K Number
K202094

Validate with FDA (Live)

Manufacturer
Date Cleared
2022-01-11

(532 days)

Product Code
Regulation Number
870.1425
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cardiovascular Suite 4.2.1 is a software program that is intended to aid trained healthcare in the quantitative analysis of vascular ultrasound images in adults, particularly for the measurement of the diameter and its changes on the brachial artery, the diameter and its changes on the carotid Intima-Media Thickness, and for carotid plaque analysis.

Device Description

The Cardiovascular Suite 4.2.1 is a software indicated for estimating early cardiovascular parameters by identifying and tracking the edges of the arteries by analyzing sequences of ultrasound images or single images of the longitudinal section of the vessel. The software consists of two main functional measurement modules: 1) the FMD-Studio for measuring Flow-Mediated-Dilation (FMD) of the brachial artery, by processing sequences of ultrasound image 2) The Carotid-Studio for measuring, by processing sequences of ultrasound images, the thickness of the carotid intima-media and the instantaneous carotid diameter that, associated with a pressure estimate, can provide arterial elasticity parameters. On single images, the software also provides a tool for Plaque Measurement and Quantification. The system is able to process previously recorded video files or directly process the video output of an ultrasound system in real time.

AI/ML Overview

Here's a summary of the acceptance criteria and the study details for the Cardiovascular Suite 4.2.1, based on the provided document:

Acceptance Criteria and Device Performance

The acceptance criteria are implied by the precision and accuracy metrics evaluated during the validation testing. The device is considered to meet these criteria if its performance falls within acceptable ranges for repeatability and agreement with an expert's measurements. The table below summarizes the reported device performance, which serves as the fulfillment of these implied criteria.

Table of Acceptance Criteria and Reported Device Performance

Measurement/MetricAcceptance Criterion (Implied by reported performance)Reported Device Performance (Cardiovascular Suite 4.2.1)
FMD Studio PrecisionCoefficient of Variation (CV)
Intra-observer intra-session FMD% variability≤ 10%9.9% ± 8.4% (reported as 10%)
Intra-observer inter-session FMD% variability≤ 13%12.9% ± 11.6% (reported as 13%)
Shear Rate measurement precision≤ 2.3% (specifically stated)2.3%
Carotid Studio PrecisionCoefficient of Variation (CV)
Intra-session Diameter variation≤ 2%2%
Intra-session IMT≤ 6%6%
Inter-session Diameter variation≤ 3%3%
Inter-session Diameter variation (cardiac cycle)≤ 12%12%
Inter-session IMT≤ 6%6%
Plaque geometric and statistics (single image)≤ 10%< 10% for each measurement
Carotid Analyzer Accuracy (IMT)Bias ± SD for IMT accuracy (specifically stated)0.006 ± 0.039mm
Carotid Analyzer Accuracy (Diameter)Bias ± SD for Diameter accuracy (specifically stated)0.060 ± 0.110mm
Carotid Analyzer Accuracy (Distension)Bias ± SD for Distension accuracy (specifically stated)0.016 ± 0.039mm
FMD Analyzer Accuracy (% diameter variation)Error ≤ 0.013% (specifically stated)0.013%

Study Information

  1. Sample sizes used for the test set and the data provenance:

    • FMD Studio Precision: 135 healthy volunteers. The data provenance is from seven Italian centers (prospective study, country of origin: Italy).
    • Carotid Studio Precision: 10 healthy volunteers. The provenance is not explicitly stated but implies a controlled, prospective study.
    • Compatible Ultrasound Devices Accuracy: Not explicitly stated as a numerical sample size but refers to "full set of images" and "each sub-set of images coming from each of the 15 devices." This suggests a test set composed of images from 15 different ultrasound devices, with a total of 120 carotid artery images (60 online/60 offline) and 120 brachial artery images (60 online/60 offline). The provenance is not explicitly stated.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Carotid Studio Precision: Two Clinical Operators (Opr 1 and Opr 2) performed measurements. Their specific qualifications (e.g., "radiologist with 10 years of experience") are not provided.
    • Compatible Ultrasound Devices Accuracy: An unspecified "expert" provided the gold-standard manual measurements. The specific qualifications of this expert are not provided.
  3. Adjudication method for the test set:

    • Carotid Studio Precision: For the "precision" part of the Carotid Studio module, Opr 1 and Opr 2 each measured vessels three times. Opr 1 repeated the analysis during a second session. This suggests direct comparison of individual measurements for variability, rather than a formal adjudication of a ground truth.
    • Compatible Ultrasound Devices Accuracy: The device's measurements were "compared with gold-standard measurements manually obtained by an expert." This implies a direct comparison rather than a consensus or adjudication among multiple readers/experts.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study is described where human readers' performance with and without AI assistance is evaluated. The studies focus on the precision and accuracy of the device itself or comparison between the device and expert/gold-standard measurements.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the performance testing described for both FMD Studio and Carotid Studio precision, and the accuracy comparisons for both modules, appear to be evaluating the standalone performance of the Cardiovascular Suite 4.2.1 software. For instance, the "accuracy of the following measurement of the CVS software was evaluated" and "The measurements were carried out by our software and compared with gold-standard measurements manually obtained by an expert." This indicates standalone algorithm performance evaluation.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • FMD Studio Accuracy: Evaluated on "synthetic image sequences."
    • Carotid Studio Accuracy: Agreement with "RF based gold-standard" (likely referring to RadioFrequency-based quantitative ultrasound measurements) and "manual measurements manually obtained by an expert".
    • Precision Studies: The ground truth for precision studies is the measurement itself, with the focus on consistency across repeated measurements by the same or different operators.
  7. The sample size for the training set:

    • The document does not explicitly state the sample size used for training the algorithm. It focuses on validation testing.
  8. How the ground truth for the training set was established:

    • Since the training set size is not provided, the method for establishing its ground truth is also not elaborated upon in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2022

Quipu S.R.L % Dallas Thomas Principal Medical Device Regulatory Consultant Thomas Regulatory Resolutions, Inc. 5613 Tiger Way Winter Garden, Florida 34787

Re: K202094

Trade/Device Name: Cardiovascular Suite 4.2.1 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: December 27, 2021 Received: December 29, 2021

Dear Dallas Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K202094

Device Name Cardiovascular Suite 4.2.1

Indications for Use (Describe)

The Cardiovascular Suite 4.2.1 is a software program that is intended to aid trained healthcare in the quantitative analysis of vascular ultrasound images in adults, particularly for the measurement of the diameter and its changes on the brachial artery, the diameter and its changes on the carotid Intima-Media Thickness, and for carotid plaque analysis.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

5.1 General Information

11 January 2022 Preparation Date:

Primary Submission Contact

Dallas L. Thomas, RAC, MHA, MPA, SSYB

Medical Device Regulatory Consultant Thomas Regulatory Resolutions Inc. 5613 Tiger Way Winter Garden, Florida USA 34787 +1 561 898 0260

dallas@thomasregulatory.com

Manufacturer / Submitter

Vincenzo Gemignani

President

Quipu S.R.L Moruzzi 1 56124 Pisa Italy VAT IT-01995110507 Tel: 39 328 3074229 Email: gemignani@quipu.eu

5.2 Regulatory Information

Subject Device NameCardiovascular Suite 4.2.1
Classification NamesComputer, Diagnostic, Programmable
Device ClassificationII
Common NameCardiovascular Suite 4.2.1
FDA Product CodeDQK
CFR References870.1425
Review PanelCardiovascular

{4}------------------------------------------------

5.3 Identification of Predicate Device

Quipu regards the Cardiovascular Suite to be substantially equivalent to the predicate K033266, which is the "Vascular Tools 5" consisting of modules: "Brachial Analyzers5", "Carotid Analyzer 5", as commercialized by Medical Imaging Applications LLC. Additionally we are including the reference devices as follows for further consideration for substantial equivalence: Reference Device: M'Ath Std K040686 by Intelligence in Medical Technologies and Reference Device: K090461, the IMAGE-ARENA 4.0 AND IA APPLICATIONS 2D CARDIAC PERFORMANCE ANALYSIS 1.0 by TOMTEC IMAGING SYSTEMS, GMBH

5.4 Subject Device Description

The Cardiovascular Suite 4.2.1 is a software indicated for estimating early cardiovascular parameters by identifying and tracking the edges of the arteries by analyzing sequences of ultrasound images or single images of the longitudinal section of the vessel.

The software consists of two main functional measurement modules:

  1. the FMD-Studio for measuring Flow-Mediated-Dilation (FMD) of the brachial artery, by processing sequences of ultrasound image

  2. The Carotid-Studio for measuring, by processing sequences of ultrasound images, the thickness of the carotid intima-media and the instantaneous carotid diameter that, associated with a pressure estimate, can provide arterial elasticity parameters. On single images, the software also provides a tool for Plaque Measurement and Quantification.

The system is able to process previously recorded video files or directly process the video output of an ultrasound system in real time.

The software is to be used only by trained healthcare professionals, such as laboratory technicians, nurses, doctors and/or ultrasonographers.

The use of the system for analyzing people with a distorted anatomy of the examined arterial tract is not recommended.

ર્સ્ટ Indications for Use

Per the current proposed product labeling, the indications for the Cardiovascular Suite 4.2.1 are quoted as follows:

Indications for Use

The Cardiovascular Suite 4.2.1 is a software program that is intended to aid trained healthcare practitioners in the quantitative analysis of vascular ultrasound images in adults, particularly for the measurement of the diameter and its changes on the brachial artery, the diameter and its changes on the carotid artery, the Carotid Intima-Media Thickness, and for carotid plaque analysis.

{5}------------------------------------------------

Contraindications

The Cardiovascular Suite 4.2.1 device is not intended for use as a test that provides a direct diagnosis of any cardiovascular disease. It is intended to supplement, not substitute, the physician's decision-making process for diagnosis and treatment. It should be used in conjunction with knowledge of the patient's history and other clinical findings. It is not intended for use in pediatric population.

Please note that the above indication is slightly reworded compared to the already cleared indications for the predicate Vascular Tools 5 and updated accordingly per current FDA Guidance. The indications for use statement also provides further clarification that is complementary to the cleared predicate indications for use.

{6}------------------------------------------------

Substantial Equivalence Discussion 5.6

Any modifications between the predicate device are provided in while the table below. The review of the indications for use and comparison characteristics provided in Error! Reference source not found. demonstrate that 4.2.1 is substantially equivalent o the predicate device, Vascular Tools 5. A reference device was also included.

Table 1. Substantial Equivalence Discussion

Please note a number of published article references are provided in this table.

ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
1510 (k)NumberK202094K033266K040686K090461N/ADifference in 510k numbers do notimpact safety or efficacy of theproduct.
2Device Name,ModelCardiovascular SuiteVascular Tools 5M'Ath® StdIMAGE-ARENA 4.0AND IA APPLICATIONS2D CARDIACPERFORMANCEANALYSIS 1.0N/ADifference in Names do not impactsafety or efficacy of the product.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
3ManufacturerQUIPU SRLMEDICALIMAGINGAPPLICATIONSLLCIntelligence inMedicalTechnologiesTOMTEC IMAGINGSYSTEMS, GMBHN/ADifferences do not impact safety orefficacy.
4CFR Reference870.1425870.1425CFR 892.2050892.2050N/ADifferences do not impact safety orefficacy.
5FDA ReviewPanelCardiovascularCardiovascularRadiologyRadiologyN/ADifferences do not impact safety orefficacy.
6FDA DeviceNameComputer, Diagnostic,ProgrammableProgrammablediagnostic computerSystem, Imageprocessing,RadiologicalUltrasonic pulsed dopplerimaging systemN/ADifferences do not impact safety orefficacy.
7FDA ProductCodeDQKDQKLLZLLZ, DQKN/ADifferences do not impact safety orefficacy.
8ClassIIIIIIIIIdenticalN/A
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M' Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
9Indications foruseThe Cardiovascular Suite is asoftware program for thequantitative analysis ofvascular ultrasound images,particularly for themeasurement of the diameterand its changes on thebrachial artery, the diameterand its changes on the carotidartery, the Carotid Intima-Media Thickness, and forcarotid plaque analysisVascular Tools 5"software programhas been developedto aid in quantitativeanalysis oflongitudinalvascular ultrasoundimages, particularlyto determinevascular diameterand intima-mediathickness, as well astheir changes asdepicted in brachialand carotid arterialultrasound images.M'Ath Stdsoftware is aWindows-basedapplicationprogram runningon a personalcomputer that isintended to aid thephysician in theorganization ofpatient datarelating to theultrasound imagesor video acquiredduring echo-cardiology examsof thecardiovascularsystem, includingthe patient'sThe Image-Arena PlatformSoftware is intended toserve as a datamanagement platform forclinical applicationpackages. It providesinformation that is used forclinical diagnosispurposes. The software issuited for stand-aloneworkstations as well as fornetworked multisysteminstallations and thereforeis an image managementsystem for research androutine use in bothphysician practices andhospitals. It is intended asa general purpose digitalmedical image processingSubstantially EquivalentDifferences do not impact safety orefficacy.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
characteristics.Additionally, thesoftware allows thephysician to makemeasurements todetermine theintima-mediathickness of thecarotid artery fromthe acquiredimages and storesthem with thepatient file.tool for cardiology. As theImage-Arena Applicationssoftware tool package ismodular structured,clinical applicationspackages with differentindications for use can beconnected. Echo-Comsoftware is intended toserve as a versatilesolution for Stress Echoexaminations in patientswho may not be receivingenough blood or oxygenbecause of blockedarteries. Image-Cornsoftware is intended forreviewing, measuring andreporting of DICOM dBtaof thi batdiah rodadlities.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
US and XA. -It can be'driven by Image;Arena orother third party platformsand is intended to launchother clinical applications.The clinical applicationpackage 2D CardiacPerformance Analysis isindicated for cardiacquantification based onechocardiographic data. Itprovides measurements ofmyocardial function(displacement, velocityand strain) that is used forclinical diagnosis purposesof patients with suspectedheart disease.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
10Anatomicalstructure of useCarotid and brachial arterySameCardio vascularbedCardiovascular BedIdentical to PredicateN/A - Cardiovascular Suite analyzescarotid and brachial artery. Thesesites are included in the predicate'sanatomical structure of use.
11CardiovascularPlaqueMeasurementfunctionincludedYesNoYesNoIdentical to ReferenceDeviceN/A-Cardiovascular Suite' tool forcarotid plaque includes geometricaland density measurementssubstantially equivalent to those ofthe reference device
12Flow-MediatedDilatation(FMD)functionincludedYesYesNoNoIdentical to PredicateN/A
13IMT - IntimaMediaThicknessYesYesYesYesIdentical to Predicate andReference DevicesN/A
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
functionincluded
14HardwarerequirementsLaptop or PC runningMac OS X, MicrosoftWindowsOperating System (OS)Laptop or PCrunning MicrosoftWindowsOperating System(OS)MicrosoftWindowsoperating systemMicrosoft Windowsoperating system on alaptop or PCModifiedDifferences do not impact safety orefficacy. Cardiovascular Suite can beused on Laptop or PC runningMicrosoft Windows OperatingSystem (OS) as the predicate.Cardiovascular Suite can be usedalso on Mac OS X, which presentsperformances comparable to theMicrosoft Windows OperatingSystem and is considered one of themost robust and user friendlyoperating system available on themarket. Therefore, this addition withrespect to the predicate device, doesnot result in differences in theperformance and safety of the device.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
15On-line / off-line operationOn-line and off-line.SameNot PubliclyAvailableOffline OnlyIdentical to PredicateN/A
16Supported dataformatsCardiovascular Suite workson A) Video formats:DICOM, AVI, MP4, MOVB)Image formats: DICOM,PNG, JPG, BMP, TIF . Thesedata formats are similar tothose of predicate andreference devices.DICOM, AVI, TIF,CRI, JPEG, BMP,RAW.AVI, JPEG, GIF,TIFF, BMP, PCX,PCD, TGA, EPS,IMG, DICOMDICOMModifiedDifferences in supported data formatsdo not impact safety or efficacy.
17CalibrationMandatory, manuallyperformedSamemanuallyperformed andautomaticallyDICOMNot Publicly AvailableIdentical to PredicateN/A- Calibration is mandatory for allthe systems
18Information foroperator in caseofWarning popups and includedin instructions for use.SameNot PubliclyAvailableNot Publicly AvailableIdentical to PredicateDifferences do not impact safety orefficacy.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
inappropriateuse
19Patient/ datastorageYes: Data archive availableSameSamesameIdentical to Predicate andReference DeviceN/A-Patients data and images arestored into the archive of thesoftware
20Data exportReporting in Microsoft Excel,RTF, RTFd, PDF, TABSeparated Values, CommaSeparated Values, HTMLformat.Reporting inMicrosoft Excel orSAS format.DICOM and pdfformatData can be exported inPDF format.ModifiedDifferences do not impact safety orefficacy. Cardiovascular Suitereporting in Microsoft Excel , PDF,TAB Separated Values, CommaSeparated Values. The additionalformat with respect to thepredicate/references can be read bythe same software program(Microsoft Excel and SAS).Moreover, the software architectureand more specifically the data exportmodule, has been tested andvalidated according to IEC 62304
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
and taking into consideration riskanalysis and reduction. Therefore,this addition with respect to thepredicate device, does not result indifferences in the performance andsafety of the device.
21ModulesTwo: one for FMD analysisand the other for CarotidanalysisSameNot PubliclyAvailableMultiple Modules basedon Publicly Available datafrom 510k Summary.Identical to Predicate andReference DeviceN/A-Two modules both forCardiovascular Suite and thepredicate device.
22Multi-userFunctionalityYes: protected user access bypassword to the softwareNoNot PubliclyAvailableYes, multi userfunctionality included.Substantially Equivalent toReference device.N/A- Comment: CardiovascularSuite provides a protected useraccess by password to the software.This feature has been introduced toguarantee safety in data access.Moreover, the multi-user access ofthe software has been tested andvalidated according to IEC 62304and taking into consideration risk
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M' Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
analysis and reduction. Therefore,this addition with respect to thepredicate device, does not result indifferences in the performance andsafety of the device.
23LicensingIssuanceDongle Licensekey and webbasedWeb basedLICENSE KEYNot Publicly AvailableSubstantially Equivalent toattributes of bothPredicate & ReferenceDevicesThe systems provide a licensingapproach web-based. In addition,Cardiovascular Suite can providelicense to the customer by dongle-key use. Dongle form is tightlycontrolled licensing. Moreover, thelicensing process of the software hasbeen tested and validated and it isfurther verified during the productionprocedure. Therefore, this additionwith respect to the predicate device,does not result in differences in theperformance and safety of the device.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
24FMD Inputimage modalityUltrasound image sequencesin B-mode modality;Ultrasound image sequencesin Duplex mode(simultaneous acquisition ofB-mode and Doppler)Ultrasound imagesequences in B-mode modality;Doppler FlowAnalysis notsupportedB-mode and colorflowNot IncludedSubstantially Equivalent toPredicateDifferences do not impact safety orefficacy. Cardiovascular Suitesupports B-mode as the predicate andreference device. In addition Dopplerflow analysis is also supported partof the predicate device.
25Measured/computedparametersFlow Mediated Dilation(FMD).Shear rate detectionFlow MediatedDilation (FMD)Not PubliclyAvailableNot Publicly AvailableModifiedBoth devices provide instantaneousbrachial diameter that can be used toevaluate Flow-Mediated-Dilation. Inaddition, FMD Studio, thanks to theDoppler Flow Analysis can provideShear rate estimation, which is notexpected to result in differences insafety and efficacy.
26Exam typeMeasurement performed onB-mode longitudinal sectionof the artery, which is imagedSameNot PubliclyAvailableNot Publicly Available,however based on theproduct website they doinclude multipleIdentical to predicateN/A- Measurement performed on B-mode longitudinal section of theartery, which is imaged above the
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
above the antecubital fossa inthe longitudinal planemeasurementfunctionalities.antecubital fossa in the longitudinalplane
27Detectionalgorithm /otheralgorithmsContour tracking algorithmbased on edge detectionoperator.Globally optimalgraph search borderdetection approachNot PubliclyAvailableNot Publicly AvailableModifiedDifferences do not impact safety orefficacy. Contour tracking approach.Both devices include an algorithmfor the automatic detection of theborder of the vessel, which is usedfor the computation of the vesseldiameter.The brachial analysis module ofVascular Tools 5 performs anautomatic detection of the edges ofthe vessel using a globally optimalgraph search border detectionapproach. [REF] The operator drawa region of interest (ROI) in whichthe best definition of the edges of thevessel is detectable on the first frameof the individual frame sequences or
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
movie clips. The analysis proceedsautomatically in supervision mode
thus allowing the operator to exclude
from the analysis poor-quality frames
or frames in which the edges of the
vessel are not correctly detected.
Similarly, using FMD Studio, the
approximate position of the edges of
the vessel is manually located before
starting the examination. After this
procedure, an automatic contour
tracking algorithm, based on a
mathematical edge detector operator
(first order absolute central moment
[REF]), locates and tracks the edges,
supplying information about quality
and time course of measurements in
real time. On completion of the
analysis, both methods automatically
generate a report, with all the
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
recorded measurements. The resultsobtained using the two systems havebeen evaluated in a comparison study[Faita et Al, REF] performed on 60participants, which showed anexcellent level of agreement in vesseledge location and subsequentdiameter evaluation. Therefore thedifference in the detection algorithmdoes not result in differences in theperformance and safety of the device.
28ECG gatingNot requiredNot requiredNot PubliclyAvailableNot IncludedIdentical to predicateN/A
29Output(displayed /stored data andresults)Instantaneous diameter chartMean diameter chartTime averaged positive ShearRate Charttime averaged positive shear1 - Instantaneousdiameter chart2 – Third orderpolynomial fit chart3 - EKG gatedNot PubliclyAvailableNot Publicly Available,but they do includemultiple data formatsbased on product details.ModifiedDifferences in output displays ofstored data and results, do not impactsafety or efficacy.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
ratetime averaged negative shearrateDoppler flowDoppler Waveformenvelop detection:
Mean diameter1 measure per frame
D Baseline [mm]: value of– Flow Integral
the baseline diameterAUC on Cycle
D Maximum [mm]: value of1 measure per frame
the maximum diameter– Flow Maximum in
during vasodilationCycle
D Recovery [mm]: diameter1 measure per frame
value in the recovery phase- Flow Average in
after vasodilationCycle
MD [%]: Flow Mediated
Dilation
FMDr [%]: Flow Mediated
Dilation calculated in relation
to the diameter in the
recovery phase after
vasodilation
SR Maximum [s-
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
1]:maximum value of theshear rateSR Baseline[s-1]:maximumvalue of the shear rateSR Area [dimensionless]:area under the curve of theshear rate,Time Average Wall ShearRate (computed usingDoppler flow velocitywaveform)GTN [%]: GTN InducedDilation
30Report reviewYesYesNot PubliclyAvailableYes report review in PDFis available.Identical to Predicate and areference device.N/A
31CAROTIDInput datamodalityUltrasound B-mode imagesequencesUltrasound B-modeimage sequencesNot PubliclyAvailableUltrasound B-mode imagesequencesIdentical to predicate and areference device.N/A - Cardiovascular Suite supportsB-mode as the predicate does
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
32Measurementtype / OutputIntima media thickness(IMT), diameter and elasticityof carotid arteries:mean and maximum IMTreported for each framediameter and IMTmeasurement in imagesequencesmean distension, cross-sectional compliancecoefficient, cross-sectionaldistensibility coefficient,Stiffness, Young moduleDoppler Flow VelocityIntima mediathickness (IMT) anddiameter of carotidarteries:mean and maximumIMT reported foreach framediameter and IMTmeasurement inimage sequences.IMT, Plaque(geometric andstatistics))andelasticityparametersIMT and Diameter ofCarotid Arteries.Additional measurementsare also available based onthe product website. Alsoincluded the doppler FlowVelocity.Modified betweenpredicate and referencedevices.Differences do not impact safety orefficacy. WITH RESPECT TO THEPREDICATE: The predicate andreference devices evaluate diameterand IMT, starting from a region ofinterest defined by the operator, in B-mode image sequences. In additiondiameter data from Carotid Studio,when combined with an estimate ofpressure, obtained through legallymarketed devices, provideparameters of arterial elasticity.Carotid intima-media thickness(IMT) and diameter elasticity aremarkers of structural and functionalvessel wall properties. Bothparameters have been found inpopulation-based studies to beassociated with cardiovascular riskfactors and prevalent cardiovascular
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
disease. The estimation of carotidelasticity by Carotid Studio has beenimplemented by adopting equationsand modalities suggested by opinionleaders guidelines . CardiovascularSuite measurement methods havebeen validated in terms of accuracyand reproducibility . CardiovascularSuite usability requirements aredefined to avoid confusion in resultsvisualization, highlighting IMT andclearly identifying all the parameters.Finally, it should be mentioned thatthe Cardiovascular Suite device isintended to supplement, notsubstitute, the physician's decision-making process; it should be used inconjunction with knowledge of thepatient's history and other clinicalfindings.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M' Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
As a conclusion, this technologicalcharacteristic does change the safetyor effectiveness of the device. WITHRESPECT TO THE REFERENCEDevice M'Ath: both systems evaluatearterial elasticity (distensibility) bycontour tracking algorithm andprovide a tool for plaque analysisbased on similar approaches.
33Exam typeMeasurement performed onB-mode longitudinal sectionof the artery.Samelongitudinal andcross sectionallongitudinal and crosssectionalIdentical to predicateN/A
34Detectionalgorithm /otheralgorithmsContour tracking algorithmbased on edge / borderdetection operator, incombination with patternrecognition approachGlobally optimalgraph search borderdetection approach[SON98, SON02]Not PubliclyAvailableNot Publicly AvailableSubstantially Equivalent toPredicateN/A
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
35Identificationof cardiac cycleYesYesYesYesIdentical to predicate andreference deviceN/A
36Identificationof carotidsectionYesYesYesYesIdentical to predicate andreference deviceN/A
37Report ReviewYesYesYesYesIdentical to predicate andreference deviceN/A
38Exam durationtime9 minutes brachial analysis/ 5minute carotid analysisSamesame for carotidanalysissame for carotid analysisIdentical to predicateN/A
39Target areaCarotid and brachial arterySameCarotidCarotid and other targetssuch as the heart.Identical to predicateN/A
40Results of theuse of thediagnosticdeviceResults are related to themeasurement of well-knownand accepted cardiovascularbiomarkers.SameNot PubliclyAvailableNot Publicly AvailableIdentical to predicateN/A
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M' Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
A clinical evaluation has beenconducted to review thesafety and effectiveness of theuse of these biomarkers forcardiovascular diseases riskstratification.Results can supplement thephysician's decision-makingprocess. They should be usedin conjunction withknowledge of the patient'shistory and other clinicalfindings.
41Carotidanalyzer -accuracyAgreement with RF basedgold-standard [JUM 2010BIANCHINI et Al]:Bland&Altman analysis (biasvalue ± standard deviation):IMT 0.006 ± 0.039mmSigned error meanIMT -0.007 ± 0.07mmSigned error maxIMT -0.07± 0.09mmNot PubliclyAvailableNot Publicly AvailableModifiedDifferences are not expected to haveimpact on safety or efficacy.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
Diameter 0.060 ± 0.110mm
Distension 0.016 ± 0.039mmD and Stiffnessaccuracy and
precision not found
42Carotidanalyzer -precision1) Intra-observer intra-sessionvariability: 7%±6% for IMT,2%±1% for diastolicdiameter, 11%±7% fordistension, 11%±7% for crosssectional compliancecoefficient and 13%±8% forcross sectional distensibilitycoefficient.2) Inter-observer intra-sessionvariability: 8%±8% for IMT,3%±2% for diastolicdiameter, 11%±10% fordistension, 12%±11% forcross sectional complianceIntra-subjectreproducibility forthe mean carotidIMThad a coefficient ofvariation of 3% witha mean absolutedifference of 0.02mm (SD 0.01).[PAT10]Not PubliclyAvailableNot Publicly AvailableModifiedCarotid Studio precision expressed ascoefficient of variation is 2% for thediameter, 11% for the diametervariation during the cardiac cycle,6% for IMT for intra-observer intra-session measurements and 3% for thediameter, 12% for the diametervariation during the cardiac cycle,6% for IMT for intra-observer inter-session measurements. As regardsplaque geometric and statistics datathe precision of the results expressedas coefficient of variation resultedlower than 10% for each
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
coefficient and $13%\pm12%$ forcross sectional distensibilitycoefficient. 3) intra-observerinter-session variability:$6%\pm6%$ for IMT, $3%\pm2%$ fordiastolic diameter, $12%\pm10%$for distension, $16%\pm11%$ forcross sectional compliancecoefficient and $17%\pm12%$ forcross sectional distensibilitycoefficient.Data from a single centerstudyPlaque alsoNB for compliance anddistensibility rep also dependson pressure estimationmeasurement obtained on a singleimage by the same operator.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M' Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
43FMD analyzer- accuracyAccuracy evaluated onsynthetic image sequences:error in assessing % diametervariation equal to 0.013%.Comparison with anon-automatedmethod (ProsoundSystem, JetPropulsionLaboratory, USA).Brachial ultrasoundtapes from 12patients undergoingendothelial functionassessmentThe correlationbetween the twoapproaches wasexcellent for boththe measurement ofabsolute diameters(r=0.995, P<0.001)and percentagediameter changesNot PubliclyAvailableNot Publicly AvailableModifiedAccuracy evaluated on syntheticimage sequences: error in assessing% diameter variation equal to0.013%.Please note as mentionedabove that: The results obtainedusing the two systems have beenevaluated in a comparison study[Faita et Al] performed on 60participants, which showed anexcellent level of agreement in vesseledge location and subsequentdiameter evaluation . Thereforedifferences do not change safety orefficacy profile doe to this low rateof difference.
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
(r=0.973, P<0.001).The BrachialAnalyzerdemonstrated nobias compared withthe other method[MAN02]
44FMD analyzer- precisionPrecision was assessed interms of reproducibility.Intra-observer intra-sessionvariability: 9.9±8.4%Intra-observer inter-sessionvariability: 12.9±11.6%Data from a multicenter study[REF].In the abovementionedexperimentationBrachial Analyzerdemonstratedexcellent precision(0.07 mm and 1.62percentage diameterchange). [MAN02]Not PubliclyAvailableNot Publicly AvailableModifiedFMD-Studio precision, expressed ascoefficient of variation, is 10% forintra-observer intra-sessionmeasurements and 13% for intra-observer inter-session measurementsof FMD%.
45IEC 62366Standard Applied to ProductSameNot PubliclyAvailableNot Publicly AvailableIdentical to PredicateN/A
ItemNo.DeviceCharacteristic.Proposed DeviceCardiovascular SuitePrimary PredicateDevice VascularTools 5Reference Device:M'Ath Std.Reference Device:K090461, the IMAGE-ARENA 4.0 AND IAAPPLICATIONS 2DCARDIACPERFORMANCEANALYSIS 1.0 byTOMTEC IMAGINGSYSTEMS, GMBHComparisonAnalysis: Identical/ SubstantiallyEquivalent /Modified / CannotBe Determined /Not ApplicableRationale as to whyModification orDifference from Predicateto Subject Device DoesNot Impact Safety andEffectiveness.
46IEC 62304Standard Applied to ProductSameNot PubliclyAvailableNot Publicly AvailableIdentical to PredicateN/A

{7}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{8}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{9}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{10}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{11}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{12}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{13}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{14}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{15}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{16}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{17}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{18}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{19}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{20}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{21}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{22}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{23}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{24}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{25}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{26}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{27}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{28}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{29}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{30}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{31}------------------------------------------------

K202094 Supplement 2

Traditional 510K

{32}------------------------------------------------

K202094 Supplement 2

Traditional 510K

Cardiovascular Suite 4.2.1

The modifications between the Cardiovascular Suite 4.2.1 and the Vascular Tools 5 predicate are examined in determined that each difference between the devices resulted in no impact to the performance, safety, or efficacy of Cardiovascular Suite 4.2.1 when compared to Vascular Tools 5.

{33}------------------------------------------------

5.7 Sterilization and Shelf Life

The Cardiovascular Suite 4.2.1 is a software only and therefore, does not require sterilization and has no patient contacting surface. Therefore no Sterilization or Shelf life testing was conducted.

5.8 Biocompatibility

The Cardiovascular Suite 4.2.1 is a software only and therefore, does not include any patient contactials. Therefore no biocompatibility testing was conducted.

5.9 Product Performance Testing - Software

Performance Software tests of Cardiovascular Suite 4.2.1 have been performed. The passing results from the performance bench testing demonstrate that Cardiovascular Suite 4.2.1 has met the functional requirements and is substantially equivalent to the predicate device.

5.10 Application of Standards

The following standards are applicable to the Cardiovascular Suite 4.2.1.

OriginalOrderStandards Applied No.Standards TitleEquivalent ISO / IECVersion
1EN ISO 13485:2016Medical devices - Qualitymanagement systems -Requirements for regulatorypurposesISO 13485:2016
2EN ISO 14971:2012Application of riskmanagement to medicaldevicesISO 14971:2007
3EN 1041:2008 +AMD1:2013Information supplied by themanufacturer of medicaldevicesN/A
4EN 62304:2006 +AMD1:2015Medical device software -Software life cycle processesIEC 62304:2006 +AMD1:2015
5EN 62366:2008Medical devices -Application of usabilityengineering to medicaldevicesIEC 62366:2007

Table 5: Applied Standards

{34}------------------------------------------------

Section 5: 510(k) Summary K202094 Supplement 2 Traditional 510K

Cardiovascular Suite 4.2.1

OriginalOrderStandards Applied No.Standards TitleEquivalent ISO / IECVersion
6EN ISO 15223-1:2016Medical devices - Symbolsto be used with medicaldevice labels, labeling andinformation to be supplied -Part 1: General requirementsISO 15223-1:2016

5.11 Validation Testing of Measurements for the Cardiovascular Suite 4.2.1

Validation testing of the measurements available in the Cardiovascular Suite 4.2.1 were performed and are summarized here.

Precision of the FMD Studio module

The precision of the FMD Studio was assessed in terms of repeatability of the measurement and tested in seven Italian centers, recruiting 135 healthy volunteers aged between 20 and 60. On each subject two different types of repeatability tests were performed:

i) intra-observer intra-session: in this case the same operator has analyzed the same subject twice within the same session, with a time interval of one hour;

ii) intra-observer inter-session: in this case the operator has analyzed twice within two different sessions, with a time interval of three days.

Precision of the Carotid Studio module

For the Carotid Studio module, sequences of the right/left common carotid artery of 10 healthy volunteers were analyzed during two separate sessions after 7 days. In the first session two Clinical Operators (Opr 1 and Opr 2) measured each vessel three times each. During the second session, only Opr 1 repeated the analysis. On each sequence of images, the following elements were calculated automatically through the Carotid Studio: Intima-media thickness (IMT), diastolic diameter (Dd), distension (ΔD), cross-sectional compliance (CC) coefficient and cross-sectional distensibility coefficient (DC).

Compatible ultrasound devices.

A further test was carried out to verify that the claimed accuracy of the CardioVascular Suite (CVS) software is maintained with claimed compatible ultrasound devices.

Scans of both the Carotid artery and the Brachial artery in longitudinal section were used for the and offline analysis setup were tested. For the online setup, the ultrasound images were acquired by the CVS software using

{35}------------------------------------------------

Section 5: 510(k) Summary K202094 Supplement 2 Traditional 510K

Cardiovascular Suite 4.2.1

the Epiphan AV.io HD frame grabber. For the offline setup, images were exported as multimedia file from the ultrasound device and imported in the CVS software by using an external USB flash memory.

Site of scanUS modality# of images (online / offline)
Carotid arteryB-mode120 (60/60)
Brachial arteryDual (B-mode + PWD)120 (60/60)

The accuracy of the following measurement of the CVS software was evaluated:

  • . Diameter (D)
  • Intima Media Thickness (IMT) ●
  • Time Average of the Doppler Flow Velocity (TADFV) .

The accuracy of the Shear Rate (SR) was estimated as the root mean square of the TADFV and the D measurements.

The measurements were carried out by our software and compared with gold-standard measurements manually obtained by an expert.

The analysis of accuracy was carried out on the full set of images. The agreement between the CVS software and the Gold Standard was tested by the analysis of the Coefficient of the standard deviation to the mean of the measurements). In addition, each sub-set of images coming from each of the 15 devices was singularly analyzed in order to better identify specific issues that might be related to the single ultrasound device.

Intra-session variability between measurements of the Doppler Flow Velocity was also evaluated.

RESULTS

FMD-Studio precision, expressed as coefficient of variation, is 10% for intra-session measurements and 13% for intra-observer inter-session measurements of FMD%. For the Shear Rate measurement, the estimated precision is 2,3%.

Carotid Studio precision expressed as coefficient of variation is 2% for the diameter variation during the cardiac cycle, 6% for IMT for intra-session measurements and 3% for the diameter, 12% for the diameter variation during the cardiac cycle, 6% for IMT for intra-observer inter-session measurements. As regards geometric and statistics data the precision of the results expressed as coefficient of variation resulted lower than 10% for each measurement obtained on a single image by the same operator.

{36}------------------------------------------------

These results are maintained for each compatible ultrasound device.

5.12 Conclusion

The subject device Cardiovascular Suite 4.2.1 is substantially equivalent to the predicate device. Cardiovascular Suite 4.2.1 shares a substantially equivalent design, indications for use and technology (i.e. features, materials, and principles of operation) with the predicate device and no new elements pertaining to change in safety or effectiveness have been identified.

§ 870.1425 Programmable diagnostic computer.

(a)
Identification. A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.(b)
Classification. Class II (performance standards).