K Number
K202119

Validate with FDA (Live)

Date Cleared
2020-10-07

(69 days)

Product Code
Regulation Number
892.1560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization.

The AcuSee AS-P1000 system is intended to be used in a clinical setting.

Device Description

The AcuSee AS-P1000 system is a medical device that tracks instrument positioning during ultrasoundguided procedures to provide instrument guidance to the end user. With accessories attached to the transducer probe and needle or needle-like rigid device, the AcuSee AS-P1000 system identifies and tracks the probe and the device within the field of view through optical detection technology. By coupling positioning information with the ultrasound image, the projected instrument pathway is displayed to the user for guiding instrumentation. The AcuSee AS-P1000 system is an accessory to any compatible ultrasound machine and transducer probe.

AI/ML Overview

The provided text is an FDA 510(k) summary for the AcuSee AS-P1000 System. It describes the device, its intended use, and a comparison to a predicate device (Clear Guide ONE). However, it does not contain the detailed performance testing results required to fill out a table of acceptance criteria and reported device performance, nor does it provide information on sample sizes, ground truth establishment methods (especially for a test set), number of experts, adjudication methods, MRMC studies, or standalone performance.

The document mainly focuses on demonstrating substantial equivalence based on technological characteristics and adherence to safety standards, rather than providing specific quantitative performance metrics of the device's accuracy in tracking and predicting needle paths.

Therefore, I cannot fulfill all parts of your request. I will indicate where information is missing.

Here's a breakdown of what can and cannot be extracted from the provided text:


Device: AcuSee AS-P1000 System
Intended Use: Augmenting the ultrasonic image of an interventional needle or needle-like rigid device and predicting its future path on a display.


1. Table of Acceptance Criteria and Reported Device Performance

This information is largely missing from the provided document. The document states that "Performance data was collected to demonstrate that the AcuSee AS-P1000 system achieves its intended function in a manner that is as effective as the predicate device" and that "The results of performance testing show that the AcuSee AS-P1000 system is as safe and as effective as the predicate device." However, no specific quantitative acceptance criteria or reported performance metrics (e.g., accuracy of needle tip prediction, tracking precision, latency) are provided.

Acceptance CriterionReported Device Performance
Not specified in the documentNot specified in the document
(e.g., Accuracy of needle tip prediction within X mm)(e.g., Achieved Y mm accuracy)
(e.g., Real-time tracking performance)(e.g., Real-time, but no specific metric given)

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: Not specified. The document mentions "Performance Testing" and "Bench testing" but does not detail the datasets or sample sizes used for these tests.
  • Data Provenance: The document generally refers to "Performance data" and "Bench testing." No information is provided regarding the country of origin of data or whether it was retrospective or prospective.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable/Not specified. The device tracks physical objects (needles) and their projected paths. "Ground truth" in this context would likely refer to highly precise physical measurements rather than expert consensus on imaging interpretation. The document does not mention the use of experts for establishing ground truth for the performance tests.

4. Adjudication method for the test set

Not applicable/Not specified. As the ground truth would likely be established by physical measurement systems, an adjudication method in the context of human expert review is not relevant here, and none is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. The document explicitly states "Clinical Testing: None." This device is a real-time intra-procedural guidance system, not an AI for image interpretation or diagnosis. Therefore, an MRMC study related to human reader improvement with AI assistance would not be relevant and was not conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, implicitly. The "Performance Testing" and "Bench testing" described would be a standalone evaluation of the algorithm's tracking and prediction capabilities against a known physical ground truth, without a human operator's influence on the measured performance of the tracking. However, no specific metrics from these tests are provided.

7. The type of ground truth used

Inferred: Likely physical measurements/engineering ground truth established through precise measurement systems during "Bench testing." The device tracks physical objects (needles and probes) and overlays their perceived position and predicted path onto an ultrasound image. The ground truth for such a system would typically involve highly accurate optical or mechanical tracking references.

8. The sample size for the training set

Not specified. The document does not mention any training data or training sets, as it describes an optical tracking system, not a machine learning model that typically requires a distinct training phase with labeled data. If there are internal models or calibrations, the data used for those is not disclosed.

9. How the ground truth for the training set was established

Not applicable/Not specified. As no training set is mentioned in the context of typical machine learning, this information is not provided.

{0}------------------------------------------------

October 7, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Weipeng (Suzhou) Medical Devices Co., Ltd. % Geetha Rao, Ph.D. Regulatory and Quality Representative Springborne Life Sciences 750 Menlo Avenue, Suite 200 MENLO PARK CA 94025

Re: K202119

Trade/Device Name: AcuSee AS-P1000 System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO Dated: July 24, 2020 Received: July 30, 2020

Dear Dr. Rao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202119

Device Name AcuSee AS-P1000 System

Indications for Use (Describe)

The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization.

The AcuSee AS-P1000 system is intended to be used in a clinical setting.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)X
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Submitter

NameWeipeng (Suzhou) Medical Devices Co., Ltd.Unit 207, Building B2
Address218 Xinghu St., Suzhou Industrial ParkSuzhou, 215123, P.R. China
Phone Number+86 0512-67311267
Fax NumberNA
Contact PersonBin Yang
Emailbinyang@vibronixinc.com
Phone Number+86 0512-67311267
Fax NumberNA
Date PreparedJuly 15, 2020

Device Information

Trade NameAcuSee AS-P1000 System
Common NameTracking or Guidance System
Product CodeIYO, Ultrasonic pulsed echo imaging system
Regulation21 CFR 892.1560, Ultrasonic pulsed echo imaging system
Device ClassII

Predicate Information

Device NameClear Guide ONE
510(k) NumberK141806

Device Overview

The AcuSee AS-P1000 system is a medical device that tracks instrument positioning during ultrasoundguided procedures to provide instrument guidance to the end user. With accessories attached to the transducer probe and needle or needle-like rigid device, the AcuSee AS-P1000 system identifies and tracks the probe and the device within the field of view through optical detection technology. By coupling positioning information with the ultrasound image, the projected instrument pathway is displayed to the user for guiding instrumentation. The AcuSee AS-P1000 system is an accessory to any compatible ultrasound machine and transducer probe.

Indications for Use

The AcuSee AS-P1000 system is indicated for augmenting the ultrasonic image of an interventional needle or needle-like rigid device, such as a biopsy needle, an aspiration needle, and for predicting its future path on a display, which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended to be used in procedures where ultrasound is currently used for visualization.

{4}------------------------------------------------

VibroniX

The AcuSee AS-P1000 system is intended to be used in a clinical setting.

Technological Characteristics

The AcuSee AS-P1000 System applies optical tracking technology to augment the ultrasonic image of an interventional needle or needle-like rigid device and for predicting its future path on a display and also show the image of a B-scan (or similar display) of a medical ultrasound imaging system. The relative position and trajectory of the needle is calculated using stereo-vision optical detection of markers on tracking mounts attached to the ultrasound probe and needle device.

FeatureSubject DeviceAcuSee AS-P1000 SystemPredicate DeviceClear Guide OneSE Comparison/Comment
510(k) NumberNAK141806NA
Product CodeIYOIYOSame
Regulation No.892.1560892.1560Same
Risk ClassIIIISame
RegulatoryPathway510(k)510(k)Same
Indications forUseThe AcuSee AS-P1000 Systemis indicated for augmenting theultrasonic image of aninterventional needle orneedle-like rigid device, such asa biopsy needle, an aspirationneedle, or ablation needle, andfor predicting its future pathon a display, which also showsthe image of a B-scan (orsimilar display) of a medicalultrasound imaging system.The device is intended to beused in procedures whereultrasound is currently used forvisualization.The Clear Guide ONE isindicated for augmenting theultrasonic image of aninterventional needle orneedle-like rigid device, such asa biopsy needle, an aspirationneedle, or ablation needle, andfor predicting its future pathon a display, which also showsthe image of a B-scan (orsimilar display) of a medicalultrasound imaging system.The device is intended to beused in procedures whereultrasound is currently used forvisualization.Identical
Intended UsersPhysicianPhysicianSame
UseEnvironmentClinical SettingClinical SettingSame
SterilityNon-SterileNon-SterileSame
MovabilityMain system mounted onmovable cartMain system mounted onmovable cartSame
FeatureSubject DeviceAcuSee AS-P1000 SystemPredicate DeviceClear Guide OneSE Comparison/Comment
FundamentalTechnologyOptical DetectionOptical DetectionSame
Device FormFactorMain system contains camera,and PC with software onmovable cart, with power cordfor AC power,Separate tracking accessoriesconsist of tracking mountsMain system (Clear GuideCore) contains hardware andsoftware with power adapterand cord,Separate tracking accessoriesinclude camera (Clear GuideSuperProbe) and custominstrumentsSimilar primaryconcept with differentphysical configuration
DetectionMechanismCamera located at a distancetracks custom tracking mountson ultrasound probe andneedleCamera is directly mounted onultrasound probe and trackscustom needleSimilar concept, witha variant spatialgeometry
TrackingAlgorithmAutomaticAutomaticSimilar
Timing fortrackingReal-timeReal-TimeSame
InformationDisplayedOverlay of the detectedinstrument positioning andprojected path onto theultrasound imageOverlay of the detectedinstrument positioning andprojected path onto theultrasound imageSame
ReusabilityReusable main system withconsumable tracking mountsReusable main system andprobe with consumabletracking needleSimilar
InstrumentstrackedAny compatible needle-likeinstrumentRequires custom needle (notknown what kind of needle)Similar
Tracking ofinstrument tipAny compatible instrumentOnly with custom instrumentwith CLEAR GUIDE PercepTIPSimilar
Compatibilitywith probeTracking mount fits anycompatible probeOptical head with camera fitsany compatible probeSame functionality
Trackinginstruments inand out-of-planeWithin +/- 90 degWithin +/- 45 degSubject device haswider performancerange
FeatureSubject DeviceAcuSee AS-P1000 SystemPredicate DeviceClear Guide OneSE Comparison/Comment
Electrical SafetyandElectromagneticCompatibilityIEC 60601-1:2005+A1:2012 -Medical electrical equipment –Part 1: General requirementsfor basic safety and essentialperformanceIEC 60601-1-2:2014 - Medicalelectrical equipment - Part 1-2:General requirements for basicsafety and essentialperformance - CollateralStandard: Electromagneticdisturbances - Requirementsand tests.IEC 60601-1 3rd Edition:Medical Electrical Equipment;Part 1 General Requirementsfor SafetyIEC 60601-1-2 :2007 – Medicalelectrical equipment - Part 1-2:General requirements forsafety – Collateral standard:Electromagnetic compatibilitySimilar - Subjectdevice meets updatedstandard
PerformanceTestingBench testingBench testingSimilar
Animal TestingNoneNoneSame
Clinical TestingNoneNoneSame

Device Comparison Table

{5}------------------------------------------------

VibroniX

{6}------------------------------------------------

Substantial Equivalence Overview

The AcuSee AS-P1000 System has the same intended use and indications for use, as well as similar technological characteristics and principles of operation compared to the previously legally cleared predicate device: Clear Guide ONE (K141806, cleared in 2014).

The AcuSee AS-P1000 System and the predicate device both apply the same optical detection technology in a similar device form factor. The AcuSee AS-P1000 system overlays the detected instrument positioning data onto an existing ultrasound image through proprietary software algorithms, In the same manner as the predicate device. Both devices track targeted instrumentation and provide real-time visual guidance during a clinical procedure. Both devices have reusable main components and some consumable accessories. The AcuSee AS-P1000 System was tested to demonstrate compliance to the same applicable safety standards as the predicate device, but for updated versions of the standards.

The AcuSee AS-P1000 System uses the same primary concept for the detection mechanism but uses a different physical configuration for the camera and tracked targets, resulting in a different spatial geometry. This variation does not alter the fundamental mechanism and raises no new questions of safety. Performance data was collected to demonstrate that the AcuSee AS-P1000 system achieves its intended function in a manner that is as effective as the predicate device.

The AcuSee AS-P1000 system uses tracking mounts that are designed to fit ultrasound probes and instruments that are available on the market. Although specific models of probes (and the associated ultrasound machines with which they are used) and instruments are individually validated to be compatible, the mounts themselves are standardized and may be used with any compatible probe or instrument. Hence unlike the predicate device, the system does not require special patterns to be marked or printed to the shaft of the needle or other tracked instrument.

{7}------------------------------------------------

VibroniX

Although there are minor differences in technological characteristics of the AcuSee AS-P1000 System, they do not raise new issues of safety or effectiveness.

Summary of Performance Testing

The performance testing for the AcuSee AS-P1000 System includes software verification/validation testing, system verification/validation testing and testing to compliance standards for electrical and electromagnetic safety. Traceability has been documented between the system specification to verification/validation protocols. The results of performance testing show that the AcuSee AS-P1000 system is as safe and as effective as the predicate device.

The AcuSee AS-P1000 system complies with the following recognized consensus standards:

    1. IEC 60601-1:2005+A1:2012 (Medical electrical equipment Part 1: General requirements for basic safety and essential performance).
    1. IEC 60601-1-2:2014 (Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests).

Conclusion

The performance of the AcuSee AS-P1000 System is substantially equivalent to that of the Clear Guide ONE system and raises no safety or effectiveness issues and performs as well or better than the predicate device.

§ 892.1560 Ultrasonic pulsed echo imaging system.

(a)
Identification. An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.