K Number
K202005

Validate with FDA (Live)

Manufacturer
Date Cleared
2020-12-19

(151 days)

Product Code
Regulation Number
880.5570
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indications for Use: NovoFine® Plus 32 G 4 mm needles are intended for use with pen injector devices for the subcutaneous injection of drugs.

Device Description

NovoFine® Plus 32 G 4 mm needle is a sterile single use needle for subcutaneous injection of drugs with a pen injector device. Prior to giving an injection, the protective tab is removed from the outer needle cap of the single-use disposable needle. With the disposable needle remaining in the outer needle cap, it is then carefully screwed onto the injection delivery device until tight and then the needle outer and inner caps are removed. Use the needles as described in the instructions for use that comes with the pen-injector device and as instructed by the healthcare professional.

After the injection, the needle is removed from the skin. The needle is detached from the injection device and disposed of in accordance with national/local regulations. For each subsequent injection, a new disposable needle must be used.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device, the NovoFine® Plus 32G 4 mm needle, and details why it is substantially equivalent to a previously cleared predicate device (NovoFine® Plus 32G 4 mm ETW cleared under K133738). This type of submission focuses on demonstrating equivalence rather than establishing new performance criteria for novel devices.

Therefore, the document does not contain information about acceptance criteria for a novel device, nor a study proving a device meets such criteria. Instead, it focuses on demonstrating that minor changes to an existing, already cleared device do not introduce new safety or effectiveness concerns, and that the updated device performs similarly to the predicate.

Here's an analysis of the provided text in relation to your request:

  1. A table of acceptance criteria and the reported device performance

    • Not Applicable. The document provides a "Side by side comparison of the subject device and predicate device" (Table 1 on page 7). This table compares specifications like outer diameter, inner diameter, length, gauge, and material composition between the current device (K202005) and the predicate device (K133738). It also lists existing standards that apply to both. However, these are device specifications and standards compliance for an equivalent product, not new acceptance criteria established for a novel device and corresponding performance data from a specific study against those criteria.
    • For example, it states "Inner diameter: Minimum 0.146 mm" for the subject device and "0.145-0.16 mm" for the predicate. This is a specified parameter, not an acceptance criterion with a calculated performance output from a study.
  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not provided. The document states "Biocompatibility and performance tests have been performed and the results are in compliance with existing domestic and international standards" (page 8). It also mentions "functional compatibility according to ISO 11608-2:2012 with all leading pen injectors available on the market." However, it does not detail sample sizes, data provenance, or whether the testing was retrospective or prospective.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. This question typically pertains to studies involving expert review for diagnostic or screening devices where a "ground truth" is established by human experts (e.g., in medical image analysis). The NovoFine® Plus is a hypodermic needle; its performance is assessed through engineering and biocompatibility testing against established standards, not expert-adjudicated clinical outcomes.
  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. As explained above, this device and its clearance process do not involve expert adjudication methods typically seen in AI/diagnostic device studies.
  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a device for drug delivery, not a diagnostic or AI-assisted interpretation device, so MRMC studies are not relevant.
  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm, so "standalone algorithm performance" is not relevant.
  7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable. The "ground truth" for this device's performance would be compliance with manufacturing specifications, established engineering standards (like ISO 11608-2:2012 for needle-based injection systems, and ISO 10993-1 for biocompatibility), and functionality testing (e.g., flow rate, bond strength, sterility). This is not typically referred to as "ground truth" in the context of clinical outcomes or expert consensus.
  8. The sample size for the training set

    • Not Applicable. This is a physical device, not an AI model, so there is no "training set."
  9. How the ground truth for the training set was established

    • Not Applicable. As above, there is no "training set" or corresponding ground truth establishment process for this type of device.

In summary, the provided document is a 510(k) submission for substantial equivalence of a hypodermic needle. It describes product specifications and compliance with established ISO standards, but it is not a study that establishes new acceptance criteria for a novel device or provides detailed performance data against such criteria in the way your request outlines for AI or diagnostic devices. The closest information available is the comparison table of specifications between the subject device and the predicate device, implying that the acceptance criteria are met by conforming to these specifications and relevant ISO standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 19, 2020

Novo Nordisk Inc. Hiral Shah Associate Manager, Regulatory Affairs P.O Box 846 Plainsboro, New Jersey 08536

Re: K202005

Trade/Device Name: NovoFine Plus 32G Tip x 4 mm Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: November 23, 2020 Received: November 25, 2020

Dear Hiral Shah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) NA

Device Name NovoFine® Plus 32 G 4 mm

Indications for Use (Describe)

Indications for Use: NovoFine® Plus 32 G 4 mm needles are intended for use with pen injector devices for the subcutaneous injection of drugs.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

CONFIDENTIAL

Date: Version: Status: Page:

18 December 2020 | Novo Nordisk 1.0 Final 1 of 6

Traditional 510(k)

NovoFine® Plus 32G 4 mm needle

Section 5 Summary

Author Regulatory Affairs Novo Nordisk Inc.

{4}------------------------------------------------

Table of contents

Page

Table of contents
510(k) Summary
1 Submitter's Name and Address:
2 Name of Device:
3 Substantial Equivalence:
4 Device Description:
5 Intended Use:
6 Technological Characteristics:
6.1Non-Clinical Tests Performed:
6.2Clinical Tests Performed:
6.3Conclusion drawn from the non-clinical and clinical tests:

{5}------------------------------------------------

510(k) Summary

21 CFR 807.87(h)

As required by 21 CFR 807.92(a)

Date Prepared: December 18, 2020

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. The NovoFine® Plus 32 G 4 mm needle meets all applicable product and quality standards for hypodermic single lumen needle products.

1 Submitter's Name and Address:

Novo Nordisk Inc. 800 Scudders Mill Road Plainsboro, New Jersey 08536

Contact Person: Hiral Shah Palkhiwala Associate Manager, Regulatory Affairs Tel: 609-787-7603 Email: hpaw@novonrdisk.com

2 Name of Device:

Proprietary Name: Common or usual name: Classification:

Product Code:

NovoFine® Plus 32 G 4 mm Pen Needle Class II device; 21 CFR 880.5570, (hypodermic single lumen needle) FMI (hypodermic single lumen needle)

Substantial Equivalence: 3

Predicate Device: NovoFine® Plus 32 G 4 mm ETW cleared under K133738 on May 21, 2014.

The main purpose of this submission is to provide clarification to the intended use which is similar amongst the cited predicate device.

Additionally, following minor changes have been implemented between the subject device and the predicate NovoFine Plus.

{6}------------------------------------------------

a. Tighten specification for minimum inner diameter of the cannula

  • b. Change in hub and inner cap polypropylene sub-type
  • c. Addition of Contract Manufacturing facility
  • d. Addition of packaging configurations (4 needles (trade) and 6 needles (trade and sample))

The above changes do not affect intended use, indications for use, technological characteristics, the general type of materials, principle of operation, shelf life, and biocompatibility. Therefore, NovoFine® Plus that is the subject of this 510(k) is considered substantial equivalent to the NovoFine® Plus cleared on May 21, 2014, 2006 (K133738) the predicate device.

The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug, and Cosmetic Act, as amended and as supplied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statement related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent laws or their application by the court.

Device Description: ব

NovoFine® Plus 32 G 4 mm needle is a sterile single use needle for subcutaneous injection of drugs with a pen injector device. Prior to giving an injection, the protective tab is removed from the outer needle cap of the single-use disposable needle. With the disposable needle remaining in the outer needle cap, it is then carefully screwed onto the injection delivery device until tight and then the needle outer and inner caps are removed. Use the needles as described in the instructions for use that comes with the pen-injector device and as instructed by the healthcare professional.

After the injection, the needle is removed from the skin. The needle is detached from the injection device and disposed of in accordance with national/local regulations. For each subsequent injection, a new disposable needle must be used.

5 Intended Use:

NovoFine® Plus 32 G 4mm is intended for use with pen injector devices for the subcutaneous injection of drugs.

Comparison with Predicate Devices:

The subject device is considered substantially equivalent in device design, general type of materials, fundamental scientific technology and device performance as the predicate devices (K133738). The purpose of this submission is to provide clarification to the intended use which is similar amongst the cited predicate device. The table below provides a side by side comparison of the subject device compared to its predicate.

{7}------------------------------------------------

Table 1Side by side comparison of the subject device and predicate device
FeaturesNovoFine® Plus 32G 4 mmSubject deviceNovoFine® Plus 32G 4 mmPredicate device
ParameterK202005K133738
Intended UseIntended for use with pen injectordevices for the subcutaneousinjection of drugsIntended for use with pen injectordevices for the subcutaneousinjection of drugs, including insulin,GLP-1 analogs, and somatropin.
Product typeHypodermic single lumen needleHypodermic single lumen needle
Durability5 years from production date5 years from production date
BiocompatibilityEN ISO 10993-1EN ISO 10993-1
ReuseSingle useSingle use
LabelingSee Section 13: Proposed LabelingLabeling configuration:100 needles (trade)7 needles (sample)6 needles (trade and sample)4 needles (trade)See Section 13: Proposed LabelingLabeling configuration:100 needles (trade)7 needles (sample)
Specifications
Outer diameter, tip0.22-0.25 mm0.22-0.25 mm
Outer diameter, cylindricalpart0.25-0.27 mm0.25-0.27 mm
Inner diameterMinimum 0.146 mm0.145-0.16 mm
Length from hub4 mm4 mm
Gauge32G32G
Tip configuration1st and 2nd grinding and glassblasting1st and 2nd grinding and glassblasting
Hub/≤needle ≥ bondstrengthFmin = 11.2 NFmin = 11.2 N
Materials
HubPolypropylene (PP) with PE whitemaster batchColor: WhitePolypropylene (PP) with PE whitemaster batchColor: White
CannulaStainless steelAISI / SUS 304DIN-kurzname: X 5 CrNi 18 10*Stainless steelAISI / SUS 304DIN-kurzname: X 5 CrNi 18 9
GlueSingle component epoxy adhesiveColor: Light yellowSingle component epoxy adhesiveColor: Light yellow
Inner needle capPolypropylene (PP)Color: TransparentPolypropylene (PP)Color: Transparent
Outer needle capPolypropylene (PP)Color: Transparent whitePolypropylene (PP)Color: Transparent white
Materials
Sealing paperMedical grade paper for steamsterilizationMedical grade paper for steamsterilization
Lubricating oil for patientneedle endMedical grade siliconeMedical grade silicone
Lubricating oil for backneedle endMedical grade siliconeMedical grade silicone

{8}------------------------------------------------

6 Technological Characteristics:

The NovoFine® Plus needle is considered substantially equivalent to the predicate NovoFine® Plus disposable needle in intended use, technology, principles of operation, general type of materials and performance. Differences between the devices do not raise any significant issues of safety and effectiveness. For a detailed side by side comparison to the predicate needles cleared by the Agency, please see the table: Comparison to a legally marketed device in the substantial equivalence.

6.1 Non-Clinical Tests Performed:

The NovoFine® Plus needle will be manufactured in accordance with current Good Manufacturing Practices for Medical Devices. Biocompatibility and performance tests have been performed and the results are in compliance with existing domestic and international standards.

The clarification to the intended use of the subject pen needle device, does not introduce critical differences or new risks to the intended use of the device. The proposed clarification reflects current testing practice and aligns with compatibility summary provided in the submission. As part of the Quality System, NovoFine® Plus has been tested for functional compatibility according to ISO 11608-2:2012 with all leading pen injectors available on the market. This testing has shown connectivity to NovoFine® Plus and maintenance of the dose accuracy through the needle of all leading pen-injectors on the market covering different types of drugs. Thus, subject needle device meets requirements for its intended use and supports the proposed clarification.

List of all ISO standards used in non-clinical performance testing (sterility, biocompatibility, and performance):

  • ISO 10993-1: 2018 Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process.
  • ISO 17665-1: 2006 Sterilization of health care products Moist heat-Part 1: Requirements ● for the development, validation and routine control of a sterilization process for medical devices
  • ISO 11608-2: 2012 Needle-based injection systems for medical use Requirements and test ● methods -- Part 2: Needles
  • ISO 9626: 2016 Stainless steel needle tubing for the manufacture of medical devices ●

Clinical Tests Performed: 6.2

No clinical tests are required.

6.3 Conclusion drawn from the non-clinical and clinical tests:

Based on the design equivalency and the functional testing, Novo Nordisk has determined that the NovoFine® Plus is substantially equivalent to predicate NovoFine Plus , K133738, which are currently marketed in the United States. Differences between the devices do not raise any significant issues of safety and effectiveness.

§ 880.5570 Hypodermic single lumen needle.

(a)
Identification. A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.(b)
Classification. Class II (performance standards).