K Number
K201824

Validate with FDA (Live)

Device Name
Nerivio
Date Cleared
2020-10-23

(114 days)

Product Code
Regulation Number
882.5899
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Nerivio is intended for acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura.

Device Description

The Nerivio is a wearable, battery-powered device that is controlled by a mobile application. The system delivers low energy electrical pulses to the upper arm for 45 minutes per treatment, after which the device turns off automatically. The device hardware consists of an armband intended to be worn on a user's upper arm. The armband contains the electronic circuitry and the battery in a plastic storage case as well as two electrodes that are attached to the interior of the armband and placed against the user's skin. The device is operated and controlled via software that is installed and run on a user's personal mobile device such as a mobile phone or tablet. The device hardware communicates with the mobile application through a Bluetooth protocol. This mobile application software allows the user to control the stimulation intensity from 0 to 100% (representing intensity levels of 0- 40mA), to start or stop the stimulation program, and to view device status such as the device's connection state, a progress bar for stimulation duration, battery level, and user notifications. The patient is instructed to adjust the intensity to the strongest stimulation level below the perceived pain level. Treatments with Nerivio are intended to be self-administered by the user immediately after the onset of migraine headache or aura.

AI/ML Overview

The provided document describes the Nerivio device and its substantial equivalence to a predicate device (Nerivio Migra), primarily based on clinical studies rather than an AI/ML algorithm's performance. Therefore, many standard AI/ML acceptance criteria and study elements (like training set size, ground truth for training, expert adjudication, MRMC studies, standalone performance with metrics like sensitivity/specificity, or specific effect sizes of AI with human readers) are not present in this document.

However, I can extract the acceptance criteria and performance data related to the clinical studies conducted for device efficacy and safety as described in the provided text. The "acceptance criteria" here refer to the primary efficacy endpoints defined for the clinical trials, which demonstrate the device's effectiveness.


1. Table of Acceptance Criteria and Reported Device Performance

Note: The document outlines clinical study endpoints as measures of device performance, which serve as "acceptance criteria" for demonstrating efficacy and safety. The primary endpoint for efficacy was "pain relief at 2 hours post-treatment."

Acceptance Criteria (Primary Efficacy Endpoint)Reported Device Performance (Study 1)Reported Device Performance (Study 2)
Pain relief at 2 hours post-treatment50.0% (19/38 participants; CI95% 33.4-66.6%) achieved pain relief.59.3% (54/91 participants; CI95% 48.5-69.5%) achieved pain relief.
Additional Metrics (Secondary Endpoints)
Pain-free at 2 hours26.3% (10/38 participants; CI95% 13.4-43.1%) achieved pain-free.20.9% (19/91 participants; CI95% 13.0-30.6%) achieved pain-free.
Sustained pain relief for 24 hours83.3% (10/12 participants; CI95% 51.6-97.9%) of those with 2-hr pain relief.71.1% (32/45 participants; CI95% 51.6-97.9%) of those with 2-hr pain relief.
Disappearance of Nausea at 2 hours58.8% (10/17 participants; CI95% 32.9-81.6%).48.8% (20/41 participants; CI95% 32.8-64.8%).
Disappearance of Photophobia at 2 hours37.5% (9/24 participants; CI95% 18.8-59.4%).40.5% (30/74 participants; CI95% 29.2-52.5%).
Disappearance of Phonophobia at 2 hours50.0% (8/16 participants; CI95% 24.7-75.3%).44.6% (29/65 participants; CI95% 32.2-57.4%).
Improvement in functional ability at 2 hours46.7% (14/30 participants; CI95% 28.3-65.7%).59.4% (19/32 participants; CI95% 40.6-76.3%).
Improvement in functional ability at 24 hours72.7% (16/22 participants; CI95% 49.8-89.3%).50.0% (7/14 participants; CI95% 23.0-76.9%).
Device-related Adverse Events1 reported (1.8% of patients [1/42] or 0.003% of treatments [1/296]). Mild, resolved within 48 hours with drug.1 reported (1.0% [1/99]). Mild, resolved within 24 hours without medication.

2. Sample size used for the test set and the data provenance

The clinical studies acted as the primary "test set" for validating the device's performance.

  • Study 1 Sample Size:
    • 42 patients recruited initially.
    • 38 participants completed at least one treatment in response to a migraine.
    • 296 qualifying migraine headaches treated.
  • Study 2 Sample Size:
    • 97 participants completed at least one treatment.
    • 91 participants formed the final analysis set for test treatment.
    • 493 evaluable treatments of qualifying migraine headaches conducted.
  • Data Provenance: The document states that Study 2 was conducted at 9 sites in the USA. Study 1 was conducted at 2 sites, but the country of origin is not explicitly stated, although it's within the context of FDA submission, suggesting US or international sites adhering to US standards. Both studies were prospective, open-label, single-arm, multicenter studies.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The concept of "experts" establishing ground truth in the traditional sense of medical image analysis (e.g., radiologists) is not applicable here. The ground truth for efficacy was self-reported by the patients using an electronic diary application:

  • Pain scores: Self-reported by participants (e.g., severe or moderate pain to mild or none, or mild pain to none).
  • Absence/presence of associated migraine symptoms: Self-reported.
  • Functional disability: Self-reported.

The diagnostic criteria for chronic migraine (ICHD-3) were applied to select eligible participants, implying medical professionals (neurologists/headache specialists) were involved in patient selection. However, their role wasn't in establishing ground truth for individual treatments but in determining study eligibility.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

No specific adjudication method (like 2+1 or 3+1 for expert review) is mentioned for the patient-reported outcomes. Given that the data collected was patient self-reported via an electronic diary, it implies direct measurement of patient experience, not subject to expert adjudication of the outcome itself, although clinical staff would have monitored data collection.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not an AI/ML device, and no MRMC comparative effectiveness study involving human readers or AI assistance was conducted or reported. This device is a direct treatment device, not a diagnostic imaging or AI-assisted diagnostic tool.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not an AI/ML algorithm. The device, Nerivio, operates as a standalone therapeutic device that patients self-administer. Its "performance" refers to its direct therapeutic effect on migraine symptoms.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for the efficacy studies was primarily patient-reported outcomes data. This includes:

  • Self-reported pain levels (e.g., severe, moderate, mild, none).
  • Self-reported presence or absence of associated migraine symptoms (nausea, photophobia, phonophobia).
  • Self-reported functional ability.

8. The sample size for the training set

The concept of a "training set" in the context of an AI/ML algorithm is not applicable here. The device does not employ an AI/ML model that requires training data. The "training treatment" mentioned in Study 2 (97 participants completed at least one training treatment) refers to an initial treatment session to familiarize participants with the device, not an AI model training phase.

9. How the ground truth for the training set was established

Not applicable, as there is no AI/ML training set. The "training treatment" phrase refers to the initial use of the device by patients.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health and Human Services logo, and on the right is the FDA logo. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

10/23/2020

Theranica Bio-Electronics LTD. % Janice Hogan Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103

Re: K201824

Trade/Device Name: Nerivio Regulation Number: 21 CFR 882.5899 Regulation Name: Trunk and limb electrical stimulator to treat headache Regulatory Class: Class II Product Code: QGT Dated: September 28, 2020 Received: September 28, 2020

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Xiaorui Tang -S

Xiaorui Tang Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201824

Device Name

Nerivio

Indications for Use (Describe)

The Nerivio is intended for acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

Theranica Bio-Electronics LTD.'s Nerivio

Submitter

Theranica Bio-Electronics LTD. 4 Ha-Omanut St. Netanya, ISRAEL, 4250438

Phone: +972-72-3909755

Facsimile: +972-72-3909762

Contact Person: Dagan Harris

Date Prepared: October 22, 2020

Name of Device: Nerivio

Common or Usual Name: Nerivio

Classification Name: Trunk and limb electrical stimulator to treat headache

Regulatory Class: Class II

Product Code: QGT

Predicate Device:

Device Name: Nerivio Migra Manufacturer: Theranica Bio-Electronics LTD. 510(K) Number: DEN180059

Device Description:

The Nerivio is a wearable, battery-powered device that is controlled by a mobile application. The system delivers low energy electrical pulses to the upper arm for 45 minutes per treatment, after which the device turns off automatically.

The device hardware consists of an armband intended to be worn on a user's upper arm. The armband contains the electronic circuitry and the battery in a plastic storage case as well as two electrodes that are attached to the interior of the armband and placed against the user's skin.

The device is operated and controlled via software that is installed and run on a user's personal mobile device such as a mobile phone or tablet. The device hardware communicates with the mobile application through a Bluetooth protocol. This mobile application software allows the user to control the stimulation intensity from 0 to 100% (representing intensity levels of 0- 40mA), to start or stop the stimulation program, and to view device status such as the device's connection state, a progress bar for stimulation duration, battery level, and user notifications.

The patient is instructed to adjust the intensity to the strongest stimulation level below the perceived pain level. Treatments with Nerivio are intended to be self-administered by the user immediately after the onset of migraine headache or aura.

{4}------------------------------------------------

Intended Use / Indications for Use

The Nerivio is intended for acute treatment of migraine with or without aura in patients 18 years of age or older. It is a prescription use, self-administered device for use in the home environment at the onset of migraine headache or aura.

Summary of Technological Characteristics

Both the device and the predicate device function as remote electrical neuromodulation (REN) devices that utilize electro-stimulation that relieves migraine headache, using equivalent output parameters. The basic pulse structure is biphasic, with symmetrical interleaving phases and rectangular shape. The amplitude shift signal alternates between a nominal maximum and a nominal minimum of the amplitude signal. The maximal output current is 40mA. The assumed impedance is 1K ohm +/- 500 ohms.

CharacteristicSubject DevicePredicate DeviceComparison
Submission NumberK201824DEN180059N/A
Device NameNerivioNerivio MigraMinor modificationfor marketing
ManufacturerTheranica Bio-Electronics LTD.Theranica Bio-Electronics LTD.Same
Indications for UseThe Nerivio is intendedfor acute treatment ofmigraine with or withoutaura in patients 18years of age or older. Itis a prescription use,self-administereddevice for use in thehome environment atthe onset of migraineheadache or aura.The Nerivio Migra isindicated for acutetreatment of migrainewith or without aura inpatients 18 years of ageor older who do nothave chronic migraine.It is a prescription use,self-administereddevice for use in thehome environment atthe onset of migraineheadache or aura.Modified toremove limitationfor the treatmentof chronicmigraine
Prescription or OTCPrescriptionPrescriptionSame
Electrical waveformBiphasic rectangular,modulatedBiphasic rectangular,modulatedSame
Electrical output
Max output voltage500 Ω2 KΩ10 ΚΩ20V (measured)60V (measured)60V (measured)20V (measured)60V (measured)60V (measured)Same
Max output current500 Ω2 KΩ10 KΩ40 mA30 mA6mA40 mA30 mA6mASame
Maximum phase charge8µC8µCSame
Maximum averagecurrent (500Ω)1.76mA1.76mASame
Maximum currentdensity (peak) (500Ω)1.6mA/cm21.6mA/cm2Same
Maximum currentdensity (r.m.s) (500Ω)0.34mA/cm0.34mA/cmSame
Maximum averagecurrent density (absvalue) (500Ω)0.07mA/cm20.07mA/cm2Same
Maximum averagepower density (500Ω)1.41mW/cm21.41mW/cm2Same
Frequency
Primary phase duration[µSec]200200Same
Pulse Duration [µSec]400400Same
Electrode Area25 cm225 cm2Same
Treatment locationUpper armUpper armSame
Treatment duration45 min.45 min.Same
ReusableYesYesSame
# of treatments per onedevice12 treatments8 treatmentsAllows for moretreatments prior toreplacement.Does not raisedifferent questionsof safety andeffectiveness.
Power sourceLiMnO2 cell batteryLithium-Ion batteryDoes not raisedifferent questionsof safety andeffectiveness.
On/off buttonPower push-buttonON/OFF switchDoes not raisedifferent questionsof safety andeffectiveness.
DimensionsDevice - 12.0 x 7.5 x 1.5cmArmband - 48.0 x 10.0x 0.3 cmDevice+Armband =49.0 X 11.0 X 3.0 cmDoes not raisedifferent questionsof safety andeffectiveness.
WeightDevice - 50 grArmband - 33 gr125 gr (device +armband)Does not raisedifferent questionsof safety andeffectiveness.
Shelf life24 months9 monthsDoes not raisedifferent questionsof safety andeffectiveness.
Mobile ApplicationYesYesSame function for
Different GUI.Does not raisedifferent questionsof safety andeffectiveness.
BiocompatibilityYes.YesDifferent armbandcoating material.Does not raisedifferent questionsof safety andeffectiveness.
SterileNoNoSame
Processor controlYes.YesSame
Wireless controlYes.YesSame
Automaticoverload tripYes.YesSame
Automatic noload tripYes.YesSame
Automatic shutoffYes.YesSame
Stimulationintensity controlYes.YesSame

{5}------------------------------------------------

{6}------------------------------------------------

Table 1 provides a comparison between the key functional features of the Nerivio and predicate device.

CharacteristicSubject DevicePredicate DeviceComparison
Submission NumberK201824DEN180059N/A
Device NameNerivioNerivio MigraMinor modificationfor marketing
ManufacturerTheranica Bio-Electronics LTD.Theranica Bio-Electronics LTD.Same
Indications for UseThe Nerivio is intendedfor acute treatment ofmigraine with or withoutaura in patients 18years of age or older. Itis a prescription use,self-administereddevice for use in thehome environment atthe onset of migraineheadache or aura.The Nerivio Migra isindicated for acutetreatment of migrainewith or without aura inpatients 18 years of ageor older who do nothave chronic migraine.It is a prescription use,self-administereddevice for use in thehome environment atthe onset of migraineheadache or aura.Modified toremove limitationfor the treatmentof chronicmigraine
Prescription or OTCPrescriptionPrescriptionSame
Electrical waveformBiphasic rectangular,modulatedBiphasic rectangular,modulatedSame
Electrical output
Max output voltageSame
500 Ω20V (measured)20V (measured)
2 KΩ60V (measured)60V (measured)
10 KΩ60V (measured)60V (measured)
Max output currentSame
500 Ω40 mA40 mA
2 KΩ30 mA30 mA
10 KΩ6mA6mA
Maximum phase charge8µC8µCSame
(500Ω)
Maximum averagecurrent (500Ω)1.76mA1.76mASame
Maximum currentdensity (peak) (500Ω)1.6mA/cm21.6mA/cm2Same
Maximum currentdensity (r.m.s) (500Ω)0.34mA/cm0.34mA/cmSame
Maximum averagecurrent density (absvalue) (500Ω)0.07mA/cm20.07mA/cm2Same
Maximum averagepower density (500Ω)1.41mW/cm21.41mW/cm2Same
Frequency
Primary phase duration[µSec]200200Same
Pulse Duration [µSec]400400Same
Electrode Area25 cm²25 cm²Same
Treatment locationUpper armUpper armSame
Treatment duration45 min.45 min.Same
ReusableYesYesSame
# of treatments per onedevice12 treatments8 treatmentsAllows for moretreatments prior toreplacement.Does not raisedifferent questionsof safety andeffectiveness.
Power sourceLiMnO2 cell batteryLithium-Ion batteryDoes not raisedifferent questionsof safety andeffectiveness.
On/off buttonPower push-buttonON/OFF switchDoes not raisedifferent questionsof safety andeffectiveness.
DimensionsDevice – 12.0 x 7.5 x 1.5cmArmband - 48.0 x 10.0x 0.3 cmDevice+Armband =49.0 X 11.0 X 3.0 cmDoes not raisedifferent questionsof safety andeffectiveness.
WeightDevice - 50 grArmband - 33 gr125 gr (device +armband)Does not raisedifferent questionsof safety andeffectiveness.
Shelf life24 months9 monthsDoes not raisedifferent questionsof safety andeffectiveness.
Mobile ApplicationsoftwareYesYesSame function fortreatment control.Different GUI.Does not raisedifferent questionsof safety andeffectiveness.
BiocompatibilityYes.YesDifferent armbandcoating material.Does not raisedifferent questionsof safety andeffectiveness.
SterileNoNoSame
Processor controlYes.YesSame
Wireless controlYes.YesSame
Automaticoverload tripYes.YesSame
Automatic noload tripYes.YesSame
Automatic shutoffYes.YesSame
Stimulationintensity controlYes.YesSame

{7}------------------------------------------------

{8}------------------------------------------------

Table 1 – Comparison between subject and predicate devices

Performance Data

Non-Clinical Tests:

Nerivio non-clinical bench tests addressed verification and validation of the hardware and software. Device performance testing is summarized in Table 2.

{9}------------------------------------------------

TestTest description
Safety testsAccording to ANSI AAMI ES60601-1, IEC 60601-1-11 and IEC60601-2-10
EMC and radio testsAccording to EN IEC 60601-1-2
Wireless coexistenceAccording to ANSI C63.27
FCCAccording to 47CFRpart 15.347part 15.205part 15.207 part 15.209
BiocompatibilityAccording to ISO 10993

Table 2 – Nerivio non-clinical tests list

In addition, the company conducted internal bench tests to verify and validate the device battery's lifetime and safety, firmware testing, system testing, mobile application software testing and usability testing. In all instances, the Nerivio functioned as intended and expected.

Clinical Tests:

Two clinical studies of the Nerivio device in chronic migraine patients were performed to assess the Nerivio safety and clinical efficacy in the chronic migraine population. Specifically, it assessed the capability of the Nerivio device to relieve migraine headache pain and associated migraine symptoms in patients with chronic migraine. All studies were in compliance with 21 CFR parts 50, 56, and 812

The first study was a prospective, open-label, single arm, multicenter study conducted at 2 sites. 42 patients were recruited in this study. Eligible participants were adults (18–75 years old) who met the International Classification of Headache Disorders (ICHD-3) criteria for chronic migraine (at least 15 headache days a month, with at least eight days a month on which their headaches and associated symptoms meet diagnostic criteria for migraine).

Participants treated their migraine attacks at home for 4 weeks (treatment phase), within one hour from migraine symptom onset. Participants were instructed to avoid taking rescue medications prior or within two hours post-treatment. Pain scores, absence/presence of associated migraine symptoms, and functional disability were recorded at baseline. 2- and 24-hours post-treatment using the electronic diary application installed on the participants' smartphones.

The primary efficacy endpoint was the proportion of participants who achieved pain relief at 2 hours post-treatment in the test treatment, defined as improvement from severe or moderate pain to mild or none, or, improvement from mild pain to none.

38 participants completed at least one treatment in response to a migraine. A total of 296 qualifying migraine headaches were treated with Nerivio. Pain relief and pain-free at 2 hours were achieved by 50.0% (19/38; Cl95% 33.4-66.6%) and 26.3% (10/38; Cl95% 13.4-43.1%) participants, respectively. Pain relief was sustained for 24 hours in 83.3% (10/12; C195% 51.6-97.9%) of the participants (7 participants did not report pain level at 24 hours and were thus excluded from the analysis). Nausea, photophobia, and phonophobia disappeared at 2 hours in 58.8% (10/17; Cl95% 32.9-81.6%)), 37.5% (9/24; Cl95% 18.8-59.4%), and 50.0% (8/16; Cl95% 24.7-75.3%) participants, respectively. Furthermore, 46.7% (14/30; Cl95% 28.3-65.7%) participants experienced improvement in functional

{10}------------------------------------------------

ability at 2 hours and 72.7% (16/22; C/95% 49.8-89.3%) participants experienced improvement in functional ability at 24 hours (8 participants with missing data at 24 hours were excluded from the analysis). Consistency analyses across all attacks (excluding the training treatment) demonstrated that 73.7% (28/38) of the participants experienced pain relief in at least 50% of their treated attacks

One device-related adverse event was reported (1.8% of patients [1/42] or 0.003% of treatments [1/296]). This adverse event included bilateral tingling in the temples, disturbed and double vision. The event resolved within 48 hours following drug therapy. There were no device-related serious adverse events and none of the participants withdrew from the study due to device-related adverse events.

The second study was also a prospective, open-label, single arm, multicenter study conducted at 9 sites in the USA. Eligible participants were adults (18–75 years old) who met the International Classification of Headache Disorders (ICHD-3) criteria for chronic migraine (at least 15 headache days a month, with at least eight days a month on which their headaches and associated symptoms meet diagnostic criteria for migraine).

Following a 4 weeks "Run-in" phase, eligible participants were asked to treat their migraine attacks at home for 4 weeks with their optimal stimulation intensity, as soon as possible after migraine headache began and always within one hour of attack onset. Participants were instructed to avoid taking rescue medications prior or within the first two hours post-treatment. Pain scores, absence/presence of migraine associated symptoms, and functional disability were recorded at baseline, 2- and 24-hours post-treatment using the electronic diary application.

The primary efficacy endpoint was the proportion of participants who achieved pain relief at 2 hours post-treatment in the test treatment, defined as improvement from severe or moderate pain to mild or none, or, improvement from mild pain to none.

97 participants completed at least one treatment of a qualifying migraine headache (the training treatment) and 91 participants completed the test treatment with evaluable data at baseline and at 2 hours, forming the final analysis set (5 participants did not have qualifying migraine headaches and 3 participants had missing data in the test treatment at baseline or at 2 hours).

A total of 493 evaluable treatments (excluding the training treatment) of qualifying migraine headaches were conducted by the 91 participants included in the analyses, with an average of 5.4±2.8 evaluable treatments per patient per 4 weeks.

The primary, secondary, and exploratory endpoints of a single attack were conducted on the test treatment of the final analysis set of 91 participants. Pain relief and pain-free at 2 hours were achieved by 59.3% (54/91; Clss% 48.5-69.5%) and 20.9% (19/91; Clss% 13.0-30.6%) of the participants, respectively. Pain relief was sustained for 24 hours in 71.1% (32/45; C195% 51.6-97.9%) of the participants (9 participants did not report pain level at 24 hours and were thus excluded from the analysis). Nausea, photophobia, and phonophobia disappeared at 2 hours in 48.8% (20/41; Cl9s% 32.8-64.8%), 40.5% (30/74; Clos% 29.2-52.5%), and 44.6% (29/65; Clg5% 32.2-57.4%) participants, respectively.

Furthermore, 59.4% (19/32; Clos% 40.6-76.3%) of the participants experienced improvement in functional ability at 2 hours (participants with missing data at baseline or at 2 hours were excluded

{11}------------------------------------------------

from the analysis) and 50.0% (7/14: Clg5% 23.0-76.9%) of the participants experienced improvement in functional ability at 24 hours (participants with missing data at baseline or at 24 hours were excluded from the analysis).

Consistency analyses across all attacks (excluding the training treatment) demonstrated that 57.1% (52/91) of the participants experienced pain relief in at least 50% of their treated attacks.

One device-related adverse event was reported (1.0% (1/99) in which pain in the arm was felt following the use of the device on that arm. The device-related adverse event was mild, resolved within 24 hours without medication.

Based on the clinical performance that was documented in these studies, the Nerivio has a safety and effectiveness profile that is similar to the predicate device.

Conclusions

The Nerivio has the same intended use and similar indications (with the modification to remove the limitation for the treatment of chronic migraine patients), technological characteristics, and principles of operation as its predicate device. The minor differences in the indications do not alter the intended therapeutic use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the Nerivio and its predicate devices raise no different questions of safety, efficacy, or usability. Performance data demonstrate that the Nerivio is as safe and effective as the Nerivio Migra. Thus, the Nerivio is substantially equivalent to its predicate.

§ 882.5899 Trunk and limb electrical stimulator to treat headache.

(a)
Identification. A trunk and limb electrical stimulator to treat headache is a device intended to treat headache through the application of electrical stimulation anywhere on the body of the patient apart from the patient's head or neck through electrodes placed on the skin. The stimulation may be provided transcutaneously or percutaneously.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. This testing must include:
(i) Characterization of the electrical stimulation, including the following: Waveforms; output modes; maximum output voltage and maximum output current (at 500Ω, 2kΩ, and 10kΩ loads); pulse duration; frequency; net charge per pulse; and maximum phase charge, maximum current density, maximum average current, and maximum average power density (at 500Ω);
(ii) Characterization of the impedance monitoring system; and
(iii) Characterization of the electrode performance including the electrical performance, adhesive integrity, shelf-life, reusability, and current distribution of the electrode surface area.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate electromagnetic compatibility and electrical, mechanical, and thermal safety in the intended use environment.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) Instructions for use, including the typical sensations experienced during treatment;
(ii) A detailed summary of the electrical stimulation output, and the device technical parameters, including any wireless specifications;
(iii) A shelf life for the electrodes and reuse information; and
(iv) Instructions on care and cleaning of the device.