K Number
K201635

Validate with FDA (Live)

Date Cleared
2021-02-06

(235 days)

Product Code
Regulation Number
870.2300
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MouthLab Vital Signs Monitoring System is a hand-held, wireless device intended to record, transfer, store and display single lead electrocardiography (ECG), heart rate variability (HRV), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), respiration rate (RR), respiration pattern (RP), oral temperature (TEMP) and basic lung function measurements: peak expiratory flow (PEF) and forced expiratory volume in one second (FEV1).

This system is for spot checking and does not have continuous monitoring capability or any alarm features. The device comes in contact with the user for approximately 60 seconds at each use and the captured medical parameters shall be displayed on a mobile application.

It is intended for use by adults in the home environment and in healthcare facilities. This system makes no specific diagnosis. Respiration rate is not intended for adults with underlying or suspected medical conditions. The device is for single user use.

It is intended for use with users who are well perfused and during no motion condition. Users with implanted pacemakers and/or implanted cardioverter-defibrillators (ICDs) are not recommended to use the device.

Device Description

The MouthLab Vital Signs Monitoring System is a non-invasive, battery-powered, handheld, wireless, personalized single-user, vital sign measuring apparatus. The device is intended to intermittently monitor the user vitals through the use of an integrated SpO2 sensor, ECG electrodes, microphones and thermistor, which measures the users' functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), oral temperature (TEMP), ECG, heart rate (HR) and heart rate variability (HRV). The users' respiration rate (RR) measurements, respiration pattern (RP) and the basic lung functions (FEV1 and PEF) measurements are obtained are obtained through the acoustic measurement of air turbulence. The vital signs data are transferred to the Aidar Cloud™ Application via a cellular network for processing and storage. This processed data is then displayed on the Aidar Care™ Mobile Application.

AI/ML Overview

The provided document, a 510(k) Summary for the MouthLab Vital Signs Monitoring System (K201635), outlines the device's acceptance criteria and the studies conducted to demonstrate its performance. The device measures multiple physiological parameters, and therefore, different studies were conducted for each.

1. Table of Acceptance Criteria and Reported Device Performance:

ParameterAcceptance Criteria (from predicate/standard)Reported Device Performance
Heart Rate (HR) Accuracy± 2 bpm or ± 2%, whichever is larger (IEC 60601-2-27)± 2 bpm or ± 2%, whichever is larger (Same as criteria)
SpO2 Accuracy± 3% (ISO 80601-2-61)± 3% (Same as criteria)
Pulse Rate (PR) Accuracy± 2 ppm or ± 2%, whichever is larger (ISO 80601-2-61)± 2 ppm or ± 2%, whichever is larger (Same as criteria)
Respiration Rate (RR) Accuracy± 2 breaths/min (from Connex Monitor K171621)± 2 breaths/min (Same as criteria)
Temperature Accuracy± 0.6 °F (± 0.3°C) (from ISO 80601-2-56)± 0.6 °F (± 0.3°C) (Conforms to ISO standard requirements)
Peak Expiratory Flow (PEF) Accuracy± 0.33 L/s or 10%, whichever is larger (ATS)± 0.33 L/s or 10%, whichever is larger (Same as criteria)
Forced Expiratory Volume (FEV1) Accuracy± 0.1 L or 5%, whichever is larger (ATS)± 0.1 L or 5%, whichever is larger (Same as criteria)
ECG Input ImpedanceConforms to IEC 60601-2-27 impedance requirements> 2.5MΩ (Conforms to IEC 60601-2-27 impedance requirements)

2. Sample Sizes Used for the Test Set and Data Provenance:

  • SpO2 Measurements: "a controlled desaturation study with adults." The specific sample size is not provided. The data provenance is implied to be prospective clinical testing, likely conducted in a controlled environment as it refers to a "controlled desaturation study." The country of origin is not specified.
  • Respiration Rate Measurements: "20 healthy adults." The data provenance is prospective clinical testing. The country of origin is not specified.
  • For other parameters (ECG/HR, Pulse Rate, Temperature, PEF/FEV1), bench testing was performed, meaning synthetic data or physical test equipment was used. No human subject sample size is applicable.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Respiration Rate Measurements: Ground truth was established by "a clinician's manually scored capnography (EtCO2) waveforms." The number of clinicians is not specified, and their specific qualifications beyond "clinician" are not detailed.
  • SpO2 Measurements: Ground truth was established by "reference co-oximeter analysis of arterial blood gas." This implies laboratory analysis rather than human expert interpretation of raw data. No human experts were explicitly stated to establish this ground truth in the provided text.

4. Adjudication Method for the Test Set:

  • For SpO2, and Temperature, the comparison was against a reference standard (co-oximeter, calibrated thermometer) implying direct comparison, not a human adjudication process.
  • For Respiration Rate, the comparison was against "a clinician's manually scored capnography waveforms." This suggests a single clinician's scoring was used as ground truth, implying no adjudication method (e.g., 2+1 or 3+1) was employed for establishing this ground truth.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC comparative effectiveness study was done involving human readers with and without AI assistance (the MouthLab system). The studies described are validation studies comparing the device's measurements to reference standards, not studies on human reader performance improvement.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Yes, the performance data provided describes the standalone performance of the MouthLab Vital Signs Monitoring System in measuring various physiological parameters. The clinical and bench tests evaluate the device's ability to accurately capture and report these values.

7. The type of ground truth used:

  • ECG and Heart Rate: Bench testing compliant with IEC 60601-2-27:2012. (Simulated physiological signals)
  • Pulse Rate: Bench testing compliant with ISO 80601-2-61:2017. (Simulated physiological signals)
  • Temperature: Bench agreement study as per ISO 80601-2-56:2017. (Reference temperature measurements)
  • Peak Flow Measurements (FEV1 and PEF): Bench testing according to the ATS Standardization of Spirometry, 2005 Update. (Reference flow/volume measurements)
  • SpO2 Measurements: Reference co-oximeter analysis of arterial blood gas. (Objective clinical reference standard)
  • Respiration Rate Measurements: Clinician's manually scored capnography (EtCO2) waveforms. (Expert interpretation of surrogate physiological data)

8. The sample size for the training set:

  • The document does not explicitly mention a training set sample size. The performance data section focuses on "bench testing" and "clinical testing" for validation, not on the training of machine learning models. If the device uses algorithms trained on data, that information is not detailed in this summary.

9. How the ground truth for the training set was established:

  • As the document does not mention a training set, the method for establishing its ground truth is not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 6, 2021

MultiSensor Diagnostics (dba Aidar Health) % Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave, Suite B-257 San Diego, California 92110

Re: K201635

Trade/Device Name: MouthLab Vital Signs Monitoring System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DPS, FLL, BZQ, BZH, DPS Dated: December 23, 2020 Received: December 28, 2020

Dear Allison Komiyama:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201635

Device Name MouthLab Vital Signs Monitoring System

Indications for Use (Describe)

The MouthLab Vital Signs Monitoring System is a hand-held, wireless device intended to record, transfer, store and display single lead electrocardiography (ECG), heart rate variability (HRV), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), respiration rate (RR), respiration pattern (RP), oral temperature (TEMP) and basic lung function measurements: peak expiratory flow (PEF) and forced expiratory volume in one second (FEV ).

This system is for spot checking and does not have continuous monitoring capability or any alarm features. The device comes in contact with the user for approximately 60 seconds at each use and the captured medical parameters shall be displayed on a mobile application.

It is intended for use by adults in the home environment and in healthcare facilities. This system makes no specific diagnosis. Respiration rate is not intended for adults with underlying or suspected medical conditions. The device is for single user use.

It is intended for use with users who are well perfused and during no motion condition. Users with implanted pacemakers and/or implanted cardioverter-defibrillators (ICDs) are not recommended to use the device.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

Device Common Name:Monitor, Physiological, Patient (Without Arrhythmia Detection or Alarms)
Device Trade Name:MouthLab™ Vital Signs Monitoring System
ApplicantMultisensor Diagnostics, LLC. (DBA Aidar Health)3402 Birch Hollow Rd,Pikesville, MD 21208Phone: (443) 875-6456Website: www.aidar.com
Contact:Sathya ElumalaiFounder & Chief Executive Officerselumalai@aidar.com
Representative/Consultant:Allison C. Komiyama, Ph.D., R.A.C.AcKnowledge Regulatory Strategies, LLCTelephone: +1(619) 458-9547Email: akomiyama@acknowledge-rs.comWebsite: www.AcKnowledge-RS.com
Date Prepared:February 4, 2021
Classification Regulation:Class II
Classification Name:870.2300: Cardiac monitor (including cardiotachometer and rate alarm)
Panel:Cardiovascular
Primary Product Code:MWI
Secondary Product Codes:DQA, DPS, FLL, BZQ, DXN, BZH
Predicate Devices:Primary:CheckMe Pro Health Monitor (K150869)Secondary: Connex Vital Signs 6000 Monitor (K171621)Wing Smart FEV1 and Peak Flow Meter (K152276)
Reference Devices:• Masimo Acoustic Respiration Sensor (K120984)• ReadMyHeart - Model RMHI3.0 (K050620)• Peak me (K180487)

{4}------------------------------------------------

INDICATIONS FOR USE: 1.

The MouthLab Vital Signs Monitoring System is a hand-held, wireless device intended to record, transfer, store and display single lead electrocardiography (ECG), heart rate (HR), heart rate variability (HRV), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), respiration rate (RR), respiration pattern (RP), oral temperature (TEMP) and basic lung function measurements: peak expiratory flow (PEF) and forced expiratory volume in one second (FEV1).

This system is for spot checking and does not have continuous monitoring capability or any alarm features. The device comes in contact with the user for approximately 60 seconds at each use and the captured medical parameters shall be displayed on a mobile application.

It is intended for use by adults in the home environment and in healthcare facilities. This system makes no specific diagnosis. Respiration rate is not intended for adults with underlying or suspected medical conditions. The device is for single user use.

It is intended for use with users who are well perfused and during no motion condition. Users with implanted pacemakers and/or implanted cardioverter-defibrillators (ICDs) are not recommended to use the device.

DEVICE DESCRIPTION: 2.

The MouthLab Vital Signs Monitoring System is a non-invasive, battery-powered, handheld, wireless, personalized single-user, vital sign measuring apparatus. The device is intended to intermittently monitor the user vitals through the use of an integrated SpO2 sensor, ECG electrodes, microphones and thermistor, which measures the users' functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), oral temperature (TEMP), ECG, heart rate (HR) and heart rate variability (HRV). The users' respiration rate (RR) measurements, respiration pattern (RP) and the basic lung functions (FEV1 and PEF) measurements are obtained are obtained through the acoustic measurement of air turbulence. The vital signs data are transferred to the Aidar Cloud™ Application via a cellular network for processing and storage. This processed data is then displayed on the Aidar Care™ Mobile Application.

3. BIOCOMPATIBILITY:

The MouthLab Vital Signs Monitoring System is a surface device that includes components that have prolonged contact duration (> 24h and less than 30 days) with the

{5}------------------------------------------------

user's mucosal membrane and intact skin. Testing was performed in accordance with ISO 10993-1.

4. PERFORMANCE DATA:

4.1. Bench Testing:

The following bench testing was provided to support the substantial equivalence of the MouthLab Vital Signs Monitoring System:

  • . ECG and Heart Rate Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to capture ECG waveforms and heart rate measurements, bench testing as per IEC 60601-2-27:2012 was conducted.
  • Pulse Rate Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to capture pulse rate measurements, bench agreement testing as per ISO 80601-2-61:2017 was conducted.
  • Temperature Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to calculate temperature measurements, a bench agreement study was conducted, as per ISO 80601-2-56:2017.
  • Peak Flow Measurements: To validate the ability of the MouthLab Vital Signs ● Monitoring System to measure the basic lung functions (FEV1 and PEF), bench testing according to the ATS Standardization of Spirometry, 2005 Update was conducted.

4.2. Clinical Testing:

The following clinical testing was provided to support the substantial equivalence of the MouthLab Vital Signs Monitoring System:

  • . SpO2 Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to make pulse oximetry measurements, a clinical agreement study as per ISO 80601-2-61:2017 was conducted. MouthLab SpO2 measurements were validated in a controlled desaturation study with adults and compared to reference co-oximeter analysis of arterial blood gas.
    Respiration Rate Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to calculate respiration rate, a clinical agreement study was conducted with 20 healthy adults and MouthLab respiration rate measurements were compared against a clinician's manually scored capnography (EtCO2) waveforms.

{6}------------------------------------------------

5. SOFTWARE DOCUMENTATION:

Software documentation for a Moderate Level of Concern device was provided in support of the MouthLab Vital Signs Monitoring System as recommended by FDA's Guidance, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005 and IEC 62304:2006 + A1:2015.

6. ELECTRICAL SAFETY TESTING:

The MouthLab Vital Signs Monitoring System was found to be compliant with the following standards:

  • IEC 60601-1:2005 + A1:2012 .
  • IEC 60601-1-6:2010 + A1:2013 o
  • IEC 60601-1-11:2015 ●

7. ELECTROMAGNETIC COMPATIBILITY TESTING:

The MouthLab Vital Signs Monitoring System was found to be compliant with IEC 60601-1-2:2014 standard.

HUMAN FACTORS/USABILITY TESTING: 8.

Human factors/usability testing was conducted to evaluate the ability of lay people to read and understand the MouthLab Vital Signs Monitoring System instructions for use, and subsequently simulate normal use of the MouthLab Vital Signs Monitoring System with limited training. The MouthLab Vital Signs Monitoring System was also tested as per IEC 60601-1-11:2015, for suitability of the system in a home environment.

SUBSTANTIAL EQUIVALENCE DISCUSSION: 9.

The MouthLab Vital Signs Monitoring System is substantially equivalent to the predicate devices based on the information summarized here.

The subject device has the same intended use and similar technological characteristics as the predicate devices cleared in K150869, K171621, and K152276. The primary predicate, CheckMe Pro Health Monitor (K150869) is also a handheld device for monitoring physiological parameters in the home or healthcare environment. The subject device has seven primary medical device functions and each of them have been compared to predicate devices. There are no differences in the technological characteristics for SpO2, Pulse Rate, Heart Rate, ECG, Temperature, Respiration Rate, Respiration Pattern, PEF and FEV1. The subject device also measures the heart rate variability (HRV) in R-R interval, which is a mathematical analysis of the variation in the

{7}------------------------------------------------

heart rate and is not intended to produce any interpretation of those measurements or be used for any kind of diagnosis.

The similar indications for use, technological characteristics, and performance characteristics for the proposed MouthLab Vital Signs Monitoring System are assessed to be substantially equivalent to the predicate devices. A comparison of the technological characteristics between the subject device and the predicate devices is provided in the tables below.

Table 1: Technological

Subject DevicePrimary PredicateSubstantialEquivalence toPredicate Device
510(k) NumberK201635K150869NA
ApplicantMultisensorDiagnostics, LLCViatom Technology Co.Ltd.NA
Device NameMouthLab Vital SignsMonitoring SystemCheckMe Pro HealthMonitorNA
ClassificationRegulation870.2300 -Physiological PatientMonitor870.2300 -Physiological PatientMonitorSame
Device ClassClass IIClass IISame
PhysiologicalParametersMonitoredECG, HR, HRV, SpO2,PR, RR, RP,TEMP, PEF, FEV1ECG, HR, SpO2, PR,TEMPSimilar
LocationHome and HealthcarefacilitiesHome and HospitalSame
Rx or OTCRxRxSame
Power Supply2000mAh rechargeablelithium-polymer battery560mAh rechargeablelithium-ion polymerbatterySimilar
UserPopulationAdultsAdults and PediatricSubject device onlysupports adultusers
AlarmNo AlarmsNo AlarmsSame
DataCollectionMinimum of 10,000measurements storage100 measurementsstorageSimilar
Memory
Subject DevicePrimary PredicateSubstantialEquivalence toPredicate Device
OperatingModesSpot-CheckContinuous Recordingand Spot-CheckSubject device onlysupports spot-check
Comparison of ECG and Heart Rate Measurements
MeasuredECGParametersECG Waveform, HeartRate (HR), Heart RateVariability (HRV)ECG Waveform, HeartRate (HR)Similar
ECG RhythmClassificationNot IncludedNot IncludedSame
ECG LeadTypeSingle Lead, 2 ContactsExternalECG CableandElectrodesSimilar
InputImpedance> 2.5MΩ> 10MΩConforms to IEC60601-2-27impedancerequirements
Input DynamicRange± 2mV± 3mVSubject devicesupports this range
Bandwidth0.5 - 50 Hz0.05 - 40HzSimilar
A/DConversion24 bit16 bitSimilar
Sampling Rate1000 Hz500 HzSimilar
MeasurementTime≤ 60 seconds30 secondsSimilar
DisplayMobile Application400*240 Dot-matrixLCD DisplaySubject device onlysupports mobileapplication.
InputDry ConductiveElectrodesDry ConductiveElectrodes and/orExternal AuxiliaryElectrodesSimilar
Heart Rate(HR) Range30 - 200 bpm30 - 250 bpmConforms to therequirements of theIEC 60601-2-27heart ratemeasurements
Subject DevicePrimary PredicateSubstantialEquivalence toPredicate Device
Heart Rate(HR) Accuracy$\pm$ 2 bpm or $\pm$ 2%,whichever is larger$\pm$ 2 bpm or $\pm$ 2%,whichever is largerSame
Comparison of Blood Oxygen Saturation (SpO2) Measurements
Display DataModeSpO2, Pulse Rate (PR)Spot-CheckSpO2, Pulse Rate (PR)Continuous Recordingand Spot-CheckSameSubject device onlysupports spot-check
Sensor TypesIntegratedIntegrated and ExternalSubject device onlysupports integratedsensors
SpO2 Range70 – 100%70 – 100%Same
SpO2Accuracy$\pm$ 3%Integrated: $\pm$ 3%Same
SpO2Resolution1%1%Same
Pulse Rate(PR) Range30 – 250 ppm30 – 250 ppmSame
Pulse Rate(PR) Accuracy$\pm$ 2 ppm or $\pm$ 2%,whichever is larger$\pm$ 2 ppm or $\pm$ 2%,whichever is largerSame
Pulse Rate(PR)Resolution1 ppm1 ppmSame

Comparison – MouthLab System vs CheckMe Pro

{8}------------------------------------------------

{9}------------------------------------------------

Table 2: Technological Comparison – MouthLab System vs Connex Monitor

Comparison of Respiration Rate Measurement
Subject DeviceSecondaryPredicateSubstantialEquivalence toPredicate Device
510(k) NumberK201635K171621NA
ApplicantMultisensorDiagnostics, LLCWelch Allyn, IncNA
Device NameMouthLab VitalSigns MonitoringSystemConnex Vital SignsMonitorNA
ClassificationRegulation870.2300 -870.2300 -Same

{10}------------------------------------------------

Subject DeviceSecondaryPredicateSubstantialEquivalence toPredicate Device
PhysiologicalPatientMonitorPhysiologicalPatient Monitor
Device ClassClass IIClass IISame
LocationHome andHealthcareFacilitiesHospitalSimilar
MeasurementTechniquesAcousticMeasurementsAcousticMeasurementsSame
Display Range6 – 50 breaths/min4 – 70 breaths/minSubject device supportsthis range
MeasurementRange6 – 50 breaths/min4 – 70 breaths/minSubject device supportsthis range
Respiration Rate(RR) Accuracy$\pm$ 2 breaths/min$\pm$ 1 breath/minSubject device supportsthis accuracy
Comparison of Temperature Measurement
MeasurementTechniqueThermo-resistiveSensorThermo-resistiveSensorSame
Measurement SiteOralOral, Axillary,RectalSubject device onlysupports oralmeasurements
ModeDirect ModeDirect or PredictiveModeSubject device onlysupports direct mode
Unit°F or °C°F or °CSame
Measuring Range93.2°F – 109.4°F(34°C – 43.0°C)80°F – 110°F(26.7°C – 43.3°C)Conforms to the ISOstandard requirements
Accuracy$\pm$ 0.6 °F( $\pm$ 0.3°C)$\pm$ 0.2°F( $\pm$ 0.1°C)Conforms to the ISOstandard requirements

Table 3: Technological Comparison – MouthLab System vs Wing

Comparison of Lung Function Measurements
510(k) NumberSubject DeviceSecondaryPredicateSubstantialEquivalence
K201635K152276NA
Subject DeviceSecondaryPredicateSubstantialEquivalence
ApplicantMultisensorDiagnostics, LLCSparo LabsNA
Device NameMouthLab VitalSigns MonitoringSystemWing Smart FEV1and Peak FlowMeterNA
ClassificationRegulation870.1860 – PeakFlow Meter forSpirometry870.1860 – PeakFlow Meter forSpirometrySame
Device ClassClass IIClass IISame
PhysiologicalParametersMonitoredFEV1, PEFFEV1, PEFSame
LocationHome and HospitalHomeSimilar
User InterfaceMobile ApplicationMobile ApplicationSame
Rx or OTCRxOTCSubject device is onlyapplicable forprescription use basedon other parameters itmeasures
Use PopulationAdultsAdults and pediatricusers above 5years of ageSubject device is foradult users
Operating ModesSpot-CheckSpot-CheckSame
MeasurementTechniqueAcousticMeasurement usingMicrophonesAcousticMeasurement usingMicrophonesSame
Peak ExpiratoryFlow (PEF) Range2 – 15 L/s50 – 900 L/minSubject devicesupports this range
Peak ExpiratoryFlow (PEF)Accuracy± 0.33 L/s or 10%,whichever is larger±20 L/min or 10%,whichever is largerSame
PEF Resolution0.01 L/s1 L/minSame
Forced ExpiratoryVolume (FEV1)Range1.00 – 7.50 L0.01 – 9.99 LSubject devicesupports this range
Forced ExpiratoryVolume (FEV1)Accuracy± 0.1 L or 5%,whichever is larger± 0.1 L or 5%,whichever is largerSame
Subject DeviceSecondaryPredicateSubstantialEquivalence
FEV1 Resolution0.01 L0.01 LSame

{11}------------------------------------------------

{12}------------------------------------------------

10. CONCLUSION:

Based on the information provided in this 510(k) premarket notification, the MouthLab Vital Signs Monitoring System is considered to be substantially equivalent (safe and effective) to its predicate devices: CheckMe Pro Health Monitor (K150869), the Welch Allyn Monitor (K171621), and Wing Smart FEV1 and Peak Flow Meter (K152276). The MouthLab Vital Signs Monitoring System has the same intended uses, similar indications, technological characteristics and principles of operation as its predicate devices. Through the performance and clinical testing conducted as mentioned herein, the minor technological differences between MouthLab Vital Signs Monitoring System and its predicate devices raise no new issues of safety and effectiveness. Hence, the MouthLab Vital Signs Monitoring System is substantially equivalent to its predicate devices.

§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).

(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).