K Number
K201628

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-02-12

(241 days)

Product Code
Regulation Number
880.5130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate (measured by a SpO2 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Device Description

The Panda iRes Warmer and Giraffe Warmer are infant radiant warmers with a heating source intended to maintain the thermal balance of an infant patient by direct radiation of energy in the infrared region of the electromagnetic spectrum. This device also provides optional integrated SpO2 Monitoring and Resuscitation.

Radiant heat from an infrared heat source is focused onto the bed to warm the patient. The operator may select either the heater power or skin temperature control method. Depending on the control method selected, the heater is either regulated at the operator selected power level or the heater output is modulated to maintain the patient's temperature at the value selected by the operator.

The subject devices have two control modes, Manual Mode and Baby Mode. In the manual Mode, the warmer controls radiant heater output from a heater power percentage setting that you enter using the control panel. In baby mode, the warmer controls radiant heater output based on temperature readings from a probe attached to the baby's skin (skin temperature probe and the reflective probe patch) and a set temperature (set temp) you enter using the control panel.

AI/ML Overview

This document describes the 510(k) summary for the Panda iRes Warmer and Giraffe Warmer, which are infant radiant warmers. The submission is for modifications to strengthen the bedside panels and latches.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state acceptance criteria for each test in a numerical format that can be directly compared to reported device performance. Instead, it indicates that testing was "completed with passing results per the acceptance criteria defined in the test cases." The specific quantitative acceptance criteria are not provided in this summary.

Test NameReported Device Performance (as stated in document)
South Door Pull Test (includes Wall 20N IEC Push Test)Test (completed with passing results)
South Wall 660N Static Load VerificationTest (completed with passing results)
Impact Test - Door shock frame test per IEC 60601-1 – 15.3.5.c.Test (completed with passing results)
Wall Threshold Test – Modified threshold test per IEC 60601-1 –9.4.2.4.3 with operator pulling on wall instead of pushing on handleTest (completed with passing results)
East and West side panels are interchangeableAnalysis (completed with passing results)
Wall Ergonomics – Caregiver ProximityAnalysis (completed with passing results)
Panel Corner SpacingAnalysis (completed with passing results)
Drawer AccessAnalysis (completed with passing results)
Side Wall RemovalAnalysis (completed with passing results)
Walls are Cleanable3rd Party Test (completed with passing results)
Reliability - Wall Latch / Bed Pin Wear TestTest (completed with passing results)
Reliability - Wall Pull Fatigue TestTest (completed with passing results)
Reliability - Wall Lean Push Fatigue – Wall in Up PositionAnalysis (completed with passing results)
Reliability - Wall Knee Push Fatigue – Wall in Down PositionTest (completed with passing results)

2. Sample Size Used for the Test Set and the Data Provenance

The document does not specify the sample sizes used for each test. The tests are described as "controlled engineering test fixtures" and "mathematical and statistical analyses," suggesting these were bench tests rather than tests on a specific patient or clinical dataset.

The data provenance is not explicitly stated in terms of country of origin or retrospective/prospective; however, the testing appears to be non-clinical performance and reliability testing conducted internally by GE Healthcare as part of their quality system.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not applicable. The testing described is non-clinical engineering and reliability testing of mechanical components. "Ground truth" in the context of clinical or diagnostic performance is not relevant here. The ground truth for these tests would be the established engineering specifications and regulatory standards (e.g., IEC 60601-1).

4. Adjudication Method for the Test Set

This information is not applicable as the tests are non-clinical engineering and reliability tests, not assessments that require human adjudication of a "ground truth" based on expert consensus.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The document states, "The Panda iRes Warmer and Giraffe Warmer did not require clinical tests to support substantial equivalence." This submission is for modifications to a mechanical component (bedside panel and latches) of an existing device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This question is not applicable. The device is a physical medical device (infant radiant warmer) and not an AI algorithm. The testing focuses on the mechanical integrity and safety of the bedside panels and latches.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical testing appears to be based on:

  • Engineering Specifications: Internal design requirements for the strength, durability, and functionality of the bedside panels and latches.
  • Regulatory Standards: Compliance with relevant international standards such as AAMI / ANSI ES60601-1, AAMI / ANSI / IEC 60601-1-2, and IEC 60601-2-21, which define performance and safety criteria for medical electrical equipment and infant radiant warmers.

8. The Sample Size for the Training Set

This information is not applicable. There is no "training set" as this is a modification to a physical medical device and not an AI/ML algorithm development.

9. How the Ground Truth for the Training Set was Established

This information is not applicable for the reasons mentioned in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line.

February 12, 2021

GE Healthcare Michelle Johnson Sr. Regulatory Affairs Leader 9900 Innovation Drive Wauwatosa, Wisconsin 53226

Re: K201628

Trade/Device Name: Panda iRes Warmer, Giraffe Warmer Regulation Number: 21 CFR 880.5130 Regulation Name: Infant Radiant Warmer Regulatory Class: Class II Product Code: FMT Dated: December 21, 2020 Received: December 22, 2020

Dear Michelle Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K201628

Device Name Panda iRes Warmer , Giraffe Warmer

Indications for Use (Describe)

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous nonitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate (measured by a SpO2 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K201628 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:
Date:February 12, 2021
Primary ContactPerson:Michelle JohnsonSr. Regulatory Affairs LeaderDatex-Ohmeda, Inc.9900 Innovation DriveWauwatosa, WI 53226Phone: 414-429-9263Email: Michelle.Johnson@ge.com
Secondary ContactPerson:Lee BushRegulatory Affairs DirectorDatex-Ohmeda, Inc.Phone: 262-309-9429Email: Lee.Bush@ge.com
Device Trade Name:Panda iRes Warmer, Giraffe Warmer
Common/Usual Name:Infant Warmer
Classification Names:Warmer, Infant Radiant
Regulation21 CFR 880.5130 Infant radiant warmer.
ClassificationII
Product Code:FMT
Predicate Device:Panda Warmer, Giraffe Warmer (K122267)

{4}------------------------------------------------

Indications for Use:Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp02) and pulse rate (measured by a SpO2 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.
Comparison ofIntended UseThere is no change to the intended use of the Panda iRes Warmer or Giraffe Warmer.

5.1 Device Description:

The Panda iRes Warmer and Giraffe Warmer are infant radiant warmers with a heating source intended to maintain the thermal balance of an infant patient by direct radiation of energy in the infrared region of the electromagnetic spectrum. This device also provides optional integrated SpO2 Monitoring and Resuscitation.

Radiant heat from an infrared heat source is focused onto the bed to warm the patient. The operator may select either the heater power or skin temperature control method. Depending on the control method selected, the heater is either regulated at the operator selected power level or the heater output is modulated to maintain the patient's temperature at the value selected by the operator.

The subject devices have two control modes, Manual Mode and Baby Mode. In the manual Mode, the warmer controls radiant heater output from a heater power percentage setting that you enter using the control panel. In baby mode, the warmer controls radiant heater output based on temperature readings from a probe attached to the baby's skin (skin temperature probe and the reflective probe patch) and a set temperature (set temp) you enter using the control panel.

5.2 Comparison of Technological Characteristics with the Predicate Device

The modified Panda iRes Warmer and Giraffe Warmer are infant radiant warmers

{5}------------------------------------------------

developed on the same platform as predicate Panada iRes Warmer and Giraffe Warmer (K122267). The proposed modifications to the Panda iRes Warmer and Giraffe Warmer include changes to strengthen the bedside panels and latches (including changing the material of the bedside panel latch from plastic to metal), to better withstand forces in the event of misuse by maneuvering the warmer by pulling the bedside panel wall. The bedside panel design is modified to accommodate the changed latch. These modifications to the bedside panel are mechanical in nature.

5.3 Determination of Substantial Equivalence:

Summary of Non-Clinical Testing:

The following performance data was provided in support of the substantial equivalence determination:

Compliance with Voluntary Standards

The Panda iRes Warmer and Giraffe Warmer is designed and tested for compliance with the following performance standards:

    1. AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012. C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
    1. AAMI / ANSI / IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance --Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests
    1. IEC 60601-2-21 Edition 2.1 2016-04, CONSOLIDATED VERSION Medical Electrical Equipment - Part 2-21: Particular Requirements For The Basic Safety And Essential Performance Of Infant Radiant Warmers [Including: Amendment 1 (2016)]

The Panda iRes and Giraffe Warmer was designed and is manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485 and the following quality assurance measures were applied to the development of subject features:

  • . Risk Analysis
  • . Requirements Reviews
  • Design Reviews .
  • Performance Testing (Verification .
  • . Safety/Reliability Testing (Verification)
  • . Summative Usability Testing (Validation)

{6}------------------------------------------------

Verification and Validation Testing

The proposed device has successfully completed all testing per our quality system.

Non-Clinical Testing

Evaluations of safety and effectiveness of the modified bedside panel used controlled engineering test fixtures to substantiate the performance of the new panel to worstcase conditions. Various mathematical and statistical analyses were performed to demonstrate that each performance item was successfully verified. The type of test and method of verification are listed in the table below.

Test NameVerificationMethod
South Door Pull Test (includes Wall 20N IEC Push Test)Test
South Wall 660N Static Load VerificationTest
Impact Test - Door shock frame test per IEC 60601-1 – 15.3.5.c.Test
Wall Threshold Test – Modified threshold test per IEC 60601-1 –9.4.2.4.3 with operator pulling on wall instead of pushing on handleTest
East and West side panels are interchangeableAnalysis
Wall Ergonomics – Caregiver ProximityAnalysis
Panel Corner SpacingAnalysis
Drawer AccessAnalysis
Side Wall RemovalAnalysis
Walls are Cleanable3rd Party Test
Reliability - Wall Latch / Bed Pin Wear TestTest
Reliability - Wall Pull Fatigue TestTest
Reliability - Wall Lean Push Fatigue – Wall in Up PositionAnalysis
Reliability - Wall Knee Push Fatigue – Wall in Down PositionTest

Animal Study

The Panda iRes Warmer and Giraffe Warmer did not require animal tests to support substantial equivalence.

Clinical Studies

The Panda iRes Warmer and Giraffe Warmer did not require clinical tests to support substantial equivalence.

Reprocessing

The device modifications have been validated for processing in accordance with recommended evaluations as listed in AAMI TIR12 and Guidance for Industry and FDA Staff -

{7}------------------------------------------------

Processing/Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling.

Human Factors Analysis

The modifications to the Panda iRes Warmer and Giraffe Warmer required summative usability for the bedside panel instructions for use.

Conclusion:

GE Healthcare considers the modified Panda iRes Warner and Giraffe Warmer to be as safe, as effective, and the performance to be substantially equivalent to the predicate device. Nonclinical tests have been summarized in the verification testing. The testing was completed with passing results per the acceptance criteria defined in the test cases. Based on development under GE Healthcare's quality system, the successful verification and engineering bench testing, GE Healthcare believes that the Panda iRes Warmer and Giraffe Warmer perform in a substantially equivalent manner to the predicate devices (K122267) and hence is safe and effective for its intended use.

§ 880.5130 Infant radiant warmer.

(a)
Identification. The infant radiant warmer is a device consisting of an infrared heating element intended to be placed over an infant to maintain the infant's body temperature by means of radiant heat. The device may also contain a temperature monitoring sensor, a heat output control mechanism, and an alarm system (infant temperature, manual mode if present, and failure alarms) to alert operators of a temperature condition over or under the set temperature, manual mode time limits, and device component failure, respectively. The device may be placed over a pediatric hospital bed or it may be built into the bed as a complete unit.(b)
Classification. Class II (Special Controls):(1) The Association for the Advancement of Medical Instrumentation (AAMI) Voluntary Standard for the Infant Radiant Warmer;
(2) A prescription statement in accordance with § 801.109 of this chapter (restricted to use by or upon the order of qualified practitioners as determined by the States); and
(3) Labeling for use only in health care facilities and only by persons with specific training and experience in the use of the device.