K Number
K202072

Validate with FDA (Live)

Date Cleared
2020-08-26

(30 days)

Product Code
Regulation Number
882.4800
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Expandable Port is intended to provide access, allow visualization of the surgical field, and retraction of soft tissue during neurological cranial surgery.

The device is indicated for use in surgery during which subcortical access is required.

Device Description

The MindsEye Expandable Port is a single use, minimally invasive intracranial medical device designed to provide access and allow visualization to the surgical field while retracting subcortical soft tissues during neurological cranial surgery. Its design allows the user to place the port while collapsed in its minimal form for atraumatic access to the subcortical space. After placement, the port can be expanded radially to create a working channel to the target surgical location. The user is able to expand the port to the desired working channel diameter appropriate for the surgical plan then collapse the port upon surgical completion for safe removal. The working channel diameter may be adjusted as needed by the user throughout surgery.

The MindsEye device is expected to be used in conjunction with other standard, commercially available intracranial medical devices, such as an aspirator, micro/endo/exoscope, irrigation, electrosurgical systems/tools and/or hemostatic agents. In addition, MindsEye is compatible with existing optical navigation systems for accurate target acquisition and may be attached to other commercial retraction systems according to user preference.

The MindsEye device is made of biocompatible, medical grade materials (metals and polymers) with no software or electrical components. The MindsEye device and its components are provided sterile and not reusable.

MindsEye is comprised of the following components and is available in three lengths (Table 1).

  • Port with attached Sheath
  • Obturator and Guide
  • Optional Support System consisting of support clips, post, and legs
AI/ML Overview

The provided text is a 510(k) summary for the MindsEye™ Expandable Port, a medical device. It does not include information about acceptance criteria or a study proving the device meets those criteria, as one would typically see for performance claims related to diagnostic accuracy or clinical effectiveness.

Instead, this document focuses on demonstrating substantial equivalence to a predicate device based on non-clinical testing, biocompatibility, sterility, and shelf-life performance. The device is a physical tool used by surgeons, not an AI/ML-driven diagnostic or prognostic device that would typically have performance metrics like sensitivity, specificity, or AUC against a ground truth.

Therefore, many of the requested points are not applicable to the information contained in this 510(k) summary. I can, however, extract the information that is present regarding testing and validation.


1. A table of acceptance criteria and the reported device performance

The document provides a "Summary of Testing" rather than a formal table of acceptance criteria with numerical performance targets. The acceptance criterion for each test is simply "Pass," indicating that the device met the required standards for that specific test.

Non-Clinical TestMindsEye ComponentAcceptance Criteria implicitly met (Reported Performance)
Cytotoxicity: MEM elutionAll potential patient contacting componentsPass - non-cytotoxic
Sensitization: (Maximum)All potential patient contacting componentsPass - non-sensitizer
Irritation: (Intracutaneous Reactivity)All potential patient contacting componentsPass - non-irritant
Acute System ToxicityAll potential patient contacting componentsPass
Material Mediated PyrogenicityAll potential patient contacting componentsPass
Packaging and Shelf LifeAllPass (including shipping/distribution simulation, environmental conditions, aging, visual packaging inspection, bubble and seal strength packaging testing)
Sterilization (B&F testing, BI for SAL 10-6, Endotoxin testing)AllPass
Specification ReviewAllPass
Damage ResistanceSheathPass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document. The tests described are non-clinical (e.g., in-vitro, material durability, packaging), not clinical studies involving human patients or complex data sets. Therefore, typical "sample sizes" for test sets of clinical images or patient data, or data provenance (country, retrospective/prospective) are not applicable or detailed here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable/provided. The ground truth for these non-clinical tests would be established by standardized testing protocols and laboratory measurements, not by expert consensus from medical professionals.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable/provided. Adjudication methods like 2+1 or 3+1 are used for clinical studies involving multiple expert readers to resolve disagreements in interpretations (e.g., of medical images). The tests described are non-clinical, laboratory, and engineering tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable/provided. An MRMC study is relevant for evaluating the impact of AI on human diagnostic performance. The MindsEye™ Expandable Port is a physical surgical retractor, not an AI/ML diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable/provided. Standalone performance studies are for AI algorithms. The MindsEye™ Expandable Port is a physical surgical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the non-clinical tests, the "ground truth" implicitly refers to established scientific and engineering standards and validated test methodologies. For example, for biocompatibility, the ground truth for "non-cytotoxic" is determined by the results of standardized cytotoxicity assays within acceptable limits. For physical properties, it would be the pre-defined engineering specifications and ASTM/ISO standards.

8. The sample size for the training set

This information is not applicable/provided. This device is a physical product, not an AI/ML model that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable/provided. (See response to #8).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 26, 2020

Minnetronix Neuro, Inc. % Prithul Bom Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K202072

Trade/Device Name: MindsEye Expandable Port Regulation Number: 21 CFR 882.4800 Regulation Name: Self-Retaining Retractor For Neurosurgery Regulatory Class: Class II Product Code: GZT Dated: July 24, 2020 Received: July 27, 2020

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K202072

Device Name MindsEye™ Expandable Port

Indications for Use (Describe)

The Expandable Port is intended to provide access, allow visualization of the surgical field, and retraction of soft tissue during neurological cranial surgery.

The device is indicated for use in surgery during which subcortical access is required.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

ln accordance with 21 CFR 807.92 and FDA Guidance Document The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] dated July 28, 2014, Minnetronix Neuro provides this 510(k) Summary.

  1. 510(k) NUMBER: K202072

5.1 SUBMITTER INFORMATION

Company Name:Minnetronix Neuro, Inc.
Company Address:1625 Energy Park Drive, St. Paul MN 55108
Company Phone:(651) 917-4060
Company Facsimile:(651) 917-4066
Contact Person:Matt Adams, VP / GM
Date Summary Prepared:August 25, 2020

5.2 DEVICE IDENTIFICATION

Device Trade Name:MindsEye™ Expandable Port
Common Name:Neurological surgical port
Classification Name:Self-retaining retractor for neurosurgery
(21 CFR 882.4800, Product Code GZT)
Device Class:Class II
Advisory Panel:Neurological Devices Panel

5.3 IDENTIFICATION OF PREDICATE DEVICE

The predicate device for the MindsEye Expandable Port is the Nico® BrainPath® (K172433) with reference to the Vycor VBAS (K060973).

5.4 DEVICE DESCRIPTION

The MindsEye Expandable Port is a single use, minimally invasive intracranial medical device designed to provide access and allow visualization to the surgical field while retracting subcortical soft tissues during neurological cranial surgery. Its design allows the user to place

{4}------------------------------------------------

the port while collapsed in its minimal form for atraumatic access to the subcortical space. After placement, the port can be expanded radially to create a working channel to the target surgical location. The user is able to expand the port to the desired working channel diameter appropriate for the surgical plan then collapse the port upon surgical completion for safe removal. The working channel diameter may be adjusted as needed by the user throughout surgery.

The MindsEye device is expected to be used in conjunction with other standard, commercially available intracranial medical devices, such as an aspirator, micro/endo/exoscope, irrigation, electrosurgical systems/tools and/or hemostatic agents. In addition, MindsEye is compatible with existing optical navigation systems for accurate target acquisition and may be attached to other commercial retraction systems according to user preference.

The MindsEye device is made of biocompatible, medical grade materials (metals and polymers) with no software or electrical components. The MindsEye device and its components are provided sterile and not reusable.

MindsEye is comprised of the following components and is available in three lengths (Table 1).

  • Port with attached Sheath
  • Obturator and Guide
  • · Optional Support System consisting of support clips, post, and legs

Table 1: MindsEye Expandable Port Sizes

Model NumberPort Length (mm)
N-EXP1A-05252
N-EXP1A-06262
N-EXP1A-07777

5.5 INDICATIONS FOR USE

The Expandable Port is intended to provide access, allow visualization of the surgical field, and retraction of soft tissue during neurological cranial surgery.

The device is indicated for use in surgery during which subcortical access is required.

{5}------------------------------------------------

TECHNOLOGICAL CHARACTERISTICS COMPARISON 5.6

While there are expected technological differences due to the ability of MindsEve to be placed in collapsed form then expanded (dilatable working channel) as compared to the static size working channel offered by the predicate; the overall indication for use, principles of operation and placement as well as general size of the device are similar to the predicate and reference devices.

Similar Characteristics:

Like the predicate and reference devices, the MindsEye port is placed using an obturator for atraumatic introduction. Placement can be performed using a commercial optical navigation system often utilized for surgical planning. The obturator is then removed leaving a working channel to allow for visualization and surgical access within the brain tissue. MindsEye can accommodate a minimum inner diameter (i.e. working channel) of 6mm to a maximum diameter of 21mm. The inner diameter range of MindsEye (6 - 21mm) is smaller then or within the diameter range of the predicate and references devices (11 – 28mm). Similarly, MindsEye is available in three port lengths (52 – 77mm) to accommodate various surgical locations which is also within the length range of the predicate and references devices (30 - 95mm). The transparent sheath allows users to see cerebral vasculature, structures and tissue adjacent to the sheath and accommodates the visualization preferences (i.e. use of an endoscope, microscope, loupes, etc.) of the user during surgery. Finally, MindsEye is compatible with typical adjunct surgical instruments such as larger retraction systems, aspirators, forceps, bipolar and other electrosurgical and hemostasis devices.

Different Characteristics:

To allow for the radial dilation of the MindsEye port, it was designed with a flexible, compliant sheath and controlled mechanism to expand and contract as intended. This is the primary difference from the predicate and reference devices which have a hard, static size sheath. To achieve this, different materials of manufacture are used as compared to the predicate and reference devices, however comprehensive biocompatibility testing was performed to maintain the safety profile of the device. Characterization and performance testing were completed to provide assurance that the sheath of the MindsEye device is robust enough to withstand the normal conditions of brain surgery. A summary of the predicate and reference device comparison is provided in Table 2.

{6}------------------------------------------------

Table 2: Predicate Comparison

Predicate NICO BrainPathMindsEye Expandable PortReference Vycor VBAS
510(k)NumberK172433Not yet assignedK060973
IntendedUseTo provide for access andallow for visualization of thesurgical field during brainand spinal surgery.Provide access and allowvisualization to the surgicalfield and to retract soft tissueduring neurological cranialsurgery.Same as predicate
Device Components
SheathLengthsL: 50, 60, 75, 95mmL: 52, 62, 77mmTC model:L: 30, 50, 70, 90mm
SheathInnerDiameterOptionsID: 11, 13.5mmID Expandable: 6 - 21mmTC model:ID: 12, 17, 21, 28mm
ObturatorTip LengthL: 8mm and 15mmL: 9 mmUnknown
Principle of Primary Operation
Consists of an "obturator-like" component and a"sheath-like" componentwhich are assembled,inserted, and disassembledto provide corridor accessSame as PredicateSame
Operation and Placement
Handheld and can beassisted by third-partynavigation (if desired)Same as PredicateSame
Shipping Configuration
Obturator and sheathpackaged and shippedseparately and paired duringsurgical caseThe obturator and sheath arepackaged and shippedtogether ready for useThe port and introducerare packagedassembled ready for use
Reusable or Single Patient Use
Single patient use andreusableSingle patient use onlySingle patient use only
Method of Sterilization
Disposables: GammaReusables:Autoclave/hydrogenperoxide gas plasmaEOReusables: NoneGammaReusables: None
Predicate NICO BrainPathMindsEye Expandable PortReference Vycor VBAS
BiocompatibilityDemonstrated based onexternally communicatingdevice in direct contact withtissue/bone/dentin for alimited durationSame as PredicateSame
Materials of ManufactureObturator: AluminumSheath: COCSheath - Chronoprene 5AGuide - ABSObturator – 303 SSSupport system – 303 SSUnknown
Cross Sectional Analysis of Obturator / SheathObturator / Sheathcombination has a circularcross sectionObturator / sheath (port)combination has a circularcross section when placedSheath (port) has a hexagonalcross section when dilatedVarious depending onmodel from circular toelliptical
General Shape of Obturator TipDistal end of obturator has aconical shape with arounded tip and no openingSame as PredicateVarious depending onthe model
Third-Party InstrumentationObturator componentinterfaces with third partyinstrumentsSame as PredicateSame
Navigation Locking MechanismIncluded as part of theobturatorSame as PredicateNone
Surface of SheathInner diameter andhorizontal proximal surfaceof the knurled ring aretexturedInner and outer surface ofsheath is non-reflective,transparent, and flexiblePort is transparent andnon-compliant
Proximal End of SheathUpdated knurled ring withconcave geometry to reduceoptical glare, still includesholes, slots and tabs.The proximal end has a non-reflective head and dial for portexpansion with tabs that canbe used to connect to thesupport systemUnknown

{7}------------------------------------------------

{8}------------------------------------------------

5.7 SUMMARY OF TESTING

A series of non-clinical tests, biocompatibility, sterility, and shelf life testing were performed to demonstrate that MindsEye met the applicable design and performance requirements to support a determination of substantial equivalence related to the safety and effectiveness of the MindsEye device as compared to the predicate and reference devices.

Non-Clinical TestMindsEye ComponentResult
Cytotoxicity: MEM elutionAll potential patientcontacting componentsPass - non-cytotoxic
Sensitization: (Maximum)All potential patientcontacting componentsPass - non-sensitizer
Irritation: (Intracutaneous Reactivity)All potential patientcontacting componentsPass - non-irritant
Acute System ToxicityAll potential patientcontacting componentsPass
Material Mediated PyrogenicityAll potential patientcontacting componentsPass
Packaging and Shelf Life• shipping/distribution simulation,• environmental conditions, aging, visualpackaging inspection,• bubble and seal strength packaging testing,• environmental and shipping simulationAllPass
Sterilization• B&F testing,• BI for SAL 10-6,• Endotoxin testingAllPass
Specification ReviewAllPass
Damage ResistanceSheathPass

Table 3: Non-Clinical Testing Summary

5.8 CONCLUSION

A review of the risk assessment, biocompatibility and other non-clinical testing, in addition to demonstrated compliance with recognized standards, have established that the MindsEye Expandable Port does not raise different questions of safety or effectiveness for the described intended use when compared to the predicate and reference devices. Therefore, the results of these analyses provide reasonable assurance that the MindsEye Expandable Port has a similar safety and effectiveness profile as compared to the predicate and supports a determination of substantial equivalence.

§ 882.4800 Self-retaining retractor for neurosurgery.

(a)
Identification. A self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery.(b)
Classification. Class II (performance standards).