K Number
K201778

Validate with FDA (Live)

Date Cleared
2021-01-08

(192 days)

Product Code
Regulation Number
866.5830
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The i-STAT TBI Plasma test is a panel of in vitro diagnostic immunoassays for the quantitative measurements of glial fibrillary acidic protein (GFAP) and ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1) in plasma and a semiquantitative interpretation of test results derived from these measurements, using the i-STAT Alinity Instrument. The interpretation of test results is used, in conjunction with other clinical information, to aid in the evaluation of patients, 18 years of age or older, presenting with suspected mild traumatic brain injury (Glasgow Coma Scale score 13-15) within 12 hours of injury, to assist in determining the need for a CT (computed tomography) scan of the head. A 'Not Elevated' test interpretation is associated with the absence of acute traumatic intracranial lesions visualized on a head CT scan.

The test is to be used with plasma prepared from EDTA anticoagulated specimens in clinical laboratory settings by a healthcare professional. The i-STAT TBI Plasma test is not intended to be used in point of care settings.

Device Description

The i-STAT TBI Plasma cartridge is a multiplex immunoassay that contains assays for both ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1) and glial fibrillary acidic protein (GFAP). The assays test for the presence of these biomarkers in a plasma sample and yield a semi-quantitative test interpretation based on measurements of both UCH-L1 and GFAP in approximately 15 minutes. The i-STAT TBI Plasma cartridge is designed to be run only on the i-STAT Alinity instrument.

The i-STAT Alinity instrument is a handheld, in vitro diagnostic device designed to run only i-STAT test cartridges. The instrument is the main user interface of the i-STAT System and functions as the electro-mechanical interface to the test cartridge. The instrument executes the test cycle, acquires and processes the electrical sensor signals converting the signals into quantitative results. These functions are controlled by a microprocessor.

The i-STAT Alinity System is comprised of the i-STAT Alinity instrument, the i-STAT test cartridges and accessories (i-STAT Alinity Base Station, Electronic Simulator and Printer).

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the i-STAT TBI Plasma cartridge with the i-STAT Alinity System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in a table format. However, it presents clinical performance parameters for sensitivity, specificity, and negative predictive value (NPV). The implied "acceptance criteria" are derived from comparison to the predicate device and the clinical utility for reducing unnecessary CT scans.

Performance ParameterAcceptance Criteria (Implied)Reported Device Performance (Pivotal Study, N=1901)Reported Device Performance (Supplemental Fresh Specimen Study, N=88)
Clinical SensitivityComparable to predicate device and high enough to identify true positive cases of intracranial lesions.95.8% (95% CI: 90.6%, 98.2%)100.0% (95% CI: 88.3%, 100.0%)
Clinical SpecificityComparable to predicate device and sufficient to potentially reduce unnecessary CT scans.40.4% (95% CI: 38.2%, 42.7%)23.7% (95% CI: 14.7%, 36.0%)
Negative Predictive Value (NPV)High enough to confidently rule out the absence of acute traumatic intracranial lesions when the test is 'Not Elevated'.99.3% (95% CI: 98.5%, 99.7%)100.0% (95% CI: 80.2%, 100.0%) (Adjusted NPV at 6% prevalence: 100.0% (95% CI: 96.9%, 100.0%))
False Negative RateLow, especially for lesions requiring surgical intervention.4.2% (5/120). No FN for surgical intervention cases.0% (0/29)
False Positive RateTolerable given the clinical benefit of potentially reducing unnecessary CT scans.59.6% (1061/1781)76.2% (45/59)

Note: The document explicitly states that the device was deemed "substantially equivalent" to the predicate, and a "benefit-risk assessment was performed," suggesting that the performance metrics achieved were considered acceptable for its intended use and comparative to the existing predicate.


2. Sample Size Used for the Test Set and Data Provenance

  • Pivotal Study:
    • Sample Size: 1901 subjects (120 with positive CT, 1781 with negative CT).
    • Data Provenance: Prospectively collected and archived (frozen) plasma specimens. Subjects enrolled at 22 clinical sites in three countries: United States, Germany, and Hungary.
  • Supplemental Fresh Specimen Study:
    • Sample Size: 88 subjects (29 with positive CT, 59 with negative CT).
    • Data Provenance: Freshly collected plasma specimens. Subjects enrolled across 4 clinical sites of the TRACK-TBI study in the United States.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: At least two neuroradiologists, with a third neuroradiologist for adjudication if necessary.
  • Qualifications of Experts: Neuroradiologists (specific years of experience or subspecialty certification not detailed, but implied by the term "neuroradiologist").

4. Adjudication Method for the Test Set

The adjudication method used was consensus interpretation between two neuroradiologists, with adjudication by a third neuroradiologist if necessary. This is commonly referred to as a "2+1" or "multiple reader, with adjudication" method.


5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not described in this document. The clinical studies focused on the standalone diagnostic performance of the device itself (i-STAT TBI Plasma test interpretation) against a CT scan ground truth, not on evaluating human reader performance with or without the device. The device's output is an "interpretation of test results...to aid in the evaluation of patients...to assist in determining the need for a CT scan," suggesting it's designed to be used by a healthcare professional as an aid, but the study design doesn't directly measure the improvement of human readers through its use.


6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the clinical studies presented here (Pivotal and Supplemental) are effectively standalone performance studies for the i-STAT TBI Plasma test. The results (sensitivity, specificity, NPV) are reported for the device's interpretation ("Elevated" or "Not Elevated") directly against the CT scan ground truth, without measuring the impact of a human healthcare professional's subsequent decision-making. The device provides "a semiquantitative interpretation of test results derived from these measurements," which is then used "in conjunction with other clinical information, to aid in the evaluation of patients...to assist in determining the need for a CT scan." So, while it's an aid to a human, the performance metrics reported are for the device's output itself.


7. The Type of Ground Truth Used

The primary ground truth used for the clinical studies was the presence or absence of acute traumatic intracranial lesions visualized on a head CT (Computed Tomography) scan. This ground truth was established by consensus interpretation of neuroradiologists.


8. The Sample Size for the Training Set

  • Assay Cutoff Determination: A training set of 420 subjects (274 males and 146 females) with suspected mild TBI was used to determine the assay cutoffs for GFAP and UCH-L1.

9. How the Ground Truth for the Training Set Was Established

For the 420 subjects in the training set used to establish assay cutoffs:

  • Subjects had suspected mild traumatic brain injury (Glasgow Coma Scale score of 13-15).
  • Blood was drawn within 12 hours of injury.
  • A head CT scan determination was performed.
  • The ground truth would have been established by the head CT scan results (presence or absence of acute traumatic intracranial lesions), similar to the clinical study's ground truth, though the specific process of expert review for these 420 cases is not detailed beyond "head CT scan determination." It's reasonable to infer a similar process of expert radiologist interpretation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 24, 2022

Abbott Laboratories Brian Ma Sr. Regulatory Affairs Specialist 400 College Road East Princeton, New Jersey 08540

Re: K201778

Trade/Device Name: i-STAT TBI Plasma cartridge with the i-STAT Alinity System Regulation Number: 21 CFR 866.5830 Regulation Name: Brain trauma assessment test Regulatory Class: Class II Product Code: QAT

Dear Brian Ma:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter for your device cleared on January 8, 2021. Specifically, FDA is updating this SE Letter due to a typographical error in the clearance date, which was incorrectly dated January 8, 2020.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ying (Katelin) Mao, OHT7: Office of In Vitro Diagnostics and Radiological Health, 301-796-6635, ying.mao(@)fda.hhs.gov .

Sincerely,

Ying Mao-S

Ying (Katelin) Mao, Ph.D. Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo has a blue square with the letters FDA in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 8, 2020

Abbott Laboratories Brian Ma Sr. Regulatory Affairs Specialist 400 College Road East Princeton, New Jersey 08540

Re: K201778

Trade/Device Name: i-STAT TBI Plasma cartridge with the i-STAT Alinity System Regulation Number: 21 CFR 866.5830 Regulation Name: Brain Trauma Assessment Test Regulatory Class: Class II Product Code: OAT Dated: December 4, 2020 Received: December 7, 2020

Dear Brian Ma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{2}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K201778

Device Name

i-STAT TBI Plasma cartridge with the i-STAT Alinity System

Indications for Use (Describe)

The i-STAT TBI Plasma test is a panel of in vitro diagnostic immunoassays for the quantitative measurements of glial fibrillary acidic protein (GFAP) and ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1) in plasma and a semiquantitative interpretation of test results derived from these measurements, using the i-STAT Alinity Instrument. The interpretation of test results is used, in conjunction with other clinical information, to aid in the evaluation of patients, 18 years of age or older, presenting with suspected mild traumatic brain injury (Glasgow Coma Scale score 13-15) within 12 hours of injury, to assist in determining the need for a CT (computed tomography) scan of the head. A 'Not Elevated' test interpretation is associated with the absence of acute traumatic intracranial lesions visualized on a head CT scan.

The test is to be used with plasma prepared from EDTA anticoagulated specimens in clinical laboratory settings by a healthcare professional. The i-STAT TBI Plasma test is not intended to be used in point of care settings.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

The information in this 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

1.Submitter Information
OwnerAbbott Point of Care Inc.
400 College Road East
Princeton, NJ 08540
ContactPrimary: Brian Ma, PhD
Senior Specialist Regulatory Affairs
Phone: 613-688-5949
Secondary: Susan Tibedo
Director Regulatory Affairs
Phone: 609-213-8514

Date Prepared

January 6, 2021

2. Device Information

Proprietary Name i-STAT TBI Plasma Cartridge with the i-STAT Alinity System Glial fibrillary acidic protein (GFAP) Common Name Ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1)

ProductcodeDevice ClassificationnameRegulationNumberClassPanel
QATBrain trauma assessmenttest866.5830II (Specialcontrols)Immunology

{5}------------------------------------------------

3. Predicate Device

Proprietary Name Banyan Brain Trauma Indicator (BTI)

510(k) Number DEN170045

ProductcodeDevice ClassificationnameRegulationNumberClassPanel
QATBrain traumaassessment test866.5830II(Specialcontrols)Immunology(82)

4. Device Description

The i-STAT TBI Plasma cartridge is a multiplex immunoassay that contains assays for both ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1) and glial fibrillary acidic protein (GFAP). The assays test for the presence of these biomarkers in a plasma sample and yield a semi-quantitative test interpretation based on measurements of both UCH-L1 and GFAP in approximately 15 minutes. The i-STAT TBI Plasma cartridge is designed to be run only on the i-STAT Alinity instrument.

The i-STAT Alinity instrument is a handheld, in vitro diagnostic device designed to run only i-STAT test cartridges. The instrument is the main user interface of the i-STAT System and functions as the electro-mechanical interface to the test cartridge. The instrument executes the test cycle, acquires and processes the electrical sensor signals converting the signals into quantitative results. These functions are controlled by a microprocessor.

The i-STAT Alinity System is comprised of the i-STAT Alinity instrument, the i-STAT test cartridges and accessories (i-STAT Alinity Base Station, Electronic Simulator and Printer).

{6}------------------------------------------------

5. Intended Use Statement

The i-STAT TBI Plasma test is a panel of in vitro diagnostic immunoassays for the quantitative measurements of glial fibrillary acidic protein (GFAP) and ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1) in plasma and a semiquantitative interpretation of test results derived from these measurements, using the i-STAT Alinity Instrument. The interpretation of test results is used, in conjunction with other clinical information, to aid in the evaluation of patients, 18 years of age or older, presenting with suspected mild traumatic brain injury (Glasgow Coma Scale score 13-15) within 12 hours of injury, to assist in determining the need for a CT (computed tomography) scan of the head. A 'Not Elevated' test interpretation is associated with the absence of acute traumatic intracranial lesions visualized on a head CT scan.

The test is to be used with plasma prepared from EDTA anticoagulated specimens in clinical laboratory settings by a healthcare professional. The i-STAT TBI Plasma test is not intended to be used in point of care settings.

{7}------------------------------------------------

Similarities and Differences: System (Test and Instrument)
Feature orCharacteristici-STAT TBI Plasma Cartridge(Candidate)Banyan BTI (DEN170045)(Predicate)
Intended Use /Indications for UseThe i-STAT TBI Plasma test is apanel of in vitro diagnosticimmunoassays for the quantitativemeasurements of glial fibrillary acidicprotein (GFAP) and ubiquitincarboxyl-terminal hydrolase L1 (UCH-L1) in plasma and a semi-quantitativeinterpretation of test results derivedfrom these measurements, using thei-STAT Alinity Instrument. Theinterpretation of test results is used, inconjunction with other clinicalinformation, to aid in the evaluation ofpatients, 18 years of age or older,presenting with suspected mildtraumatic brain injury (Glasgow ComaScale score 13-15) within 12 hours ofinjury, to assist in determining theneed for a CT (computedtomography) scan of the head. A 'NotElevated' test interpretation isassociated with the absence of acutetraumatic intracranial lesionsvisualized on a head CT scan.The test is to be used with plasmaprepared from EDTA anticoagulatedspecimens in clinical laboratorysettings by a healthcare professional.The Banyan BTI is an in vitro diagnosticchemiluminescent enzyme-linkedimmunosorbent assay (ELISA). The assayprovides a semi-quantitative measurementof the concentrations of ubiquitinC-terminal hydrolase-L1 (UCH-L1) andglial fibrillary acidic protein (GFAP) inhuman serum and is used with theSynergy 2 Multi-mode Reader.The assay results obtained from serumcollected within 12 hours of suspectedhead injury are used, along with otheravailable clinical information, to aid in theevaluation of patients 18 years of age andolder with suspected traumatic brain injury(Glasgow Coma Scale score 13-15). Anegative assay result is associated withthe absence of acute intracranial lesionsvisualized on a head CT (computedtomography) scan.
Intended UserClinical LaboratoryClinical Laboratory
MeasurandsGFAP and UCH-L1GFAP and UCH-L1
Assay TechnologyEnzyme-linked immunosorbent assayEnzyme-linked immunosorbent assay
Assay FormatSingle use multiplex cartridge (bothassays (GFAP and UCH-L1) in onecartridge)Two test kits on separate 96-well microtiterplates - one for GFAP and one for UCH-L1
DetectionTechnologyElectrochemicalChemiluminescence
Specimen TypePlasmaSerum
Sample Volume20 µLGFAP kit: 150 µLUCH-L1 kit: 100 µL
Similarities and Differences: System (Test and Instrument)
Feature orCharacteristici-STAT TBI Plasma Cartridge(Candidate)Banyan BTI (DEN170045)(Predicate)
PreparationReady to UseManual preparation of reagents and plate.
AutomationTest and wash cycles are fullyautomated after sample loading stepManual sample loading, reagent mixing, aswell as transfer from incubation andwashing to read steps
AnalyticalMeasuring IntervalGFAP: 30 – 10,000 pg/mLUCH-L1: 200 – 3,200 pg/mLGFAP: 10 – 320 pg/mLUCH-L1: 80 – 2,560 pg/mL
Time to Result~ 15 minutes~ 4 hours
Reportable ResultQuantitative results for GFAP andUCH-L1 and semi-quantitativeinterpretationQuantitative results for GFAP and UCH-L1and semi-quantitative interpretation
Instrument Platformi-STAT AlinitySynergy 2 Multi-Mode Microplate Reader(BioTek Instruments, Inc.)
ControlsGFAP and UCH-L1 combined: 2levels (Control 1, Control 2)GFAP: 2 levels (Control 1, Control 2)UCH-L1: 2 levels (Control 1, Control 2)
CalibrationNo calibration needed by the enduser, calibration is pre-set duringmanufacture of the cartridgeCalibration curve generated by end userfor each run using six standards

6. Summary Comparison of Technological Characteristics

{8}------------------------------------------------

7. Performance Characteristics

1. Analytical Performance

a. Precision/Reproducibility

Semi-quantitative precision: The precision of the GFAP and UCH-L1 assays in the i-STAT TBI plasma cartridge with the i-STAT Alinity System was evaluated using plasma samples representing nine (9) levels of GFAP and seven (7) levels of UCH-L1 spanning the reportable range as well as the i-STAT TBI Controls (L1 and L2). This single-site study was based on CLSI document EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline-Third Edition. Each sample was tested for at least 20 days with two (2) runs per day and two (2) results per run for a total of 80 measurements per sample per cartridge lot. Runs were separated by a minimum of 2 hours. The plasma samples were made of pooled plasma from healthy normal donors spiked with recombinant GFAP and UCH-L1 antigens or spiked with

{9}------------------------------------------------

native antigens from pooled TBI patient plasma as shown in the tables below. The components of variability were estimated for GFAP and UCH-L1 and the precision results are shown in Table 1, Table 2, and Table 3.

Table 1: Estimate of GFAP Assay Precision
SampleNMean(pg/mL)RepeatabilityBetween-RunBetween-DayBetween-LotWithin-Laboratory
SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)
1238‡17.01.7610.4%0.915.4%0.613.6%1.317.7%2.4614.5%
2238‡30.82.498.1%0.000.0%0.000.0%0.521.7%2.558.3%
3238‡65.63.214.9%0.871.3%1.031.6%0.620.9%3.545.4%
4238*§104.93.373.2%2.082.0%0.000.0%1.501.4%4.244.0%
5238‡962.922.422.3%13.611.4%17.331.8%21.172.2%37.903.9%
61602029.539.181.9%26.301.3%19.100.9%94.894.7%107.695.3%
72403139.575.982.4%35.921.1%49.341.6%97.093.1%137.574.4%
8160*†5713.3143.962.5%42.680.7%65.721.2%170.293.0%236.364.1%
9159†7537.2129.571.7%133.301.8%35.890.5%187.572.5%266.513.5%

Table 1: Estimate of GFAP Assay Precision

*Additional GFAP result(s) was obtained due to cartridge re-run due to a UCH-L1 star-out

†one (1) outlier removed from analysis

キtwo (2) outliers removed from analysis

Sthree (3) outliers removed from analysis

Table 2: Estimate of UCH-L1 Assay Precision
RepeatabilityBetween-RunBetween-DayBetween-LotWithin-Laboratory
SampleNMean(pg/mL)SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)
1238‡72.54.886.7%1.732.4%0.000.0%3.935.4%6.509.0%
2239†300.115.125.0%5.942.0%0.000.0%15.545.2%22.487.5%
3240519.929.565.7%1.540.3%13.382.6%8.211.6%33.516.4%
4238‡1058.956.885.4%22.592.1%15.131.4%33.63.2%71.446.7%
5159*1639.691.575.6%8.720.5%15.741.0%28.461.7%97.566.0%
62402067.4111.095.4%54.992.7%46.012.2%15.000.7%133.066.4%
7239†2849.7145.45.1%100.563.5%0.000.0%15.160.5%177.446.2%

*one (1) result was unavailable due to star-out

*one (1) outlier removed from analysis

‡two (2) outliers removed from analysis

{10}------------------------------------------------

Table 3: Estimate of GFAP and UCH-L1 Assay Precision with i-STAT TBI Controls
SampleNMean(pg/mL)RepeatabilityBetween-RunBetween-DayBetween-Lot**Within-Laboratory
SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)SD(pg/mL)CV(%)
GFAP Assay
L1238*§196.79.945.1%2.691.4%2.251.1%5.702.9%11.986.1%
L2242*5153.8236.894.6%94.931.8%28.100.5%183.003.6%315.296.1%
UCH-L1 Assay
L1239†562.629.795.3%9.571.7%11.922.1%13.212.3%36.006.4%
L22401624.790.145.5%53.683.3%0.000.0%32.252.0%109.766.8%
  • Additional GFAP result was obtained due to cartridge re-run due to a UCH-L1 star-out

*one (1) outlier removed from analysis

‡two (2) outliers removed from analysis

§three (3) outliers removed from analysis

**This refers to precision estimates calculated between cartridge lots. A single lot of i-STAT TBI controls was used for this study.

Qualitative precision: The qualitative agreement of cartridge results relative to the expected sample result (mean) was evaluated for the 80 measurements per sample per cartridge lot for each assay from the semi-quantitative analysis above. The mean, total number of replicates, total number of elevated results, and % correct call for each sample level are presented in Table 4 for GFAP and Table 5 for UCH-L1.

Table 4: GFAP assay results for qualitative precision analysis
PlasmaSampleLotMean(pg/mL)Total Numberof ReplicatesTotal Number ofResults at orabove the cut-off% CorrectCall
1AA18.5800100%
B15.6800100%
C17.1800100%
2BD30.9805974%
E30.5805063%
F31.5806986%
3CD65.78080100%
E64.88080100%
F66.08080100%
4CA106.38080100%
B102.78080100%
C105.18181100%
5CA939.38080100%
B979.68080100%
C973.88080100%
6CG2096.88080100%
H1962.28080100%
7CA3050.28080100%

{11}------------------------------------------------

Table 4: GFAP assay results for qualitative precision analysis
PlasmaSampleLotMean(pg/mL)Total Numberof ReplicatesTotal Number ofResults at orabove the cut-off% CorrectCall
B3244.78080100%
C3123.68080100%
8CG5579.98080100%
H5832.88383100%
9cG7404.28080100%
H7685.78080100%
i-STAT TBI Controls
A203.48181100%
L1CB193.38080100%
C195.18080100%
L2CA5162.08282100%
B5335.28080100%
C4964.28080100%

A Below cut-off

^ Below cut-off
B Near cut-off (mean ± 25%)
© Above cut-off

^c Above cut-off

Table 5: UCH-L1 assay results for qualitative precision analysis
PlasmaSampleLotMean(pg/mL)Total Numberof ReplicatesTotal Number ofResults at orabove the cut-off% CorrectCall
1AD70.7800100%
E73.5800100%
F77.1800100%
2BD295.6800100%
E289.080199%
F317.5800100%
3CA525.18080100%
B509.18080100%
C525.68080100%
4CA1039.88080100%
B1039.08080100%
C1098.88080100%
5CG1617.97979100%
H1661.08080100%
6CA2078.78080100%
B2040.28080100%
C2083.28080100%
7CA2865.08080100%
B2819.58080100%
C2855.78080100%
i-STAT TBI Controls
L1CA556.88080100%
B575.28080100%
C552.68080100%
L2CA1601.28080100%
B1664.78080100%

{12}------------------------------------------------

Table 5: UCH-L1 assay results for qualitative precision analysis
PlasmaSampleLotMean(pg/mL)Total Numberof ReplicatesTotal Number ofResults at orabove the cut-off% CorrectCall
C1608.18080100%

A Below cut-off

B Near cut-off (mean ± 25%)

C Above cut-off

Semi-quantitative multi-site precision: The precision performance of the GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge on the i-STAT Alinity System was evaluated using seven (7) test materials representing six (6) levels of GFAP and five (5) levels of UCH-L1 at three (3) external clinical sites. At each site, each test material was tested once per day for five (5) days by two (2) different operators, with each operator using three (3) i-STAT Alinity Instruments. The plasma samples were made of pooled plasma from healthy normal donors spiked with recombinant GFAP and UCH-L1 antigens or spiked with native antigens from pooled TBI patient plasma as shown in the tables below. Thirty (30) replicates were obtained for the applicable assay analyzed for each test material at each of the three (3) sites for a total of 90 replicates for each test material. The estimates of GFAP and UCH-L1 precision are shown in Table 6 and Table 7.

{13}------------------------------------------------

ble 6: i-STAT TBI Plasma Multi-site Precision - GFAP (pg/mL) - 2 Test - - - - - - - - MoannL) - All Sites
etween-OperatoVithin-Site (TotalBetween-SiteOveral
Test MaterialN95% Cl%CVSD 95% CI)%CV95% Cli%CV95% Ci%CV95% Cl%CVəsində aid bitki növü. İstinadlar SED%CV
ಕ್ಕಾ9019.4 7, 262.60 14, 3.37ਟ13.4%0.00 0.00, 0.00)0.0%0.00 (0.00, 0.000.0%2.60 32, 2.9713.4%2.88 66, 12.7114.9%3.8820.0%
б9030.8 23, 361.76, 2.77 ,2.157.0%0.55 (0.41, 0.83)1.8%0.20 (0.13, 0.59)0.7%2.22 97, 2.53`7.2%2.51 45, 11.078.1%3.3510.9%
10B9066.2 54, 772.97 (2.44, 3.844.5%1.3%0.00 (0.00, 0.000.0%3.09 76, 3.54,ਟ4.7%4.00 (2.31, 17.646.0%5.057.6%
890150.1 37, 1684.48 (3.73, 5.643.0%1.4%1.41 88, 4.13)0.9%4.98 (4.38, 5.78)3.3%4.09 (2.37, 18.08)2.7%6.454.3%
દિવેલી કરવામાં આવેલું એક ગામના લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામમાં મુખ્યત્વે ખેત-ઉત064504.7 4257, 481388.75 72.82, 114.752.0%62.32 39.53, 79.29) 139.53, 79.29) 142.89, 18.2.14 142.89, 286.61.2%0.00 0.00, 0.00,0.0%103.03 0.40, 120.202.3%74.34 [42.95, 328.251.7%127.052.8%
7A909196.8 [8575, 10099193.85 161.79, 243.682.1%2.1%63.22 (39.17, 184.600.7%270.84 229.00, 33.562.9%167.65 (96.86, 740.241.8%318.533.5%
pooled normal donor plasma spiked with recombinant GFAPand UCH-L1 antiger
esamp
patient
TB
ledpool
3on.100
antigens
SllWITH
spiked:
maplasma
Controllers (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1OlCdo
al414no.
oledาดเ>0StD
TestMaterialNMean(Min, Max)Within-DayBetween-DayBetween-OperatorWithin-Site (Total)Between-SiteOverall
SD(95% CI)%CVSD(95% CI)%CVSD(95% CI)%CVSD(95% CI)%CVSD(95% CI)%CVSD(95% CI)%CV
5a9060.5(44, 83)6.03(5.10, 7.44)10.0%1.57(1.19, 2.38)2.6%2.69(1.67, 7.85)4.4%6.23(5.52, 7.18)10.3%0.00(0.00, 0.00)0.0%6.2310.3%
11c90297.9(257, 348)16.53(13.68, 21.08)5.6%5.01(3.79, 7.59)1.7%3.71(2.30, 10.84)1.2%17.28(15.36, 19.79)5.8%9.22(5.33, 40.71)3.1%19.586.6%
8a90538.0(467, 613)28.88(24.66, 35.04)5.4%9.52(7.19, 14.43)1.8%17.19(10.65, 50.19)3.2%30.41(26.63, 35.57)5.7%3.25(1.88, 14.36)0.6%30.585.7%
6a905143.2(4417, 6539)311.62(263.93, 382.79)6.1%58.76(44.39, 89.03)1.1%148.37(91.93, 433.25)2.9%317.11(280.91, 365.15)6.2%137.41(79.39, 606.70)2.7%345.606.7%
7a909538.7(7946, 11016)516.39(436.60, 636.06)5.4%107.70(81.36, 163.20)1.1%236.10(146.28, 689.40)2.5%527.50(467.50, 607.02)5.5%351.68(203.19, 1552.81)3.7%633.996.6%

pooled normal don't plasma spiked with room
pooled TBI patient and normal donor plasma

{14}------------------------------------------------

Qualitative multi-site precision: The qualitative agreement of cartridge results relative to the expected sample result (mean) was evaluated for the 90 measurements per test material for each assay from the semi-quantitative analysis above. The mean, total number of replicates, total number of elevated results, and % correct call for each test material are presented in Table 8 for GFAP and Table 9 for UCH-L1.

Table 8: Qualitative Precision Analysis - GFAP Assay - All Sites
TestMaterialMean(pg/mL)Total Numberof ReplicatesTotal Number ofGFAP Results at orabove the cut-off% Correct Call
5 A19.4900100%
9 B30.8906067%*
10 C66.29090100%
8 C150.19090100%
6 C4504.79090100%
7 C9196.89090100%

A Below cut-off: B Near cut-off (overall mean +/- 25%): ^ Above cut-off

*Test Material range includes cut-off. % Correct call based on test material overall mean for all sites.

Table 9: Qualitative Precision Analysis - UCH-L1 Assay - All Sites
TestMaterialMean(pg/mL)Total Numberof ReplicatesTotal Number of UCH-L1 Results at orabove the cut-off% CorrectCall
5 A60.5900100%
11 B297.9900100%*
8 C538.09090100%
6 C5143.29090100%
7 C9538.79090100%

A Below cut-off; B Near cut-off (overall mean +/- 25%); ^ Above cut-off

*Test Material range includes cut-off. % Correct call based on test material overall mean for all sites.

b. Linearity/assay reportable range

Linearity: The linearity of the GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge with the i-STAT Alinity System was established by testing plasma samples of varying antigen levels that range from below the lower limit of the reportable range to above the upper reportable range for both GFAP and UCH-L1. The study was designed based on CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures. The study was conducted using plasma samples spiked with either native antigens from a TBI patient sample exhibiting high levels of both GFAP and UCH-L1 or with recombinant GFAP and UCH-L1 antigens. Plasma samples of varying GFAP and UCH-L1 levels were prepared through proportional mixing of low and high antigen concentration samples.

The regression equation for the linear range of the GFAP assay is y=1.02x-6.7. The regression equation for the linear range of the UCH-L1 assay is y=1.04x-17.7. Regression summary of the results obtained for each assay in the i-STAT TBI Plasma cartridge (y-

{15}------------------------------------------------

Table 10: Linearity Across Reportable Range
AssayReportableRange(pg/mL)SlopeInterceptr2
GFAP30 - 100001.02-6.70.9985
UCH-L1200 – 32001.04-17.70.9869

axis) versus the expected values (x-axis) is provided in Table 10.

Recovery: The recovery evaluation was performed using data collected as part of the linearity studies. The % recovery was calculated using the means of the GFAP and UCH-L1 assay results and the expected values of the plasma samples as determined by the i-STAT Alinity instruments. The % recovery is presented in Table 11 for the GFAP assay and Table 12 for the UCH-L1 assay.

Table 11: GFAP Assay % Recovery
AntigenSourceNMean GFAP(pg/mL)GFAP Expected(pg/mL)% Recovery
Native Plasma63245.53245.5100.0%
62607.82599.6100.3%
61990.31953.6101.9%
51353.21307.7103.5%
6634.8661.795.9%
6346.7338.7102.4%
6169.5177.395.6%
698.996.5102.5%
649.456.188.1%
615.815.8100.0%
RecombinantPlasma510686.510686.5100.0%
69612.29618.999.9%
68581.28551.3100.3%
67630.57483.8102.0%
66117.76416.195.3%
65653.35348.6105.7%
64539.14281.0106.0%
63595.33213.4111.9%
62355.72145.8109.8%
61206.61078.3111.9%
6571.8608.694.0%
6294.5309.695.1%
5153.6160.295.9%
581.585.495.4%
639.447.782.6%
623.529.280.5%
616.419.982.4%

{16}------------------------------------------------

Table 12: UCH-L1 Assay % Recovery
AntigenSourceNMean UCH-L1(pg/mL)UCH-L1 Expected(pg/mL)% Recovery
Native Plasma62286.62286.6100.0%
61945.61844.1105.5%
61408.61401.6100.5%
51012.6959.2105.6%
6453.2516.787.7%
6296.4295.4100.3%
6183.1184.899.1%
RecombinantPlasma64950.84298.9115.2%
63598.33245.7110.9%
62492.82192.4113.7%
61216.91139.1106.8%
5592.1651.490.9%
6345.1368.693.6%
5213.1227.293.8%
5154.4156.598.7%

Hook Effect: The GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge on the i-STAT Alinity System were evaluated for high dose hook effect. The testing was conducted using plasma samples spiked to a high antigen level for each assay (>100,000 pg/mL). Each sample was tested to verify that the measured current response (nA)' is greater than that of a nominal GFAP target of 10,000 pg/mL and a nominal UCH-L1 target of 4,000 pg/mL. Hook effect was not observed for the GFAP and UCH-L1 assays as the current responses of high dose samples were significantly greater than 10,000 pg/mL for the GFAP assay and 4,000 pg/mL for the UCH-L1 assay.

Traceability, Calibration, and Reference Interval C.

Traceability and Calibration: The i-STAT System test for glial fibrillary acidic protein (GFAP) or ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1) measures GFAP and UCH-L1 amount-of-substance concentration in plasma (units of measure: pg/mL) for in vitro diagnostic use.

There are no internationally recognized standard reference materials available for either glial fibrillary acidic protein (GFAP) or ubiquitin carboxyl-terminal hydrolase L1 (UCH-L1). GFAP and UCH-L1 values assigned to i-STAT controls and calibration verification materials are traceable to Abbott Point of Care's working calibrators prepared using recombinant GFAP and UCH-L1 (expressed and purified from E. coli). The working calibrators are traceable to an in-house Reference Standard prepared from recombinant GFAP and UCH-L1 (expressed and purified from E. coli).

1 The i-STAT TBI Plasma cartridge measures assay current response in nanoamps (nA).

{17}------------------------------------------------

i-STAT System controls and calibration verification materials are validated for use only with the i-STAT System and assigned values may not be commutable with other methods.

Reference Interval: A reference interval study was conducted with a US-based general population. Plasma specimens from 225 self-declared healthy subjects between the ages of 18 and 79 years reporting no history of neurological disease within 1 year were tested with the i-STAT TBI Plasma cartridge with the i-STAT Alinity system to determine GFAP and UCH-L1 levels. Based on the test results, a 95% reference interval of an apparently healthy population of each biomarker was determined to be as follows:

Table 13: Reference Interval
BiomarkerNMean(pg/mL)SD(pg/mL)Median(pg/mL)Reference Interval(2.5th to 97.5thPercentile) (pg/mL)
GFAP2251916.2152 – 51
UCH-L12258142.47121 – 204

d. Detection Limit

Limit of Quantitation (LoQ): The LoQ was determined for the GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge in a study based on CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition. The testing was conducted on five (5) days using four (4) lots of i-STAT TBI cartridges with plasma from normal donors containing six (6) low levels of GFAP and UCH-L1. The LoQ for the GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge tested on i-STAT Alinity instruments was determined to be below the lower limit of the reportable range for each assay as shown in Table 14.

Table 14: LoQ Results for the GFAP and UCH-L1 Assays
AssayLower limit of thereportable range(pg/mL)LoQ(pg/mL)
GFAP3023
UCH-L120070

e. Analytical Specificity

Endogenous and Exogenous Interferences: The interference performance of the GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge on the i-STAT Alinity System was evaluated using plasma samples in a study based on CLSI EP07 ED3: Interference Testing in Clinical Chemistry, Third Edition. The effect of each substance was evaluated by comparing the performance of a control sample, spiked with blank solvent solution, with the test results from a sample spiked with the potentially interfering substance at the

{18}------------------------------------------------

toxic/pathological concentration based on CLSI EP37 ED1: Supplemental Tables for Interference Testing in Clinical Chemistry, First Edition, as applicable. A substance was identified as an interferent if the difference between the control and test samples was outside of a pre-determined acceptable range for each assay. Table 15 contains the list of potentially interfering substances tested for the GFAP and UCH-L1 assays and the interference results.

Test ConcentrationInterferenceInterference
Substanceμmol/Lmg/dLAssay(Yes/No)Results
Albumin150 g/L15 g/dLGFAPUCH-L1No
Bilirubin68440GFAPUCH-L1No
Bilirubin (conjugated)47540GFAPUCH-L1No
Hemoglobin10 g/L1000GFAPUCH-L1GFAPNo
Human anti-mouse antibody (HAMA) a> 160x bN/AUCH-L1YesHighest concentration tested where no interference observed: 40xTesting above this level showed decreased resultsc
Intralipid (Intralipid 20%)N/A4747GFAPUCH-L1GFAPNo
Rheumatoid Factor (RF) a1000 IU/mLN/AUCH-L1YesHighest concentration tested where no interference observed: 500 IU/mLTesting above this level showed decreased resultsc
Triglycerides a33.88 mmol/L3000GFAPUCH-L1No
Acetaminophen a1.324 mmol/L15.6GFAPUCH-L1No
Sodium Ascorbate2985.25GFAPUCH-L1No
Caffeine55610.8GFAPUCH-L1No
Clopidogrel a21.49 μg/mLGFAPUCH-L1No
Dopamine4.060.114GFAPUCH-L1No
Ethanol130 mmol/L600GFAPUCH-L1YesHighest concentration
Table 15: Interfering Substance TestingTest ConcentrationAssayInterference (Yes/No)Interference Results
Substanceμmol/Lmg/dLtested where no interference observed: 65 mmol/L d Testing above this level showed decreased results
Erythromycin1880.720GFAPUCH-L1NoNo
Nicotine5.970.00240GFAPUCH-L1NoNo
Metoprolol Tartrate a18.7128.06GFAPUCH-L1NoNo
Acetylsalicylic acid a3.62 mmol/L6521.79GFAPUCH-L1NoNo
Chloramphenicol2417.80GFAPUCH-L1NoNo
Diclofenac812.40GFAPUCH-L1NoNo
Ibuprofen a2.425 mmol/L50.0GFAPUCH-L1NoNo
Phenytoin2386GFAPUCH-L1NoNo
Amphetamine2.440.033GFAPUCH-L1NoNo
Benzoylecgonine a8.642.5 μg/mLGFAPUCH-L1NoNo
Nicardipine hydrochloride0.970.0465GFAPUCH-L1NoNo
EDDP† perchlorate a0.3308125 ng/mLGFAPUCH-L1NoNo
Methadone10.30.318GFAPUCH-L1NoNo
Methaqualone a32.368.1 μg/mLGFAPUCH-L1NoNo
d-Methamphetamine a1.865278.4 ng/mLGFAPUCH-L1NoNo
Morphine27.30.78GFAPUCH-L1NoNo
Oxazepam15.10.432GFAPUCH-L1NoNo
Phencyclidine a0.03578.7 ng/mLGFAPUCH-L1NoNo
Secobarbital66.8159.17GFAPUCH-L1NoNo
Cocaine a11.4063.46 μg/mLGFAPUCH-L1NoNo
Propoxyphene a9.4632.11GFAPUCH-L1NoNo
Warfarin2437.5GFAPUCH-L1NoNo
Table 15: Interfering Substance Testing
SubstanceTest ConcentrationInterference(Yes/No)InterferenceResults
μmol/Lmg/dLAssay
UCH-L1No
Diazepam1050.330GFAPNo
UCH-L1No

{19}------------------------------------------------

{20}------------------------------------------------

  • 2-Ethylidene-1,5-dimethyl-3,3-diphenylpyrrolidine

a The test concentration used for this substance is not from CLSI guideline EP37 1* edition

b The 'x' factor listed indicates the number of times more activity than a known negative sample, for its ability to crosslink antibodies in a mouse system assay.

C One out of the five samples enriched for the presence of HAMA and two out of the five samples enriched for presence of RF exhibited an interference effect. See note regarding HAMA or other heterophile antibodies in Limitations of the Procedure section above.

d Note that the ethanol level is well above the CLSI highest therapeutic level of 43.4 mmol/L (200 mg/dL)

Cross-reactivity: The i-STAT TBI Plasma cartridge is specific to the measurement of GFAP and UCH-L1. The following proteins in Table 16 with significant homology to GFAP or UCH-L1 were tested at highest known physiological levels and none were found to have significant cross-reactive effects on measured the GFAP or UCH-L1 levels.

Table 16: Interfering Substance Testing
SubstanceTestConcentrationpg/mLAssayCross-reactivity(Yes/No)
Keratin type II10 000GFAPNo
Internexin77 000GFAPNo
Neurofilament medium8600GFAPNo
Neurofilament heavy77 000GFAPNo
Neurofilament light68GFAPNo
Peripherin5000GFAPNo
Desmin127 000GFAPNo
Vimentin354 000GFAPNo
Ubiquitin Carboxyl-TerminalHydrolase L3 (UCH-L1)354 000UCH-L1No

Cross-talk: The GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge were evaluated for potential cross-talk to determine if high levels of the antigen (GFAP or UCH-L1) of one assay have potential to the impact the result of the other assay. Plasma samples spiked to low and high GFAP and UCH-L1 levels were evaluated in the presence of a single high level of the other antigen being evaluated for potential cross-talk. No cross-talk effect was observed as the results demonstrated that the GFAP result is not affected when UCH-L1 is present in a sample, and that the UCH-L1 result is not affected when GFAP is present in a sample.

{21}------------------------------------------------

Table 17: Interfering Substance Testing
SubstanceTestConcentrationpg/mLAssayCross-talk(Yes/No)
UCH-L1100 000GFAPNo
GFAP100 000UCH-L1No

f. Assay cutoff

Cutoff: The assay cutoffs were determined by analyzing a training set with GFAP and UCH-L1 results from a total of 420 (274 males and 146 females) with suspected mild traumatic brain injury (TBI; Glasgow Coma Scale score of 13-15). Subjects who had blood drawn within 12 hours of injury and a head CT scan determination, were included in the analysis. Using a 10-fold cross validation and bootstrapping method, the cutoff values of 30 pg/mL (GFAP assay) and 360 pg/mL (UCH-L1 assay) were selected for the i-STAT TBI Plasma Cartridge using the selection criteria with an adjusted NPV (prevalence 6%)≥98.5% and sensitivity ≥98%.

2. Clinical Study

Clinical Sensitivity and Specificity: A pivotal study using prospectively collected and archived (frozen) plasma specimens was conducted to establish the clinical performance of the i-STAT TBI Plasma test. The testing of the archived plasma specimens was conducted at three clinical sites in the United States.

The specimens were originally collected in a prospective, multi-center clinical study that enrolled consenting men and women 18 years of age or older who presented to emergency departments (ED) with suspected traumatic brain injury with initial Glasgow Coma Scale (GCS) scores of 13-15 and who had a computed tomography (CT) scan performed per the clinical site's standard of care. Subjects were enrolled at 22 clinical sites in three countries: United States, Germany and Hungary.

CT scans were performed in accordance with the clinical site's standard of care. Images were transmitted to a central neuroimaging processing center. Images were interpreted by at least two neuroradiologists who were masked to other clinical and laboratory data; procedures for scoring images were established before conducting image review. The clinical outcome was based on the consensus interpretation between two neuroradiologists with adjudication by a third neuroradiologist if necessary. Outcomes were positive or negative as defined by the presence or absence of acute traumatic intracranial lesions, respectively. Acute intracranial lesion was defined as any trauma induced or related finding visualized upon head CT scan.

Whole blood was collected into K2EDTA blood collection tubes from each subject using venipuncture and centrifuged to obtain plasma. Specimens were collected within 12 hours of head injury. The plasma specimens were divided into aliquots and frozen in cryovials before being provided to testing sites.

{22}------------------------------------------------

Of the 1994 subjects with GCS scores of 13 to 15 enrolled in the original study, specimens from 93 subjects were not included in the performance analysis due to subject discontinuation, lack of consent for specimen archiving for future testing, inconclusive or unreadable CT scan results, and/or unknown time from injury to blood collection. Specimens from 1901 subjects were included in the analysis.

The demographic characteristics of the subjects represented in the performance analysis are summarized in Table 18 below.

{23}------------------------------------------------

CharacteristicHead CT Scan ResultTotal
PositiveNegative
N12017811901
Age1 (Years)
Mean58.848.549.1
Median58.548.049.0
Standard Deviation18.2921.0120.99
Range(20, 95)(18, 98)(18, 98)
Gender, N (%)
Male70(58.3%)1005(56.4%)1075(56.6%)
Female50(41.7%)776(43.6%)826(43.5%)
Race2, N (%)
White98(81.7%)1245(69.9%)1343(70.6%)
Black or African American16(13.3%)483(27.1%)499(26.2%)
Asian5 (4.2%)24 (1.3%)29 (1.5%)
Native Hawaiian/Pacific Islander1 (0.8%)2 (0.1%)3 (0.2%)
American Indian or Alaska Native1 (0.8%)9 (0.5%)10 (0.5%)
Unknown1 (0.8%)27 (1.5%)28 (1.5%)
Ethnicity, N (%)
Hispanic or Latino1 (0.8%)89 (5.0%)90 (4.7%)
Not Hispanic or Latino118(98.3%)1691(94.9%)1809(95.2%)
Not Reported1 (0.8%)1 (0.1%)2 (0.1%)

Table 18: Demographic Characteristics

1 Age was calculated relative to the date of informed consent.

² Subjects could have indicated more than one race.

The head injury characteristics of the subjects represented by the 1901 specimens included in the performance analysis were tabulated. Information regarding time from head injury to exam, head injury to CT scan, and head injury to blood draw, as well as GCS, neurological assessment and physical evidence of trauma, categorized by head CT scan results, are shown in Table 19.

{24}------------------------------------------------

Table 19: Head Injury Characteristics
-----------------------------------------------
Head CT Scan Result
CharacteristicPositiveNegativeTotal
N12017811901
Time from head injury to exam (hours)1
Mean1.91.61.6
Median1.21.01.1
Standard Deviation1.731.711.71
Range(0.3, 7.8)(0.1, 10.7)(0.1, 10.7)
Time from head injury to CT scan (hours)1
Mean2.82.72.7
Median2.12.22.1
Standard Deviation1.951.931.93
Range(0.5, 8.9)(0.2, 13.3)(0.2, 13.3)
Time from head injury to blood draw (hours)1
Mean3.83.53.5
Median3.33.13.2
Standard Deviation1.911.881.89
Range(0.3, 9.3)(0.3, 11.9)(0.3, 11.9)
Glasgow Coma Score - N (%)
137 (5.8%)15 (0.8%)22 (1.2%)
1419 (15.8%)71 (4.0%)90 (4.7%)
1594 (78.3%)1695 (95.2%)1789 (94.1%)
Neurological assessment - N (%) of subjects experiencing:
Loss of Consciousness (LOC)82 (68.3%)721 (40.5%)803 (42.2%)
Alteration of Consciousness (AOC)92 (76.7%)978 (54.9%)1070 (56.3%)
Confusion44 (36.7%)313 (17.6%)357 (18.8%)
Vomiting14 (11.7%)128 (7.2%)142 (7.5%)
Post Traumatic Amnesia (PTA)81 (67.5%)546 (30.7%)627 (33.0%)
Post Traumatic Seizures2 (1.7%)11 (0.6%)13 (0.7%)
Subjects With Drug or AlcoholIntoxication at the Time ofPresentation to Facility33 (27.5%)369 (20.7%)402 (21.1%)
Dangerous Mechanism of Injury227 (22.5%)369 (20.7%)396 (20.8%)
Physical Evidence3
Visible Trauma Above the Clavicle101 (84.2%)1102 (61.9%)1203 (63.3%)
Suspected Open or Depressed SkullFracture14 (11.7%)46 (2.6%)60 (3.2%)

1 Based on time subject was initially examined at the medical facility

2 Dangerous mechanism of injury was pedestrian struck by a motor vehicle, an occupant ejected from a motor vehicle, or a fall from an elevation of 3 or more feet or 5 stairs

3 Prior to head CT

The i-STAT TBI Plasma test clinical performance estimates are shown in Table 20. Of the 1901 specimens, 120 were associated with positive CT scan results. Of these 120 specimens, 115 had an 'elevated' i STAT TBI Plasma test interpretation (115/120, clinical sensitivity = 95.8%). Five specimens associated with CT scan positive results had an i-STAT TBI Plasma test interpretation that was 'not elevated'. The rate of false

{25}------------------------------------------------

negative (FN) results was 4.2% (5/120). Five subjects in the study were identified with lesion requiring surgical intervention; none of these five subjects had a FN result, suggesting that the i-STAT TBI Plasma test correctly classified all these five CTpositive subjects with a test interpretation of 'elevated.' Of the 1781 specimens associated with negative CT scan results, 720 had an i-STAT TBI Plasma test interpretation that was 'not elevated' (720/1781, clinical specificity = 40.4%). The rate of False Positive (FP) results was 59.6% (1061/1781).

Overall, there were 725 specimens with i-STAT TBI Plasma test interpretations of 'not elevated'. Of these. 720 specimens were associated with negative CT scan results. The Negative Predictive Value (NPV) of the assay was 99.3% (720/725).

Table 20: Clinical Performance
i-STAT TBI PlasmaTest InterpretationAdjudicated CT Scan ResultPositiveAdjudicated CT Scan ResultNegativeTotal
Elevated11510611176
Not Elevated5720725
Total12017811901
Clinical Performance ParametersN=190195% Confidence Interval
Clinical Sensitivity95.8%90.6%, 98.2%
Clinical Specificity40.4%38.2%, 42.7%
Negative Predictive Value (NPV)99.3%98.5%, 99.7%
Positive Predictive Value (PPV)9.8%9.2%, 10.2%
Likelihood Ratio Negative (LRN)0.100.04, 0.23
Likelihood Ratio Positive (LRP)1.611.51, 1.69

To supplement the results of the pivotal study (N=1901) described above, a study was conducted using freshly collected plasma specimens from consenting men and women 18 years of age or older who presented to Level 1 trauma center emergency departments (ED) with suspected traumatic brain injury, with initial Glasgow Coma Scale(GCS) scores of 13-15, and who had a computed tomography (CT) scan of the head performed per the clinical site's standard of care. A total of 88 subjects were enrolled across 4 clinical sites of the Transforming Research and Clinical Knowledge in Traumatic Brain Injury (TRACK-TBI) study in the United States.

Similar to the pivotal study. CT scans were performed in accordance with the clinical site's standard of care. Images were interpreted by at least two neuroradiologists who were masked to other clinical and laboratory data; procedures for scoring images were established before conducting image review. The clinical outcome was based on the consensus interpretation between two neuroradiologists, with adjudication by a third neuroradiologist if necessary. Outcomes were positive or negative as defined by the presence or absence of acute traumatic intracranial lesions, respectively. Acute intracranial lesion was defined as any trauma induced or related finding visualized upon head CT scan.

Whole blood was collected into K3EDTA blood collection tubes from each subject

{26}------------------------------------------------

using venipuncture and centrifuged to obtain plasma. Specimens were collected within 12 hours of head injury. The demographic characteristics of the subjects represented in the performance analysis are summarized in Table 21 below:

Table 21: Demographic Characteristics - Supplemental Fresh SpecimenStudy
CharacteristicHead CT Scan ResultTotal
PositiveNegative
N295988
Age (Years)
Mean49.239.342.5
Median473641
Standard Deviation16.9215.4316.52
Range(24, 85)(18, 76)(18, 85)
Gender
Male234063
Female61925

The head injury characteristics of the subjects in the supplemental plasma fresh specimen study including information regarding time from head injury to exam, head injury to CT scan, and head injury to blood draw, as well as GCS, neurological assessment and physical evidence of trauma, categorized by head CT scan results, are shown in Table 22.

Table 22: Head Injury Characteristics – Supplemental Fresh Specimen Study
CharacteristicHead CT Scan ResultTotal
PositiveNegative
N295988
Time from head injury to CT scan (hours)
Mean2.52.22.3
Median2.01.91.9
Standard Deviation1.761.391.51
Range(0.7, 8.7)(0.7, 7.5)(0.7, 8.7)
Time from head injury to blood draw (hours)
Mean6.64.45.1
Median6.03.94.3
Standard Deviation2.931.962.54
Range(2.3, 11.8)(2.0, 9.9)(2.0, 11.8)
Glasgow Coma Score – N (%)1

{27}------------------------------------------------

Table 22: Head Injury Characteristics - Supplemental Fresh Specimen Study
CharacteristicHead CT Scan ResultPositiveNegativeTotal
131 (1.1%)0 (0.0%)1 (1.1%)
146 (6.8%)9 (10.2%)15 (17.0%)
1522 (25.0%)50 (56.8%)72 (81.8%)
Neurological assessment - N (%) of subjects experiencing:
Loss of Consciousness (LOC)23 (79.3%)37 (62.7%)60 (68.2%)
Presence of Confusion19 (65.5%)40 (67.8%)59 (67.0%)
Vomiting2---
Post-traumatic Amnesia (PTA)22 (75.9%)38 (64.4%)60 (68.2%)
Post-traumatic Seizures0 (0%)0 (0%)0 (0%)
Subjects with Drug Intoxication atthe Time of Presentation to Site3 (10.3%)2 (3.4%)5 (5.7%)
Subjects with Alcohol Intoxicationat the Time of Presentation to Site6 (20.7%)4 (6.8%)10 (11.4%)
Physical Evidence - N (%)
Signs of Skull Fracture9 (31.0%)1 (1.7%)10 (11.4%)
Mechanism of Injury - N (%)
Acceleration/ Deceleration24 (82.8%)41 (69.5%)65 (73.9%)
Blow to Head4 (13.8%)8 (13.6%)12 (13.6%)
Head Against Object24 (82.8%)42 (71.2%)66 (75.0%)
Fall19 (65.5%)21 (35.6%)40 (45.5%)

1 Percent based on total subjects

2 Information not collected

The i-STAT TBI Plasma test clinical performance estimates from the supplemental fresh plasma specimen study are shown in Table 23. Of the 88 subjects tested, 29 were associated with positive head CT scan results. Of these 29 subjects, 29 had an 'elevated' i-STAT TBI Plasma test interpretation (29/29, clinical sensitivity = 100.0%). The rate of False Negatives (FN) was 0% (0/29). Of the 59 subjects associated with negative CT scan results, 14 had an i-STAT TBI Plasma test interpretation that was 'not elevated' (14/59, clinical specificity = 23.7%). The rate of False Positive (FP) results was 76.2% (45/59).

Overall, there were 14 specimens with i-STAT TBI Plasma test interpretations of 'not elevated'. All 14 specimens were associated with negative head CT scan results. The Negative Predictive Value (NPV) of the assay was 100% (14/14).

Table 23: Clinical Performance – Supplemental Fresh Specimen Study
i-STAT TBI PlasmaAdjudicated CT Scan ResultTotal

{28}------------------------------------------------

Test InterpretationPositiveNegative
Elevated294574
Not Elevated01414
Total295988
Clinical Performance ParametersN=8895% Confidence Interval
Clinical Sensitivity100.0%88.3%, 100.0%
Clinical Specificity23.7%14.7%, 36.0%
Negative Predictive Value (NPV)*100.0%80.2%, 100.0%
Positive Predictive Value (PPV)*39.2%35.9%, 43.4%
Likelihood Ratio Negative (LRN)0.000.00, 0.50
Likelihood Ratio Positive (LRP)1.311.14, 1.56

*NPV and PPV estimated at 33.0% prevalence of CT scan positive rate for suspected mild TBI subjects. Adjusted NPV and PPV at 6% prevalence (to be comparable to the pivotal study) are 100.0% (95% Cl: 96.9%, 100.0%) and 7.7% (95% CI: 6.8%, 9.1%), respectively.

8. Conclusion

Per special controls requirements of 21 CFR 866.5830, clinical studies were performed to support substantial equivalence. A pivotal study (N=1901) using archived plasma specimens from the same cohort of subjects that was used to demonstrate clinical performance of the predicate device showed similar clinical specificity and specificity with a similar number of false negative results. None of the false negative results were of subjects identified with a lesion requiring surgical intervention. The i-STAT TBI Plasma test showed high clinical sensitivity when validated in a supplemental fresh plasma specimen study (N=88). The clinical data demonstrates that the i-STAT TBI Plasma cartridge with the i-STAT Alinity System performs comparably to the predicate Banyan BTI device. In addition, a benefit-risk assessment was performed to help determine substantial equivalence for the device. Although similar clinical specificity to the predicate device was observed in the pivotal study, this was not as high in the supplemental fresh plasma study. However, the device shows potential for some reduction in the number of unnecessary CT scans. The i-STAT TBI Plasma cartridge offers significant benefit over the predicate device by establishing clinical workflow efficiencies through ease of use and a reduce test time. The assessment concludes that the benefits outweighs the risks. Overall, the results of analytical and clinical studies demonstrate that performance of the GFAP and UCH-L1 assays in the i-STAT TBI Plasma cartridge with the i-STAT Alinity System are substantially equivalent to the comparative method.

§ 866.5830 Brain trauma assessment test.

(a)
Identification. A brain trauma assessment test is a device that consists of reagents used to detect and measure brain injury biomarkers in human specimens. The measurements aid in the evaluation of patients with suspected mild traumatic brain injury in conjunction with other clinical information to assist in determining the need for head imaging per current standard of care.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The 21 CFR 809.10(b) compliant labeling must include detailed descriptions of and results from performance testing conducted to evaluate precision, accuracy, linearity, analytical sensitivity, interference, and cross-reactivity. This information must include the following:
(i) Performance testing of device precision must, at minimum, use one unmodified clinical specimen from the intended use population with concentration of the brain injury biomarker(s) near the medical decision point. Contrived specimens that have been generated from pooling of multiple samples or spiking of purified analyte to cover the measuring range may be used, but the contrived samples must be prepared to mimic clinical specimens as closely as possible. This testing must evaluate repeatability and reproducibility using a protocol from an FDA-recognized standard.
(ii) Device performance data must be demonstrated through a clinical study and must include the following:
(A) Data demonstrating clinical validity including the clinical sensitivity and specificity, and positive and negative predictive value of the test in the intended use population of patients with suspected mild traumatic brain injury (
i.e., Glasgow Coma Score (GCS) of 13-15), or equivalent standard of care for determination of severity of traumatic brain injury (TBI).(B) Study must be performed using the operators and in settings that are representative of the types of operators and settings for which the device is intended to be used.
(C) All eligible subjects must meet the well-defined study inclusion and exclusion criteria that define the intended use population. The prevalence of diseased or injured subjects in the study population must reflect the prevalence of the device's intended use population, or alternatively, statistical measures must be used to account for any bias due to enrichment of subpopulations of the intended use population.
(D) All eligible subjects must have undergone a head computerized tomography (CT) scan or other appropriate clinical diagnostic standard used to determine the presence of an intracranial lesion as part of standard of care and must also be evaluated by the subject device. All clinical diagnostic standards used in the clinical study must follow standard clinical practice in the United States.
(E) Relevant demographic variables and baseline characteristics including medical history and neurological history. In addition, head injury characteristics, neurological assessments, and physical evidence of trauma must be provided for each subject. This information includes but is not limited to the following: Time since head injury, time from head injury to CT scan, time from head injury to blood draw, GCS score or equivalent, experience of loss of consciousness, presence of confusion, episodes of vomiting, post-traumatic amnesia characteristics, presence of post-traumatic seizures, drug or alcohol intoxication, mechanism of injury, acute intracranial lesion type, neurosurgical lesion, and cranial fracture.
(F) Each CT scan or other imaging result must be independently evaluated in a blinded manner by at least two board-certified radiologists to determine whether it is positive or negative as defined by the presence or absence of acute intracranial lesions. This independent review must be conducted without access to test results of the device. Prior to conducting the review, the criteria and procedures to be followed for scoring the images must be established, including the mechanism for determining consensus.
(G) All the clinical samples must be tested with the subject device blinded to the TBI status and the neurological-lesion-status of the subject.
(H) Details on how missing values in data are handled must be provided.
(I) For banked clinical samples, details on storage conditions and storage period must be provided. In addition, a specimen stability study must be conducted for the duration of storage to demonstrate integrity of archived clinical samples. The samples evaluated in the assay test development must not be used to establish the clinical validity of the assays.
(iii) Performance testing of device analytical specificity must include the most commonly reported concomitant medications present in specimens from the intended use population. Additionally, potential cross-reacting endogenous analytes must be evaluated at the highest concentration reported in specimens from the intended use population.
(iv) Expected/reference values generated by testing a statistically appropriate number of samples from apparently healthy normal individuals.
(2) The 21 CFR 809.10(a) and (b) compliant labeling must include the following limitations:
(i) A limiting statement that this device is not intended to be used a stand-alone device but as an adjunct to other clinical information to aid in the evaluation of patients who are being considered for standard of care neuroimaging.
(ii) A limiting statement that reads “A negative result is generally associated with the absence of acute intracranial lesions. An appropriate neuroimaging method is required for diagnosis of acute intracranial lesions.”
(iii) As applicable, a limiting statement that reads “This device is for use by laboratory professionals in a clinical laboratory setting.”