(162 days)
EMS: The Compex Sport Elite 3.0 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the Compex Sport Elite 3.0 programs is not suitable for relabilitation or physiotherapy.
TENS: The Compex Sport Elite TENS is intended for:
-
Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities:
-
The symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
The Compex Sport Elite 3.0 is an Over-the-Counter device to be used by adults only.
The Compex Sport Elite 3.0 is an electrical stimulation device, which stimulates healthy muscles by means of electrical impulses transmitted by electrodes. The electrical pulses generated by the Compex Sport Elite 3.0 stimulates motor nerves to stimulate a muscular response. Depending on the parameters of theelectrical impulses (pulse frequency. duration of contraction, duration of rest, and totalsession duration). different types of muscle work can be imposed on the stimulated muscles. The Compex Sport Elite 3.0 may therefore be considered a technique of muscletraining.
TENS (Transcutaneous Electrical Nerve Stimulation) and NMES (NeuroMuscular Electrical Stimulation) target different nerve groups of the body. TENS specifically targets the sensory nerves, which are responsible for sending pain signals to the brain. NMES targets the muscle itself, specifically through the motor nerves. This allows the NMES machine to create a muscle contraction to recruit more muscle fibers when training; warming up or recovering.
The Compex Sport Elite 3.0 system consists of these components:
- 1x Stimulator
- 1x Lead WireSet
- 2x Small Performance Snap Electrodepackages ●
- 2x Large Performance Snap Electrodepackages
- 1x Charging Cable ●
- 1x Carrying Case ●
These components are packaged together in a carrying case along with the user guide and a battery charger.
The electrodes are supplied by Axelgaard Mfg. Co., Inc. and cleared under K130987.
The stimulator is a microprocessor controlled 4-channel electro-stimulator. The stimulator drives each output channel independently based upon the parameters predefined for a program selected by the user operates the device through a User Interface (UI) consisting of a graphic LCD, keypad controls and supporting software. The User Interface is made up of different menus that allow the following functions: set up the device (Options Menu), select a desired category (Category Menu), select a preset program (Treatments ) Menu), tune a selected program (Stimulation Menu), andadiust the stimulation intensities (Level Menu).
The provided document is a 510(k) summary for a medical device, the Compex® Sport Elite 3.0. This type of submission to the FDA focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving de novo safety and effectiveness through extensive clinical trials. Thus, the information requested in your prompt regarding acceptance criteria, sample sizes, expert ground truth, MRMC studies, standalone performance, and training set details for AI/ML-based devices is not applicable to this specific document.
The document primarily addresses acceptance criteria in the context of non-clinical testing and comparison to the predicate device. The study proving the device meets these "acceptance criteria" is a demonstration of substantial equivalence rather than a clinical performance study.
Here's an analysis of the provided information within the context of a 510(k) submission, addressing the questions where applicable:
1. A table of acceptance criteria and the reported device performance:
The document doesn't present "acceptance criteria" in the traditional sense of a performance study with specific numerical thresholds for accuracy, sensitivity, or specificity for an AI/ML device. Instead, the acceptance criteria are implicitly defined by the safety and performance standards the device was tested against and by its substantial equivalence to the predicate device.
The document provides a "Comparison Analysis" (Table 1: Characteristics, Table 2: Output Specifications, Table 3: Software Comparison Analysis) between the subject device (Compex® Sport Elite 3.0) and its predicate device (Compex Sport Elite, K170918). This comparison itself serves as the "reported device performance" against the "acceptance criteria" of being substantially equivalent to the predicate and meeting relevant safety standards.
Table 1: Comparison Analysis - Characteristics (Excerpt)
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis (HAZ-PRO-130) |
|---|---|---|---|---|---|
| 1 | Electrodes connection | [Description] | [Description] | Same | No additional risk. |
| 2 | Power Source | Rechargeable Ni-MH battery 4.8V (4 cells AA=R6); replaceable. | Rechargeable Li-ion battery 3.7V (one cell); not replaceable. | Different | Smaller size; reduced weight. User will not handle battery. The battery conforms to IEC 62133-2 and UN DOT 38.3. No additional risk. |
| ... | ... | ... | ... | ... | ... |
| 24 | Ingress Protection | IP20 | IP22 | Different | IP22 is required by IEC 60601-1-11. Reduced risk. |
| ... | ... | ... | ... | ... | ... |
Table 2: Output Specifications Comparison Analysis (Excerpt)
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk analysis |
|---|---|---|---|---|---|
| 33 | Maximum Output Voltage (±10%) | 60 V @ 500 Ω, 165 V @ 2 kΩ, 165 V @ 10 kΩ | 60 V @ 500 Ω, 165 V @ 2 kΩ, 165 V @ 10 kΩ | Same | No additional risk |
| 35 | Pulse Width | NMES: 200 to 400 [µs], TENS: 70 to 300[µs] | NMES: 200 to 400 [µs], TENS: 70 to 300[µs] | Same | No additional risk |
| ... | ... | ... | ... | ... | ... |
Table 3: Software Comparison Analysis (Excerpt)
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis |
|---|---|---|---|---|---|
| 41 | Dual MCU design | [Description] | [Description] | Different master MCU, Same protection MCU | Subject device's master MCU has improved performance for control and user interface. Same protection MCU is used. No additional risk. |
| 47 | UI Screens - Placements selection screen | Does not exist, need reference to printed user guide | Using graphic display to inform user the proper location of the treatment electrode pads on a human body diagram. | Different | Subject device offers clearer usage guide. Reduced risk. |
| ... | ... | ... | ... | ... | ... |
The "acceptance criteria" here are that the subject device performs identically or similarly to the predicate device in terms of critical functional and safety parameters, and any differences do not raise new issues of safety or effectiveness. For "Different" or "Similar" characteristics, the risk analysis justifies why these differences do not introduce new risks or even reduce existing ones (e.g., improved ingress protection).
2. Sample sizes used for the test set and the data provenance:
- Sample Size: Not applicable. This is a medical device (muscle stimulator), not an AI/ML diagnostic or prognostic system that processes clinical data. The "testing" refers to bench testing and comparison to a predicate, not performance on a clinical test set of patient data.
- Data Provenance: Not applicable for clinical data. The provenance for the testing data would be from laboratory bench tests conducted by a Nationally Recognized Test Laboratory (Nemko Shanghai Ltd. Shenzhen Branch).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable. Ground truth as in expert consensus on disease status is not relevant for this type of device and submission. The "ground truth" for the device's technical specifications and safety conformity would be the industry standards (e.g., IEC 60601 series) and the specifications of the predicate device.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. Adjudication methods are used in clinical studies, particularly for subjective assessments or when establishing ground truth from multiple readers. This is a technical comparison and safety conformity assessment.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is an electrical stimulator, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This is not an algorithm-only device. Its function is direct physical electrical stimulation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for this 510(k) submission is based on:
- International safety and performance standards: e.g., ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-10, IEC 62133-2, ANSI/AAMI/IEC 62304.
- Specifications and performance of the legally marketed predicate device (Compex Sport Elite, K170918): The core of the 510(k) submission is to demonstrate that the new device is "substantially equivalent" to this predicate.
8. The sample size for the training set:
Not applicable. There is no AI/ML component with a training set for this device.
9. How the ground truth for the training set was established:
Not applicable. There is no AI/ML component with a training set for this device.
In summary: The provided document is an FDA 510(k) clearance letter and summary for a physical medical device (muscle stimulator). The "study that proves the device meets the acceptance criteria" refers to bench testing against recognized electrical safety and performance standards and a detailed comparison to a legally marketed predicate device to demonstrate substantial equivalence, rather than a clinical trial or performance study typical for AI/ML-driven diagnostic devices.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 27, 2020
DJO, LLC Michael Davis Regulatory Affairs Consultant 1430 Decision Street Vista. California 92081
Re: K201653
Trade/Device Name: Compex® Sport Elite 3.0 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX, NUH, NYN Dated: June 16, 2020 Received: June 18, 2020
Dear Michael Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K201653
Device Name Compex® Sport Elite 3.0
Indications for Use (Describe)
EMS: The Compex Sport Elite 3.0 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the Compex Sport Elite 3.0 programs is not suitable for relabilitation or physiotherapy.
TENS: The Compex Sport Elite TENS is intended for:
-
Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities:
-
The symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis.
The Compex Sport Elite 3.0 is an Over-the-Counter device to be used by adults only.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ------------------------------------------------- | -- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DJO, a company that specializes in powering motion. The logo features a red circle with a white curved line inside, followed by the letters "DJO" in black. Below the logo, the text "DJO, LLC / 1430 Decision Street / Vista, CA 92081 / 800.321.9549 / djoglobal.com" is displayed in gray.
510(k) Summary
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k)number is
| Submitted by: | DJO, LLC1430 Decision StreetVista, CA 92081FDA Establishment Registration: 2020737 |
|---|---|
| Contact Person: | Michael DavisRegulatory Affairs Site Lead(205) 789-8154 |
| Date Summary Prepared: | November 24, 2020 |
| Trade Name: | Compex® Sport Elite3.0 |
| Common Name: | Stimulator, Muscle |
| Classification Name: | Stimulator, muscle, powered (21 CFR890.5850)Stimulator, Nerve, Transcutaneous (21 CFR 882.5890) |
| Product Code: | Powered Muscle Stimulator, For Muscle Conditioning - 21CFR 890.5850; Product Code NGX; Review Panel:Physical Medicine (primary product code) |
| Stimulator, Nerve, Transcutaneous, Over-The-Counter- 21CFR 882.5890; Product Code NUH; Review Panel:Neurology | |
| Stimulator, Electrical, Transcutaneous, For Arthritis 21CFR 890.5850; Product Code NYN (subsequent code);Review Panel: Neurology | |
| Regulatory Class: | Class II |
| Predicate Device: | Compex Sport Elite (K170918) - ClassII(Manufacturer's own predicate device) |
{4}------------------------------------------------
Device Description:
The Compex Sport Elite 3.0 is an electrical stimulation device, which stimulates healthy muscles by means of electrical impulses transmitted by electrodes. The electrical pulses generated by the Compex Sport Elite 3.0 stimulates motor nerves to stimulate a muscular response. Depending on the parameters of theelectrical impulses (pulse frequency. duration of contraction, duration of rest, and totalsession duration). different types of muscle work can be imposed on the stimulated muscles. The Compex Sport Elite 3.0 may therefore be considered a technique of muscletraining.
TENS (Transcutaneous Electrical Nerve Stimulation) and NMES (NeuroMuscular Electrical Stimulation) target different nerve groups of the body. TENS specifically targets the sensory nerves, which are responsible for sending pain signals to the brain. NMES targets the muscle itself, specifically through the motor nerves. This allows the NMES machine to create a muscle contraction to recruit more muscle fibers when training; warming up or recovering.
The Compex Sport Elite 3.0 system consists of these components:
- 1x Stimulator
- 1x Lead WireSet
- 2x Small Performance Snap Electrodepackages ●
- 2x Large Performance Snap Electrodepackages
- 1x Charging Cable ●
- 1x Carrying Case ●
These components are packaged together in a carrying case along with the user guide and a battery charger.
The electrodes are supplied by Axelgaard Mfg. Co., Inc. and cleared under K130987.
The stimulator is a microprocessor controlled 4-channel electro-stimulator. The stimulator drives each output channel independently based upon the parameters predefined for a program selected by the user operates the device through a User Interface (UI) consisting of a graphic LCD, keypad controls and supporting software. The User Interface is made up of different menus that allow the following functions: set up the device (Options Menu), select a desired category (Category Menu), select a preset program (Treatments ) Menu), tune a selected program (Stimulation Menu), andadiust the stimulation intensities (Level Menu).
See Appendix 5 Illustrated Device Description for a detailed description.
Indications for Use:
EMS: The Compex Sport Elite 3.0 is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The work imposed on the muscles by the Compex Sport Elite 3.0 programs is not suitable for rehabilitation or physiotherapy.
{5}------------------------------------------------
TENS: The Compex Sport Elite 3.0 TENS is intended for:
- Temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal householdandworkactivities;
- The symptomatic relief and management of chronic, intractable pain and relief of pain associated witharthritis.
The Compex Sport Elite 3.0 is an Over-the-Counter device to be used by adults only.
Summary of Non-clinical Tests:
The subject device has been evaluated for safety and performance by lab bench testing in accordance with the following standards:
-
- ANSI/AAMI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- ANSI/AAMI/IEC 60601-1-2:2014. Medical Electrical Equipment -- Part 1-2: 2. General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests
- IEC 60601-1-6 Edition 3.1 2013-10, Medical Electrical Equipment Part 1-6: 3. General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usability
- IEC 60601-1-11 Edition2.0 2015-01, Medical electrical equipment Part 1-11: 4. General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment Part 2-10: 5. Particular Requirements For The Basic Safety And Essential Performance Of Nerve And MuscleStimulators
- IEC 62133-2 Edition 1.0 2017-02, Secondary Cells And Batteries Containing 6. Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications - Part 2: LithiumSystems
- ANSI/AAMI/IEC 62304:2006/A1:2016. Medical Device Software Software 7. LifeCycle Processes [Including Amendment 1 (2016)]
Comparison to the Predicate Device and Conclusion:
The changes do not affect the intended use and do not alter the fundamental scientific technology of the predicate device: Compex Sport Elite (K170918). The subject device Compex Sport Elite 3.0 technological characteristics, features, specifications, materials, mode of operation and intended use are substantially equivalent to the predicate device named above. The differences between the subject device and predicate device do not raise new issues of safety or effectiveness.
{6}------------------------------------------------
The subject device has undergone testing by a Nationally Recognized Test Laboratory, Nemko Shanghai Ltd. Shenzhen Branch, to assure safety and conformance to standards for Electromagnetic Compatibility. Further, the subject device conforms to existing design controls procedures per §820.30. Below is a comparison analysis between the subject and predicatedevices: Table 1 Comparison Analysis
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis(HAZ-PRO-130) |
|---|---|---|---|---|---|
| 1 | Electrodesconnection | Lead wires including femalecustom SNAP plugged on thecustom Compex female SNAPassembled in the electrode.Entire electronic circuit for four(4) Stimulation Channels andUser Interface is combined intosame casing, connected to theelectrodes with 6-pole cables. | Lead wires including femalecustom SNAP plugged on thecustom Compex female SNAPassembled in the electrode.Entire electronic circuit for four(4) Stimulation Channels andUser Interface is combined intosame casing, connected to theelectrodes with 6-pole cables. | Same | No additional risk. |
| 2 | Power Source | Rechargeable Ni-MHbattery 4.8V (4 cellsAA=R6);replaceable. | Rechargeable Li-ion.battery 3.7V (one cell); notreplaceable. | Different | Smaller size; reduced weight.User will not handle battery.The battery conforms to IEC62133-2 and UN DOT 38.3.No additional risk |
| 3 | Battery ChargerRegulation | Discrete components withMCU regulate the batterycharging | Independent charging IC toregulate battery charging withprotection monitor. | Different | Subject device utilizes anindependent charging IC toperform protection workduring charging ordischarging process, thisreduces MCU involvement inregulation and simplifies thedesign.No additional risk |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis(HAZ-PRO-130) |
| 4 | Transformer | 1. Input voltage: 4.8V2. Size: EFD153. N1: 2.7uH, N2: 350uH5. Output power: 10W | 1. Input voltage: 3.7V2. Size: EFD15 (same)3. N1: 2.6uH, N2:650uH5. Output power: 10W(same) | Similar | Same mechanical structureand output.Lower input voltage (3.7V) ofsubject device leads to theinput current and inductancevalue increase to achieve thesame output power.No additional risk. |
| 5 | Number ofOutput Modes | Two (NMES/TENS) | Two (NMES/TENS) | Same | No additional risk. |
| 6 | Power Adapter | DC jack male charging port,9V/400mA power adapter | Micro-USB male charging port,5V/2A USB power adapter or PC | Different | 1. Thesubject deviceprovidesthe Micro-USB charging portwhich allows to connect toUSB power adapter or PC(user convenience)2. USB charging is morereliable method whichreduces hazard ID1.2.15'battery overcharging'.3. No additional risk. |
| 7 | Number ofOutputChannels | Four | Four | Same | No additional risk. |
| 8 | LCD | 1. LCD type: MonoSTNLCD, 128 * 64pixel2. Data Bus: 8-Bits Busby MCU system bus3. Backlight: 2 serial 3parallel (total 6 LEDs) | 1. LCD type: ColorTFTLCD,320 * 240 pixel2. Data Bus: 16-Bits Busbyspecial LCD 8080port3. Backlight: 4 Parallel LEDs | Different | 1. The backlight are LEDs2. The drive and controlmethod all are PWMtechnology3. The method for maxcurrent limiting for LEDareall same by hardware4. TFT display has betterperformance on resolutionand viewing angle than |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis(HAZ-PRO-130) |
| 9 | LED Indicator | No LEDs for UserInterface | 1. LED: Five indicating LEDs,four are for channel buttonsand the other one is for centralconfirmation button.2. LED Drive & Power: 5V Powersupply with lightadjusting | Different | 1. Common LED2. The drive circuit limits themax current of LED lowerthan 20mA in hardware.Software PWM adjusts theLED brightness. No additionalrisk |
| 10 | SynchronousOr Alternating | Synchronous, but never 2channels activated at the sametime | Synchronous, but never 2channels activated at the sametime | Same | No additional risk |
| 11 | Method ofChannelIsolation | Each channel is the middle ofan H-Bridge. Except when it isactivated, each channel isalways in high impedancestate. | Each channel is the middle ofan H-Bridge. Except when it isactivated, each channel isalways in high impedancestate. | Same | No additional risk |
| 12 | RegulatedCurrent orRegulatedVoltage | Regulated current (all channels) | Regulated current (all channels) | Same | No additional risk |
| 13 | SoftwareMicroprocessorControl | Yes | Yes | Same | No additional risk |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis(HAZ-PRO-130) |
| 14 | AutomaticOverload Trip | Yes | Yes | Same | No additional risk |
| 15 | Automatic No-Load Trip | Yes | Yes | Same | No additional risk |
| 16 | Automatic ShutOff | "On/Off" switch | "On/Off" switch | Same | No additional risk |
| 17 | Patient OverrideControl | Yes, push on On/Off buttondirectly pauses the program | Yes, push on On/Off buttondirectly pauses the program | Same | No additional risk |
| 18 | Indicator Display- On/Off Status | Yes | Yes | Same | No additional risk |
| 19 | Indicator Display- Low BatteryDetection | Yes | Yes | Same | No additional risk |
| 20 | Indicator DisplayVoltage/Current Level | Yes, unit = [Energy]The Max/Min level showed ondisplay is 999/0 | Yes, unit = [Energy]The Max/Min level showed ondisplay is 999/0 | Same | No additional risk |
| 21 | Timer Range(minutes) | Yes, unit= [seconds],max = 55 [minutes] | Yes, unit= [seconds],max = 55 [minutes] | Same | No additional risk |
| 22 | HousingMaterials andConstruction | Casing: Plastic(ABS, with PMMA on thewindows)Buttons: Silicon rubberABS housing around thebattery cells Battery contacts:SK5 steel | Casing: Plastic(ABS, withPMMA on thewindows)Buttons: Plastic ABSSilicone sleeve: Silicone + Gasphase glue | Similar | 1. Silicone sleeve is designedto meet IP22.2. uses biocompatiblematerials used on otherdevices3. No additional risk |
| 23 | Dimensions | 1429936mm | Without silicone sleeve:136mm76mm21mmWith silicone sleeve:140mm80mm24.6mm | Different | Subject device is more compact.No additional risk |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis (HAZ-PRO-130) |
| 24 | Ingress Protection | IP20 | IP22 | Different | IP22 is required by IEC 60601-1-11. Reduced risk. |
| 25 | Type of Charging Port | DC Jack female charging port | USB type-A customized 5 pinfemale connector | Different | 1. The subject device providecustomized charging portwhich reduces theoccurrence of hazard ID1.1.11 'incorrect connectionterminals'2. No additional risk |
| 26 | BOSS feature | Yes | Yes | Same | No additional risk |
| 27 | Part number of MCU | Renesas: M30620SPGP, up to24Mhz, RAM 10K (ExternalROM 512K) VCC= 3V and 5V,LQFP100 | ST: STM32F103VCT, 32bit up to72Mhz, RAM 48K ROM 256K,VCC= 3V, LQFP100 | Different | The performance of ARMMCU of subject device isimproved. No additional risk. |
| 28 | Data library | Shared with MCU ROM | External 64M bit SPI Flashmemory | Different | The subject device stores amuch larger data library inexternal flash memory. Noadditional risk |
| 29 | РСВ | PCB layers: 4 layersSize: 75mm * 120mm * T:1.6mmSurface treatment: gold platingFlammability rating: UL 94V0 | PCB layers: 4 layersSize: 72mm * 125mm *T:1.6mmSurface treatment: gold platingFlammability rating: UL 94 V0 | Similar | The only difference is thePCB size. No additional risk. |
| 30 | Library Update Port | UART | USB | Different | The communication port isdifferent, but the function issame. All with CRCverification. No additional risk. |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk analysis |
| 31 | Waveform | Symmetrical Biphasic | Symmetrical Biphasic | Same | No additional risk |
| 32 | Shape | Rectangular | Rectangular | Same | No additional risk |
| 33 | Maximum OutputVoltage (±10%) | 60 V @ 500 Ω165 V @ 2 kΩ165 V @ 10 kΩ | 60 V @ 500 Ω165 V @ 2 kΩ165 V @ 10 kΩ | Same | No additional risk |
| 34 | Maximum OutputCurrent (±10%) | 120 mA @ 500 Ω82 mA @ 2 kΩ16 mA @ 10 kΩ | 120 mA @ 500 Ω82 mA @ 2 kΩ16 mA @ 10 kΩ | Same | No additional risk |
| 35 | Pulse Width | NMES:200 to 400 [µs] (microseconds)TENS:70 to 300[µs] (measured at 50%of positive pulse) | NMES:200 to 400 [µs](microseconds)TENS:70 to 300[µs] (measuredat 50% of positive pulse) | Same | No additional risk |
| 36 | Frequency | 1 to 120 Hz | 1 to 120 Hz | Same | No additional risk |
| 37 | Net Charge[µC/pulse] | 0 [µC] @ 500ΩExcitation pulse fullycompensated | 0 [µC] @ 500ΩExcitation pulse fullycompensated | Same | No additional risk |
| 38 | Maximum PhaseCharge [µC] | 48 [µC] @ 500Ω | 48 [µC] @ 500Ω | Same | No additional risk |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk analysis |
| 39 | MaximumCurrent Density[mA/cm2] | 1.49 [mA/cm2] @ 500Ω | 1.49 [mA/cm2] @ 500Ω | Same | No additional risk |
| 40 | Maximum PowerDensity[mW/cm2] | 27.6 [mW/cm2] @ 500Ω | 27.6 [mW/cm2] @ 500Ω | Same | No additional risk |
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
Table 2 Output Specifications Comparison Analysis
{12}------------------------------------------------
Table 3 Software Comparison Analysis
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis |
|---|---|---|---|---|---|
| 41 | Dual MCU designMaster MCU isused to controltreatmentprogram and userinterface.Protection MCUis used to monitorthe operation ofthe Master MCUand the treatmentoutput. | Master MCU: M30620SPGP, up to24Mhz, RAM 10K (External ROM512K) VCC= 3V and 5V, LQFP100Protection MCU:MSP430F1101 | Master MCU: STM32F103VCT,32bit, up to 72Mhz, RAM 48KROM 256K, VCC= 3V, LQFP100Protection MCU: MSP430F1101 | Differentmaster MCUSameprotectionMCU | Subject device'smaster MCU hasimprovedperformance forcontrol and userinterface. Sameprotection MCU isused.No additional risk. |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis |
| 42 | DevelopmentEnvironment | Boot: Renesas IDEMain firmware: Renesas IDEProtection system: IAR | Boot: IARMain firmware: KeilProtection system: IAR | Similar | 1. The functionalityofboot and mainfirmware for bothdevices are thesame.2. No additional risk |
| 43 | Software/FirmwareOrganization | Software is divided into Boot,Main Firmware. | Software is divided into Boot,Main Firmware, Libraries,Manufacturing and BrandingRecords, Live Records | Different | Subject device usesa more updatedsoftware organizationtool. Reduced risk |
| 44 | Keyboard-Buttons design | Power ON buttonInformation button4 channel selection buttons.4 up/down buttons fortreatment strength adjustment | Power On buttonFive button guide wheel forinformation, user set up, andprogram strength adjustment.4 channel selection buttons | Similar | Subject device offerseasier userinteraction. Noadditional risk. |
| 45 | UI Screens -Favorites screen | Does not exist | Allows user to set up favoritesscreen | Different | Subject device offersuser a quicker devicestart. No additionalrisk. |
| 46 | UI Screens -Program screen | Allows user to select varioususer settings for the device,such as language, soundvolume, display contrast, etc. | Allows user to select various usersettings for the device, such astime, date, language, powersaving mode, etc. | Similar | Subject device offersa better userexperience. Noadditional risk. |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis |
| 47 | UI Screens -Placementsselection screen | Does not exist, need referenceto printed user guide | Using graphic display to informuser the proper location of thetreatment electrode pads on ahuman body diagram. | Different | Subject device offersclearer usage guide.Reduced risk. |
| 48 | UI Screens -Level selectionscreen | 5 treatment levels are selectedby the up-down button andviewable on the display | 5 treatment levels are selected bythe guiding wheel and theviewable on the display | Same | No additional risk |
| 49 | UI Screens -Channelsselection screen | Insert the treatment lead wire tothe corresponding output port,then press the channel button toadd corresponding channels. | Insert the treatment lead wire tothe corresponding output port,then press the channel button toadd corresponding channels. | Same | No additional risk |
| 50 | UI Screens -Stimulationscreen | Text and graphic information | Text and graphic information | Similar | Subject device offersbetter treatmentprogress informationto user. No additionalrisk |
| 51 | UI Screens - Stimpause screenand Stimcompleted screen | Text and graphic information | Text and graphic information | Similar | Subject device offersbetter treatmentprogress informationto user. No additionalrisk |
| 52 | UI Screens -Setting screen | Offers sound, language, displaycontrast selection | Offers sound, language, powersaving setting selection. Chineseis added. | Similar | Subject device offersbetter userexperience. Noadditional risk |
| 53 | UI Screens -Charging screen | Uses an icon on the display toshow the device batterycharging progress. | Uses an icon on the display toshow the device battery chargingprogress. | Same | No additional risk |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis |
| 54 | UI Screens -Electrode faultscreen | Graphic and text information"display electrode faultsymbol on screen, a twinklingarrow points to the faultchannel" | Graphic and text information"display electrode fault symbol onscreen, a twinkling arrow pointsto the fault channel, LED of faultchannel flashes" | Similar | 1. Subject deviceoffers better userexperience.2. No additional risk |
| 55 | UI Screens -Lowbattery screen | Graphic and test informationThere is a battery icon toindicate the charge state (6bars mean the battery is fullycharged), if the battery iconcontains just two bars, thismeans the power is runninglow, need to recharge the unit | Graphic and test informationThere is a battery icon to indicatethe charge state (4 bars mean thebattery is fully charged).4 bars indicate full battery (4.0-4.2V)3 bars indicate almost full (3.8-4.0V)2 bars indicate half-empty (3.6-3.8V)1 white bar indicates almost empty(3.5-3.6V)1 red bar indicates empty(<3.5V) | Similar | 1. Subject deviceoffers better userexperience.2. No additional risk |
| 56 | Firmware &Library Update | Firmware and library can beupdated from UART port inthe factory or service center | Firmware and library can beupdated from customized USBport in the factory or servicecenter. Unit charging or upgradesuse same USB port, which isconvenient in production andupgrades is not available for user | Different | 1.Subject deviceprovides the USBdata transmitconnector which iskeyed and can onlybe connected in oneorientation andreduces theoccurrence HazardID 1.1.11 'incorrectconnection terminals'2.User cannotupdatethe firmware andlibrary via USB portwithout permission |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis |
| from manufacture,there is no impact onits safety andeffectiveness. Noadditional risk | |||||
| 57 | External FlashLibrary DataStructure | Does not exist | Uses external flash to storelibraries to graphics, treatmentprogram and multi languageinterface. | Different | Subject device offersbetter userexperience.No additional risk |
| 58 | CRC verificationfor master MCUcode | CRC algorithm is used forsoftware integrity control. | CRC algorithm is used forsoftware integrity control. | Same | No additional risk |
| 59 | CommunicationProtocol | UART communicationprotocol that is developed forthe production software anddevice control | USB communication protocol thatis developed for the productionsoftware and device control | Different | 1. Bothcommunicationprotocols are used forproduction purposeand do not impactsafety andeffectiveness on endusers2. Subject deviceuses USBcommunicationprotocol which isconvenient to servicepersonnel (reduceshazard ID 7.4.1 'bador inadequatemaintenance')3. No additional risk |
| No | Characteristic | Predicate Device | Subject Device | Same/Similar/Different | Risk Analysis |
| 60 | Programs | Muscle Building1-Endurance2-Resistance3-Strength4-Explosive strengthWarm Up & Recovery5-Prewarm Up6-Training recovery7-Competition recovery8-Muscle relaxation9-PotentiationPain Management10-Pain Relief TENS | Muscle Building1-Endurance2-Resistance3-Strength4-Explosive strengthWarm Up & Recovery5-Prewarm Up6-Training recovery7-Competition recovery8-Muscle relaxation9-PotentiationPain Management10-Pain Relief TENS | Same | No additional risk |
| 61 | MSP430 ProtectSystemDevelopmentEnvironment | IAR | IAR | Same | No additional risk |
| 62 | ProtectionSystem ReportError | 12 error codes | 13 error codes.The first 12 error codes are thesame as predicate device.The No. 13th error code is toensure both ST32 (main MCU)and MSP430 (protection systemMCU) are working normally at thesame time. | Similar | 1. The new 13therrorcode is to checkMCU workingproperly2. No additional risk |
| 63 | CRC Verificationof ProtectionSystem Code | CRC algorithm is used for softwareintegrity control. | CRC algorithm is used forsoftware integrity control. | Same | No additional risk |
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
{17}------------------------------------------------
N/A