K Number
K201935

Validate with FDA (Live)

Date Cleared
2021-06-22

(344 days)

Product Code
Regulation Number
890.5650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Air Pressure Therapy System VU-IPC04B is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.

Device Description

Air Pressure Therapy System has a pump. The pump is connected to the dedicated cuff via the hose series, one cuff has 6 compression chambers. The pump compresses and inflates air into the chambers continuously to produce compression from the body tail end to body center and release the air after compression as one cycle process. Cycle time and pressure can be adjusted by purpose to avoid any discomfort.

Air Pressure Therapy System intermittent pneumatic compression could be used in family for improving blood circulation in the treated area, in exercise rehabilitation field for muscle relaxation before exercise and muscle rehabilitation after exercise to relieve muscle soreness.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called "Air Pressure Therapy System: VU-IPC04B". The FDA has determined the device is substantially equivalent to a legally marketed predicate device, "NormaTec Pulse and NormaTec Pulse Pro (K160608)".

Based on the provided text, the device in question is a Powered Inflatable Tube Massager (Product Code IRP), which falls under Class II medical devices.

Here's an analysis of the acceptance criteria and study proving the device meets them, according to the document:

1. A table of acceptance criteria and the reported device performance:

The document directly states: "Performance testing - clinical: Not Applicable." and "Performance testing - animal: Not Applicable." This indicates that for this specific type of device and its indications for use, no clinical or animal performance studies were required by the FDA or provided by the manufacturer to demonstrate meeting acceptance criteria for its therapeutic claims (temporary relief of muscle aches/pains, increased circulation).

Instead, the substantial equivalence hinges on:

  • Indications for Use: The subject device's indications for use are identical to the predicate device.
  • Technological Characteristics: The core technology (compressor and valve system for sequential inflation) is the same.
  • Non-clinical Performance Data: This is where the "acceptance criteria" lie in this context, focusing on safety and technical specifications rather than direct clinical efficacy.

Here's a table summarizing the relevant "performance characteristics" and acceptance based on this document:

Acceptance Criteria (Based on comparison to Predicate Device & Applicable Standards)Reported Device Performance (VU-IPC04B)Meets Criteria?
Indications for Use (Same as predicate)Temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.Yes (Identical)
Regulation Number (Same as predicate)21 CFR 890.5650Yes (Same)
Classification Name (Same as predicate)Massager, Powered Inflatable TubeYes (Same)
Regulation Class (Same as predicate)2Yes (Same)
Product Code (Same as predicate)IRPYes (Same)
Rx or OTC Use (Same as predicate)OTCYes (Same)
Pressure Range (Same as predicate)30-110mmHgYes (Same)
Mode of Compression (Same as predicate)SequentialYes (Same)
Safety Feature (Same as predicate)Power button on main unit allows user to stop therapy session at any time.Yes (Same)
Technology (Same as predicate)Compressor and valve system which sequentially inflates inflatable chambers.Yes (Same)
Standard Compliance (Non-clinical safety)Meets IEC 60601-1-11: 2015, IEC 60601-1-2: 2014, ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/R)2012, ISO 10993-5: 2009, ISO 10993-10: 2010, IEC 62133-2: 2017.Yes
Treatment time (Similar to predicate without raising safety concerns)5-99 mins (Predicate: 10 mins to continuous / up to 175 mins)Yes (Similar, with justification for no adverse impact)
Power source (Similar to predicate without raising safety concerns)110-240V; 50Hz/60Hz (Predicate: 100-240 VAC input)Yes (Similar, implied no adverse impact)
Power consumption (Similar to predicate without raising safety concerns)35W (Predicate: 14W)Yes (Similar, implied no adverse impact)
Dimensions/Weight (Similar to predicate without raising safety concerns)Various (Predicate: 4"x5"x9" / 3.6 lbs)Yes (Similar, implied no adverse impact)
Number of chambers (Similar to predicate without raising safety concerns)6 chambers (Predicate: 5 chambers)Yes (Similar, implied no adverse impact)
Work modes (Similar to predicate without raising safety concerns)Mode A (sequential), Mode B (gradual fill), Mode C (pulse), Mode D (all chambers inflate together) (Predicate: Sequential, Pulse)Yes (Similar, with justification for no adverse impact)

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable. No clinical or animal test set data was used for performance validation as per the document. The substantial equivalence is based on comparison to a predicate device and engineering/bench testing against recognized standards.
  • Data Provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No expert-adjudicated test set was used for performance assessment in this 510(k).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is an Air Pressure Therapy System, not an AI-powered diagnostic or assistive technology for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. The device is not an algorithm; it's a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • The "ground truth" for this 510(k) relies on the established safety and effectiveness of the legally marketed predicate device (NormaTec Pulse and NormaTec Pulse Pro), combined with compliance to recognized international and national electrical, safety, and biocompatibility standards (e.g., IEC 60601 series, ISO 10993 series) through non-clinical (bench) testing. There is no "ground truth" derived from patient outcomes or expert consensus for this specific device's clinical performance, as clinical studies were deemed unnecessary for substantial equivalence.

8. The sample size for the training set:

  • Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 22, 2021

Xiamen Weiyou Intelligent Technology Co., Ltd % Sam Lin Official Correspondent Shanghai Spica Management Consulting Co., Ltd. 609 Room, No.133 Shengang Avenue, Pudong New District Shanghai. 201306 China

Re: K201935

Trade/Device Name: Air Pressure Therapy System: VU-IPC04B Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: February 3, 2021 Received: June 2, 2021

Dear Sam Lin:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Heather Dean, PhD Assistant Director. Acute Injurv Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201935

Device Name

Air Pressure Therapy System: VU-IPC04B

Indications for Use (Describe)

Air Pressure Therapy System VU-IPC04B is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Type of submissionTraditional
Date preparedFebruary 3, 2021
Submission sponsor
Manufacturer nameXiamen Weiyou Intelligent Technology Co., Ltd.
AddressUnit 101-401, No.6 Xianghong Road, Xiang'an District, Xiamen,Fujian, China
Tel86-0592-6251545
Email17916220@qq.com
Contact personYong Sun
Device identification
Classification nameMassager, Powered Inflatable Tube
Trade nameAir Pressure Therapy System: VU-IPC04B
Device classificationClass II
Regulation number21 CFR 890.5650
PanelPhysical Medicine
Product codeIRP
Previous submissionsNone
Application correspondent
Company nameShanghai Spica Management Consulting Co., Ltd.
Address609 Room, No.133 Shengang Avenue, Pudong New District,Shanghai, China
Tel86-15626132181
Emailsam@spicagloble.com
Contact personSam Lin
Predicate and reference device information
SponsorNORMATEC INDUSTRIES, LP X
Trade/Device nameNormaTec Pulse and NormaTec Pulse Pro
510K numberK160608
Regulation number21 CFR 890.5650

{4}------------------------------------------------

Indications for use of the device

Air Pressure Therapy System VU-IPC04B is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.

Device description

Air Pressure Therapy System has a pump. The pump is connected to the dedicated cuff via the hose series, one cuff has 6 compression chambers. The pump compresses and inflates air into the chambers continuously to produce compression from the body tail end to body center and release the air after compression as one cycle process. Cycle time and pressure can be adjusted by purpose to avoid any discomfort.

Air Pressure Therapy System intermittent pneumatic compression could be used in family for improving blood circulation in the treated area, in exercise rehabilitation field for muscle relaxation before exercise and muscle rehabilitation after exercise to relieve muscle soreness.

Performance testing - clinical

Not Applicable.

Performance testing - animal

Not Applicable.

{5}------------------------------------------------

Subject devicePredicate deviceDifferences discussion
Device nameAir Pressure Therapy System: VU-IPC04BNormaTec Pulse and NormaTec PulsePro
510(k) numberK201935K160608N/A
ManufacturerXiamen Weiyou Intelligent Technology Co., Ltd.NORMATEC INDUSTRIES, LP XN/A
Product regulation21 CFR 890.565021 CFR 890.5650Same
Classification nameMassager, Powered Inflatable TubeMassager, Powered Inflatable TubeSame
Regulation class22Same
Product codeIRPIRPSame
Indications for useAir Pressure Therapy System VU-IPC04B isintended to temporarily relieve minor muscle achesand/or pains, and to temporarily increase circulationto the treated areas.The NormaTec Pulse and Pulse Pro isan air pressure massager intended totemporarily relieve minor muscleaches and/or pains, and to temporarilyincrease circulation to the treatedareas.Same
Rx or OTCOTCOTCSame
Pressure range30-110mmHg30-110mmHgSame
Treatment time5-99minsStays on until the user turns it off orcan be set up to turn off in a range of10 mins to continuous / Usercontrolled 10 minutes to 175 minutesor continuous -total time over 4SimilarThe treatment time of subjectdevice is smaller thanpredicate device (K160608)so the difference of treatment
segments.time would not raise adverselyimpact on safety andeffectiveness.
StandardIEC 60601-1-11,ES60601-1:2005/(R)2012C1:2009/(R)2012 and60601-1-2, ISO 10993-10,62133-2ANSI AAMI A1:2012,and A2:2010/(R)2012, IECISO 10993-5, IECES 60601-1, IEC 60601-1-2, IEC60601-1-11Similar
Mode of compressionSequentialSequentialSame
Power source110-240V; 50Hz/60Hz100- 240 VAC inputSimilar
Power consumption35W14WSimilar
Dimensions(WHD)22149CM4" x 5" x 9"Similar
Weight1.6KG3.6 poundsSimilar
PhotoImage: a black device with buttonsN/ASimilar
Size and appearance ofsleeves (leg part)Image: two black sleeves and a black round objectM: 91x65cmXL: 110x70cm(overlapping)XXL: 125x76cm(overlapping)Short: 14" x 43"Standard: 14" x 48"Tall: 14" x 60"Similar
Housing materialsMolded ABS enclosureMolded ABS enclosureSame
Number of chambers6 chambers5 chambersSimilar
Mode A:Sequential mode:Although the subject device
In this mode, only a single chamber is inflated at atime. Starting from the chamber ① and working upStarting with the distal chamber andprovides 4 kinds of work
to the chamber ⑥. Then the cycle repeats.progressing up the proximal chamber,mode, the Mode A and Mode
Mode B:each section compresses and theC are the same with predicate
In this mode, the chamber ① stays inflated. Itpressure gradually rises to thedevice (K160608), while the
gradually adds a chamber until all six chambers arepre-determined air pressure level,other work modes of subject
filled with air. Then the cycle repeats.then decompresses and the airpressure drops.device just have difference
Mode C:Once the top section decompresses,the cycle begins again. again.about inflatable order of the
In this mode, chamber ① inflates & deflates 4 times,then holds pressure; next chamber ② inflates &Image: Diagram of sequential modedifferent chambers. The
deflates 4 times, chamber ①② hold pressure;treatment pressure range are
chamber ③ inflates & deflates 4 times, chamberthe same under different work
Work mode②③ hold pressure, chamber ① deflates; chamber ④modes, so the difference of
inflates & deflates 4 times, chamber ③④ holdpressure range would not raise
pressure, chamber ② deflates; working as this wayadversely impact on safety
up to chamber ⑥. Then the cycle repeats.and effectiveness.
Note: Under this mode, for the FIRST CYCLE, itworks as mode B to warm up; from the 2nd cycle on,it works as PULSE Mode C.
Mode D:Normatec Pulse mode:Starting with the distal zone andprogressing up the proximal zone, onezone compresses and the pressuregradually rises to the pre-determinedair pressure level, holds the air ofprevious two zone, the other zones donot hold, until the last zone finished,deflate the all last three zone thenenter into next cycle.
In this mode, all chambers inflates together, anddeflates together. Then the cycle repeats.
A:In this mode, only a single chamber is inflated at a time. Starting from the chamber 1and working up to the chamber 6 Then the cycle repeats.Image: Diagram of chambers inflating one at a time, numbered 1 through 6B:In this mode, the chamber 1 stays inflated. It gradually adds a chamber until allsix chambers are filled with air. Then the cycle repeats.Image: Diagram of chambers inflating sequentially, numbered 1 through 6C:In this mode, chamber 1 inflates & deflates 4 times, then holds pressure;next chamber 2 inflates & deflates 4 times, chamber 12 hold pressure;chamber 3 inflates & deflates 4 times, chamber 23 hold pressure, chamber 1 deflates;chamber 4 inflates & deflates 4 times, chamber 34 hold pressure, chamber 2 deflates;working as this way up to chamber 6 Then the cycle repeats.Image: Diagram of chambers inflating and deflating, numbered 1 through 6, with a pulse iconNote: Under this mode, for the FIRST CYCLE, it works as mode B to warm up;from the 2nd cycle on, it works as PULSE mode C.D:In this mode, all chambers inflates together, and deflates together.Then the cycle repeats.Image: Diagram of chambers inflating and deflating together, numbered 1 through 6Image: Diagram of zones pulsing and holding
Safety featurePower button on main unit allows user to stoptherapy session at any timePower button on main unit allowsuser to stop therapy session at anytimeSame
TechnologyCompressor and valve system which sequentiallyinflates inflatable chambersCompressor and valve system whichsequentially inflates inflatablechambersSame

Table 6A: Summary of Comparison

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Performance Characteristic

The device meets all the applicable technical requirements of :

IEC 60601-1-11: 2015 - Medical electrical equirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for medical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-1-2: 2014 - Medical electrical equipments for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility

ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/R)2012 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

ISO 10993-5: 2009 - Biological Evaluation of Medical Device - Part 5: Tests for in vitro Cytotoxicity

ISO 10993-10: 2010 - Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization

IEC 62133-2: 2017 - Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems

Conclusion

Based on the indications for use, technological characteristics, and non-clinical performance data, "Air Pressure Therapy System: VU-PC04B (K20193)' is as safe, as effective, and performs as well as the legally marketed predicate devices, "NormaTec Pulse Pro (K160608)". Therefore, the subject device is substantially equivalent to the predicate device.

N/A