The differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.

Note 2:

Although the "Particulate Filtration Efficiency" and "Differential Pressure" of subject device is with predicate device, it meets the requirement of essential performance standard ASTM 2100. The differences between the predicate device and subject device will not affectiveness of the subject device.

7. Summary of Non-Clinical Tests Performed

Test PerformanceTest methodPass criteriaTest results /Verdict
Bacterial filtrationefficiencyASTM F2101-14 Standard Test Method forEvaluating the Bacterial FiltrationEfficiency (BFE) of Medical Face MaskMaterials, Using a Biological Aerosol ofStaphylococcus aureus according to ASTMF2100:2019$\geq 98%$99.65% / Pass

{6}------------------------------------------------

Differential pressure(Delta-P)EN 14683: 2019, Annex C Medical face masks -Requirements and test methods according toASTM F2100:2019<6.0 mmH2O/cm²3.72 mmH2O/cm² / Pass
Sub-micronparticulate filtrationefficiency at 0.1 µm ofPolystyrene LatexSpheresASTM F2299-03 Standard Test Method forDetermining the Initial Efficiency of MaterialsUsed in Medical Face Masks to Penetration byParticulates Using Latex Spheres according toASTM F2100:2019≥ 98%99.9% / Pass
Resistance to penetrationby synthetic blood,minimum pressure in mmHg for pass resultASTM F1862/F1862M-17 Standard TestMethod for Resistance of Medical Face Masksto Penetration by Synthetic Blood (HorizontalProjection of Fixed Volume at a KnownVelocity) according to ASTM F2100:2019Fluidresistantclaimed at160 mm HgFluid Resistantclaimed at 160 mmHg / Pass
Flame spread16 CFR Part 1610 Standard for theFlammability of Clothing according to ASTMF2100:2019Class 1Class 1 / Pass

8. Summary of Clinical Tests Performed

Not applicable.

9. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission K202139, the Medical surgical mask is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K093179.

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.