K Number
K201752

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-01-29

(217 days)

Product Code
Regulation Number
882.4560
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Disposable Pre-calibrated Suction is an accessory of the Cranial Image Guided Surgery System and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial Image Guided Surgery System may be indicated.

Surgical example procedures include:

  • · Cranial resection of tumors and other lesions
  • · Resection of skull base tumors or other lesions
  • · AVM Resection
Device Description

Disposable Pre-calibrated Suction is an accessory of the Cranial IGS system and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial IGS system may be indicated.

The Brainlab Disposable Suction is an accessory for the currently released and developed Brainlab optical IGS systems for cranial procedures. The device will be pre-calibrated, i.e. it will be automatically recognized by the system and is immediately ready to use. By tracking the flat markers attached to the integrated tracking array, the instrument and thereby position of the tip can be located by the Brainlab optical IGS systems for cranial procedures.

The device is intended for single short term invasive use on an individual patient during a single procedure. This invasive device is used for a short-term limited contact (<24 hours) and can be in direct contact with the central nervous system (CNS) tissues and the cerebrospinal fluid (CSF).

AI/ML Overview

The provided text describes a 510(k) submission for a medical device called "Disposable Pre-calibrated Suction." It is an accessory to a Cranial Image Guided Surgery System. The document focuses on demonstrating substantial equivalence to predicate devices rather than proving a specific AI algorithm's performance against acceptance criteria for a diagnostic task.

Therefore, the information required to populate the fields related to AI model acceptance criteria, test set details (sample size, provenance, expert consensus, adjudication, MRMC studies), training set details, and ground truth establishment is not present in the provided text. The document primarily details mechanical, accuracy, shelf-life, biocompatibility, and sterility testing for a physical medical instrument.

However, I can provide the acceptance criteria and reported performance for the instrument tracking accuracy, which is a key technical performance aspect mentioned in the document.

Here's a table based on the provided "Performance Data" section:

Acceptance Criteria and Reported Device Performance (Instrument Tracking Accuracy)

MetricAcceptance Criteria (Not explicitly stated as criteria, but implied by reported performance)Reported Device Performance (REF 52184-01)Reported Device Performance (REF 52185-01)
Locational Error (Mean)Not explicitly stated; implied to be low0.45 mm0.51 mm
Locational Error (Std Dev)Not explicitly stated; implied to be low0.11 mm0.09 mm
Locational Error (95th percentile)Not explicitly stated; implied to be below a certain threshold0.93 mm1.01 mm
Locational Error (99% confidence interval)Not explicitly stated; implied to be low0.53 mm0.59 mm
Angular Error (Mean)Not explicitly stated; implied to be low0.19 °0.23 °
Angular Error (Std Dev)Not explicitly stated; implied to be low0.05 °0.06 °
Angular Error (95th percentile)Not explicitly stated; implied to be below a certain threshold0.27 °0.31 °
Angular Error (99% confidence interval)Not explicitly stated; implied to be low0.22 °0.28 °

Study Proving Device Meets Acceptance Criteria:

The study described is primarily a technical performance verification aimed at demonstrating substantial equivalence for a physical medical instrument, not an AI algorithm.

  1. Sample sized used for the test set and the data provenance:

    • The document does not specify the sample size (number of measurements or trials) used for the instrument tracking accuracy testing.
    • Data Provenance: Not explicitly stated, but implied to be from internal testing by Brainlab AG given the context of a 510(k) summary. No country of origin for the data is mentioned, nor whether it was retrospective or prospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is not applicable as the "ground truth" here refers to the actual physical location and orientation of the instrument during accuracy testing, which would be established by high-precision measurement systems, not human experts.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for instrument performance testing.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a navigated suction instrument, not an AI-powered diagnostic tool requiring human reader studies.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The "standalone" performance here refers to the instrument's accuracy as measured by the tracking system, independent of a human's surgical skill. The table above presents this standalone performance.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For instrument tracking accuracy, the ground truth would be established by a highly accurate (e.g., optical or mechanical) reference measurement system, often referred to as a "gold standard" measurement setup. It is not expert consensus, pathology, or outcomes data.
  7. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that requires a training set. The "pre-calibrated" aspect refers to factory calibration, not AI model training.
  8. How the ground truth for the training set was established:

    • Not applicable as there is no training set for an AI model.

In summary, the provided document focuses on the engineering and biocompatibility validation of a physical medical device accessory rather than the clinical or diagnostic performance of an AI algorithm. Therefore, many of the requested details related to AI model evaluation are not present.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

January 29, 2021

Brainlab AG Chiara Cunico Manager Regulatory Affairs Olof-Palme-Str. 9 Munich, Bavaria 81829 Germany

Re: K201752

Trade/Device Name: Disposable Pre-calibrated Suction Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: December 28, 2020 Received: December 31, 2020

Dear Chiara Cunico:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201752

Device Name Disposable Pre-Calibrated Suction

Indications for Use (Describe)

Disposable Pre-calibrated Suction is an accessory of the Cranial Image Guided Surgery System and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial Image Guided Surgery System may be indicated.

Surgical example procedures include:

  • · Cranial resection of tumors and other lesions
  • · Resection of skull base tumors or other lesions
  • · AVM Resection
Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

January 29, 2021

General Information
ManufacturerBrainlab AG, Olof-Palme-Str. 9, 81829, Munich, Germany
Establishment Registration8043933
Device NameNeurological Stereotaxic Instrument
Trade NameDisposable Pre-calibrated suction
Classification NameStereotaxic instrument
Product CodesHAW
Regulation Number882.4560
Regulatory ClassII
PanelNeurology
Predicate Devices and K1. K082060 Cranial ENT IGS System
Numbers2. K092467 Disposable Stylet
ManufacturerBrainlab AG
Contact Information
Primary ContactAlternate Contact
Chiara CunicoRegulatory Affairs Brainlab
Manager RAPhone: +49 89 99 15 68 0
Phone: +49 89 99 15 68 0Fax: +49 89 99 15 68 5033
Email: chiara.cunico@brainlab.comEmail: regulatory.affairs@brainlab.com

1. Indications for Use

Disposable Pre-calibrated Suction is an accessory of the Cranial Image Guided Surgery System and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial Image Guided Surgery System may be indicated.

Surgical example procedures include:

  • · Cranial resection of tumors and other lesions
  • Resection of skull base tumors or other lesions
  • AVM Resection

{4}------------------------------------------------

2. Device Description

Disposable Pre-calibrated Suction is an accessory of the Cranial IGS system and intended to be used as a navigated suction device in any surgical procedure in which the use of the Cranial IGS system may be indicated.

The Brainlab Disposable Suction is an accessory for the currently released and developed Brainlab optical IGS systems for cranial procedures. The device will be pre-calibrated, i.e. it will be automatically recognized by the system and is immediately ready to use. By tracking the flat markers attached to the integrated tracking array, the instrument and thereby position of the tip can be located by the Brainlab optical IGS systems for cranial procedures.

The device is intended for single short term invasive use on an individual patient during a single procedure. This invasive device is used for a short-term limited contact (<24 hours) and can be in direct contact with the central nervous system (CNS) tissues and the cerebrospinal fluid (CSF).

3. Substantial Equivalence

The reusable ENT suction has the same intended use and indications for use as the Disposable Suction and hence, is considered to be a suitable primary predicate device. Furthermore, the secondary predicate - Disposable stylet, was cleared under K092467 and found to be substantially equivalent with the predicate device Cranial ENT IGS (K082060). It has comparable features in terms of: sterility, single-use and tracking technology as the Subject Device and is also intended for invasive patient contact, including CNS and CSF.

At a high level, the subject and primary predicate device are based on the following same technological elements:

  • Optical infrared light based reflective marker tracking for instrument localization a)
  • b) Suction and pointing functionality combined in one device
  • Stainless steel suction tube that is in direct contact with patient c)
  • Accessory to the Cranial ENT IGS system d)

The following technological differences exist between the subject and primary predicate device, but are all similar to the secondary predicate device:

  • (1) Pre-calibrated versus manual calibrated instrument
  • (2) Single use sterile versus reusable instrument
  • (3) Flat Marker versus Spherical Marker Tracking
  • (4) Integrated tracking array versus (de)attachable tracking array
  • (5) Polycarbonate handle glued to suction tube versus total stainless steel instrument

The Subject Device has larger tube diameters of 2.7 mm and 4mm when compared with the predicate device (2 mm and 3.3 mm). This larger size does not introduce an additional risk to patient safety since these dimensions are identical to suction tubes of the cleared reference device K162929 "Stryker Navigation System with CranialMap software application". K162929 has similar indications for use as the subject device.

{5}------------------------------------------------

Disposable Pre-calibrated Suction(Brainlab)Posterior-FossaSuction Tube (Stryker)Comparison
Tube diameter2.7mm and 4.0 mm2.7mm and 4.0 mmIdentical

These differences do not raise any new questions of safety or effectiveness compared to the predicate device.

4. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

The substantial equivalence of the Disposable Suction is furthermore extensively verified. Technical performance has been tested on the following main aspects:

  • Mechanical stability testing: mechanical stability of polycarbonate handle glued to a) suction tube has proven to be sufficient for intended use.
  • Instrument tracking accuracy testing: instrument accuracy with flat marker technology b) has proven to meet accuracy requirements.
  • Instrument system integration testing: pre-calibration of the Disposable Suction has C) proven to be effective within the Cranial IGS system.
  • Shelf-life testing: technical performance (mechanical stability and accuracy) have proven d) to still meet requirement after shelf-life.
  • Biocompatibility testing: biocompatibility has been stated for all materials of the e) Disposable Suction, also considering the manufacturing process.
  • Sterility testing: sterilization has effectively shown that the sterility level of 106 f) has been reached.
  • Sterile barrier system integrity testing: packaging validation has proven the integrity of g) the sterile barriers system of the double blister packaging.
REF 52184-01REF 52185-01
Locational errorAngular errorLocational errorAngular error
Mean0.45 mm0.19 °0.51 mm0.23 °
Standard deviation0.11 mm0.05 °0.09 mm0.06 °
95th percentile0.93 mm0.27 °1.01 mm0.31 °
99% confidence interval0.53 mm0.22 °0.59 mm0.28 °

Table: Locational and angular accuracy Disposable Suction (REF 52184-01 and REF 52185-01)

Biocompatibility testing

The biological safety as a part of Disposable Suction basic safety has been assessed during biological safety evaluation. For a description of the used materials characterization, biological endpoint evaluation as well as for details about the performed verification tests refer to the Biological Risk Assessment by NAMSA performed according to ISO 10993-1 FDA Guidance "Use of International Standard ISO10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk

management process". The relevant "Biological Effect" endpoints have been assessed to identify

{6}------------------------------------------------

existing physical and/or chemical information and assess necessity of additional testing in order to demonstrate biological safety for the device acc. to its intended use..

Following biocompatibility tests have been performed (by ISO 17025 accredited laboratories) on the release candidate products manufactured, packaged and sterilized according to final manufacturing process:

  • Cytotoxicity on tube test acc. to ISO 10993-5 ●
  • Cytotoxicity on handle test acc. to ISO 10993-5
  • Acute Systemic Toxicity test acc. to ISO 10993-11 ●
  • Irritation test acc. to ISO 10993-10
  • Pyrogen test acc. to European Pharmacopeia, 9th edition, 2016 .
  • . Pyrogen test acc. to USP 42 - NF 37
  • . Sensitization test acc. to ISO 10993-10
  • Particle test acc. to USP [788] ●
  • Bioburden test before and after packaging acc. to ISO 11737-1 .
  • EO residual gas testing acc. to ISO 10993-7 ●
  • Hemolysis tests acc. to ISO 10993-4 ●

5. Conclusion

The comparison of the Subject Device with the predicate devices shows that the Subject Device has similar functionality, intended use, technological characteristics, and typical users as the predicate devices. Verification and validation activities ensured that the design specifications are met and that the Device does not introduce new issues concerning safety and effectiveness. Hence, the Subject Device is substantially equivalent to the predicate device.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).